K Number
K150961

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-12-04

(238 days)

Product Code
Regulation Number
884.6160
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The EmbryoSlide culture dish is intended for preparing, storing, and transferring human embryos. It is intended to be used only with the EmbryoScope device.

Device Description

The Vitrolife A/S EmbryoSlide culture dish is intended for preparing, storing, and transferring human embryos. The Vitrolife A/S EmbryoSlide culture dish is intended to be used with the EmbryoScope device that was previously cleared via K092183, K111715 and K133712. A previous version of the EmbryoSlide culture dish was cleared via K092183. Minor modifications have been made to the original EmbryoSlide culture dish as described below.

The proposed EmbryoSlide culture dish contains two types of wells which are only used for rinsing and handling the embryos before or after incubation and wells in which the embryos reside during incubation. The original EmbryoSlide culture dish included only wells for which the embryos reside during incubation.

This EmbryoSlide culture dishfits exactly on the culture dish holder in the incubator and holds a maximum of 12 embryos in separate micro wells. As with the original EmbryoSlide culture dish, the proposed EmbryoSlide culture dish contains a large oil reservoir with 12 wells for the incubation of 12 individual embryos. Each well has a volume of 25 ul. Inside each well there is a central depression where the embryo rests. The well has a diameter of approximately 250 um.

As described above, the proposed EmbryoSlide culture dish includes 4 special wells, two at each end of the culture dish that are intended for rinsing and general handling of the embryos either before or after incubation. Like the other wells, these handling/rinsing wells are designed to hold a volume of 25ul.

AI/ML Overview

The provided document is a 510(k) summary for the Vitrolife A/S EmbryoSlide Culture Dish. It details the device, its intended use, and performance testing conducted to demonstrate substantial equivalence to a predicate device.

Here's an analysis of the acceptance criteria and study information provided:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
1-cell Mouse Embryo Assay (MEA)
Blastocyst rate of at least 80% at 96 hoursThe EmbryoSlide® culture dish is tested per LOT for performance by 1-cell Mouse Embryo Assay (MEA) testing, which requires a blastocyst rate of at least 80% at 96 hours for adequate performance. (Implies that the device meets this, as it is stated as a requirement for adequate performance)
In vitro cytotoxicity (USP <87>, ISO 10993-5)
Cytotoxicity grade ≤ 2The cytotoxicity grade was ≤ 2, meeting the requirements of ISO 10993-5 and USP <87>.
Bacterial endotoxins (USP <85>, <161>)
Less than 20 EU/deviceThe test solution contained less than 20 EU/device, meeting the maximum stated in USP <85> and <161>.

2. Sample size used for the test set and the data provenance

  • MEA Testing: "tested per LOT" is mentioned, indicating ongoing quality control testing rather than a specific sample size for a single clearance study. The document does not specify a numerical sample size for the MEA test set used for this 510(k) submission.
  • Cytotoxicity Testing: Not specified.
  • Bacterial Endotoxins Testing: Not specified, also "tested per LOT".
  • Data Provenance: Not explicitly stated but implied to be from laboratory testing conducted by the manufacturer, Vitrolife A/S. No country of origin for specific data or whether it was retrospective/prospective is detailed.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Not applicable/Not mentioned. The testing described (MEA, cytotoxicity, endotoxins) are lab-based performance tests with objective endpoints, not requiring expert ground truth or interpretation in the way clinical studies or image analysis algorithms would.

4. Adjudication method for the test set

  • Not applicable/Not mentioned. The tests are analytical/quantitative laboratory assays.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study was conducted or mentioned. This device is a culture dish, not an AI-powered diagnostic or assistive tool for human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable. This device is a physical labware product, not an algorithm.

7. The type of ground truth used

  • The "ground truth" for the performance tests mentioned are defined by established biological and chemical assay standards:
    • MEA: Successful development of embryos to the blastocyst stage.
    • Cytotoxicity: Absence of adverse effects on cultured mammalian cells per USP <87> and ISO 10993-5.
    • Bacterial Endotoxins: Concentration of endotoxins below a specified threshold per USP <85> and <161>.

8. The sample size for the training set

  • Not applicable. This is a medical device (culture dish), not a machine learning model, so there is no training set in the AI context.

9. How the ground truth for the training set was established

  • Not applicable. As above, there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus or a representation of the human form.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 04, 2015

Vitrolife A/S Belinda Dueholm Regulatory Affairs Specialist Jens Juuls Vej 20 DK-8260 Viby J Denmark

Re: K150961

Trade/Device Name: Embryoslide Culture Dish Regulation Number: 21 CFR 884.6160 Regulation Name: Assisted Reproduction Labware Regulatory Class: Class II Product Code: MOK Dated: October 2, 2015 Received: October 7, 2015

Dear Belinda Dueholm,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Benjamin R. Fisher -S

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K150961

Device Name

EmbryoSlide culture dish

Indications for Use (Describe)

The EmbryoSlide culture dish is intended for preparing, storing, and transferring human embryos. It is intended to be used only with the EmbryoScope device.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."


Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary for the Vitrolife A/S EmbryoSlide Culture Dish

(per 21CFR 807.92(c))

1. SUBMITTER/510(K) HOLDER

Vitrolife A/S Jens Juuls Vej 20 DK-8260 Viby J Denmark

Contact Person: Ms. Belinda Dueholm Telephone: +45 7221 7900 (main) +45 2076 3707 (direct)

Date Prepared: April 1, 2015

2. DEVICE NAME

EmbryoSlide culture dish Proprietary Name: Common/Usual Name: IVF Incubator and Accessories Classification Name: Assisted Reproduction Labware (21 CFR 884.6160)

3. PREDICATE AND REFERENCE DEVICES

  • K092183. Vitrolife A/S EmbryoSlide culture dish (Predicate device) ●
  • K070047, Nunc IVF 4-Well Dish (Reference device) ●

4. DEVICE DESCRIPTION

The Vitrolife A/S EmbryoSlide culture dish is intended for preparing, storing, and transferring human embryos. The Vitrolife A/S EmbryoSlide culture dish is intended to be used with the EmbryoScope device that was previously cleared via K092183, K111715 and K133712. A previous version of the EmbryoSlide culture dish was cleared via K092183. Minor modifications have been made to the original EmbryoSlide culture dish as described below.

The proposed EmbryoSlide culture dish contains two types of wells which are only used

{4}------------------------------------------------

for rinsing and handling the embryos before or after incubation and wells in which the embryos reside during incubation. The original EmbryoSlide culture dish included only wells for which the embryos reside during incubation.

This EmbryoSlide culture dishfits exactly on the culture dish holder in the incubator and holds a maximum of 12 embryos in separate micro wells. As with the original EmbryoSlide culture dish, the proposed EmbryoSlide culture dish contains a large oil reservoir with 12 wells for the incubation of 12 individual embryos. Each well has a volume of 25 ul. Inside each well there is a central depression where the embryo rests. The well has a diameter of approximately 250 um.

As described above, the proposed EmbryoSlide culture dish includes 4 special wells, two at each end of the culture dish that are intended for rinsing and general handling of the embryos either before or after incubation. Like the other wells, these handling/rinsing wells are designed to hold a volume of 25ul.

ട. INTENDED USE

The EmbryoSlide culture dish is intended for preparing, and transferring human embryos. It is intended to be used only with the EmbryoScope device.

6. TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE

The subject and predicate device have the following differences in technological characteristics: addition of rinsing wells to the culture dish, changes to the dimensions of the dish, the use of a different polystyrene formulation, and modifications to the device packaging. The differences in technological characteristics do not raise different questions of safety and effectiveness.

PERFORMANCE TESTING 7.

The EmbryoSlide® culture dish is tested per LOT for performance by 1-cell Mouse Embryo Assay (MEA) testing, which requires a blastocyst rate of at least 80% at 96 hours for adequate performance.

The EmbryoSlide® culture dish was tested for in vitro cytotoxicity in cultured mammalian cells. The test was performed in accordance with the United States Pharmacopeia (USP) <87> and the ISO 10993-5 guidelines. Based on the results, EmbryoSlide® culture dish passes the requirements of ISO 10993-5 and USP <87> as the cytotoxicity grade was ≤ 2.

The EmbryoSlide® culture dish is tested per LOT for bacterial endotoxins using the kinetic chromogenic method. The test of the undiluted extract (the test solution) was performed in

{5}------------------------------------------------

accordance with the US Pharmacopeia <85> and <161>. The test solution contained less than 20 EU/device, which is the maximum, as stated in the USP <85> and <161>.

The EmbryoSlide® culture dish devices are sterilized under controlled conditions via electron beam radiation. The minimum dose has been validated according to ISO 11737 to a sterility assurance level of 10-6.

Shelf life was determined through an accelerated aging study evaluating the package integrity and MEA.

Conclusion

The performance testing demonstrates that the subject device is substantially equivalent to the proposed predicate device.

§ 884.6160 Assisted reproduction labware.

(a)
Identification. Assisted reproduction labware consists of laboratory equipment or supplies intended to prepare, store, manipulate, or transfer human gametes or embryos for in vitro fertilization (IVF), gamete intrafallopian transfer (GIFT), or other assisted reproduction procedures. These include syringes, IVF tissue culture dishes, IVF tissue culture plates, pipette tips, dishes, plates, and other vessels that come into physical contact with gametes, embryos or tissue culture media.(b)
Classification. Class II (special controls) (mouse embryo assay information, endotoxin testing, sterilization validation, design specifications, labeling requirements, and clinical testing). The device, when it is a dish or plate intended for general assisted reproduction technology procedures, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.