K Number
K150942

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-11-25

(231 days)

Product Code
Regulation Number
862.1345
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Contour Next USB blood glucose monitoring system is an over the counter (OTC) device utilized by persons with diabetes in home settings for the quantitative measurement of glucose in whole blood, and is for single-patient use only and should not be shared.

The system consists of a Contour Next USB blood glucose meter, Contour Next test strips and Contour Next control solutions.

The Contour Next USB blood glucose monitoring system is indicated for use with fresh capillary whole blood samples drawn from the fingertip and palm only. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly).

The Contour Next USB blood glucose monitoring system is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

The Contour Next Test Strips are for use with the Contour Next USB blood glucose monitoring system for the quantitative measurement of glucose in whole blood.

Device Description

The Contour Next USB Blood Glucose Meter consists of a small handheld blood glucose meter that utilizes dry reagent test strips for the measurement of glucose in capillary whole blood by persons with diabetes. Liquid control solution is used to check the performance of the system. The meter, together with the test strips and control solutions, is referred to as the Contour Next USB Blood Glucose Monitoring System.

The chemical principle utilized for both the predicate and modified devices is based on measurement of electrical current caused by the reaction of glucose in the blood with chemicals on the reagent strip. The blood sample is drawn into the tip of the reagent strip through capillary action. Glucose in the sample reacts with FAD glucose dehydrogenase (FAD-GDH) enzyme on the reagent strip. The electrons generated by this reaction are shuttled to an electrode by a mediator chemical, producing a current that is proportional to the glucose in the sample. After a fixed reaction time, the glucose concentration in the sample is calculated and displayed.

AI/ML Overview

This document is a 510(k) premarket notification for the Contour Next USB Blood Glucose Monitoring System, which primarily describes modifications to an existing predicate device (K121087) rather than providing a detailed study of the modified device's performance against specific acceptance criteria. The modifications focus on improved error detection capabilities.

Therefore, much of the requested information regarding acceptance criteria, specific performance metrics, sample sizes, expert involvement, and ground truth for a full performance study is not explicitly detailed in the provided text. The submission focuses on demonstrating substantial equivalence to the predicate device due to specific enhancements.

Here's an attempt to extract the available information based on your request, with noted limitations where data is not provided:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state numerical acceptance criteria in a table format for blood glucose measurement accuracy for the modified device. Instead, it describes functional improvements related to error detection. The "reported device performance" focuses on the functionality of these new error checks and equivalency to the predicate device.

Acceptance Criteria (Inferred from improvements)Reported Device Performance (Modified Device)
Detect test strips exposed to degrading chemicalError message given for exposed strips
Detect un-mixed control solutionError message given for un-mixed control solution
Detect sample 'perturbation' during a testError message given for disturbed samples
For overall glucose measurement accuracy:
Not statistically different from predicate device (K121087)Equivalency testing confirmed performance not statistically different.
No good results classified as errors by modificationsSoftware verification ensured no good results were misclassified.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Test Set Sample Size: Not specified. The document mentions "bench testing" and an "internal user study" but does not give sample sizes for either.
  • Data Provenance: Not specified, but likely proprietary internal data from Bayer Healthcare. It doesn't specify country of origin or whether it's retrospective or prospective. Given the context of 510(k) submissions, such studies are typically prospective tests conducted by the manufacturer.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

Not applicable or not specified. This type of submission for a blood glucose meter does not typically involve expert consensus to establish ground truth in the same way an imaging or diagnostic AI device would. The "ground truth" for a blood glucose meter is usually established by a laboratory reference method. The document mentions "internal user study," suggesting lay users may have been involved in some testing, but no details on expert involvement for ground truth are provided.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable or not specified. Adjudication methods like 2+1 are typically used for subjective assessments (e.g., image interpretation) where multiple experts review and reconcile differences. For a quantitative device like a blood glucose meter, ground truth is established by a reference method, not by expert adjudication.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. MRMC studies are relevant for devices that assist human readers (e.g., AI for medical imaging). This is a standalone diagnostic device. The study mentions an "internal user study" but not in the context of an MRMC design or comparing human performance with/without AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Yes, in essence. The bench testing and software verification of the error detection mechanisms would fall under "standalone" algorithm testing to ensure the meter correctly identifies errors and does not misclassify good results. The equivalency testing would also implicitly assess the algorithm's accuracy compared to the predicate.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For blood glucose meters, the ground truth for accuracy studies is typically established by a laboratory reference method (e.g., YSI glucose analyzer) for whole blood glucose concentration. While not explicitly stated for this specific submission, this is the standard for blood glucose meter validation. The text focuses on the functional correctness of its new error checks rather than primary accuracy data against a reference method, as it's demonstrating equivalence to an already cleared device.

8. The sample size for the training set

Not applicable. As a medical device (blood glucose meter) and not an AI/ML algorithm that requires a "training set" in the conventional sense, this information is not relevant or provided. The "training" for such a device is its design and manufacturing process, and its performance is validated through testing.

9. How the ground truth for the training set was established

Not applicable, as there is no "training set" in the context of AI/ML for this device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The symbol is composed of three abstract human profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 25, 2015

BAYER HEALTHCARE JENNIFER GREGORY REGULATORY AFFAIRS SPECIALIST 430 SOUTH BEIGER ST. MISHAWAKA IN 46544

Re: K150942

Trade/Device Name: Contour Next USB Blood Glucose Monitoring System Regulation Number: 21 CFR 862.1345 Regulation Name: Glucose test system Regulatory Class: II Product Code: NBW. LFR Dated: October 29, 2015 Received: October 30, 2015

Dear Jennifer Gregory:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For:

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150942

Device Name

Contour Next USB Blood Glucose Monitoring System

Indications for Use (Describe)

The Contour Next USB blood glucose monitoring system is an over the counter (OTC) device utilized by persons with diabetes in home settings for the quantitative measurement of glucose in whole blood, and is for single-patient use only and should not be shared.

The system consists of a Contour Next USB blood glucose meter, Contour Next test strips and Contour Next control solutions.

The Contour Next USB blood glucose monitoring system is indicated for use with fresh capillary whole blood samples drawn from the fingertip and palm only. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly).

The Contour Next USB blood glucose monitoring system is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

The Contour Next Test Strips are for use with the Contour Next USB blood glucose monitoring system for the quantitative measurement of glucose in whole blood.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☑ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image is the Bayer company logo. The logo consists of the word "BAYER" stacked vertically in a circle. The circle is colored with green at the top and blue at the bottom.

510(k) Summary

Date prepared: November 23, 2015

According to the requirements of 21 CFR 807.92, the following information is being submitted in sufficient detail to provide an understanding of the basis for a determination of substantial equivalence.

1) SubmitterJennifer GregoryRegulatory Affairs SpecialistBayer Healthcare LLC430 South Beiger StreetMishawaka, IN 46544Telephone: (574) 256-3447Fax: (574) 256-3519
2) Device name:Trade name: Bayer Contour® Next USB Blood GlucoseMonitoring SystemCommon name: Blood Glucose MeterClassification name: Blood Glucose Test System, Over-the-Counter, 75 NBW, LFR (21 CFR § 862.1345)
3) Predicatedevice:Contour Next USB Blood Glucose Meter (Reference:Contour Next USB Blood Glucose MonitoringSystem (K121087))
4) Devicedescription:The Contour Next USB Blood Glucose Meter consists of asmall handheld blood glucose meter that utilizes dryreagent test strips for the measurement of glucose incapillary whole blood by persons with diabetes. Liquidcontrol solution is used to check the performance of thesystem. The meter, together with the test strips andcontrol solutions, is referred to as the Contour Next USBBlood Glucose Monitoring System.The chemical principle utilized for both the predicate andmodified devices is based on measurement of electricalcurrent caused by the reaction of glucose in the blood withchemicals on the reagent strip. The blood sample is drawninto the tip of the reagent strip through capillary action.Glucose in the sample reacts with FAD glucosedehydrogenase (FAD-GDH) enzyme on the reagent strip.The electrons generated by this reaction are shuttled to an

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" written twice, once vertically and once horizontally, with the letters intersecting in the middle. The logo is surrounded by a circle that is green on the top and blue on the bottom.

electrode by a mediator chemical, producing a current that is proportional to the glucose in the sample. After a fixed reaction time, the glucose concentration in the sample is calculated and displayed.

    1. Intended Use: The Contour Next USB blood glucose monitoring system is an over the counter (OTC) device utilized by persons with diabetes in home settings for the quantitative measurement of glucose in whole blood, and is for singlepatient use only and should not be shared.
      The system consists of a Contour Next USB blood glucose meter, Contour Next test strips and Contour Next control solutions.

The Contour Next USB blood glucose monitoring system is indicated for use with fresh capillary whole blood samples drawn from the fingertip and palm only. Alternative site testing should be done only during steady state times (when glucose is not changing rapidly).

The Contour Next USB blood glucose monitoring system is not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.

The Contour Next Test Strips are for use with the Contour Next USB blood glucose monitoring system for the quantitative measurement of glucose in whole blood.

Data demonstrating substantial equivalence

The Contour Next USB Blood Glucose Meter consists of a small handheld blood qlucose meter that is substantially equivalent to the predicate device, the Contour Next USB Blood Glucose Meter (K121087). The modified and predicate devices use the same glucose calculation algorithm. Both devices also use dry reagent test strips for the measurement of glucose in capillary whole blood by persons with diabetes and liquid controls to check the performance of the system. The same Contour Next test strips and Contour Next control solutions are used by both the modified and predicate devices.

A detailed comparison of the characteristics featured between the modified and predicate devices is provided in the tables below:

{5}------------------------------------------------

Image /page/5/Picture/1 description: The image is the logo for Bayer. The logo is a circle with the word "BAYER" written vertically in the center. The letters are arranged in a cross shape. The circle is green at the top and fades to blue at the bottom. The letters are gray with a silver outline.

Modified Device (Contour Next USB meter)

Image /page/5/Picture/3 description: The image shows a Bayer Contour Next One blood glucose meter. The meter's display shows a reading of 93 mg/dL, taken before a meal at 1:33 PM on November 10th. The meter has a sleek, black design with a gray button on the right side that has the Bayer logo.

Predicate Device (Contour Next USB meter)

Image /page/5/Picture/5 description: The image shows a Bayer Contour Next blood glucose meter. The meter displays a reading of 93 mg/dL, taken before a meal. The display also shows the time as 1:33 PM and the date as 11/10. The meter has buttons for reminder and notes.

Summary of the Technological Characteristics of the Modified Device Compared to Predicate
SIMILARITIES to Predicate
CharacteristicPredicateContour Next USB(K121087)Contour Next USB(Modified Device)
Test StripContour Next Test StripsSame as Predicate
Control SolutionContour Next Control Solution(Level 1 and 2)Same as Predicate
Detection MethodAmperometricSame as Predicate
Measuring Range20-600 mg/dLSame as Predicate
Sample Volume0.6 µLSame as Predicate
Countdown time displayed5 SecondsSame as Predicate
Illuminated Strip PortYesSame as Predicate
Operational Buttons4Same as Predicate
Battery TypeRechargeable (3.4-4.2V)Same as Predicate
Operating TemperatureRange41°-113° FSame as Predicate
Operating Humidity Range10-93% RHSame as Predicate
Hematocrit Range15%-65%Same as Predicate
Meter life5 YearsSame as Predicate
Validated Product Used forCleaning and DisinfectionClorox Germicidal wipesSame as Predicate
Meal MarkersYesSame as Predicate

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" arranged in a cross shape, with the letters stacked vertically and horizontally. The letters are in a sans-serif font and are colored gray. The logo is surrounded by a circle that transitions from green at the top to blue at the bottom.

Summary of the Technological Characteristics of the Modified Device Compared toPredicate
SIMILARITIES to Predicate
PredicateContour Next USB
CharacteristicContour Next USB(K121087)(Modified Device)
Calibration/CodingAutocoding (no coding forusers)Same as Predicate
User InterfaceAlphanumeric, Iconic, NativeLanguageSame as Predicate
Display (technology)Graphical (OLED)Same as Predicate
Display VisibilityDay and nightSame as Predicate
Communication PortUSB InterfaceSame as Predicate
Communication Link toComputerDirect USB connection oroptional USB cableSame as Predicate
Test Results in Memory2000 ResultsSame as Predicate
User ContactMaterials/Surface FinishPolycarbonite/ABS with ASplastics with mirror or mattefinishSame as Predicate

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the Bayer company logo. The logo consists of the word "BAYER" written twice, once horizontally and once vertically, forming a cross shape. The logo is surrounded by a circle that is green on the top half and blue on the bottom half.

DIFFERENCES from Predicate
CharacteristicPredicateContour NextUSB (K121087)Contour Next USB(Modified Device)Risk Assessment Summary
Improved detection of test strips thatmay have been exposed to a chemicalthat can degrade the mediator.NoYesThe error check improves the ability of themodified meter to detect exposed test strips andprovide an error message instead of a highbiased result.
Improved detection of un-mixed controlsolution.NoYesThe error check improves the ability of themodified meter to detect un-mixed control testsolutions and provide an error message insteadof a high biased result.
Improved detection of sample'perturbation' during a test.NoYesThe error check improves the ability of themodified meter to detect a sample that isdisturbed during the countdown period andprovide an error message instead of a biasedresult.

{8}------------------------------------------------

Summary of Performance testing

Bench testing was conducted to ensure that the error checks acted as intended and gave error messages for:

  • . test strips that had been exposed to a reducing agent
  • . un-mixed control test solution
  • . samples that had been disturbed during the test countdown

Software verification testing was conducted to ensure that no good results were classified as errors as a result of the modifications.

Equivalency testing was conducted to ensure the performance of the modified and predicate meters was not statistically different.

An internal user study was conducted to ensure the modifications did not impact the blood glucose result accuracy of the Contour Next USB system.

Conclusions from Performance Evaluations

Based on the outcome of the performance testing conducted, the modified Contour Next USB Blood Glucose Meter is substantially equivalent to the predicate Contour Next USB Blood Glucose Meter (K121087).

§ 862.1345 Glucose test system.

(a)
Identification. A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.(b)
Classification. Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.