(256 days)
MReye Embolization Coils are intended for use in peripheral arterial and venous vessel embolization procedures.
The MReye Embolization Coils are constructed from coiled Inconel (nickel chromium alloy) wire and synthetic fibers. The wire forms primary coil diameters from 0.025 to 0.038 inches, and the extended embolus lengths range from 1 to 20 cm. Upon exiting from the cartridge and the catheter. the coil forms a secondary curl diameter ranging from 0 (straight) to 45 mm.
This document is a 510(k) premarket notification for a medical device called "MReye® Embolization Coils." The purpose of the 510(k) is to demonstrate that the new device is substantially equivalent to a predicate device already on the market.
Based on the provided text, there is no information about acceptance criteria or a specific study proving the device meets acceptance criteria in the context of diagnostic performance (e.g., sensitivity, specificity, accuracy).
The document focuses on:
- Device Description: The physical composition and characteristics of the embolization coils.
- Intended Use: For peripheral arterial and venous vessel embolization procedures.
- Comparison to Predicate Device: States that the "MReye® Embolization Coils are identical in terms of intended use, principles of operation, materials of construction, and basic technological characteristics to the predicate device."
- Performance Data: Mentions "MRI Testing" was performed to verify that the implant will be labeled as "MR Conditional" with applicable parameters described in the Instructions for Use. This is a safety performance test related to MRI compatibility, not a diagnostic accuracy study for a device that interprets or processes medical images.
Therefore, I cannot populate the requested table and information beyond stating that the available text does not contain this kind of data.
Based on the provided text, the answer to your request is as follows:
There is no information regarding the acceptance criteria, device performance metrics (e.g., sensitivity, specificity, or accuracy), or a study design to assess such performance for a diagnostic device in the provided document.
This document is a 510(k) premarket notification for a vascular embolization device (MReye® Embolization Coils) and focuses on demonstrating substantial equivalence to a predicate device, primarily through material, design, and intended use comparison, along with MRI compatibility testing for safety. It is not a submission for an AI/ML-driven diagnostic device that would typically involve the type of performance studies you are asking about.
Therefore, I cannot provide the requested information for the following points as they are not present in the document:
- A table of acceptance criteria and the reported device performance
- Sample sizes used for the test set and data provenance
- Number of experts used to establish ground truth
- Adjudication method
- Multi-reader multi-case (MRMC) comparative effectiveness study
- Standalone performance study
- Type of ground truth used
- Sample size for the training set
- How ground truth for the training set was established
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird symbol, composed of three overlapping profiles facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Cook Incorporated Nozomi Yagi Regulatory Affairs Specialist 750 Daniels Way Bloomington, Indiana 47404
Re: K150931
Trade/Device Name: MReye® Embolization Coils Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: October 19, 2015 Received: October 20, 2015
Dear Nozomi Yagi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kenneth J. Cavanaugh -S
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K150931
Device Name MReye® Embolization Coils
Indications for Use (Describe)
MReye Embolization Coils are intended for use in peripheral arterial and venous vessel embolization procedures.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in white, block letters on a red background. Below the word "COOK" is the word "MEDICAL" in smaller, white, block letters, also on a red background. The registered trademark symbol is located to the right of the word "COOK".
10.0 510(k) SUMMARY
K150931
As required by 21 CFR §807.92 Date Prepared: April 3, 2015
I. SUBMITTER
| Applicant: | Cook Incorporated |
|---|---|
| Contact: | Nozomi Yagi |
| Applicant Address: | Cook Incorporated |
| 750 Daniels Way | |
| Bloomington, IN 47404 | |
| Contact Phone Number: | (812) 335-3575 ext. 104511 |
| Contact Fax Number: | (812) 332-0281 |
II. DEVICE
Trade Name: Common Name: Classification Name: Regulation: Product Code
MReye® Embolization Coils Vascular Embolization Device Device, Vascular, For Promoting Embolization 21 CFR §870.3300 KRD
III. PREDICATE DEVICE
The device subject of this submission is believed to warrant a determination of substantial equivalence to the predicate device, the Cook MReye Embolization Coils (K090624). This predicate has never been subject to a design-related recall, and no reference device was used in this submission.
IV. DEVICE DESCRIPTION
The MReye Embolization Coils are constructed from coiled Inconel (nickel chromium alloy) wire and synthetic fibers. The wire forms primary coil diameters from 0.025 to 0.038 inches, and the extended embolus lengths range from 1 to 20 cm. Upon exiting from the cartridge and the catheter. the coil forms a secondary curl diameter ranging from 0 (straight) to 45 mm.
V. INDICATIONS FOR USE
MReye Embolization Coils are intended for use in peripheral arterial and venous vessel embolization procedures.
{4}------------------------------------------------
VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
The modification included in this submission expands the using existing technology and does not include any new technological characteristics. The MReye® Embolization Coils are comparable to the predicate device. Specifically, the MReye® Embolization Coils are identical in terms of intended use, principles of operation, materials of construction, and basic technological characteristics to the predicate device.
In addition to product expansion, a revision to the MRI labeling to update MR Conditional language has been included in this submission. The modifications are adequately supported by testing.
VII. PERFORMANCE DATA
The following test was performed to determine the appropriate Magnetic Resonance Imaging information for the proposed device:
- . MRI Testing - MRI testing verifies that the implant will be labeled as "MR Conditional" with the applicable parameters described in the Instructions for Use.
VIII. CONCLUSIONS
The result of the testing provides reasonable assurance that the proposed device has been designed and tested to assure conformance to the requirements of its intended use. The proposed device also does not raise new questions of safety or effectiveness as compared to the predicate device and thus is substantially equivalent to the predicate device.
§ 870.3300 Vascular embolization device.
(a)
Identification. A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.(b)
Classification. Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).