K Number
K150890

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-05-29

(57 days)

Product Code
Regulation Number
888.3560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Exactech® Optetrak Logic® CC is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.

The Exactech® Optetrak Logic® CC is indicated for cemented use only.

Device Description

Optetrak Logic CC is a cemented total knee prosthesis system that provides femoral components, tibial inserts, stems, augments, and surgical instrumentation for use in primary and revision total knee arthroplasty. Like corresponding predicate Optetrak CC system femoral components and tibial inserts, Optetrak Logic CC femoral components and tibial inserts are constrained condylar prostheses featuring an increased level of constraint compared to other Optetrak and Optetrak Logic knee system devices. Optetrak Logic CC also provides various stem extensions and augment blocks surgeons can use to address bone defects and bone loss.

AI/ML Overview

The provided text is a 510(k) premarket notification for the Exactech® Optetrak Logic® CC knee prosthesis. This document does not contain the detailed information requested regarding acceptance criteria and the study proving the device meets them.

Specifically, the document does not include:

  • A table of acceptance criteria and reported device performance.
  • Information on sample size, data provenance, number or qualifications of experts, or adjudication methods for any test set.
  • Details on a multi-reader multi-case (MRMC) comparative effectiveness study or any effect sizes.
  • Information on standalone algorithm performance.
  • The type of ground truth used, sample size for the training set, or how ground truth for the training set was established.

Instead, the document primarily focuses on:

  • Regulatory approval: It is a letter from the FDA determining substantial equivalence to predicate devices.
  • Device description: It describes the Exactech® Optetrak Logic® CC as a cemented total knee prosthesis system with various components.
  • Intended use: It specifies the indications for use in skeletally mature individuals for total knee replacement due to various conditions and for revision surgery.
  • Substantial Equivalence: It claims substantial equivalence to existing cleared Exactech Optetrak devices.
  • Testing referenced: It mentions "mechanical testing" including "Constraint testing" and "Stem extension fatigue testing" but does not provide results, acceptance criteria, or details of these tests within the provided text. The statement "Results of engineering studies referenced in this 510(k) submission demonstrate proposed Optetrak Logic CC Total Knee System devices are substantially equivalent to cited cleared predicate Optetrak devices" is a conclusion of testing, not a detailed report of the testing itself.

Therefore, I cannot provide the requested information based on the given text.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name in a circular arrangement around an emblem. The emblem consists of three stylized human profiles facing right, stacked on top of each other, representing health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 29, 2015

Exactech, Inc. Patrick Hughes Senior Regulatory Affairs Specialist 2320 NW 66th Court Gainesville, Florida 32653

Re: K150890 Trade/Device Name: Exactech® Optetrak Logic® CC Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer/Metal/Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH Dated: March 30. 2015 Received: April 2, 2015

Dear Patrick Hughes,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K150890

Device Name Exactech® Optetrak Logic® CC

Indications for Use (Describe)

The Exactech® Optetrak Logic® CC is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.

The Exactech® Optetrak Logic® CC is indicated for cemented use only.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Exactech® Optetrak Logic® CC Traditional 510(k) - 510(k) Summary of Safety and Effectiveness

Sponsor:Exactech, Inc.2320 N.W. 66th CourtGainesville, FL 32653
---------------------------------------------------------------------------

Phone: (352) 377-1140 (352) 378-2617 Fax:

FDA Establishment Number 1038671

  • Contact: Patrick Hughes Senior Regulatory Affairs Specialist
  • Date: March 30, 2015

Trade or Proprietary or Model Name(s):

Exactech® Optetrak Logic® CC

Common Name:

Cemented Total Knee Prosthesis

Classification Name:

Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Information on devices to which substantial equivalence is claimed:

510(k) NumberTrade or Proprietary Model NameManufacturer
K954208Optetrak CCExactech, Inc
K093360Optetrak LogicExactech, Inc
K933610Optetrak Trapezoidal TraysExactech, Inc.
K012251Optetrak Stem ExtensionsExactech, Inc

Indications for Use:

The Optetrak Logic CC Total Knee System is indicated for use in skeletally mature individuals undergoing primary surgery for total knee replacement due to osteoarthritis, osteonecrosis, rheumatoid arthritis and/or post-traumatic degenerative problems. They are also indicated for revision of failed previous reconstructions where sufficient bone stock and soft tissue integrity are present.

The Optetrak Logic CC Total Knee System is indicated for cemented use only.

{4}------------------------------------------------

Exactech® Optetrak Logic® CC Traditional 510(k) - 510(k) Summary of Safety and Effectiveness

Device Description:

Optetrak Logic CC is a cemented total knee prosthesis system that provides femoral components, tibial inserts, stems, augments, and surgical instrumentation for use in primary and revision total knee arthroplasty. Like corresponding predicate Optetrak CC system femoral components and tibial inserts, Optetrak Logic CC femoral components and tibial inserts are constrained condylar prostheses featuring an increased level of constraint compared to other Optetrak and Optetrak Logic knee system devices. Optetrak Logic CC also provides various stem extensions and augment blocks surgeons can use to address bone defects and bone loss.

Testing Description:

This submission includes references to the following mechanical testing:

  • Constraint testing
  • Stem extension fatigue testing .

Substantial Equivalence Conclusion:

Results of engineering studies referenced in this 510(k) submission demonstrate proposed Optetrak Logic CC Total Knee System devices are substantially equivalent to cited cleared predicate Optetrak devices.