K Number
K150868

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-06-23

(83 days)

Product Code
Regulation Number
866.2680
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Solana™ GAS Assay is a rapid in vitro diagnostic test for the qualitative detection of Group A B-hemolytic Streptococus (Streptococcus pyogenes) nucleic acids isolated from throat swab specimens obtained from patients with signs and symptoms of pharyngitis, such as sore throat. The Solana™ GAS Assay is intended for use only with the Solana™ instrument.

Device Description

The Solana™ GAS Assay amplifies and detects GAS DNA present in throat swab specimens obtained from symptomatic patients.

The assay consists of two major steps: 1) specimen preparation, and 2) amplification and detection of target sequence specific to GAS using isothermal Helicase-Dependent Amplification (HDA) in the presence of target-specific fluorescence probe.

Patient specimen on a throat swab is transferred to a Lysis Tube and subjected to heattreatment at 95°C for 5 minutes. The heat-treated sample is added to a Dilution Tube, and then transferred to a Reaction Tube. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of GAS-specific target sequence. In Solana, the target sequence is amplified by GAS specific primers and detected by a GAS specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by GAS specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target and PRC probes carry a ribonucleic acid. Upon annealing to GAS or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and it can print out the results via a printer.

AI/ML Overview

The Solana™ GAS Assay is a rapid in vitro diagnostic test for the qualitative detection of Group A β-hemolytic Streptococcus (Streptococcus pyogenes) nucleic acids isolated from throat swab specimens obtained from patients with signs and symptoms of pharyngitis, such as sore throat. The assay is intended for use only with the Solana™ instrument.

1. Table of Acceptance Criteria and Reported Device Performance

Performance MetricAcceptance Criteria (Implicit for predicate device)**Reported Device Performance (Solana™ GAS Assay)
Clinical Sensitivity96.5% (95% CI: 91.3% - 98.6%)*98.2% (95% CI: 95.5% - 99.3%)
Clinical Specificity98.0% (95% CI: 97.0% - 98.6%)*97.2% (95% CI: 95.9% - 98.1%)
Analytical Sensitivity (LoD)Not explicitly stated but expected to be comparable to predicate for substantial equivalence6.81 x 10⁴ CFU/mL

*The predicate device's performance (Lyra™ Direct Strep Assay) in the "Comparison with predicate" section is used as a proxy for implicit acceptance criteria, as the submission aims to demonstrate substantial equivalence. Explicit acceptance criteria beyond this are not detailed in the provided text for the Solana™ GAS Assay's clinical performance.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: 1082 fresh throat swab specimens.
  • Data Provenance: Prospective study conducted from December 2014 to February 2015 at four distinct geographical sites across the United States.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

The document does not explicitly state the number of experts or their specific qualifications (e.g., radiologists with 10 years of experience). However, the ground truth was established by "conventional streak-stab culture technique onto a trypticase soy agar plate containing 5% horse red blood cells" and "typed as Lancefield group A by latex agglutination." This implies standard microbiology laboratory procedures and interpretation, which would typically be performed by trained clinical microbiologists or laboratory technologists.

4. Adjudication Method for the Test Set

The document describes discordant specimens. Of the 24 specimens that were Solana™ GAS Assay positive but Composite Culture negative, 16 were tested with an additional FDA-cleared molecular device, and 8 were negative. Of the 4 specimens that were Solana™ GAS Assay negative but Composite Culture positive, 3 were tested with an additional FDA-cleared molecular device. This indicates a form of secondary testing or adjudication for discordant results, though a formal 2+1 or 3+1 method is not explicitly named.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

No, an MRMC comparative effectiveness study involving human readers and AI assistance was not done. This device is an in-vitro diagnostic test for qualitative detection of nucleic acids, not an imaging or diagnostic aid for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Yes, the clinical performance study evaluated the Solana™ GAS Assay as a standalone device. The "Solana instrument measures and interprets the fluorescent signal, using on-board method-specific algorithms" and "report the test results to the user on its display screen."

7. The Type of Ground Truth Used

The ground truth used was composite bacterial culture. A specimen was considered positive if either the swab or the transport media was positive for β-hemolytic Streptococcus and typed as Lancefield group A by latex agglutination.

8. The Sample Size for the Training Set

The document does not explicitly mention a separate "training set" for the clinical performance evaluation. The clinical sensitivity and specificity were established using a prospective clinical study where 1082 fresh throat swab specimens were collected and tested. This dataset largely serves as the evaluation or test set. For analytical performance, various specified quantities of bacterial strains were used.

9. How the Ground Truth for the Training Set Was Established

Given that a specific "training set" for clinical performance with a separate ground truth establishment method is not described, the approaches described for the test set (composite bacterial culture as detailed in point 7) would apply if any training data were used from clinical samples. For analytical studies, ground truth was established by using quantified (CFU/mL) cultures of known bacterial strains.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with the profiles overlapping each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 1st, 2015

QUIDEL CORPORATION RONALD H. LOLLAR SENIOR DIRECTOR, CLINICAL AND REGULATORY AFFAIRS 2005 EAST STATE STREET, SUITE 100 ATHENS, OH 45701

Re: K150868

Trade/Device Name: Solana™ Gas Assay, Solana™ Instrument Regulation Number: 21 CFR 866.2680 Regulation Name: Streptococcus spp. nucleic acid-based assay Regulatory Class: II Product Code: PGX Dated: March 31, 2015 Received: April 1, 2015

Dear Mr. Lollar:

This letter corrects our substantially equivalent letter of June 23, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{1}------------------------------------------------

medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Sally A. Hojvat -S

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150868

Device Name Solana™ GAS Assay, Solana™ Instrument

Indications for Use (Describe)

The Solana™ GAS Assay is a rapid in vitro diagnostic test for the qualitative detection of Group A B-hemolytic Streptococus (Streptococcus pyogenes) nucleic acids isolated from throat swab specimens obtained from patients with signs and symptoms of pharyngitis, such as sore throat. The Solana™ GAS Assay is intended for use only with the Solana™ instrument.

Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Applicant:

Quidel Corporation 12544 High Bluff Drive, Suite 200 San Diego, California 92130 Telephone: 858-552-7910 Fax: 858-646-8045

Contact Information:

Ronald H. Lollar, Senior Director Clinical and Regulatory Affairs 2005 East State Street, Suite 100 Athens, Ohio 45701 740-589-3300 – Corporate number 740-589-3373 – Desk phone 858-552-6451—Fax Ron.Lollar@quidel.com

Date of preparation of 510(k) summary:

March 31, 2015

A. 510(k) Number:

K150868

B. Purpose for Submission:

To obtain substantial equivalence for the Solana™ GAS Assay when performed on the Solana™ instrument

C. Measurand:

DNase B (sdaB) sequence of Streptococcus pyogenes (Group A Streptococcus)

D. Type of Test:

Helicase-dependent amplification (HDA)

{4}------------------------------------------------

E. Applicant:

Quidel Corporation

F. Proprietary and Established Names:

Solana™ GAS Assay Solana™ instrument

G. Regulatory Information:

Product CodeClassificationRegulation SectionPanel
PGX – Groups A, C and G Beta-HemolyticStreptococcusNucleic AcidAmplificationClass II (Non-exempt)21 CFR 866.2680 – Streptococcus spp. Nucleic Acid-Based AssayMicrobiology (83)

H. Intended Use:

    1. Intended Use(s):
      The Solana™ GAS Assay is a rapid in vitro diagnostic test for the qualitative detection of Group A β-hemolytic Streptococcus (Streptococcus pyogenes) nucleic acids isolated from throat swab specimens obtained from patients with signs and symptoms of pharyngitis, such as sore throat. The Solana™ GAS Assay is intended for use only with the Solana™ instrument.
    1. Indication(s) for Use:
      Same as intended Use
    1. Special conditions for use statement(s):
    • For in vitro diagnostic use only
    • . For prescription use only

{5}------------------------------------------------

4. Special instrument requirements:

Solana™ instrument

l. Device Description:

The Solana™ GAS Assay amplifies and detects GAS DNA present in throat swab specimens obtained from symptomatic patients.

The assay consists of two major steps: 1) specimen preparation, and 2) amplification and detection of target sequence specific to GAS using isothermal Helicase-Dependent Amplification (HDA) in the presence of target-specific fluorescence probe.

Patient specimen on a throat swab is transferred to a Lysis Tube and subjected to heattreatment at 95°C for 5 minutes. The heat-treated sample is added to a Dilution Tube, and then transferred to a Reaction Tube. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of GAS-specific target sequence. In Solana, the target sequence is amplified by GAS specific primers and detected by a GAS specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by GAS specific primers and detected by a PRC specific fluorescence probe.

The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target and PRC probes carry a ribonucleic acid. Upon annealing to GAS or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and it can print out the results via a printer.

Materials Provided:

  • 48 Tests per Kit
ComponentQuantityStorage
Dilution Buffer48 tubes/kit 0. 5 mL2° to 8°C
Lysis Buffer48 tubes/kit 0. 5 mL2° to 8°C
Reaction Tubes48 tubes/kit2° to 8°C

{6}------------------------------------------------

Materials required but not provided:

  • . External controls for Group A Streptococcus (e.g. Quidel Molecular A + G Streptococci Control Set #M111, which contains positive and negative controls, serves as an external processing and extraction control)
  • Sterile DNAse-free filter-blocked or positive displacement micropipettor tips
  • Micropipettor
  • Stopwatch or timer
  • Scissors or a blade
  • Micro tube tray ●
  • Heat block capable of 95° C ± 2° C temperature ●
  • Solana Instrument
  • Thermometer

J. Substantial Equivalence Information:

    1. Predicate device name(s):
      Lyra™ Direct Strep
    1. Predicate 510(k) number(s): K133883
    1. Comparison with predicate:
Similarities
ItemSolana™ GAS AssayLyra™ Direct Strep Assay(K133883)
Intended UseThe Solana™ GAS Assay is arapid in vitro diagnostic testfor the qualitative detectionof Group A β-hemolyticStreptococcus( Streptococcus pyogenes )nucleic acids isolated fromthroat swab specimensobtained from patients withsigns and symptoms ofpharyngitis, such as sorethroat. The Solana™ GASThe Lyra Direct StrepAssay is a Real-Time PCRin vitro diagnostic test forthe qualitative detectionand differentiation ofGroup A β-hemolyticStreptococcus( Streptococcus pyogenes )and pyogenic Group Cand G β-hemolytic
Similarities
ItemSolana™ GAS AssayLyra™ Direct Strep Assay(K133883)
Assay is intended for useonly with the Solana™instrument.Streptococcus nucleicacids isolated fromthroat swab specimensobtained from patientswith signs and symptomsof pharyngitis, such assore throat. The assaydoes not differentiatebetween pyogenicGroups C and G β-hemolytic Streptococcus .All negative test resultsshould be confirmed bybacterial culture,because negative resultsdo not preclude Group A,C or G Strep infectionand should not be usedas the sole basis fortreatment.The assay is intended foruse in hospital, reference,or state laboratorysettings. The device is notintended for point-of-careuse.
Similarities
ItemSolana™ GAS AssayLyra™ Direct Strep Assay(K133883)
Sample TypesThroat swab specimensSame
Sample Heat LysisManualSame
Detection TechniqueAutomatically detectsfluorescence afterdissociation of fluorophorefrom quencher duringamplificationSame
Testing Time45 minutes60 - 70 minutes

{7}------------------------------------------------

{8}------------------------------------------------

Differences
ItemSolana™ GAS AssayLyra™ Direct Strep Assay(K133883)
DNA AmplificationTechnologyHelicase-dependentamplification (HDA); self-containedReal time polymerasechain reaction
InstrumentSolana™ABI 7500 Fast DXThermocycler
Target Sequence Detected78 base pair (bp) sequenceS. pyogenes genome,resident in the DNase B( sdaB ) geneGAS* – 99bp product inthe putative competence( comX1.1 ) genePyo GCS/GGS* – 188bpproduct in the tagatose-6-phosphate kinase ( lacC )gene
Clinical Sensitivity98.2% (95% CI: 95.5-99.3%)GAS Sensitivity:96.5%[95% CI: 91.3% -

{9}------------------------------------------------

Differences
ItemSolana™ GAS AssayLyra™ Direct Strep Assay (K133883)
98.6%Pyo GCS/GGS Sensitivity:95.7%[95% CI: 88.1% - 98.5%]
Clinical Specificity97.1% (95% CI: 95.7-98.0%)GAS Specificity:98.0%[95% CI: 97.0% - 98.6%]Pyo GCS/GGS Specificity:98.3%[95% CI: 97.4% - 98.9%]

K. Standard/Guidance Document Referenced (if applicable):

screen, and it can print out the results via a printer.

Not applicable

L. Test Principle:

Patient specimen on a throat swab is transferred to a Lysis Tube and subjected to heattreatment at 95°C for 5 minutes. The heat-treated sample is added to a Dilution Tube, and then transferred to a Reaction Tube. The Reaction Tube contains lyophilized HDA reagents, dNTPs, primers and probes. Once rehydrated with the diluted sample, the Reaction Tube is placed in Solana for amplification and detection of GAS-specific target sequence. In Solana, the target sequence is amplified by GAS specific primers and detected by a GAS specific fluorescence probe included in the Reaction Tube. A competitive process control (PRC) is included in the Lysis Tube to monitor sample processing, inhibitory substances in clinical samples, reagent failure or device failure. The PRC target is amplified by GAS specific primers and detected by a PRC specific fluorescence probe. The target and PRC probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target and PRC probes carry a ribonucleic acid. Upon annealing to GAS or PRC amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. Solana measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display

{10}------------------------------------------------

M. Performance Characteristics:

1. Analytical performance:

  • a. Precision/Reproducibility:

Reproducibility

In order to confirm the reproducibility of the Solana™ GAS Assay, a blinded and randomized study panel containing Streptococcus pyogenes negative and positive samples were tested at three test sites (one in-house laboratory and two clinical sites) with three (3) instruments. Each site tested a reproducibility panel and Assay Controls for five days in triplicate. Testing was done by two operators at each site. Each operator ran the panel once a day using one lot of Solana™ GAS Assay. A total of 540 specimens were tested (including controls). The Solana™ GAS Assay generated the following reproducible results in this study.

CategorySITEOverall Percent Positive95% Confidence Interval
Site #1Site #2Site #3
Detected:#positive /# tested% PositiveDetected:#positive /# tested% PositiveDetected:#positive /# tested% Positive
GAS High Negative24/3080%20/3067%14/3047%58/9064%54-74%
GAS Low Positive30/30100%30/30100%30/30100%90/90100%96%-100%
GAS Moderate Positive30/30100%30/30100%30/30100%90/90100%96%-100%
GAS Negative0/300%0/300%0/300%0/900%0%-4%
GAS Positive Control30/30100%30/30100%30/30100%90/90100%96%-100%
GAS Negative Control0/300%0/300%0/300%0/900%0%-4%

The results suggest that there are no significant differences between different users using different instruments at different sites on different days.

  • b. Linearity/assay reportable range:
    Not applicable - This assay is qualitative.

  • c. Traceability, Stability, Expected values (controls, calibrators, or methods):

{11}------------------------------------------------

Traceability:

Not applicable. This assay is qualitative.

Specimen Stability:

A study was performed to determine the stability of samples collected in a number of routinely used swab systems: nylon flocked swabs in amies media, Rayon swab in amies media, polyester swab in amies media, Rayon swab in stuart media, polyester swab in stuart media, and Rayon in amies gel.

A freshly grown stock of GAS bacteria of known titer was used to spike the swabs listed above. The spiked samples were stored at 25°C ± 2°C for 2 days and then at 2 to 8°C for up to 8 more days prior to being tested in the Solana™ GAS Assay. A separate study was performed where the spiked samples were stored at ≤-15℃ or <-70°C for a minimum of 32 days before testing.

Based on this study, specimens collected using various collection/transport systems listed above can be stored at 25°C ± 2°C for 2 days and then at 2 to 8°C for up to 8 more days before testing or at ≤-15°C or ≤-70°C for up to 32 days before testing.

Controls:

Controls (Quidel Molecular A + G Streptococci Control Set #M111, which contains positive and negative controls, serves as an external processing and extraction control) were run on the Solana™ GAS Assay each day of testing. These controls are described as follows:

  • a. The process control is used to monitor sample processing, to detect HDA inhibitory specimens and to confirm the integrity of assay reagents and cassette detection. The process control is included in the Lysis Buffer tube.
  • b. The external positive control may be treated as a patient specimen. The control should be sampled and tested as if it were a swab specimen and processed as described in the Assay Procedure. The external positive control is intended to monitor substantial reagent and cassette failure.

{12}------------------------------------------------

Quidel Corporation

  • c. The external neqative control may be treated as a patient specimen. The control should be sampled and tested as if it were a swab specimen and processed as described in the Assay Procedure. The external negative control is used to detect reagent or environmental contamination (or carry-over) by GAS DNA or amplicon.
  • d. Detection limit:

The analytical sensitivity (limit of detection or LoD) of the Solana™ GAS Assay was determined using quantified (CFU/mL) cultures of two Streptococcus pyoqenes strains by serial dilution. The LoD is defined as the lowest concentration at which 95% of all replicates tested positive.

The GAS bacterial strains were freshly grown. The cell density of these bacterial suspensions was estimated using the OD600 reading. After a cell suspension with OD600 of 0.1 (0.5 McFarland units) were established, the bacteria were serially diluted in to densities ranging from 3x to 0.3x LoD levels based on preliminary studies.

Each test dilution was run as 20 replicates in the Solana™ assay and plated on 20 TSA + 5% blood plates. The study was performed in 5 experiments of 14 assays per strain. For each experiment, 4 replicates of each of the three bacterial dilutions were performed, along with a positive and a negative run control. All five experiments of the LoD study for each strain were completed within 8 hours. The stocks of cells were stored on ice or at 4 °C when not in use.

The highest dilution where at least 19 of 20 replicates show detection of GAS (95% positivity) was assigned the Limit of Detection of the strain. The CFU/mL was calculated based on the average bacterial plate count of the dilution.

The LoD for the 2 Streptococcus pyogenes strains tested were 2.44 x10ª CFU/mL (ATCC #19615) and 6.81 x10 CFU/mL (ATCC #12344). Based on this data the reported LoD for the Solana™ GAS Assay is 6.81 x104 CFU/mL.

{13}------------------------------------------------

e. Analytical specificity:

Cross Reactivity:

An in silico BLAST analysis of primers used in the Solana™ GAS Assay against sixtyone (61) potential interfering organisms (see below) did not show evidence of crossreactivity.

Arcanobacterium sp.Human adenovirus FLactobacillus sp.1
Bacillus sp.Human adenovirus GLegionella pneumophila
Bacteroides sp.2Human coronavirus 229EMeasles virus
Bordetella sp.Human coronavirus HKU1Human Metapneumovirus
Branhamella sp.Human coronavirus NL63Moraxella sp.
Burkholderia sp.Human enterovirus AMumps virus
Campylobacter sp.3Human enterovirus BMycoplasma pneumoniae
Candida sp.Human enterovirus CNeisseria sp.
Corynebacterium sp.Human enterovirus DPeptostreptococcus sp.
CytomegalovirusHuman herpesvirus 1Proteus sp.
Enterobacterio phage MS2Human herpesvirus 2Pseudomonas sp.
Enterococcus sp.Human herpesvirus 4Respiratory syncytial virus Type B
Escherichia coliHuman parainfluenza virus 1Saccharomyces cerevisiae
Fusobacterium sp.Human parainfluenza virus 2Serratia sp.
Haemophilus sp.Human parainfluenza virus 3Staphylococcus sp.
Human adenovirus AHuman parainfluenza virus4a and 4bTreponema sp.
Human adenovirus BInfluenza virus AVeillonella sp.
Human adenovirus CInfluenza virus BYersinia sp.
Human adenovirus DInfluenza virus CPrevotella oralis4
Human adenovirus EKlebsiella sp.Parvimonas micra5
Veillonella parvula

¹ Includes L. acidophilus

2 Includes B. ovatus

3 Includes C. rectus

4 In NCBI, Bacteroides oralis is Prevotella oralis.

5 In NCBI, Peptostreptococcus micros is Parvimonas micra.

{14}------------------------------------------------

A study was performed to evaluate the performance of the Solana™ GAS Assay in the presence of forty-six (46) other microorganisms commonly found in throat specimens. Each potentially interfering microorganism was tested in the presence of 2 x LoD Group A Streptococcus (2 strains) in the presence of clinically relevant levels of viruses (10 pfu/ml) and bacteria (10 cfu/mL) or higher. All strain combinations were spiked on to swabs. The strains included in the cross-reactivity study are shown in the table below.

Acinetobacter IwoffiiStaphylococcus epidermidis MRSE
Arcanobacterium haemolyticumStenotrophomonas maltophilia
Bacillus cereusStreptococcus agalactiae
Bordetella pertussisStreptococcus anginosus
Burkholderia cepaciaStreptococcus bovis
Corynebacterium diphtheriaStreptococcus canis
Enterococcus faecalisStreptococcus dysgalactiae subsp equisimilis
Escherichia coliStreptococcus gordonii (Virdans type)
Fusobacterium necrophorumStreptococcus intermedius
Haemophilus influenza type AStreptococcus mitis
Klebsiella pneumoniaStreptococcus mutans
Lactobacillus acidophilusStreptococcus oralis
Lactococcus lactisStreptococcus pneumoniae
Legionella jordanisStreptococcus salivarius
Legionella micdadeiStreptococcus sanguinis
Legionella pneumophilaStreptococcus suis
Moraxella cartarrhalisCandida albicans
Neisseria gonorrhoeaeAdenovirus Type 1
Neisseria subflavaAdenovirus Type 11 (Slobitski)
Peptostreptococcus micros (aka Parvimonasmicra)Influenza A
Pseudomonas aeruginosaInfluenza B
Serratia marcescensParainfluenza Type 4B (VR-1377)
Staphylococcus aureus MRSARhinovirus Type 15 (1734)

None of the organisms or viruses tested above cross-reacts with the performance of the Solana™ GAS Assay.

{15}------------------------------------------------

Interference:

A study was conducted using two strains of Streptococcus pyogenes (ATCC 19615 and 12344) tested near LoD to evaluate the Solana™ GAS Assay for potential interference using a panel consisting of twenty-eight (28) common biological and chemical substances found in throat samples. Substances were introduced into the assay dilution tubes at concentrations which were medically relevant. Each of the strains was tested for each substance. None of the substances tested were found to interfere with the Solana™ GAS Assay.

Substance NameTestConcentrationInterference?(Y/N)
Children's DimetappDM Cold & CoughElixir25% v/vNo
Chloraseptic Max:Sore Throat Relief10% v/vNo
BreathSavers 3 HourMint-Spearmint10% w/vNo
Cepacol Sore Throat:Cherry Flavor5% w/vNo
Robitussin Cough &Cold-CF Max10% v/vNo
Ricola MountainHerb ThroatDrops-Sugar Free15% w/vNo
Human Saliva10% v/vNo
Robitussin NighttimeCold, & Flu10% v/vNo
Crest Pro-HealthNight Mint25% v/vNo
CVS Tussin CF15% v/vNo
Chloraseptic ThroatCherry lozenge10% w/vNo
Halls CherryMentholyptus15% w/vNo
Tic Tac Freshmints10% w/vNo
Zicam® Oral Mist0.625% v/vNo

{16}------------------------------------------------

Substance NameTestConcentrationInterference?(Y/N)
SucretsComplete-VaporCherry5% w/vNo
Acetaminophen19.5 mg/mLNo
Aspirin12.3 mg/mLNo
Ibuprofen15.6 mg/mLNo
Benadryl2.7 mg/mLNo
Crest® CompleteToothpaste5% w/vNo
Contac® Cold + FluCaplets Night10% w/vNo
Children's Wal-TapElixir Cold & Allergy(Dimetap Children'sCold and Allergy)25% v/vNo
Children's Wal-TapDM Elixir Cold &Cough25% v/vNo
Robitussin NighttimeCough, Cold, & Flu(peak cold)10% v/vNo
Halls Mentholyptus(not cherry flavor)15% w/vNo
Listerine Cool MintAntiseptic15% v/vNo
Whole Blood5% v/vNo
Mucin (BovineSubmaxillary Gland,type I-S)5.0 mg/mLNo

Analytical Reactivity (Inclusivity):

The reactivity of the Solana™ GAS Assay was evaluated against an additional seven (7) strains of Streptococcus pyogenes (GAS) at concentrations near the limit of detection (LoD) of the assay.

{17}------------------------------------------------

Each strain was tested as three replicates in the Solana™ GAS Assay. The study was performed in multiple experiments. For each experiment, three replicates of up to three strains were performed, along with a positive and a negative run control. All seven strains were detected by the Solana™ GAS Assay.

Bacterial StrainConcentrationCFU/mLStrain Detected(Yes/No)
ATCC 12384$6.81 x10^4$Yes
NCIMB 13285$6.81 x10^4$Yes
CCUG 33061$6.81 x10^4$Yes
CCUG 33409$6.81 x10^4$Yes
CCUG 39158$6.81 x10^4$Yes
ATCC 49399$6.81 x10^4$Yes
CCUG 53553$6.81 x10^4$Yes
  • f. Assay cut-off:
    Not applicable.

2. Comparison studies:

  • a. Method comparison with predicate device:
    Not applicable

  • b. Matrix comparison:
    A comparison study was conducted between negative clinical matrix and the contrived negative matrices used in the analytical studies in order to validate their use. The matrix comparison study results are shown in the table below.

{18}------------------------------------------------

Contrived NegativeMatrixPooled Negative ClinicalMatrix
Detected% PosDetected% Pos
StreptococcuspyogenesATCC 196151 x LoD20/20100%20/20100%

These studies demonstrate that the contrived negative matrices do not alter the performance of the device in the context of these analytical studies.

3. Clinical studies:

a. Clinical Sensitivity:

Performance characteristics of the Solana GAS Assay were established during a prospective study conducted December 2014 to February 2015. One thousand eighty-two (1082) fresh throat swab specimens from female and male patients were collected at four distinct geographical sites across the United States. A single specimen was collected per patient. Samples were collected on Polyester, Nylon or Rayon Swab with liquid Amies or Polyester Swab or Rayon with liquid Stuart's. The swabs were inoculated by conventional streak-stab culture technique onto a trypticase soy agar plate containing 5% horse red blood cells. Testing with the Solana device was performed at the four external laboratories using the same swab that was plated for the culture. All residual specimen transport media from the samples was shipped daily (with cold packs) to a central location. The transport media was cultured using the same testing protocol as that employed by the clinical sites.

One thousand eighty-two (1082) fresh throat specimens were cultured for Group A β-hemolytic Streptococcus and tested with the Solana GAS Assay. The specimens were cultured at the testing sites and the transport media was cultured at a central location. The specimen was considered positive if either the swab or the transport media was positive for ß-hemolytic Streptococcus (Composite Culture) and typed as Lancefield group A by latex agglutination. The table below details the overall performance using composite culture results as a reference.

{19}------------------------------------------------

Performance Results of Solana GAS Assay for Group A β-hemolytic Streptococcus
Overall Performance (All Sites)
Solana GAS AssayComposite Culture PositiveNegativeTotal
Positive22024*244
Negative4**833837
Total2248571081
95% CI
Sensitivity220/22498.2%95.5% to 99.3%
Specificity833/85797.2%95.9% to 98.1%
* Of the twenty-four (24) discordant specimens, sixteen (16) of these specimens werepositive for GAS when tested with an additional FDA-cleared molecular device, eight (8)were negative.** Of the four (4) discordant specimen, three (3) were negative when tested with anadditional FDA-cleared molecular device.
Site 1 Performance
Solana® GAS AssayComposite Culture PositiveNegativeTotal
Positive60565
Negative1333334
Total61338399
95% CI
Sensitivity60/6198.4%91.3% to 99.7%
Specificity333/33898.5%96.6 % to 99.4%
Site 2 Performance
Solana® GAS AssayComposite Culture PositiveNegativeTotal
Positive69978
Negative1134135
Total70143213
95% CI
Sensitivity69/7098.6%92.3% to 99.7%
Specificity134/14393.7%88.5 % to 96.7%
Site 3 Performance
Solana® GAS AssayComposite Culture PositiveNegativeTotal
Positive29635
Negative0186186
Total29192221
95% CI
Sensitivity29/29100%88.3% to 100%
Specificity186/19296.9%93.4 % to 98.6%

{20}------------------------------------------------

  • b. Clinical specificity:
    See Section 3a.

  • ﻥ Other clinical supportive data (when a. and b. are not applicable):
    Not applicable

4. Clinical cut-off:

Not applicable

5. Expected values:

The overall prevalence of Group A ß-hemolytic Streptococcus in patients tested during this study was 20.7% (224/1081) based on composite bacterial culture and 22.6% (244/1081) based on the Solana™ GAS Assay. All clinical specimens collected during this study were collected between December 2014 and February 2015.

N. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:

Instrument: Solana™ Instrument

O. System Descriptions:

1. Modes of Operation:

The Solana instrument heats each reaction tube to 64°C. If present, the target sequence is amplified by GAS specific primers and detected by a GAS specific fluorescence probe included in the Reaction Tube. The target probes are labeled with a quencher on one end and a fluorophore on the other end. In addition, the target probes carry a ribonucleic acid. Upon annealing to GAS amplicons, the fluorescence probes are cleaved by RNaseH2 and the fluorescence signal increases due to physical separation of fluorophore from quencher. The Solana instrument measures and interprets the fluorescent signal, using on-board method-specific algorithms. Solana instrument will then report the test results to the user on its display screen, and it can print out the results via a printer.

{21}------------------------------------------------

2. Software:

FDA has reviewed applicant's Hazard Analysis and software development processes for this line of product types:

Yes __________________________________________________________________________________________________________________________________________________________________________ No ___________________________________________________________________________________________________________________________________________________________________________

P. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10, 21 CFR 801.109, and the special controls.

§ 866.2680

Streptococcus spp. nucleic acid-based assay.(a)
Identification. AStreptococcus spp. nucleic acid-based assay is a qualitative in vitro diagnostic device intended to simultaneously detect and identify variousStreptococcus spp. nucleic acids extracted directly from clinical specimens. The device detects specific nucleic acid sequences for organism identification. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genusStreptococcus and provides epidemiological information on these diseases. Pathogenic streptococci are associated with infections, such as sore throat, impetigo (an infection characterized by small pustules on the skin), urinary tract infections, rheumatic fever, and kidney disease.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Premarket notification submissions must include detailed device description documentation, including the device components, ancillary reagents required but not provided, and a detailed explanation of the methodology including primer/probe sequence, design, and rationale for sequence selection.
(2) Premarket notification submissions must include detailed documentation from the following analytical and clinical performance studies: Analytical sensitivity (Limit of Detection), reactivity, inclusivity, precision, reproducibility, interference, cross reactivity, carry-over, and cross contamination.
(3) Premarket notification submissions must include detailed documentation from a clinical study. The study, performed on a study population consistent with the intended use population, must compare the device performance to results obtained from well-accepted reference methods.
(4) Premarket notification submissions must include detailed documentation for device software, including, but not limited to, software applications and hardware-based devices that incorporate software.
(5) Premarket notification submissions must include database implementation methodology, construction parameters, and quality assurance protocols, as appropriate.
(6) The device labeling must include limitations regarding the need for culture confirmation of negative specimens, as appropriate.
(7) A detailed explanation of the interpretation of results and acceptance criteria must be included in the device's 21 CFR 809.10(b)(9) compliant labeling.
(8) Premarket notification submissions must include details on an end user device training program that will be offered while marketing the device, as appropriate.