K Number
K150855

Validate with FDA (Live)

Date Cleared
2015-10-21

(204 days)

Product Code
Regulation Number
888.3353
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The H-MAX S Femoral Hip System consists of the H-MAX S femoral stem, a modular femoral head, and an acetabular cup. In total hip arthroplasty, the H-Max S Stems are intended for use with modular CoCrMo femoral heads or Biolox Delta ceramic femoral heads and a compatible Limacorporate Cemented Cup or uncemented Delta TT acetabular cup.

Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis
  • and hip dysplasia;
  • rheumatoid arthritis;
  • osteoarthritis after femoral heads fractures;
  • revisions in cases of good remaining femoral bone stock.

The Limacorporate Cemented Cup is intended for cemented use in hip arthroplasty where the acetabular socket needs reconstruction. The Limacorporate Delta TT Cup is intended for uncemented use.

Device Description

The H-MAX S Femoral Hip System consists of the H-MAX S femoral stem, a modular femoral head, and an acetabular cup. In total hip arthroplasty, the H-Max S Stems are intended for use with modular CoCrMo femoral heads or Biolox Delta ceramic femoral heads and a compatible Limacorporate Cemented Cup or uncemented Delta TT acetabular cup.

The H-MAX S femoral stem is a monolithic, cementless stem made from Ti6Al4V (ISO 5832-3. ASTM F1472). The external surface has a macro-roughened surface and a layer of hydroxyapatite is applied along the length of the stem. The H-MAX S stem is manufactured in 11 sizes for each of two configurations, standard and lateralizing, that vary in the CCD angle.

The Biolox Delta ceramic head is a modular femoral ceramic head with a female 12/14 taper for coupling to the male 12/14 taper of the H-Max S femoral stem. The Biolox Delta ceramic heads are available in diameters of 28, 32, and 36 mm in three (3) neck lengths (S, M, L); the 36 mm diameter head is also available with a XL neck length.

The Delta TT Cup is a hemi-spherical, 2-piece modular cup consisting of a metal shell and a polyethylene liner. The metal shell is manufactured using an EBM (Electron Beam Melting) process with titanium alloy powder (Ti6Al4V, ASTM F1472 - ISO 5832-3) producing a non-porous bulk interior surface with a Trabecular Titanium porous structure on the external surface. A polar threaded hole is used for attachment of an introducer instrument; the hole mates with the polar peg of the polyethylene liner when the liner is inserted into the cup. Three threaded holes in lateral positions allow additional fixation of the cup using bone screws; polyethylene plugs are used to fill the holes when bone screws are not required. The Delta TT Cup metal shell is coupled with a liner made of either conventional or cross-linked ultra high molecular weight polyethylene (UHMWPE) by means of a taper coupling. The Delta TT Cup metal shell is available in external diameters of 44, 46 and 48 mm, for use with Small liners; 50 and 52 mm, for use with Medium liners; and 54, 56, 58, 60, 62, 64, 66, 68, 70, 72, 74 and 76mm mm for use with Large liners. The polyethylene liners are available in three (3) sizes, Small, Medium and Large, with inner diameters of 28, 32, and 36mm.

Femoral head devices are used by surgeons to replace the head of the femur during total hip surgery. They are characterized by a spherical shape and are coupled with the acetabular cup inserted in the acetabulum.

The femoral heads are coupled with the H-Max S stems by means of a 12/14 Morse taper. Modular heads are made of CoCrMo (ASTM F1537) and are available in diameters of 28, 32 and 36 mm and in S. M. L and XL offsets.

Cemented cups are used to completely replace the acetabular cavity, using cement as the fixation agent. Cemented acetabular cups are manufactured entirely from UHMWPE (ASTM F648, ISO 5834-2). An AISI 316-L stainless steel ring is inserted in the equatorial region to allow the evaluation of the device position through radiography.

The acetabular cups are available in two different versions:

  • standard
  • protruded

The standard version is completely hemispherical and is designed for use in deep acetabula. The protruded version has a lateral portion which provides greater coverage of the femoral head to resist dislocation. The cemented cups are available with articular diameters of 28 and 32 mm.

AI/ML Overview

The provided document is a 510(k) premarket notification letter from the FDA regarding a medical device, the H-MAX S Femoral Hip System. This type of document focuses on establishing substantial equivalence to existing legally marketed devices rather than presenting detailed clinical trial data for acceptance criteria.

Therefore, the information typically required to fill out the requested table regarding acceptance criteria and the study that proves the device meets them, including sample sizes, ground truth establishment, expert qualifications, and MRMC studies, is not present in this document.

Here's what can be inferred or stated based on the document's content:

1. A table of acceptance criteria and the reported device performance:

Acceptance CriteriaReported Device Performance
Not explicitly stated as quantifiable acceptance criteria. The document focuses on demonstrating substantial equivalence to predicate devices rather than meeting specific performance metrics from a new clinical study.The device (H-MAX S Femoral Hip System) is described as being identical in certain components (femoral stems, specific modular heads) or having identical design, material, and sizes to components already cleared by the FDA in predicate devices.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not applicable. No new test set data from a clinical or independent performance study is presented. The substantial equivalence relies on comparison to predicate devices and their existing clearances.
  • Data Provenance: Not applicable in the context of new device performance data. The "data" here refers to pre-existing information and clearance of predicate devices, which would have their own provenance from previous submissions.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. No new test set requiring expert ground truth establishment is described. The FDA's review process involves internal experts, but the document does not detail this.

4. Adjudication method for the test set:

  • Not applicable. No new test set requiring adjudication is described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No. This device is a mechanical implant (hip system), not an AI-powered diagnostic or assistive technology. Therefore, an MRMC study or evaluation of human reader improvement with AI assistance is entirely irrelevant and was not performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • No. Similar to point 5, this is a mechanical device, not an algorithm. Standalone algorithm performance is not applicable.

7. The type of ground truth used:

  • Not applicable for new device performance in this context. The "ground truth" for substantial equivalence is effectively the established safety and effectiveness of the identified predicate devices, as determined by the FDA in their prior clearances.

8. The sample size for the training set:

  • Not applicable. No training set is mentioned as this is not an AI/machine learning device.

9. How the ground truth for the training set was established:

  • Not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of three stylized human profiles facing to the right, representing the department's focus on health and human well-being. The profiles are depicted in a simple, abstract manner.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 21, 2015

Limacorporate S.P.A. Stephen Peoples Principal Consultant Via Nazionale 52 33038 Villanova Di San Daniele Udine-Italy

Re: K150855

Trade/Device Name: H-MAX S Femoral Hip System Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: MEH, JDI, KWY, LPH Dated: September 15, 2015 Received: September 22, 2015

Dear Mr. Peoples:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

Page 2 - Ms. Jayana Kenana

(21 CFR Part 807); labeling (21 CFR Part 801 ); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

H-MAX S Femoral Hip System Indications for Use

The H-MAX S Femoral Hip System consists of the H-MAX S femoral stem, a modular femoral head, and an acetabular cup. In total hip arthroplasty, the H-Max S Stems are intended for use with modular CoCrMo femoral heads or Biolox Delta ceramic femoral heads and a compatible Limacorporate Cemented Cup or uncemented Delta TT acetabular cup.

Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis ●
  • and hip dysplasia; ●
  • rheumatoid arthritis; ●
  • osteoarthritis after femoral heads fractures;
  • revisions in cases of good remaining femoral bone stock. ●

The Limacorporate Cemented Cup is intended for cemented use in hip arthroplasty where the acetabular socket needs reconstruction. The Limacorporate Delta TT Cup is intended for uncemented use.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

{3}------------------------------------------------

Date: October 19, 2015

Manufacturer:U.S. Contact Person:
Limacorporate S.p.A.Dr. Stephen J. Peoples
Via Nazionale, 52Principal Consultant
33038 – Villanova di San DanielePhone: 260-645-0327
Udine - ItalyFAX: +39 0432945512
ProductProduct CodeRegulation and Classification Name
H-Max S femoral hipsystemMEHHip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncementedprosthesis per 21 CFR 888.3353
JDIHip joint metal/polymer semi-constrainedcemented prosthesis per 21 CFR 888.3350
KWYHip joint femoral (hemi-hip) metal/polymercemented or uncemented prosthesis per 21 CFR888.3390
LPHProsthesis, Hip, Semi-Constrained,Metal/Polymer, Porous Uncemented per 21CFR 888.3358

Description

The H-MAX S Femoral Hip System consists of the H-MAX S femoral stem, a modular femoral head, and an acetabular cup. In total hip arthroplasty, the H-Max S Stems are intended for use with modular CoCrMo femoral heads or Biolox Delta ceramic femoral heads and a compatible Limacorporate Cemented Cup or uncemented Delta TT acetabular cup.

The H-MAX S femoral stem is a monolithic, cementless stem made from Ti6Al4V (ISO 5832-3. ASTM F1472). The external surface has a macro-roughened surface and a layer of hydroxyapatite is applied along the length of the stem. The H-MAX S stem is manufactured in 11 sizes for each of two configurations, standard and lateralizing, that vary in the CCD angle.

The Biolox Delta ceramic head is a modular femoral ceramic head with a female 12/14 taper for coupling to the male 12/14 taper of the H-Max S femoral stem. The Biolox Delta ceramic heads are available in diameters of 28, 32, and 36 mm in three (3) neck lengths (S, M, L); the 36 mm

{4}------------------------------------------------

diameter head is also available with a XL neck length.

The Delta TT Cup is a hemi-spherical, 2-piece modular cup consisting of a metal shell and a polyethylene liner. The metal shell is manufactured using an EBM (Electron Beam Melting) process with titanium alloy powder (Ti6Al4V, ASTM F1472 - ISO 5832-3) producing a non-porous bulk interior surface with a Trabecular Titanium porous structure on the external surface. A polar threaded hole is used for attachment of an introducer instrument; the hole mates with the polar peg of the polyethylene liner when the liner is inserted into the cup. Three threaded holes in lateral positions allow additional fixation of the cup using bone screws; polyethylene plugs are used to fill the holes when bone screws are not required. The Delta TT Cup metal shell is coupled with a liner made of either conventional or cross-linked ultra high molecular weight polyethylene (UHMWPE) by means of a taper coupling. The Delta TT Cup metal shell is available in external diameters of 44, 46 and 48 mm, for use with Small liners; 50 and 52 mm, for use with Medium liners; and 54, 56, 58, 60, 62, 64, 66, 68, 70, 72, 74 and 76mm mm for use with Large liners. The polyethylene liners are available in three (3) sizes, Small, Medium and Large, with inner diameters of 28, 32, and 36mm.

CoCrMo Femoral heads

Femoral head devices are used by surgeons to replace the head of the femur during total hip surgery. They are characterized by a spherical shape and are coupled with the acetabular cup inserted in the acetabulum.

The femoral heads are coupled with the H-Max S stems by means of a 12/14 Morse taper. Modular heads are made of CoCrMo (ASTM F1537) and are available in diameters of 28, 32 and 36 mm and in S. M. L and XL offsets.

Cemented cups

Cemented cups are used to completely replace the acetabular cavity, using cement as the fixation agent. Cemented acetabular cups are manufactured entirely from UHMWPE (ASTM F648, ISO 5834-2). An AISI 316-L stainless steel ring is inserted in the equatorial region to allow the evaluation of the device position through radiography.

The acetabular cups are available in two different versions:

  • · standard
  • · protruded

The standard version is completely hemispherical and is designed for use in deep acetabula. The protruded version has a lateral portion which provides greater coverage of the femoral head to resist dislocation. The cemented cups are available with articular diameters of 28 and 32 mm.

The following table shows the availability of the cemented cups diameters for each type of cup.

{5}------------------------------------------------

Cup TypeInner diameterRange of Outer Diameter
Standard Cups28 mm44-58 mm with increments of 2 mm
32 mm46-58 mm with increments of 2 mm
Protruded Cups28 mm44-58 mm with increments of 2 mm
32 mm46-58 mm with increments of 2 mm

Intended Use/Indications

The H-MAX S Femoral Hip System consists of the H-MAX S femoral stem, a modular femoral head, and an acetabular cup. In total hip arthroplasty, the H-Max S Stems are intended for use with modular CoCrMo femoral heads or Biolox Delta ceramic femoral heads and a compatible Limacorporate Cemented Cup or uncemented Delta TT acetabular cup.

Hip arthroplasty is intended for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

  • non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis
  • and hip dysplasia;
  • rheumatoid arthritis;
  • osteoarthritis after femoral heads fractures; ●
  • revisions in cases of good remaining femoral bone stock. ●

The Limacorporate Cemented Cup is intended for cemented use in hip arthroplasty where the acetabular socket needs reconstruction. The Limacorporate Delta TT Cup is intended for uncemented use.

Predicate Devices

  • H-MAX S Femoral Hip System (Limacorporate, K112091)
  • Minima S System (Limacorporate, K141327)
  • Master SL Femoral Stem (Limacorporate, K143509) 0
  • Modulus Femoral Stem (Limacorporate, K112158) ●
  • Delta TT Acetabular System (Limacorporate, K112898, K141395). 0

Comparable Features to Predicate Device(s)

The subject H-MAX S Femoral Hip Stems are identical to the H-MAX S Femoral Hip Stems cleared in K112091. In addition:

  • The subject CoCrMo modular heads (36 mm) are identical to CoCrMo modular heads (36 ●

{6}------------------------------------------------

mm) approved for use with Master SL femoral stem (K143509) and with Minima S femoral stem (K141327) in combination with Delta TT Acetabular cups;

  • . The subject Biolox Delta heads (36 mm) are identical to the Biolox Delta Heads cleared in K141327 (Minima S) and K143509 (Master SL) in combination with Delta TT Acetabular cups;
  • The subject CoCrMo heads (28, 32 mm) are identical to CoCrMo heads (28, 32 mm) ● cleared in K141327 (Minima S) and K143509 (Master SL) in combination with Delta TT Acetabular cups;
  • The subject Biolox Delta heads (28, 32 mm) are identical to the Biolox Delta Heads ● cleared in K141327 (Minima S) and K143509 (Master SL) in combination with Cemented cups or Delta TT Acetabular cups.

Non-Clinical Testing:

The materials, design, femoral head sizes and acetabular components remain the same as described in previously cleared submissions and no new issues of safety and effectiveness are introduced by adding:

  • CoCrMo 0 36 mm modular heads coupled with Delta TT system for use with the Ho MAX S Femoral Hip System;
  • Biolox Delta Ø 36 mm modular heads coupled with Delta TT system for use with the Ho MAX S Femoral Hip System;
  • CoCrMo 0 28, 32 mm modular heads coupled with Delta TT system for use with the H-. MAX S Femoral Hip System.
  • Biolox Delta Ø 28, 32 mm modular heads coupled with Delta TT system and Cemented ● cups for use with the H-MAX S Femoral Hip System.

The tests provided in previously cleared submissions are therefore applicable to the new couplings proposed to be used with the H-MAX S Femoral Hip Stem.

In detail:

  • Coupling of H-MAX S Femoral Hip system with CoCrMo modular heads (36 mm) and ● Delta TT acetabular cups : the design, material, and sizes of the CoCrMo modular heads (36 mm) articulating with compatible Delta TT acetabular system are identical to the design, material, and sizes of CoCrMo modular heads (36 mm) cleared for use with the Minima S femoral stem (K141327) and the Master SL femoral stem (K143509) articulating with Delta TT acetabular system.
  • Coupling of H-MAX S Femoral Hip system with Biolox Delta heads (36 mm) and Delta TT acetabular cups : the design, material, and sizes of the Biolox Delta Femoral Heads for use with the H-MAX S Femoral Hip System Stems are identical to the design, material, and

{7}------------------------------------------------

sizes of the Biolox Delta Femoral Heads cleared for use with the Minima S Hip System (K141327) and Master SL femoral stem (K143509) articulating with Delta TT acetabular system. The coupling between the Biolox Delta Head and the H-MAX S Femoral Hip Stem is achieved through the same taper used to couple the Biolox Delta ceramic head to the Minima Hip Stem (K141327) and Master SL femoral stem (K143509). The H-MAX S Femoral Hip Stem (K112091) is manufactured from the same material (Ti6Al4V allov) as the Minima S Hip Stem and Master SL femoral stem cleared for use with Biolox Delta ceramic heads ( in K141327 and K143509 respectively).

  • Coupling of H-MAX S Femoral Hip system with CoCrMo heads (28, 32 mm) and Delta TT ● acetabular cups : CoCrMo modular heads sizes 28 and 32 in offset S, M, L, XL are identical in design, material and sizes to the ones cleared in the original 510(k) K112091, for the use with H-MAX S stem and cemented cups. These devices are coupled with cemented cups in the original submission of H-MAX S (K112091): the coupling of these heads with Delta TT cups has already been cleared with Master SL (K140975) and Minima S (K141327) stems.
  • Coupling of H-MAX S Femoral Hip system with Biolox Delta heads (28, 32 mm) and Cemented cups / Delta TT acetabular cups : the design, material, and sizes of the Biolox Delta Femoral Heads for use with the H-MAX S Femoral Hip System Stems are identical to the design, material, and sizes of the Biolox Delta Femoral Heads cleared for use with the Minima S Hip System (K141327) and Master SL femoral stem (K143509) articulating with Cemented cups / Delta TT acetabular system. The coupling between the Biolox Delta Head and the H-MAX S Femoral Hip Stem is achieved through the same taper used to couple the Biolox Delta ceramic head to the Minima Hip Stem (K141327) and Master SL femoral stem (K143509). The H-MAX S Femoral Hip Stem (K112091) is manufactured from the same material (Ti6Al4V alloy) as the Minima S Hip Stem and Master SL femoral stem cleared for use with Biolox Delta ceramic heads ( in K141327 and K143509 respectively).

Clinical Testing: Clinical testing was not necessary to demonstrate substantial equivalence.

§ 888.3353 Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis.

(a)
Identification. A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.(b)
Classification. Class II.