K Number
K150791

Validate with FDA (Live)

Date Cleared
2015-04-24

(30 days)

Product Code
Regulation Number
862.3150
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Healgen Secobarbital Test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test may yield preliminary positive results even when the prescription drug Secobarbital is ingescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Buprenorphine Test is an immunochromatographic assay for the qualitative determination of Buprenorplaine in human urine at a Cut-Off concentration of 10 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test may yield preliminary positive results even when the prescription drug Burrenorphine is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Buprenorphine in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Methadone Test is an immunochromatographic assay for the qualitative determination of Methadone in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Device Description

Healgen Secobarbital Test. Healgen Buprenorphine Test and Healgen Methadone Test are immunochromatographic assays for Secobarbital, Buprenorphine and Methadone. Each assay test is a lateral flow system for the qualitative detection of Secobarbital, Buprenorphine and Methadone (target analyte) in human urine. The products are in vitro diagnostic devices, which come in the form of: Strips, Cassettes, DipCards, or Cups. Each product contains a Test Device (in one of the four formats), and a package insert. Each test device is sealed with a desiccant in an aluminum pouch.

AI/ML Overview

Here's an analysis of the acceptance criteria and study findings for the Healgen Secobarbital, Buprenorphine, and Methadone Tests, structured as requested:

Acceptance Criteria and Device Performance

The provided document describes the analytical and comparison studies performed to demonstrate the substantial equivalence of the Healgen drug tests. The acceptance criteria are implied by the expected performance in these studies, particularly the accuracy relative to GC/MS at and around the cutoff concentrations.

Table of Acceptance Criteria (Implied) and Reported Device Performance

Test TypePerformance MetricAcceptance Criteria (Implied)Reported Device Performance (Summary)
PrecisionAccuracy at various concentrations around the cutoffConsistent and correct results (negative below cutoff, positive above) at +/- 50%, +/- 75%, +/- 100% of the cutoff. Mixed results expected only very close to the cutoff (+/- 25%).For all three drugs and all four formats, tests performed by three different operators over 25 days showed: - 100% accuracy (50-/0+ or 50+/0-) at -100%, -75%, -50%, +50%, +75%, +100% of cutoff. - Mixed results (e.g., 22-/28+, 25-/25+, 20-/30+) within the +/- 25% range of the cutoff, which is expected for qualitative tests at the decision point.
Cut-off VerificationAbility to distinguish between drug concentrations at and around the cut-offAll samples at or above +25% cut-off should be positive; all samples at or below -25% cut-off should be negative.For all three drugs, across three different device lots, and three different operators: - All results were positive at and above +25% cut-off. - All results were negative at and below -25% cut-off.
InterferenceAbsence of interference from common urine substancesNo interference at 100µg/mL for listed substances.No differences observed across different formats for a comprehensive list of compounds at 100µg/mL for all three drugs.
SpecificityCross-reactivity with related compounds and non-reactivity with unrelated compounds.Acceptable levels of cross-reactivity for structurally similar compounds and no detection for unrelated compounds (where applicable).- Secobarbital: 100% cross-reactivity with Amobarbital; varying degrees with Alphenol (40%), Aprobarbital (120%), Butabarbital (12%), Butathal (12%), Butalbital (12%), Cyclopentobarbital (60%), Pentobarbital (12%), Phenobarbital (1%). - Buprenorphine: 100% with Buprenorphine -3-D-Glucuronide; 50% with Norbuprenorphine and Norbuprenorphine-3-D-Glucuronide. No detection of Morphine, Oxymorphone, Hydromorphone at 100,000 ng/mL. - Methadone: 6% cross-reactivity with Doxylamine. No detection of EDDP, EMDP, LAAM HCI, Alpha Methadol at 100,000 ng/mL.
Effect of Urine Specific Gravity and pHConsistent results across physiological ranges of specific gravity and pH.All samples at or above +25% cut-off should be positive; all samples at or below -25% cut-off should be negative.For all three drugs, across various specific gravity (1.000-1.035) and pH (4-9) levels: - All results were positive for samples at and above +25% cut-off. - All results were negative for samples at and below -25% cut-off.
Method Comparison (Clinical Samples)Agreement with GC/MS results for negative, low negative, near cutoff negative/positive, and high positive samples.High agreement expected, especially for definitive negative and high positive samples. Discordant results are noted and acceptable near the cutoff.Across all three drugs and all four formats, for each of three viewers: - 100% agreement for negative, low negative, and high positive samples. - Discordant results (some positives read as negative, or vice versa) observed mainly in the "Near Cutoff Negative" and "Near Cutoff Positive" categories, which is expected for qualitative tests near their decision point. Specific discordant sample numbers and GC/MS values (e.g., 302 ng/mL, 303 ng/mL for Secobarbital, 10.2 ng/mL for Buprenorphine with a 10 ng/mL cutoff) are provided, showing the device performs qualitatively as expected around the cutoff.
Lay-User StudyHigh percentage of correct results by untrained users, ease of understanding instructions.High accuracy across various concentrations, especially for definitive positive/negative samples. Instructions should be easily understood.For all three drugs and all four formats, performed by over 1680 lay persons (560 per drug): - 90-100% correct results for samples at -25% Cutoff, +25% Cutoff. - 100% correct results for samples at -100%, -75%, -50% Cutoff, and +50%, +75% Cutoff. - Package insert instructions were easily followed, and Flesch-Kincaid read-out at Grade Level 7.

Study Details

  1. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

    • Precision Studies:
      • For each of the three drugs (Secobarbital, Buprenorphine, Methadone) and each of the four device formats (Strip, Cassette, Dip Card, Cup): 9 drug concentrations (including the cutoff, +/-25%, +/-50%, +/-75%, +/-100%) were tested with 50 replicates per concentration per lot.
      • This means (9 concentrations * 50 replicates * 3 lots) = 1350 tests per format.
      • Total precision tests per drug: 1350 tests/format * 4 formats = 5400 tests.
      • Total precision tests for all three drugs: 5400 tests * 3 drugs = 16,200 individual tests.
      • Data Provenance: Not explicitly stated, but the study was "carried out for samples... prepared by spiking drug in negative samples." This suggests a controlled laboratory setting, likely prospective. The country of origin is not mentioned.
    • Cut-off Verification Studies:
      • For each drug: 150 samples (equally distributed across -50%, -25%, cut-off, +25%, +50% cut-off) per drug. These were tested using three different lots of each device by three different operators.
      • Total samples per drug: 150.
      • Total cut-off verification tests for all three drugs: 150 * 3 drugs = 450 samples.
      • Data Provenance: Controlled laboratory setting, prepared by spiking drugs into negative samples; likely prospective. Country of origin not mentioned.
    • Interference Studies:
      • Test Set Description: Drug-free urine and target drug urine (at 25% above cut-off) spiked with various potential interfering substances at 100 µg/mL.
      • Sample Size: Not explicitly stated as a number of individual samples, but tested "three batches of each device for all formats" against a comprehensive list of compounds.
      • Data Provenance: Controlled laboratory setting; likely prospective. Country of origin not mentioned.
    • Specificity Studies (Cross-Reactivity):
      • Test Set Description: Related drug metabolites and other components.
      • Sample Size: Not explicitly stated as a number of individual samples, but tested "three batches of each device for all formats."
      • Data Provenance: Controlled laboratory setting; likely prospective. Country of origin not mentioned.
    • Effect of Urine Specific Gravity and pH Studies:
      • Test Set Description: Urine samples with specific gravity of 1.000 to 1.035, or pH 4 to 9, spiked with target drugs at 25% below and 25% above cut-off levels.
      • Sample Size: Not explicitly stated but tested "three batches of each device for all formats."
      • Data Provenance: Controlled laboratory setting; likely prospective. Country of origin not mentioned.
    • Method Comparison (Clinical Samples):
      • For each of the three drugs and each of the four device formats: 80 unaltered clinical samples (40 negative and 40 positive).
      • Total clinical samples: 80 samples/format * 4 formats * 3 drugs = 960 samples.
      • Each sample was run by three different laboratory assistants.
      • Data Provenance: "unaltered clinical samples," blind labeled. The origin (country, retrospective/prospective) is not specified.
    • Lay-User Study:
      • Sample Size: 560 lay persons for each drug test (Secobarbital, Buprenorphine, Methadone). Total individuals performing the study: 1680.
      • For each drug, urine samples were prepared at 7 different concentrations per drug: negative, +/-25%, +/-50%, +/-75%, +/-100% of the cutoff. Each concentration had 20 samples. (7 concentrations * 20 samples = 140 samples per drug tested by lay users).
      • Data Provenance: Conducted at "three intended user sites." The demographic breakdown (age, gender, educational/professional backgrounds) is provided for the participants. Likely a prospective study. Country of origin not mentioned.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

    • The ground truth for all analytical and comparison studies was established by GC/MS (Gas Chromatography/Mass Spectrometry). This is a highly accurate and widely accepted "gold standard" analytical method for drug detection and quantification in urine.
    • The document implies that the GC/MS results themselves serve as the ground truth, rather than human experts interpreting the GC/MS. The qualifications of the individuals operating the GC/MS are not specified, but they would typically be trained laboratory technicians or chemists.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set

    • For the Method Comparison (Clinical Samples), the samples were "blind labeled and compared to GC/MS results." The results from the three "Viewer" (laboratory assistants) were individually recorded and compared to the GC/MS. There is no explicit "adjudication method" described where multiple human readers' opinions are combined to form a consensus before comparison to ground truth. Instead, individual human readings (referred to as "Viewer A, B, C") were directly compared to the GC/MS ground truth.
    • For the Lay-User Study, each participant interpreted their own test and results were compared to GC/MS. No adjudication method among lay users is mentioned.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    • This document describes the performance of immunochromatographic assay devices for drug detection, which are qualitative diagnostic tests, not AI-powered image analysis tools.
    • Therefore, an MRMC comparative effectiveness study comparing human readers with and without AI assistance was NOT performed or applicable for this device type. The "Viewers" mentioned in the method comparison are human operators reading a physical test strip/cassette/cup/dip card.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

    • This device is an immunochromatographic assay. It is a physical test that produces a visual result (colored lines) which is then interpreted by a human. There is no "algorithm only" performance component. The device's performance is intrinsically linked to human interpretation of its visual output.
    • The closest to "standalone" performance are the analytical studies (precision, cut-off, interference, specificity) where the chemical reaction and visual readout are evaluated, but even then, a human reads the result.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

    • The primary ground truth used for all performance evaluations (precision, cut-off, interference, specificity, method comparison, lay-user study) was GC/MS (Gas Chromatography/Mass Spectrometry). This is a highly accurate analytical measurement of drug concentration.
  7. The sample size for the training set

    • This document describes the performance evaluation of a rapid diagnostic test (immunochromatographic assay). These types of tests are typically developed through chemical and biological engineering, not through machine learning models that require "training sets" in the computational sense.
    • Therefore, there is no "training set" sample size listed in this context.
  8. How the ground truth for the training set was established

    • As there is no "training set" for a machine learning model, this question is not applicable to the described device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized graphic of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 24, 2015

HEALGEN SCIENTIFIC LLC C/O JOE SHIA REGULATORY CONSULTANT 504 EAST DIAMOND AVE. SUITE F GAITHERSBURG MD 20877

Re: K150791

Trade/Device Name: Healgen Secobarbital Test (Strip, Cassette, Cup, Dip Card), Healgen Buprenorphine Test (Strip, Cassette, Cup, Dip Card), Healgen Methadone Test (Strip, Cassette, Cup, Dip Card) Regulation Number: 21 CFR 862.3150 Regulation Name: Barbiturate test system Regulatory Class: II Product Code: DIS, DJR, DJG Dated: March 15, 2015 Received: March 25, 2015

Dear Mr. Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D.

Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

Indications for Use

510(k) Number (if known) K150791

Device Name Healgen Secobarbital Test (Strip, Cassette, Cup, Dip Card) Healgen Buprenorphine Test (Strip, Cassette, Cup, Dip Card) Healgen Methadone Test (Strip, Cassette, Cup, Dip Card)

Indications for Use (Describe)

Healgen Secobarbital Test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test may yield preliminary positive results even when the prescription drug Secobarbital is ingescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Buprenorphine Test is an immunochromatographic assay for the qualitative determination of Buprenorplaine in human urine at a Cut-Off concentration of 10 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test may yield preliminary positive results even when the prescription drug Burrenorphine is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Buprenorphine in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Methadone Test is an immunochromatographic assay for the qualitative determination of Methadone in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive. For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Type of Use (Select one or both, as applicable)
☒ Prescription Use (Part 21 CFR 801 Subpart D) ☒ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.FORM FDA 3881 (8/14)

PSC Publishing Services (301) 443-6740

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

    1. Date: April 21, 2015
    1. Submitter: HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX 77401
    1. Contact person: Jiangiu Fang HEALGEN SCIENTIFIC LLC 5213 Maple St Bellaire, TX 77401 Telephone: 713-733-8088 Fax: 713-733-8088 Email: bryan@healgen.com
    1. Device Name: Healgen Secobarbital Test (Strip, Cassette, Cup, Dip Card) Healgen Buprenorphine Test (Strip, Cassette, Cup, Dip Card) Healgen Methadone Test (Strip, Cassette, Cup, Dip Card)

Classification:

Product CodeCFR #Panel
DIS21 CFR, 862.3150Barbiturate Test SystemToxicology
DJR21 CFR, 862.3620Methadone Test SystemToxicology
DJG21 CFR, 862.3650Opiate Test SystemToxicology
    1. Predicate Devices: K132812 UCP Multi-Drug Test Key Cups
    1. Intended Use / Indications for Use

Healgen Secobarbital Test is an immunochromatographic assay for the qualitative determination of Secobarbital in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test may yield preliminary positive results even when the prescription drug Secobarbital is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Secobarbital in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and

{5}------------------------------------------------

professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Buprenorphine Test is an immunochromatographic assay for the qualitative determination of Buprenorphine in human urine at a Cut-Off concentration of 10 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test may yield preliminary positive results even when the prescription drug Buprenorphine is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for Buprenorphine in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

Healgen Methadone Test is an immunochromatographic assay for the qualitative determination of Methadone in human urine at a Cut-Off concentration of 300 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for prescription and for over-the-counter use.

    1. Device Description
      Healgen Secobarbital Test. Healgen Buprenorphine Test and Healgen Methadone Test are immunochromatographic assays for Secobarbital, Buprenorphine and Methadone. Each assay test is a lateral flow system for the qualitative detection of Secobarbital, Buprenorphine and Methadone (target analyte) in human urine. The products are in vitro diagnostic devices, which come in the form of: Strips, Cassettes, DipCards, or Cups. Each product contains a Test Device (in one of the four formats), and a package insert. Each test device is sealed with a desiccant in an aluminum pouch.

8. Substantial Equivalence Information

A summary comparison of features of the Healgen Secobarbital Test, Healgen Buprenorphine Test and Healgen Methadone Test and the predicate device is provided in Table 1, Table 2 and Table 3.

Table 1: Features Comparison of Healgen Secobarbital Test and the Predicate Device

{6}------------------------------------------------

ItemDevicePredicate – K132812
Intended UseFor the qualitative determination of drugsof abuse in human urine.Same
Drug AnalyteSecobarbitalthe class ofbarbituratesincludingsecobarbital
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values300 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.Same
ConfigurationsStrip, Cassette, Cup, Dip CardCup

Table 2: Features Comparison of Healgen Buprenorphine Test and the Predicate Device

ItemDevicePredicate -K132812
Intended UseFor the qualitative determination ofdrugs of abuse in human urine.Same
Drug AnalyteBuprenorphineSame
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values10 ng/mLSame
IntendedPopulationFor over-the-counter and prescriptionuses.Same
ConfigurationsStrip, Cassette, Cup, Dip CardCup

Table 3: Features Comparison of Healgen Methadone Test and the Predicate Device

{7}------------------------------------------------

ItemDevicePredicate – K132812
Intended UseFor the qualitative determination of drugs of abuse in human urine.Same
Drug AnalyteMethadoneSame
MethodologyCompetitive binding, lateral flow immunochromatographic assays based on the principle of antigen antibody immunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values300 ng/mLSame
Intended PopulationFor over-the-counter and prescription uses.Same
ConfigurationsStrip, Cassette, Cup, Dip CardCup

9. Test Principle

Healgen Secobarbital Test, Healgen Buprenorphine Test and Healgen Methadone Test are rapid tests for the qualitative detection of Secobarbital, Buprenorphine and Methadone in urine samples. Each assay test is a lateral flow chromatographic immunoassay. During testing, a urine specimen migrates upward by capillary action. If target drugs are present in the urine specimen below its cut-off concentration, it will not saturate the binding sites of its specific antibody (monoclonal mouse antibody) coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cut-off concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample to indicate that the test has been performed properly.

10. Performance Characteristics

    1. Analytical Performance
    • a. Precision

Precision studies were carried out for samples with concentrations of -100% cut-off, -50% cut-off, -25% cut-off, at the cut-off, +25% cut-off, +75% cut-off, +75% cut-off and +100% cut-off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blind labeled and randomized by the person who prepared samples and did not take part in the sample testing. For each concentration, tests were performed two runs per day for 25 days by three different operators for each format of devices. Different set of operators tested each format. The results obtained are summarized in the following tables:

{8}------------------------------------------------

Secobarbital

Strip Format

Result
Drug-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot: BAR120100150-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot: BAR120100250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot: BAR120100350-/0+50-/0+50-/0+50-/0+23-/27+50+/0-50+/0-50+/0-50+/0-

Cassette Format

DrugResult-100% Cut-off-75% Cut-off-50% Cut-off-25% Cut-offCut-off+25% Cut-off+50% Cut-off+75% Cut-off+100% Cut-off
Lot: BAR120100450-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot: BAR120100550-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot: BAR120100650-/0+50-/0+50-/0+50-/0+17-/33+50+/0-50+/0-50+/0-50+/0-

Dip Card Format

Result -100% -75%-50% --25%+25%+50% +75% +100%
DrugCut-off Cut-off Cut-off Cut-offCut-offCut-off Cut-off Cut-off Cut-off
Lot: BAR120100750-/0+ 50-/0+ 50-/0+ 50-/0+ 23-/27+ 50+/0- 50+/0- 50+/0-
Lot: BAR120100850-/0+ 50-/0+ 50-/0+ 50-/0+ 21-/29+ 50+/0- 50+/0- 50+/0- 50+/0-
Lot: BAR120100950-/0+ 50-/0+ 50-/0+ 50-/0+ 23-/27+ 50+/0- 50+/0- 50+/0- 50+/0-

CUP Format

Result
Drug-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot: BAR120101050-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-
Lot: BAR120101150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot: BAR120101250-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

Buprenorphine

Strip Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot: BUP120400150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot: BUP120400250-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot: BUP120400350-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-50+/0-

Cassette Format

{9}------------------------------------------------

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot: BUP120400450-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot: BUP120400550-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot: BUP120400650-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-

Dip Card Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot: BUP120400750-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-
Lot: BUP120400850-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot: BUP120400950-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-

CUP Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
DrugLot: BUP120401050-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-50+/0-
Lot: BUP120401150-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-
Lot: BUP120401250-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-

Methadone

Strip Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot: MTD120200150-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-
Lot: MTD120200250-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-
Lot: MTD120200350-/0+50-/0+50-/0+50-/0+26-/24+50+/0-50+/0-50+/0-

Cassette Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot: MTD120200450-/0+50-/0+50-/0+50-/0+19-/31+50+/0-50+/0-50+/0-50+/0-
Lot: MTD120200550-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot: MTD120200650-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

Dip Card Format

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot: MTD120200750-/0+50-/0+50-/0+50-/0+28-/22+50+/0-50+/0-50+/0-50+/0-
Lot: MTD120200850-/0+50-/0+50-/0+50-/0+24-/26+50+/0-50+/0-50+/0-50+/0-

{10}------------------------------------------------

DrugResult-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Lot: MTD120200950-/0+50-/0+50-/0+50-/0+27-/23+50+/0-50+/0-50+/0-

CUP Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot: MTD120201050-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot: MTD120201150-/0+50-/0+50-/0+50-/0+25-/25+50+/0-50+/0-50+/0-50+/0-
Lot: MTD120201250-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

b. Linearity

Not applicable.

  • c. Stability
    The devices are stable at 4-30°C for 24 months based on the accelerated stability study at 45°C and real time stability determination at both 4 ℃ and 30℃.

Control materials are not provided with the device. The labeling provides information on how to obtain control materials.

  • d. Cut-off
    A total of 150 samples equally distributed at concentrations of -50% cut-off; -25% cut-off; cut-off; +25% cut-off; +50% cut-off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% cut-off and all negative at and below -25% cut-off for Secobarbital, Buprenorphine and Methadone. The following cut-off values for the test devices have been verified.
TestCalibratorCut-off(ng/mL)
Secobarbital TestSecobarbital300
Buprenorphine TestBuprenorphine10
Methadone TestMethadone300

e. Interference

Potential interfering substances found in human urine of physiological or pathological conditions were added to drug-free urine and target drugs urine with concentration at 25% above cut-off levels. These urine samples were tested using three batches of each device for all formats.

{11}------------------------------------------------

Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different formats.

AcetophenetidinEthyl-p-aminobenzoatePhencyclidine
N-AcetylprocainamideFenoprofenPhenelzine
Acetylsalicylic Acid(Aspirin)FurosemidePhentermine
AminopyrineGentisic acidPhenylephrine-L
AmitriptylineHemoglobinPhenylethylamine
AmoxicillinHydralazinePhenylpropanolamine
D-Amphetamine(+/-)-4-HydroxyamphetamineHCLPrednisolone Acetate
L-AmphetamineHydrochlorothiazidePrednisone
Amphetamine SulfateHydrocodoneProcaine(Novocaine)
Ampicinine(Ampicillin)HydrocortisonePromazine
Apomorphinea -Hydroxyhippuric acidPromethazine
L-Ascorbic Acidp-HydroxymethamphetaminePropoxyphene,d-
AspartameIbuprofenPropranolol
AtropineImipraminePseudoephedrine HCL
Benzilic acidIsoxsuprineQuinidine
BenzphetamineIsoproterenol-(+/-)Quinine
Bezoic AcidKetamineRanitidine(Zantac)
BilirubinLabetalolSalicylic Acid
CaffeineLevorphanolSecobarbital
ChloramphenicolLoperamideSerotonin
ChlordiazepoxideHCLMaprotilineSulfamethazine
ChloroquineMeprobamateSulindac
ChlorothiazideMethadoneTemazepam
ChlorpheniramineMethoxyphenamine11-Nor-△9-Tetrahydrocannabinol
Chlorpromazine(+/-)-Methylenedioxyamphetamine(MDA)Tetracycline
CholesterolMethylphenidateTetrahydrozoline
ClomipramineNalbuphineThiamine
Clonidine hydrochlorideNalidixic acidL-Thyroxine
CortisoneNaloxone hydrochlorideThioridazineHydrochloride
Cotinine(-)Naltrexone hydrochlorideTriamterene
CreatinineNaproxenTriflupromazineHydrochloride

Secobarbital

{12}------------------------------------------------

DeoxyepinephrineNiacinamideTrimethoprim
DextromethorphanNifedipineTrimipramine
DiazepamNorethindroneTryptamine
DiflunisalNorpropoxypheneDL-Tryptophan
DigoxinNoscapineTyramine
DoxylamineOxazepamD/L-Tyrosine
Ecgonine methylesterOxymetazolineUric Acid
R(-)-EpinephrinePapaverineVerapamil
ErythromycinPenicillinZomepirac
Estrone-3-sulfatePerphenazine

Buprenorphine

AcetophenetidinEthyl-p-aminobenzoatePhencyclidine
N-AcetylprocainamideFenoprofenPhenelzine
Acetylsalicylic Acid(Aspirin)FurosemidePhenobarbital
AminopyrineGentisic acidPhentermine
AmitriptylineHemoglobinPhenylephrine-L
AmoxicillinHydralazinePhenylethylamine
Amobarbital(+/-)-4-HydroxyamphetamineHCLPhenylpropanolamine
D-AmphetamineHydrochlorothiazidePrednisolone Acetate
L-AmphetamineHydrocodonePrednisone
Amphetamine SulfateHydrocortisoneProcaine(Novocaine)
Ampicinine(Ampicillin)a -Hydroxyhippuric acidPromazine
Apomorphinep-HydroxymethamphetaminePromethazine
L-Ascorbic AcidIbuprofenPropoxyphene,d-
AspartameImipraminePropranolol
AtropineIsoxsuprinePseudoephedrine HCL
Benzilic acidIsoproterenol-(+/-)Quinidine
BenzphetamineKetamineQuinine
Bezoic AcidLabetalolRanitidine(Zantac)
BilirubinLevorphanolSalicylic Acid
CaffeineLoperamideSecobarbital
ChloramphenicolMaprotilineSerotonin
Chlordiazepoxide HCLMeprobamateSulfamethazine
ChloroquineMethadoneSulindac
ChlorothiazideMethoxyphenamineTemazepam
Chlorpheniramine(+/-)-Methylenedioxyamphetamine (MDA)11-Nor-△9-Tetrahydrocannabinol

{13}------------------------------------------------

ChlorpromazineMethylphenidateTetracycline
CholesterolNalbuphineTetrahydrozoline
ClomipramineNalidixic acidThiamine
Clonidine hydrochlorideNaloxone hydrochlorideL-Thyroxine
CortisoneNaltrexone hydrochlorideThioridazineHydrochloride
Cotinine(-)NaproxenTriamterene
CreatinineNiacinamideTriflupromazine
Hydrochloride
DeoxyepinephrineNifedipineTrimethoprim
DextromethorphanNorethindroneTrimipramine
DiazepamNorpropoxypheneTryptamine
DiflunisalNoscapineDL-Tryptophan
DigoxinOxazepamTyramine
DoxylamineOxymetazolineD/L-Tyrosine
Ecgonine methylesterPapaverineUric Acid
R(-)-EpinephrinePenicillinVerapamil
ErythromycinPentobarbitalZomepirac
Estrone-3-sulfatePerphenazine

Methadone

Methadone
AcetophenetidinEthyl-p-aminobenzoatePhenelzine
N-AcetylprocainamideFenoprofenPhenobarbital
Acetylsalicylic Acid(Aspirin)FurosemidePhentermine
AminopyrineGentisic acidPhenylephrine-L
AmitriptylineHemoglobinPhenylethylamine
AmoxicillinHydralazinePhenylpropanolamine
Amobarbital(+/-)-4-HydroxyamphetamineHCLPrednisolone Acetate
D-AmphetamineHydrochlorothiazidePrednisone
L-AmphetamineHydrocodoneProcaine(Novocaine)
Amphetamine SulfateHydrocortisonePromazine
Ampicinine(Ampicillin)a -Hydroxyhippuric acidPromethazine
Apomorphinep-HydroxymethamphetaminePropoxyphene,d-
L-Ascorbic AcidIbuprofenPropranolol
AspartameImipraminePseudoephedrine HCL
AtropineIsoxsuprineQuinidine
Benzilic acidIsoproterenol-(+/-)Quinine
BenzphetamineKetamineRanitidine(Zantac)

{14}------------------------------------------------

Bezoic AcidLabetalolSalicylic Acid
BilirubinLevorphanolSecobarbital
CaffeineLoperamideSerotonin
ChloramphenicolMaprotilineSulfamethazine
Chlordiazepoxide HCLMeprobamateSulindac
ChloroquineMethoxyphenamineTemazepam
Chlorothiazide(+/-)-Methylenedioxyamphetamine(MDA)11-Nor-△9-Tetrahydrocannabinol
ChlorpheniramineMethylphenidateTetracycline
ChlorpromazineNalbuphineTetrahydrozoline
CholesterolNalidixic acidThiamine
ClomipramineNaloxone hydrochlorideL-Thyroxine
Clonidine hydrochlorideNaltrexone hydrochlorideThioridazineHydrochloride
CortisoneNaproxenTriamterene
Cotinine(-)NiacinamideTriflupromazineHydrochloride
CreatinineNifedipineTrimethoprim
DeoxyepinephrineNorethindroneTrimipramine
DextromethorphanNorpropoxypheneTryptamine
DiazepamNoscapineDL-Tryptophan
DiflunisalOxazepamTyramine
DigoxinOxymetazolineD/L-Tyrosine
DoxylaminePapaverineUric Acid
Ecgonine methylesterPenicillinVerapamil
R(-)-EpinephrinePentobarbitalZomepirac
ErythromycinPerphenazine
Estrone-3-sulfatePhencyclidine
  • f. Specificity
    To test the specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device for all formats. The obtained lowest detectable concentration was used to calculate the cross-reactivity. There were no differences observed for different formats.
Secobarbital(Cut-off=300 ng/mL)Result% Cross-Reactivity
SecobarbitalPositive at 300 ng/mL100%
AmobarbitalPositive at 300 ng/mL100%

{15}------------------------------------------------

AlphenolPositive at 750 ng/mL40%
AprobarbitalPositive at 250 ng/mL120%
ButabarbitalPositive at 2500 ng/mL12%
ButathalPositive at 2500 ng/mL12%
ButalbitalPositive at 2500 ng/mL12%
CyclopentobarbitalPositive at 500 ng/mL60%
PentobarbitalPositive at 2500 ng/mL12%
PhenobarbitalPositive at 25000 ng/mL1%
Buprenorphine(Cut-off=10 ng/mL)Result% Cross-Reactivity
BuprenorphinePositive at 10 ng/mL100%
Buprenorphine -3-D-GlucuronidePositive at 10 ng/mL100%
NorbuprenorphinePositive at 20 ng/mL50%
Norbuprenorphine-3-D-GlucuronidePositive at 20 ng/mL50%
MorphineNegative at 100000Not detected
OxymorphoneNegative at 100000Not detected
HydromorphoneNegative at 100000Not detected
Methadone(Cut-off=300 ng/mL)Result% Cross-Reactivity
MethadonePositive at 300 ng/mL100%
DoxylaminePositive at 5000 ng/mL6%
EDDPNegative at 100000Not detected
EMDPNegative at 100000Not detected
LAAM HCINegative at 100000Not detected
Alpha MethadolNegative at 100000Not detected
  • g. Effect of Urine Specific Gravity and Urine pH
    To investigate the effect of urine specific gravity and urine pH, urine samples with of 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above cut-off levels. These samples were tested using three batches of each device for all formats. Results were all positive for samples at and above +25% cut-off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different formats.

2. Comparison Studies

The method comparison studies for the Secobarbital Test, the Buprenorphine Test and the Methadone Test were performed in-house with three different laboratory assistants for each format

{16}------------------------------------------------

of the device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to GC/MS results. The results are presented in the tables below:

Secobarbital
StripformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)NearCutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan +50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001524
Negative10151510

Discordant Results of Secobarbital Strip

ViewerSample NumberGC/MS ResultStrip FormatViewer Results
Viewer ABAR06303Negative
Viewer ABAR72302Negative
Viewer BBAR06303Negative
Viewer BBAR72302Negative
Viewer CBAR06303Negative
CassetteformatNegativeLowNegativeby GC/MS(less than-50%)NearCutoffNegativeby GC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001324
Negative10151530
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001324
Negative10151530

Discordant Results of Secobarbital Cassette

{17}------------------------------------------------

ViewerSample NumberGC/MS ResultCassette FormatViewer Results
Viewer ABAR06303Negative
Viewer ABAR16308Negative
Viewer ABAR72302Negative
Viewer BBAR06303Negative
Viewer BBAR72302Negative
Viewer CBAR72302Negative
Viewer CBAR06303Negative
Viewer CBAR16308Negative
Cup formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cut-off)Near Cutoff Positive by GC/MS (Between the cut-off and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001324
Negative10151530
Viewer BPositive0001324
Negative10151530
Viewer CPositive0001424
Negative10151520

Discordant Results of Secobarbital Cup

ViewerSample NumberGC/MS ResultCup Format Viewer Results
Viewer ABAR06303Negative
Viewer ABAR16308Negative
Viewer ABAR72302Negative
Viewer BBAR06303Negative
Viewer BBAR16308Negative
Viewer BBAR72302Negative
Viewer CBAR06303Negative
Viewer CBAR72302Negative

{18}------------------------------------------------

Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001324
Negative10151530
Viewer BPositive0001324
Negative10151530
Viewer CPositive0001324
Negative10151530

Discordant Results of Secobarbital Dip Card

ViewerSample NumberGC/MS ResultDip Card FormatViewer Results
Viewer ABAR06303Negative
Viewer ABAR16308Negative
Viewer ABAR72302Negative
Viewer BBAR06303Negative
Viewer BBAR16308Negative
Viewer BBAR72302Negative
Viewer CBAR06303Negative
Viewer CBAR16308Negative
Viewer CBAR72302Negative

Buprenorphine

StripformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001324
Negative10151530
Viewer BPositive0001324
Negative10151530
Viewer CPositive0001424
Negative10151520

Discordant Results of Buprenorphine Strip

{19}------------------------------------------------

ViewerSample NumberGC/MS ResultStrip FormatViewer Results
Viewer ABUP2310.2Negative
Viewer ABUP6710.6Negative
Viewer ABUP1210.4Negative
Viewer BBUP2310.2Negative
Viewer BBUP6710.6Negative
Viewer BBUP1210.4Negative
Viewer CBUP2310.2Negative
Viewer CBUP1210.4Negative
CassetteformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001424
Negative10151520

Discordant Results of Buprenorphine Cassette

ViewerSample NumberGC/MS ResultCassette Format Viewer Results
Viewer ABUP2310.2Negative
Viewer ABUP1210.4Negative
Viewer BBUP2310.2Negative
Viewer BBUP1210.4Negative
Viewer CBUP2310.2Negative
Viewer CBUP1210.4Negative
Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
001324
Viewer APositive0

{20}------------------------------------------------

Negative10151530
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001324
Negative10151530
Discordant Results of Buprenorphine Dip Card
ViewerSample NumberGC/MS ResultDip Card FormatViewer Results
Viewer ABUP2310.2Negative
Viewer ABUP6710.6Negative
Viewer ABUP1210.4Negative
Viewer BBUP2310.2Negative
Viewer BBUP1210.4Negative
Viewer CBUP2310.2Negative
Viewer CBUP6710.6Negative
Viewer CBUP1210.4Negative
CupformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001324
Negative10151530
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001424
Negative10151520

Discordant Results of Buprenorphine Cup

ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer ABUP2310.2Negative
Viewer ABUP6710.6Negative
Viewer ABUP1210.4Negative
Viewer BBUP2310.2Negative
Viewer BBUP1210.4Negative
Viewer CBUP2310.2Negative
Viewer CBUP1210.4Negative

{21}------------------------------------------------

Methadone

StripformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001324
Negative10151530
Viewer CPositive0001324
Negative10151530

Discordant Results of Methadone Strip

ViewerSample NumberGC/MS ResultStrip Format Viewer Results
Viewer AMTD80302Negative
Viewer AMTD67303Negative
Viewer BMTD80302Negative
Viewer BMTD67303Negative
Viewer BMTD59306Negative
Viewer CMTD80302Negative
Viewer CMTD67303Negative
Viewer CMTD59306Negative
CassetteformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001424
Negative10151520

Discordant Results of Methadone Cassette

{22}------------------------------------------------

ViewerSample NumberGC/MS ResultCassette Format Viewer Results
Viewer AMTD80302Negative
Viewer AMTD67303Negative
Viewer BMTD80302Negative
Viewer BMTD67303Negative
Viewer CMTD80302Negative
Viewer CMTD67303Negative
Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001424
Negative10151520

Discordant Results of Methadone Dip Card

ViewerSample NumberGC/MS ResultDip Card FormatViewer Results
Viewer AMTD80302Negative
Viewer AMTD67303Negative
Viewer BMTD80302Negative
Viewer BMTD67303Negative
Viewer CMTD80302Negative
Viewer CMTD67303Negative
Cup formatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001324
Negative10151530

{23}------------------------------------------------

Viewer BPositive0001424
Negative10151520
Viewer CPositive0001424
Negative10151520
ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer AMTD80302Negative
Viewer AMTD67303Negative
Viewer AMTD59306Negative
Viewer BMTD80302Negative
Viewer BMTD67303Negative
Viewer CMTD80302Negative
Viewer CMTD67303Negative

Discordant Results of Methadone Cup

Lay-user study

A lay user study was performed at three intended user sites with 560 lay persons testing each of the Secobarbital, the Buprenorphine and the Methadone devices. Total of 1680 individuals performed the study. A total of 190 females and 370 males tested the Secobarbital samples, 217 females and 343 males tested the Buprenorphine samples and 216 females and 344 males tested the Methadone samples. They had diverse educational and professional backgrounds and ranged in age from 21 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

% of CutoffNumber ofsamplesSecobarbitalConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100%
-75% Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022511995%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%

Comparison between GC/MS and Lay Person Results (Secobarbital Strip)

Comparison between GC/MS and Lay Person Results (Secobarbital Cassette)

{24}------------------------------------------------

NumberofsamplesSecobarbitalConcentration by GC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022511995%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%

Comparison between GC/MS and Lay Person Results (Secobarbital DipCard)

NumberSecobarbitalLay person resultsThe
% of CutoffofsamplesConcentration by GC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022511995%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%

Comparison between GC/MS and Lay Person Results (Secobarbital Cup)

NumberofsamplesSecobarbitalConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
% of CutoffNo. ofPositiveNo. ofNegative
-100% Cutoff200020100%
-75% Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022511995%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%
% of CutoffNumberBuprenorphineLay person resultsThe
-------------------------------------------------------------

{25}------------------------------------------------

ofsamplesConcentration by GC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff202.5020100%
-50% Cutoff205020100%
-25% Cutoff207.511995%
+25% Cutoff2012.518290%
+50% Cutoff2015200100%
+75% Cutoff2017.5200100%

Comparison between GC/MS and Lay Person Results (Buprenorphine Cassette)

% of CutoffNumber of samplesBuprenorphine Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100%Cutoff200020100%
-75% Cutoff202.5020100%
-50% Cutoff205020100%
-25% Cutoff207.511995%
+25% Cutoff2012.519195%
+50% Cutoff2015200100%
+75% Cutoff2017.5200100%

Comparison between GC/MS and Lay Person Results (Buprenorphine DipCard)

% of CutoffNumber of samplesBuprenorphine Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
No. of PositiveNo. of Negative
-100% Cutoff200020100%
-75% Cutoff202.5020100%
-50% Cutoff205020100%
-25% Cutoff207.521890%
+25% Cutoff2012.518290%
+50% Cutoff2015200100%
+75% Cutoff2017.5200100%
Comparison between GC/MS and Lav Person Results (Buprenorphine Cup)
-------------------------------------------------------------------------------
% of CutoffNumberBuprenorphineLay person resultsThe
-------------------------------------------------------------

{26}------------------------------------------------

ofsamplesConcentration by GC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff202.5020100%
-50% Cutoff205020100%
-25% Cutoff207.521890%
+25% Cutoff2012.518290%
+50% Cutoff2015200100%
+75% Cutoff2017.5200100%

Comparison between GC/MS and Lay Person Results (Methadone Strip)

% of CutoffNumberofsamplesMethadoneConcentration by GC/MS(ng/mL)Lay person resultsNo. ofPositiveLay person resultsNo. ofNegativeThepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022511995%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%

Comparison between GC/MS and Lay Person Results (Methadone Cassette)

% of CutoffNumberofsamplesMethadoneConcentration by GC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100%Cutoff200020100%
-75% Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022511995%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%
Comparison between GC/MS and Lay Person Results (Methadone DipCard)
% of CutoffNumberMethadoneLay person resultsThe
---------------------------------------------------------

{27}------------------------------------------------

ofsamplesConcentration by GC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022511995%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%

Comparison between GC/MS and Lay Person Results (Methadone Cup)

% of CutoffNumberofsamplesMethadoneConcentration by GC/MS(ng/mL)Lay person resultsThe
No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022521890%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

    1. Clinical Studies
      Not applicable.
  • 11.Conclusion
    Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity and method comparison of the devices, it's concluded that the Healgen Secobarbital Test, Healgen Buprenorphine Test and Healgen Methadone Test are substantially equivalent to the predicate.

§ 862.3150 Barbiturate test system.

(a)
Identification. A barbiturate test system is a device intended to measure barbiturates, a class of hypnotic and sedative drugs, in serum, urine, and gastric contents. Measurements obtained by this device are used in the diagnosis and treatment of barbiturate use or overdose and in monitoring levels of barbiturate to ensure appropriate therapy.(b)
Classification. Class II (special controls). A barbiturate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).