K Number
K150627

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-05-27

(77 days)

Product Code
Regulation Number
890.5710
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Rapid Aid Infant Heel Warmer is primarily used in hospitals, doctor's offices, and other healthcare facilities. It is an instant warm pack intended to be used on an infant's heel to increase blood circulation to the area to aid in the drawing of blood for analysis. It is a single use, non-toxic, non-sterile, disposable warmer.

Device Description

The Rapid Aid Infant Heel Warmer is a self contained unit comprised of a flexible, poly/nylon outer pouch containing:

  • a) A metal disc that can be flexed during activation,
  • b) Liquid solution of food grade sodium acetate and water,
  • c) Minute crystals of sodium acetate.
    An adhesive tape is attached to the top of the unit is activated by squeezing firmly on the inner fluid pouch, this will cause the inner perforated pouch to activate. Rapid crystallization occurs when the liquid contents are exposed to the minute crystals of sodium acetate contained within the poly/nylon outer pouch. This exothermnic reaction causes the unit to heat up to 104ºF. The adhesive tape strip is used to hold the warmer in place if desired.
AI/ML Overview

Here's an analysis of the provided text regarding the Rapid Aid Instant Disposable Infant Heel Warmer, focusing on acceptance criteria and study details:

1. Table of Acceptance Criteria and Reported Device Performance

The main performance characteristic tested for this device is its temperature profile.

Acceptance Criteria CategoryAcceptance Criteria (Implicit)Reported Device Performance (New Device)Reported Device Performance (Predicate Device)
Maximum TemperatureWithin specified range102.6 ± 0.2 °F102.2 ± 0.4 °F
Dwell Time (within 101-104 °F)Within specified range8.5 ± 1.0 min8.5 ± 0.8 min

Note: The exact numerical "specified range" for maximum temperature and dwell time is not explicitly stated as a formal acceptance criterion but is implied by the comparison to the predicate device and the statement "both within the specified range for this device."

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated. The document mentions "The Infant Heel Warmer with Disc was tested against the predicate device," implying a test set, but does not provide the number of units tested.
  • Data Provenance: Not explicitly stated, but it's a bench test conducted by the manufacturer, Rapid Aid Corp. (Canada). This would be prospective data generated for the 510(k) submission.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not applicable to this type of device and study. The testing described is bench testing of physical properties (temperature, dwell time, material properties), not diagnostic accuracy requiring expert interpretation.

4. Adjudication Method for the Test Set

This information is not applicable as the study involves objective physical measurements, not subjective evaluations requiring adjudication.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not conducted. This device is a physical warmer, not an AI or diagnostic tool that would involve human readers interpreting cases.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This information is not applicable. This is a physical medical device, not an algorithm, so "standalone performance" in the context of AI does not apply. The performance evaluation was essentially a standalone bench test of the device's physical properties.

7. The Type of Ground Truth Used

The "ground truth" for the performance evaluation appears to be the established performance characteristics of the legally marketed predicate device (Rapid Aid Instant Disposable Infant Heel Warmer, K040856) and the "specified range for this device" (though this range is not explicitly defined in the document). In essence, the new device's performance was compared to that of a previously cleared, substantially equivalent device.

For material properties, the ground truth is based on ASTM standards (for poly/nylon pouch material) and toxicological acceptability (for sodium acetate).

8. The Sample Size for the Training Set

This information is not applicable. This is a physical medical device, not an AI model, so there is no "training set."

9. How the Ground Truth for the Training Set Was Established

This information is not applicable for the reasons stated in point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized depiction of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 27, 2015

Rapid Aid Corporation c/o Mandell Horwitz Consultants, LLC Diane Horwitz, Ph.D., RAC 2995 Steven Martin Drive Fairfax, VA 22031

Re: K150627

Trade Name: Rapid Aid Instant Disposable Infant Heel Warmer Regulation Number: 21 CFR 890.5710 Regulation Name: Hot or cold disposable pack Regulatory Class: Class I Product Code: MPO Dated: April 27, 2015 Received: April 27, 2015

Dear Dr. Horwitz:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code 27f Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Felipe Aquel - S

Carlos L. Peña. PhD. MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150627

Device Name

Rapid Aid Instant Disposable Infant Heel Warmer

Indications for Use (Describe)

Rapid Aid Infant Heel Warmer is primarily used in hospitals, doctor's offices, and other healthcare facilities. It is an instant warm pack intended to be used on an infant's heel to increase blood circulation to the area to aid in the drawing of blood for analysis. It is a single use, non-toxic, non-sterile, disposable warmer.

Type of Use (Select one or both, as applicable)

  • Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Rapid Aid, a company that specializes in treating bumps and bruises. The logo is made up of the words "RAPID" in blue and "AID" in red, with a plus sign between them. The tagline "Makes it Better!" is below the logo. The text "Treating the World's Bumps & Bruises" is above the logo.

510(k) Summary of Safety and Effectiveness

Manufacturer and Submitter

Company Name:Rapid Aid Corp.
Company Address:4120A Sladeview CrescentMISSISSAUGA ON L5L 5Z3 CANADA
Telephone: (905) 820-4788
Fax: (905) 820-1649
Contact Person:Rose Ren, QA/Regulatory Affairs Manager
Date Summary Prepared:March 1, 2015

Device

Trade/Device Name:Rapid Aid Instant Disposable Infant Heel Warmer - Disc
Common/Usual Name:Infant Heel Warmer
Classification Name:Infant Heel Warmer (Chemical Heat Pack)
Regulation Number:21 CFR 890.5710
Product Code:MPO
Classification Panel:Physical Medicine
Classification:Class I

Substantial Equivalence

The Rapid Aid Instant Disposable Infant Heel Warmer with Disc is substantially equivalent to the Rapid Aid Instant Disposable Infant Heel Warmer (K040856).

Device Description

The Rapid Aid Infant Heel Warmer is a self contained unit comprised of a flexible, poly/nylon outer pouch containing:

  • a) A metal disc that can be flexed during activation,
  • b) Liquid solution of food grade sodium acetate and water,
  • c) Minute crystals of sodium acetate.

An adhesive tape is attached to the top of the unit is activated by squeezing firmly on the inner fluid pouch, this will cause the inner perforated pouch to activate. Rapid crystallization occurs when the liquid contents are exposed to the minute crystals of sodium acetate contained within the poly/nylon outer pouch. This exothermnic reaction causes the unit to heat up to 104ºF. The adhesive tape strip is used to hold the warmer in place if desired.

Intended Use/Indications for Use

Rapid Aid Infant Heel Warmer is primarily used in hospitals, doctor's offices, and other healthcare facilities. It is an instant warm pack intended to be used on an infant's heel to increase blood circulation to the area to aid in the drawing of blood for analysis. It is a single use, nontoxic, non-sterile, disposable warmer.

Substantial Equivalence Analysis

The Infant Heel Warmer with Disc has the identical Intended Use and Indications for use as the predicate device. The basic technological characteristics remain the same, with the exception of the replacing an inner pouch for activation with a metal disc, to start the exothermic reaction.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for Rapid Aid, a company that specializes in treating bumps and bruises. The logo features the words "RAPID" in large, blue letters, followed by "AID" in red letters. A red cross is incorporated into the design of the word "AID". The tagline "Makes it Better!" is written in blue below the logo.

Several other currently marketed Infant Heel Warmer products cleared by FDA use this method of activation. The modified device has been tested and it has been verified that product specifications are met. It is therefore concluded that the Infant Heel Warmer with Disc is substantially equivalent to the predicate device.

Performance - Bench Testing

The Infant Heel Warmer with Disc was tested against the predicate device to ensure conformance with temperature characteristics (maximum temperature and dwell time requirements). The average maximum temperature achieved was 102.6 ± 0.2 °F (new device) and 102.2 ± 0.4 °F (predicate) and the average dwell time that the unit was between 101 °F and 104 °F was 8.5 ± 1.0 min (new device) and 8.5 ± 0.8 min (predicate). The predicate and the new devices were both within the specified range for this device.

The materials that contact the skin have not changed:

  • The outer poly/nylon pouch material has been tested following ASTM standards and is latex free and non-sensitizing.
  • . The finished packing material is tested for: thickness following ASTM D1203, tensile strength following ASTM D-882 and seal width. Pouch packaging material is subject to incoming inspection for width/length, seal integrity and burst strength.
  • . The chemical mixture claims are based on the results of testing the sodium acetate, which is a nontoxic, food grade chemical and has been found to be toxicologically acceptable for it's intended use.
  • . Finished product is subject to testing for peak temperature, seal integrity and pressure testing.

Date Summary Prepared: March 1, 2015

§ 890.5710 Hot or cold disposable pack.

(a)
Identification. A hot or cold disposable pack is a device intended for medical purposes that consists of a sealed plastic bag incorporating chemicals that, upon activation, provides hot or cold therapy for body surfaces.(b)
Classification. Class I (general controls). Except when intended for use on infants, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 890.9.