AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Emit® II Plus Buprenorphine Assay is a homogeneous enzyme immunoassay with a 5 ng/mL cutoff. The assay is intended for use in laboratories for the qualitative analyses of buprenorphine in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.

The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as LC/MS or Permitting laboratories to establish quality control procedures.

The Emit® II Plus Buprenorphine Assay provides only a preliminary analytical test result. A more specific alternative chemical method(s) must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/ MS) or LC/MS are the preferred confirmatory methods. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Emit® II Plus Specialty Drug Calibrator/Control Level 1, Emit® II Plus Specialty Drug Calibrator/Control Level 2, Emit® II Plus Specialty Drug Calibrator/Control Level 3, Emit® II Plus Specialty Drug Calibrator/Control Level 4

The Emit® II Plus Specialty Drug Calibrators/Controls are used in the calibration of the Emit® II Plus Buprenorphine Assay. These products may also be used as quality control materials based on the Buprenorphine Assay cutoff.

Emit® II Plus Specialty Drug Control Negative and Emit® II Plus Specialty Drug Control Positive

The Emit® II Plus Specialty Drug Control Negative and Control Positive are for use with the Emit® II Plus Buprenorphine Assay.

Device Description

The Emit® II Plus Buprenorphine assay is a homogeneous enzyme immunoassay with a 5 ng/mL cutoff. The assay, used for the detection of Buprenorphine in human urine, utilizes a two-reagent system. The Antibody/Substrate Reagent 1 is a liquid ready-to-use product comprised of mouse monoclonal antibodies to buprenorphine, glucose-6-phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in a diluent containing bovine serum albumin (BSA), preservatives and stabilizers. The Enzyme Reagent 2 is a liguid. ready-to-use product containing norbuprenorphine labeled bacterial recombinant glucose-6 phosphate dehydrogenase (rG6PDH) in a diluent containing bovine serum albumin (BSA), Hepes buffer, preservatives and stabilizers.

The assay kit consists of Reagent 1 and Reagent 2 in plastic containers and is available in three sizes: large kit (1L), small kit (115 mL), and 28 mL Emit® II Plus assays are designed for use with a number of chemistry analyzers.

The Emit® II Plus Buprenorphine assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result.

Emit® II Plus Specialty Drug Calibrators / Controls are in-vitro diagnostic products used in the calibration of the Emit® II Plus Buprenorphine assay. These products may also be used as quality control materials based on the Buprenorphine Assay cutoff. The matrix is pooled, drug-free, human urine based product containing buprenorphine, preservatives and surfactants. Each level of calibrator is packaged and sold separately, 1 plastic bottle per box, 10 mL in a 15 mL bottle. Values are nominal and are verified with urine based buprenorphine anchor pool and adjusted if needed. The anchor pool levels are verified by LC/MS/MS and are within 10% of nominal drug concentration for levels 2.5 and 5.0 ng/mL and within 5 % of nominal concentrations for levels 15 and 25 ng/mL.

Emit® II Plus Specialty Drug Control Negative and Emit® II Plus Specialty Drug Control Positive are in-vitro diagnostic products are for use with the Emit® II Plus Buprenorphine assay. The matrix is pooled, drug-free, human urine based product containing buprenorphine, preservatives and surfactants. The Control Negative and Control Positive are packaged separately in 15 mL plastic vials with a 10 mL fill per vial. Values are nominal and are verified with urine based buprenorphine anchor pool and adjusted if needed. Anchor pool levels are verified by LC/MS/MS.

AI/ML Overview

Here's an analysis of the acceptance criteria and the study proving the device meets them, based on the provided text:

Device: Emit® II Plus Buprenorphine Assay, Emit® II Plus Specialty Drug Calibrator/Control Level 1-4, Emit® II Plus Specialty Drug Control Negative/Positive.

Acceptance Criteria and Reported Device Performance

The acceptance criteria are not explicitly stated as numbered or bulleted points in the provided document. However, based on the performance data presented, we can infer the criteria that the device aimed to meet for substantial equivalence. The document primarily focuses on demonstrating performance against expected standards for an in vitro diagnostic device for drug screening, particularly a qualitative and semi-quantitative assay.

Here's a table summarizing the inferred acceptance criteria and the device's performance:

Acceptance Criteria Category/Inferred CriteriaReported Device Performance
Precision (Qualitative Analysis) - Agreement near cutoff - Consistent classification (Negative/Positive) for concentrations significantly below/above cutoff. - Acceptable level of concordant/discordant results near the cutoff for both qualitative and semi-quantitative modes.Precision: Qualitative Analysis (5 ng/mL cutoff) - 0, 2.5, 3, 3.75 ng/mL (below cutoff): 80 Negative out of 80 results (100% agreement). - 5 ng/mL (cutoff): 25 Negative / 55 Positive (68.75% positive, 31.25% negative). This indicates some variability at the exact cutoff, which is typical for immunoassays. - 6.25, 7, 7.5, 10 ng/mL (above cutoff): 80 Positive out of 80 results (100% agreement). - Additional Precision Study: - 1.25 ng/mL (-75% cutoff): 80 Negative out of 80 results (100% agreement). - 8.75 ng/mL (+75% cutoff): 80 Positive out of 80 results (100% agreement). Overall, good precision away from the cutoff, expected variability at the cutoff.
Precision (Semi-quantitative Analysis) - Agreement near cutoff - Consistent classification (Negative/Positive) and semi-quantitative results for concentrations significantly below/above cutoff.Precision: Semi-quantitative Analysis (5 ng/mL cutoff) - 0, 2.5, 3, 3.75 ng/mL (below cutoff): 80 Negative out of 80 results (100% agreement). - 5 ng/mL (cutoff): 25 Negative / 55 Positive (68.75% positive, 31.25% negative). - 6.25, 7, 7.5, 10 ng/mL (above cutoff): 80 Positive out of 80 results (100% agreement). - Additional Precision Study: - 1.25 ng/mL (-75% cutoff): 80 Negative out of 80 results (100% agreement). - 8.75 ng/mL (+75% cutoff): 80 Positive out of 80 results (100% agreement). Results mirror qualitative analysis, demonstrating consistent semi-quantitative performance.
Specificity and Cross-reactivity - Buprenorphine Metabolites - Appropriate cross-reactivity with buprenorphine and its primary active metabolite (norbuprenorphine). - Low to no cross-reactivity with inactive glucuronide metabolites.Buprenorphine Metabolite Recovery - Buprenorphine (5 ng/mL): 103.2% recovery (5.2 ng/mL). - Norbuprenorphine (5 ng/mL): 92.6% recovery (4.6 ng/mL). - Buprenorphine Glucuronide (1000 ng/mL): 0.1% cross-reactivity (0.9 ng/mL). - Norbuprenorphine Glucuronide (1000 ng/mL): 0.1% cross-reactivity (1.2 ng/mL). Demonstrates good detection of buprenorphine and norbuprenorphine, minimal interference from inactive glucuronide metabolites.
Specificity and Cross-reactivity - Structurally Related Compounds - No significant false positives from a panel of common opioids and other structurally related compounds at high concentrations.Cross-reactivity with Structurally Related Compounds - Over 20 common opioids and related compounds (e.g., 6-acetylcodeine, codeine, heroin, morphine, oxycodone) were tested at 100,000 ng/mL. - All tested compounds showed "Negative" qualitative results and "<0.01%" cross-reactivity (semi-quantitative). Excellent specificity, not cross-reacting with a wide range of other opioids at very high concentrations, which is critical for drug screening assays.
Interference - Structurally Unrelated Compounds - No false positive or false negative results from a panel of common medications and other substances expected in urine at clinically relevant concentrations.Interference of Structurally Unrelated Compounds - Over 50 structurally unrelated compounds (e.g., acetaminophen, amoxicillin, caffeine, ibuprofen, fluoxetine, methadone, naproxen, phencyclidine, THC) were tested. - Each interferent was spiked into urine pools at -40% cutoff (3 ng/mL buprenorphine) and +40% cutoff (7 ng/mL buprenorphine). - At the stated concentration for each interferent, all "-40% Cutoff" samples remained Negative and all "+40% Cutoff" samples remained Positive, for both qualitative and semi-quantitative results. Demonstrates good resistance to interference from a broad range of common drugs and substances.
Interference - Endogenous Substances - No false positive or false negative results due to endogenous substances, pH, or specific gravity variations.Endogenous Substances Interference, pH, and Specific Gravity - 15 endogenous substances (e.g., acetone, ascorbic acid, bilirubin, creatinine, ethanol, glucose, hemoglobin, urea) were tested at high concentrations. - pH range (pH 3 to pH 11) and Specific Gravity range (1.002 to 1.035) were tested. - For all endogenous substances, pH levels, and specific gravity levels, the -40% cutoff samples remained Negative, and the +40% cutoff samples remained Positive, for both qualitative and semi-quantitative results. Indicates robust performance across a range of physiological conditions and common urine constituents.
Recovery/Linearity - Accurate semi-quantitative recovery of buprenorphine across a range of concentrations.Recovery/Linearity - Tested nine concentrations of buprenorphine from 2 ng/mL to 25 ng/mL. - % Recovery ranged from 92.5% to 105.0%. Demonstrates good linearity and accurate recovery of buprenorphine concentrations over the tested range.
Method Comparison (Qualitative and Semi-quantitative) - Agreement with Reference Method - High agreement with a more specific, confirmatory method (LC/MS/MS). - Acceptable number of discordant results, with plausible explanations for any discrepancies.Method Comparison (Emit® II Plus Buprenorphine Assay vs. LC/MS/MS) - Total Samples: 127 unaltered human urine samples. - Qualitative Agreement: 90% positive agreement, 98% negative agreement. - Overall Agreement (from 2x2 table): (65 + 54) / 127 = 119 / 127 = 93.7%. - Discordant Results (8 samples): - 7 samples were Emit® Positive but LC/MS/MS Negative (concentration between 3.86-4.97 ng/mL, below 5 ng/mL cutoff for LC/MS/MS buprenorphine/norbuprenorphine sum). These are typically considered "near cutoff" false positives for a screening assay. - 1 sample was Emit® Negative but LC/MS/MS Positive (buprenorphine at 5.12 ng/mL). This is a single false negative near the cutoff. High overall agreement with LC/MS/MS, with expected discordance at concentrations close to the cutoff, demonstrating performance comparable to drug screening immunoassays.

Study Details:

  1. Sample Size used for the test set and data provenance:

    • Precision/Cutoff Characterization: For the main precision study, 8 urine pools were tested (0, 2.5, 3, 3.75, 5, 6.25, 7, 7.5, 10 ng/mL). Each pool was analyzed in duplicate twice a day for 20 days, resulting in 80 replicates per concentration level (total 720 replicates) for the main study. An additional precision study used 2 urine pools (1.25 and 8.75 ng/mL), with 80 replicates each (total 160 replicates). The urine was "drug-free human urine" spiked with buprenorphine.
    • Specificity and Cross-reactivity (Metabolites): Each metabolite was tested at n=5 replicates.
    • Specificity and Cross-reactivity (Structurally Related Compounds): Each compound was tested at n=5 replicates.
    • Interference (Structurally Unrelated Compounds): Each interferent was tested at n=5 replicates (at both -40% and +40% cutoff levels).
    • Interference (Endogenous Substances): Each endogenous substance, pH level, and specific gravity level was tested (N=not explicitly stated, but typically n=5 replicates as above).
    • Recovery/Linearity: Nine concentrations were analyzed in replicates of N=5.
    • Method Comparison: 127 unaltered human urine samples were used.
    • Data Provenance: The document states studies were conducted at "Siemens Healthcare Diagnostics Inc." and used "human urine" (drug-free or unaltered). No specific country of origin is mentioned. It is an internal, retrospective study using banked samples or samples collected for the purpose of the study.
  2. Number of experts used to establish the ground truth for the test set and qualifications of those experts:

    • For the Method Comparison study, the ground truth was established by LC/MS/MS (Liquid Chromatography/Mass Spectrometry/Mass Spectrometry), which is a highly specific and sensitive analytical method considered a "confirmatory method." This is a laboratory-based instrumental method, not established by human experts in the traditional sense of medical image interpretation or clinical diagnosis. The "experts" would be the laboratory personnel operating and interpreting the LC/MS/MS results, qualified in analytical chemistry and toxicology testing. Their number and specific qualifications are not specified in this document.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • For the method comparison, the ground truth was established by LC/MS/MS as a single, definitive reference method. There was no multi-expert adjudication mentioned as would be common in diagnostic imaging studies. The LC/MS/MS result serves as the "gold standard."
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic assay (a chemical test) and operates independently of human "readers" or "AI assistance" in the way that imaging devices or AI-driven diagnostic software would. The output is a numerical value or a qualitative (positive/negative) result generated by an automated analyzer.
  5. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

    • Yes, the provided performance data represents standalone performance. The Emit® II Plus Buprenorphine Assay is an automated immunoassay designed for use on chemistry analyzers. All precision, specificity, interference, linearity, and method comparison data reflect the assay's performance without "human-in-the-loop" interpretation beyond running the assay and recording the results. The ultimate interpretation and clinical judgment for drug-of-abuse test results might involve a human, but the analytical performance presented is the measurement device's standalone capability.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc):

    • The ground truth for the performance studies, particularly the crucial method comparison, was established by LC/MS/MS (Liquid Chromatography/Mass Spectrometry/Mass Spectrometry). This is an analytical reference method considered the gold standard for confirming drug presence and concentration in toxicology.
  7. The sample size for the training set:

    • This document describes performance characteristics of a diagnostics assay for regulatory submission. It does not refer to a "training set" in the context of machine learning. The "training" of such a device primarily involves the development and optimization of the reagent formulation and assay parameters during the product development phase, which isn't typically quantified by a distinct "training set sample size" like in AI/ML development. The precision studies use spiked urine, which could be considered part of the characterization but not a "training set" in the modern AI sense.
  8. How the ground truth for the training set was established:

    • Given that this is not an AI/ML device but a chemical immunoassay, the concept of a "ground truth for a training set" as typically understood in AI/ML performance studies does not directly apply. The calibration and control materials are manufactured to known (targeted) concentrations, and their values are "verified with urine based buprenorphine anchor pool and adjusted if needed," with anchor pool levels "verified by LC/MS/MS." This implies that the 'true' concentrations for calibrators and controls used to define the assay's performance characteristics are established via LC/MS/MS.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a seal for the Department of Health & Human Services - USA. The seal is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, creating a sense of depth and unity.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 23, 2015

SIEMENS HEALTHCARE DIAGNOSTICS, INC. FRANCES DILLON SR. MANAGER, REGULATORY AFFAIRS P.O. BOX 6101, M/S 514 NEWARK DE 19714-6101

Re: K150606

Trade/Device Name: Emit II Plus Buprenorphine Assay,

Emit II Plus Specialty Drug Calibrator/Control Level 1, Emit II Plus Specialty Drug Calibrator/Control Level 2, Emit II Plus Specialty Drug Calibrator/Control Level 3, Emit II Plus Specialty Drug Calibrator/Control Level 4. Emit II Plus Specialty Drug Control Negative, Emit II Plus Specialty Drug Control Positive

Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG, DLJ, LAS Dated: September 4, 2015 Received: September 8, 2015

Dear Ms. Dillon:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{1}------------------------------------------------

CFR Part 807): labeling (21 CFR Parts 801 and 809); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney H. Lias -S

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K150606

Device Name

Emit® II Plus Buprenorphine Assay, Emit® II Plus Specialty Drug Calibrator/Control Level 1, Emit® II Plus Specialty Drug Calibrator/Control Level 2, Emit® II Plus Specialty Drug Calibrator/Control Level 3, Emit® II Plus Specialty Drug Calibrator/Control Level 4, Emit® II Plus Specialty Drug Control Negative, Emit® II Plus Specialty Drug Control Positive,

Indications for Use (Describe) Emit® II Plus Buprenorphine Assay

Emit® II Plus Buprenorphine Assay is a homogeneous enzyme immunoassay with a 5 ng/mL cutoff. The assay is intended for use in laboratories for the qualitative analyses of buprenorphine in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.

The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as LC/MS or Permitting laboratories to establish quality control procedures.

The Emit® II Plus Buprenorphine Assay provides only a preliminary analytical test result. A more specific alternative chemical method(s) must be used to obtain a confirmed analytical result. Gas Chromatography/Mass Spectrometry (GC/ MS) or LC/MS are the preferred confirmatory methods. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Emit® II Plus Specialty Drug Calibrator/Control Level 1, Emit® II Plus Specialty Drug Calibrator/Control Level 2, Emit® II Plus Specialty Drug Calibrator/Control Level 3, Emit® II Plus Specialty Drug Calibrator/Control Level 4

The Emit® II Plus Specialty Drug Calibrators/Controls are used in the calibration of the Emit® II Plus Buprenorphine Assay. These products may also be used as quality control materials based on the Buprenorphine Assay cutoff.

Emit® II Plus Specialty Drug Control Negative and Emit® II Plus Specialty Drug Control Positive

The Emit® II Plus Specialty Drug Control Negative and Control Positive are for use with the Emit® II Plus Buprenorphine Assay.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(K) Summary

This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the Safe Medical Device Act of 1990 and 21 CFR 807.92.

The assigned 510(k) number is: K150606

1. Submitter

Siemens Healthcare Diagnostics Inc. P.O. Box 6101 Newark, DE 19714

Tel: 302-631-6951 FAX: 3023-631-6299

Contact Person: Frances Dillon Date of Preparation: October 20, 2015

2. Device Information

Trade Name:Emit® II Plus Buprenorphine Assay
Common Name:Buprenorphine Assay
Classification Name:Opiates test system - 21CFR 862.3650
Regulatory Class:II
Product Code:DJG - Enzyme Immunoassay, Opiates
Panel:Clinical Toxicology (91)
Trade Name:Emit® II Plus Specialty Drug Calibrator/Control Level 1Emit® II Plus Specialty Drug Calibrator/Control Level 2Emit® II Plus Specialty Drug Calibrator/Control Level 3Emit® II Plus Specialty Drug Calibrator/Control Level 4
Common Name:Buprenorphine Calibrator/Control
Classification Name:Clinical Toxicology Calibrator - 21CFR 862.3200Clinical Toxicology Control Material - 21CFR 862.3280
Regulatory Class:II (Calibrator)I (Control)
Product Code:DLJ - Calibrators, Drug SpecificLAS - Drug Specific Control Materials
Panel:Clinical Toxicology (91)
Trade Name:Emit® II Plus Specialty Drug Control NegativeEmit® II Plus Specialty Drug Control Positive
Common Name:Buprenorphine Negative Control and BuprenorphinePositive Control
Classification Name:Clinical Toxicology Control material - 21CFR 862.3280
Regulatory Class:I
Product Code:LAS - Drug Specific Control Materials
Panel:Clinical Toxicology (91)

{4}------------------------------------------------

3. Predicate Device

Assay:Microgenics CEDIA® Buprenorphine Assay, K040316.
Calibrator:Microgenics CEDIA® Buprenorphine Calibrators, K040316.
Control:Microgenics CEDIA® Buprenorphine Controls, K040316.

4. Device Description

Assav

The Emit® II Plus Buprenorphine assay is a homogeneous enzyme immunoassay with a 5 ng/mL cutoff. The assay, used for the detection of Buprenorphine in human urine, utilizes a two-reagent system. The Antibody/Substrate Reagent 1 is a liquid ready-to-use product comprised of mouse monoclonal antibodies to buprenorphine, glucose-6-phosphate (G6P), and nicotinamide adenine dinucleotide (NAD) in a diluent containing bovine serum albumin (BSA), preservatives and stabilizers. The Enzyme Reagent 2 is a liguid. ready-to-use product containing norbuprenorphine labeled bacterial recombinant glucose-6 phosphate dehydrogenase (rG6PDH) in a diluent containing bovine serum albumin (BSA), Hepes buffer, preservatives and stabilizers.

The assay kit consists of Reagent 1 and Reagent 2 in plastic containers and is available in three sizes: large kit (1L), small kit (115 mL), and 28 mL Emit® II Plus assays are designed for use with a number of chemistry analyzers.

The Emit® II Plus Buprenorphine assay provides only a preliminary analytical test result. A more specific alternative chemical method must be used to obtain a confirmed analytical result.

Calibrators / Controls

Emit® II Plus Specialty Drug Calibrators / Controls are in-vitro diagnostic products used in the calibration of the Emit® II Plus Buprenorphine assay. These products may also be used as quality control materials based on the Buprenorphine Assay cutoff. The matrix is pooled, drug-free, human urine based product containing buprenorphine, preservatives and surfactants. Each level of calibrator is packaged and sold separately, 1 plastic bottle per box, 10 mL in a 15 mL bottle. Values are nominal and are verified with urine based buprenorphine anchor pool and adjusted if needed. The anchor pool levels are verified by LC/MS/MS and are within 10% of nominal drug concentration for levels 2.5 and 5.0 ng/mL and within 5 % of nominal concentrations for levels 15 and 25 ng/mL.

Emit® II Plus Specialty Drug Calibrator/Control levels are as follows:

Specialty Drug Calibrator/ControlTargeted Buprenorphine Concentration(ng/mL
Level 12.5
Level 25
Level 315
Level 425

{5}------------------------------------------------

The Emit® Calibrator/Control Level 0, which contains no drug and was cleared under K993755 is used with the Emit® II Plus Buprenorphine Assay. There are no changes to the Calibrator Level 0 product. Controls

Emit® II Plus Specialty Drug Control Negative and Emit® II Plus Specialty Drug Control Positive are in-vitro diagnostic products are for use with the Emit® II Plus Buprenorphine assay. The matrix is pooled, drug-free, human urine based product containing buprenorphine, preservatives and surfactants. The Control Negative and Control Positive are packaged separately in 15 mL plastic vials with a 10 mL fill per vial. Values are nominal and are verified with urine based buprenorphine anchor pool and adjusted if needed. Anchor pool levels are verified by LC/MS/MS.

Emit® II Plus Specialty Drug Control Negative and Control Positive levels are as follows:

Specialty Drug ControlTargeted Buprenorphine Concentration(ng/mL)
Negative3
Positive7

5. Indications for Use

Emit® II Plus Buprenorphine Assay

Emit® II Plus Buprenorphine Assay is a homogeneous enzyme immunoassay with a 5 nq/mL cutoff. The assay is intended for use in laboratories for the qualitative and/or semiquantitative analyses of buprenorphine in human urine. Emit® II Plus assays are designed for use with a number of chemistry analyzers.

The semi-quantitative mode is for the purpose of enabling laboratories to determine an appropriate dilution of the specimen for confirmation by a confirmatory method such as LC/MS or permitting laboratories to establish quality control procedures.

The Emit® II Plus Buprenorphine Assay provides only a preliminary analytical test result. A more specific alternative chemical method(s) must be used to obtain a confirmed analytical result. GC/MS and LC/MS are the preferred confirmatory methods. Other chemical confirmation methods are available. Clinical consideration and professional judgment should be applied to any drug-of-abuse test result, particularly when preliminary positive results are used.

Emit® II Plus Specialty Drug Calibrator / Control Level 1, Level 2, Level 3, and Level 4

The Emit® II Plus Specialty Drug Calibrators/Controls are used in the calibration of the Emit® II Plus Buprenorphine Assay. These products may also be used as quality control materials based on the Buprenorphine Assay cutoff.

Emit® II Plus Specialty Drug Control Negative and Emit® II Plus Specialty Drug Control Positive

{6}------------------------------------------------

The Emit® II Plus Specialty Drug Control Negative and Control Positive are for use with the Emit® Il Plus Buprenorphine Assay.

AttributesPredicate DeviceCEDIA® Buprenorphine Assay(K040316)Proposed DeviceEmit® II PlusBuprenorphine Assay
Similarities
Intended UseA homogeneous enzymeimmunoassay for qualitative orsemi-quantitative determination ofthe presence of buprenorphine inhuman urine at a cutoffconcentration of 5 ng/mL.Preliminary analytical test result.Same
AnalyteBuprenorphineSame
AntibodyMouse monoclonal antibody tobuprenorphineSame
Test SystemHomogeneous enzymeimmunoassaySame
DetectionAbsorbance change measuredspectrophotometricallySame
Sample typeHuman urineSame
Cutoff Level5 ng/mLSame
InstrumentAutomated clinical analyzerscapable of maintaining a constanttemperature, pipetting, mixingreagents, measuring enzymaticrates, and timing the reaction.Same
Calibrator levels0 ng/mL + four (4) levelsSame
Control LevelsHigh Control 7 ng/mLLow Control 3 ng/mLSame
Differences
AssaymethodologyUses CEDIA® technologyUses EMIT® technology
DetectionAbsorbance change measuredspectrophotometrically at 660 nm.Absorbance changemeasuredspectrophotometrically at340 nm.
ReferenceMethodologyGC/MSLC/MS
Reagents FormR1 and R2: Lyophilized(Reconstitution Required)R1: Liquid - Ready to useR2: Liquid - Ready to use
Calibrator levels0, 5, 20, 50, and 75 ng/mL0, 2.5, 5, 15, and 25 ng/mL

6. Comparison of Technological Characteristics with the Predicate Device

{7}------------------------------------------------

Comparison of Calibrator/Control Similarities and Differences

AttributesPredicate DeviceCEDIA® BuprenorphineCalibrator (K040316)Proposed DeviceEmit® II Plus SpecialtyDrug Calibrator / Control
Similarities
Intended UseThe CEDIA® Buprenorphinecalibrators are used tocalibrate the CEDIA®Buprenorphine Assay inhuman urine.Same
MatrixHuman urineSame
AnalyteBuprenorphineSame
PreparationLiquid - Ready to useSame
Storage2 – 8 °CSame
Differences
TargetConcentrations forBuprenorphine0, 5, 20, 50, and 75 ng/mL2.5, 5, 15, and 25 ng/mL(0 ng/mL calibrator is Emit®Calibrator/Control Level 0)

Comparison of Control Similarities

AttributesPredicate DeviceCEDIA® BuprenorphineNegative Control andPositive Control (K040316)Proposed DeviceEmit® II Plus SpecialtyDrug Control Negativeand Control Positive
Similarities
Intended UseThe CEDIA® Buprenorphinecontrols are used to qualifythe CEDIA® BuprenorphineAssay in human urine.Same
MatrixHuman urineSame
AnalyteBuprenorphineSame
TargetConcentrations forBuprenorphineHigh Control 7 ng/mLLow Control 3 ng/mLSame
PreparationLiquid, ready to useSame
Storage2 – 8ºCSame

7. Performance Data

Performance of the Emit® II Plus Buprenorphine Assay was evaluated at Siemens Healthcare Diagnostics Inc. on a Viva-E® Analyzer.

a. Precision / Cutoff Characterization

Studies were conducted using urine pools prepared by spiking Buprenorphine into drug-free human urine at 8 concentrations (levels). The concentrations tested represent the following levels relative to the 5 ng/mL cutoff: 100, -50%,

{8}------------------------------------------------

-40%, and -25% below the cutoff, at the cutoff, and +25%, +40%, +50%, and +100% of the cutoff. The drug-free human urine pool was also tested as the 0 ng/mL buprenorphine level. The studies were performed on Viva-E® analyzer. For each level, samples were analyzed in duplicate twice a day, for 20 days (N=80).

Urine Pool(ng/mL)% ofCutoff# ofResultsRepeatabilityResultsWithin-LaboratoryResults
O-100%8080 Negative80 Negative
2.5-50%8080 Negative80 Negative
3-40%8080 Negative80 Negative
3.75-25%8080 Negative80 Negative
5cutoff8025 Negative/55 Positive25 Negative/55 Positive
6.25+25%8080 Positive80 Positive
7+40%8080 Positive80 Positive
7.5+50%8080 Positive80 Positive
10+100%8080 Positive80 Positive

Precision: Qualitative Analysis

Precision: Semi-quantitative Analysis

Urine Pool(ng/mL)% ofCutoff# ofResultsRepeatabilitvResultsWithin-LaboratoryResults
0-100%8080 Negative80 Negative
2.5-50%8080 Negative80 Negative
3-40%8080 Negative80 Negative
3.75-25%8080 Negative80 Negative
5cutoff8025 Negative/55 Positive25 Negative/55 Positive
6.25+25%8080 Positive80 Positive
7+40%8080 Positive80 Positive
7.5+50%8080 Positive80 Positive
10+100%8080 Positive80 Positive

An additional precision study was performed using urine pools prepared by spiking Buprenorphine into drug-free human urine at one concentration level

{9}------------------------------------------------

relative to the 5 ng/mL cutoff: -75% below the cutoff and +75% above the cutoff. The studies were performed on Viva-E® analyzer. The samples were analyzed in duplicate, 40 times for a total of 80 replicates over one day.

Urine Pool(ng/mL)% ofCutoff# ofResultsRepeatabilityResultsWithin-LaboratoryResults
1.25-75%8080 Negative80 Negative
8.75+75%8080 Positive80 Positive

Precision: Qualitative Analysis

Precision: Semi-Quantitative Analysis

Urine Pool(ng/mL)% ofCutoff# ofResultsRepeatabilityResultsWithin-LaboratoryResults
1.25-75%8080 Negative80 Negative
8.75+75%8080 Positive80 Positive

b. Specificity and Cross-reactivity

Buprenorphine Metabolite Recovery - Buprenorphine and the buprenorphine metabolites norbuprenorphine, buprenorphine glucuronide and norbuprenorphine glucuronide were spiked into aliquots of drug free urine at the levels shown and run at n=5 replicates. The samples were assayed and the mean recovery results were determined.

Buprenorphine and Buprenorphine Metabolite Recovery

CompoundConcentrationTested(ng/mL)Semi-quantitativeResults
Recovery(ng/mL)% Cross-reactivity
Buprenorphine55.2103.2
Norbuprenorphine54.692.6
BuprenorphineGlucuronide10000.90.1
NorbuprenorphineGlucuronide10001.20.1

{10}------------------------------------------------

Structurally Related Compounds - Samples were prepared by spiking drug-free human urine with individual cross-reactants to the targeted level. The samples were evaluated on the Viva-E® analyzer. All samples were tested in replicates of n=5.

For qualitative analysis, cross-reactivity is reported as the level of crossreactant that produces a rate response ≥ the cutoff. For semi-quantitative analysis the percent cross-reactivity was calculated.

Qualitative ResultsSemi-quantitative Results
CompoundConc. Tested (ng/mL)Qualitative ResultSemi-Quantitative Result%Cross-reactivity
6-acetylcodeine100000NegativeNegative<0.01
6-acetylmorphine100000NegativeNegative<0.01
Codeine100000NegativeNegative<0.01
Dextromethorphan100000NegativeNegative<0.01
Dihydrocodeine100000NegativeNegative<0.01
Ethyl Morphine100000NegativeNegative<0.01
Heroin100000NegativeNegative<0.01
Hydrocodone100000NegativeNegative<0.01
Hydromorphone100000NegativeNegative<0.01
Levorphanol100000NegativeNegative<0.01
Morphine100000NegativeNegative<0.01
Morphine 3-glucuronide100000NegativeNegative<0.01
Morphine 6-glucuronide100000NegativeNegative<0.01
Nalorphine100000NegativeNegative<0.01
Naloxone100000NegativeNegative<0.01
Naltrexone100000NegativeNegative<0.01
Norcodeine100000NegativeNegative<0.01
Normorphine100000NegativeNegative<0.01
Noroxycodone100000NegativeNegative<0.01
Noroxymorphone100000NegativeNegative<0.01
Oxycodone100000NegativeNegative<0.01
Oxymorphone100000NegativeNegative<0.01

Cross-reactivity with the Structurally Related Compound

{11}------------------------------------------------

c. Interference

Structurally Unrelated Compounds - Interference from the structurally unrelated compounds was evaluated by spiking each compound into a 3 ng/mL and 7 ng/mL spiked urine pool. A control (blank) was also prepared without the interferent compound. Samples were evaluated on the Viva-E® analyzer. The blank sample was run, followed by the sample containing the interfering compound. All samples were tested in replicates of N=5. The mean rate and concentration were calculated for qualitative and semiquantitative analysis, respectively. The sample containing the drug was compared to the control, which contained only buprenorphine (3 or 7 ng/mL) to determine if there was a false response relative to the assay cutoff (5 ng/mL). At the stated concentration, the samples did not give a false response relative to the 5 ng/mL cutoff.

- 40% Cutoff (3 ng/mL)+40% Cutoff (7 ng/mL)
Interference(StructurallyUnrelatedCompounds)Conc.Tested(ug/mL)QualitativeResultSemi-quantitativeResultQualitativeResultSemi-quantitativeResult
10. 11-dihydrocarbamazepine85NegativeNegativePositivePositive
Acetaminophen1000NegativeNegativePositivePositive
Acetylsalicylic Acid1500NegativeNegativePositivePositive
Amitriptyline100NegativeNegativePositivePositive
Amoxicillin500NegativeNegativePositivePositive
ZT (Zidovudine)2000NegativeNegativePositivePositive
Benzoylecgonine1000NegativeNegativePositivePositive
Brompheniramine75NegativeNegativePositivePositive
Caffeine1000NegativeNegativePositivePositive
Captopril500NegativeNegativePositivePositive
Chlordiazepoxide100NegativeNegativePositivePositive
Chlorpromazine10NegativeNegativePositivePositive
Cimetidine1000NegativeNegativePositivePositive
Clomipramine2.5NegativeNegativePositivePositive
Clonidine1000NegativeNegativePositivePositive
Cyclobenzaprine125NegativeNegativePositivePositive
d-amphetamine700NegativeNegativePositivePositive
Desipramine800NegativeNegativePositivePositive
Diazepam100NegativeNegativePositivePositive
Digoxin0.01NegativeNegativePositivePositive
Diphenhydramine1000NegativeNegativePositivePositive
d-methamphetamine500NegativeNegativePositivePositive
Doxepine100NegativeNegativePositivePositive
EDDP1000NegativeNegativePositivePositive
EMDP100NegativeNegativePositivePositive
Enalapril500NegativeNegativePositivePositive
Fluoxetine500NegativeNegativePositivePositive
Glutethimide500NegativeNegativePositivePositive
Haloperidol100NegativeNegativePositivePositive
Hydroxyzine500NegativeNegativePositivePositive
Ibuprophen1000NegativeNegativePositivePositive
Imipramine200NegativeNegativePositivePositive
Ketamine100NegativeNegativePositivePositive
KetorolacTromethamine400NegativeNegativePositivePositive
LAAM (L-a-acetylmethadol)25NegativeNegativePositivePositive
L-Cotinine100NegativeNegativePositivePositive
Levofloxacin100NegativeNegativePositivePositive
Levothyroxine (L-Thyroxine)50NegativeNegativePositivePositive
Lidocaine1000NegativeNegativePositivePositive
Lormetazepam1NegativeNegativePositivePositive
LSD10NegativeNegativePositivePositive
MDMA (Ecstasy)1000NegativeNegativePositivePositive
Meperidine800NegativeNegativePositivePositive
Methadone500NegativeNegativePositivePositive
Methaqualone600NegativeNegativePositivePositive
NAPA400NegativeNegativePositivePositive
Naproxen1000NegativeNegativePositivePositive
Nicotinic Acid500NegativeNegativePositivePositive
Nifedipine500NegativeNegativePositivePositive
Nordiazepam100NegativeNegativePositivePositive
Nortryptiline250NegativeNegativePositivePositive
Oxazepam300NegativeNegativePositivePositive
Perphenazine150NegativeNegativePositivePositive
Phencyclidine900NegativeNegativePositivePositive
Phenobarbital500NegativeNegativePositivePositive
Phenelzine100NegativeNegativePositivePositive
Phenytoin1000NegativeNegativePositivePositive
Procainamide1000NegativeNegativePositivePositive
Procyclidine800NegativeNegativePositivePositive
Promethazine100NegativeNegativePositivePositive
Propoxyphene1000NegativeNegativePositivePositive
Protriptyline200NegativeNegativePositivePositive
Pseudoephedrine1000NegativeNegativePositivePositive
Quinacrine900NegativeNegativePositivePositive
Ranitidine1000NegativeNegativePositivePositive
Ritalin1000NegativeNegativePositivePositive
Salicylic Acid500NegativeNegativePositivePositive
Scopolamine500NegativeNegativePositivePositive
Secobarbital1000NegativeNegativePositivePositive
Tapentadol100NegativeNegativePositivePositive
THC100NegativeNegativePositivePositive
Thioridazine100NegativeNegativePositivePositive
Tramadol1000NegativeNegativePositivePositive
Trazodone5NegativeNegativePositivePositive
Trimethoprim1000NegativeNegativePositivePositive
Triprolidine (zymine)50NegativeNegativePositivePositive
Tyramine100NegativeNegativePositivePositive
Verapamil500NegativeNegativePositivePositive
Zolpidem100NegativeNegativePositivePositive

Interference of Structurally Unrelated Compounds

{12}------------------------------------------------

{13}------------------------------------------------

Endogenous Substances Interference - Each compound was spiked into a -40% cutoff and a +40% cutoff which was prepared by spiking buprenorphine to aliquots of drug-free human urine. The results show that the endogenous substances, pH and sG caused no interference relative to the 5 ng/mL cutoff. Qualitative and semi-quantitative results are provided.

-40% Cutoff (3 ng/mL)+40% Cutoff (7 ng/mL)
Interferences(EndogenousSubstances)Conc.TestedQualitativeResultSemi-QuantitativeResultQualitativeResultSemi-QuantitativeResult
Acetone1.0 g/dLNegativeNegativePositivePositive
Ascorbic Acid1.5 g/dLNegativeNegativePositivePositive
Conjugated Bilirubin2.0 mg/dLNegativeNegativePositivePositive
Unconjugated Bilirubin2.0 mg/dLNegativeNegativePositivePositive
Creatinine0.5 g/dLNegativeNegativePositivePositive
Ethanol1.0 g/dLNegativeNegativePositivePositive
Immuno GammaGlobulin (IgG)0.5 g/dLNegativeNegativePositivePositive
Glucose2.0 g/dLNegativeNegativePositivePositive
Galactose1.0 g/dLNegativeNegativePositivePositive
Hemoglobin115 mg/dLNegativeNegativePositivePositive
Human Serum Albumin0.5 g/dLNegativeNegativePositivePositive
Oxalic Acid0.1 g/dLNegativeNegativePositivePositive
Riboflavin7.5 mg/dLNegativeNegativePositivePositive
Sodium Chloride6.0 g/dLNegativeNegativePositivePositive
Urea6.0 g/dLNegativeNegativePositivePositive
Sodium Azide1% w/vNegativeNegativePositivePositive
Sodium Fluoride1% w/vNegativeNegativePositivePositive

{14}------------------------------------------------

PotentialInterference(pH)-40% Cutoff (3 ng/mL)QualitativeResultSemi-QuantitativeResult+40% Cutoff (7 ng/mL)QualitativeResultSemi-QuantitativeResult
pH 3NegativeNegativePositivePositive
pH 4NegativeNegativePositivePositive
pH 5NegativeNegativePositivePositive
pH 6NegativeNegativePositivePositive
pH 7NegativeNegativePositivePositive
pH 8NegativeNegativePositivePositive
pH 9NegativeNegativePositivePositive
pH 11NegativeNegativePositivePositive
PotentialInterference(Specific Gravity)-40% Cutoff (3 ng/mL)Qualitative Result-40% Cutoff (3 ng/mL)Semi-Quantitative Result+40% Cutoff (7 ng/mL)Qualitative Result+40% Cutoff (7 ng/mL)Semi-Quantitative Result
sG 1.002NegativeNegativePositivePositive
sG 1.005NegativeNegativePositivePositive
sG 1.010NegativeNegativePositivePositive
sG 1.015NegativeNegativePositivePositive
sG 1.020NegativeNegativePositivePositive
sG 1.025NegativeNegativePositivePositive
sG 1.030NegativeNegativePositivePositive
sG 1.035NegativeNegativePositivePositive

d. Recovery/Linearity

Drug free urine pools were spiked with nine concentrations of buprenorphine at levels 2-25 ng/mL and analyzed semi-quantitatively in replicates of N= 5 on a Viva-E® analyzer. The mean observed Buprenorphine concentration was compared to the expected Buprenorphine concentration and percent recovery calculated.

ExpectedBuprenorphineConcentration(ng/mL)Mean BuprenorphineConcentration by Emit®Il Plus BuprenorphineAssay (ng/mL)% Recovery
22.1105.0
33.1103.3
43.997.5
55.0100.0
87.796.3
1211.192.5
1817.798.3
2221.095.5
2523.995.6

{15}------------------------------------------------

e. Method Comparison

Split sample method comparison was conducted with one-hundred twenty seven (127) unaltered human urine samples, which were analyzed by the Emit® II Plus Buprenorphine Assay on a Viva-E® Analyzer vs. LC/MS/MS. The results are presented below:

LC/MS/MS%Agreement
Negative(<2.5 ng/mL)Negative Within 50%below thecutoff (2.5-4.9 ng/mL)Positive Within 50%above thecutoff (5.0-7.5 ng/mL)Positive(>7.5 ng/mL)
Qualitative
Emit® Positive07164990%
Emit® Negative4591098%
Semi-quantitative
Emit® Positive07164990%
Emit® Negative4591098%

Emit® II Plus Buprenorphine Assay vs. LC/MS/MS

Comparison Table for Qualitative and Semi-quantitative Assay Performance

LC/MS/MS
+-
Emit® II PlusBuprenorphine+657
-154

Discordant Result Summary

SampleIDEmit® II PlusBuprenorphine(ng/mL)LC/MS/MSEmit+/-LC/MS/MS+/-
Bup(ng/mL)NorBup(ng/mL)Bup+NorBup(ng/mL)
1905.603.923.92+-
1936.104.974.97+-
1955.404.064.06+-
2265.804.214.21+-
2505.104.134.13+-
775.504.604.60+-
3165.703.863.86+-
3384.35.1205.12-+

Bup = Buprenorphine; NorBup = Norbuprenorphine

{16}------------------------------------------------

Calibrator/Control and Cutoff Control Traceability f.

The Emit® II Plus Specialty Drug Calibrator/Control and Emit® II Plus Specialty Drug Control Neqative and Control Positive are traceable to Cerilliant Buprenorphine Cat. No. B-902. When stored refrigerated at 2-8°C the calibrators/controls are stable opened or unopened until the expiration date printed on the vial label.

8. Conclusion

The information provided in this premarket notification demonstrates the Emit® II Plus Buprenorphine Assay, Emit® II Plus Specialty Drug Calibrator / Control, Emit® II Plus Specialty Drug Control Negative and Control Positive are substantially equivalent to the legally marketed predicate devices for their general intended use. Substantial equivalence was demonstrated through comparison of intended use and technological features to the commercially available predicate devices and confirmed by liquid chromatography/ tandem mass spectrometry (LC/MS/MS), a reference analytical method. The information in this pre-market notification provides reasonable assurance that the Emit® II Plus Buprenorphine Assay. Emit® II Plus Specialty Drug Calibrator / Control, Emit® II Plus Specialty Drug Control Negative and Control Positive demonstrate safety and effectiveness similar to the predicate devices - Microgenics CEDIA® Buprenorphine Assay, CEDIA® Buprenorphine Calibrators, and CEDIA® Buprenorphine Controls (cleared under K040316).

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).