K Number
K150557

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-10-08

(218 days)

Product Code
Regulation Number
890.5050
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Podimetrics Remote Temperature Monitoring System™ (RTM System™) is intended to be used by a patient in conjunction with a healthcare professional or caretaker for periodic evaluation of the temperature over the soles of the feet for signs of inflammation. It will provide information indicating when the patient and healthcare provider should communicate for further evaluation and treatment regarding any persistent localized inflammation observed on the feet via the electronic sensing system and remote visualization of its data. The Podimetrics RTM System™ is intended to be used under the direction of a healthcare professional as an adjunct to, and not in replacement of, self-examination and periodic foot care and examination conducted by a healthcare professional and does not diagnose any specific disease state.

Device Description

The Podimetrics RTM System™ is designed to help a patient and healthcare professional periodically evaluate the temperature over the soles of the feet for signs of inflammation. This evaluation may be helpful for determining when the patient and healthcare provider should communicate for physical examinations or care in addition to normally scheduled visits.

The system is comprised of the Podimetrics Mat and the Thermogram Explorer™ internet application. The Podimetrics Mat is about the same size and shape as a floor mat and placed on the floor of the patient's home. The patient stands on the mat once per day for about 20 seconds to collect a thermal scan of the soles of the feet. The thermal scan is recorded electronically via 2-dimensional array of temperature-sensitive resistors covering the top surface of the mat and the data is transmitted by an internal cellular modem to a remote server via cellular network.

The thermal scan data is processed and stored on the remote server and is displayed in Thermogram Explorer™, an internet application. The patient and healthcare professional can use Thermogram Explorer™ to access the thermograms and evaluate them for signs of inflammation. If any signs are observed, the patient can see the healthcare professional for a physical examination.

AI/ML Overview

The Podimetrics Remote Temperature Monitoring System™ underwent a series of bench tests to demonstrate its safety and effectiveness.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance CriteriaReported Device Performance
AccuracyAccurate within 1°F over the range 59 to 104°F, with a precision of 0.5°F and a temperature resolution of 0.2°F.
Scan TimeApproximately 20 seconds to complete each thermal scan.
Battery LifeGreater than one month of daily use.
Battery SafetyMet test requirements for protection against overheating, over-discharging, and over-charging.
Service LifeGreater than 2 years of daily use (demonstrated through cyclic loading, battery charge/discharge cycles, and cleaning tests).
Electrical SafetyMet all requirements of IEC 60601-1 (3rd edition), 60601-1-2 (3rd edition), and 60601-1-11 (1st edition) for electromagnetic compatibility and electrical safety in the home healthcare environment.
Wireless Data TransmissionSecure and without loss of data. Data saved locally and transmitted upon reconnection if cellular network is inaccessible.
Human Factors (Mat Use)No patients reported problems with setup, thermal scans, or charging during a 14-day in-home study. No issues identified.
Human Factors (Web Application)Users verified to display thermograms for a particular patient and compare temperatures between two locations.
BiocompatibilityPatient-contacting material met individual test requirements and is considered biocompatible for intended use.
Floor SafetyTop and bottom surfaces met non-slip properties test criterion and are not considered a trip hazard.
ShippingDevice maintained performance after shipping tests.

2. Sample size used for the test set and the data provenance:

  • Human Factors Study (Mat Use): A "14-day in-home human factors study" was conducted. The document does not specify the exact number of participants, but it implies a sample of patients. No specific country of origin is mentioned, but "in-home" suggests users in typical environments. The study was prospective in nature.
  • Other tests (Accuracy, Scan Time, Battery Life, Service Life, Electrical Safety, Wireless Data Transmission, Biocompatibility, Floor Safety, Shipping): These appear to be bench testing or laboratory-based tests. The document does not specify sample sizes for these tests, which is typical for engineering and material performance assessments rather than clinical trials. The data provenance is laboratory testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • The document does not mention the use of experts to establish ground truth for the performance characteristics. Most of the tests are objective engineering or physical performance measurements (e.g., temperature accuracy, battery life, electrical safety standards).
  • For the human factors studies, users themselves were the evaluators of their experience, not experts establishing a medical ground truth.

4. Adjudication method for the test set:

  • No adjudication method is described, as the tests primarily involve objective measurements or user feedback rather than subjective interpretations requiring expert consensus.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC comparative effectiveness study was done. The device is a monitoring system and not primarily an AI algorithm for diagnosis or interpretation that would typically require such a study to demonstrate improvement in human reader performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • The document describes the device as a "Remote Temperature Monitoring System™" and an "adjunct to, and not in replacement of, self-examination and periodic foot care and examination conducted by a healthcare professional." While the system processes and visualizes data, it is not presented as a standalone diagnostic AI algorithm. The performance characteristics focus on the accuracy and reliability of the physical device and its data transmission, not the diagnostic accuracy of an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • The ground truth for most of the performance characteristics (e.g., accuracy, battery life, electrical safety) is based on established engineering and physical measurement standards.
  • For the human factors studies, the "ground truth" was user experience and their ability to use the device as intended, reported directly by the users. There is no mention of medical ground truth (like pathology or clinical outcomes) being established for these specific performance tests, as the device's main function is data acquisition and presentation, not medical diagnosis.

8. The sample size for the training set:

  • The document does not describe an AI algorithm or machine learning component that would require a distinct "training set" in the context of typical AI device development. The device's function is explained as an electronic sensing system that records and displays thermal scan data.

9. How the ground truth for the training set was established:

  • As no training set for an AI algorithm is described, this question is not applicable based on the provided text.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure in profile, with three faces overlapping each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 8, 2015

Podimetrics, Inc. Jonathan Bloom, MD CEO and Co-Founder 231 R Holland Street Somerville, Massachusetts 02144

Re: K150557

Trade/Device Name: Podimetrics Remote Temperature Monitoring System™ Regulation Number: 21 CFR 890.5050 Regulation Name: Daily Activity Assist Device Regulatory Class: I Product Code: OIZ Dated: September 1, 2015 Received: September 2, 2015

Dear Dr. Bloom:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Tina Kiang -S

for Erin I. Keith, MS Director Division of Dental. Anesthesiology, General Hospital, Respiratory, and Infection Control Devices Office of Device Evaluation Center for Devices and Radiological Devices

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K150557

Device Name

Podimetrics Remote Temperature Monitoring System

Indications for Use (Describe)

The Podimetrics Remote Temperature Monitoring System™ (RTM System™) is intended to be used by a patient in conjunction with a healthcare professional or caretaker for periodic evaluation of the temperature over the soles of the feet for signs of inflammation. It will provide information indicating when the patient and healthcare provider should communicate for further evaluation and treatment regarding any persistent localized inflammation observed on the feet via the electronic sensing system and remote visualization of its data. The Podimetrics RTM System™ is intended to be used under the direction of a healthcare professional as an adjunct to, and not in replacement of, self-examination and periodic foot care and examination conducted by a healthcare professional and does not diagnose any specific disease state.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

008 Summary

Image /page/3/Picture/3 description: The image shows the logo for Podimetrics. The logo consists of a circular emblem with two white footprints inside, along with a shield in the center of the footprints. To the right of the emblem, the word "PODIMETRICS" is written in a sans-serif font.

510(k) Summary for the Podimetrics Remote Temperature Monitoring System™

8.1 Submitter/510(k) Holder

Podimetrics, Inc. 49 Day St., Suite A Somerville, MA 02144

Contact: Jonathan Bloom (p) 617-990-6597 (f) 617-507-3496 jon@podimetrics.com

Date Prepared: February 23, 2015

8.2 Device Name

Trade Name: Podimetrics Remote Temperature Monitoring System™ Common Name: Foot Temperature Monitor Classification Name: Daily Activity Assist Device, (Product Code OIZ, 21 CFR 890.5050) Device Classification: Class I

8.3 Predicate Device

Visual Footcare Technologies TempStat™ (K080816, Product Code OIZ)

8.4 Indications for Use

The Podimetrics Remote Temperature Monitoring System™ (RTM System™) is intended to be used by a patient in conjunction with a healthcare professional or caretaker for periodic evaluation of the temperature over the soles of the feet for signs of inflammation. It will provide information indicating when the patient and

{4}------------------------------------------------

healthcare provider should communicate for further evaluation and treatment regarding any persistent localized inflammation observed on the feet via the electronic sensing system and remote visualization of its data. The Podimetrics RTM System™ is intended to be used under the direction of a healthcare professional as an adjunct to, and not in replacement of, self-examination and periodic foot care and examination conducted by a healthcare professional and does not diagnose any specific disease state.

8.5 Device Description

The Podimetrics RTM System™ is designed to help a patient and healthcare professional periodically evaluate the temperature over the soles of the feet for signs of inflammation. This evaluation may be helpful for determining when the patient and healthcare provider should communicate for physical examinations or care in addition to normally scheduled visits.

The system is comprised of the Podimetrics Mat and the Thermogram Explorer™ internet application. The Podimetrics Mat is about the same size and shape as a floor mat and placed on the floor of the patient's home. The patient stands on the mat once per day for about 20 seconds to collect a thermal scan of the soles of the feet. The thermal scan is recorded electronically via 2-dimensional array of temperaturesensitive resistors covering the top surface of the mat and the data is transmitted by an internal cellular modem to a remote server via cellular network.

The thermal scan data is processed and stored on the remote server and is displayed in Thermogram Explorer™, an internet application. The patient and healthcare professional can use Thermogram Explorer™ to access the thermograms and evaluate them for signs of inflammation. If any signs are observed, the patient can see the healthcare professional for a physical examination.

8.6 Technological Characteristics and Substantial Equivalence

The Podimetrics RTM System is substantially equivalent to the Visual Footcare Technologies, LLC TempStat™. Both devices include a sensor pad upon which the feet are placed to measure the temperature across the plantar surface of the feet.

The TempStat™ uses liquid crystals embedded in a polycarbonate plastic which change color corresponding to the skin temperature they are in contact with. The Podimetrics RTM System™ uses electronic temperature-sensitive resistors encased in a polyurethane elastomer which record the temperature applied and display it on a computer. Although the technological method differs, both devices generate a 2dimensional heat map, or thermogram, of the feet which can be used to evaluate for signs of inflammation in the same manner. Both devices produce thermograms that are evaluated by the patient, and the Podimetrics RTM System™ can additionally be accessed by the healthcare provider for remote evaluation. This may lead to more

{5}------------------------------------------------

frequent communication and physical examination by the healthcare professional compared to the standard of care.

The differences in the intended use between the Podimetrics RTM System™ and the TempStat™ are minor and do not affect the device's intended use.

The following table contains a summary of the descriptive and technological characteristics of the Podimetrics RTM System™ compared to the predicate device.

CharacteristicTempStat™(K080816)Podimetrics RTMSystem™Discussion ofequivalency
ManufacturerVisual FootcareTechnologiesPodimetrics
Product CodeOIZOIZSame
CFR890.5050890.5050Same
Intended UseFoot examination toolused for potentialinflammatory changesFoot examination toolused for potentialinflammatory changesSame
CharacteristicTempStat™(K080816)Podimetrics RTMSystem™Discussion ofequivalency
Indications foruseTempStat is a daily activity assist device that is intended to assist a patient to visualize the image and condition of the soles of their feet in connection with a daily self-examination for signs of inflammation on the sole of the foot. TempStat will assist the patient in determining when they should contact their healthcare provider for further diagnosis and treatment regarding any skin changes seen by a mirror and/or highlighted by the liquid crystal foot pads. TempStat is intended to be used as an adjunct to, and not in replacement of, periodic foot care and examinations conducted by a health care professional and used under the direction of a health care professional and does not diagnose any specific disease state.The Podimetrics Remote Temperature Monitoring (RTM) System™ is intended to be used by a patient for periodic evaluation of the temperature over the soles of the feet for signs of inflammation in conjunction with a healthcare professional or caretaker. It will assist in determining when the patient and healthcare provider should communicate for further diagnosis and treatment regarding any persistent localized inflammation observed on their feet via the electronic sensing system and remote visualization of its data. The Podimetrics RTM System is intended to be used under the direction of a healthcare professional as an adjunct to, and not in replacement of, self-examination and periodic foot care and examination conducted by a healthcare professional and does not diagnose any specific disease state.Similar. Whereas the predicate device displays the thermal data qualitatively in a color-map on the device, the Podimetrics RTM System measures temperature electronically and displays the same type of data quantitatively for evaluation by the patient and healthcare provider through a web application.
Prescription UseThe TempStat™ is for prescription only.The Podimetrics RTM System™ is for prescription only.Same
TargetpopulationMarketed for diabetics, especially those with neuropathy and risk for foot ulcers or other inflammatory conditions.Marketed for diabetics, especially those with neuropathy and risk for foot ulcers or other inflammatory conditions.Same
CharacteristicTempStat™(K080816)Podimetrics RTMSystem™Discussion ofequivalency
Anatomical siteThe plantar aspect of the feet.The plantar aspect of the feet.Same
Where usedThe patient's home.The patient's home.Same
PerformanceLiquid crystal pads areaccurate to within 1°Fwith respect to colorchanges.Electronic sensors areaccurate to within 1°F.Same
Data formatProvides a color-codedthermogram on thesurface of the device.Provides a color-codedand quantitativethermogram with actualtemperature in the webapplication.Both devices generateand display a color-coded thermogram ofthe feet. ThePodimetrics RTMSystem™additionally displaysthe temperaturequantitatively inThermogramExplorer™.
Energy usedand/or deliveredNo energy is used ordelivered.The Podimetrics Mat ispowered by arechargeable battery butdoes not deliver anyenergy.Neither devicedelivers any energy tothe patient.
Human FactorsThe user places their feeton the TempStat sensingpads for 60 secondswhile sitting.The user stands on thePodimetrics Mat for 20seconds.Both devices areplaced on the floorand require the userto place their feet onthe top surface for ashort measurement.
Materials andBiocompatibilityFoot contacting material:Liquid crystals encasedwithin polycarbonateplastic.Foot contacting material:Polyurethane film withsilver ion additiveAlthough the specificmaterials differbetween thePodimetrics andpredicated device, thepatient-contactingmaterial of thePodimetrics producthas been tested inaccordance with ISO10993 parts 5, 10, and11.
CharacteristicTempStat™(K080816)Podimetrics RTMSystem™Discussion ofequivalency
Compatibilitywith theenvironment andother devicesTempStat™ includes noelectronics or chemicalsso there are nocompatibility concerns.The Podimetrics Matincludes electronics toperform thermal scansand which use RFwireless technology totransmit the data.The Podimetrics Mathas been tested forEMC in accordancewith IEC 60601-1-2and is subject to FCCcertification beforemarketing.
Electrical SafetyThe TempStat™includes no electronics.The Podimetrics Matincludes electronics fortemperature sensing,user interface, datastorage, and datatransmission.The Podimetrics Mathas been tested forelectrical safety inaccordance with IEC60601-1 and IEC60601-1-11.
SoftwareThe TempStat™includes no software.The Podimetrics RTMSystem™ uses softwareto generate, store,process, and visualizethermometric data.The PodimetricsRTM System™includes softwarewhich adheres to"General Principles ofSoftware Validation;Final Guidance forIndustry and FDAStaff" issued onJanuary 11, 2002.

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

8.7 Performance Characteristics of the Podimetrics RTM System™

The Podimetrics RTM System™ underwent a series of bench testing to demonstrate it is as safe and effective as the predicate for its intended use. Testing included:

  • Accuracy: The Podimetrics RTM System™ is accurate within 19F over ● the range 59 to 104°F, with a precision of 0.5°F and a temperature resolution of 0.2°F.
  • Scan time: Each thermal scan requires approximately 20 seconds to . complete using the Podimetrics Mat as demonstrated through laboratory bench testing and the human factors testing performed.
  • Battery life: The Podimetrics Mat is cordless, has a battery life of greater . than one month of daily use.

{9}------------------------------------------------

  • . Battery safety: The rechargeable battery was tested for protection against overheating, over-discharging, and over-charging and met test requirements.
  • Service life: The service life of the Podimetrics Mat is greater than 2 years . of daily use as demonstrated through cyclic loading tests, battery charge/discharge cycles, and cleaning tests.
  • Electrical safety: The Podimetrics Mat has been tested for electromagnetic . compatibility and electrical safety in the home healthcare environment in accordance with IEC 60601-1 (3rd edition), 60601-1-2 (3rd edition), and 60601-1-11 (1st edition). The device met all requirements of the testing.
  • · Wireless data transmission: The Podimetrics RTM System™ transmits data wirelessly securely and without loss of data. In the event of an inaccessible cellular network, the data is saved locally and transmitted upon reconnection.
  • Human Factors: Podimetrics has conducted a 14-day in-home human . factors study to assess the ability of users to use the Podimetrics Mat at home. No patients reported any problems setting up the device, taking and transmitting thermal scans, or charging the batteries. No issues were identified. Podimetrics has also conducted a human factors study for the Thermogram Explorer web application to verify users can display a thermogram for a particular patient and compare temperatures between two locations on the feet.
  • Biocompatibility: The patient-contacting material of the Podimetrics Mat ● has been tested for biocompatibility for the type of contact and duration of contact. The testing demonstrated that the patient contacting material met the individual test requirements and is considered biocompatible for the product's intended use.
  • Floor safety: The top and bottom surface of the Podimetrics Mat were ● tested for non-slip properties. The product met the test criterion and is not considered a trip hazard.
  • Shipping: The Podimetrics Mat packaging was tested to ensure the device ● maintains performance after shipping. The device functioned as intended post ship testing.

All testing performed demonstrates that the Podimetrics RTM System™ is as safe and effective as the predicate as an aid in detecting inflammation on the soles of a patient's foot.

{10}------------------------------------------------

8.8 Conclusion

Podimetrics believes that, based on the descriptive information and testing provided in this submission, the Podimetrics RTM System™ is substantially equivalent to its predicate device.

§ 890.5050 Daily activity assist device.

(a)
Identification. A daily activity assist device is a modified adaptor or utensil (e.g., a dressing, grooming, recreational activity, transfer, eating, or homemaking aid) that is intended for medical purposes to assist a patient to perform a specific function.(b)
Classification. Class I (general controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 890.9. If the device is not labeled or otherwise represented as sterile, the device is also exempt from the current good manufacturing practice requirements of the quality management system regulation in part 820 of this chapter, except for requirements concerning records and complaint files under § 820.35 of this chapter.