K Number
K150505

Validate with FDA (Live)

Device Name
Cellfina System
Manufacturer
Date Cleared
2015-07-29

(154 days)

Product Code
Regulation Number
878.4790
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cellfina™ System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits up to 2 vears of observation.

Device Description

The Cellfina System is intended to provide precise focal release of subcutaneous tissue for improvement in the appearance of cellulite. The release of subcutaneous tissue for improvement in the appearance of cellulite is a minimally invasive surgical technique by physicians using of manual surgical instruments and accessories. The Cellfina device consists of a powered cutting blade and a means for instrument guidance to control the depth, size and shape of the tissue release. The system consists of a sterile single use disposable kit and a non-sterile reusable motor module.

AI/ML Overview

The provided text describes specific acceptance criteria and performance data for the Cellfina™ System, a device intended for the improvement of the appearance of cellulite.

Here's a breakdown of the requested information:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Primary Endpoints)Reported Device Performance (2-year follow-up)
Achievement of ≥1 point average reduction in the 0-5 point Cellulite Severity Scale as determined by independent physician assessment of subject photographs taken before and 2 years after treatment.The average improvement was 2.0 points (p<0.0001) with the 97.5% confidence limit of a 1.8 point improvement. (Met)
Improvement of one grade or more in severity (none, mild, moderate, severe) in >60% of treated subjects as determined by independent physician assessment of subject photographs taken before and 2 years after treatment.88.5% of treated subjects had improvement of 1 grade with a 95% confidence limit of 79.2%. (Met)
Freedom from Serious Adverse Events (SAE) directly attributable to the Cellfina™ System or procedure.100% of subjects were free from Serious Adverse Events (SAE) directly attributable to the Cellfina™ System or procedure. (Met)

Additional Secondary Measures Achieved:

  • Correct Selection by Independent Physician Assessment: Average rate of correct selection of blinded subject photographs (before and two years after treatment) was 98.1% (95% confidence limit of 94.5%).
  • Noticeable Improvement (GAIS) and Marked Improvement: 100% of subjects had noticeable improvement by the Global Aesthetic Improvement Scale (GAIS). 52% were characterized as having marked improvement or better by independent physician assessment.
  • Subject Satisfaction: 96% of subjects were either satisfied or very satisfied at the 2-year follow-up.
  • Subject Reported Pain: Average pain on a 0-10 numerical rating scale was 4.5 for anesthesia delivery and 3.7 for tissue release. At follow-ups, 71% of subjects rated pain ≤ 3 at 3 days, and over 95% rated pain ≤ 3 thereafter. No subjects reported pain at the 2-year follow-up.

2. Sample size used for the test set and the data provenance

  • Sample Size for Test Set: The document indicates that "All subjects served as their own control" and mentions endpoints being achieved for "all subjects." However, the exact number of subjects in the pivotal clinical study (test set) is not explicitly stated in the provided text.
  • Data Provenance: The study was a "prospective, multi-center, non-randomized open label, safety and effectiveness study." The location (country of origin) of the study centers is not specified, but it was conducted under IDE G120116.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: An "independent and blinded reviewer of the photographs" was used to verify effectiveness, and "independent physician assessment" determined the reduction in cellulite severity and grade improvement. The exact number of independent physicians/reviewers is not specified.
  • Qualifications of Experts: The text refers to "independent physician assessment." No further details on their specific qualifications (e.g., specialty, years of experience) are provided.

4. Adjudication method for the test set

  • The document mentions an "independent and blinded reviewer" for photographs to verify effectiveness. For the primary and powered secondary endpoints, "independent physician assessment" was used. The text does not explicitly detail an adjudication method (e.g., 2+1, 3+1) if there were multiple independent assessments or if there were disagreements. It implies a single "independent physician assessment" for each metric.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No, a multi-reader multi-case (MRMC) comparative effectiveness study focusing on AI assistance for human readers was not mentioned. The study evaluated the effectiveness of the Cellfina System itself, not an AI assistant.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • No, a standalone (algorithm only) performance study was not mentioned. The Cellfina System is a physical device that is used by physicians, and the study evaluated its direct clinical performance on patients.

7. The type of ground truth used

  • The ground truth was established through expert consensus/assessment of clinical data. This included:
    • Independent physician assessment of subject photographs (before and 2 years after treatment) for cellulite severity and grade improvement.
    • Global Aesthetic Improvement Scale (GAIS) based on independent physician assessment.
    • Subject-reported satisfaction and pain levels.
    • Adverse event reporting.

8. The sample size for the training set

  • The text does not mention a separate "training set" in the context of device development or any machine learning/AI models. The clinical study described is a pivotal trial evaluating the device's safety and efficacy.

9. How the ground truth for the training set was established

  • As no "training set" (in the context of AI/machine learning) is described, this question is not applicable based on the provided text. The clinical study described served as the basis for demonstrating the device's effectiveness.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized human figure with three faces in profile, representing health and well-being. The figure is enclosed within a circular border that contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA".

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Ulthera Incorporated Ms. Suzon Lommel Vice President Regulatory and Quality Affairs 1840 South Stapley Drive, Suite 200 Mesa, Arizona 85204

Re: K150505 Trade/Device Name: Cellfina™ System Regulation Number: 21 CFR 878.4790 Regulation Name: Powered surgical instrument for improvement in the appearance of cellulite Regulatory Class: Class II Product Code: OUP Dated: June 24, 2015 Received: June 30, 2015

Dear Ms. Lommel:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

July 29, 2015

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Joshua C. Nipper -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S.

For Director

Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150505

Device Name Cellfina™ System

Indications for Use (Describe)

The Cellfina™ System is intended for long term improvement in the appearance of cellulite in the buttocks and thigh areas of adult females as supported by clinical data demonstrating no significant reduction in treatment benefits up to 2 vears of observation.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5. 510(k) Summary

This 510(k) Summary for the Cellfina™ System is submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990 and follows the Office of Device Evaluation (ODE) guidance concerning the organization and content of a 510(k) summary.

Applicant:Ulthera, Inc.
Address:1840 South Stapley DriveSuite 200Mesa, AZ 85204
Contact Person:Suzon Lommel, Vice President Regulatory and Quality Affairs
Telephone:(408) 645-4979
Fax:(480) 619-4071
Submission Date:February 20, 2015
Device Trade Name:Cellfina System
Common Name:Powered surgical instrument for improvement in the appearance ofcellulite.
Classification:Regulatory Class II
Classification Name:Powered surgical instrument for improvement in the appearance ofcellulite.
Regulation Number:21CFR 878.4790
Product Code:OUP
Legally Marketed Predicate:Cellfina System (trade name previously Cabochon System), 510(k):K134010
Applicable Guidance:The following guidance special controls are applicable to the CellfinaSystem:
Predicate DeviceK134010, Cellfina SystemSubject DeviceCellfina System
IntendedUse/Indications forUseThe Cellfina System is intended for longterm improvement in the appearanceof cellulite in the buttocks and thighareas of adult females as supported byclinical data demonstrating nosignificant reduction in treatmentbenefits up to 1 year of observation.The Cellfina System is intended for longterm improvement in the appearanceof cellulite in the buttocks and thighareas of adult females as supported byclinical data demonstrating nosignificant reduction in treatmentbenefits up to 2 years of observation.Clinical performance data provided forexpanded indication
Where usedClinic/Doctor's OfficeClinic/Doctor's Office
Same
Anatomical SiteButtocks and thighsButtocks and thighs
Type of EnergyMechanical movement driven by wall powered motor moduleMechanical movement driven by wall powered motor module
Biological EffectControlled mechanical cutting of the fibrous tissue (septae) which contributes to the appearance of cellulite.The Cellfina System controls and stabilizes the cutting motion used by physicians with the needle based release procedure.Controlled mechanical cutting of the fibrous tissue (septae) which contributes to the appearance of cellulite.The Cellfina System controls and stabilizes the cutting motion used by physicians with the needle based release procedure.
Location of Motor Module LabelAttached to the DC power cableAdheres to the bottom of the motor module
SterilizationDisposable Kit, CK1:Ethylene Oxide, SAL 10-6Motor Module, CM1: Sterile dome provided; use with disinfecting wipesNo change to contentDisposable Kit, CK1:Ethylene Oxide, SAL 10-6Motor Module, CM1: Sterile dome provided; use with disinfecting wipes
Shelf LifeDisposable Kit, CK1: 6 monthsMotor Module, CM1: Undefined usesDisposable Kit, CK1: 6 monthsMotor Module, CM1: 15 uses
PackagingDisposable Kit, CK1:• 3 Tyvek pouchesCleaning and Disinfection Data providedDisposable Kit, CK1:• 1 Thermoformed tray with Tyvek sealed outer tray
Inner shipping chipboard box Outer corrugated shipper box Motor Module, CM1: Packaged with disposable kitFoam insert within outer corrugated shipper box Outer corrugated shipper box Larger size to accommodate tray Motor Module, CM1: individually packagedPackaging Testing Provided
BiocompatibilityBiocompatibleBiocompatibleBiocompatibility Testing Provided
ElectromagneticCompatibilityStandardsCompatibleCompatible
Medical ElectricalEquipment SafetyStandardsCompatibleSameCompatibleMedical Electrical Equipment SafetyTesting Provided
Motor Module(CM1) DesignModificationsMotor Controller; Mechanical fuse in secondary box (motor controller)3 piece enclosure designMachinedMotor Controller; Firmware fuse integrated into motor module2 piece enclosure designMachinedFunctional Testing Provided
Disposable Kit(CK1) DesignModificationsMachinedInternal adhesive, Loctite 33211 PTFE cover (tube) for both the anesthesia needle and the bladeMolded; minor dimensional changes to bottom plate to mate with CM1 due to transition from machined to molded.Internal adhesive, Loctite 43112 individual PTFE covers (tubes); one for the anesthesia needle and 1 for bladeFunctional Testing Provided
DemonstratedSafety and Efficacyin treated areaClinical performance data provided for safety and efficacy at 3 months underIDE G120116 – De Novo K101231,cleared July 12, 2013Clinical performance data provided for safety and efficacy at 1 year under IDEG120116 – 510(k) K134010, clearedApril 14, 2014Extended follow-up clinicalperformance data provided for safetyand efficacy at 2 years under IDEG120116Clinical Follow-up Data Provided

Image /page/3/Picture/5 description: The image shows the logo for Ulthera, a company that specializes in medical aesthetic devices. The logo features the word "ulthera" in a sans-serif font, with the "u" and "l" in a light blue color and the rest of the letters in a darker shade of blue. To the right of the word, there are three orange circles arranged horizontally, followed by the trademark symbol.

{4}------------------------------------------------

Class II (special controls). The special controls for this device are:

  • (1) Non-clinical testing must be performed to demonstrate that the device meets all design specifications and performance requirements, and to demonstrate durability and mechanical integrity of the device.
  • (2) In vivo evaluation of the device must demonstrate device performance, including the safety of the release methodology and blood loss at the treatment sites.
  • (3) All elements of the device that may contact the patient must be demonstrated to be biocompatible.
  • (4) Electrical safety and electromagnetic compatibility of the device must be demonstrated.
  • The labeling must include a summary of in vivo evaluation data and all the device specific (2) warnings, precautions, and/or contraindications.
  • (6) Sterility and shelf-life testing for the device must demonstrate the sterility of patient contacting components and the shelf life of these components.

Device Description:

The Cellfina System is intended to provide precise focal release of subcutaneous tissue for improvement in the appearance of cellulite. The release of subcutaneous tissue for improvement in the appearance of cellulite is a minimally invasive surgical technique by physicians using of manual surgical instruments and accessories. The Cellfina device consists of a powered cutting blade and a means for instrument guidance to control the depth, size and shape of the tissue release. The system consists of a sterile single use disposable kit and a non-sterile reusable motor module.

Substantial Equivalence Comparison:

Image /page/4/Picture/12 description: The image shows the logo for Ulthera, a company that specializes in medical aesthetic devices. The logo features the word "ulthera" in a sans-serif font, with the "u" and "l" in a light blue color and the rest of the letters in a darker shade of blue. Above the "era" portion of the word, there are three orange circles arranged horizontally. The TM symbol is located to the right of the word.

{5}------------------------------------------------

Image /page/5/Picture/2 description: The image shows the word "ulthera" in a sans-serif font. The letters are a light blue color. To the right of the word are three yellow circles arranged horizontally. To the right of the circles is the trademark symbol.

{6}------------------------------------------------

Image /page/6/Picture/2 description: The image shows the logo for Ulthera. The word "ulthera" is written in a sans-serif font, with the "u" and "l" in a light blue color and the rest of the letters in a darker blue. To the right of the word, there are three orange circles arranged horizontally, followed by the trademark symbol.

{7}------------------------------------------------

Bench Performance Data:

Comprehensive bench testing has been successfully completed for the Cellfina System on aged product. Testing was performed on finished, sterile, aged devices that were exposed to environmental and transportation conditioning prior to testing which included simulated use, durability, and mechanical integrity. All devices were shown to meet the predetermined acceptance criteria.

Clinical Performance Data:

The safety and effectiveness of the Cellfina System was evaluated in a pivotal clinical study conducted under IDE G120116. The IDE was modified to allow for extended follow-up clinical performance data to be gathered for safety and efficacy at 2 years. No additional treatments were performed.

A prospective, multi-center, non-randomized open label, safety and effectiveness study was conducted for improvement of the appearance of cellulite. All subjects served as their own control and underwent a single treatment with the Cellfina System. Subjects underwent follow-up assessments at 3 days, 14 days, 1 year and 2 years post treatment. The subjects were asked to rate their satisfaction with their appearance and pain. Photographs were taken in accordance with a protocol-specific procedure at baseline and each follow-up. An independent and blinded reviewer of the photographs before treatment and at 2 years was used to verify the effectiveness of the procedure. A DSMB was formed and independently managed to provide safety oversight for the study.

All endpoints were achieved. The primary safety endpoint, defined as freedom from serious adverse events attributable to the procedure or device was achieved for all subjects (100%). The overall study success criteria was met with a clinically significant, long term improvement in the appearance of cellulite in the treated subjects defined by the following:

  • The primary endpoint was met: achievement of ≥1 point average reduction in the 0-5 point Cellulite Severity Scale as determined by independent physician assessment of subject photographs taken before and 2 years after treatment.
    • o The average improvement was 2.0 points (p<0.0001) with the 97.5% confidence limit of a 1.8 point improvement.

{8}------------------------------------------------

  • The powered secondary endpoint was met: improvement of one grade or more in severity (none, mild, moderate, severe) in >60% of treated subjects as determined by independent physician assessment of subject photographs taken before and 2 years after treatment.
    • 88.5% of treated subjects had improvement of 1 grade with a 95% confidence limit of o 79.2%.
  • . The safety endpoint was met: 100% of subjects were free from Serious Adverse Events (SAE) directly attributable to the Cellfina™ System or procedure.

All additional secondary measures were also achieved or acceptable at the 2 year timepoint:

  • The average rate of correct selection by independent physician assessment of blinded subject photographs taken before and was 98.1% with a 95% confidence limit of 94.5%.
  • . 100% of the subjects had noticeable improvement by the GAIS and 52% were characterized as having marked improvement or better by independent physician assessment of blinded subject photographs taken before and 2 years after treatment.
  • . 96% of the subjects were either satisfied or very satisfied as evaluated by a 5 point Likert type scale at the 2 year follow-up.
    • . Subject reported pain on a 0-10 numerical rating scale was 4.5 for the delivery of anesthesia and 3.7 for the tissue release portion of the procedure. For the follow-ups, 71% of subjects rated pain ≤ 3 at 3 days, and over 95% rated pain ≤ 3 thereafter. No subjects reported pain at the 2 year follow-up.

Conclusion Statements:

This 510(k) Premarket Notification is for updated labeling of the Cellfina System based on completion of the 2 year follow-up from the pivotal trial. The clinical data demonstrates that the Cellfina System is both safe and effective for long term improvement in the appearance of cellulite. In addition, this 510(k) also includes modifications to the design of the Cellfina System for manufacturability and ease of customer use. Through comprehensive bench testing it has been confirmed that these changes do not pose any new questions of safety, product output or efficacy.

As demonstrated through nonclinical and clinical tests, the Cellfina System is as safe, as effective, and performs as well as or better than the legally marketed predicate device (K134010) which was cleared on April 14, 2014, and as such, is substantially equivalent.

Image /page/8/Picture/12 description: The image shows the logo for Ulthera. The word "ulthera" is written in lowercase letters, with the "u" and "l" in a light blue color and the rest of the letters in a darker blue. To the right of the word, there are three yellow circles arranged horizontally. The letters are stacked vertically, and the circles are to the right of the letters.

§ 878.4790 Powered surgical instrument for improvement in the appearance of cellulite.

(a)
Identification. A powered surgical instrument for improvement in the appearance of cellulite is a prescription device that is used for the controlled release of subcutaneous tissue for improvement in the appearance of cellulite. The device consists of a cutting tool powered by a motor and a means for instrument guidance to control the areas of subcutaneous tissue cutting underneath the cellulite depressions or dimples.(b)
Classification. Class II (special controls). The special controls for this device are:(1) Non-clinical testing must be performed to demonstrate that the device meets all design specifications and performance requirements, and to demonstrate durability and mechanical integrity of the device.
(2) In vivo evaluation of the device must demonstrate device performance, including the safety of the release methodology and blood loss at the treatment sites.
(3) All elements of the device that may contact the patient must be demonstrated to be biocompatible.
(4) Electrical safety and electromagnetic compatibility of the device must be demonstrated.
(5) The labeling must include a summary of in vivo evaluation data and all the device specific warnings, precautions, and/or contraindications.
(6) Sterility and shelf-life testing for the device must demonstrate the sterility of patient contacting components and the shelf life of these components.