K Number
K150502

Validate with FDA (Live)

Date Cleared
2015-10-09

(225 days)

Product Code
Regulation Number
862.3650
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The DRI® Hydrocodone Assay is intended for the qualitative and semi-quantitative detection and estimation of Hydrocodone and its metabolites in human urine at a cutoff of 300 ng/mL. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of specimen for confirmatory method such as LC-MS/MS or GC-MS and permitting laboratories to establish quality control measures.

This assay provides a preliminary analytical test result. A more specific alternative chemical method must be used in order to confirm an analytical result. Gas chromatography/mass spectrometry (GC/MS) and Liquid Chromatography/ tandem mass spectrometry (LC-MS/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The DRI® Hydrocodone Assay Calibrators are intended for the DRI® Hydrocodone Assay. For In Vitro Diagnostic Use Only.

The DRI® Hydrocodone Controls are unassayed quality control material intended for use in the DRI Hydrocodone Assay to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. For In Vitro Diagnostics Use Only

Device Description

The DRI® Hydrocodone Assay is supplied as a liquid ready-to-use homogeneous enzyme immunoassay. The assay uses specific antibodies that can detect Hydrocodone and its metabolites without any significant cross-reactivity to other opiate compounds. The assay is based on competition between a drug labeled with glucose-6-phosphate dehydrogenase (G6PDH) and free drug from the urine sample, for a fixed amount of specific antibody binding sites. In the absence of free drug from the sample, the specific antibody binds the drug labeled with G6PDH and causes a decrease in enzyme activity. In the presence of free drug, the free drug occupies the antibody binding sites, allowing the drug bound G6PDH to interact with the substrate, resulting in enzyme activity. This phenomenon creates a direct relationship between the drug concentration in urine and enzyme activity. The enzyme activity is determined spectrophotometrically at 340 nm by measuring the conversion of nicotinamide adenine dinucleotide (NAD) to NADH.

The DRI® Hydrocodone Assay is a kit comprised of two reagents, Reagent A and Reagent E, which are bottled separately but sold together within the same kit.

The Reagent A solution contains: mouse monoclonal anti-hydrocodone antibody, glucose-6phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in Tris buffer with Sodium Azide (≤0.09%) as a preservative). The Reagent E solution contains: glucose-6-phosphate dehydrogenase (G6PDH) in Tris buffer with Sodium Azide (≤0.09%) as preservative.

The DRI® Hydrocodone Enzyme Immunoassay calibrators designated for use at the 300 ng/mL cutoff contain 0 (negative), 100, 300, 500, and 1,000 ng/mL of hydrocodone in human urine matrix with sodium azide (≤0.09%) as preservative. The controls are provided at a concentration of 225 and 375 ng/mL. The calibrators are sold separately and the two controls are sold as a kit.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the DRI® Hydrocodone Assay, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Performance CharacteristicAcceptance Criteria (Implicit)Reported Device Performance (DRI® Hydrocodone Assay)
Precision (Qualitative)All samples below cutoff should read negative. All samples above cutoff should read positive. Samples at cutoff may show variability.All samples tested recovered accurately. Samples below cutoff read negative, samples above cutoff read positive. At the 300 ng/mL cutoff, for samples spiked at 300 ng/mL (100% of cutoff), 46/80 were negative and 34/80 were positive within run and total run. For samples spiked at 225 ng/mL (-25% of cutoff), all 80 were negative. For samples spiked at 375 ng/mL (+25% of cutoff), all 80 were positive.
Precision (Semi-Quantitative)Similar to qualitative precision for classification, with quantitative results expected to be close to spiked values.All samples tested recovered accurately. Samples below cutoff read negative, samples above cutoff read positive. At the 300 ng/mL cutoff, for samples spiked at 300 ng/mL (100% of cutoff), 40/80 were negative and 40/80 were positive within run and total run. For samples spiked at 225 ng/mL (-25% of cutoff), all 80 were negative. For samples spiked at 375 ng/mL (+25% of cutoff), all 80 were positive.
Accuracy (vs. LC-MS/MS)High concordance with the reference method, especially for samples not near the cutoff.Overall concordance between DRI® Hydrocodone Assay and LC-MS/MS was 93%.
Linearity/Analytical RecoveryRegression equation close to y=x, with a high R² value. Recovery percentages within an acceptable range (e.g., 80-120%).Regression equation: y=1.0341-1.9933. R² value: 0.9965. Recovery % for spiked hydrocodone concentrations ranged from 94% to 114%. The results "passed the acceptance criteria" and "demonstrated the values were within the acceptance criteria."
Specificity and Cross-ReactivitySpecific detection of hydrocodone and its key metabolites, with minimal or no cross-reactivity from other opiates or common substances at expected concentrations.Hydrocodone and its active metabolites (Hydromorphone, Hydromorphone-3β-glucuronide) showed high cross-reactivity (102-122%). Other substances tested showed very low (e.g., Norhydrocodone 3.1%, Dihydrocodeine 2.7%, Levorphanol 1.7%, Naloxone 2.0%, NorOxycodone 0.3%, Oxycodone 2.5%, Oxymorphone-6β-D-glucuronide 2.2%, Oxymorphone 2.5%) or negligible (<0.2%-<0.4%) cross-reactivity.
Interference (pH, Endogenous Substances)No interference in qualitative or semi-quantitative results when potentially interfering substances are present.Controls (low and high) detected accurately in the presence of listed interfering substances (e.g., Acetaminophen, Caffeine, Creatinine, Ethanol, Glucose, Hemoglobin, Ibuprofen, various pH levels, etc.), indicating no interference.
Interference (Specific Gravity)No interference in qualitative or semi-quantitative results across a range of specific gravity values.No interference observed for drug-free urine samples or spiked low/high control samples across a specific gravity range of 1.000 to 1.028 g/mL (data showed up to 1.036 g/mL).
Stability (Open Vial Calibrators & Controls)Maintain performance for specified duration.Supported a claim of 60 days at 2-8°C for qualitative and semi-quantitative modes.
Stability (Real-Time Reagent, Calibrators & Controls)Maintain performance for specified duration.Ongoing, carried out up to 289 days at 2-8°C.
Stability (Accelerated Reagents, Calibrators & Controls)Predict longer-term stability based on accelerated conditions.Low control detected negative and high control detected positive for 4 months at 23°C (equivalent to 13 months according to Q10 model). Six-month accelerated stability also confirmed these results.

2. Sample Size Used for the Test Set and Data Provenance

  • Precision Studies: 80 replicates (tested twice per day for 20 days, n=80) for each spiked concentration level (0, 75, 150, 225, 300, 375, 450, 525, 600 ng/mL).
  • Accuracy Study (Patient Samples): 100 patient samples.
  • Linearity Study: 5 replicates for each spiked concentration level (12 levels including 0).
  • Specificity & Cross-Reactivity: Not explicitly stated but implied to be multiple tests for each substance and concentration.
  • Interference (pH, Endogenous Substances, Specific Gravity): Not explicitly stated, but "controls were detected accurately" implies sufficient testing for each condition.

Data Provenance: The document does not explicitly state the country of origin for the data or whether the studies were retrospective or prospective. Given it's a 510(k) submission to the FDA, it can be inferred the data was generated under conditions suitable for regulatory review, likely from laboratory settings. The "patient samples" for the accuracy study would be clinical specimens.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

  • For the Accuracy study, the ground truth for the 100 patient samples was established using LC-MS/MS (Liquid Chromatography/tandem mass spectrometry), which is specified as the "preferred confirmatory method" and is an analytical chemical method, not based on expert consensus. Therefore, no human experts were used to establish the ground truth in this context. LC-MS/MS itself is considered a highly sensitive and specific gold standard for drug confirmation.
  • For the Precision study, ground truth was based on the known spiked concentrations using LC-MS/MS for confirmation.
  • For the Linearity study, ground truth was based on the known spiked concentrations confirmed with LC-MS/MS data.
  • For Specificity, Cross-Reactivity, and Interference studies, ground truth was based on known concentrations of spiked substances.

4. Adjudication Method for the Test Set

Not applicable. The ground truth for the accuracy study was established by LC-MS/MS, an objective analytical method. There was no need for human expert adjudication of results against another human interpretation.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

No, an MRMC comparative effectiveness study was not done. This device is an in-vitro diagnostic assay for detecting substances in urine, not an imaging or diagnostic AI tool that assists human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done

Yes, the studies presented are all standalone performance evaluations of the assay system (device and analyzer) without human-in-the-loop performance influencing the primary analytical result. The assay directly reports qualitative (positive/negative) or semi-quantitative results. Clinical interpretation and professional judgment are advised after the preliminary analytical test result.

7. The Type of Ground Truth Used

The primary ground truth used for performance evaluation (e.g., accuracy, precision, linearity) was the objective analytical measurement by LC-MS/MS (Liquid Chromatography/tandem mass spectrometry), which is considered a gold standard confirmatory method (sometimes referred to as a "Reference Method"). For spiked samples, the ground truth was the known spiked concentration confirmed by LC-MS/MS.

8. The Sample Size for the Training Set

This document does not specify a separate "training set" or its size. In the context of an immunoassay, the concept of a training set as used in machine learning (where algorithms learn from data) is typically not applicable in the same way. The development of an immunoassay involves optimizing reagent formulations, antibody specificity, and assay conditions rather than training a predictive algorithm on a dataset.

9. How the Ground Truth for the Training Set was Established

Since a "training set" in the machine learning sense is not explicitly mentioned or applicable for this type of immunoassay development, the method for establishing its ground truth is not described. The manufacturer would have performed extensive R&D and optimization during method development to achieve the desired performance characteristics which were then validated in the studies described.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling a stylized eagle or bird with three human profiles incorporated into its design.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 9, 2015

MICROGENICS CORPORATION LAURIE WONG MANAGER, REGULATORY AFFAIRS 46500 KATO ROAD FREMONT CA 94538

Re: K150502

Trade/Device Name: DRI Hydrocodone Assay, DRI Hydrocodone Calibrators, DRI Hydrocodone Controls Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate Test System Regulatory Class: II Product Code: DJG, DLJ, LAS Dated: August 14, 2015 Received: August 19, 2015

Dear Ms. Wong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150502

Device Name DRI® Hydrocodone Assay DRI® Hydrocodone Calibrators DRI® Hydrocodone Controls

Indications for Use (Describe)

DRI® Hydrocodone Assay

The DRI® Hydrocodone Assay is intended for the qualitative and semi-quantitative detection and estimation of Hydrocodone and its metabolites in human urine at a cutoff of 300 ng/mL. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of specimen for confirmatory method such as LC-MS/MS or GC-MS and permitting laboratories to establish quality control measures.

This assay provides a preliminary analytical test result. A more specific alternative chemical method must be used in order to confirm an analytical result. Gas chromatography/mass spectrometry (GC/MS) and Liquid Chromatography/ tandem mass spectrometry (LC-MS/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

DRI® Hydrocodone Calibrators

The DRI® Hydrocodone Assay Calibrators are intended for the DRI® Hydrocodone Assay. For In Vitro Diagnostic Use Only.

DRI® Hydrocodone Controls

The DRI® Hydrocodone Controls are unassayed quality control material intended for use in the DRI Hydrocodone Assay to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. For In Vitro Diagnostics Use Only

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) Summary of Safety and Effectiveness

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.

Introduction

According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for a determination of substantial equivalence.

Submitter Name, Address and Content:

Microgenics 46500 Kato Road Fremont, CA 94538 Phone: (510) 979-5000 Fax: (510) 979-5002 Email: laurie.wong@thermofisher.com

Contact: Laurie Wong Regulatory Affairs Manager

Date Summary Originally Prepared: February 25, 2015 Date Summary updated per FDA request of October 1, 2015

Device Name and Classification

Trade or Proprietary Name:

DRI® Hydrocodone Assay DRI® Hydrocodone Assay Calibrators DRI® Hydrocodone Assay Controls

Common Name: Homogeneous Hydrocodone Enzyme Immunoassay

Classification Name:Enzyme Immunoassay, Hydrocodone
Class II, DJG (91 Toxicology)
21 CFR 862.3650

Drug Specific Calibrators Class II, DLJ (91 Toxicology) 21 CFR 862.3200

Drug Specific Controls Class I, Reserved LAS (91 Toxicology) 21 CFR 862.3280

{4}------------------------------------------------

Legally Marketed Predicate Device(s)

The DRI® Hydrocodone Assay is substantially equivalent to the predicate Lin-Zhi International, Inc., LZI Hydrocodone Enzyme Immunoassay (K141055). The DRI® Hydrocodone Assay, calibrators, and controls are identical or similar to the predicate in terms of intended use, method principle, device components, and clinical performance.

Device Description

The DRI® Hydrocodone Assay is supplied as a liquid ready-to-use homogeneous enzyme immunoassay. The assay uses specific antibodies that can detect Hydrocodone and its metabolites without any significant cross-reactivity to other opiate compounds. The assay is based on competition between a drug labeled with glucose-6-phosphate dehydrogenase (G6PDH) and free drug from the urine sample, for a fixed amount of specific antibody binding sites. In the absence of free drug from the sample, the specific antibody binds the drug labeled with G6PDH and causes a decrease in enzyme activity. In the presence of free drug, the free drug occupies the antibody binding sites, allowing the drug bound G6PDH to interact with the substrate, resulting in enzyme activity. This phenomenon creates a direct relationship between the drug concentration in urine and enzyme activity. The enzyme activity is determined spectrophotometrically at 340 nm by measuring the conversion of nicotinamide adenine dinucleotide (NAD) to NADH.

The DRI® Hydrocodone Assay is a kit comprised of two reagents, Reagent A and Reagent E, which are bottled separately but sold together within the same kit.

The Reagent A solution contains: mouse monoclonal anti-hydrocodone antibody, glucose-6phosphate (G6P) and nicotinamide adenine dinucleotide (NAD) in Tris buffer with Sodium Azide (≤0.09%) as a preservative). The Reagent E solution contains: glucose-6-phosphate dehydrogenase (G6PDH) in Tris buffer with Sodium Azide (≤0.09%) as preservative.

The DRI® Hydrocodone Enzyme Immunoassay calibrators designated for use at the 300 ng/mL cutoff contain 0 (negative), 100, 300, 500, and 1,000 ng/mL of hydrocodone in human urine matrix with sodium azide (≤0.09%) as preservative. The controls are provided at a concentration

{5}------------------------------------------------

of 225 and 375 ng/mL. The calibrators are sold separately and the two controls are sold as a kit. Intended Use

The DRI® Hydrocodone Assay is intended for the qualitative and semi-quantitative detection and estimation of Hydrocodone and its metabolites in human urine at a cutoff of 300 ng/mL. The semi-quantitative mode is for purposes of enabling laboratories to determine an appropriate dilution of specimen for confirmation by a confirmatory method such as LC-MS/MS or GC-MS and permitting laboratories to establish quality control measures.

This assay provides a preliminary analytical test result. A more specific alternative chemical method must be used in order to confirm an analytical result. Gas chromatography/mass spectrometry (GC/MS) and Liquid Chromatography/tandem mass spectrometry (LC-MS/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be applied to any drug of abuse test result, particularly when preliminary positive results are used.

The DRI® Hydrocodone Assay Calibrators are intended for the calibration of the DRI® Hydrocodone Assay. For In Vitro Diagnostic Use Only.

The DRI® Hydrocodone Controls are unassayed quality control material intended for use in the DRI Hydrocodone Assay to detect and monitor systematic deviations from accuracy resulting from reagent or instrument defects. For In Vitro Diagnostics Use Only

Comparison to Predicate Device

The DRI® Hydrocodone Assay is substantially equivalent to the predicate Lin-Zhi International, Inc., LZI Hydrocodone Enzyme Immunoassay, Calibrators and Controls cleared by the FDA under the premarket notification K141055 for its stated intended use.

The following table (next page) compares the DRI® Hydrocodone Assay to the predicate device.

{6}------------------------------------------------

DeviceSubject DevicePredicate Device
CharacteristicsDRI Hydrocodone Assay, Calibrators, and ControlsLZI Hydrocodone Enzyme Immunoassay, Calibrators and Controls (K141055)
Intended UseThe DRI® Hydrocodone Assay isintended for the qualitative andsemi-quantitative detectionand estimation of Hydrocodone andits metabolites in human urine at acutoff of 300 ng/mL. Thesemi-quantitative mode is forpurposes of enabling laboratories todetermine an appropriatedilution of specimen forconfirmation by a confirmatorymethod such as LC-MS/MS or GC-MS and permitting laboratories toestablish quality control measures.This assay provides a preliminaryanalytical test result. A morespecific alternative chemicalmethod must be used in order toconfirm an analytical result. Gaschromatography/massspectrometry (GC/MS) and LiquidChromatography/tandem massspectrometry (LC-MS/MS)are the preferred confirmatorymethods. Clinical consideration andprofessional judgmentshould be applied to any drug ofabuse test result, particularly whenpreliminary positive results are used.The DRI® Hydrocodone AssayCalibrators are intended for thecalibration of the DRI®Hydrocodone Assay. The DRI®Hydrocodone Controls are used tovalidate the DRI® HydrocodoneAssay calibration. For In VitroDiagnostics Use Only.The LZI Hydrocodone EnzymeImmunoassay, when used inconjunction with the AU480automated clinical system analyzers,is intended for the qualitative andsemi-quantitative determination ofhydrocodone in human urine atcutoff values of 100 or 300 ng/mL.The assay is designed forprofessional use with a number ofautomated clinical chemistryanalyzers.This assay provides a rapidscreening procedure for determiningthe presence of hydrocodone andhydromorphone in urine. The assayprovides only a preliminaryanalytical result. A more specificalternative chemical method must beused in order to obtain a confirmedanalytical result Gas or liquidchromatography/mass spectrometry(GC/MS or LC/MS) is the preferredconfirmatory method. Clinicalconsideration and professionaljudgment should be exercised withany drug of abuse test result,particularly when the preliminarytest result is positive.
AnalyteHydrocodoneHydrocodone
Cutoff Level300 ng/mL100 or 300 ng/mL
Sample MatrixHuman Urine MatrixHuman Urine Matrix
Calibrators Level300 ng/mL Cutoff: 5 levels: 0, 100,300, 500 and 1,000 ng/mL100 ng/mL Cutoff: 5 levels: 0, 50,100, 150, and 300 ng/mL300 ng/mL Cutoff: 5 levels: 0, 150,300, 500, and 800 ng/mL
Controls Level300 ng/mL Cutoff: 2 Levels (225ng/mL and 375 ng/mL)100 ng/mL Cutoff: 2 Levels (75ng/mL, 125 ng/mL)
Test PrincipleThe DRI® Hydrocodone Assay issupplied as a liquid ready-to-usehomogeneous enzymeimmunoassay. The assay usesspecific antibodies that can detectHydrocodone without anysignificant cross-reactivity to otheropiate compounds. The assay isbased on competition between adrug labeled with glucose-6-phosphate dehydrogenase (G6PDH)and free drug from the urine sample,for a fixed amount of specificantibody binding sites. In theabsence of free drug from thesample, the specific antibody bindsthe drug labeled with G6PDH andcauses a decrease in enzymeactivity. This phenomenon creates adirect relationship between the drugconcentration in urine and enzymeactivity. The enzyme activity isdetermined spectrophotometricallyat 340 nm by measuring theconversion of nicotinamide adeninedinucleotide (NAD) to NADH.The LZI Hydrocodone EnzymeImmunoassay is a liquid ready-to-use homogeneous enzymeimmunoassay. The assay usesspecific antibodies that can detectHydrocodone without anysignificant cross-reactivity to otheropiate compounds.
Storage2-8°C until expiration date2-8°C until expiration date

{7}------------------------------------------------

{8}------------------------------------------------

Performance Characteristics Summary:

AU680 Analyzer

Precision: 300 ng/mL Cutoff

Samples spiked with various amounts of Hydrocodone were tested in both qualitative and semiquantitative mode using a Clinical Laboratory and Standards Institute (CLSI) protocol. Results presented below were generated by testing all samples in replicates of two, twice per day for 20 days, total n=80.

Within Run Precision (n=80)Total Run Precision (n=80)
HydrocodoneSpikeConcentration(ng/mL)% ofCutoffLC-MS/MS(ng/mL)Number ofdeterminantsImmunoassayResultsNumber ofdeterminantsImmunoassayResults
0-100%08080 Neg8080 Neg
75-75%878080 Neg8080 Neg
150-50%1718080 Neg8080 Neg
225-25%2558080 Neg8080 Neg
300100%3458046 Neg/34 Pos8046 Neg/34 Pos
375+25%4428080 Pos8080 Pos
450+50%5358080 Pos8080 Pos
525+75%5618080 Pos8080 Pos
600+100%6648080 Pos8080 Pos

Qualitative Results using CLSI:

All samples tested recovered accurately. Samples at levels below the cutoff read as negative and samples at levels above the cutoff read as positive.

Semi-Quantitative Results Using CLSI:

Within Run Precision (n=80)Total Run Precision (n=80)
HydrocodoneSpikeConcentration(ng/mL)% ofCutoffLC-MS/MS(ng/mL)Number ofdeterminantsImmunoassayResultsNumber ofdeterminantsImmunoassayResults

{9}------------------------------------------------

0-100%08080 Neg8080 Neg
75-75%878080 Neg8080 Neg
150-50%1718080 Neg8080 Neg
225-25%2558080 Neg8080 Neg
300100%3458040 Neg/40 Pos8040 Neg/40 Pos
375+25%4428080 Pos8080 Pos
450+50%5358080 Pos8080 Pos
525+75%5618080 Pos8080 Pos
600+100%6648080 Pos8080 Pos

Accuracy

One hundred patient samples were analyzed by the DRI® Hydrocodone Assay in both qualitative and semi-quantitative modes and the results were compared to the LC-MS/MS. The overall concordance between LC-MS/MS and the DRI Hydrocodone Assay was 93%.

Qualitative Results with LC-MS/MS as reference method

DRIHydroco-doneAssayNegativeLess than50% ofcutoffconcentration byLC/MSanalysis(<150ng/mL)NearCutoffNegative(Between50%below thecutoff andthe cutoffconcentration)(150- 299ng/mL)NearCutoffPositive(Betweenthe cutoffand 50%above thecutoffconcentration)(300-450ng/mL)HighPositive(greaterthan 50%above thecutoffconcentration)>450ng/mL
Positive01*5**1039
Negative316710

{10}------------------------------------------------

*and ** 1 Sample in each bin has oxycodone concentrations greater than >37,000 ng/mL

DRIHydroco-doneAssayNegativeLess than50% ofcutoffconcentration byPredicateanalysis(<150ng/mL)NearCutoffNegative(Between50%below thecutoff andthe cutoffconcentration)150- 299ng/mLNearCutoffPositive(Betweenthe cutoffand 50%above thecutoffconcentration)300-450ng/mLHighPositive(greaterthan 50%above thecutoffconcentration)>450ng/mL
Positive01*5**1039
Negative316710

Semi-Quantitative Results with LC-MS/MS as reference method

*and ** 1 Sample in each bin has oxycodone concentrations greater than >37,000 ng/mL

Discordant Result Table for the Discrepant Samples near cutoff

Sample #Qualitative EIALC-MS/MS (ng/mL)Semi-Quantitative EIA (ng/mL)
HydrocodoneHydromorphoneHydormorphone 3β-D glucuronideAdjusted Total#
33Positive143.3<LLOQ♦67.6210.9369
70*Positive138.4<LLOQ<LLOQ1381280
75**Positive216.7<LLOQ<LLOQ217974
76Positive198.8<LLOQ42.6241319
83Positive78.4<LLOQ110.1188.5392

{11}------------------------------------------------

89Positive192.3<LLOQ56.2248.5406
96Negative303.3<LLOQ50.1353.4286

adjusted total LC-MS/MS (ng/mL) Hydrocodone + Hydromorphone + Hydromorphone 38-D glucuronide.

  • and **Oxycodone positive samples >37,000 ng/mL.

*LLOO = Lowest Limit of Quantitation is 40 ng/mL

Analytical Recovery and Linearity

To demonstrate linearity for the purpose of sample dilution and quality control over the entire assay range, a drug free urine pool was spiked with Hydrocodone across the range of the calibration curve. Each sample was run in replicates of five on the AU680 instrument in semiquantitative mode and the average was used to determine percent recovery compared to the expected target value. When comparing the result (y) and target (x) value, using the least squares regression technique, the regression equation and correlation are as follows y=1.0341-1.9933 and r2 value was 0.9965.

Linearity Results for Lot 3

TargetHydrocodone(ng/mL)ObservedRecovery(ng/mL)N=5Recovery %
0N/AN/A
504794
7576101
100108108
150171114
225250111
300302101
375398106
450472105
500527105
750844113
10001014101

The results demonstrated that the data passed the acceptance criteria. The LC-MS/MS data demonstrated the values were within the acceptance criteria. The results indicated that the

{12}------------------------------------------------

dilution linearity exists across the calibrator range.

Specificity and Cross-Reactivity

Several opiates (structurally similar or dissimilar to Hydrocodone) were tested for crossreactivity in the DRI® Hydrocodone Assay. Cross-reactant solutions were prepared by adding known amounts of stock solution to the drug free urine.

Hydrocodone, and its metabolites listed in the table were titrated to the lowest levels yielding a positive result in the assay.

Structurally related and unrelated opiates listed in the table were tested and results are shown below.

Hydrocodone and itsmetabolitesTestedConcentration(ng/mL)Pos/Neg% Cross-reactivity
Neg urine0Pos0%
Hydrocodone300Pos102%
Hydromorphone250Pos122%
Hydromorphone-3β-glucuronide250Pos122%
Norhydrocodone10,000Pos3.1%
Dihydrocodeine11,000Pos2.7%

Cross-Reactivity Results for Lot 3

Structurally relatedcompounds and otheropiatesTestedConcentration(ng/mL)Pos/Neg% Cross-reactivity
6-Acetyl Morphine100,000Neg< 0.3%
Buprenorphine100,000Neg< 0.3%
Buprenorphine-3β-D-glucuronide100,000Neg< 0.3%
Codeine150,000Neg< 0.2%
Dextromethorphan250,000Neg< 0.2%
EDDP150,000Neg< 0.2%
Fentanyl100,000Neg< 0.3%
Heroin100,000Neg< 0.3%
Levorphanol18,000Pos1.7%

{13}------------------------------------------------

Structurally relatedcompounds and otheropiatesTestedConcentration(ng/mL)Pos/Neg% Cross-reactivity
Methadone100,000Neg< 0.3%
Meperidine100,000Neg< 0.3%
Morphine150,000Neg< 0.2%
Morphine-3β-D-glucuronide70,000Neg< 0.4%
Morphine-6β-D-glucuronide75,000Neg< 0.4%
Nalbuphine150,000Neg< 0.3%
Naloxone15,000Pos2.0%
Naltrexone100,000Neg< 0.3%
Norbuprenorphine100,000Neg< 0.3%
Norcodeine150,000Neg< 0.2%
Normorphine150,000Neg< 0.2%
NorOxycodone100,000Pos0.3%
Oxycodone12,000Pos2.5%
Oxymorphone-6β-D-glucuronide14,000Pos2.2%
Oxymorphone12,000Pos2.5%
Tapentadol100,000Neg< 0.3%
Thebaine100,000Neg< 0.3%
Tramadol100,000Neg< 0.3%

Interference

The potential interference of pH and endogenous physiologic substances on recovery of Hydrocodone using DRI® Hydrocodone Assay was assessed by spiking known compounds of potentially interfering substances into the low (225 ng/mL) and high (375 ng/mL) controls for 300 ng/mL cutoff. In the presence of the compounds listed below, the controls were detected accurately indicating that these compounds did not show interference in the assay. The samples were tested in both qualitative and semi-quantitative mode.

Summary of Results
Cross ReactantsSpikedConcentration (ng/mL)Negative urineSQLow ControlQUALHigh ControlQUAL
Neg urine0NegNegPos

{14}------------------------------------------------

Summary of Results
Cross ReactantsSpikedConcentration (ng/mL)Negative urineSQLow ControlQUALHigh ControlQUAL
Acetaminophen10NegNegPos
Acetone500NegNegPos
Acetyl Salicylic Acid10NegNegPos
Ascorbic Acid150NegNegPos
Caffeine10NegNegPos
Creatinine400NegNegPos
Ethanol10NegNegPos
Galactose5NegNegPos
Glucose1000NegNegPos
Hemoglobin150NegNegPos
Human SerumAlbumin200NegNegPos
Ibuprophen10NegNegPos
Oxalic acid50NegNegPos
Riboflavin3NegNegPos
Sodium Chloride1000NegNegPos
Urea1000NegNegPos
pH4NegNegPos
pH5NegNegPos
pH6NegNegPos
pH7NegNegPos

{15}------------------------------------------------

Summary of Results
Cross ReactantsSpikedConcentration (ng/mL)Negative urineSQLow ControlQUALHigh ControlQUAL
pH8NegNegPos
pH9NegNegPos
pH10NegNegPos

Specific Gravity

Drug Free urine samples with specific gravity ranging in value from 1.000 to 1.036 g/mL were split and either left unspiked or spiked to a final concentration of either 225 ng/mL or 375 ng/mL (the low and high control concentrations, respectively). These samples were then evaluated in qualitative and semi-quantitative modes. No interference was observed.

Spiked Hydrocodone Concentration
SpecificGravity(g/mL)Low Control225 ng/mLHigh Control375 ng/mL
1.000NegPos
1.006NegPos
1.007NegPos
1.010NegPos
1.013NegPos
1.018NegPos
1.021NegPos
1.025NegPos
1.028NegPos

{16}------------------------------------------------

1.034NegPos
1.036NegPos

Stability:

Open Vial Calibrators and Controls

Open vial stability studies for three lots stored at 2-8°C were carried out to support a claim of 60 days at this time for qualitative and semi-quantitative.

Real Time Stability for Reagent, Calibrators and Controls

Real time studies for reagents, calibrators and controls at 2-8°C are ongoing and have been carried out up to 289 days.

Accelerated Stability Results for Reagents, Calibrators and Controls

Accelerated testing results show that the low control was detected as negative and the high control was detected as positive for each time point for a period of 4 months at 23°C. This is equivalent to 13 months of stability according to Q10 math model. The % recoveries of the low and high control were within 80-120%. The data for six month accelerated stability also confirmed that the low control was detected as negative and the high control was detected as positive.

Summary

As summarized, the DRI® Hydrocodone Assay is substantially equivalent to the legally marketed predicate LZI Hydrocodone Enzyme Immunoassay (Lin Zhi International) for the declared intended use. Substantial equivalence has been demonstrated through a comparison of the intended use and device characteristics when comparing the subject device to the legally marketed predicate. Performance testing was completed to verify that the device functions as intended and that design specifications have been satisfied. The content of the pre-market notification for the DRI® Hydrocodone Assay provides evidence that the device is safe and effective for the intended use.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).