K Number
K150489

Validate with FDA (Live)

Date Cleared
2015-06-11

(106 days)

Product Code
Regulation Number
870.4350
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

EOS PMP Integrated: Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir
The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The venous reservoir with cardiotomy filter is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacum assisted) during nomal operation to assure the proper oxygenation capability of the device. The device is intended to be used for 6 hours or less.

EOS PMP: Hollow Fiber Oxygenator
The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The device is intended to be used for 6 hours or less.

Device Description

The EOS PMP is a high efficiency hollow fiber diffusion membrane oxygenator with integrated heat exchanger.
The device provides oxygenation and carbon dioxide removal from patient's blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia, or aids in the maintenance of normothermia during surgery.
The device can be operated at flow rates up to 5 liters per minute (I/min).
The device can be connected with, but not limited to, the D905 EOS reservoir (K043323) that collects, defoams, and filters patient's blood.
The EOS PMP is a modified version of the currently marketed D905 EOS (K043323).

AI/ML Overview

The provided text describes a 510(k) premarket notification for a medical device, the EOS PMP and EOS PMP Integrated, which are hollow fiber oxygenators. The document focuses on demonstrating substantial equivalence to a legally marketed predicate device (D905 EOS).

Here's an analysis of the acceptance criteria and the study that proves the device meets them, based only on the provided text:

Summary of Device Performance and Acceptance Criteria:

Acceptance CriteriaReported Device Performance
Functional/Performance Tests:EOS PMP successfully met all acceptance criteria for these tests.
1. Blood traumaMet acceptance criteria.
2. Biological activityMet acceptance criteria.
3. Leaching of coatingMet acceptance criteria.
4. Flaking of coatingMet acceptance criteria.
5. Uniformity of coatingMet acceptance criteria.
6. Blood volume capacityMet acceptance criteria.
7. Oxygenating performance/blood side pressure dropsMet acceptance criteria.
8. Blood, water, gas pathway integrityMet acceptance criteria.
9. Plasma leakageMet acceptance criteria.
Other Criteria:
Sterile and non-pyrogenicEffectiveness of production techniques demonstrated.
Substantial Equivalence to D905 EOSDemonstrated for relevant functional parameters and device function.

Study Details:

  1. Sample size used for the test set and the data provenance:

    • The document states that "In vitro testing was conducted on the oxygenating module that is the only device element subject to modifications with respect to the unmodified device." However, it does not specify the sample size (e.g., number of devices tested) for these in vitro tests.
    • Data provenance: The tests were "in vitro," meaning they were conducted in a controlled, non-living environment (e.g., laboratory setting), rather than with human or animal subjects. The document does not specify a country of origin for the data; the submitter is "Sorin Group Italia," suggesting the tests might have been performed in Italy or by a third-party laboratory.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • This information is not provided in the document. The tests are described as "in vitro" and against "relevant requirements of 'Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and FDA Staff, November 13, 2000'," suggesting objective, measurable criteria rather than expert interpretation of results.
  3. Adjudication method (e.g. 2+1, 3+1, none) for the test set:

    • This information is not provided as the testing involved in-vitro performance measurements against established guidances/standards, not expert adjudication of subjective findings.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • No MRMC comparative effectiveness study was done or mentioned. This device is a cardiopulmonary bypass oxygenator, not an AI-assisted diagnostic tool requiring human reader involvement.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

    • This question is not applicable. The device is a physical medical device (an oxygenator), not an algorithm or software. The "standalone performance" was assessed through the in-vitro functional/performance tests.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The ground truth for the in-vitro performance tests was based on pre-defined acceptance criteria derived from "relevant requirements of 'Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and FDA Staff, November 13, 2000'" and ISO 10993-1 for biological evaluation. These guidances and standards establish objective measurable benchmarks for safe and effective performance of such devices.
  7. The sample size for the training set:

    • This information is not applicable as this is not a machine learning/AI device that requires a training set. The "modified device" (EOS PMP) was compared to an "unmodified device" (D905 EOS) and evaluated against established performance criteria.
  8. How the ground truth for the training set was established:

    • This information is not applicable as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, stacked on top of each other.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 11, 2015

Sorin Group Italia S.r.1. % Barry Sall RAC, FRAPS Principal Consultant PAREXEL Consulting, LLC 195 West Street Waltham, Massachusetts 02451

Re: K150489

Trade/Device Name: EOS PMP, EOS PMP Integrated Regulation Number: 21 CFR 870.4350 Regulation Name: Cardiopulmonary Bypass Oxygenator Regulatory Class: Class II Product Code: DTZ Dated: April 10, 2015 Received: April 13, 2015

Dear Barry Sall:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device

{1}------------------------------------------------

related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICESFood and Drug AdministrationForm Approved: OMB No. 0910-0120Expiration Date: January 31, 2017See PRA Statement below.
Indications for Use
510(k) Number (if known)
K150489
Device Name
EOS PMP
EOS PMP Integrated

Indications for Use (Describe)

EOS PMP Integrated: Hollow Fiber Oxygenator with Integrated Hardshell Venous/Cardiotomy Reservoir The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothernia during surgery. The venous reservoir with cardiotomy filter is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacum assisted) during nomal operation to assure the proper oxygenation capability of the device. The device is intended to be used for 6 hours or less.

Patient population: Paediatric/small adults

EOS PMP: Hollow Fiber Oxygenator

The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The device is intended to be used for 6 hours or less.

Patient population: Paediatric/small adults

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

"DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a callection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

PSC Publishing Servisore (101) 443-6300 . EX

{3}------------------------------------------------

SUBMITTER:Sorin Group Italia86, Via Statale 12 Nord41037 Mirandola (MO) Italy
CONTACT PERSON:Luigi VecchiPhone: 39 0535 29811Fax: 39 0535 25229
DATE PREPARED:May 28, 2015
DEVICE TRADE NAME:EOS PMPEOS PMP Integrated
COMMON NAME:Hollow Fiber OxygenatorHollow Fiber Oxygenator with integrated hardshellvenous/cardiotomy reservoirHardshell Venous/Cardiotomy Reservoir
CLASSIFICATION NAME:Cardiopulmonary Bypass Oxygenator/Cardiopulmonary Bypass Heat Exchanger/Cardiopulmonary Bypass Blood Reservoir/Cardiopulmonary Bypass Defoamer
CLASSIFICATION CODE:DTZ
REGULATION NUMBER:870.4350
UNMODIFIED DEVICE(S):D905 EOS: hollow fiber oxygenator with integratedhardshell venous/cardiotomy reservoir (K043323)

510(k) SUMMARY

DEVICE DESCRIPTION:

The EOS PMP is a high efficiency hollow fiber diffusion membrane oxygenator with integrated heat exchanger.

The device provides oxygenation and carbon dioxide removal from patient's blood. The integrated heat exchanger controls blood temperature and allows the use of hypothermia, or aids in the maintenance of normothermia during surgery.

The device can be operated at flow rates up to 5 liters per minute (I/min).

The device can be connected with, but not limited to, the D905 EOS reservoir (K043323) that collects, defoams, and filters patient's blood.

The EOS PMP is a modified version of the currently marketed D905 EOS (K043323).

{4}------------------------------------------------

INDICATION FOR USE:

The intended uses for the device models are:

EOS PMP Integrated: Hollow Fiber Oxyqenator with Integrated Hardshell Venous/Cardiotomy Reservoir

The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The venous reservoir with cardiotomy filter is intended to collect blood aspirated from the operating field during surgical procedures and the blood from patient's veins (gravity or vacuum assisted) during normal operation to assure the proper oxygenation capability of the device is intended to be used for 6 hours or less.

EOS PMP: Hollow Fiber Oxygenator

The device is intended for use in patients who undergo cardiopulmonary bypass surgery requiring extracorporeal circulation with a maximum blood flow rate of 5 liters /minute. It provides oxygenation and carbon dioxide removal from venous blood. The integrated heat exchanger provides blood temperature control and allows the use of hypothermia or aids in the maintenance of normothermia during surgery. The device is intended to be used for 6 hours or less.

TECHNOLOGICAL CHARACTERISTICS:

The EOS PMP has the same fundamental technological characteristics, principles of operation and control mechanisms as the unmodified device.

The EOS PMP can be operated at a minimum blood flow rate of 0.5 LPM rather than 1 LPM and is provided with a different oxygenating fiber consisting of polymethylpentene (PMP, surface area approx. 1.4 m²) rather than the polypropylene fiber used for the unmodified device (PP, surface area approx. 1.1 m²). As a result the static priming volume is ≤200 ml rather than ≤210 ml.

No other changes have been made to the device. The venous/cardiotomy reservoir has not been modified and is not subject of the present 510(k).

Except for the change of the oxygenating fiber, the EOS PMP utilizes the same materials as the unmodified device.

No change to the intended use has been made as a result of these modifications.

The EOS PMP and the D905 EOS unmodified device share fundamental technological characteristics except for modifications that do not affect the basic device function. These differences do not raise any new issues of safety and effectiveness.

The devices are substantially equivalent on the basis of operating principles and basic function.

There are no differences in packaging design between modified and unmodified device.

The EOS PMP is ethylene oxide sterilized and has a non-pyroqenic fluid path. It is for single use only.

{5}------------------------------------------------

NON CLINICAL TEST RESULTS:

Applicable tests were conducted in accordance with the requirements of ISO 10993-1.

The current draft Guidance "Use of International Standard ISO-10993, "Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" and its quidelines was also considered limited to external communicating device, circulating blood, limited contact duration.

IN VITRO TEST RESULTS:

In vitro testing was conducted on the oxygenating module that is the only device element subject to modifications with respect to the unmodified device. The tests were performed to demonstrate unmodified device substantial equivalency and compliance to safety and effectiveness requirements.

The tests were conducted in accordance with the relevant requirements of "Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions; Final Guidance for Industry and FDA Staff, November 13, 2000".

For venous/cardiotomy reservoir, the performance data previously submitted and already cleared are cross referenced.

The following table lists the performance tests conducted. The EOS PMP successfully met all acceptance criteria for each of the following tests.

TESTTEST CLASSIFICATIONTEST TITLE
1Functional/PerformanceBlood trauma
2Functional/PerformanceBiological activity
3Functional/PerformanceLeaching of coating
4Functional/PerformanceFlaking of coating
5Functional/PerformanceUniformity of coating
6Functional/PerformanceBlood volume capacity
7Functional/PerformanceOxygenating performance/blood side pressure drops
8Functional/PerformanceBlood, water, gas pathway integrity
9Functional/PerformancePlasma leakage

CONCLUSIONS:

The results of in vitro studies demonstrated that the modified oxygenator performs in a manner substantially equivalent to the unmodified oxygenator device with respect to the relevant functional parameters.

Test results of this study demonstrate the EOS PMP is substantially equivalent to the D905 EOS unmodified device with respect to device function.

Additional testing has also demonstrated the effectiveness of production techniques to assure that the device is sterile and non-pyrogenic.

§ 870.4350 Cardiopulmonary bypass oxygenator.

(a)
Identification. A cardiopulmonary bypass oxygenator is a device used to exchange gases between blood and a gaseous environment to satisfy the gas exchange needs of a patient during open-heart surgery.(b)
Classification. Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Cardiopulmonary Bypass Oxygenators 510(k) Submissions.”