(212 days)
The PROPEL device is indicated for manually drilling holes in tissue and bone for orthodontic and dental operative procedures including initiation holes for drill bits, implants, screws, plates and other orthodontic appliances.
The PROPEL Device is intended for use in dental operative procedures involving the preparation of tissue and bone. The PROPEL Device can be used manually as a bone cutting instrument in accordance with the FDA regulation 21 CFR Sec. 872.4120. It CANNOT be attached to any powered driver.
The PROPEL Device has a stainless steel drill tip that can be manually adjusted to the length needed for treatment or cutting depth. The plastic handle incorporates a retractable plastic tube over the cutting drill instrument. This enables the drill tip to be locked into a specific depth at 3. 5. or 7mm, thus preventing the drill tip from penetrating the bone bevond the targeted depth when used during orthodontic and dental operative procedures. The handle contains a red LED. operated by batteries. The LED is illuminated when the tip of the drill tip reaches the set depth. There is no motor in the device. The batteries are only for signaling when the prescribed depth has been achieved.
The device is provided sterile and is for single use only.
The document provided does not contain a study that establishes acceptance criteria and then proves the device meets those criteria in the typical format of a clinical trial or performance study with detailed statistical results. Instead, it is a 510(k) Premarket Notification from the FDA to Propel Orthodontics, LLC, confirming substantial equivalence of their PROPEL Device to a predicate device.
The information primarily focuses on comparing the PROPEL device's features to a predicate device (Synthes 90° Screwdriver, K082649) to demonstrate substantial equivalence, rather than setting specific performance acceptance criteria and then validating against them with a dedicated study.
However, based on the provided text, I can extract information related to "acceptance criteria" in the sense of compliance with standards and "reported device performance" from bench testing, and information about the evidence used to support substantial equivalence.
Here's an analysis based on the structure of your request:
1. Table of Acceptance Criteria and Reported Device Performance
Given the nature of a 510(k) submission, "acceptance criteria" here refer more to compliance with recognized standards and successful completion of bench tests, rather than explicit numerical thresholds for clinical performance.
| Acceptance Criteria Category (Compliance/Test Type) | Reported Device Performance (Summary) |
|---|---|
| Biocompatibility | In compliance with ISO 10993-1 and ISO 7405. |
| Sterilization | Validated according to ISO 11137-1, -2, and -3 to ensure a Sterility Assurance Level (SAL) of 10-6. |
| Shelf-life | Validated according to ASTM F1980-07 for a 2-year shelf-life. |
| Electrical Safety | In compliance with IEC 60601-1 and IEC 60601-1-2. |
| Bench Testing (General) | Demonstrated compliance with product specifications, medical community expectations, and product labeling. |
| Bench Testing (Specific) | Bit reliability (deflection resistance, hole depth repeatability) |
| Various tensile and compression tests of components | |
| Dial and bit depth reliability | |
| LED light reliability | |
| Clinical Literature | Documented ability to create holes through gum tissue into alveolar bone in humans. |
2. Sample Size Used for the Test Set and Data Provenance
The document does not detail specific sample sizes for test sets in the context of clinical performance evaluation (e.g., number of patients or cases). The "supporting data" section refers to:
- Bench testing: Conducted to demonstrate performance, but no sample sizes (e.g., number of devices tested for reliability) are provided.
- Clinical literature article: Only one article is cited ("Alikhani M, Rapis M, Soldan B, et al. Effect of micro-osteoperforations on the rate of tooth movement. Am J Orthod Dentofacial Orthop. 144:639-48; 2013."). The provenance of data within this article (country of origin, retrospective/prospective) is not specified in the 510(k) summary. This is presented as existing published literature, not a study specifically commissioned for this 510(k) submission with a freshly collected test set.
3. Number of Experts and Qualifications for Ground Truth
Not applicable. The document does not describe a process of establishing ground truth by experts for a specific test set. The clinical literature cited (Alikhani et al.) would have its own methodology and expert involvement, but this is not detailed in the 510(k) summary.
4. Adjudication Method
Not applicable. There is no mention of a ground truth adjudication method as there is no specific test set undergoing expert review described.
5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study
No. The document presents a comparison of the PROPEL device to a predicate device (Synthes 90° Screwdriver) for substantial equivalence based on technological characteristics and intended use, not a comparative effectiveness study involving human readers with and without AI assistance. This device is a manual drilling tool, not an AI-powered diagnostic or assistive tool.
6. Standalone (Algorithm Only) Performance
No. This is a physical medical device (a manual drill), not an algorithm or AI system.
7. Type of Ground Truth Used
For the non-clinical tests (biocompatibility, sterilization, shelf-life, electrical safety, bench tests), "ground truth" is defined by compliance with established international and national standards (ISO, ASTM, IEC) and the device's own product specifications.
For the claim of being able to create holes in alveolar bone, the "ground truth" is supported by a citation to published clinical literature.
8. Sample Size for the Training Set
Not applicable. This device is not an AI/ML algorithm that requires a training set. The term "training set" is relevant for AI models, not for traditional medical devices like a manual bone drill.
9. How Ground Truth for the Training Set Was Established
Not applicable, as there is no training set for this device.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 18, 2015
Propel Orthodontics, LLC c/o Mr. Craig J. Coombs Coombs Medical Device Consulting, Inc. 1193 Sherman Street Alameda. California 94501
Re: K150392
Trade/Device Name: PROPEL Device Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: II Product Code: DZJ Dated: August 18, 2015 Received: August 19, 2015
Dear Mr. Coombs:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Craig J. Coombs
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Image /page/1/Picture/7 description: The image contains a signature and the name "Tina Kiang" in a simple, sans-serif font. The signature is complex and illegible, consisting of loops and lines. The name is positioned to the right of the signature.
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K150392
Device Name PROPEL Device
Indications for Use (Describe)
The PROPEL device is indicated for manually drilling holes in tissue and bone for orthodontic and dental operative procedures including initiation holes for drill bits, implants, screws, plates and other orthodontic appliances.
| Type of Use (Select one or both, as applicable) | |
|---|---|
| ☒ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Propel Orthodontics. The word "PROPEL" is written in a large, sans-serif font in blue. Below the word "PROPEL" is the word "ORTHODONTICS" in a smaller, sans-serif font, also in blue. Above the word "PROPEL" is a series of blue arcs that form a semi-circle.
| A. Device Information: | |
|---|---|
| Category | Comments |
| Sponsor: | Propel Orthodontics, LLC.384 South Abbott AvenueMilpitas, CA 95035Company Contact: Bryce WayChief Executive OfficerTel: 408-394-5851Email: BW@PropelOrtho.com |
| Correspondent ContactInformation: | Craig CoombsCoombs Medical Device Consulting1193 Sherman StreetAlameda, CA 94501Tel: 510-337-0140Fax: 510-337-0416 |
| Device Common Name: | Manual Bone Drill |
| Device Classification Number& Name: | 21 CFR 872.4120Bone Cutting Instrument and Accessories |
| Device Classification &Product Code: | Class II,DZJ |
| Device Proprietary Name: | PROPEL Device |
Section 5: 510(k) Summary
Predicate Device Information:
| Predicate Device: | 90° Screwdriver |
|---|---|
| Predicate Device Manufacturer: | Synthes |
| Predicate Device Common Name: | Manual Screwdriver & Bone Drill |
| Predicate Device Premarket Notification # | K082649 |
| Predicate Device Classification & Name: | 21 CFR 872.4120Bone Cutting Instrument and Accessories |
| Predicate Device Classification &Product Code: | Class II,DZJ |
B. Date Summary Prepared
14 September 2015
C. Description of Device
The PROPEL Device is intended for use in dental operative procedures involving the preparation of tissue and bone. The PROPEL Device can be used manually as a bone cutting instrument in accordance with the FDA regulation 21 CFR Sec. 872.4120. It CANNOT be attached to any powered driver.
The PROPEL Device has a stainless steel drill tip that can be manually adjusted to the length needed for treatment or cutting depth. The plastic handle incorporates a retractable plastic tube
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Propel Orthodontics. The word "PROPEL" is written in a large, sans-serif font in blue. Below it, the word "ORTHODONTICS" is written in a smaller, sans-serif font, also in blue. Above the word "PROPEL" is a semi-circular design made of several curved lines that get progressively shorter from left to right.
over the cutting drill instrument. This enables the drill tip to be locked into a specific depth at 3. 5. or 7mm, thus preventing the drill tip from penetrating the bone bevond the targeted depth when used during orthodontic and dental operative procedures. The handle contains a red LED. operated by batteries. The LED is illuminated when the tip of the drill tip reaches the set depth. There is no motor in the device. The batteries are only for signaling when the prescribed depth has been achieved.
The device is provided sterile and is for single use only.
D. Indications for Use
The PROPEL device is indicated for manually drilling holes in tissue and bone for orthodontic and dental operative procedures including initiation holes for drill bits, implants, screws, plates and other orthodontic appliances.
E. Comparison to Predicate Device
The PROPEL device is substantially equivalent in intended use, technology, design, materials, and physician use as that found in the predicate Synthes 90° Screwdriver (K082649).
Product Code Equivalence
Both the subject PROPEL device and the predicate Synthes 90° Screwdriver have identical product codes: DZJ.
Indications for Use Equivalence
The subject device is indicated "for manually drilling holes in tissue and bone for orthodontic and dental operative procedures including initiation holes for drill bits, implants, screws, plates and other orthodontic appliances."
The predicate device is indicated "for the manual and powered pre-drilling and insertion of bone fixation screws in oral/maxillofacial surgery."
Both devices are bone-cutting devices for initiation holes for screws and other devices. The PROPEL device is a manual drill, whereas the predicate device is a powered drill. The subject device does not have a screwdriver function like the predicate. As a result, the subject device is indicated for a subset of performance features of the predicate device.
Because subject device features are a subset of the predicates, they do not raise new questions of safety or efficacy; they allow the conclusion of substantial equivalence between the Indications for Use.
Technological Equivalence
Both devices use rotating steel drill bits to cut holes in the bone of the upper or lower jaw. They differ in that the predicate device is a pneumatically powered drill, and the subject device is a manually powered drill.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Propel Orthodontics. The word "PROPEL" is written in a sans-serif font in blue. Below the word "PROPEL" is the word "ORTHODONTICS" in a smaller, gray sans-serif font. Above the word "PROPEL" is a series of blue lines that form an arc.
Comparison of Technological Aspects of the Predicate and Application Devices
| Characteristic | Predicate Device:Synthes 90° ScrewdriverK082649 | Subject Device:Propel Orthodontics, LLCPROPEL Device | How do differences, if any,pertain to justification ofsubstantial equivalence? |
|---|---|---|---|
| OperatingPrinciple | Hand-held powered drill orhand-held manualscrewdriver | Hand-held manual drill | Difference in power modeand lack of screwdriverfunction in application devicedoes not raise new questionsof safety or efficacy. |
| Drill Angle | Drills holes or inserts screwsat a 90° angle from the handle | Drills holes an a 0° angle(i.e. inline) with the handle | Difference in angle does notdoes not raise new questionsof safety or efficacy for whatthe application device isindicated for |
| Powered orManual | Screwdriver intended tooperate manually:Drill intended to operateconnected to a mechanicalpower source compatiblewith a Intra couplingaccording to ISO 3964/EN 23964 | Manually operated drill | Difference in power mode andlack of screwdriver functionin application device does notraise new questions of safetyor efficacy. |
| Direct drive orgeared? | 2:1 gearing ratio whenturning proximal end | No gearing, device is turnedby rotating proximal end | No clinical difference |
| Method forcreating holes | Stainless Steel Drill Bit | Stainless Steel Drill Bit | Clinically Identical |
| Interchangeabledrill bits | Yes | No. Drill bit built-in | Identical result as far ascreating holes |
| Characteristic | Predicate:Synthes 90° Screwdriver(K082649) | Subject Device:Propel Orthodontics, LLCPROPEL Device | How do differences, if any,pertain to justification ofsubstantial equivalence? |
| Drill Bit OD | 1.0 - 2.4mm | 1.6mm | Size available in applicationdevice is a subset of thoseavailable in the predicatedevice |
| Drill Bit WorkingLength | 4 – 18mm | 3, 5, or 7 mm | The smaller length availablein the application device areto serve the techniquesdescribed in the Indication forUse. Does not raise newquestions of safety or efficacy |
| Depth Stops | Yes, buy selecting theappropriate length bit | Yes, by adjusting the depthstop on the drill. | Clinically Identical |
| Types ofScrewdriver bits | Screwholder bits for screwOD's of 1.5, 2.0 and 2.4mm | None. Device cannot beconverted into a screwdriver | Raises no questions of safetyor efficacy |
| Indicator ofAchievement ofProper Depth | Head of drill in contact withtissue | LED light comes on toprevent unwarranted pressureon tissue | Clinically Identical |
| Materials: DrillBits | Stainless Steel | Stainless Steel | Clinically Identical |
| Materials: devicebody | Metallic | ABS and polycarbonate | Clinically Identical |
| Single Use? | Reusable | Single Use Only | Clinically Identical |
| Sterilization | Provided non-sterile, can beresterilized | Provided sterile | Clinically Identical aftersterilization of predicate |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Propel Orthodontics. The word "PROPEL" is in a large, sans-serif font and is colored in a light blue. Above the word, there is a series of curved lines that are also light blue, creating an arc shape. Below the word "PROPEL", the word "ORTHODONTICS" is in a smaller, sans-serif font and is colored in gray.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Propel Orthodontics. The word "PROPEL" is in a large, blue, sans-serif font. Below it, the word "ORTHODONTICS" is in a smaller, gray, sans-serif font. Above the word "PROPEL" is a curved, dashed line that resembles an arch.
The subject and predicate devices use steel drill bits to cut holes in the bone of upper or lower jaw. The diameter of the holes created by the PROPEL device are a subset of those available from the predicate device. Both devices allow the user to control the depth of the hole. The predicate uses different length bits to achieve the intended depth. The subject device has a retractable sleeve around its drill bit that allows the physician to preselect the hole depth up to7mm. The predicate device drills holes at a 90° angle to the axis of the handle, whereas the subject device drills holes in-line with the axis of the handle.
Propel Orthodontics, LLC, concludes that the subject PROPEL device and the predicate Synthes 90° Screwdriver are substantially equivalent.
F. Summary of Supporting Data
Biocompatibility data demonstrates that the PROPEL device is in compliance with ISO 10993-1: Biological evaluation of medical devices as described in ISO 7405: Dentistry — Evaluation of biocompatibility of medical devices used in dentistry.
Sterilization validation according to ISO 11137-1, -2, and -3 to ensure a SAL of 10th and shelf-life validation according to ASTM F1980-07 were conducted to validate the 2 year shelf-life of the PROPEL device.
Electrical safety testing demonstrates that the PROPEL device is in compliance with IEC 60601-1: Medical electrical equipment - Part 1: General reauirements for basic safety and essential performance and IEC 60601-1-2: Medical electrical equipment – Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests.
Bench testing has demonstrated that the PROPEL is in compliance with its product specifications, the expectations of the medical community and the product labeling. The following bench tests were conducted to demonstrate the Propel device's performance.
Bit reliability, including deflection resistance and hole depth repeatability Various tensile and compression tests of components Dial and bit depth reliability LED light reliability
A clinical literature article has documented that the PROPEL device can create holes through the gum tissue into alveolar bone in humans.
G. 510(k) Summary Conclusion
In accordance with 21CFR807.92(b)(3), it can be concluded from the nonclinical testing and the clinical literature that the PROPEL device is as safe and as effective as the predicate device in the context of its Indications for Use and technological characteristics.
It can be concluded that the PROPEL device is substantially equivalent to the predicate Synthes 90° Screwdriver.
1 Alikhani M, Rapis M, Soldan B, et al. Effect of micro-osteoperforations on the rate of tooth movement. Am J Orthod Dentofacial Orthop. 144:639-48; 2013.
§ 872.4120 Bone cutting instrument and accessories.
(a)
Identification. A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.(b)
Classification. Class II.