K Number
K150380

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-07-17

(154 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ProMIS™ Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; deformities (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

The ProMIS™ Fixation System can be used in an open approach or a posterior, percutaneous approach with MIS instrumentation.

Device Description

The ProMIS™ Fixation System consists of 3 main parts:

    1. ProMIS™ Fusion Rods: a straight or bent rod with various lengths.
    1. Polyaxial Pedicle screws.
    1. Setscrew.

All components are manufactured from Ti6Al4V per ASTM F136.

AI/ML Overview

The provided text describes a medical device called the ProMIS™ Fixation System and its 510(k) summary for FDA clearance. The document focuses on establishing substantial equivalence to a predicate device rather than presenting a study to meet specific acceptance criteria with performance metrics like accuracy, sensitivity, or specificity. Therefore, many of the requested details about acceptance criteria and study design are not present in this document.

However, based on the Performance Data section, the acceptance criteria are implicit in the ASTM F1717 standard for spinal implant constructs. The device's performance is demonstrated by its compliance with this standard.

Here's a breakdown of the available information:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (from ASTM F1717)Reported Device Performance
Static CompressionWas subjected to testing.
Static TorsionWas subjected to testing.
Dynamic CompressionWas subjected to testing.

Note: The document states that the system was subjected to these tests, implying it met the criteria, but does not provide specific numerical performance results.

2. Sample size used for the test set and the data provenance

Not applicable. This document describes mechanical testing of a physical device, not an AI or diagnostic algorithm that would require a test set of data/images. The "data provenance" mentioned in your template (country of origin, retrospective/prospective) is relevant for clinical or AI studies, which this is not.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable. Ground truth for mechanical testing is established by the physical properties of the materials and the mechanical outcomes measured, not by expert consensus on data interpretation.

4. Adjudication method for the test set

Not applicable. Mechanical testing outcomes are typically objectively measured and do not require expert adjudication in the same way clinical data interpretation would.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is a medical device for spinal fixation, not an AI diagnostic tool that assists human readers.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

Not applicable. This is a physical medical device, not an algorithm.

7. The type of ground truth used

The "ground truth" for the performance data in this context is the results of the mechanical tests performed according to ASTM F1717. These tests directly measure the physical performance and durability of the device under stress conditions, establishing whether it meets the engineering standards for spinal implants.

8. The sample size for the training set

Not applicable. There is no training set mentioned, as this is physical device testing, not an AI/machine learning application.

9. How the ground truth for the training set was established

Not applicable. There is no training set for this type of test.

Summary of the Study:

The "study" described in the document is a pre-clinical mechanical performance assessment. The ProMIS™ Fixation System was subjected to standard mechanical tests: static compression, static torsion, and dynamic compression, as specified by the ASTM F1717 standard. The purpose of these tests was to demonstrate the substantial equivalence of the ProMIS™ System to the predicate devices (THUNDERBOLT™ and LANCER™ Pedicle Screw Systems) in terms of mechanical properties and safety/effectiveness for its intended use. The document concludes that the performance data demonstrates substantial equivalence, indicating that the device met the mechanical acceptance criteria implied by the ASTM F1717 standard.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol depicts a stylized caduceus, with a double helix intertwined with a staff, representing health and medicine.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Premia Spine Ltd. Mr. Ron Sacher CEO 7 Giborey Israel Street Ramat Poleg, 42504 Israel

Re: K150380

Trade/Device Name: ProMIS™ Fixation System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, MNI, MNH Dated: June 15, 2015 Received: June 15, 2015

Dear Mr. Sacher:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

July 17, 2015

{1}------------------------------------------------

Page 2 - Mr. Ron Sacher

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150380

Device Name ProMIS™ Fixation System

Indications for Use (Describe)

The ProMIS™ Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities of the thoracic, lumbar, and sacral spine: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; deformities (i.e., scoliosis, kyphosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

The ProMIS™ Fixation System can be used in an open approach or a posterior, percutaneous approach with MIS instrumentation.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY Premia Spine's ProMIS™ Fixation System

Premia Spine Ltd. 7 Giborey Israel Street Ramat Poleg, 42504, Israel

Phone: 972-54-4626602 Facsimile: 972-72-2281201 Contact Person: Mr. Ron Sacher

Date Prepared: June 12, 2015

Trade Name: ProMIS™ Fixation System

21 C.F.R. 888.3070 Pedicle screw spinal system. NKB - orthosis, spinal pedicle fixation, for degenerative disc disease, Class III MNH - orthosis, spondylolisthesis spinal fixation, Class II MNI - orthosis, spinal pedicle fixation, Class II

Primary Predicate Device

  1. THUNDERBOLT™ and LANCER™ Pedicle Screw Systems by Choice Spine, LP (K132049).

Intended Use / Indications for Use

The ProMIS™ Fixation System is intended to provide immobilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative disc disease (DDD; defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; tracture or dislocation); spinal stenosis; deformities or curvatures (i.e., scoliosis, and/or lordosis); tumor; pseudoarthrosis; and failed previous fusion.

The ProMIS™ Fixation System can be used in an open approach or a posterior, percutaneous approach with MIS instrumentation.

Device Description

The ProMIS™ Fixation System consists of 3 main parts:

    1. ProMIS™ Fusion Rods: a straight or bent rod with various lengths.
    1. Polyaxial Pedicle screws.
    1. Setscrew.

All components are manufactured from Ti6Al4V per ASTM F136.

{4}------------------------------------------------

Performance Data

The ProMIS™ Fixation System was subjected to static compression, static torsion, and dynamic compression testing per ASTM F1717.

Substantial Equivalence

The subject and predicate systems are made from the same materials and are composed of similar components, including pedicle screws, fusion rods, and set screws. In addition, the diameter and length range of the screws used in the subject devices are within the range of screws cleared in the predicate system. The subject devices use 6.0 mm Ti alloy rods while the predicate system uses 5.5 mm Ti alloy or CoCr rods. Finally, the ProMIS System features very similar surgical instruments compared to the predicate devices. In particular, the ProMIS, like the predicate, has surgical instruments used for percutaneous insertion of pedicle screws via a tower system.

The ProMIS™ Fixation System is substantially equivalent to the THUNDERBOLT™ and LANCER™ Pedicle Screw Systems. The ProMIS™ Fixation System has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor technological differences between the ProMIS™ Fixation System and its predicate devices raise no new questions of safety or effectiveness. Performance data demonstrate that the ProMIS™ Fixation System is substantially equivalent to THUNDERBOLT™ and LANCER™ Pedicle Screw Systems.

Conclusions

The ProMIS™ Fixation System is substantially equivalent to the predicate devices.

§ 888.3070 Thoracolumbosacral pedicle screw system.

(a)
Identification. (1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.
(b)
Classification. (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards;
(ii) Compliance with mechanical testing standards;
(iii) Compliance with biocompatibility standards; and
(iv) Labeling that contains these two statements in addition to other appropriate labeling information:
“Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.”
“Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.”
(2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls:
(i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use.
(ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant.
(iii) Device components must be demonstrated to be biocompatible.
(iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments.
(v) Labeling must include the following:
(A) A clear description of the technological features of the device including identification of device materials and the principles of device operation;
(B) Intended use and indications for use, including levels of fixation;
(C) Identification of magnetic resonance (MR) compatibility status;
(D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and
(E) Detailed instructions of each surgical step, including device removal.
(3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls:
(i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate.
(ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.