K Number
K150329

Validate with FDA (Live)

Date Cleared
2015-11-05

(268 days)

Product Code
Regulation Number
870.1025
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Spacelabs élance Vital Signs Monitor and élance Central Station is indicated for use in patient populations for:

  • Adult
  • Pediatric (1 year old and above)
    The Spacelabs élance Vital Signs Monitor and élance Central Station facilitates the monitoring of:
  • ECG with arrhythmia detection
  • Respiration
  • Non-invasive blood pressures
  • Invasive blood pressures
  • Body temperature
  • Functional arterial oxygen saturation, and
  • End tidal CO2.
    The Spacelabs élance Vital Signs Monitor and élance Central Station is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.
Device Description

The Spacelabs Healthcare (Spacelabs) élance Vital Signs Monitor is a family of portable patient monitors intended to be used by clinicians and medical qualified personnel for monitoring ECG with arrhythmia detection, respiration, NIBP, temperature, SPO2, invasive blood pressure and EtCO2. Models within the Spacelabs élance family come in two different sized viewing areas (10.2" and 12.1"), two different housing colors (white and black) and offer selected monitoring features.
The Spacelabs élance Central Station software package is available for use with a customer acquired computer based on specifications provided by Spacelabs. This package allows monitoring of the élance Vital Signs Monitor at a central workstation.

AI/ML Overview

The provided text is a 510(k) premarket notification summary for the Spacelabs Healthcare élance Vital Signs Monitor and élance Central Station. This document focuses on demonstrating substantial equivalence to predicate devices, rather than establishing de novo clinical acceptance criteria and conducting a dedicated study to prove that the device meets those criteria.

Therefore, much of the requested information (e.g., number of experts, adjudication method, MRMC comparative effectiveness study, standalone performance metrics, sample sizes, and detailed ground truth establishment for a specific "study" demonstrating meeting acceptance criteria) is not present in this document because it describes a different type of regulatory submission. Clinical trial data with specific acceptance criteria and detailed performance studies are typically required for Novel Devices or those without a clear predicate.

However, based on the provided text, I can infer some aspects and present the available information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria here are implicitly alignment with recognized standards and successful completion of verification and validation testing, demonstrating that the device performs as intended and is as safe and effective as the predicate. Specific numerical performance metrics for individual parameters (e.g., accuracy for NIBP, sensitivity/specificity for arrhythmia detection) are referred to as being assessed against these standards but are not explicitly detailed in a table format with specific numerical targets.

Performance AspectAcceptance Criteria (Implicitly from Standards)Reported Device Performance
SoftwareCompliance with FDA guidance documents: "The content of premarket submissions for software contained in medical devices," "Off-the-shelf software use in medical devices," "General principles of software validation," and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices." Implies robust design, development, verification, and validation processes meeting predetermined specifications."Test results indicated that the élance VSM and élance CS software comply with predetermined specification."
Electrical SafetyCompliance with IEC 60601-1:1988, Am1: 1991, Am2: 1995 (Medical electrical equipment – Part 1: General requirements for basic safety and essential performance)."Test results indicated that the élance VSM complies with the applicable standards."
EMCCompliance with IEC 60601-1-2: 2007 (Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests)."Test results indicated that the élance VSM complies with the applicable standards."
Parameter Performance (Bench)Compliance with various standards including: - ANSI/AAMI EC57: 1998/(R) 2003 (Cardiac rhythm and ST-segment measurement algorithms)- ANSI/AAMI SP10: 2002, Am1: 2003 (Sphygmomanometers)- EN 12470-4: 2009 (Electrical thermometers)- EN 60601-1-6: 2010 (Usability)- EN 60601-1-8 2007, Am1: 2013, AC: 2014 (Alarm systems)- EN 62366: 2008 (Usability engineering)- IEC 60601-2-27: 2005 (ECG monitoring equipment)- IEC 60601-2-30: 1999 (Non-invasive blood pressure monitoring)- IEC 60601-2-34: 2000 (Invasive blood pressure monitoring)- IEC 60601-2-49: 2011 (Multifunction patient monitoring equipment)- ISO 80601-2-55: 2011 (Respiratory gas monitors)- ISO 80601-2-61: 2011 (Pulse oximeter equipment)- Internal requirements and procedures."Test results indicated that the élance VSM and élance CS comply with these Standards. Additionally, the élance VSM and élance CS were tested in accordance with internal requirements and procedures, and test results indicated that the devices comply with the predetermined requirements."
Overall EquivalenceThe device is "as safe, as effective, and perform as well as or better than the predicate devices."The conclusion states that "the élance VSM and élance CS are as safe, as effective, and perform as well as or better than the predicate devices. Therefore, the élance VSM and élance CS are considered substantially equivalent to the predicate devices."

2. Sample Size Used for the Test Set and Data Provenance

The document refers to "test results" for compliance with various standards but does not specify sample sizes for these tests for each parameter. The data provenance is not explicitly stated as originating from a specific country or being retrospective/prospective in a clinical trial context. It refers to a series of bench tests against recognized standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications

This information is not provided in the document. For bench testing against standards, the "ground truth" often comes from reference equipment, simulated signals, or specified test conditions described within the standards themselves, rather than clinical expert consensus.

4. Adjudication Method for the Test Set

This information is not provided as it typically pertains to clinical trials requiring expert review of patient data, which is not described here.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done

A MRMC comparative effectiveness study is not mentioned or described in this document. The submission focuses on demonstrating substantial equivalence through technical characteristics and compliance with standards, not on assessing the improvement of human readers with AI assistance. The device is a monitor, not an AI diagnostic tool for human interpretation.

6. If a Standalone (Algorithm Only Without Human-in-the-Loop Performance) was done

The document describes "parameter performance" testing for the élance VSM and élance CS against various standards. This testing would inherently involve evaluating the algorithms and hardware of the device itself (standalone performance) under controlled conditions to ensure they meet the criteria specified in those standards (e.g., accuracy of ECG measurement, NIBP, SpO2). However, specific metrics like sensitivity, specificity, or PPV/NPV for arrhythmia detection are not quantified in this summary, beyond stating compliance with the relevant standard (ANSI/AAMI EC57).

7. The Type of Ground Truth Used

For the performance testing mentioned, the ground truth would be based on:

  • Reference Standards: The specifications and requirements outlined in the cited IEC, ANSI/AAMI, EN, and ISO standards define the acceptable performance.
  • Controlled Bench Test Conditions: Specific inputs (e.g., simulated ECG waveforms, pressure cuffs with known pressures, temperature probes in calibrated environments) were used to test the device's accuracy and functionality against known values.
  • Predetermined Specifications: Internal requirements and procedures of Spacelabs Healthcare.

8. The Sample Size for the Training Set

This document does not mention a training set in the context of machine learning or AI algorithm development because the device, as described, is a vital signs monitor leveraging established physiological monitoring techniques rather than a novel AI system requiring extensive machine learning training. The "software" referred to is for device operation and data processing, validated against traditional software development processes.

9. How the Ground Truth for the Training Set Was Established

As no training set for machine learning is mentioned, this information is not applicable/provided.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 5, 2015

Spacelabs Healthcare % Thomas Kroenke, Principal Consultant Speed To Market, Inc. PO Box 3018 Nederland, Colorado 80466

Re: K150329

Trade/Device Name: élance Vital Signs Monitor and élance Central Station Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: MHX Dated: October 1, 2015 Received: October 6, 2015

Dear Thomas Kroenke:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{1}------------------------------------------------

Part 807); labeling (21 CFR Part 801); medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

K150329 510(k) Number (if known):

Device Name: élance Vital Signs Monitor and élance Central Station

Indications For Use: The Spacelabs élance Vital Signs Monitor and élance Central Station is indicated for use in patient populations for:

  • Adult ।
  • -Pediatric (1 year old and above)

The Spacelabs élance Vital Signs Monitor and élance Central Station facilitates the monitoring of:

  • ECG with arrhythmia detection -
  • -Respiration
  • Non-invasive blood pressures -
  • -Invasive blood pressures
  • -Body temperature
  • Functional arterial oxygen saturation, and -
  • -End tidal CO2.

The Spacelabs élance Vital Signs Monitor and élance Central Station is a prescription device intended to be used by healthcare professionals in all areas of a healthcare facility.

Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Page 1 of 1

{3}------------------------------------------------

Submission Date:01 October 2015
Submitter:Spacelabs Healthcare35301 SE Center StSnoqualmie, WA 98065
Submitter Contact:Mr. Al Van HoudtSpacelabs HealthcarePhone: +1 (425) 363- 5970Fax: +1 (425) 363-5762Email: al.vanhoudt@spacelabs.com
ApplicationCorrespondent:Thomas KroenkeSpeed To Market, Inc.PO Box 3018Nederland, CO 80466 USAtkroenke@speedtomarket.net303 956 4232
Manufacturing Site:Spacelabs Healthcare35301 SE Center StSnoqualmie, WA 98065
Trade Name:élance Vital Signs Monitor and élance Central Station
Common Name:Patient Physiological Monitor (with arrhythmia detection or alarms)
Classification Name:Patient Physiological Monitor (with arrhythmia detection or alarms)
PrimaryClassificationRegulation:21 CFR §870.1025
Primary ProductCode:MHX
SubstantiallyEquivalent Devices:New Spacelabs ModelPredicate510(k) NumberPredicateManufacturer / Model
Spacelabs Healthcare élanceVital Signs Monitor andélance Central StationK093501Spacelabs Healthcare élanceVital Signs Monitor and élanceCentral Station
K033296Masimo Corporation MasimoSET® Rad-5 Pulse Oximeter
K012891Nellcor Puritan BennettIncorporated OxiMAX PulseOximetry System with N-595 PulseOximeter and OxiMAX Sensorsand Cables

{4}------------------------------------------------

Device Description:The Spacelabs Healthcare (Spacelabs) élance Vital Signs Monitor is afamily of portable patient monitors intended to be used by cliniciansand medical qualified personnel for monitoring ECG with arrhythmiadetection, respiration, NIBP, temperature, SPO2, invasive bloodpressure and EtCO2. Models within the Spacelabs élance family comein two different sized viewing areas (10.2" and 12.1"), two differenthousing colors (white and black) and offer selected monitoringfeatures.
The Spacelabs élance Central Station software package is available foruse with a customer acquired computer based on specificationsprovided by Spacelabs. This package allows monitoring of the élanceVital Signs Monitor at a central workstation.
Intended Use:The Spacelabs élance Vital Signs Monitor and élance Central Station isindicated for use in patient populations for:- Adult- Pediatric (1 year old and above)The Spacelabs élance Vital Signs Monitor and élance Central Stationfacilitates the monitoring of:- ECG with arrhythmia detection- Respiration- Non-invasive blood pressures- Invasive blood pressures- Body temperature- Functional arterial oxygen saturation, and- End tidal CO2.The Spacelabs élance Vital Signs Monitor and élance Central Station isa prescription device intended to be used by healthcare professionals inall areas of a healthcare facility.

{5}------------------------------------------------

Technology Comparison:

The Spacelabs Spacelabs élance Vital Signs Monitor and élance Central Station employs the same technological characteristics as the predicate device.

CharacteristicPredicate DeviceProposed Device
ParametersECG with arrhythmiadetectionRespirationNon-invasive blood pressureInvasive blood pressureBody temperatureFunctional arterial oxygensaturation, andEnd tidal CO2.Same
Display TypeLCDSame
Display Sizes10.2" and 12.1"Same
VSM Variants5, 5i, 5c, 5 elite, 5i elite, 5celite7, 7i, 7c, 7 elite, 7i elite, 7celiteSame
Additional Options-93300-M, Masimo SpO293300-N, Nellcor SpO293300-A: Arrhythmia93300-S: ST Segment Analysis

Summary of Performance Testing:

Software

The élance VSM and élance CS contain MAJOR level of concern software. Software was designed and developed according to a robust software development process, and was rigorously verified and validated. Software information is provided in accordance with

  • · FDA guidance: The content of premarket submissions for software contained in medical devices, 11 May 05;
  • · FDA guidance: Off-the-shelf software use in medical devices, 09 Sep 99:
  • FDA guidance: General principles of software validation; Final guidance for industry and FDA staff, 11 Jan 02; and
  • · FDA guidance: Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, 14 Jun 13

Test results indicated that the élance VSM and élance CS software comply with predetermined specification.

{6}------------------------------------------------

Electrical SafetyThe élance VSM was tested for patient safety in accordance with:IEC 60601-1:1988, Am1: 1991, Am2: 1995, Medical electrical equipment – Part 1: General requirements for basic safety and essential performance. Test results indicated that the élance VSM complies with the applicable standards.
Electromagnetic CompatibilityThe élance VSM was tested for EMC in accordance with:IEC 60601-1-2: 2007, Medical electrical equipment – Part 1-2: General requirements for basic safety and essential performance – Collateral standard: Electromagnetic compatibility – Requirements and tests. Test results indicated that the élance VSM complies with the applicable standards.
Performance Testing – BenchThe élance VSM was tested for parameter performance in accordance with:ANSI/AAMI EC57: 1998/(R) 2003, Testing and reporting performance results of cardiac rhythm and ST-segment measurement algorithms; ANSI/AAMI SP10: 2002, Am1: 2003, Manual, electronic, or automated sphygmomanometers; EN 12470-4: 2009, Clinical thermometers – Part 4: Performance of electrical thermometers for continuous measurement; EN 60601-1-6: 2010, Medical electrical equipment – Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability; EN 60601-1-8 2007, Am1: 2013, AC: 2014, Medical electrical equipment - Part 1-8: General requirements for basic safety and essential performance – Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems; EN 62366: 2008, Medical devices - Application of usability engineering to medical devices; IEC 60601-2-27: 2005, Medical electrical equipment – Part 2-27: Particular requirements for the safety, including essential performance, of electrocardiographic monitoring equipment;

{7}------------------------------------------------

  • Performance Testing – Bench (continued)
  • · IEC 60601-2-30: 1999, Medical electrical equipment Part 2-30: Particular requirements for the safety, including essential performance, of automatic cycling non-invasive blood pressure monitoring equipment;
  • · IEC 60601-2-34: 2000, Medical electrical equipment Part 2-34: Particular requirements for the safety, including essential performance, of invasive blood pressure monitoring equipment;
  • · IEC 60601-2-49: 2011, Medical electrical equipment Part 2-49: Particular requirements for the basic safety and essential performance of multifunction patient monitoring equipment;
  • ISO 80601-2-55: 2011, Medical electrical equipment Part 2-55: Particular requirements for the basic safety and essential performance of respiratory gas monitors; and
  • ISO 80601-2-61: 2011, Medical electrical equipment Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.

The élance CS was tested for parameter performance in accordance with:

  • EN 60601-1-6: 2010, Medical electrical equipment Part 1-6: General requirements for basic safety and essential performance -Collateral standard: Usability;
  • EN 60601-1-8 2007, Am1: 2013, AC: 2014, Medical electrical equipment – Part 1-8: General requirements for basic safety and essential performance – Collateral Standard: General requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems;
  • · EN 62366: 2008, Medical devices Application of usability engineering to medical devices;
  • · Test results indicated that the élance VSM and élance CS comply with these Standards. Additionally, the élance VSM and élance CS were tested in accordance with internal requirements and procedures, and test results indicated that the devices comply with the predetermined requirements.

Verification and validation activities were conducted to establish the Conclusion performance and safety characteristics of the élance VSM and élance CS. The results of these activities demonstrate that the élance VSM and élance CS are as safe, as effective, and perform as well as or better than the predicate devices. Therefore, the élance VSM and élance CS are considered substantially equivalent to the predicate devices.

§ 870.1025 Arrhythmia detector and alarm (including ST-segment measurement and alarm).

(a)
Identification. The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.(b)
Classification. Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.