The CD HORIZON® Growth Rod Conversion Set is indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, early-onset spinal deformities associated with thoracic insufficiency including early-onset scoliosis. The CD HORIZON® Growth Rod Conversion Set may be used with any cleared traditional CD HORIZON® Spinal System rod construct ranging in diameter from 3.5mm to 5.5mm, with the exception of PEEK Rod constructs. The CD HORIZON® Growth Rod Conversion Set may not be used with PEEK Rods, SPIRE™ Spinous Process Plates, or Shape Memory Alloy (SMA) Staples.
The CD HORIZON® Growth Rod Conversion Set consists of a variety of connectors designed to convert a traditional fusion construct into a non-fusion growth enabling construct which can be surgically lengthened on a periodic basis as the patient grows. The CD HORIZON® Growth Rod Conversion Set components are manufactured from stainless steel or titanium alloy and are designed to interact with rod-based pedicle screw/hook constructs in which the rods range in diameter from 3.5mm to 5.5mm. The CD HORIZON® Growth Rod Conversion Set is specifically to be used with any traditional CD HORIZON® Spinal System (3.5mm) fusion construct cleared for pediatric usage. The CD HORIZON® Growth Rod Conversion Set may not be used with PEEK Rods, SPIRE® Spinous Process Plates and/or Shape Memory Alloy (SMA) Staples. The CD HORIZON® Growth Rod Conversion Set is limited to a posterior approach. The CD HORIZON® Growth Rod Conversion Set connectors are provided in both sterile and non-sterile form. The sterile implants are sterilized via gamma irradiation.
This device is a CD HORIZON® Growth Rod Conversion Set and the 510(k) summary (K150200) describes a submission to add sterile connectors and set screws to the existing device, utilizing gamma irradiation and sterile packaging. This submission is about a modification to an existing medical device, specifically regarding its packaging and sterilization process, rather than a new device with performance metrics related to diagnostic accuracy or clinical outcomes. Therefore, the standard elements for performance criteria and a study proving those criteria (as might be expected for an AI/ML device or a device with clinical efficacy claims) are not applicable in the usual sense.
However, I can extract the relevant information regarding the "acceptance criteria" and the "study" that proves the device meets them, reinterpreting these terms in the context of this specific 510(k) submission.
Acceptance Criteria and Device Performance (in the context of sterilization and packaging):
| Acceptance Criteria Category | Specific Criteria/Standard Adhered To | Reported Device Performance/Compliance |
|---|---|---|
| Material Biocompatibility | Manufactured from stainless steel (ASTM F138) or titanium alloy (ASTM F136). Materials must have a well-characterized level of local biological response and a long history of clinical use. | Subject devices use identical materials to predicate device (K133904), which are in accordance with ASTM F138 and ASTM F136. No additional biocompatibility testing required due to established history and standards compliance. |
| Mechanical Performance | No design changes made to the implants. Device must not represent a "new worst case" compared to the predicate device. | No mechanical testing performed on subject connectors and set screws; deemed not necessary as no design changes were made. The evaluation determined that subject devices do not represent a new worst case. |
| Sterilization Efficacy | Sterilization must meet recognized standards for radiation sterilization. | Sterilization is via gamma irradiation. Compliance with ANSI/AAMI/ISO 11137-1:2006/A1:2013 and ANSI/AAMI/ISO 11137-2:2013 demonstrated. |
| Packaging Integrity | Packaging must maintain sterility and withstand shipping. | Compliance with ASTM F88/F88M-09 (Seal Strength), ASTM D4169-09 (Performance Testing of Shipping Containers), and ASTM F1929-98 (Detecting Seal Leaks in Porous Medical Packaging) demonstrated. |
| Substantial Equivalence | The device (with new sterilization/packaging) must be substantially equivalent to the predicate device. | Based on risk analysis, packaging and sterilization rationales, and supporting documentation, the subject components are substantially equivalent to the predicate devices in K133904. |
Study Information (reinterpreted for sterilization and packaging validation):
-
Sample size used for the test set and the data provenance:
- Test Set Sample Size: The document does not specify a "sample size" for a clinical or performance test set in the traditional sense, as this submission is focused on sterilization and packaging validation. Instead, the "testing" involved adherence to established standards and protocols for manufacturing, sterilization, and packaging. These standards (e.g., ISO, ASTM) would have their own sample size requirements for their validation methods, but these are not enumerated here.
- Data Provenance: The data is based on adherence to international and US standards for medical device manufacturing, materials, sterilization, and packaging. This is generally laboratory-based testing conducted by the manufacturer or contracted labs, rather than data from specific countries or retrospective/prospective studies on patient populations.
-
Number of experts used to establish the ground truth for the test set and the qualifications of those experts:
- Not applicable in the context of this specific submission. The "ground truth" here is adherence to established engineering, materials science, and sterilization standards, validated by standard test methods rather than expert consensus on clinical data. The FDA reviewers are experts in regulatory compliance and relevant scientific fields.
-
Adjudication method (e.g., 2+1, 3+1, none) for the test set:
- Not applicable. This is not a study requiring adjudication of expert interpretations.
-
If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
- Not applicable. This is a spinal implant component, not an AI/ML device.
-
If a standalone (i.e. algorithm only without human-in-the loop performance) was done:
- Not applicable. This is not an algorithm.
-
The type of ground truth used (expert consensus, pathology, outcomes data, etc):
- The "ground truth" is adherence to validated engineering standards and material specifications (ASTM F136, ASTM F138, ISO 11137, ASTM D4169, ASTM F88, ASTM F1929) for material composition, sterilization efficacy, and packaging integrity. The primary "ground truth" for the overall submission is demonstrating substantial equivalence to a previously cleared predicate device (K133904) based on these technical considerations.
-
The sample size for the training set:
- Not applicable. This is not an AI/ML device and does not involve a training set.
-
How the ground truth for the training set was established:
- Not applicable.
Summary of the study/evidence presented:
The "study" in this context refers to the engineering and validation documentation provided by Medtronic Sofamor Danek USA, Incorporated to demonstrate that the changes (sterile connectors with pre-loaded set screws and gamma irradiation sterilization) to the CD HORIZON® Growth Rod Conversion Set do not alter its fundamental scientific technology or safety and effectiveness compared to the predicate device.
The evidence relies on:
- Material Equivalence: Stating that the materials (stainless steel per ASTM F138 or titanium alloy per ASTM F136) are identical to those in the predicate device and have a long history of safe clinical use.
- Design Equivalence: Confirming that no design changes were made to the implants themselves.
- Sterilization Validation: Referring to compliance with ISO 11137-1 and ISO 11137-2 for gamma irradiation.
- Packaging Validation: Referring to compliance with ASTM F88/F88M, ASTM D4169, and ASTM F1929 for seal strength, shipping, and leak detection.
- Risk Analysis: Performed to support the claim that the modified device does not represent a new worst case.
The conclusion is that, based on this documentation and adherence to standards, the modified device is substantially equivalent to the predicate device due to identical design, materials, method of implantation, and intended use, with only the sterilization and packaging method being updated.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and the word "ADMINISTRATION" in a smaller font below.
July 29, 2022
Medtronic Sofamor Danek USA, Incorporated Mr. Lee Grant Distinguished Regulatory Affairs Advisor 1800 Pyramid Place Memphis, Tennessee 38132
Re: K150200
Trade/Device Name: CD HORIZON® Growth Rod Conversion Set Regulation Number: 21 CFR 888.3070 Regulation Name: Thoracolumbosacral pedicle screw system Regulatory Class: Class II Product Code: PGM
Dear Mr. Grant:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated February 25, 2015. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under regulation number, 21 CFR 888.3070.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Ronald Jean, OHT6: Office of Orthopedic Devices, (301)796-5650, Ronald.Jean@fda.hhs.gov
Sincerely,
Ronald P. Jean -S
Ronald P. Jean, Ph.D. Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing right, arranged in a stacked formation. The profiles are rendered in black and have a flowing, abstract design. Encircling the profile graphic is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
February 25, 2015
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Medtronic Sofamor Danek USA, Incorporated Mr. Lee Grant Distinguished Regulatory Affairs Advisor 1800 Pyramid Place Memphis, Tennessee 38132
Re: K150200
Trade/Device Name: CD HORIZON® Growth Rod Conversion Set Regulatory Class: Unclassified Product Code: PGM Dated: January 20, 2015 Received: January 29, 2015
Dear Mr. Grant:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
Page 2 - Mr. Lee Grant
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name CD HORIZON® Growth Rod Conversion Set
Indications for Use (Describe)
The CD HORIZON® Growth Rod Conversion Set is indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, lifethreatening, early-onset spinal deformities associated with thoracic insufficiency including early-onset scoliosis. The CD HORIZON® Growth Rod Conversion Set may be used with any cleared traditional CD HORIZON® Spinal System rod construct ranging in diameter from 3.5mm, with the exception of PEEK Rod constructs. The CD HORLZON® Growth Rod Conversion Set may not be used with PEEK rods, SPIRE™ Spinous Process Plates, or Shape Memory Alloy (SMA) Staples.
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
| CD HORIZON® Growth Rod Conversion Set | ||
|---|---|---|
| 510(k) Summary – K150200 | ||
| February 2015 | ||
| I. | Submitter: | Medtronic Sofamor Danek USA, Inc. |
| 1800 Pyramid Place | ||
| Memphis, TN 38132 | ||
| Telephone: (901)396-3133 | ||
| Fax: (901) 346-9738 | ||
| Contact: | Lee Grant | |
| Distinguished Regulatory Affairs Advisor | ||
| Date Prepared: | January 16, 2015 | |
| II. | Device | |
| Name of Device: | CD HORIZON® Growth Rod Conversion Set | |
| Common Name: | Spinal Fixation Appliance | |
| Classification Name: | Growing Rod System | |
| Regulatory Class: | Unclassified | |
| Product Codes: | PGM | |
| III. | Predicate Device | K133904 (SE 2/25/2014) CD HORIZON® |
| Growth Rod Conversion Set – Primary Predicate | ||
| The predicate device has not been subject |
to a design related recall.
IV. Device Description:
The CD HORIZON® Growth Rod Conversion Set consists of a variety of connectors designed to convert a traditional fusion construct into a non-fusion growth enabling construct which can be surgically lengthened on a periodic basis as the patient grows. The CD HORIZON® Growth Rod Conversion Set components are manufactured from stainless steel or titanium alloy and are designed to interact with rod-based pedicle screw/hook constructs in which the rods range in diameter from 3.5mm to 5.5mm. The CD HORIZON® Growth Rod Conversion Set is specifically to be used with any traditional CD HORIZON® Spinal System (3.5mm) fusion construct cleared for pediatric usage. The CD HORIZON® Growth Rod Conversion Set may not be used with PEEK Rods, SPIRE® Spinous Process Plates and/or Shape Memory Alloy (SMA) Staples. The CD HORIZON® Growth Rod Conversion Set is limited to a posterior approach. The CD HORIZON® Growth Rod Conversion Set connectors are provided in both sterile and non-sterile form. The sterile implants are sterilized via gamma irradiation.
{5}------------------------------------------------
The purpose of this 510(k) submission is to add sterile connectors and set screws to the CD HORIZON® Growth Rod Conversion Set utilizing gamma irradiation and sterile packaging. The subject components are manufactured out of medical grade titanium alloy per ASTM F136 or medical grade stainless steel per ASTM F138.
V. Indications for Use:
The CD HORIZON® Growth Rod Conversion Set is indicated in patients with potential for additional spinal growth under 10 years of age who require surgical treatment to obtain and maintain correction of severe, progressive, life-threatening, early-onset spinal deformities associated with thoracic insufficiency, including early-onset scoliosis. The CD HORIZON® Growth Rod Conversion Set may be used with any cleared traditional CD HORIZON® Spinal System rod construct ranging in diameter from 3.5mm to 5.5mm, with the exception of PEEK Rod constructs. The CD HORIZON® Growth Rod Conversion Set may not be used with PEEK Rods, SPIRE™ Spinous Process Plates, or Shape Memory Alloy (SMA) Staples.
VI. Comparison of Technological Characteristics with Predicate Device
The subject CD HORIZON® Growth Rod Conversion Set implants have the same fundamental scientific technology as the predicate CD HORIZON® Growth Rod Conversion Set implants for the treatment of an identical patient population for the same indications of use. All of the implants included in this application were previously cleared in K133904 (SE 02/25/14), the only differences being that the subject implants will now be offered as pre-packaged sterile implants and the connectors will come with pre-loaded set screws. The subject connectors and set screws are identical in design, materials used to manufacture the implants, size, method of implantation and intended use. As stated previously the subject implants will be provided to the end user in a sterile state with sterilization performed via gamma irradiation.
VII. Performance Data
The following performance data are provided in support of the substantial equivalence determination.
Biocompatibilitv
The subject CD HORIZON® Growth Rod Conversion Set implants are permanent implants and are manufactured from stainless steel in accordance with ASTM F138 Standard Specification for Wrought 18Chromium-14Nickel- 2.5Molybdenum Stainless Steel Bar and Wire for Surgical Implants, or from titanium alloy in accordance with ASTM F136: Standard Specification for Wrought Ti-6A1-4V ELI Alloy for Surgical Implants. These materials are identical to the
{6}------------------------------------------------
materials used to manufacture the predicate CD HORIZON® Growth Rod Conversion Set implants cleared in K133904. According to the aforementioned standards, all of the materials used to manufacture the subject devices have a well characterized level of local biological response and a long history of clinical use in medical devices; therefore, no additional biocompatibility testing is required.
Mechanical Testing
No mechanical testing was performed on the subject connectors and set screws as no design changes were made to the implants. In accordance with, Guidance for Industry and FDA Staff -Spinal System 510(k)'s", Medtronic has evaluated the subject devices to demonstrate substantial equivalence to predicate device. It was determined that subject devices do not represent a new worst case. Packaging and sterilization assessment/rationales were used to demonstrate substantial equivalence in accordance with the following standards:
ANSI/AAMI/ISO 11137-1:2006/ A1:2013, Sterilization of health care products - Radiation -Part 1: Requirements for development, validation and routine control of a sterilization process for medical devices.
ANSI/AAMI/ISO 11137-2:2013, Sterilization of health care products - Radiation - Part 2: Establishing a sterilization dose.
ASTM F88/F88M-09: 2009 Standard Test Method for Seal Strength of Flexible Barrier Materials ASTM D4169-09 - 2009: Standard Practice for Performance Testing of Shipping Containers and Systems (Sterility).
ASTM F1929-98: 2004 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration.
As a new worst case has not been indicated, the sterilization and packaging rationales were deemed adequate to prove equivalence to the predicate device and no additional mechanical testing is required.
VIII. Conclusion
Based on a risk analysis, packaging and sterilization rationales, and additional supporting documentation provided in the pre-market notification, the subject CD HORIZON® Growth Rod Conversion Set components are substantially equivalent to the predicate devices found in K133904 (SE 02/25/14).