K Number
K150178

Validate with FDA (Live)

Date Cleared
2015-02-11

(16 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The CD HORIZON® SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined), spondylolisthesis, trauma, and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

Device Description

The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

The purpose of this 510(k) submission is to add additional transverse hooks to the CD HORIZON® Spinal System. These additional transverse hooks contain two changes from the predicate 2, CD HORIZON Spinal System, 1) a modified hook blade angle, and 2) packaged sterile via gamma irradiation. The subject sterile transverse hooks along with other components such as metal rods, screws, and other connecting components are intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of thoracic, lumbar, and/or sacral spine. The subject transverse hooks are manufactured out of medical grade titanium alloy per ASTM F136.

AI/ML Overview

The provided document is a 510(k) Summary for the Medtronic CD HORIZON® Spinal System, specifically focusing on the addition of new transverse hooks. This document outlines the device's characteristics, indications for use, and a comparison to predicate devices, but it does not contain information about acceptance criteria or a study proving the device meets specific performance criteria in the way a clinical trial or algorithm validation study would.

Instead, this is a regulatory submission for substantial equivalence to already approved devices. The "performance data" section focuses on biocompatibility and mechanical testing, which are assessed through engineering rationale and comparison to existing standards and predicate devices, rather than a study with acceptance criteria and reported device performance metrics in the context of diagnostic accuracy or clinical outcomes.

Here's an analysis based on the information provided, explicitly stating where information is not present:

1. A table of acceptance criteria and the reported device performance

  • Not Applicable in this document's context. This 510(k) submission does not present acceptance criteria and reported performance in the manner of a clinical or algorithm study (e.g., sensitivity, specificity, accuracy). Instead, it demonstrates substantial equivalence through:
    • Biocompatibility: Conformance to FDA's Draft Guidance for ISO-10993 and use of ASTM F136 compliant material with a long clinical history.
    • Mechanical Testing: Engineering rationale demonstrating that the subject devices do not represent a new worst-case compared to predicate devices, and thus no additional testing was required.

2. Sample size used for the test set and the data provenance (e.g., country of origin of the data, retrospective or prospective)

  • Not Applicable. This document describes a submission for device modification based on substantial equivalence, relying on engineering rationale and existing material standards rather than human subject data or a test set as defined for AI/algorithm performance.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g., radiologist with 10 years of experience)

  • Not Applicable. Ground truth in the context of a clinical or algorithm study is not relevant here. The evaluation is based on material science, mechanical properties, and regulatory guidance.

4. Adjudication method (e.g., 2+1, 3+1, none) for the test set

  • Not Applicable.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Not Applicable. This is not a study involving human readers or AI assistance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not Applicable. This involves a medical implant, not an algorithm.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

  • Not Applicable. The "truth" in this context is adherence to biomechanical principles, material standards (ASTM F136), and biological safety standards (ISO-10993), as well as being demonstrably equivalent to previously cleared predicate devices.

8. The sample size for the training set

  • Not Applicable. This is a hardware modification for a spinal implant, not an AI/algorithm.

9. How the ground truth for the training set was established

  • Not Applicable.

Summary of what the document does provide regarding device performance and acceptance:

The document states that the new transverse hooks for the CD HORIZON® Spinal System met the requirements for substantial equivalence by:

  • Biocompatibility: The devices were evaluated in accordance with FDA's Draft Guidance for ISO-10993. Since they are permanent implants (>30-day body contact) and manufactured from "identical materials as the predicate devices" (medical grade titanium alloy per ASTM F136), and this material has a "long clinical history of use in similar medical devices," no additional biocompatibility testing was required. This implies the acceptance criteria for biocompatibility were met by demonstrating material equivalence to a well-established, safe material.
  • Mechanical Testing: According to the "Guidance for Industry and FDA Staff - Spinal System 510(k)'s", Medtronic evaluated the devices. It was determined that the "subject devices do not represent a new worst case" and an "engineering rationale was used to demonstrate substantial equivalence." This means the acceptance criterion for mechanical performance was met by proving, through engineering analysis, that the modified hooks perform at least as well as the predicate devices and do not introduce new mechanical risks. No additional mechanical testing was required because the engineering rationale was deemed adequate.

The "study" that proves the device meets (or is equivalent to) the functional and safety requirements is the engineering rationale and comparison to predicate devices, along with adherence to recognized standards for materials (ASTM) and biological evaluation (ISO). There are no specific quantitative metrics or statistical analyses described in this particular section of the 510(k) summary for such a "study" as one might find for a diagnostic or AI device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, overlapping each other to create a sense of depth and unity. The faces are simple and abstract, with no distinct features other than the outline of the head and neck.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Medtronic Sofamor Danek USA, Incorporated Mr. Michael A. Scott Senior Regulatory Affairs Specialist 1800 Pyramid Place Memphis, Tennessee 38132

February 11, 2015

Re: K150178

Trade/Device Name: CD HORIZON® Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNI, MNH, KWP, KWQ Dated: January 23, 2015 Received: January 26, 2015

Dear Mr. Scott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Mr. Michael A. Scott

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K150178

Device Name CD HORIZON Spinal System

Indications for Use (Describe)

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain with degeneration of the disc confirmed by history and radios), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

Except for hooks, when used as an anterolateral the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spiral System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

When used for posterior non-cervical pediatic patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scolosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following spondylolisthesis/spondy/olysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devilt autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The CD HORIZON® SPIRE™ Plate is a posterior, single-level, non-pedicle supplemental fixation device intended for use in the noncervical spine (TI-St) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined), spondylolisthesis, trauma, and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

ວຍອາ ໄດ້ ປະຊາ ແລະ ແລະ ແລະ ອອກ ການ ຂະແດງສາ ປີ ທຸການ ຂອງປີ ປະຊາທິປຣາ ແລະວິຣະ ລາມ

. WOJER SSERODA JIAMS 47ATS ART OT MAO3 Q3T319MOJ RUOY GN32 TON OCT

:of ,пөртиф гіп) рпіриber тої глойгердна плівиколі ,пойзепоз поліветної під то to de la mar a sismit se nebud sint publisher stremator bases . noitemorini to noitest pre ətəlqmoz baş bərən ətəb ədə nisbilir. İstinadlar Şərqar siləsi Şarat və Şərqibəyləri və və və və və və və və və və və və və və və və və v ədi bildir. Əsnodsən rəq suran et balamizə zi növlərməlir. 10 növü. İstinadlar Sinfi 10 əmil nəbud əsr

vog.syy.spy@guglys.com ​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​​ Office of Chief Information Office Food and Drug Administration รรวมของ กรุงเทพมหานคร ที่ได้รับ to เทือบทางของ .

to noilsellos e 'o' broqser of beingen ton si nosted a pre ',osupros lon yem your your your your your your you " เอตเทนต BMO bilev บุปกษามว в ระธุรไตรเb วัน รรอไทม แดนิธยาวอังก

{4}------------------------------------------------

CD HORIZON® Spinal System 510(k) Summary

January 2015

I.Submitter:Medtronic Sofamor Danek USA, Inc.1800 Pyramid PlaceMemphis, TN 38132Telephone: (901)396-3133Fax: (901) 346-9738
Contact:Michael A. ScottSenior Regulatory Affairs Specialist
Date Prepared:January 23, 2015
II.Device
Name of Device:CD HORIZON® Spinal System
Common Name:Spinal Fixation System
Regulatory Classification:Class III (Pre-amendment)
Regulation Number,Classification Name,Product Codes:21 CFR 888.3070 Pedicle Screw Spinal System, NKB,OSH, MNI, MNH21 CFR 888.3060 Spinal Intervertebral Body FixationOrthosis, KWQ21 CFR 888.3050 Spinal Interlaminal Fixation Orthosis,KWP
III.Predicate DeviceK142591 (S.E. 12/01/2014) CD HORIZON® SpinalSystem - Primary Predicate
K132471 (S.E. 10/08/2013) CD HORIZON® SpinalSystem - Predicate 2, Additional Predicate
K132328 (S.E. 12/06/2013) CD HORIZON®Spinal System – Predicate 3, Additional Predicate
The predicate devices have not been subjectto a design related recall.

{5}------------------------------------------------

IV. Device Description:

The CD HORIZON® Spinal System consists of a variety of shapes and sizes of rods, hooks, screws, CROSSLINK® Plates, staples and connecting components, as well as implant components from other Medtronic spinal systems, which can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

A subset of CD HORIZON® Spinal System components may be used for posterior pedicle screw fixation in pediatric cases. These constructs may be comprised of a variety of shapes and sizes of rods (ranging in diameter from 3.5mm to 6.35mm), hooks, screws, CROSSLINK® Plates and connecting components. Similarly to the CD HORIZON® implants used in adult cases, these components can be rigidly locked into a variety of configurations, with each construct being tailor-made for the individual case.

The purpose of this 510(k) submission is to add additional transverse hooks to the CD HORIZON® Spinal System. These additional transverse hooks contain two changes from the predicate 2, CD HORIZON Spinal System, 1) a modified hook blade angle, and 2) packaged sterile via gamma irradiation. The subject sterile transverse hooks along with other components such as metal rods, screws, and other connecting components are intended to help provide immobilization and stabilization of spinal segments as an adjunct to fusion of thoracic, lumbar, and/or sacral spine. The subject transverse hooks are manufactured out of medical grade titanium alloy per ASTM F136.

V. Indications for Use:

The CD HORIZON® Spinal System with or without SEXTANT® instrumentation is intended for posterior, non-cervical fixation as an adjunct to fusion for the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, curvatures (i.e., scoliosis, kyphosis, or lordosis), tumor, pseudarthrosis, and/or failed previous fusion.

{6}------------------------------------------------

Except for hooks, when used as an anterolateral thoracic/lumbar system, the CD HORIZON® Spinal System may also be used for the same indications as an adjunct to fusion.

With the exception of degenerative disc disease, the CD HORIZON® LEGACY™ 3.5mm rods and the CD HORIZON® Spinal System PEEK rods and associated components may be used for the aforementioned indications in skeletally mature patients as an adjunct to fusion. The 3.5mm rods may be used for the specific pediatric indications noted below.

When used for posterior non-cervical pedicle screw fixation in pediatric patients, the CD HORIZON® Spinal System implants are indicated as an adjunct to fusion to treat progressive spinal deformities (i.e., scoliosis, kyphosis, or lordosis) including idiopathic scoliosis, neuromuscular scoliosis, and congenital scoliosis. Additionally, the CD HORIZON® Spinal System is intended to treat pediatric patients diagnosed with the following conditions: spondylolisthesis/spondylolysis, fracture caused by tumor and/or trauma, pseudarthrosis, and/or failed previous fusion. These devices are to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

The CD HORIZON® SPIRE™ Plate is a posterior, single level, non-pedicle supplemental fixation device intended for use in the non-cervical spine (T1-S1) as an adjunct to fusion in skeletally mature patients. It is intended for plate fixation/attachment to spinous processes for the purpose of achieving supplemental fixation in the following conditions: degenerative disc disease (as previously defined), spondylolisthesis, trauma, and/or tumor.

In order to achieve additional levels of fixation, the CD HORIZON® Spinal System rods may be connected to the VERTEX® Reconstruction System with the VERTEX® rod connector. Refer to the VERTEX® Reconstruction System Package Insert for a list of the VERTEX® indications of use.

{7}------------------------------------------------

VI. Comparison of Technological Characteristics with Predicate Device

The subject CD HORIZON® Spinal System transverse hooks have identical indications for use and intended use as the CD HORIZON® Spinal System primary predicate most recently cleared by the FDA in K142591 (S.E. 12/01/2014). Like the CD HORIZON Spinal System hooks found in the predicate 2 submission (K132471 {S.E. 10/08/2013}), the subject transyerse hooks can accept both 5.5mm and 6.0mm diameter rods, have an anatomic design that mimics the posterior spinal elements and are used with other connecting components to help provide immobilization and stabilization of spinal segments as an adjunct to fusion. Additionally, the new transverse hooks are provided sterile identical to that of the predicate 3 devices found in K132328 (S.E. 12/06/2013). Therefore, the technological characteristics of the subject device are the same as the technological characteristics of the predicate devices.

VII. Performance Data

The following performance data are provided in support of the substantial equivalence determination.

Biocompatibility

The biocompatibility evaluation for the CD HORIZON® Spinal System devices was conducted in accordance with FDA's Draft Guidance for Industry and FDA Staff "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing" issued April, 23, 2013.

The subject CD HORIZON® Spinal System transverse hooks are permanent implants and will be classified as permanent, >30 day body contact according to with FDA's Draft Guidance for Industry and FDA Staff "Use of International Standard ISO-10993, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing". The subject transverse hooks are manufactured from identical materials as the predicate devices, in accordance with the following ASTM standard:

{8}------------------------------------------------

ASTM F136 - Standard Specification for Wrought Titanium-6 Aluminum-4 Vanadium ELI (Extra Low Interstitial) Alloy for Surgical Implant Applications.

The titanium alloy material used for the subject CD HORIZON® Spinal System Transverse Hooks has a long clinical history of use in similar medical devices. Therefore, no additional biocompatibility testing is required.

Mechanical Testing

In accordance with, Guidance for Industry and FDA Staff - Spinal System 510(k)'s", Medtronic has evaluated the subject devices to demonstrate substantial equivalence to the predicate devices. It was determined that subject devices do not represent a new worst case. An engineering rationale was used to demonstrate substantial equivalence. As a new worst case has not been indicated and an engineering rationale was deemed adequate to prove equivalence to the predicate device. no additional mechanical testing is required.

Conclusion VIII.

Based on a risk analysis, engineering rationale, and additional supporting documentation provided in the pre-market notification, the subject CD HORIZON® Spinal System are substantially equivalent to the following predicates: K142591 (S.E. 12/01/2014) – Primary Predicate, and additional predicates K132471 (S.E. 10/08/2013) - Predicate 2, and K132328 (S.E. 12/06/2013) – Predicate 3.

N/A