K Number
K150151

Validate with FDA (Live)

Date Cleared
2015-05-21

(118 days)

Product Code
Regulation Number
878.4040
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Cardinal Health SmartGown™ surgical gown is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The Cardinal Health SmartGown™ surgical gown is a single use, disposable medical device provided sterile and non-sterile.

Device Description

The Cardinal Health SmartGown™ surgical gowns are constructed of a multi-layer construction of a nonwoven outer layer, breathable film core and a nonwoven inner layer and have been tested according to AAMI PB70:2012 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health care Facilities. The Cardinal Health SmartGown™ surgical gown is a single use, disposable medical device that will be provided in a variety of sterile and non-sterile packaging configurations. This submission covers 12 models of Cardinal Health SmartGown surgical gowns, see Table 1 below.

AI/ML Overview

This document describes the testing and acceptance criteria for the Cardinal Health SmartGown™ surgical gown. Here's a breakdown of the requested information:

1. Table of Acceptance Criteria and Reported Device Performance

The acceptance criteria for the Cardinal Health SmartGown™ surgical gown are established by various ASTM and AATCC standards, as well as the AAMI PB70:2012 Level 4 requirements. The reported device performance confirms that the gowns meet or exceed these criteria.

Feature / TestAcceptance CriteriaReported Device Performance (Mean, Min/Max) - BodyReported Device Performance (Mean, Min/Max) - SleeveReported Device Performance (Mean, Min/Max) - Back Panel
Physical Properties (Critical Zones - Body & Sleeve)
ASTM D3776 Basis Weight (gsm)Body: 50.9 Mean min; Sleeve: 65.1 Mean minBody: 52.1 (51.3 / 53.1)Sleeve: 69.6 (69.0 / 70.3)N/A (Not a critical zone for this test)
ASTM D5034 Grab Tensile CD (lb)Body: 14.4 Mean min; Sleeve: 18.0 Mean minBody: 17.8 (15.9 / 20.0)Sleeve: 23.9 (21.9 / 26.0)N/A
ASTM D5733 Trap Tear Peak MD (lb)Body: 4.0 Mean min; Sleeve: 5.4 Mean minBody: 5.6 (4.4 / 6.8)Sleeve: 7.7 (6.0 / 9.9)N/A
ASTM D5733 Trap Tear Peak CD (lb)N/A (MD Trap tear is limiting specification value)Body: 8.9 (6.9 / 10.4)Sleeve: 10.7 (9.0 / 12.5)N/A
ASTM D774 Mullen Burst (psi)Body: 30.0 Mean min; Sleeve: 36.9 Mean minBody: 34.3 (31.2 / 37.9)Sleeve: 45.1 (39.5 / 49.4)N/A
ASTM E96 WVTR, 23°C, 50%RH (g/m²/24hours)Body: 740 Mean min; Sleeve: 708 Mean minBody: 822 (752 / 866)Sleeve: 764 (724 / 814)N/A
ASTM E96 WVTR, 27°C, 50%RH (g/m²/24hours)Body: 931 Mean min; Sleeve: 833 Mean minBody: 983 (951 / 1022)Sleeve: 918 (876 / 983)N/A
ASTM E96 WVTR, 32°C, 50%RH (g/m²/24hours)Body: 1191 Mean min; Sleeve: 1117 Mean minBody: 1439 (1323 / 1618)Sleeve: 1251 (1185 / 1426)N/A
16 CFR Part 1610 FlammabilityBody: Class 1; Sleeve: Class 1Body: Class 1Sleeve: Class 1Back Panel: Class 1
AAMI PB70 Barrier Performance LevelBody: Level 4; Sleeve: Level 4Body: Level 4Sleeve: Level 4N/A (Back panel is Level 3 for specialty gown)
Physical Properties (Non-Critical Zone - SMS Back panel)
ASTM D3776 Basis Weight (gsm)Back Panel: 31.5 Mean minN/AN/ABack Panel: 34.6 (33.4/35.8)
AATCC-42 Water Impact (g)Back Panel: 0.5 Mean maxN/AN/ABack Panel: 0.12 (0.08/0.20)
ASTM D5034 Grab Tensile CD (lb)Back Panel: 10.0 Mean minN/AN/ABack Panel: 12.5 (9.2/14.8)
ASTM D5733 Trap Tear Peak MD (lb)Back Panel: 3.0 Mean minN/AN/ABack Panel: 3.7 (2.8/5.1)
AATCC 127 Hydrostatic Head (cm)Back Panel: 50 Mean minN/AN/ABack Panel: 69.0 (50.7/78.7)
Liquid Barrier Performance Classification Properties
ASTM F1671/F1671M-13 (Level 4 compliance)AAMI PB70:2012 Level 4 requirementsPassPassPass (for critical zones)

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: For almost all tests (Basis Weight, Grab Tensile, Mullen Burst, WVTR, Flammability, Trap Tear, Water Impact, Hydrostatic Pressure, ASTM F1670, ASTM F1671), the sample size used was n=32.
  • Data Provenance: The document does not specify the country of origin of the data. However, it indicates that the tests were performed on "Final finished goods, that were twice EO sterilized (Cardinal Health SmartGown™ surgical gown)". This suggests the data is prospective as it's from testing the manufactured device against established performance standards.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Their Qualifications

This is a medical device (surgical gown) performance testing, not a diagnostic AI study. Therefore, there are no human experts establishing "ground truth" in the traditional sense of medical image interpretation or clinical diagnosis. The "ground truth" here is the adherence to established internationally recognized material testing standards and regulatory guidance (ASTM, AATCC, AAMI PB70:2012, FDA Guidance). The experts involved would be the testing laboratories and their personnel following these standard protocols. The document does not specify their individual qualifications beyond executing the tests according to the standards.

4. Adjudication Method for the Test Set

Not applicable. As this is not a study involving human interpretation or clinical outcomes, there is no adjudication method in the context of resolving differing expert opinions. The outcome of each test is a direct measurement or observation against a predefined standard.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

Not applicable. This is not an AI-powered diagnostic or assistive device. It is a surgical gown, and its performance is evaluated through standardized physical and barrier protection tests, not through human reader interpretation.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This device is not an algorithm or AI. It is a physical medical device.

7. The Type of Ground Truth Used

The ground truth used for this device's acceptance is based on established material testing standards and regulatory performance classifications:

  • ASTM International Standards: Various ASTM standards (e.g., D3776, D5034, D5733, D774, E96, F1670, F1671) define the methods for testing physical properties and barrier resistance.
  • AATCC Standards: AATCC standards (e.g., 42, 127) define methods for water resistance.
  • AAMI PB70:2012 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes: This standard sets the performance levels for surgical apparel, with the device aiming for Level 4.
  • 16 CFR Part 1610 (Flammability): A regulatory standard for flammability.

The device's performance is compared directly against the specified requirements within these standards.

8. The Sample Size for the Training Set

Not applicable. This is not a machine learning or AI device that requires a training set. The device is a physical product whose performance is evaluated through empirical testing against established standards.

9. How the Ground Truth for the Training Set Was Established

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 21, 2015

Cardinal Health 200, LLC. Ms. Lavenia Ford Manager, Regulatory Affairs 1500 Waukegan Road Waukegan, IL 60685

Re: K150151 Trade/Device Name: SmartGown™ Surgical Gown Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: FYA. Dated: April 23, 2015 Received: April 24, 2015

Dear Ms. Ford,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply

{1}------------------------------------------------

with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the Cardinal Health logo. The logo consists of a stylized graphic element above the text "CardinalHealth". The graphic element appears to be a series of curved lines, possibly representing a bird in flight or a stylized representation of health and wellness. The text "CardinalHealth" is written in a bold, sans-serif font.

1500 Waukegan Road

Waukegan, IL 60085

www.cardinalhealth.com

510(k) SUMMARY SmartGown™ surgical gown

Manufacturer:

Cardinal Health 200, LLC 1500 Waukegan Road Waukegan, IL 60085

1500 Waukegan Road Waukegan, IL 60085

Regulatory Affairs Contact:

Telephone Number:

(847) 887-3323

Lavenia Ford

(847) 785-2461

April 21, 2015 Date summary Prepared:

Trade Name:

Fax Number:

SmartGown™ surgical gown

Regulation Number/Device Class: Class II per 21 CFR § 878.4040

FYA

Surgical Apparel Regulation Name:

Surgical Gown Common Name:

Product Code:

Predicate Device:

K012984 -Cardinal Health SmartGown™

{3}------------------------------------------------

Description

The Cardinal Health SmartGown™ surgical gowns are identified by Regulation 21 CFR 878.4040 with product code FYA.

The Cardinal Health SmartGown™ surgical gowns are constructed of a multi-layer construction of a nonwoven outer layer, breathable film core and a nonwoven inner layer and have been tested according to AAMI PB70:2012 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health care Facilities. The Cardinal Health SmartGown™ surgical gown is a single use, disposable medical device that will be provided in a variety of sterile and non-sterile packaging configurations. This submission covers 12 models of Cardinal Health SmartGown surgical gowns, see Table 1 below.

Indications for Use

Cardinal Health SmartGown™ surgical gown is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The Cardinal Health SmartGown™ surgical gown is a single use, disposable medical device provided sterile and non-sterile.

This submission covers 12 models of Cardinal Health SmartGown™ surgical gown, see Table 1 below. Each model is a multi-layer construction of a nonwoven outer layer, breathable film core and a nonwoven inner layer, and has been tested according to AAMI PB70:2012 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health Care Facilities.

Catalog #
SterileNon-sterile
SingleBulkHospitalBulk Small QtyModel DescriptionModel Size
89013N/AN/AN/ASmartGownTM surgical gown, Set-in sleeveX-Small
8900589005NN/AN/ASmartGownTM surgical gown, Set-in sleeveSmall/Medium
8901589015NN/A890015NSmartGownTM surgical gown, Set-in sleeveLarge
8904589045NN/A890045NSmartGownTM surgical gown, Set-in sleeveX-Large
8907589075NK89075NN/ASmartGownTM surgical gown, Set-in sleeveXX-Large
3901539015NK39015NN/ASmartGownTM surgical gown, Raglan sleeveLarge
3904539045NN/AN/ASmartGownTM surgical gown, Raglan sleeveX-Large
3904939049NK39049NN/ASmartGownTM surgical gown, Raglan sleeveX-Large, X-Long
3907539075NN/AN/ASmartGownTM surgical gown, Raglan sleeveXX-Large
3907939079NN/AN/ASmartGownTM surgical gown, Raglan sleeveXX-Large, X-Long
3909939099NN/AN/ASmartGownTM surgical gown, Raglan sleeveXXX-Large, X-Long
3247432474NK32474N324740NSmartGownTM surgical gown (specialty),Raglan sleeveX-Large, X-Long,A-Line

Table 1: Product Description and Catalog Number

The Cardinal Health SmartGown™ surgical gown is a single use, disposable medical device provided in a variety of sterile and non-sterile packaging configurations. Bulk non-sterile Cardinal Health SmartGown™ surgical gowns provided to convenience kit packers.

{4}------------------------------------------------

Non-sterile Cardinal Health SmartGown™ surgical gowns will include EO sterilization parameters on labeling as follows:

  • EO Concentration: 690 mg/L -
  • Temperature: 130 +/- 10ºF ﻴ
  • Exposure Time: 150 minutes ﻴ
  • Humidity: 50 +/- 5% ﺳ
  • Aeration Time: 18 hours 13

Device and Predicate Device Technical Characteristics

The proposed Cardinal Health SmartGown™ surgical gowns are primarily construction of nonwoven and nonwoven laminates (see Table 3).

Table 3: Proposed Cardinal Health SmartGown™ surgical gown description

DesignBody and front tieattachment AAMI PB70Critical ZoneSleeve and sleeveseam AAMI PB70Critical ZoneBack Panel AAMIPB70 Non-CriticalZoneCodes
Set-in sleeveSMS / Film / SMSMeets:ASTM F1671AAMI PB70 Level 4SB / Film / SBMeets:ASTM F1671AAMI PB70 Level 4SMS / Film / SMSMeets:ASTM F1671AAMI PB70 Level 4890138900589015890458907589005N89015N89045N89075N890015N890045NK89075N
Raglan sleeveSMS/ Film / SMSMeets:ASTM F1671AAMI PB70 Level 4SB / Film / SBMeets:ASTM F1671AAMI PB70 Level 4SMS / Film / SMSMeets:ASTM F1671AAMI PB70 Level 439015390453904939075390793909939015N39045N39049N39075N39079N39099NK39015NK39049N
Raglansleeve, AlineSMS / Film / SMSMeets:ASTM F1671AAMI PB70 Level 4SB / Film / SBMeets:ASTM F1671AAMI PB70 Level 4SMSMeets:AATCC 42 \u2264 1.0gAATCC 127 \u226550cmAAMI PB70 Level 33247432474N324740NK32474N

{5}------------------------------------------------

The Cardinal Health SmartGown™ surgical gowns consist of zones; body, sleeve and back panel. The body front and lower sleeve are critical per AAMI PB70:2012 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health care Facilities, refer to Table 3 below.

The Cardinal Health SmartGown™ surgical gowns are substantially equivalent to the predicate SmartGown surgical gown with regards to claims, safety and effectiveness, design, technology, and intended use. See Table 4 below.

The proposed Cardinal Health SmartGown™ surgical gown body and sleeve materials are a laminated structure consisting of an outer layer of polyolefin nonwoven, laminated to a monolithic breathable barrier film, which is laminated to an inner layer of polyolefin nonwoven. The body material is constructed of an outer layer of SMS polyolefin nonwoven, laminated to a monolithic breathable barrier film, which is laminated to an inner layer of SMS polyolefin nonwoven. The gown sleeves are constructed of an outer layer of SB polyolefin nonwoven, laminated to a monolithic barrier film, which is laminated to an inner layer of SB polyolefin

: nonwoven.

The proposed Cardinal Health SmartGown™ surgical gown (specialty) non-critical zone body back panels are constructed of polyolefin SMS nonwoven fabric. The surgical gown back panels are not a critical zone per AAMI PB70:2012.

Testing was performed according to the Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993 and AAMI PB70: June 21, 2012, Liquid Barrier Performance Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities. All results of testing met ASTM F1671/F1671M-13, and meets AAMI PB70:2012 Level 4 requirements.

Table 4: Side by Side Comparison of Predicate device and Proposed Cardinal Health SmartGown™ surgical gown

Element of ComparisonPredicate Cardinal HealthSmartGown™ surgical gownProposed Cardinal HealthSmartGown™ surgical gown
SmartGown™ surgical gownSmartGown™ surgical gown

{6}------------------------------------------------

Intended UseThe predicate Cardinal HealthSmartGown™ surgical gown isintended to be worn by operating roompersonnel during surgical proceduresto protect both the surgical patient andthe operating room personnel fromtransfer of microorganisms, bodyfluids, and particulate material.The predicate Cardinal HealthSmartGown™ surgical gown isintended for use in infection controlpractices to reduce the potentialexposure of the wearer to blood andbody fluids.The predicate Cardinal HealthSmartGown™ surgical gown is a singleuse, disposable medical device,provided sterile and non-sterile.The proposed Cardinal HealthSmartGown™ surgical gown isintended to be worn by operating roompersonnel during surgical proceduresto protect both the surgical patient andthe operating room personnel fromtransfer of microorganisms, bodyfluids, and particulate material.The proposed Cardinal HealthSmartGown™ surgical gown isintended for use in infection controlpractices to reduce the potentialexposure of the wearer to blood andbody fluids.The proposed Cardinal HealthSmartGown™ surgical gown is a singleuse, disposable medical device,provided sterile and non-sterile.
Material CompositionPredicate Cardinal HealthSmartGown™ surgical gown criticalzones are a multi-layer construction ofnonwoven outer (polyolefin spunmelt),monolithic film core (copolyester),nonwoven inner (carded polyester).Proposed Cardinal HealthSmartGown™ surgical gown bodycritical zone is a multi-layerconstruction of nonwoven outer(polyolefin SMS), monolithic film core(copolyester), nonwoven inner(polyolefin SMS).Proposed Cardinal HealthSmartGown™ surgical gown sleevescritical zone are multi-layerconstruction of nonwoven outer (SB),monolithic film core, nonwoven inner(SB).
Design FeaturePredicate Cardinal HealthSmartGown™ surgical gown isprovided with neck binding with hookand loop tabs, belt ties, removabletransfer tab accessory, and cuffsProposed Cardinal HealthSmartGown™ surgical gown isprovided with neck binding with hookand loop tabs, belt ties, removabletransfer tab accessory, and cuffs.Proposed Cardinal HealthSmartGown™ surgical gown(specialty) non critical zone backpanels are polyolefin SMS nonwovenfabric.
Element of ComparisonPredicate Cardinal HealthSmartGown™ surgical gownProposed Cardinal HealthSmartGown™ surgical gown
Performance(+/- 3 Sigma)Test ResultsMean (min / max)Material Specification
PhysicalProperties(Critical)ASTM D3776 Basisweight (gsm)67.81 - 88.15Body: 52.1 (51.3 / 53.1)Sleeve: 69.6 (69.0 / 70.3)Body: 50.9 Mean minSleeve: 65.1 Mean min
AAMIPB70:2012)ASTM D5034 Grabtensile CD (lb)14.9 – 20.3Body: 17.8 (15.9 / 20.0)Sleeve: 23.9 (21.9 / 26.0)Body: 14.4 Mean minSleeve: 18.0 Mean min
ASTM D5733 TrapTear Peak MD (lb)Performance values notavailable in predicate 510(k)submissionBody: 5.6 (4.4 / 6.8)Sleeve: 7.7 (6.0 / 9.9)Body: 4.0 Mean min **Sleeve: 5.4 Mean min **
ASTM D5733 TrapTear Peak CD (lb)Performance values notavailable in predicate 510(k)submissionBody: 8.9 (6.9 / 10.4)Sleeve: 10.7 (9.0 / 12.5)N/A **
ASTM D774 Mullenburst (psi)32.1 – 59.1Body: 34.3 (31.2 / 37.9)Sleeve: 45.1 (39.5 / 49.4)Body: 30.0 Mean minSleeve: 36.9 Mean min
ASTM E96 WVTR,upright cup @ 23°C,50%RH (g/m²/24hours)640 - 1007Body: 822 (752 / 866)Sleeve: 764 (724 / 814)Body: 740 Mean minSleeve: 708 Mean min
ASTM E96 WVTR,upright cup @ 27°C,50%RH (g/m²/24hours)820 - 1332Body: 983 (951 / 1022)Sleeve: 918 (876 / 983)Body: 931 Mean minSleeve: 833 Mean min
ASTM E96 WVTR,upright cup @ 32°C,50%RH (g/m²/24hours)1005 – 1707Body: 1439 (1323 / 1618)Sleeve; 1251 (1185 / 1426)Body: 1191 Mean minSleeve; 1117 Mean min
(CPSC), Part 1610FlammabilityClass 1Body: Class 1Sleeve: Class 1Body: Class 1Sleeve: Class 1
AAMI PB70 BarrierPerformance LevelPerformance standard notavailable at time of predicatesubmission.Body: Level 4Sleeve: Level 4Body: Level 4Sleeve: Level 4
PhysicalProperties(SMS Backpanel Non-CriticalZone perAAMIPB70:2012)ASTM D3776 Basisweight (gsm)Performance values notavailable in predicate 510(k)submissionBack Panel: 34.6 (33.4/35.8)Back Panel: 31.5 Meanmin
AATCC-42 WaterImpact (g)Performance values notavailable in predicate 510(k)submissionBack Panel: 0.12 (0.08/0.20)Back Panel: 0.5 Mean max
ASTM D5034 Grabtensile MD (lb)Performance values notavailable in predicate 510(k)submissionBack Panel: 18.8 (15.8/21.5)N/A *
ASTM D5034 Grabtensile CD (lb)Performance values notavailable in predicate 510(k)submissionBack Panel: 12.5 (9.2/14.8)Back Panel: 10.0 Mean min
ASTM D5733 TrapTear Peak MD (lb)Performance values notavailable in predicate 510(k)submissionBack Panel: 3.7 (2.8/5.1)Back Panel: 3.0 Meanmin
ASTM D5733 TrapTear Peak CD (lb)Performance values notavailable in predicate 510(k)submissionBack Panel: 6.0 (4.8 / 7.8)N/A **
AATCC 127Hydrostatic Head (cm)Performance values notavailable in predicate 510(k)submissionBack Panel: 69.0 (50.7/78.7)50 Mean min
(CPSC), Part 1610FlammabilityPerformance values notavailable in predicate 510(k)submissionBack Panel: Class 1Class 1
* MD Grab tensile not specified, CD Grab tensile is limiting specification value** CD Trap tear not specified, MD Trap tear is limiting specification value
Liquid Barrier PerformanceClassification PropertiesPredicate device was tested accordingto ASTM F1671-97 in previous 510(k)submission K012984.Device was tested in accordance withASTM F1671/F1671M-13, and meetsAAMI PB70:2012 Level 4

,

{7}------------------------------------------------

:

{8}------------------------------------------------

requirements.
Sterilization ModalityEthylene OxideEthylene Oxide
BiocompatibilityPass ISO 10993-1Pass ISO 10993-1

Table 4 highlights the equivalencies and comparison between proposed Cardinal Health SmartGown™ surgical gown and predicate Cardinal Health SmartGown™ surgical gown.

Final finished goods, that were twice EO sterilized (Cardinal Health SmartGown1™ surgical gown), were tested. The products were tested in the critical zones of the body and sleeve to generate test data shown in Table 4 above, including back panels where noted. Test results establish that the product meets predetermined acceptance criteria of specifications for intended use, to demonstrate device is safe and effective, as noted in Table 4 above.

Mass Per Unit Area (Basis Weight) of Woven Fabric .

ASTM D3776-09 Test Methods for Mass Per Unit Area (Weight) of Woven Fabric results were reported on the Body, Sleeve and Back Panel materials. This method was used to characterize the basis weight of each fabric. The basis weight test indicates how much a square meter of the fabric weighs. Each test specimen is cut to a specific size and the mass of the specimen is measured on a laboratory scale. The basis weight is calculated from that information. Results are reported in grams per square meter. The sample size used was n=32 and results were reported in mean, min and max.

Breaking Strength and Elongation of Textile Fabrics (Grab tensile) .

ASTM D5034-09 Standard Test Method for Breaking Strength and Elongation of Textile Fabrics (Grab Test) results were reported on the Body, Sleeve and Back Panel materials. This method was used to characterize the tensile strength of each fabric. The grab tensile test indicates the force a fabric can withstand before tearing when pulled in a given direction. The test was performed in both roll orientations; cross direction (CD) and machine direction (MD). The test specimen is placed in the clamps of a tensile testing machine. The test specimen is subjected to an increasing force as the jaws move apart at a specified constant rate. Results are reported as peak force values. Peak values reflect the maximum force the sample endured before the initiation of failure. Results are reported in pounds of force. The sample size used was n=32 and results were reported in mean, min and max.

Bursting Strength of Textile Fabrics (Mullen burst) .

{9}------------------------------------------------

ASTM D774/D774M-97 (2007) Standard Test Method for Bursting Strength of Textile Fabrics (Mullen Burst) results were reported on the Body and Sleeve materials. This method was used to characterize the burst strength of each fabric. The bursting strength test indicates how the fabric will resist puncture by a blunt object. Each fabric specimen is clamped over a flexible diaphragm. The diaphragm is expanded by pressure to the point of specimen rupture. The difference between the total pressure required to rupture the specimen and the pressure on the diaphragm at time of specimen burst with clamp released is reported as the Mullen burst strength. Results are reported in pounds per square inch. The sample size used was n=32 and results were reported in mean, min and max.

Water Vapor Transmission of Materials, Upright cup ●

ASTM E96/E96M-13 Standard Test method for Water Vapor Transmission of Materials, Upright cup (WVTR) results were reported on the Body and Sleeve materials. This method was used to characterize the water vapor transmission rate of each fabric over a range of temperatures at 50%RH (23°C, 27°C, 32°C). The water vapor transmission rate test specimen is affixed over a sample cup of water. The prepped sample cup is weighed at time 0. The prepped sample cup is then is heated to the target temperature, in chamber with constant %RH. The prepped sample cup is then measured at predetermined time points to measure the mass loss (water vapor transmission thru the sample). The water vapor transmission rate results are reported as mass of water per unit of area per period of time (g/m2/24 hours). The sample size used was n=32 and results were reported in mean, min and max.

Standard for the Flammability of Clothing Textiles ●

16 CFR Part 1610, Standard for the Flammability of Clothing Textiles, provides methods of testing the flammability of clothing and textiles intended to be used for clothing, and establishes three classes of flammability. The standard sets forth the requirements which textiles shall meet to be so classified. Class 1 is the best performing class. Textiles meeting these requirements are generally accepted by the trade as having no unusual burning characteristics. The flammability test indicates the inherent flammability classification of the surgical gown. Flammability results were reported on the Body, Sleeve and Back Panel materials. The sample size used was n=32 and results were reported as the Flammability Classification Level.

Tearing Strength of Nonwoven Fabrics by the Trapezoid Procedure (Trap tear) �

ASTM D5733-99 Standard Test Method for Tearing Strength of Nonwoven Fabrics by the Trapezoid Procedure results were reported on the Back Panel material. This method

{10}------------------------------------------------

was used to characterize the tear strength of the material. The trapezoidal tear test indicates how the fabric will resist tearing around an existing hole or tear. The test was performed in both roll orientations; cross direction (CD) and machine direction (MD). A tear of specified size and orientation is initiated on the test specimen. The test specimen is placed in the clamps of a tensile testing machine. The test specimen is subjected to an increasing force as the jaws move apart at a specified constant rate. Results are reported as average trapezoidal tear strength, in pounds of force. The sample size used was n=32 and results were reported in mean, min and max.

. Water Resistance: Impact Penetration Test

AATCC 42-2013 Water Resistance: Impact Penetration Test results were reported on the Back Panel material. This method was used to characterize the water resistance to impact penetration of the material. The impact penetration test measures the fabric's resistance to penetration by liquids, which is a key barrier property. The test specimen and a fluid blotter paper are cut to a predetermined size. The blotter is then weighed prior to testing. The test specimen and blotter are then clamped onto the test fixture (test specimen on top of blotter). A predetermined amount of water is allowed to spray on the sample from a fixed distance and spray head. The blotter is then weighed to determine the amount of water penetration thru the sample. The amount of liquid penetration is the final blotter weight less the initial blotter weight. The results are reported in grams. The sample size used was n=32 and results were reported in mean. min and max.

Water Resistance: Hydrostatic Pressure Test .

AATCC 127:2013 Water Resistance: Hydrostatic Pressure Test results were reported on the Back Panel material. This method was used to characterize the resistance of a fabric to the penetration of water under hydrostatic pressure, which is a key barrier property. The test specimen is clamped the test chamber, which is filled with water to contact the sample. The water pressure is gradually increased, exerting greater pressure onto the fabric. The unexposed side of the fabric is observed for water droplets, signaling failure of the barrier feature of the test fabric. Eventually, leakage occurs, marking the end of the test. The greater water pressure, the more repellent the fabric. Results are reported in centimeters of water. The sample size used was n=32 and results were reported in mean, min and max.

  • Resistance of Materials Used in Protective Clothing to Penetration by Synthetic . Blood - Gown

{11}------------------------------------------------

ASTM F1670-08, Standard Test Method for Resistance of Materials Used in Protective Clothing to Penetration by Synthetic Blood, evaluates the effectiveness of materials used in protective clothing for protecting the wearer against contact with body fluids that potentially contain blood-borne pathogens. This test method is intended to identify protective clothing material candidates for further testing according to a more rigorous procedure involving a surrogate for blood-borne pathogens. The test specimen is clamped onto the open face of pressure chamber of defined size, with a retaining ring around the outer circumference of the specimen. The pressure chamber is filled with synthetic blood. The specimen is exposed to the synthetic blood for a defined dwell period of time at ambient pressure. Then, chamber is pressurized, at a specified constant rate, to a maximum of 2 psi, for a defined period of time. The chamber is then allowed to revert to ambient pressure, and the specimen is exposed to the synthetic blood for a defined dwell period of time, to the conclusion of the test. A mesh screen may be used to support the specimen. Throughout the cycle, the reverse side of the fabric is observed for synthetic blood droplets, signaling failure of the barrier feature of the test fabric. If no droplets are observed thru the entire cycle, the sample passes. Results are reported as pass/fail. Procedure B was used, with a mesh screen to support the sample during testing. The sample size used was n=32 and results were reported as pass/fail.

Resistance of Materials Used in Protective Clothing to Penetration by Blood-Borne Pathogens Using Phi-X174 Bacteriophage Penetration as a Test System - Gown

ASTM F1671/F1671M-13, Standard Test Method for Resistance of Materials Used in Protective Clothing to Penetration by Blood-Borne Pathogens Using Phi-X174 Bacteriophage Penetration as a Test System, evaluates the effectiveness of materials used in protective clothing for protecting the wearer against contact with blood-borne pathogens using a surrogate microbe suspended in a body fluid simulant under conditions of continuous contact. The test specimen is clamped onto the open face of a pressure chamber, with a retaining ring around the outer circumference of the specimen. The pressure chamber is filled with fluid test media containing a surrogate microbe. The specimen is exposed to the fluid media for a defined dwell period of time under ambient pressure. Then, chamber is pressurized, at a specified constant rate, to a maximum of 2 psi, for a defined period of time. The chamber is then allowed to revert to ambient pressure, and the specimen is exposed to the fluid media for a defined dwell period of time, to the conclusion of the test. A mesh screen may be used to support the specimen. Throughout the cycle, the reverse side of the fabric is observed for fluid media droplets, signaling failure of the barrier feature of the test fabric. If no droplets are observed thru the entire cycle, the sample will then be plated for bacteriological assay. A specimen showing liguid penetration constitutes a failure. A specimen assay identifying evidence of viable microbe that penetrates the material, even when liquid penetration is not visible, constitutes a failure. Results are reported as passffail.

{12}------------------------------------------------

This method was used to characterize the AAMI PB70;2012 Liquid Barrier Performance Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities performance level, as compliant to level 4. Procedure B was used, with a mesh screen to support the sample during testing. The sample size used was n=32 and results were reported as pass/fail.

Nonwoven and nonwoven laminate materials are characterized by basis weight as opposed to thickness as basis weight is a more precise method to characterize nonwoven and nonwoven laminate performance. This is in compliance with physical specification requirements outlined in Section 3.a.1 of Guidance on Premarket Notification [510(k)] Submissions for Surgical Gowns and Surgical Drapes, issued on August 1, 1993, where it states that weight per square yard or thickness, may be used as applicable.

Based on ASTM F2407-06R13. Table 2. Note B. there are no generally accepted test methods for snag or puncture resistance available at this time for nonwoven laminates.

The proposed product does not affect the substantial equivalence of the device, as explained below:

  • · Intended use
    The proposed Cardinal Health SmartGown™ surgical gown intended use is substantially equivalent to the predicate Cardinal Health SmartGown™ surgical gown. Intended use of proposed is the same as the predicate. The Cardinal Health SmartGown™ surgical gown is a sterile, single use, disposable medical device. Cardinal Health SmartGown™ surgical gown is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material.

Conclusion Statement

Based on the results of the biocompatibility and physical performance testing the Cardinal Health SmartGown™ surgical gowns are as safe and as effective for their intended use as the predicate device. The Cardinal Health SmartGown™ surgical gowns are substantially equivalent to the predicate device, in terms of general intended use performance testing, material composition, configuration/dimensions and safety and effectiveness.

{13}------------------------------------------------

Indications for Use

510(k) Number (if known) K150151

Device Name SmartGownTM surgical gown

Indications for Use (Describe)

Cardinal Health SmartGown™ surgical gown is intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. The Cardinal Health SmartGown™ surgical gown is a single use, disposable medical device provided sterile and non-sterile.

This submission covers 12 models of Cardinal Health SmartGown™ surgical gown, see Table 1, Each model is a multilaver construction of a nonwoven outer layer, breathable film core and a nonwoyen inner laver, and has been tested according to AAMI PB70:2012 Liquid Barrier and Performance Classification of Protective Apparel and Surgical Drapes Intended for Use in Health Care Facilities.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{14}------------------------------------------------

Table 1: Product Description and Catalog Number

Catalog #
SterileNon-sterile
SingleBulkHospitalBulk Small QtyModel DescriptionModel Size
89013N/AN/AN/ASmartGownTM surgical gown, Set-in sleeveX-Small
8900589005NN/AN/ASmartGownTM surgical gown, Set-in sleeveSmall/Medium
8901589015NN/A890015NSmartGownTM surgical gown, Set-in sleeveLarge
8904589045NN/A890045NSmartGownTM surgical gown, Set-in sleeveX-Large
8907589075NK89075NN/ASmartGownTM surgical gown, Set-in sleeveXX-Large
3901539015NK39015NN/ASmartGownTM surgical gown, Raglan sleeveLarge
3904539045NN/AN/ASmartGownTM surgical gown, Raglan sleeveX-Large
3904939049NK39049NN/ASmartGownTM surgical gown, Raglan sleeveX-Large, X-Long
3907539075NN/AN/ASmartGownTM surgical gown, Raglan sleeveXX-Large
3907939079NN/AN/ASmartGownTM surgical gown, Raglan sleeveXX-Large, X-Long
3909939099NN/AN/ASmartGownTM surgical gown, Raglan sleeveXXX-Large, X-Long
3247432474NK32474N324740NSmartGownTM surgical gown (specialty), Raglan sleeveX-Large, X-Long, A-Line

The Cardinal Health SmartGown™ surgical gown is a single use, disposable medical device provided in a variety of sterile and non-sterile packaging configurations. Bulk non-sterile Cardinal Health SmartGown™ surgical gowns provided to convenience kit packers.

Non-sterile Cardinal Health SmartGown™ surgical gowns will include EO sterilization parameters on labeling as follows:

  • EO Concentration: 690 mg/L -
  • Temperature: 130 +/- 10ºF -
  • Exposure Time: 150 minutes
  • Humidity: 50 +/- 5% -
  • Aeration Time: 18 hours

§ 878.4040 Surgical apparel.

(a)
Identification. Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.(b)
Classification. (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.