K Number
K150150

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-03-27

(63 days)

Product Code
Regulation Number
892.1680
Age Range
All
Reference & Predicate Devices
N/A
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

1717SGN and 1717SCN Digital Flat Panel X-Ray Detector are indicated for digital imaging solution designed for general radiographic system for human anatomy. They are intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

Device Description

1717SGN and 1717SCN digital solid state X-ray detectors are based on flat-panel technology. Both radiographic image detector and processing units consist of a scintillator coupled to an a-Si TFTsensor. A digital flat panel X-ray detector needs to be integrated with a radiographic imaging system. It can be utilized to capture and digitalize X-ray images for radiographic diagnosis. The RAW files can be further processed as DICOM compatible image files by separate console SW (not part of this 510k submission) for a radiographic diagnosis and analysis.

AI/ML Overview

The provided text describes a 510(k) premarket notification for the Rayence Co., Ltd.'s 1717SGN / 1717SCN Digital Flat Panel X-ray Detectors. The submission aims to demonstrate substantial equivalence to predicate devices (LLX240AB01 and LTX240AA01-A manufactured by Samsung Mobile Display Co., Ltd.).

This document outlines the testing conducted to support the claim of substantial equivalency but does not provide specific, quantifiable acceptance criteria in a table format with corresponding pass/fail results or detailed device performance metrics relevant to an AI/Machine Learning context. The focus is on demonstrating that the new devices perform comparably to (or better than) existing predicate devices for general radiographic use, primarily through non-clinical bench testing and a limited clinical consideration.

Therefore, many of the requested elements for an AI/ML device's acceptance criteria and study proving performance cannot be fully extracted or are not applicable from this document, as it pertains to a hardware medical device submission.

Here's an attempt to address the points based on the provided text, highlighting where information is missing or not applicable:


Acceptance Criteria and Study Proving Device Meets Acceptance Criteria

This document describes the 510(k) submission for a digital flat panel X-ray detector, not an AI/Machine Learning diagnostic algorithm. Therefore, the typical acceptance criteria and study designs for AI/ML devices (e.g., sensitivity, specificity, AUC, MRMC studies, ground truth establishment for AI training/testing sets) are not directly applicable or detailed in this submission.

The general acceptance criteria for this hardware device revolve around demonstrating substantial equivalence to predicate devices in terms of:

  • Safe and effective operation.
  • Having the same intended use.
  • Similar technological characteristics.
  • Performance that is comparable or superior to the predicate.

1. A table of acceptance criteria and the reported device performance

The document does not present acceptance criteria in a quantitative table format suitable for an AI/ML context (e.g., minimum sensitivity, specificity). Instead, performance is reported comparatively against the predicate devices.

Acceptance Criterion (Implicit)Device Performance (1717SGN / 1717SCN)
Non-clinical Performance
MTF (Modulation Transfer Function)Performed "better" than predicate devices (LLX240AB01/LTX240AA01-A).
DQE (Detective Quantum Efficiency)1717SGN: Higher DQE than LLX240AB01 at all spatial frequencies and superior Signal-to-Noise Ratio (SNR) transfer.
1717SCN: Higher DQE at high spatial frequencies (1 lp/mm to 3 lp/mm).
NPS (Noise Power Spectrum)Test results provided (details not explicitly stated as comparative data).
Clinical Performance (Qualitative Comparison)
Spatial Resolution"comparable or superior" to predicate. Specifically, "superior" for 1717SCN/1717SCN (likely a typo, refers to subject devices).
Soft Tissue Contrast"comparable or superior" to predicate. Specifically, "superior" for 1717SCN/1717SCN. Soft tissues on extremity films seen with "better clarity".
Evaluation of Anatomic StructuresNo difficulty in evaluating a wide range of anatomic structures necessary to provide a correct conclusion.
Safety & Electrical PerformancePassed IEC 60601-1:2005 + CORR.1(2006) + CORR(2007) and IEC 60601-1-2:2007 (EMC).

2. Sample size used for the test set and the data provenance

  • Test Set (Clinical Consideration): "sample radiographs of similar age groups and anatomical structures" were taken. A specific number of cases or images is not provided.
  • Data Provenance: Not explicitly stated, but given the manufacturer is Rayence Co., Ltd. (Korea) and the consulting firm is in Houston, TX, the data could be from various geographies. The clinical consideration involved a "licensed US radiologist", suggesting the review was done in the U.S. There is no indication of retrospective or prospective study design, but the description "After comparing a broad review of plain radiographic images taken with..." suggests an observational review rather than a formal prospective clinical trial.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: "a licensed US radiologist" (implies one expert).
  • Qualifications: "licensed US radiologist." No specific experience level (e.g., 10 years) is provided.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Adjudication Method: "reviewed by a licensed US radiologist to render an expert opinion." This implies no formal adjudication among multiple readers, as only one expert is mentioned. The comparison was a qualitative expert review.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • No MRMC study or AI assistance: This submission is for a digital X-ray detector, not an AI-powered diagnostic tool. Therefore, an MRMC study related to AI assistance was not applicable and not performed.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • Not applicable: This is a hardware device. No diagnostic algorithm is being evaluated in a standalone capacity.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Type of Ground Truth (for clinical consideration): The "ground truth" was the expert opinion of a single licensed US radiologist comparing images from the new devices against images from the predicate devices. This is a qualitative comparison, not a formal diagnostic ground truth like pathology.

8. The sample size for the training set

  • Not applicable: This is a hardware device; there is no mention of a "training set" in the context of an AI/ML algorithm.

9. How the ground truth for the training set was established

  • Not applicable: No training set for an AI/ML algorithm was used.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name arranged in a circular fashion around a symbol. The symbol is a stylized representation of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 27, 2015

Rayence Co., Ltd. % Mr. Dave Kim Medical Device Regulatory Affairs Mtech Group 8310 Buffalo Speedway HOUSTON TX 77025

Re: K150150

Trade/Device Name: 1717SGN / 1717SCN Digital Flat Panel X-ray Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: II Product Code: MQB Dated: January 20, 2015 Received: January 23, 2015

Dear Mr. Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150150

Device Name

1717SGN / 1717SCN Digital Flat Panel X-ray Detector

Indications for Use (Describe)

1717SGN and 1717SCN Digital Flat Panel X-Ray Detector are indicated for digital imaging solution designed for general radiographic system for human anatomy. They are intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CER 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

1. 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.

Date 510k summary prepared: March 19, 2015

Submitter's Name, address, telephone number, a contact person:

Submitter's Name :Rayence Co., Ltd.
Submitter's Address:14, Samsung 1-ro 1-gil, Hwaseong-si, Gyeonggi-do, Korea
Submitter's Telephone:+82-31-8015-6459
Contact person:Mr. Kee Dock Kim / RA Team Manager / +82-31-8015-6459
Official Correspondent:Dave Kim (davekim@mtech-inc.net)
(U.S. Designated agent)
Address:8310 Buffalo Speedway, Houston, TX 77025
Telephone:+713-467-2607
Fax:+713-583-8988

Name of the device, including the trade or proprietary name if applicable, the common or usual name and the classification name, if known:

Trade/proprietary name:1717SGN / 1717SCN
Common Name:Digital Flat Panel X-ray Detector
Classification Name :21CFR892.1680 / Stationary x-ray system
Product Code:MQB

Predicate Device :

Manufacturer: Samsung Mobile Display Co., Ltd.
Device: LLX240AB01 & LTX240AA01-A
510(k) Number: K102587 & K090742
Decision Date: DEC 1, 2010 & SEP 18, 2009

{4}------------------------------------------------

2. Device Description

1717SGN and 1717SCN digital solid state X-ray detectors are based on flat-panel technology. Both radiographic image detector and processing units consist of a scintillator coupled to an a-Si TFTsensor. A digital flat panel X-ray detector needs to be integrated with a radiographic imaging system. It can be utilized to capture and digitalize X-ray images for radiographic diagnosis. The RAW files can be further processed as DICOM compatible image files by separate console SW (not part of this 510k submission) for a radiographic diagnosis and analysis.

3. Indication for use

1717SGN and 1717SCN Digital Flat Panel X-Ray Detector is indicated for digital imaging solution designed for general radiographic system for human anatomy. It is intended to replace film or screen based radiographic systems in all general purpose diagnostic procedures. Not to be used for mammography.

4. Summarv of Design Control Risk management

1717SGN and 1717SCN digital X-ray detectors are the modification version of LLX240AB01(K102587) and LTX240AA01-A(K090742), respectively.

1717SGN and 1717SCN are slightly smaller in dimensions and substantially lighter in weight than LLX240AB01(K102587) and LTX240AA01-A(K090742).

The risks and the hazardous impact of the device modification were analyzed with FMEA method. The specific risk control and protective measures to mitigate the risks from the modification were reviewed and implemented in the new product design phase. The overall assessment concluded that all risks and hazardous conditions identified arising from the design change were successfully mitigated and accepted.

5. Summary of the technological characteristics of the device compared to the predicate device:

1717SGN and 1717SCN detector described in this 510(k) have the same indications for use and similar technical characteristics as each predicate device, LLX240AB01(K102587) and LTX240AA01-A(K090742) flat panel detector, manufacturerd by Samsung Mobile Display Co., Ltd.

{5}------------------------------------------------

5.1 Scintillator layer

1717SGN and 1717SCN detector have same Hardware, Software and components except the type of the scintillator layer.

( *scintillator : a phosphor that produces scintillations)

Scintillator Layer TypeProposedPredicate
Gd2O2S:Tb(Gadolinium Oxysulfide)1717SGNLLX240AB01
CsI(Cesium Iodide)1717SCNLTX240AA01-A

Dimensions and Weight 5.2

Proposed1717SGNPredicateLLX240AB01
1717SCNLTX240AA01-A
DetectorModel name1717SGNLLX240AB01
W x L x H460 X 460 X 15.5500 x 496.6 x 45mm
Weight4 kg14.5kg
Power supplyModel nameRS1717SPS01
W x L x H188 X 92 X 41.5290 x 245 x 68 mm
Weight0.5 kg4 kg

5.3 Power Requirements (Power supply)

ProposedPredicate
1717SGN1717SCNLLX240AB01LTX240AA01-A
PowersupplyModel nameRS1717SPS01
W x L x HAC 100 - 240 V~,50/60HzAC100-120 / 220-240 V~,50/60Hz

5.4 Proposed Labeling

ProposedPredicate
1717SGNLLX240AB01
1717SCNLTX240AA01-A

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image shows labels for Rayence and Samsung Mobile Display Co., Ltd products. The Rayence labels include models 1717SGN, 1717SCN, and RS1717, identifying medical image processing units and a power supply, with ratings around 24V and compliance certifications. The Samsung labels feature models LLX240AB01, LTX240AA01-A, and SPS01, detailing digital flat panel X-ray detectors and a power supply, with weight and dimension specifications, along with caution notices for power supply usage.

5.5 Comparison table

ProposedPredicate
Characteristic1717SGNLLX240AB01
1717SCNLTX240AA01-A

{7}------------------------------------------------

510(k) numberLLX240AB01K102587
LTX240AA01-AK090742
Intended Use1717SGN and 1717SCN DigitalFlat Panel X-Ray Detector isindicated for digital imagingsolution designed for generalradiographic system for humananatomy. It is intended to replacefilm or screen based radiographicsystems in all general purposediagnostic procedures. Not to beused for mammography.Digital Flat Panel X-Ray Detector isindicated for digital imaging solutiondesigned for general radiographicsystem for human anatomy. It isintended to replace film or screen basedradiographic systems in all generalpurpose diagnostic procedures. Not tobe used for mammography.
ManufacturerRayence Co., Ltd.Samsung Mobile Display Co., Ltd.
Detector TypeAmorphous Silicon, TFTAmorphous Silicon, TFT
Scintillator1717SGN1717SCNGd2O2S:TbCsI:TlLLX240AB01LTX240AA01-AGd2O2S:TbCsI:Tl
Imaging Area17 x 17 inches
Pixel matrix3072 x 3072 (9 million)
Pixel pitch127 µm143 µm
Resolution3.9 lp/mm3.5 lp/mm
A/D conversion14 / 16 bit14 bit
Preview time≤25 seconds per Image
Data outputRAW*The RAW files are convertible intoDICOM 3.0 by console S/WRAW*The RAW files are convertible into DICOM3.0 by console S/W
Dimensions460 × 460 × 15.5 mm500 x 496.6 x 45 mm
Weight4 kg14.5 kg
ApplicationGeneral Radiology system or PortablesystemAvailable with upright stand, table,universal stand.General Radiology system or Portable systemAvailable with upright stand, table, universalstand.

6.System requirements to operate with other radiographic system components

    1. Recommended Generator Specification:
      Energy range: 40~150kVp
  • mA range 10~1000mA(depending on the generator power)
    [NOTE] To our best knowledge, the detector is compatible with the X-ray generators with the specifications described above. If you have questions regarding the compatibility issue for other generators, please contact your Rayence representative.

    1. Software Development Kit (Library's Software Interface) for third party viewing software Peripheral hardware: DaVinci detector connected via Ethernet card.

{8}------------------------------------------------

CPU/Motherboard: 1 GHz (or faster) CPU, cache L2≥256KB RAM: ≥512MB, 2GB recommended Hard drive: no special requirements Display: At least 17-inch monitor is required. 21-inch monitor is suggested. OS: Windows 2K or XP professional Development environment: MS Visual C++ 6.0 or MS Visual Studio 2005

7. Summary of Performance Testing

1717SGN and 1717SCN Digital Flat Panel X-Ray Detector have the same indications for use, material, form factor, performance, and safety characteristics compared to the predicate devices, LLX240AB01 and LTX240AA01-A, respectively.

The non-clinical test report and clinical consideration report for each subject device were prepared and submitted to FDA separately to demonstrate the substantial equivalency of the subject devices compared to each respective predicate device. The non-clinical test report contains the MTF, DQE and NPS test results of 1717SGN and 1717SCN by using the identical test equipment and same analysis method described by IEC 6220-1. The comparative result of the MTF test for 1717SGN and 1717SCN detector with respective predicate demonstrated that the MTF of the both 1717SGN and 1717SCN performed better compared with the predicate devices. The DQE represents the ability to visualize object details of a certain size and contrast. 1717SGN demonstrated higher DQE performance than LLX240AB01 at all spatial frequencies as well as superior Signal-to Noise Ratio (SNR) transfer from the input to the output of the detector as a function of frequency. 1717SCN has higher DQE performance at high spartial frequencies, especially from 11p/mm to 3 lp/mm.

To further demonstrate the substantial equivalency of two devices, clinical images are taken from both subject devices and reviewed by a licensed US radiologist to render an expert opinion. Both the test subject (1717SGN and 1717SCN) and control group (LLX240AB01 and LTX240AA01-A) are evaluated and compared by taking sample radiographs of similar age groups and anatomical structures in accordance with the test protocol of diagnostic radiography evaluation procedure.

After comparing a broad review of plain radiographic images taken with 1717SGN / 1717SCN and LLX240AB01 / LTX240AA01-A, the images obtained with the 1717SGN / 1717SCN are comparable or superior to the same view obtained from a similar patient with the LLX240AB01 / LTX240AA01-A. In general, both the spatial resolution and soft tissue contrast are superior

{9}------------------------------------------------

using the 1717SCN / 1717SCN digital flat panel detector. Specifically, the soft tissues on extremity films were seen with better clarity. There is no difficulty in evaluating a wide range of anatomic structures necessary to provide a correct conclusion.

Clinical images were provided; these images were not necessary to establish substantial equivalence based on the modifications to the device but they provide further evidence in addition to the laboratory performance data to show that the subject devices operate as indicated.

Based on the non-clinical bench testing supplemented by the clinical consideration and the outcome of a comparative review by an expert for both devices, the sponsor demonstrated that the subject devices operate as indicated and are substantially equivalent to the predicate devices in terms of safety and effectiveness.

The manufacturing facility is in conformance with the design control procedure requirements and the relevant EPRC standards as specified in 21 CFR 802.30 and the records are available for review.

8. Summary for any testing in the submission:

Electrical, mechanical, environmental safety and performance testing according to standard IEC 60601-1: 2005 + CORR.1(2006) + CORR(2007) (Medical electrical equipment Part 1:General requirements for basic safety and essential performance) was performed, and EMC testing were conducted in accordance with standard IEC 60601-1-2:2007.

Non-clinical & Clinical considerations according to FDA Guidance for the Submissions of 510(k)'s for Solid State X-ray Imaging Devices" was performed. All test results were satisfactory.

All test results were satisfactory

9. Conclusions:

In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807 and based on the information provided in this premarket notification Rayence Co., Ltd. concludes that 1717SGN / 1717SCN are safe and effective and substantially equivalent in comparison with LLX240AB01 / LTX240AA01-A, the predicate devices as described herein.

N/A