K Number
K150076

Validate with FDA (Live)

Date Cleared
2015-06-02

(139 days)

Product Code
Regulation Number
872.4120
Panel
Dental
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are piezoelectric devices for surgery that enable mechanical ultrasound treatment, osteotomy and osteoplasty techniques to be applied in dental use.

Device Description

Sonic Surgeon 310L. Sonic Surgeon 600L and Sonic Surgeon 800L are ultrasonic surgical units intended for mechanical ultrasound treatment in prophylaxis, periodontics. The normal mode is tooth scaling applications, and the Boost Mode of Sonic Surgeon 310L. Sonic Surgeon 600L and Sonic Surgeon 800L can be used for surgical procedures, including osteotomy, osteoplasty, periodontal surgery and implantation.

The tips can easily be changed during the treatment and must also be cleaned and autoclaved.

AI/ML Overview

This document is a 510(k) Summary for the Sonic Surgeon 310L, 600L, and 800L devices, seeking to demonstrate substantial equivalence to predicate devices, not an AI/ML device. Therefore, it does not contain the information requested in the prompt, such as acceptance criteria for AI performance, details of a study proving a device meets such criteria, sample sizes for test/training sets in an AI context, expert ground truth establishment, adjudication methods, or MRMC studies.

The document focuses on the technological characteristics and performance parameters comparing the new devices to existing predicate devices (Sonic Surgeon 300 and UBS ultrasonic bone surgery/UDD ultrasonic debridement device) to assert substantial equivalence.

Here's an analysis of the available information in relation to the prompt:

  1. Table of acceptance criteria and the reported device performance:
    The document does not present "acceptance criteria" in the sense of predefined thresholds for performance to be met by the device itself beyond functional equivalence to predicates. Instead, it compares specific technological characteristics and performance parameters to predicate devices. These comparisons serve as the basis for claiming substantial equivalence.

    Characteristic/ParameterPredicate Device (Sonic Surgeon 300 - K110881)Predicate Device (UBS/UDD - K080220)Proposed Devices (310L, 600L, 800L)
    Intended UseUltrasonic Surgical Unit for osteotomy & osteoplasty in dental useBone cutting instruments for oral surgeryPiezoelectric devices for mechanical ultrasound treatment, osteotomy, and osteoplasty in dental use
    Sterilization MethodSame as Proposed Devices (implied, not explicitly detailed for predicate)Same as Proposed Devices (implied, not explicitly detailed for predicate)Autoclaved (for tips)
    Materials UsedEnclosure: PC, ABS; Hose/Seals: Silicone; Hand Piece: PES; Tips: TrimRite stainless; Tip Coating: ZrN, ZrN+DiamondNot detailedEnclosure: PC, ABS; Hose/Seals: Silicone; Hand Piece: PPSU; Tips: TrimRite stainless; Tip Coating: ZrN, ZrN+Diamond
    Electrical Input PowerAC 95-115V, AC 210-240VAC 230V±10%AC 100-240V
    Auto Scanning Freq.26 ± 3 kHz26 kHz ± 10%Automatic scanning, 26 ± 3 kHz
    Power Output (Normal Mode)(Not specified for 300)(Not specified for UBS/UDD Normal)310L: 15-17W (L1), 17-21W (L2), 21-27W (L3)600L: 15-17W (L1), 17-21W (L2), 21-27W (L3)800L: 15-17W (L1), 17-21W (L2), 21-27W (L3)
    Power Output (Boost Mode)(Not specified for 300)90 W (overall power mentioned)310L: 30-50W600L: 50-70W800L: 70-90W
    Water Flow (approx. ml/min)Level 1: 10; Level 2: 30; Level 3: 50; Level 4: 70; Level 5: 90Not detailedLevel 1: 7-13; Level 2: 15-40; Level 3: 40-60; Level 4: 60-80; Level 5: 80-110
    LED Brightness (lux, at 20mm)XNot detailed310L: 1900-2600600L: 2000-2800800L: 2000-2800 (across 3 levels)
    Foot Switch IP ClassIPX1Not detailedIPX8

    The "acceptance criteria" here are implicitly that the proposed devices must demonstrate similar performance and safety characteristics to the legally marketed predicate devices, supporting the claim of substantial equivalence. For example, the maximum output power of the Sonic Surgeon 800L (70-90W) is compared to the predicate UBS device (90W) to establish equivalence in this aspect.

  2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective):
    This document describes a medical device, not an AI/ML product. The "test set" in this context refers to standard engineering and electrical safety tests, and potentially material biocompatibility tests. It does not involve a "test set" of clinical data for algorithmic performance evaluation. No information regarding sample size or data provenance in the context of clinical studies for performance metrics is provided. The tests mentioned are for compliance with IEC standards.

  3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience):
    Not applicable. Ground truth and expert adjudication are concepts relevant to AI/ML performance evaluation, which is not the subject of this 510(k) submission. The "ground truth" for this device would be established by physical measurements and compliance with engineering and safety standards.

  4. Adjudication method (e.g. 2+1, 3+1, none) for the test set:
    Not applicable. This is not an AI/ML filing.

  5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:
    Not applicable. This is not an AI/ML filing; no human-in-the-loop performance or reader studies are mentioned.

  6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:
    Not applicable. This is not an AI/ML filing.

  7. The type of ground truth used (expert consensus, pathology, outcomes data, etc):
    Not applicable in the context of AI/ML. The "ground truth" for this device's performance would be derived from physical measurements (e.g., power output, frequency, LED brightness, water flow) and compliance with established engineering and safety standards (IEC 60601-1, IEC 60601-1-2).

  8. The sample size for the training set:
    Not applicable. This is not an AI/ML filing. There is no concept of a "training set" for this type of device.

  9. How the ground truth for the training set was established:
    Not applicable. This is not an AI/ML filing.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus, which is a symbol often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 2, 2015

Dong Il Technology Ltd. c/o Mr. Peter Chung Plus Global 300 Atwood Street Pittsburgh, Pennsylvania 15213

Re: K150076

Trade/Device Name: Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L Regulation Number: 21 CFR 872.4120 Regulation Name: Bone Cutting Instrument and Accessories Regulatory Class: II Product Code: DZI Dated: April 29, 2015 Received: April 30, 2015

Dear Mr. Chung:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150076

Device Name

Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L

Indications for Use (Describe)

Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are piezoelectric devices for surgery that enable mechanical ultrasound treatment, osteotomy and osteoplasty techniques to be applied in dental use.

Type of Use (Select one or both, as applicable)X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K150076

510(k) Summary[as required by 807.92(c)]

A. Submission Information

    1. Submission: 510(k) Submission
    1. Submission Date: June 1, 2015
    1. Submission Type: Traditional
    1. 510(k) number: K150076

B. Applicant:

    1. Company name: DONG IL TECHNOLOGY LTD.
    1. Address: 28, Namyang-ro, 930(gubaeksamsip)beon-gil, Bugyang-dong, Hwaseong-si, Gyeonggi-do, Republic of Korea
    1. Tel: +82(31)229-5500
    1. Fax: +82(31)357-2610
    1. Homepage: http://www.dit-med.co.kr
    1. Contact person: Peter Chung 412-687-3976
    1. Contact person address: 300, Atwood ST, Pittsburgh, PA, 15213

C. Proprietary and Established Names:

    1. Trade Name: Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L
    1. Common Name: Bone cutting instrument and accessories
    1. Regulation Name: Sonic surgical instrument and accessories/attachments
    1. Regulation Number: 872.4120
    1. Classification: Class 2
    1. Panel: Dental
    1. Product Code: DZI
    1. Identify contact classification:

Category: External communicating device Contact: Tissue/bone/dentin Contact duration: Limited exposure (A)

D. legally marketed predicate devices:

Primary: Sonic Surgeon 300 (K110881) Reference: UBS ultrasonic bone surgery and UDD ultrasonic debridement device (K080220)

E. Device Description:

Sonic Surgeon 310L. Sonic Surgeon 600L and Sonic Surgeon 800L are ultrasonic surgical units intended for mechanical ultrasound treatment in prophylaxis, periodontics. The normal mode is tooth scaling applications, and the Boost Mode of Sonic Surgeon 310L. Sonic Surgeon 600L and Sonic Surgeon 800L can be used for surgical procedures, including osteotomy, osteoplasty, periodontal surgery and implantation.

The tips can easily be changed during the treatment and must also be cleaned and autoclaved. Materials for Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are as follows:

ComponentMaterial
Enclosure (Generator Unit, FootSwitch)PC(Polycarbonate)
Hose, and sealsABS(Acrylonitrile-Butadiene-Styrene Terpolymer)
Enclosure (Handpiece)Silicone
TipPES(Polyethersulfone)
PPSU(Polyphenylsulfone)

{4}------------------------------------------------

F. Intended use

Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are piezoelectric devices for surgery that enable mechanical ultrasound treatment, osteotomy and osteoplasty techniques to be applied to use.

G. Technological Characteristics:

Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L have the following features:

  • . Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L transform generated ultrasonic (26±3kHz) energy to the kinetic energy and transmits it to the tip.
  • . The generator of the Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L performs automatic tuning of the operating frequency, the efficiency of the piezoelectric transducer in the handpiece. This feature makes it possible to perform bone cutting, grinding and drilling action.

Our device is very similar to predicate devices, Sonic Surgeon 300 (K110881), UBS ultrasonic bone surgery and UDD ultrasonic debridement device(K080220), because Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L have the following equivalent characteristics: intended use, sterilization method, material used, electrical input power, frequency, and power output as predicate devices.

ManufacturerModelInsulationClassInput VoltageInput PowerCurrent orPowerFrequencyUltrasonicOperationFrequency
K150076
DONG ILTECHNOLOGY LTDSonic Surgeon310L, 600L, 800LClass I,Type BAC 100-240 VAC Max 2A50 / 60 HzAutomaticscanning,$26 \pm 3$ KHz
K110881
DONG ILTECHNOLOGY LTDSonic Surgeon300Class I,Type BAC 95-115 VAC 210-240 VAC Max 1A50 / 60 HzAutomaticscanning,$26 \pm 3$ KHz
K080220
ITALIAMEDICAS.R.LUBS ultrasonicbone surgery andUDD ultrasonicdebridementdeviceClass I,Type CFAC 230V±10%140 VA50 / 60 Hz$26$ kHz $\pm$10%

Summary Comparison table

ManufacturerModelIndications for Use
K150076
DONG ILTECHNOLOGYLTDSonic Surgeon 310LSonic Surgeon 600LSonic Surgeon 800LSonic Surgeon 310L, Sonic Surgeon 600L, Sonic Surgeon 800L are piezoelectric devices for surgery that enable mechanical ultrasound treatment, osteotomy and osteoplasty techniques to be applied in dental use.
K110881
DONG ILTECHNOLOGYLTDSonic Surgeon 300Ultrasonic Surgical Unit is a piezoelectric device for bone surgery that enables osteotomy and osteoplasty techniques to be applied to in dental use.
K080220
ITALIA MEDICAS.R.LUBS ultrasonic bone surgery and UDD ultrasonicdebridementThe UBS Ultrasonic Bone Surgery and the UDD Ultrasonic Debridement Device are bone cutting instruments intended for use in oral surgery.

{5}------------------------------------------------

(1) Comparison between proposed Device and predicate Sonic Surgeon 300(K110881)

Predicate DeviceProposed Deviceremark
Model NameSonic Surgeon 300Sonic Surgeon 310LSonic Surgeon 600LSonic Surgeon 800L
Features
LED LightXOOO
LCDLCD DisplayTouch LCDTouch LCDTouch LCD
Input VoltageAC 95 - 115V(50/60Hz)AC 210 – 240 V(50/60Hz)AC100V -AC240V(50/60Hz)AC100V - AC240V(50/60Hz)AC100V - AC240V(50/60Hz)
Pump CoverAutomaticManualManualManual
Power SupplyToroidal TransformerSMPSSMPSSMPS
Foot Switch IP classIPX1IPX8IPX8IPX8
Operating Frequency26 ± 3 kHz26 ± 3 kHz26 ± 3 kHz26 ± 3 kHz
Other features are same as Predicate.
Dimensions
Hand pieceA type:$Ø$ 20×154.2mm(L)A type:$Ø$ 20×154.2mm(L)B type(LED):$Ø$ 20×152.7mm(L)C type(LED):$Ø$ 22.5×155.4mm(L)C type(LED):$Ø$ 22.5×
GeneratorH:106mmL:300mmW:284mmH:106mmL:300mmW:298mmH:106mmL:300mmW:298mmH:106mmL:300m
Foot SwitchH: 37 mmL: 140 mmW: 240 mmH: 37 mmL: 140 mmW: 240 mmH: 37 mmL: 140 mmW: 240 mmH: 37 mmL: 140 mmW: 240 mm
Tip HolderH: 37.9 mmL: 41 mmW: 58 mmH: 38 mmL: 41 mmW: 58 mmH: 38 mmL: 41 mmW: 58 mmH: 38 mmL: 41 mmW: 58 mm
Torque Wrench$Ø$ 40×31mm(L)$Ø$ 40×31mm(L)$Ø$ 40×31mm(L)$Ø$ 40×31mm(L)
Wire Water Pack$Ø$ 5×477mm(L)$Ø$ 5×470mm(L)$Ø$ 5×470mm(L)$Ø$ 5×470mm(L)
Storage tray-not for sterilizer useH: 61 mmL: 180 mmW: 200 mmH: 61 mmL: 180 mmW: 200 mmH: 61 mmL: 180 mmW: 200 mmH: 61 mmL: 180 mmW: 200 mm
Wire Hand pieceH: 128 mmL: 66 mmW: 83 mmH: 128 mmL: 66 mmW: 83 mmH: 128 mmL: 66 mmW: 83 mmH: 128 mmL: 66 mmW: 83 mm
TipsRefer to Table 4Table.1
Refer to Table 5 & Table 6The material of handpiece enclosure is different from Sonic Surgeon 300.Other materials are same as predicate device.
Materials
Performance parameters
LED Light ControlXOOOTable.2
Output PowerMAX 40W15 ~ 50W15 ~ 70W15 ~ 90WTable.3
Irrigation10~90 ml/min.7~110 ml/min.7~110 ml/min.7~110 ml/min.Table.4

{6}------------------------------------------------

(2) Comparing 'output Power' between proposed Device and predicate model 'UBS ultrasonic bone surgery and UDD ultrasonic debridement device (K080220)'

Predicate DeviceProposed Deviceremark
ManufacturerITALIA MEDICAS.R.LDONG IL TECHNOLOGY LTD.
Model NameModel UltrasonicDebridementSonic Surgeon310LSonic Surgeon600LSonic Surgeon800L
Performance parameters
OutputPower90 W30~50 W50~70 W70~90 WTable.3

Table.1 Tips (Sonic Surgeon 300/310L/600L/800L)

No.Sonic Surgeon 300 / 310LSonic Surgeon 600L / 800L
NameSizeNameSize
1DG3-001Ø 4.5x28.26mm(L)BTG-B017Ø 5.5x28.5mm(L)
2DG3-002Ø 4.5×30.51mm(L)BTG-R018Ø 5.5x29.5mm(L)
3DM3-003Ø 4.5x31.86mm(L)BTM-D019Ø 5.5x31.5mm(L)
4DE3-004Ø 4.5×28.29mm(L)BTE-K020Ø 5.5x29.0mm(L)
5DH6-005Ø 4.5x30.75mm(L)BTH-S021Ø 5.5x30.1mm(L)
6DD6-016Ø 4.5x27.7mm(L)BTD-F022Ø 5.5x27.5mm(L)
7DD6-028iØ 4.5x28.2mm(L)BTD-F023Ø 5.5x28.5mm(L)
8DC6-SawØ 4.5x31.5mm(L)BTC-I024Ø 5.5x31.5mm(L)
9DC6-Saw LØ 4.5x30.9mm(L)BTC-I025Ø 5.5×31.0mm(L)
10DC6-Saw RØ 4.5x30.9mm(L)BTC-I026Ø 5.5x31.0mm(L)

Table.2 LED Brightness

LEDlevelLED brightness (lux)Remark
Sonic Surgeon 310LSonic Surgeon 600LSonic Surgeon 800L
1$1,900\pm100$$2,000\pm100$$2,000\pm100$
2$2,400\pm100$$2,500\pm100$$2,500\pm100$Distance:20mm
3$2,600\pm100$$2,800\pm100$$2,800\pm100$

{7}------------------------------------------------

Table.3 Output Power

[Unit: W]
Model NameNormal ModeBoost Mode
Level1Level2Level3Level1Level2Level3
Sonic Surgeon 310L15~1717~2121~2730~5030~5030~50
Sonic Surgeon 600L15~1717~2121~2750~7050~7050~70
Sonic Surgeon 800L15~1717~2121~2770~9070~9070~90

Table.4 Water flow

Pump levelWater flow (ml/min approximately)Remark
Sonic Surgeon 300Sonic Surgeon 310L/600L/800L
1107 ~ 13
23015 ~ 40
35040 ~ 60
47060 ~ 80
59080 ~ 110

Table.5 Materials (Sonic Surgeon 300)

NoNameMaterialsCompanyCAS No.Grade
1Enclosure(GeneratorUnit)PC(Polycarbonate)ABS(Acrylonitrile-Butadiene-StyreneTerpolymer)LG ChemicalPC(103598-77-2)ABS(9003-56-9)LUPOYGN5001RF
2Hose, andsealsSiliconeElkem7440-21-3Syntheticrubber
3Enclosure(Hand Piece)PES(Polyethersulfone)BASF25608-63-3UltrasonE2010
4TipsTrimRite stainlessCARPENTER-UNS S42010
5Tip CoatingZrN(Zirconium Nitride)ZrN+DiamondChemetall(Zirconium)ProSciTech25658-42-8Diamond(77820-40-3)CAM24

{8}------------------------------------------------

Table.6 Materials (Sonic Surgeon 310L/600L/800L)
NoNameMaterialsCompanyCAS No.Grade
1Enclosure(GeneratorUnit)PC(Polycarbonate)ABS(Acrylonitrile-Butadiene-StyreneTerpolymer)LG ChemicalPC(103598-77-2)ABS(9003-56-9)LUPOYGN5001RF
2Hose, andsealsSiliconeElkem7440-21-3Syntheticrubber
3Enclosure(Hand Piece)PPSU(Polyphenylsulfone)SolvayPolyphenylsulfone(25608-64-4)Titanium dioxide(13463-67-7)Carbon black(1333-86-4)DuradexD-3000
4TipsTrimRite stainlessCARPENTER-UNS S42010
5Tip CoatingZrN(Zirconium Nitride)ZrN+DiamondChemetall(Zirconium)ProSciTech25658-42-8Diamond(77820-40-3)CAM24

H. Performance Information

The Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L have been manufactured and tested to meet the safety requirements of several IEC standards. The Sonic Surgeon 310L. Sonic Surgeon 600L and Sonic Surgeon 800L comply with IEC 60601-1 Medical electrical equipment - Part 1: General requirements for safety and IEC 60601-1-2:2007, Medical electrical equipment - Part 1: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests.

l. Conclusion:

The performance tests and evidence demonstrate that Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L function in a substantially equivalent manner to the predicate device.

In terms of output power, Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are substantially equivalent to the predicate device UBS ultrasonic bone surgery and UDD ultrasonic debridement device (K080220). The maximum output power of Sonic Surgeon 800L is 70~90W and that of UBS ultrasonic bone surgery and UDD ultrasonic debridement device (K080220)' is 90W. In terms of light power, the capacity of the device's LED light is demonstrated in 'Biocompatibility' section and an LED report was attached as evidence.

Because our proposed device has the same intended use, the same sterilization method, same materials used, same electrical input power, same frequency, and same power output as predicate devices, as well as similarity of product configuration and administration, it can be concluded the Sonic Surgeon 310L, Sonic Surgeon 600L and Sonic Surgeon 800L are substantially equivalent to the identified predicate devices.

§ 872.4120 Bone cutting instrument and accessories.

(a)
Identification. A bone cutting instrument and accessories is a metal device intended for use in reconstructive oral surgery to drill or cut into the upper or lower jaw and may be used to prepare bone to insert a wire, pin, or screw. The device includes the manual bone drill and wire driver, powered bone drill, rotary bone cutting handpiece, and AC-powered bone saw.(b)
Classification. Class II.