K Number
K150063

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-10-02

(262 days)

Product Code
Regulation Number
862.1155
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The ACCU NEWS One Step Pregnancy Test Strip: The ACCU NEWS One Step Pregnancy Test Strip is a lateral flow immunoassay to qualitatively detect the presence of human Chorionic (hCG) hormone in urine specimen to aid early detection of pregnancy for both prescription use and OTC use.

ACCU NEWS One Step Pregnancy Test Cassette: The ACCU NEWS One Step Pregnancy Test Cassette is a lateral flow immunoassay to qualitatively detect the presence of human Chorionic (hCG) hormone in urine specimen to aid early detection of pregnancy for both prescription use and OTC use.

ACCU NEWS One Step Pregnancy The ACCU NEWS One Step Pregnancy Test Midstream is a lateral flow immunoassay to qualitatively detect the presence of human Chorionic Gonadotropin (hCG) hormone in urine specimen to aid early detection of pregnancy for both prescription use and OTC use.

Device Description

ACCU NEWS "M One Step hCG Pregnancy Test is a qualitative lateral flow test for the detection of hCG (human chorionic gonadotropin) hormone, which is present in urine during the early stage of pregnancy. It is built on a basic lateral flow immunoassay test strip model containing 6 components as illustrated below:

The nitrocellulose membrane is coated with Goat anti-mouse IgG polyclonal antibody on the control line and anti-alpha hCG mouse monoclonal antibody on the test line. The conjugate pad contains colloidal-gold-labeled anti-beta mouse monoclonal hCG antibody. When urine specimen is drawn through the sample pad, the hCG present in the urine will react with the anti-beta monoclonal hCG antibody on the conjugate pad and the antialpha monoclonal hCG antibody on the test line to develop visible lines in the test region.

The ACCU NEWS TM One Step hCG Pregnancy Test is available in three different formats: 1) strip, 2) cassette, and 3) midstream.

AI/ML Overview

The provided text describes the acceptance criteria and a study demonstrating the performance of the ACCU NEWS™ One Step hCG Pregnancy Test.

1. Table of Acceptance Criteria and Reported Device Performance:

Acceptance Criteria / Performance MetricPredicate Device Performance (K043443 Wondfo One Step HCG Urine Pregnancy Test)ACCU NEWS™ One Step hCG Pregnancy Test Performance
Detection Limit25 mIU/mL25 mIU/mL
Read Time Window3-5 minutes5 minutes (optimal)
High Dosage Hook EffectNo high dosage hook effect up to 100,000 mIU/mLNo high dosage hook effect up to 500,000 mIU/mL
Specificity (LH)LH at 300 mIU/mLLH at 1000 mIU/mL
Specificity (FSH)FSH at 300 mIU/mLFSH at 1000 mIU/mL
Specificity (TSH)TSH at 1000 μIU/mLTSH at 1000 μIU/mL
pH InterferenceNo interference for urine with pH 4-9No interference for urine with pH 2-9
Specific Gravity InterferenceNo interference for urine with Specific Gravity 1.000-1.050No interference for urine with Specific Gravity 1.003-1.030
Comparison Study AgreementNot explicitly stated as an acceptance criterion for the predicate, but implied 100% agreement between subject and predicate device tests100% agreement with predicate device results
OTC Lay-User AgreementNot applicable (not for predicate comparison)100% agreement between lay user and professional results
Shelf-life StabilityNot explicitly stated2 years at 4-30°C (accelerated testing at 60°C)

2. Sample Size Used for the Test Set and Data Provenance:

  • Precision/Sensitivity Study:

    • Test Set Sample Size: For each of the three formats (Strip, Cassette, Midstream), and for each of 6 hCG concentrations (0, 12.5, 18.75, 25, 50, 100 mIU/mL), at least 30 urine samples were tested. This means a minimum of 3 (formats) * 6 (concentrations) * 30 (samples) = 540 samples were used in total across the three lots for each format.
    • Data Provenance: Urine samples were used. The study was carried out at three sites (one internal site and two external sites), implying a mix of internal and potentially external, real-world data points. The text doesn't specify the country of origin of the data, but the company is based in Carlsbad, CA, USA. The study appears to be prospective in nature, as specific experiments were designed and performed to evaluate the device.
  • Comparison Study:

    • Test Set Sample Size: 250 female donors. Urine samples were collected from these donors.
    • Data Provenance: Urine samples were randomly collected from 250 female donors from three sites at various times throughout the day. The text doesn't specify the country of origin. This appears to be a prospective study.
  • OTC Lay-User Test Result:

    • Test Set Sample Size: 250 female donors.
    • Data Provenance: Urine samples were randomly collected from 250 female donors at three sites. This appears to be a prospective study.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts:

  • Precision/Sensitivity Study: The "ground truth" for the hCG concentrations (0, 12.5, 18.75, 25, 50, 100 mIU/mL) was established by external calibration against the hCG standard WHO 3rd IS. This implies laboratory standards rather than clinical expert consensus for ground truth on concentration. The results were then read by individuals at the testing sites, but their specific qualifications as "experts" for establishing ground truth are not detailed beyond "each site tested a unique lot of test device."

  • Comparison Study & OTC Lay-User Test Result:

    • For the comparison study and the OTC lay-user study, the "ground truth" for the pregnancy status of the 250 urine samples was primarily established by the predicate hCG test. Healthcare professionals also tested the urine samples with the ACCU NEWS™ devices, and their results served as a professional ground truth for the lay-user comparison. The qualifications of these "healthcare professionals" are not specified (e.g., radiologist, lab technician, nurse), nor is the exact number of such professionals.

4. Adjudication Method for the Test Set:

  • No explicit adjudication method (like 2+1 or 3+1) is described for resolving discrepancies in any of the studies.
  • In the Precision/Sensitivity study, the results are presented as counts for '+' and '-' readings for each lot and concentration, and there's no mention of adjudication if readings differed.
  • In the Comparison Study and OTC Lay-User Test Result studies, 100% agreement was observed between the subject device and the predicate device/healthcare professional readings, so no adjudication would have been needed to reconcile conflicting results.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC comparative effectiveness study was done using AI. This device is a lateral flow immunoassay (a rapid diagnostic test), not an AI-powered diagnostic tool requiring human interpretation with or without AI assistance. The "readers" are either the device itself (for objective lines) or human users interpreting the lines.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • This device is not an algorithm. It is a physical diagnostic test (lateral flow immunoassay). The "standalone" performance is essentially what is measured in the Precision/Sensitivity study, where the device's ability to detect various hCG concentrations is assessed directly.

7. The Type of Ground Truth Used:

  • Precision/Sensitivity Study: The ground truth was known hCG concentrations in urine samples, calibrated against the WHO 3rd International Standard. This can be considered a form of a reference standard or analytical ground truth.
  • Comparison Study: The ground truth was the results from a legally marketed predicate hCG test device. This is a comparative ground truth established against an existing, approved method.
  • OTC Lay-User Test Result: The ground truth for lay-user performance was the results obtained by healthcare professionals using the ACCU NEWS™ device. This is a form of expert consensus/reference reading as ground truth for evaluating ease of use and interpretation by non-professionals.

8. The Sample Size for the Training Set:

  • The document does not describe a "training set" in the context of machine learning or AI models. This device is a chemical immunoassay, not an AI-driven system. The development of the device would have involved internal R&D and optimization, but not formal "training sets" in the AI sense.

9. How the Ground Truth for the Training Set was Established:

  • As there is no "training set" in the context of an AI/ML model for this device, this question is not applicable. The development of the device's components and design would have been based on established clinical chemistry principles and validated through iterative testing and quality control processes.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 2, 2015

CORETESTS, INC. WEN LI REGULATORY AFFAIRS SPECIALIST 6190 YARROW DRIVE CARLSBAD CA 92011

Re: K150063

Trade/Device Name: ACCU NEWS One Step hCG Pregnancy Test Strip ACCU NEWS One Step hCG Pregnancy Test Cassette ACCU NEWS One Step hCG Pregnancy Test Midstream Regulation Number: 21 CFR 862.1155 Regulation Name: Human chorionic gonadotropin (HCG) test system Regulatory Class: II Product Code: LCX Dated: August 14, 2015 Received: August 17, 2015

Dear Wen Li:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Katherine Serrano -S

For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K150063

Device Name

ACCU NEWS One Step hCG Pregnancy Test Strip ACCU NEWS One Step hCG Pregnancy Test Cassette ACCU NEWS One Step hCG Pregnancy Test Midstream

Indications for Use (Describe)

ACCU NEWS One Step Pregnancy Test Strip: The ACCU NEWS One Step Pregnancy Test Strip is a lateral flow immunoassay to qualitatively detect the presence of human Chorionic (hCG) hormone in urine specimen to aid early detection of pregnancy for both prescription use and OTC use.

ACCU NEWS One Step Pregnancy Test Cassette: The ACCU NEWS One Step Pregnancy Test Cassette is a lateral flow immunoassay to qualitatively detect the presence of human Chorionic (hCG) hormone in urine specimen to aid early detection of pregnancy for both prescription use and OTC use.

ACCU NEWS One Step Pregnancy The ACCU NEWS One Step Pregnancy Test Midstream is a lateral flow immunoassay to qualitatively detect the presence of human Chorionic Gonadotropin (hCG) hormone in urine specimen to aid early detection of pregnancy for both prescription use and OTC use.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

|× Prescription Use (Part 21 CFR 801 Subpart D)

|X Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510k Summary

A. Submitter Information

Coretests Inc. 6190 Yarrow Drive, Carlsbad, CA 92011, USA. Phone Number: 858-333-1122 Contact Person: Wen Li Summary Prepared on September 29th, 2015

B. Trade Name, common name, classification name

Trade Names: ACCU NEWSTM One Step hCG Pregnancy Test Strip, ACCU NEWS™ One Step hCG Pregnancy Test Cassette, and ACCU NEWS™ One Step hCG Pregnancy Test Midstream. Common Name: Qualitative Lateral Flow Immunoassay Classification: Class II Regulation Number: 21 CFR 862.1155 Human chorionic gonadotropin test system Product code: LCX Panel: Clinical Chemistry (75)

C. Predicate device

K043443 Wondfo One Step HCG Urine Pregnancy Test

D. Device description

ACCU NEWS "M One Step hCG Pregnancy Test is a qualitative lateral flow test for the detection of hCG (human chorionic gonadotropin) hormone, which is present in urine during the early stage of pregnancy. It is built on a basic lateral flow immunoassay test strip model containing 6 components as illustrated below:

Image /page/3/Figure/9 description: The image shows a diagram of a lateral flow assay test strip. The test strip consists of several components, including a sample pad labeled as '1', a conjugate pad labeled as '2', and a nitrocellulose membrane labeled as '3'. The test strip also has a test line labeled as '4(a)', a control line labeled as '4(b)', an absorbent pad labeled as '5', and a plastic adhesive backing card labeled as '6'.

The nitrocellulose membrane is coated with Goat anti-mouse IgG polyclonal antibody on the control line and anti-alpha hCG mouse monoclonal antibody on the test line. The conjugate pad contains colloidal-gold-labeled anti-beta mouse monoclonal hCG antibody. When urine specimen is drawn through the sample pad, the hCG present in the urine will react with the anti-beta monoclonal hCG antibody on the conjugate pad and the antialpha monoclonal hCG antibody on the test line to develop visible lines in the test region.

{4}------------------------------------------------

Test Result Interpretation:

Not Pregnant: No visible pink line appearing in the test region.

Pregnant: A visible pink line appearing in the test region.

Invalid: No visible line in control region, regardless if there is a visible line in the test region or not.

The ACCU NEWS TM One Step hCG Pregnancy Test is available in three different formats: 1) strip, 2) cassette, and 3) midstream.

Strip Format

The Strip format is a test strip used by itself without any housing components. Urine specimen is collected in a cup and the sample pad is dipped into the urine directly.

Cassette Format

The Cassette format includes a test strip housed in a plastic cassette. Urine specimen is collected in a cup and a dropper is provided to transfer the urine specimen to the sample well of the cassette, wetting the sample pad to activate the test.

Midstream Format

The Midstream format includes a test strip housed in a plastic case with an absorbent tip contacting the sample pad of the test strip. When the absorbent tip is placed in a urine steam it draws urine specimen onto the sample pad of the test strip to activate the test.

All three formats are built on the basic test strip model containing 6 components as illustrated above. The strips used in the three different formats contains the same components except that the plastic adhesive backing card, absorbent pad and sample pad are in different sizes. This is specified in the table below:

ComponentStripCassetteMidstream
Plastic Adhesive Backing Card300 x 80 mm300 x 60 mm300 x 85 mm
Absorbent pad298×33 mm298x23 mm298x31mm
Sample Pad300 x 24 mm300 x 20 mm300 x 40 mm

E. Intended use

ACCU NEWSTM One Step Pregnancy Test Strip: The ACCU NEWS™ One Step Pregnancy Test Strip is a lateral flow immunoassay to qualitatively detect the presence of human Chorionic Gonadotropin (hCG) hormone in urine specimen to aid early detection of pregnancy for both prescription use and OTC use.

ACCU NEWS™ One Step Pregnancy Test Cassette: The ACCU NEWS™ One Step Pregnancy Test Cassette is a lateral flow immunoassay to qualitatively detect the presence of human Chorionic Gonadotropin (hCG) hormone in urine specimen to aid early detection of pregnancy for both prescription use and OTC use.

ACCU NEWSTM One Step Pregnancy Test Midstream: The ACCU NEWSTM One Step

{5}------------------------------------------------

Pregnancy Test Midstream is a lateral flow immunoassay to qualitatively detect the presence of human Chorionic Gonadotropin (hCG) hormone in urine specimen to aid early detection of pregnancy for both prescription use and OTC use.

Similarities
FeatureSubject DevicePredicate Device
Intended UseIt is intended for bothprofessional and over-the-counter use for qualitativedetection of hCG hormone inurine to aid early detection ofpregnancy.It is intended for bothprofessional and over-the-counter use for qualitativedetection of hCG hormone inurine to aid early detection ofpregnancy.
Specimen TypeUrineUrine
TechnologyLateral Flow ImmunoassayLateral Flow Immunoassay
Basic ComponentsTest strip containingnitrocellulose membrane,sample pad, alpha and betahCG antibodiesTest strip containingnitrocellulose membrane,sample pad, alpha and beta hCGantibodies
Detection Limit25 mIU/mL25 mIU/mL
Storage Temperature4-30°C4-30°C
FormatsStrip, Cassette, MidstreamStrip, Cassette, Midstream
TraceabilityWHO 3rd InternationalStandardWHO 3rd International Standard
Differences
FeatureSubject DevicePredicate Device
Read Timeat 5 minutes3-5 minutes
High Dosage HookEffectNo high dosage hook effectfor hCG up to 500,000mIU/mLNo high dosage hook effect forhCG up to 100,000 mIU/mL
SpecificityLH at 1000mIU/mL, FSH at1000mIU/mL, and TSH at1000μIU/mLLH at 300mIU/mL, FSH at 300mIU/mL, and TSH at 1000mIU/mL
pH InterferenceNo interference for urine withpH 2-9No interference for urine withpH 4-9
Specific GravityInterferenceNo interference for urine withSpecific Gravity 1.003-1.030No interference for urine withSpecific Gravity 1.000-1.050
  • F. Summary of the technological characteristics of your device compared to the predicate device

G. Non-clinical performance data

a. Read Time Window

To determine the optimal time to interpret the test results of the ACCU NEWS™ One Step hCG Pregnancy Test, 10 devices from each format (Strip, Cassette, and Midstream) were tested with each of the urine samples containing 0, 12.5, 18.75, 25, 50, and 100mIU/mL hCG. Data indicated that test result was most accurate when it

{6}------------------------------------------------

was interpreted at 5 minutes.

b. Precision/Sensitivity

    1. Three lots of devices for each format were used to test blind-labelled urine samples containing 0, 12.5, 18.75, 25, 50, and 100mIU/mL of hCG. All urine samples were calibrated against the hCG standard WHO 3td IS. The study was carried out at three sites (one internal site and two external sites representative of the intended use setting). Each site tested a unique lot of test device. For each format, at least 30 urine samples were tested for each concentration.
hCG conc. (mIU/mL)Lot #1Lot #2Lot #3Total
+-+-+-+-
0010010010030
12.5010010010030
18.75010010010030
25100100100300
50100100100300
100100100100300

Table 1: Precision Study Data for Strip

Table 2: Precision Study Data for Cassette
hCG conc. (mIU/mL)Lot #1Lot #2Lot #3Total
+-+-+-+-
0010010010030
12.5010010010030
18.75010010010030
25100100100300
50100100100300
100100100100300
+-+-+-+-
0010010010030
12.5010010010030
18.75010010010030
25100100100300
50100100100300
100100100100300

Table 3: Precision Study Data for Midstream

hCG conc. (mIU/mL)Lot #1Lot #2Lot #3Total
+-+-+-+-
0010010010030
12.5010010010030
18.75010010010030
25100100391591
50100100100300
100100100100300

Test result demonstrated that all three formats of ACCU NEWS™ One Step hCG Pregnancy Test were able to produce consistent and precise test results.

    1. Linearity/Assay reportable range:
      Not applicable. This is a qualitative test.
    1. Detection Limit:
      The Precision/Sensitivity study demonstrated that the detection limit for the test was 25 mIU/mL.

{7}------------------------------------------------

Traceability, Stability, Expected values (controls, calibrators, or methods) C. The test is calibrated against reference material traceable to WHO International Standard 3rd edition.

A shelf-life stability test of the device was performed in accelerated temperature at 60°C. The results support that the device was stable for 2 years when stored at 4-30°C.

Comparison Study d.

Urine samples were randomly collected from 250 female donors from three sites at various times throughout the day. About 50% of the donors were suspected to be pregnant. Donor age range was from 18 to 43. Healthcare professionals tested each urine sample with each format of ACCU NEWSTM One Step hCG Pregnancy Test and a predicate hCG test. Test results were compared.

Table 4. Data Summary for Comparison Study
PredicateDeviceResultACCU NEWS™ One Step hCG Pregnancy Test
MidstreamCassetteStrips
PositiveNegativePositiveNegativePositiveNegative
Positive124012401240
Negative012601260126
Total124126124126124126
% Agreement100%100%100%

Table 4: Data Summary for Comparison Study

Test result demonstrated that test results from all three formats of the ACCU NEWS"M One Step hCG Pregnancy Test had 100% agreement with that of the predicate device.

e. Specificity

Specificity was determined by testing the ACCU NEWS™ One Step hCG Test with structurally related compounds spiked in negative urine samples and positive (hCG 25 mIU/mL) urine samples. The compounds analyzed were luteinizing hormone (LH), follicle-stimulating hormone (FSH), and thyroid stimulating hormone (TSH). Five tests were performed for each compound. Result demonstrated that ACCU NEWSTM One Step hCG Test was not interfered by FSH, TSH, and LH at the levels of 1000mIU/mL, 1000µIU/mL, and 1000mIU/mL respectively. However, LH, FSH, and TSH may cause false negative readings at concentrations of >1000 mIU/mL, >1000 mIU/mL, and >1000 µIU/mL respectively.

f. Interference

1. Exogenous Compounds

To determine if some exogenous compounds would react and interfere with the ACCU NEWS™M One Step hCG Pregnancy Test, positive urine samples (hCG 25 mIU/mL) and negative urine samples were spiked with individual chemicals to specific concentrations. Each spiked urine sample was tested with five ACCU NEWS TM One Step hCG Pregnancy Test devices from each format. Results showed that ACCU NEWS™ One Step hCG Pregnancy Test was not interfered

{8}------------------------------------------------

SubstanceConcentration
(1R,2S)-(-)-Ephedrine20 mg/dL
(1S,2R)-(+)-Ephedrine20 mg/dL
Acetylsalicylate Acid20 mg/dL
Albumin (Human)2000 mg/dL
Ascorbic Acid20 mg/dL
Atropine20 mg/dL
Acetaminophen20 mg/dL
Ampicillin5.3 mg/dL
Bilirubin2 mg/dL
Caffeine20 mg/dL
Cannabinol10 mg/dL
Gentisic Acid20 mg/dL

by the substances listed below at their specific concentrations.

Substance Concentration 2000 mg/dL Glucose Hemoglobin 250 mg/dL Ibuprofen 40 mg/dL Methadone 10 mg/dL Morphine 10 mg/dL 20 mg/dL Nicotine 1.5 mg/dL Tetracycline 20 mg/dL Acetone 20 mg/dL Phenylpropanolamine Salicylic acid 20 mg/dL Uric Acid 20 mg/dL DL-ß-Hydroxybutyric Acid 2000 mg/dL

2. pH Study

To determine if different pH values in urine would have an effect on the ACCU NEWS™ One Step hCG Pregnancy Test, positive (hCG 25 mIU/mL) and negative urine samples with the following pH values: 2.0, 3.0, 4.0, 5.0, 6.0, 7.0, 8.0, and 9.0 were tested. Each urine sample was tested with three test devices from each format of the ACCU NEWS™M One Step hCG Test. Result showed that pH values 2.0-9.0 in urine did not interfere with the performance of the test.

3. Specific Gravity

To determine if different specific gravity values in urine can affect the ACCU NEWSTM One Step hCG Pregnancy Test, positive (25 mIU/mL) and negative hCG urine samples with the following specific gravity levels: 1.003, 1.005, 1.010, 1.015, 1.020, and 1.030 were tested. Each urine sample was tested with three test devices from each format of the ACCU NEWS 110 One Step hCG Pregnancy Test. Result showed that the test functioned properly with no interference from these various specific gravity values in urine.

4. High Dose/Hook Effect

To determine if high concentrations of hCG would cause false negative result on the ACCU NEWS™ One Step hCG Pregnancy Test. Urine samples spiked with 50, 100, 200, 300, and 500 IU/mL of hCG were tested with the device. Each urine sample was tested with three test devices from each format of the ACCU NEWSTM One Step hCG Pregnancy Tes. Result showed that the device did not exhibit high dose/hook effect for urine containing up to 500 IU/mL (or 500,000 mIU/mL) of hCG.

5. hCG ß-core fragment

To determine the effects of ß-core hCG fragment on the ACCU NEWS™ One Step hCG Test, the device was tested with negative and positive (hCG 25mIU/mL) hCG urine samples spiked with ß-core hCG fragment at the concentrations of 20,000 pmol/L, 25,000 pmol/L, 50,000 pmol/L, 62,500 pmol/L, 100,000 pmol/L, 125,000 pmol/L, 250,000 pmol/L, 500,000 pmol/L

{9}------------------------------------------------

and 1000.000 pmol/L. Result showed that ß-core hCG fragments did not interfere with the test result of the device at or below the concentration of 100,000pmol/L. However, at a concentration of >100,000 pmol/L. hCG ß-core Fragment may cause false negative reading.

H. Clinical performance data

  • a. Clinical Sensitivity: Not Applicable
  • b. Clinical Specificity: Not Applicable

Other Clinical Supportive Data: ﻥ OTC lay-user Test Result

Urine samples were randomly collected from 250 female donors at three sites. About 50% of the donors were suspected to be pregnant. Donor age range was from 18 to 43. Donors were given all three formats of the ACCU NEWSTM One Step hCG Pregnancy Test to test the collected urine samples. The same urine samples were tested by healthcare professionals using the ACCU NEWSTM One Step hCG Pregnancy Test. The recorded test results from the lay users were compared with that from the healthcare professionals.

ACCU NEWS™ One Step hCG Pregnancy Test Strip
PositiveNegative
ProfessionalTest Result(StripFormat)Positive1240
Negative0126
Total124126
% Agreement100%

Table 5: Data Summary of OTC Lay-User Test Result for Strip

Table 6: Data Summary of OTC Lay-User Test Result for Cassette

ACCU NEWS™ Step hCG Pregnancy Test Cassette
PositiveNegative
ProfessionalTest Result(CassetteFormat)Positive1240
Negative0126
Total124126
% Agreement100%

Table 7: Data Summary of OTC Lay-User Test Result for Midstream

ACCU NEWS™ One Step hCG Pregnancy Test Midstream
PositiveNegative
ProfessionalTest Result(MidstreamFormat)Positive1240
Negative0126
Total124126
% Agreement100%

{10}------------------------------------------------

Data showed 100% agreement between lay user results and professional result. This demonstrated that the device was easy enough to be used by untrained lay users. In addition, each participant was given an English questionnaire to assess the readability of the labeling. The result showed that lay users found the product instruction to be easy to understand and the device to be easy to use.

I. Conclusion

The performance characteristics studies performed demonstrated substantial equivalency between ACCU NEWS™ One Step hCG Pregnancy Test and the predicate device.

§ 862.1155 Human chorionic gonadotropin (HCG) test system.

(a)
Human chorionic gonadotropin (HCG) test system intended for the early detection of pregnancy —(1)Identification. A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class II.(b)
Human chorionic gonadotropin (HCG) test system intended for any uses other than early detection of pregnancy —(1)Identification. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.(2)
Classification. Class III.(3)
Date PMA or notice of completion of a PDP is required. As of the enactment date of the amendments, May 28, 1976, an approval under section 515 of the act is required before the device described in paragraph (b)(1) may be commercially distributed. See § 862.3.