K Number
K150037

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-10-16

(280 days)

Product Code
Regulation Number
892.5900
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SRT-100 Vision System is a low energy x-ray system, with ultrasound imaging capability, intended for superficial radiotherapy and electronic brachytherapy treatments of primary malignant epithelial neoplasms of the skin and keloids. Applications include: (a) basal cell carcinoma; (b) squamous cell carcinoma; (c) Metatypic carcinoma; (d) cutaneous appendage carcinoma (e) Kaposi's Sarcoma; and (f) the treatment of keloids are benign fibrous growths that arise from proliferation of dermal tissue typically arising from injuries to skin tissue.

The ultrasound capability, employed in a non-diagnostic mode, is used to assist the physician in the selection of the correct cone applicator size. The Derma-Scan C Ultrasound component was initially cleared with an indication for use as an ultrasonic scanning system used to visualize the layers of skin, including bold vessels, and to make approximate measurements of dimensions in layers of skin and blood vessels, by ultrasonic means.

The red-diode laser assembly is a commercial pointer device employed by physicians for improving the alignment of the focused beam.

Device Description

The Sensus Healthcare SRT-100 Vision is a complete, stand-alone, x-ray radiation therapy system. It consists of five separate components: (a) control console; (b) base unit; (c) ultrasound imaging; (d) red-diode laser; and (e) applicators.

Control Console: Specifically designed module housing the switches and indicators used by the operator to set up and execute x-ray exposures. The controls adjust the machine functions and settings only! There is no treatment planning capability. The Control Console is connected, through a cable, to the Base Unit.

Base Unit: The base unit consists of a cabinet containing the high voltage generator, power supply components, cooling system, and an arm/positioning mechanism on which the x-ray tube housing assembly is mounted. A series of Applicators are included, which are affixed to the x-ray port on the x-ray tube housing assembly to limit the x-ray beam and provide fixed Source-to-Skin Distance (SSD). The X-ray Tube-Housing Assembly contains a motorized filter mechanism, which moves the appropriate beam filter: (a) 0.10 mm Al at 20 to 30 kV; (b) 0.10 mm Cu at 50 to 100 kV; (c) 0.45 mm Al at 50kV; (d) 0.75 mm Al at 70 kV; (e) 1.15 mm Al at 100 kV; and (f) 4.0 mm Al at 50 to100 kV; into the beam path depending on the kV setting selected by the operator.

Ultrasound Imaging: The Derma-Scan C Ultrasound System component is integrated with the SRT-100 Vision computer and contains: (a) scanning main unit; (b) handheld probe and (c) a medical grade power supply to provide power to the computer. The ultrasound component is designed to meet international safety requirements.

Red-Diode Laser: A red-diode laser is integrated with the SRT-100 Vision System. The laser is manufactured by U.S. Laser and is classified as FDA Laser Class 3A. The application of the red-diode laser with the Sensus SRT-100 Vision has been tested in accordance with IEC 60825-1.

Applicators: The system is shipped with a set of interchangeable treatment applicators, which define the source to skin distance (SSD) and the diameter of the treatment beam's exposure. The applicator size, therefore, determines the amount of total dose delivered per minute to the lesion and the actual area that will be treated by the system's x-ray beam. Each applicator is embedded with a unique magnet binary combination, which allows the system to automatically detect an applicator as it is mounted on the x-ray port. This provides the system with the information about the applicator's SSD and diameters. which allows it to correlate the applicable dose rate for each applicator that is attached to the x-ray port, thus allowing for a precise and user-error-free dose rate per minute calculation. There are a variety of applicator sizes available for use with the Sensus Healthcare SRT-100 Vision System, driven by the treatment modality.

AI/ML Overview

Here's the breakdown of the acceptance criteria and study information for the Sensus Healthcare Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capabilities (SRT-100 Vision System), based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance:

The document does not explicitly provide a table of acceptance criteria with specific numerical performance metrics for the device's efficacy in treating medical conditions. Instead, it focuses on demonstrating substantial equivalence to a predicate device and adherence to safety and electrical standards.

The key "performance" demonstrated is that the device meets these standards and is "as safe and effective" as the predicate.

Here's a table based on the information provided, inferring the acceptance criteria from the testing described:

Acceptance Criterion (Inferred from testing)Reported Device Performance
Conformance to IEC 60601-1 (General Safety)Meets IEC 60601-1:2007 (3rd edition)
Conformance to IEC 60601-1-2 (EMC)Meets IEC 60601-1-2:2007
Conformance to IEC 60601-1-4 (Programmable Systems)Meets IEC 60601-1-4:1996
Conformance to IEC 60601-1-6 (Usability)Meets IEC 60601-1-6:2010
Conformance to IEC 60601-2-8 (Therapeutic X-ray)Meets IEC 60601-2-8:2010
Conformance to IEC 60601-2-32 (X-ray Associated Equipment)Meets IEC 60601-2-32:1994
Conformance to IEC 60601-2-37 (Ultrasonic Diagnostic/Monitoring)Meets IEC 60601-2-37:2007
Conformance to IEC 60825-1 (Laser Safety)Meets IEC 60825-1:2007
Conformance to IEC 62366 (Usability Engineering)Meets IEC 62366:2007
Non-Clinical Performance Testing (Functional, Bench, System)All testing was performed in accordance with defined test cases with clearly delineated acceptance criteria. Bench testing, functional testing, and overall system performance testing were conducted using FDA consensus standards and recognized ISO and IEC standards.
Substantial Equivalence to Predicate Device (SRT-100 Vision K131582)No functional difference other than the addition of electronic brachytherapy. Changes to x-ray tube and technology improvements have no impact on form, fit, function, safety, and/or efficacy. Device is as safe and effective and performs as well as the predicate device.

2. Sample Size for the Test Set and Data Provenance:

The document primarily describes non-clinical testing (bench, functional, system performance, safety standards adherence). Therefore, there isn't a "test set" in the sense of a clinical patient cohort.

  • Sample Size for Test Set: Not applicable for a clinical test set; refers to the device itself and its components.
  • Data Provenance: The testing was conducted by "qualified and accredited independent laboratories" (page 8). The exact country of origin is not specified, but the standards cited are international (IEC, ISO). The nature of this testing is prospective in the sense that the device was designed and then tested against established engineering and safety standards.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications:

This question is largely not applicable for the type of non-clinical testing described. The "ground truth" for compliance with electrical and safety standards is the standard itself, as interpreted and verified by the independent testing laboratories.

  • Number of Experts: Not specified, as this wasn't a clinical study requiring expert review of patient data.
  • Qualifications of Experts: Not specified, but implied to be engineers and specialists from "qualified and accredited independent laboratories" who are experts in the listed IEC/ISO standards.

4. Adjudication Method for the Test Set:

None. Adjudication methods (like 2+1 or 3+1) are used in clinical studies where multiple human readers independently interpret data, and conflicts are resolved by a tie-breaking reader or consensus. This document describes engineering and performance testing against objective standards, not interpretation of clinical data by multiple readers.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study was done, and effect size:

No. The document does not describe an MRMC comparative effectiveness study where human readers' performance with and without AI assistance was evaluated. The device itself is an X-ray radiation therapy system with ultrasonic imaging capabilities, not an AI-assisted diagnostic tool for human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) was done:

The device is a medical device (X-ray radiation therapy system) that delivers treatment and assists with treatment planning (ultrasound for cone selection). It is not an algorithm that performs a diagnostic task independently. While its components (like the ultrasound) have defined functionalities, the concept of "standalone algorithm performance" as typically applied to AI/CADe devices for diagnosis is not relevant in this context. The core function is therapy delivery, which is inherently human-controlled.

7. The Type of Ground Truth Used:

For the non-clinical performance and safety testing, the "ground truth" was:

  • Established engineering specifications and product requirements for the Sensus Healthcare SRT-100 Vision System.
  • International electrical safety and medical device standards (e.g., IEC 60601 series, IEC 60825-1, IEC 62366).
  • The performance and safety profile of the predicate device (Sensus Healthcare SRT-100 Vision K131582), to which substantial equivalence was claimed.

For the clinical literature review referenced (page 7), the ground truth for efficacy would have come from the outcomes reported in those studies (e.g., tumor regression, keloid reduction, recurrence rates). However, this review is not a primary study conducted for this 510(k) but rather a summary of existing literature.

8. The Sample Size for the Training Set:

Not applicable. The device is an X-ray radiation therapy system. It is not an AI/machine learning algorithm that requires a "training set" of data in the typical sense. Its design and engineering are based on established physics, medical knowledge, and safety standards, rather than statistical learning from a dataset.

9. How the Ground Truth for the Training Set was Established:

Not applicable, as there is no training set for this type of device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing, ribbon-like shape at the bottom. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 16, 2015

Sensus Healthcare % Mr. Kal Fishman Chief Technical Officer 851 Broken Sound Pkwy NW, Suite #215 BOCA RATON FL 33487

Re: K150037

Trade/Device Name: Sensus Healthcare Superficial X-rav Radiation Therapy System with Ultrasonic Imaging Capabilities

Regulation Number: 21 CFR 892.5900 Regulation Name: X-ray radiation therapy system Regulatory Class: II Product Code: JAD Dated: September 18, 2015 Received: September 21, 2015

Dear Mr. Fishman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert Ools

Robert A. Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K150037

Device Name

Superficial X-ray Radiation Therapy System with Ultrasonic Imaging Capability

Indications for Use (Describe)

The SRT-100 Vision System is a low energy x-ray system, with ultrasound imaging capability, intended for superficial radiotherapy and electronic brachytherapy treatments of primary malignant epithelial neoplasms of the skin and keloids. Applications include: (a) basal cell carcinoma; (b) squamous cell carcinoma; (c) Metatypic carcinoma; (d) cutaneous appendage carcinoma (e) Kaposi's Sarcoma; and (f) the treatment of keloids are benign fibrous growths that arise from proliferation of dermal tissue typically arising from injuries to skin tissue.

The ultrasound capability, employed in a non-diagnostic mode, is used to assist the physician in the selection of the correct cone applicator size. The Derma-Scan C Ultrasound component was initially cleared with an indication for use as an ultrasonic scanning system used to visualize the layers of skin, including bold vessels, and to make approximate measurements of dimensions in layers of skin and blood vessels, by ultrasonic means.

The red-diode laser assembly is a commercial pointer device employed by physicians for improving the alignment of the focused beam.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Section Five (5) - 510(k) Summary

510(k) SUMMARY

[As Required by 21 CFR 807.92(c)]

Submitter's Name & Address:Sensus Healthcare851 Broken Sound Parkway NWSuite 215Boca Raton, FL 33487
Contact Person:Kal Fishman, CTOTelephone (561) 922-5808Fax (561) 948-2071kal@sensushealthcare.com
Date Summary Prepared:October 9, 2015
Device Name:Trade/Proprietary Name – SRT-100 VisionCommon/Usual Name – Superficial X-ray RadiationTherapy System with UltrasonicImaging CapabilityClassification Name – X-ray Radiation Therapy System(892.5900)
Classification:Class II
Product Code:JAD
Regulation Number:892.5900
Predicate Device:Sensus Healthcare SRT-100 Vision (K131582)

{4}------------------------------------------------

Reference Devices:

Cortex Technology Derma-Scan C Ultrasound System (K983945)

Nucletron Esteya Electronic Brachytherapy System (K132092)

Previous FDA Submissions & Clearances

Table 1.0 contains a compilation of previously cleared versions of the SRT-100 and SRT-100 Vision Systems.

ClearanceDateClassProductCodeSystemIndications for Use
K13158228 August2013IIJADSRT-100VisionThe SRT-100 Vision is a low energy x-ray system, with ultrasound imaging and red-diode laser capabilities, intended for superficial radiotherapy treatment of primary malignant epithelial neoplasms of the skin and keloids.
K12398514 May2013IIJADSRT-100The SRT-100 is a low energy x-ray system intended for superficial radiotherapy treatment of primary malignant epithelial neoplasms of the skin and keloids.
K063456*9 January2007IIJADSRT-100The SRT-100 is a low energy x-ray system intended for superficial radiotherapy treatment of primary malignant epithelial neoplasms of the skin.

Table 1.0 – Previously Cleared SRT-100 Systems

*Note: This original clearance was provided to Topex, Inc. in 2007. It was also the predicate device referenced for the Esteya System.

Device Description

The Sensus Healthcare SRT-100 Vision is a complete, stand-alone, x-ray radiation

therapy system. It consists of five separate components: (a) control console; (b) base unit; (c)

ultrasound imaging; (d) red-diode laser; and (e) applicators.

Control Console: Specifically designed module housing the switches and indicators used

by the operator to set up and execute x-ray exposures. The controls adjust the machine

functions and settings only! There is no treatment planning capability. The Control

Console is connected, through a cable, to the Base Unit.

{5}------------------------------------------------

Base Unit: The base unit consists of a cabinet containing the high voltage generator, power supply components, cooling system, and an arm/positioning mechanism on which the x-ray tube housing assembly is mounted. A series of Applicators are included, which are affixed to the x-ray port on the x-ray tube housing assembly to limit the x-ray beam and provide fixed Source-to-Skin Distance (SSD). The X-ray Tube-Housing Assembly contains a motorized filter mechanism, which moves the appropriate beam filter: (a) 0.10 mm Al at 20 to 30 kV; (b) 0.10 mm Cu at 50 to 100 kV; (c) 0.45 mm Al at 50kV; (d) 0.75 mm Al at 70 kV; (e) 1.15 mm Al at 100 kV; and (f) 4.0 mm Al at 50 to100 kV; into the beam path depending on the kV setting selected by the operator.

Ultrasound Imaging: The Derma-Scan C Ultrasound System component is integrated with the SRT-100 Vision computer and contains: (a) scanning main unit; (b) handheld probe and (c) a medical grade power supply to provide power to the computer. The ultrasound component is designed to meet international safety requirements.

Red-Diode Laser: A red-diode laser is integrated with the SRT-100 Vision System. The laser is manufactured by U.S. Laser and is classified as FDA Laser Class 3A. The application of the red-diode laser with the Sensus SRT-100 Vision has been tested in accordance with IEC 60825-1.

Applicators: The system is shipped with a set of interchangeable treatment applicators, which define the source to skin distance (SSD) and the diameter of the treatment beam's exposure. The applicator size, therefore, determines the amount of total dose delivered per minute to the lesion and the actual area that will be treated by the system's x-ray beam. Each applicator is embedded with a unique magnet binary combination, which allows the system to automatically detect an applicator as it is mounted on the x-ray port.

{6}------------------------------------------------

This provides the system with the information about the applicator's SSD and diameters. which allows it to correlate the applicable dose rate for each applicator that is attached to the x-ray port, thus allowing for a precise and user-error-free dose rate per minute calculation. There are a variety of applicator sizes available for use with the Sensus Healthcare SRT-100 Vision System, driven by the treatment modality.

Intended Use:

The SRT-100 Vision System is a low energy x-ray system. with ultrasound imaging capability, intended for superficial radiotherapy and electronic brachytherapy treatments of primary malignant epithelial neoplasms of the skin and keloids. Applications include: (a) basal cell carcinoma; (b) squamous cell carcinoma; (c) Metatypic carcinoma; (d) cutaneous appendage carcinoma (e) Kaposi's Sarcoma; and (f) the treatment of keloids are benign fibrous growths that arise from proliferation of dermal tissue typically arising from injuries to skin tissue.

The ultrasound capability, employed in a non-diagnostic mode, is used to assist the physician in the selection of the correct cone applicator size. The Derma-Scan C Ultrasound component was initially cleared with an indication for use as an ultrasonic scanning system used to visualize the layers of skin, including bold vessels, and to make approximate measurements of dimensions in layers of skin and blood vessels, by ultrasonic means.

The red-diode laser assembly is a commercial pointer device employed by physicians for improving the alignment of the focused beam.

Prescriptive Statement

Caution: Federal law restricts this device to sale by or on the order of a physician.

{7}------------------------------------------------

Technological Characteristics/Principles of Operation

The SRT-100 Vision produces and emits filtered, low energy (20 to 100 kV) x-ray radiation, which is electrically generated using a conventional ceramic x-ray tube (Comet MXR-100). Provision is made to limit the x-radiation to a specified treatment field, and to control the radiation dose to the patient through selection and monitoring of energy, emission level and duration of emission. To mitigate effects of ionizing radiation on healthy cells, and to accumulate more damage in the neoplastic cells and keloids associated with scar tissue, the total dose is fractionated, which means distributing the total dose over a period of time. Typically, 8 to 12 fractions at a rate of 1 to 5 per week are used to deliver a total dose of 40-60 Gy, although larger PMENs may require up to 40 fractions over an 8-week period for a total dose of 80 Gy. When treating keloids, typically 1 to 4 fractions are employed, delivering a total dose in the range of 10 to 40 Gy.

Review of Clinical Literature

A summary of multiple clinical studies for the treatment of keloids, electronic brachytherapy, and supporting literature has been collected, reviewed, and a clinical evaluation report scripted in support of this 510(k). The literature collected and reviewed supports Sensus Healthcare's claims of safety and efficacy for the application of superficial radiotherapy and electronic brachytherapy treatments of primary malignant epithelial neoplasms of the skin and keloids. Applications include: (a) basal cell carcinoma; (b) squamous cell carcinoma; (c) Metatypic carcinoma; (d) cutaneous appendage carcinoma (e) Kaposi's Sarcoma; and (f) the treatment of keloids. Keloids are benign fibrous growths that arise from proliferation of dermal tissue typically arising from injuries to skin tissue.

{8}------------------------------------------------

Summary of Non-Clinical Performance Testing

The Sensus Healthcare SRT-100 Vision has identical capabilities with the SRT-100 Vision System predicate device; and similar capabilities with (b) the Esteya Electronic Brachytherapy System. The Sensus Healthcare SRT-100 Vision System, as configured, has been engineered and tested to meet Sensus Healthcare product requirements, required electrical and mechanical safety standards, and meet clinical expectations. All testing of the SRT-100 Vision System was performed in accordance with defined test cases with clearly delineated acceptance criteria. Additionally, FDA consensus standards and recognized ISO and IEC standards (e.g., IEC 60601-1 3th edition) were employed for the bench testing, functional testing, and overall system performance testing of the SRT-100 Vision. Furthermore, all testing was performed by qualified and accredited independent laboratories.

Non-clinical Safety Tests

The Sensus Healthcare SRT-100 Vision has been designed and constructed to meet the following electrical and mechanical safety standards:

  • IEC 60601-1:2007 Medical electrical equipment Part 1: General o requirements for basic safety and essential performance (3td edition)
  • IEC 60601-1-2:2007 Medical equipment Part 1: General requirements o for safety. Collateral standard: electromagnetic compatibility requirements and tests
  • IEC 60601-1-4:1996 Medical electrical equipment--Part O 1-4: General requirements for safety - Collateral Standard: Programmable electrical medical systems
  • IEC 60601-1-6:2010 Medical electrical equipment Part o 1-6: General requirements for safety - Collateral Standard: Usability
  • IEC 60601-2-8:2010 Medical equipment Part 2-8: о Particular requirements for the safety of therapeutic X-ray equipment operating in the range 10 kV to 1MV

{9}------------------------------------------------

  • IEC 60601-2-32:1994 Medical electrical equipment Part o 2: Particular requirements for the safety of associated equipment of X-ray equipment
  • IEC 60601-2-37:2007 Particular requirements for the o safety of ultrasonic medical diagnostic and monitoring equipment
  • IEC 60825-1:2007 Safety of laser products Part 1: O Equipment classification and requirements
  • IEC 62366:2007 Medical devices -- Application of o usability engineering to medical devices

Substantial Equivalence Discussion

Sensus Healthcare SRT-100 Vision (K131582)

Other than the addition of electronic brachytherapy capability as an additional indication for use, there is no functional difference between the SRT-100 Vision versus the previously cleared SRT-100 Vision. Changes to the x-ray tube and other technology improvements have no impact on the form, fit, function, safety, and/or efficacy of the SRT-100 Vision.

Nucletron, B.V Esteya System (K132092)

The Nucletron, B.V. Esteya System was cleared by FDA on 26 September 2013 (K132092).

The Esteya System referenced the Topex SRT-100 System as one of their predicate devices. The

Topex SRT-100 and associated intellectual properties (IP) was acquired by Sensus Healthcare in

  1. The Topex SRT-100 was the predicate for the first Sensus Healthcare SRT-100 as the

technology (IP) belongs to Sensus Healthcare.

Conclusion Statement

The SRT-100 Vision System has the same intended use as the predicate device. Any technological changes to the device do not raise new questions of safety or effectiveness. Performance testing, along with verification and validation activities demonstrate that SRT-100

{10}------------------------------------------------

Vision is as safe and effective, and performs as well as the predicate device. Therefore, SRT-100 Vision can be considered substantially equivalent to the predicate device.

§ 892.5900 X-ray radiation therapy system.

(a)
Identification. An x-ray radiation therapy system is a device intended to produce and control x-rays used for radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.(b)
Classification. Class II.