K Number
K150035

Validate with FDA (Live)

Device Name
V10 system
Manufacturer
Date Cleared
2015-05-01

(112 days)

Product Code
Regulation Number
878.4400
Age Range
All
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Viora V10 system is intended for dermatological procedures.

The V-ST Handpiece is indicated for the temporary relief of minor muscle aches and pain, temporary relief of muscle spasm, and temporary improvement of local blood circulation.

The V-Form Handpiece (with BC and FC applicators) is indicated for delivering non thermal RF combined with massage:

  • · relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation and
  • . temporary reduction in the appearance of cellulite.
Device Description

The Viora V10 system is a RF multi application platform with two available treatment Handpieces:

  • V-ST Handpiece Bi-polar radiofrequency (RF) Handpiece.
  • · V-Form Handpiece (with BC and FC applicators) mechanical vacuum massage and Bi-polar radiofrequency (RF) Handpiece.

The Main Unit (console) provides the operational and safety function of the system. The operator can modify the treatment parameters to achieve specific tissue effects depending on individual patient's skin condition and anatomical structure.

AI/ML Overview

This document describes the Viora V10 system, an RF multi-application platform intended for dermatological procedures. It leverages existing predicate devices (Viora V20 system K142093 and Viora Reaction™ system K090221) to establish substantial equivalence.

Here's a breakdown of the requested information based on the provided text:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance CriteriaReported Device Performance
Safety StandardsComplies with:
IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.Confirmed compliance.
IEC 60601-1-2: Medical Electrical Equipment Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.Confirmed compliance.
IEC 60601-2-2: Medical Electrical Equipment - Part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories.Confirmed compliance.
Effectiveness (Bench Testing)Bench testing demonstrated that the V10 system is as safe and effective as the cleared predicate devices.
Substantial EquivalenceThe V10 system console has similar technological characteristics as the predicate device (Viora V20). The Viora V10 V-ST Handpiece has the same intended use, technological features, and performance characteristics as the cleared Viora V20 V-ST Handpiece (K142093). The Viora V10 V-Form Handpiece (with BC and FC applicators) and the Viora Reaction™ B-contour and F-contour applicators (K090221) have identical intended use and similar technological features.
No New Questions of Safety & EffectivenessPerformance data demonstrates that the differences between the V10 console and the predicate do not raise any new questions of safety and effectiveness. Any differences in the V-Form Handpiece design do not raise any new questions of safety and effectiveness, as verified by performance testing.

2. Sample size used for the test set and the data provenance

The document does not describe a specific test set with a sample size for evaluating the V10 system through human subject data. Instead, it relies on predicate device data and bench testing.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

Not applicable, as no dedicated human subject test set is described with ground truth established by experts.

4. Adjudication method for the test set

Not applicable, as no dedicated human subject test set is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC comparative effectiveness study is mentioned. This device is an RF multi-application platform, not an AI-assisted diagnostic tool typically assessed with MRMC studies.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

This is not applicable as the device is not an algorithm performing a diagnostic function. The "standalone" performance here refers to the device's functional and safety performance through bench testing, which was done.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.)

For the V10 system itself, the "ground truth" for demonstrating safety and effectiveness relies on:

  • Compliance with recognized international standards (IEC 60601 series).
  • Bench testing results comparing its performance to predicate devices.
  • Demonstrated equivalence of technological characteristics and intended use to previously cleared predicate devices.

8. The sample size for the training set

Not applicable. This document describes a medical device seeking 510(k) clearance based on substantial equivalence to predicate devices, not an AI/ML algorithm that requires a training set.

9. How the ground truth for the training set was established

Not applicable, as no training set for an AI/ML algorithm is mentioned.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing right, with flowing lines representing hair or clothing.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 1, 2015

Viora Ltd. Omri Kesler Chief Operating Officer 3 Maskit Street 4673303 Herzliya Israel

Re: K150035

Trade/Device Name: V10 system Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: PBX. ISA Dated: March 31, 2015 Received: April 2, 2015

Dear Mr. Kesler:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K150035

Device Name V10 system

Indications for Use (Describe)

The Viora V10 system is intended for dermatological procedures.

The V-ST Handpiece is indicated for the temporary relief of minor muscle aches and pain, temporary relief of muscle spasm, and temporary improvement of local blood circulation.

The V-Form Handpiece (with BC and FC applicators) is indicated for delivering non thermal RF combined with massage: · relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation and

· temporary reduction in the appearance of cellulite.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for VIORA. The logo consists of a stylized infinity symbol above the word "VIORA" in a bold, sans-serif font. Below the word "VIORA" is the tagline "Shape the Future" in a smaller, lighter font.

510(k) Summary

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act of 1990 and 21

CFR 807.92.

Submitter Nameand Address:Viora Ltd.3 Maskit StreetHerzliya, Israel 4673303Israel
Contact Person:Omri KeslerCOOEmail: omri@vioramed.comPhone Number: +972 9955 1344Fax Number: +972 9955 1345
EstablishmentRegistrationNumber:3005695724
Date Prepared:May 01, 2015
Device TradeName(s):V10 system
Device CommonName:Multi application RF device
Classification:Name: Electrosurgical cutting and coagulation device andaccessoriesProduct code: PBX, ISARegulation No: 21CFR878.4400, 21 CFR 890.5660Class: IIPanel: General and plastic surgery devices
PredicateDevice(s):Viora V20 system (K142093)Viora Reaction™ system (K090221)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for VIORA. The logo consists of a stylized infinity symbol above the word "VIORA" in a bold, sans-serif font. Below "VIORA" is the tagline "Shape the Future" in a smaller, lighter font. The logo is black and white.

Device description

The Viora V10 system is a RF multi application platform with two available treatment Handpieces:

  • V-ST Handpiece Bi-polar radiofrequency (RF) Handpiece.
  • · V-Form Handpiece (with BC and FC applicators) mechanical vacuum massage and Bi-polar radiofrequency (RF) Handpiece.

The Main Unit (console) provides the operational and safety function of the system. The operator can modify the treatment parameters to achieve specific tissue effects depending on individual patient's skin condition and anatomical structure.

Intended use and indication for use statement

The Viora V10 system is intended for dermatological procedures.

The V-ST Handpiece is indicated for the temporary relief of minor muscle aches and pain, temporary relief of muscle spasm, and temporary improvement of local blood circulation.

The V-Form Handpiece (with BC and FC applicators) is indicated for delivering non thermal RF combined with massage:

  • · relief of minor muscle aches and pain, relief of muscle spasm, temporary improvement of local blood circulation and
  • . temporary reduction in the appearance of cellulite.

Predicade Devices

Substantial equivalence to the following predicate devices is claimed:

Device name510k No.Date of Clearance
Viora V20 systemK142093November 14, 2014
Viora Reaction™ systemK090221July 1st, 2009

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for a company called VIORA. The logo consists of the word "VIORA" in a bold, sans-serif font, with the tagline "Shape the Future" underneath in a smaller font. Above the word "VIORA" is a stylized graphic that resembles an infinity symbol with a curved line extending upwards from the left side of the symbol. The logo is simple, modern, and professional.

Substantial Equivalence to Predicate Device

The V10 system console has similar technological characteristics as the predicate device, performance data demonstrate that the differences between the V10 console and the predicate do not raise any new questions of safety and effectiveness. The Viora V10 V-ST Handpiece has the same intended use, the same technological features and the same performance characteristics as the cleared Viora V20 V-ST Handpiece (K142093). The Viora V10 V-Form Handpiece (with BC and FC applicators) and the Viora ReactionTM B-contour and F-contour applicators (K090221) have identical intended use and similar technological features. Any differences in the V-Form Handpiece design do not raise any new questions of safety and effectiveness, as verified by performance testing.

Therefore, the V10 system is substantially equivalent to its predicate devices.

Performance standards

The V10 system complies with:

  • . IEC 60601-1: Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance.
  • IEC 60601-1-2: Medical Electrical Equipment Part 1-2: General . Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Compatibility - Requirements And Tests.
  • . IEC 60601-2-2: Medical Electrical Equipment - Part 2-2: Particular Requirements For The Basic Safety And Essential Performance Of High Frequency Surgical Equipment And High Frequency Surgical Accessories.

Performance Bench Tests

Bench testing demonstrated that the V10 system is as safe and effective as the cleared predicate devices.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for VIORA. The logo consists of a stylized infinity symbol above the word "VIORA" in a bold, sans-serif font. Below the word "VIORA" is the tagline "Shape the Future" in a smaller, lighter font. The overall design is clean and modern.

Pre-Clinical and clinical study

Since the technological parameters of the Viora V10 system are well within the previously cleared Viora V20 (console and V-ST Handpiece) and Reaction (Bcontour and F-contour applicators) systems, Viora believes that animal and clinical studies are not required to determine the safety and efficacy of the V10 system.

Conclusion

Based on the technological characteristics of the devices and the intended use, Viora believes that the V10 system and the predicate devices are substantially equivalent. The differences do not raise any new issues of safety or effectiveness.

§ 878.4400 Electrosurgical cutting and coagulation device and accessories.

(a)
Identification. An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.(b)
Classification. Class II.