(65 days)
IMMULITE® 2000 Androstenedione Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Androstenedione assay on the IMMULITE 2000 systems
IMMULITE® 2000 Troponin I Calibration Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Troponin I assay on the IMMULITE 2000 systems
The IMMULITE® 2000 Androstenedione Calibration Verification Material (CVM) contains one set of four vials each 2mL in liquid form. CVM1 contains processed human serum with preservatives. CVM2, CVM3 and CVM4 contain androstenedione in processed human serum matrix with preservative.
The IMMULITE® 2000 Troponin I Calibration Verification Material (CVM) contains one set of four vials, 2mL each after reconstitution. CVM1 contains equine serum with 0.88% sodium azide and preservative. CVM2, CVM3 and CVM4 contain human Troponin I and rabbit Troponin C in equine serum matrix with 0.88% sodium azide and preservative.
Here's a breakdown of the acceptance criteria and the studies for both the IMMULITE® 2000 Androstenedione Calibration Verification Material and the IMMULITE® 2000 Troponin I Calibration Verification Material, based on the provided document:
Device 1: IMMULITE® 2000 Androstenedione Calibration Verification Material (CVM) (K143636)
1. Table of Acceptance Criteria and Reported Device Performance
The provided document details the stability acceptance criteria. Performance data is presented within the context of demonstrating compliance with these stability criteria rather than as a separate "reported performance" table.
| CVM Level | Dose at Time 0 (ng/mL) | Guideline Criteria % difference to time 0 | Acceptable Dose Range (ng/mL) | Reported Performance (from stability study, implicitly deemed compliant) |
|---|---|---|---|---|
| LAOCVM1 | 0.00 | N/A | ≤ 0.3 | For LAOCVM1, the dose at time 0 is 0.00 ng/mL, and the acceptable range is ≤ 0.3 ng/mL. The study implicitly found values within this range over the stability period, as it states the device "meets the performance Specifications for its intend use." However, specific numerical results for each time point are not provided in this summary. |
| LAOCVM2 | 1.93 | ±10% | 1.74 - 2.12 | For LAOCVM2, the dose at time 0 is 1.93 ng/mL. The acceptable range is 1.74 - 2.12 ng/mL. The stability study results are described as supporting the claimed stability, indicating results fell within this range. |
| LAOCVM3 | 6.68 | ±10% | 6.01 - 7.35 | For LAOCVM3, the dose at time 0 is 6.68 ng/mL. The acceptable range is 6.01 - 7.35 ng/mL. The stability study results are described as supporting the claimed stability, indicating results fell within this range. |
| LAOCVM4 | 15.4 | ±13% | 13.4 - 17.4 | For LAOCVM4, the dose at time 0 is 15.4 ng/mL. The acceptable range is 13.4 - 17.4 ng/mL. The stability study results are described as supporting the claimed stability, indicating results fell within this range. |
2. Sample Size for Test Set and Data Provenance
- Test Set Description: The "test set" in this context refers to the samples used in the stability studies and value assignment.
- Stability Studies: Multiple samples of each CVM level (LAOCVM1, LAOCVM2, LAOCVM3, LAOCVM4) from lot 006 were tested.
- Real-time Stability: Samples were tested in duplicate at Day 0, 15, 27, and 38 months (Table 2). The specific number of individual vials/runs for each time point beyond "duplicate (as a minimum)" isn't detailed.
- Open Component Stability: CVM lot 006 was tested at 2-hourly intervals for up to 9 hours. The number of samples per interval is not explicitly stated beyond "in duplicate (as a minimum)".
- Value Assignment Validation: 28 samples (10 normal female serum samples and 18 spiked samples) and four levels of commercially available controls were used.
- Expected Values/Target Values/Reference Range: Each CVM level was tested for a total of 27 replicates (9 runs and 3 replicates per run) using 3 different reagent kit lots and 8 IMMULITE 2000 systems.
- Data Provenance: Not explicitly stated as "country of origin," but "internal material" and "gravimetrically prepared" suggest internal laboratory data. It is prospective testing designed to characterize the device's performance.
3. Number of Experts and Qualifications for Ground Truth
- Not Applicable. This device is an in vitro diagnostic reagent (calibration verification material) for laboratory instruments, not an AI or imaging device requiring expert interpretation for ground truth. Ground truth is established through metrological traceability and standard reference methods (gravimetric preparation, assigned reference calibrators).
4. Adjudication Method for Test Set
- Not Applicable. See point 3.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Not Applicable. See point 3.
6. Standalone Performance (Algorithm Only)
- Yes, in essence. The performance testing described (stability, value assignment) evaluates the intrinsic performance of the CVM as a standalone device (material), independent of human interpretation. Its function is to verify instrument calibration, not to provide diagnostic interpretations.
7. Type of Ground Truth Used
- Metrological Traceability: The CVMs are traceable to an internal material that has been gravimetrically prepared. They are also value assigned using assigned reference calibrators, which themselves would have established traceability.
- Reference Calibrators: The CVMs are described as a subset of 7-level Androstenedione calibrators, and their dose values are generated using curves from these assigned reference calibrators. These calibrators serve as the "ground truth" reference for verifying the CVM values.
8. Sample Size for Training Set
- Not Applicable in the conventional sense of machine learning. This is not an AI/ML device. The "training" here refers to the process of establishing the reference values and behavior of the CVM and the associated assay. The value assignment process describes the materials and procedures used to characterize the CVMs, which inherently involves multiple measurements and statistical analysis rather than a distinct "training set" like in AI.
9. How Ground Truth for Training Set Was Established
- Not Applicable in the AI sense. For this device, "ground truth" (or more appropriately, reference values) for the CVMs are established through:
- Gravimetric Preparation: The internal material to which the CVMs are traceable is gravimetrically prepared, a fundamental method for establishing concentration.
- Assigned Reference Calibrators: The CVMs' dose values are derived from curves generated by assigned reference calibrators. These reference calibrators represent the established "ground truth" for the assay.
- Validated Materials and Procedures: The CVMs are manufactured using qualified materials and measurement procedures, ensuring their accuracy relative to the established reference values.
Device 2: IMMULITE® 2000 Troponin I Calibration Verification Material (CVM) (K143636)
1. Table of Acceptance Criteria and Reported Device Performance
Similar to the Androstenedione CVM, the document specifies stability acceptance criteria and indicates successful performance without providing detailed numerical results for each time point beyond the initial dose.
| CVM Level | Dose at Time 0 (ng/mL) | Guideline Criteria % difference to Time 0 | Acceptable Dose Range (ng/mL) | Reported Performance (from stability study, implicitly deemed compliant) |
|---|---|---|---|---|
| LTICVM1 | 0.00 | N/A | ≤0.20 | For LTICVM1, the dose at time 0 is 0.00 ng/mL, and the acceptable range is ≤ 0.20 ng/mL. The stability study supports the claimed stability, indicating results fell within this range. |
| LTICVM2 | 0.56 | ±10% | 0.50 - 0.62 | For LTICVM2, the dose at time 0 is 0.56 ng/mL. The acceptable range is 0.50 - 0.62 ng/mL. The stability study supports the claimed stability, indicating results fell within this range. |
| LTICVM3 | 13.8 | ±10% | 12.4 - 15.2 | For LTICVM3, the dose at time 0 is 13.8 ng/mL. The acceptable range is 12.4 - 15.2 ng/mL. The stability study supports the claimed stability, indicating results fell within this range. |
| LTICVM4 | 159 | ±10% | 143 - 175 | For LTICVM4, the dose at time 0 is 159 ng/mL. The acceptable range is 143 - 175 ng/mL. The stability study supports the claimed stability, indicating results fell within this range. |
2. Sample Size for Test Set and Data Provenance
- Test Set Description:
- Stability Studies: Multiple samples of each CVM level (LTICVM1, LTICVM2, LTICVM3, LTICVM4) from lot 010 were tested.
- Real-time Stability: Samples were run in duplicate (as a minimum) at Day 0, 3, 9, and 10 months (Table 2). Specific counts beyond "duplicate" are not specified.
- Open Component Stability: CVM lot 090 was tested at 2-hourly intervals for up to 9 hours. Specific counts not detailed.
- Value Assignment Validation: Three levels of commercially available controls and 25 spiked samples were used.
- Expected Values/Target Values/Reference Range: Tested on 15 replicates in total (5 runs and 3 replicates per run) using 4 IMMULITE 2000 systems and 2 different reagent kit lots.
- Data Provenance: Not explicitly stated as "country of origin," but "internal standard" and "gravimetrically prepared" suggest internal laboratory data. It is prospective testing.
3. Number of Experts and Qualifications for Ground Truth
- Not Applicable. See point 3 for Androstenedione CVM.
4. Adjudication Method for Test Set
- Not Applicable. See point 3 for Androstenedione CVM.
5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study
- Not Applicable. See point 3 for Androstenedione CVM.
6. Standalone Performance (Algorithm Only)
- Yes, in essence. Similar to the Androstenedione CVM, the performance testing described evaluates the intrinsic performance of the Troponin I CVM as a standalone material.
7. Type of Ground Truth Used
- Metrological Traceability: The CVMs are traceable to an internal standard that has been gravimetrically prepared.
- Reference Calibrators: The CVMs are described as a subset of 10-level Troponin I calibrators, and their dose values are generated using curves from these assigned reference calibrators, which serve as the "ground truth" reference.
8. Sample Size for Training Set
- Not Applicable in the conventional sense of machine learning. See point 8 for Androstenedione CVM.
9. How Ground Truth for Training Set Was Established
- Not Applicable in the AI sense. For this device, "ground truth" (reference values) are established through:
- Gravimetric Preparation: The internal standard to which the CVMs are traceable is gravimetrically prepared.
- Assigned Reference Calibrators: The CVMs' dose values are derived from curves generated by assigned reference calibrators, which represent the established "ground truth" for the assay.
- Validated Materials and Procedures: The CVMs are manufactured using qualified materials and measurement procedures, ensuring their accuracy relative to the established reference values.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of a stylized caduceus, which is a symbol of medicine and healthcare. The caduceus is composed of three intertwined lines that form the shape of a bird or a wing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
SIEMENS HEALTHCARE DIAGNOSTICS, INC. DONNA VELASQUEZ 5210 PACIFIC CONCOURSE DRIVE LOS ANGELES CA 90045
February 25, 2015
Re: K143636 Trade/Device Name: IMMULITE® 2000 Androstenedione Calibration Verification Material, IMMULITE® 2000 Troponin I Calibration Verification Material Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: I. Reserved Product Code: JJX Dated: December 19, 2014 Received: December 22, 2014
Dear Ms. Donna Velasquez:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Katherine Serrano -A
For : Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Indications for Use
510(k) Number (if known) K143636
Device Name
IMMULITE® 2000 Androstenedione Calibration Verification Material
Indications for Use (Describe)
IMMULITE® 2000 Androstenedione Calibration Verification Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Androstenedione assay on the IMMULITE 2000 systems
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Indications for Use
510(k) Number (if known) K143636
Device Name
IMMULITE® 2000 Troponin I Calibration Verification Material
Indications for Use (Describe)
IMMULITE® 2000 Troponin I Calibration Material (CVM) is for in vitro diagnostic use in the verification of calibration of the IMMULITE Troponin I assay on the IMMULITE 2000 systems
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: K143636
| 1. | Submitter | |
|---|---|---|
| Mailing Address: | Siemens Healthcare Diagnostics Inc.5210 Pacific Concourse DriveLos Angeles, CA 90045 | |
| Contact Person: | Donna VelasquezRegulatory Technical Specialist | |
| Phone Number: | (310)-645-8200 x7403 | |
| Fax Number: | (310)-645-9999 | |
| E-mail Address: | donna.velasquez@siemens.com | |
| Date Prepared: | January 22, 2015 | |
| 2. Device Name | ||
| Proprietary Name: | IMMULITE® 2000 Androstenedione Calibration VerificationMaterial | |
| Measurand: | Quality Control materials for IMMULITE® 2000 Androstenedioneassay | |
| Type of Test: | Calibration Verification Material (CVM) for IMMULITE® 2000Androstenedione assay | |
| Regulation Section: | 21 CFR 862.1660, Quality Control Material | |
| Classification: | Class I Reserved | |
| Products Code: | JJX - Single (Specified) Analyte Controls (Assayed andUnassayed) | |
| Panel: | Clinical Chemistry (75) | |
| 3. Predicate Device Name | IMMULITE® 2000 Cortisol Calibration Verification Material(CVM) | |
| Predicate 510(k) No: | K141444 | |
| 4. Device Description: | The IMMULITE® 2000 Androstenedione Calibration VerificationMaterial (CVM) contains one set of four vials each 2mL in liquidform. CVM1 contains processed human serum with preservatives.CVM2, CVM3 and CVM4 contain androstenedione in processedhuman serum matrix with preservative. | |
| 5. Intended Use:Indication for Use: | See Indications for Use Statement belowThe IMMULITE® 2000 Androstenedione CalibrationVerification Material (CVM) is for in vitro diagnostic use in theverification of calibration of the IMMULITE Androstenedioneassay on the IMMULITE 2000 systems | |
| Special Conditions for | ||
| Use Statement(s): | For prescription use only |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The letters are evenly spaced and the word is horizontally oriented. The image appears to be a logo or brand name.
Special Instrument Requirements:
IMMULITE® 2000 Systems
6. Technological Characteristics and Substantial Equivalence Comparison with Predicate:
A comparison of the device features, intended use, and other information demonstrates that the IMMULITE® 2000 Androstenedione Calibration Verification Material (CVM) is substantially equivalent to the predicate device as summarized in Table 1.
| SIMILARITIES | ||
|---|---|---|
| Candidate DeviceIMMULITE 2000 AndrostenedioneCVM | Predicate DeviceIMMULITE 2000 Cortisol CVM | |
| IntendedUse | The IMMULITE® 2000Androstenedione CalibrationVerification Material(CVM) is for in vitro diagnostic usein the verification of calibration of theIMMULITE Androstenedione assayon the IMMUL-DPC2000 systems. | The IMMULITE® CortisolCalibration Verification Material(CVM) is for in vitro diagnostic use inthe verification of calibration of theIMMULITE Cortisol assay on theIMMULITE 2000 systems |
| Storage | Same | |
| Form | Liquid | Same |
| Stability | Stable unopened until theexpiration date | Same |
| Levels | 4 | Same |
| Matrix | Human Serum with preservatives | Same |
| Use | Single Use Only | Same |
| DIFFERENCES | ||
| Candidate DeviceIMMULITE 2000 AndrostenedioneCVM | Predicate DeviceIMMULITE 2000 Cortisol CVM | |
| Analyte | Androstenedione | Cortisol |
Table 1: Substantial Equivalence Comparison
7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance Specifications for its intend use. The following tests were performed on the candidate device.
Stability Summary:
The stability study was conducted to validate real-time shelf life and open component (in-use or open vial) claim for the IMMULITE 2000 Androstenedione Calibration Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 3 years when stored at -20℃ prior to opening.
{6}------------------------------------------------
Open Component stability studies presents results that support 8 hours of stability at ambient or room temperature (15-25°C).
Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. The stability calibrators/CVMs are run in duplicate (as a minimum) at the time points shown in Table 2, and the dose value is determined from the reference calibrator curve.
| CVM Level | Time-Points (Months) | |||
|---|---|---|---|---|
| LAOCVM1 | Day 0 | 15 | 27 | 38 |
| LAOCVM2 | Day 0 | 15 | 27 | 38 |
| LAOCVM3 | Day 0 | 15 | 27 | 38 |
| LAOCVM4 | Day 0 | 15 | 27 | 38 |
Table 2: Stability Protocol Summary - LAOCVM Lot 006
For Open Component testing, the results were determined from a 2-point adjustment. Using IMMULITE 2000 Androstenedione (L2KAO) kit lot 336, CVM lot 006 was tested at 2-hourly intervals for up to 9 hours at ambient or room temperature (15-25°C) conditions.
Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE Androstenedione Calibration Verification Material (CVM) real time and open component stability testing is applied to dose compared to dose at time point zero. The result is required to fall between ±10% for CVM level 2 and 3 and ±13% for CVM level 4 as shown in Table 3.
| Table 3: Acceptance criteria for stability of IMMULITE 2000 Androstenedione CVM Lot 006 |
|---|
| CVM Level | Dose attime 0(ng/mL) | GuidelineCriteria% differenceto time 0 | Acceptabledose range(ng/mL) |
|---|---|---|---|
| LAOCVM1 | 0.00 | N/A | ≤ 0.3 |
| LAOCVM2 | 1.93 | ±10% | 1.74 - 2.12 |
| LAOCVM3 | 6.68 | ±10% | 6.01 - 7.35 |
| LAOCVM4 | 15.4 | ±13% | 13.4 - 17.4 |
Traceability:
The IMMULITE Androstenedione CVMs are traceable to an internal material which has been gravimetrically prepared. The CVMs are manufactured using qualified materials and measurement procedures.
{7}------------------------------------------------
Value Assignment:
The IMMULITE Androstenedione CVMs are 4 level materials which are a subset of 7 level Androstenedione calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Androstenedione reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned reference calibrators. The CVMs are manufactured using qualified materials and measurement procedures. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each instrument independently. CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples, spiked patient samples and controls using the assigned CVM values. Four levels of commercially available controls and 28 samples (10 normal female serum samples and 18 spiked samples) were used to validate calibrator/CVM value assignments.
Expected Values/Target Values/Reference Range:
Each CVM level was tested for a total of 27 replicates; 9 runs and 3 replicates per run, 3 different reagent kit lots and 8 IMMULITE 2000 systems. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The target values are provided in the IMMULITE® 2000 CVM Calibration Material lotspecific value card. The expected assay range is 0.3 to 10 ng/mL. The target values in Table 4 can be considered as guidelines.
| Analyte targetlevels | CVM Level | TargetMean(ng/mL) | StandardDeviation(SD) | Guideline ±2SD Range(ng/mL) | ||
|---|---|---|---|---|---|---|
| LAOCVM1 | 0.00 | - | 0.00 ≤0.30 | |||
| LAOCVM2 | 1.31 | 0.10 | 1.11 1.51 | |||
| LAOCVM3 | 5.00 | 0.25 | 4.50 5.50 | |||
| LAOCVM4 | 10.0 | 0.65 | 8.70 11.3 | |||
| Assay Range | 0.3 to 10 ng/mL |
| Table 4: Analyte Target Range Values | ้ | ี | ||
|---|---|---|---|---|
| -- | --------------------------------------- | --- | --- | -- |
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
- CEN 13640 Stability Testing of In Vitro Diagnostic Reagents
- Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators ●
- Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control Material ●
Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "SIEMENS" in a sans-serif font. The letters are all capitalized and are a teal color. The word is displayed horizontally and is centered in the image.
Conclusion: 8.
The IMMULITE® 2000 Androstenedione Calibration Verification Material is substantially equivalent to the predicate device intended for similar use. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 Androstenedione Calibration Verification Material does not raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
{9}------------------------------------------------
510(k) Summary
Introduction: According to the requirements of 21 CFR 807.92, the following information provides sufficient details to understand the basis for determination of substantial equivalence.
The assigned 510(k) Number: K143636
| 1. SubmitterMailing Address: | Siemens Healthcare Diagnostics Inc.5210 Pacific Concourse DriveLos Angeles, CA 90045 |
|---|---|
| Contact Person:Phone Number:Fax Number:E-mail Address:Date Prepared: | Donna VelasquezRegulatory Technical Specialist(310) 645-8200 x7403(310) 645-9999donna.velasquez@siemens.comJanuary 22, 2015 |
| 2. Device NameProprietary Name:Measurand:Type of Test:Regulation Section:Classification:Products Code:Panel: | IMMULITE® 2000 Troponin I Calibration Verification MaterialQuality Control materials for IMMULITE® 2000 Troponin I assayCalibration Verification Material (CVM) for IMMULITE® 2000Troponin I assay21 CFR 862.1660, Quality Control MaterialClass I ReservedJJX - Single (Specified) Analyte Controls (Assayed andUnassayed)Clinical Chemistry (75) |
| 3. Predicate Device NamePredicate 510(k) No: | IMMULITE 2000 Prolactin Calibration Verification Material(CVM)K140818 |
| 4. Device Description: | The IMMULITE® 2000 Troponin I Calibration VerificationMaterial (CVM) contains one set of four vials, 2mL each afterreconstitution. CVM1 contains equine serum with 0.88% sodiumazide and preservative. CVM2, CVM3 and CVM4 contain humanTroponin I and rabbit Troponin C in equine serum matrix with0.88% sodium azide and preservative. |
| 5. Intended Use:Indication for Use: | See Indications for Use Statement belowThe IMMULITE® 2000 Troponin I Calibration VerificationMaterial (CVM) is for in vitro diagnostic use in the verificationof calibration of the IMMULITE Troponin I assay on theIMMULITE 2000 systems |
| Special Conditions forUse Statement(s): | For prescription use only |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a bright teal. The letters are evenly spaced and the word is horizontally oriented.
Special Instrument Requirements:
IMMULITE® 2000 Systems
-
- Technological Characteristics and Substantial Equivalence Comparison with Predicate:
A comparison of the device features, intended use, and other information demonstrates that the IMMULITE® 2000 Troponin I Calibration Verification Material (CVM) is substantially equivalent to the predicate device as summarized in Table 1.
- Technological Characteristics and Substantial Equivalence Comparison with Predicate:
| SIMILARITIES | ||
|---|---|---|
| Candidate Device | Predicate Device | |
| IMMULITE 2000 Troponin I CVM | IMMULITE 2000 Prolactin CVM | |
| IntendedUse | The IMMULITE® 2000 Troponin ICalibration Verification Material (CVM)is for in vitro diagnostic use in theverification of calibration of theIMMULITE Troponin I assay on theIMMULITE 2000 systems | The IMMULITE® ProlactinCalibration Verification Material(CVM) is for in vitro diagnostic usein the verification of calibration ofthe IMMULITE Prolactin assay onthe IMMULITE 2000 systems |
| Form | Lyophilized | Same |
| Matrix | Equine serum with preservatives | Same |
| Levels | 4 | Same |
| Stability | Stable unopened until the expiration date | Same |
| Use | Single Use Only | Same |
| DIFFERENCES | ||
|---|---|---|
| Candidate DeviceIMMULITE 2000 Troponin I CVM | Predicate DeviceIMMULITE 2000 Prolactin CVM | |
| Storage | ≤-20°C | 2-8°C |
| Analyte | Troponin I | Prolactin |
7. Non-Clinical Performance Testing
Performance testing has been carried out to demonstrate that this device meets the performance specifications for its intend use. The following tests were performed on the candidate device.
Stability Summary:
The stability study was conducted to validate real-time shelf life and open component (in-use or open vial) claim for the IMMULITE 2000 Troponin I Calibration Material (CVM) to ensure that it maintains optimal product performance on IMMULITE 2000 platforms throughout the established shelf life of the CVM. The Calibration Verification Materials are stable up to 10 months when stored at -20°C prior to opening, and for 2 hours at ambient or room temperature (15-25°C) after reconstitution.
{11}------------------------------------------------
Stability Protocol Summary:
The CVM study protocols are run as part of the calibrator stability testing. The stability CVMs and reference CVMs are run in duplicate (as a minimum) at the time points shown in Table 2 and the dose value determined from the reference calibrator curve.
| CVM Level | Time-Points (Months) | |||
|---|---|---|---|---|
| LTICVM1 | Day 0 | 3 | 9 | 10 |
| LTICVM2 | Day 0 | 3 | 9 | 10 |
| LTICVM3 | Day 0 | 3 | 9 | 10 |
| LTICVM4 | Day 0 | 3 | 9 | 10 |
| Table 2: Stability Protocol Summary - LTICVM Lot 010 | |
|---|---|
For Open Component testing, the results were determined from a 2-point adjustment. Using IMMULITE 2000 Troponin I (L2KT12) kit lot 270, CVM lot 090 was tested at 2-hourly intervals for up to 9 hours at ambient or room temperature (15-25°C) conditions.
Stability Acceptance Criteria Summary:
The Acceptance Criteria for the IMMULITE 2000 Troponin I CVM for real time stability and Open-Component testing is applied to dose compared to dose at time point zero. The result is required to fall between ±10% difference for CVM levels 2, 3 and 4 as shown in Table 3.
Table 3: Acceptance criteria for stability of IMMULITE 2000 Troponin I CVM Lot 010
| CVM level | Dose atTime 0(ng/mL) | GuidelineCriteria% difference toTime 0 | Acceptabledose range(ng/mL) |
|---|---|---|---|
| LTICVM1 | 0.00 | N/A | ≤0.20 |
| LTICVM2 | 0.56 | ±10% | 0.50 - 0.62 |
| LTICVM3 | 13.8 | ±10% | 12.4 - 15.2 |
| LTICVM4 | 159 | ±10% | 143 - 175 |
Traceability:
The IMMULITE Troponin I CVMs are traceable to an internal standard which has been gravimetrically prepared. The CVMs are manufactured using qualified materials and measurement procedures.
Value Assignment:
The IMMULITE Troponin I CVMs are 4 level materials which are a subset of 10 level Troponin I calibrators. Calibrators are not commercialized but are used internally during manufacture and release testing of Troponin I reagents and two point adjustors. The IMMULITE calibrators and therefore CVMs are value assigned using assigned reference calibrators. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. The CVM values are then averaged
{12}------------------------------------------------
across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The controls must fall within their target ranges. Three levels of commercially available controls and 25 spiked samples were used to validate calibrator/CVM value assignments.
Expected Values/Target Values/Reference Range:
The CVMs are manufactured using qualified materials and measurement procedures. The Troponin I CVMs were tested on 15 replicates in total comprised of 5 runs and 3 replicates per run on 4 IMMULITE 2000 systems and 2 different reagent kit lots. The CVMs dose values are generated using curve generated by assigned reference calibrators. The CVM values are calculated based on the recovered values for each run on each instrument independently. The CVM values are then averaged across all systems. Quality control is performed by calculating the recovery of patient samples and controls using the assigned CVM values. The Guideline Range (95% confidence interval) for each CVM level was established based on the Target Mean and ± 2 Standard Deviation (SD). The expected assay range is up to 180 ng/mL. The target values in Table 4 can be considered as guidelines.
| Analyte targetlevels | CVM Level | TargetMean(ng/mL) | StandardDeviation(SD) | Guideline ±2SD Range(ng/mL) | |
|---|---|---|---|---|---|
| LTICVM1 | 0.00 | - | 0.00 $\leq$ 0.20 | ||
| LTICVM2 | 0.575 | 0.092 | 0.391 | 0.759 | |
| LTICVM3 | 14.2 | 0.90 | 12.4 | 16.0 | |
| LTICVM4 | 159 | 15.0 | 129 | 189 | |
| Assay Range | Up to 180 ng/mL |
Table 4: Analyte Target Range Levels
Each laboratory should establish their limits for acceptability based on methodology, clinical significance and medical decision levels of the test analyte. The representative, total precision tabulated in the respective assay instructions for use may be considered as one factor when establishing local, acceptable ranges. The values provided above may be considered as guidelines. Value assignment is lot specific.
Standard/Guidance Documents Referenced:
- CEN 13640 Stability Testing of In Vitro Diagnostic Reagents ●
- Guidance for Industry Abbreviated 510(k) Submissions for In Vitro Diagnostic Calibrators ●
- Guidance for Industry and FDA Staff Assayed and Unassayed Quality Control Material ●
Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10
8. Conclusion:
The IMMULITE® 2000 Troponin I Calibration Verification Material is substantially equivalent to the predicate device intended for similar use. The substantial equivalence of the device is supported by the non-clinical testing and was found to be comparable and supports the claims of substantial equivalence, product safety and effectiveness. Based on the testing completed and the comparisons with predicate device, the IMMULITE® 2000 Troponin I Calibration Verification Material does not
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a teal or cyan color. The text is horizontally oriented and appears to be a logo or brand name.
raise any new questions on safety and effectiveness and the results support a determination of substantial equivalence.
§ 862.1660 Quality control material (assayed and unassayed).
(a)
Identification. A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.(b)
Classification. Class I (general controls). Except when intended for use in donor screening tests, quality control materials (assayed and unassayed) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.