K Number
K143576

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-02-27

(72 days)

Product Code
Regulation Number
888.3060
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Exactech Ambassador™ Cervical Plate System is intended for anterior cervical fixation (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e.. scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Device Description

The Ambassador™ Cervical Plate System consists of a variety of bone plates and screws. Fixation is achieved by inserting bone screws through the openings in the plate into the vertebral bodies of the cervical spine. The Ambassador™ Cervical Plate System implant components are made from titanium alloy such as described by ASTM F136. Plates and screws are provided nonsterile with instructions for sterilization.

AI/ML Overview

I am sorry, but the provided text describes a medical device called "Exactech Ambassador™ Cervical Plate System" and its 510(k) premarket notification. This document focuses on demonstrating substantial equivalence to predicate devices based on technological characteristics and mechanical performance data, rather than acceptance criteria and a study proving device performance as typically understood for AI/ML-based medical devices.

Therefore, I cannot extract the information required in your request, such as a table of acceptance criteria and reported device performance (in the context of clinical metrics), sample sizes for test sets, expert qualifications, adjudication methods, MRMC studies, standalone performance, ground truth types, or training set details. This type of information is usually associated with studies evaluating the diagnostic or predictive performance of algorithms, which is not the subject of this document.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, stacked one behind the other. The faces are rendered in a simple, abstract style.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 27, 2015

Exactech, Incoporated % Kenneth C. Maxwell II Regulatory and Quality Specialist Empirical Consulting 4628 Northpark Drive Colorado Springs, Colorado 80918

Re: K143576

Trade/Device Name: Exactech Ambassador™ Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: December 16, 2014 Received: December 18, 2014

Dear Mr. Maxwell:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

Page 2 - Mr. Kenneth Maxwell

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143576

Device Name Exactech Ambassador™

Indications for Use (Describe)

The Exactech Ambassador™ Cervical Plate System is intenor cervical fixation (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e.. scoliosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D) ☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

5.510(K) SUMMARY

Submitter's Name:Exactech, Inc.
Submitter's Address:2320 NW 66th CourtGainesville, FL 32658
Submitter's Telephone:352.377.1140
Contact Person:Kenneth C. MaxwellEmpirical Testing Corp.719.291.6874
Date Summary was Prepared:20 February 2015
Trade or Proprietary Name:Exactech Ambassador™
Common or Usual Name:Anterior Cervical Plate System
Classification:Class II per 21 CFR §888.3060 Spinal Intervertebral BodyFixation Orthosis
Product Code:KWQ
Classification Panel:Orthopaedic and Rehabilitation Devices Panel

DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION:

The Ambassador™ Cervical Plate System consists of a variety of bone plates and screws. Fixation is achieved by inserting bone screws through the openings in the plate into the vertebral bodies of the cervical spine. The Ambassador™ Cervical Plate System implant components are made from titanium alloy such as described by ASTM F136. Plates and screws are provided nonsterile with instructions for sterilization.

INDICATIONS FOR USE

The Exactech Ambassador™ Cervical Plate System is intended for anterior cervical fixation (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion.

TECHNOLOGICAL CHARACTERISTICS

The Ambassador™ Cervical Plate System is made from titanium that conforms to F136. The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are the same between the subject and predicates:

  • Indications for Use
  • Materials of manufacture
  • Structural support mechanism .

{4}------------------------------------------------

Table 5-1 Predicate Devices

510kNumberTrade or Proprietary or Model NameManufacturerPredicateType
K071667VECTRA ACPSynthesPrimary
K042798C-TEK™ C-Thru Anterior Cervical PlateSystemInterporeAdditional
K070891Classic ACP SystemAltiva®Additional

Performance Data

The Ambassador™ Cervical Plate System has been tested in the following test modes:

  • Static axial compression bending per modified ASTM F1717-13 .
  • . Static torsion per modified ASTM F1717-13
  • Dynamic axial compression bending fatigue per modified ASTM F1717-13 ●

The results of this non-clinical testing show that the strength of the Ambassador™ Cervical Plate System is sufficient for its intended use and is substantially equivalent to legally marketed predicate devices.

CONCLUSION

The overall technology characteristics and mechanical performance data lead to the conclusion that the Ambassador™ Cervical Plate System is substantially equivalent to the predicate device.

N/A