The Sensititre HP MIC Susceptibility plate with Oritavancin (0.0005-8mcg/ml)
K143526 · Thermo Fisher Scientific · JWY · Mar 5, 2015 · Microbiology
Device Facts
| Record ID | K143526 |
| Device Name | The Sensititre HP MIC Susceptibility plate with Oritavancin (0.0005-8mcg/ml) |
| Applicant | Thermo Fisher Scientific |
| Product Code | JWY · Microbiology |
| Decision Date | Mar 5, 2015 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 866.1640 |
| Device Class | Class 2 |
Intended Use
The Sensititre HP MIC Susceptibility plate is an in vitro diagnostic product for clinical susceptibility testing of fastidious isolates. This 510(k) is for the newly approved Oritavancin in the dilution range of 0.0005-8ug/ml to the Sensitire HP MIC Susceptibility plate for testing Streptococcus spp. The approved primary "Indications for Use" and clinical significance for Streptococus spp. is for the following species: Streptococcus pyogenes Streptococcus agalactiae Streptococcus dysgalactia Streptococcus anginosus
Device Story
Sensititre HP MIC Susceptibility plate is an in vitro diagnostic device used in clinical laboratories for antimicrobial susceptibility testing. It utilizes microbroth dilution to determine the Minimum Inhibitory Concentration (MIC) of Oritavancin against fastidious Streptococcus species. The device consists of plates containing specific concentrations of the antimicrobial agent. Laboratory technicians inoculate the plates with patient isolates; after incubation, growth is observed to determine the MIC. The resulting MIC value assists clinicians in selecting appropriate antibiotic therapy for patients with bacterial infections. The device provides standardized, quantitative susceptibility data to guide clinical decision-making and improve patient outcomes.
Clinical Evidence
Bench testing only. The device performance was evaluated by comparing MIC results obtained using the Sensititre HP MIC Susceptibility plate to a reference broth microdilution method. Testing included clinical isolates of Streptococcus pyogenes, Streptococcus agalactiae, Streptococcus dysgalactiae, and Streptococcus anginosus to demonstrate essential agreement and categorical agreement between the subject device and the reference method.
Technological Characteristics
Microbroth dilution susceptibility plate. Contains dehydrated Oritavancin in a dilution range of 0.0005-8ug/mL. Designed for in vitro diagnostic use in clinical laboratory settings. Standardized plate format for automated or manual inoculation and reading.
Indications for Use
Indicated for clinical susceptibility testing of fastidious Streptococcus spp. (S. pyogenes, S. agalactiae, S. dysgalactiae, S. anginosus) to Oritavancin. For prescription use only.
Regulatory Classification
Identification
An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.
Related Devices
- K143381 — Sensititre Susceptibility Plates · Thermo Fisher Scientific · Feb 19, 2015
- K141651 — SENSITITRE SUSCEPTIBILITY PLATES · Thermo Fisher Scientific · Oct 21, 2014
- K062681 — SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE (HP) MIC PLATE · Trek Diagnostic Systems, Ltd. · Oct 13, 2006
- K062722 — SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE (HP) MIC PLATE · Trek Diagnostic Systems, Inc. · Oct 20, 2006
- K140985 — SENSITITRE HAEMOPHILUS/STREPTOCOCCUS PNEUMONIAE(HP) SUSCEPTIBILITY PLATES · Thermo Fisher Scientific · Jun 10, 2014
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
THERMO FISHER SCIENTFIC CYNTHIA KNAPP DIRECTOR OF CLINICAL OPERATIONS 1 THERMO FISHER WAY OAKWOOD VILLAGE OH 44014-6
March 5, 2015
Re: K143526
Trade/Device Name: The Sensititre HP MIC Susceptibility plate with Oritavancin in the dilution range of 0.0005-8ug/mL
Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial Susceptibility Test Powder Regulatory Class: II Product Code: JWY, LRG, LTT, LTW Dated: December 10, 2014 Received: December 12, 2014
Dear Ms. Knapp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Uwe Scherf –S for
Sally Hojvat, M.Sc., PhD. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) k143526
Device Name
The Sensititre HP MIC Susceptibility plate with Oritavancin in the dilution range of 0.0005-8ug/ml
Indications for Use (Describe)
The Sensititre HP MIC Susceptibility plate is an in vitro diagnostic product for clinical susceptibility testing of fastidious isolates.
This 510(k) is for the newly approved Oritavancin in the dilution range of 0.0005-8ug/ml to the Sensitire HP MIC Susceptibility plate for testing Streptococcus spp.
The approved primary "Indications for Use" and clinical significance for Streptococus spp. is for the following species: Streptococcus pyogenes Streptococcus agalactiae Streptococcus dysgalactia Streptococcus anginosus
Type of Use (Select one or both, as applicable)
X | Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."