Sensititre Susceptibility Plates

K143381 · Thermo Fisher Scientific · JWY · Feb 19, 2015 · Microbiology

Device Facts

Record IDK143381
Device NameSensititre Susceptibility Plates
ApplicantThermo Fisher Scientific
Product CodeJWY · Microbiology
Decision DateFeb 19, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Intended Use

The Sensititre HP MIC Susceptibility plate is an in vitro diagnostic product for clinical susceptibility testing of fastidious isolates. This 510(k) is for the addition of newly approved Dalbavancin for the dilution range of 0.0005-2ug/ml to the Sensititre HP MIC Susceptibility plate for testing Streptococcus spp. The approved primary "Indications for Use" and clinical significance for Streptococus spp. is for the following species: Streptococcus pyogenes Streptococcus agalactiae Streptococcus anginosus

Device Story

Sensititre HP MIC Susceptibility plate is an in vitro diagnostic tool for determining antimicrobial susceptibility of fastidious bacteria; specifically Streptococcus species. Device utilizes microbroth dilution method to determine Minimum Inhibitory Concentrations (MIC) of Dalbavancin. Used in clinical microbiology laboratories by trained technicians. Plate contains pre-dispensed, dried antimicrobial agent (Dalbavancin) at specific concentrations. Bacterial isolates are inoculated into the plate; incubated; and growth is observed. MIC results assist clinicians in selecting appropriate antibiotic therapy for patients with Streptococcus infections. Benefits include standardized susceptibility testing to guide effective treatment.

Clinical Evidence

No clinical data provided in the document. The submission relies on the established performance of the Sensititre MIC plate platform and the addition of a new antimicrobial agent (Dalbavancin) for testing Streptococcus species.

Technological Characteristics

Microbroth dilution susceptibility test plate. Contains dried antimicrobial agent (Dalbavancin) in concentrations ranging from 0.0005-2ug/ml. Designed for use with fastidious isolates (Streptococcus spp.).

Indications for Use

Indicated for in vitro clinical susceptibility testing of fastidious isolates, specifically Streptococcus pyogenes, Streptococcus agalactiae, and Streptococcus anginosus, using Dalbavancin at concentrations of 0.0005-2ug/ml.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around it. Inside the circle is an emblem featuring a stylized image of three human profiles facing right, with a design that resembles a caduceus or flowing fabric beneath them. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 THERMO FISHER SCIENTIFIC CYNTHIA KNAPP DIRECTOR OF CLINICAL OPERATIONS 1 THERMO FISHER WAY OAKWOOD VILLAGE OH 44014-6 February 19, 2015 Re: K143381 Trade/Device Name: Sensititre Haemophilus/Streptococcus pneumoniae (HP) MIC plates with Dalbavancin (0.0005- 2 ug/ml) Regulation Number: 21 CFR 866.1640 Regulation Name: Antimicrobial susceptibility test powder Regulatory Class: II Product Code: JWY, LRG, LTT, LTW Dated: November 24, 2014 Received: November 25, 2014 Dear Ms. Knapp: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {1}------------------------------------------------ If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Uwe Scherf -S for Sally Hojvat, M.Sc., PhD. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K143381 ## Device Name Sensititre Haemophilus / Streptococcus pneumoniae (HP) MIC plates with Dalbavancin at a concentration of 0.0005-2ug/ml Indications for Use (Describe) The Sensititre HP MIC Susceptibility plate is an in vitro diagnostic product for clinical susceptibility testing of fastidious isolates. This 510(k) is for the addition of newly approved Dalbavancin for the dilution range of 0.0005-2ug/ml to the Sensititre HP MIC Susceptibility plate for testing Streptococcus spp. The approved primary "Indications for Use" and clinical significance for Streptococus spp. is for the following species: Streptococcus pyogenes Streptococcus agalactiae Streptococcus anginosus | Type of Use (Select one or both, as applicable) | |-------------------------------------------------| |-------------------------------------------------| 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%