K Number
K143440

Validate with FDA (Live)

Device Name
Natus Quantum
Date Cleared
2015-04-20

(139 days)

Product Code
Regulation Number
882.1400
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks™/SleepWorks™ software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals. The amplifier is designed to facilitate functional mapping using a Digital Switch Matrix. The Digital Switch Matrix portion of the headbox is a combination of hardware relays and software controls allowing the user (physician or technologist) to switch electrode pairs between the EEG recording amplifier and the external cortical stimulator for stimulus delivery.

The Natus Quantum Amplifier is intended to be used by trained medical professionals, and is designed for use in clinical environments such as hospital rooms, epilepsy monitoring units, intensive care units, and operating rooms. It can be used with patients of all ages, but is not designed for fetal use.

Device Description

The Natus Quantum amplifier is comprised of a base unit and several breakout boxes. It is part of a system that is made up of a personal computer, a photic stimulator, an isolation transformer, video and audio equipment, networking equipment, and mechanical supports. The amplifier also contains an internal switch matrix to allow for a connection to an external cortical stimulator.

EEG and other physiological signals, from scalp electrodes, grid or needle electrodes, and other accessories such as pulse oximeters can be acquired by the Natus Quantum amplifier. These signals are digitized and transmitted to the personal computer running the Natus NeuroWorks software. The signals are displayed on the personal computer and can be recorded to the computer's local storage or to remote networked storage for later review.

AI/ML Overview

The provided text describes the Natus Quantum Amplifier, an electroencephalograph, and its regulatory submission (K143440). However, the document does not contain a study that directly proves the device meets specific acceptance criteria in terms of clinical performance metrics like sensitivity, specificity, or accuracy.

The document focuses on demonstrating substantial equivalence to predicate devices (EMU128S and NeuroLink IP 256) primarily through technical specifications and compliance with various safety, EMC, and quality standards. The "Performance Tests" section is very brief and refers to non-clinical verification testing rather than clinical efficacy studies.

Therefore, the following information is based on what is available or can be inferred from the provided text. Many requested fields will be marked as "Not Applicable" or "Not Provided" because the document does not describe the kind of clinical study you're asking about.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (from a clinical study perspective)Reported Device Performance (from the document)
Clinical performance metrics (e.g., sensitivity, specificity, accuracy in detecting electrophysiological signals)Not provided. The document focuses on technical specifications and functional verification.
Technical Specifications (Comparison to Predicate Devices):
EEG Channels64-256 (Subject Device, Predicate NeuroLink IP); 128 (Predicate EMU128S)
Reference ChannelsDedicated separate reference and ground (All devices)
Input Impedance>1000 MOhm (Subject Device); >100 MOhms (Predicate NeuroLink IP); >47 MOhms (Predicate EMU128S)
Input Noise< 1.5uV pk to pk @ .1-100Hz bandwidth (<0.53uV rms@1-100Hz bandwidth) (Subject Device); < 1uV rms @ 1-70Hz bandwidth (Predicate NeuroLink IP); < 2uV pk to pk @ full bandwidth (Predicate EMU128S)
Maximum Operational DC input voltage electrode offset±300V (All devices)
Input Bias Current<1nA (Subject Device, Predicate NeuroLink IP); < 20pA (Predicate EMU128S)
Common mode Rejection Ratio>110dB@60Hz (Subject Device, Predicate EMU128S); >40dB@60Hz (Predicate NeuroLink IP)
Sampling Frequency256, 512, 1024, 2048, 4096, 8192, 16384 Hz (Subject Device); 256, 512, 1024 Hz (Predicate NeuroLink IP); 256, 512, 1024, 2048 Hz (Predicate EMU128S)
Sampling Resolution - EEG channels24 bits (Subject Device); 16 bits (Predicate NeuroLink IP); 22 bits (Predicate EMU128S)
Sampling Quantization - EEG channels305nV (Subject Device); 179 nV (Predicate NeuroLink IP); 310 nV (Predicate EMU128S)
Storage Resolution - EEG Channels16 bits (All devices)
Functional / Design Verification Tests:
Signal Quality Verification TestPass
Functionality Verification TestPass

Note on Acceptance Criteria: The document implies that meeting the specified technical characteristics that are substantially equivalent or superior to the predicate devices, and passing internal design verification tests, are the "acceptance criteria" for regulatory clearance based on substantial equivalence. It does not provide clinical acceptance criteria.

2. Sample size used for the test set and the data provenance

  • Sample Size: Not Applicable. The document describes non-clinical verification testing of the device hardware/software, not a clinical study on patient data.
  • Data Provenance: Not Applicable. No patient data was used for the described performance tests.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • Number of Experts: Not Applicable. Ground truth for clinical data is not relevant to the described non-clinical verification tests.
  • Qualifications of Experts: Not Applicable.

4. Adjudication method for the test set

  • Adjudication Method: Not Applicable. No clinical test set requiring adjudication was described.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • MRMC Study: No. This document describes an EEG amplifier, not an AI-assisted diagnostic tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Standalone Performance: Not Applicable. This is a hardware device (EEG amplifier) with associated software for data acquisition, display, storage, and archiving. It is not an algorithm for standalone diagnostic performance.

7. The type of ground truth used

  • Type of Ground Truth: For the "Performance Tests" (Signal Quality Verification Test, Functionality Verification Test), the ground truth would be the design specifications and expected operational parameters of the device. These tests verify if the actual output matches the designed output. No clinical "ground truth" (e.g., pathology, outcomes data) for diagnosis is mentioned for these tests.

8. The sample size for the training set

  • Sample Size: Not Applicable. This is not an AI/machine learning device that requires a training set.

9. How the ground truth for the training set was established

  • Ground Truth Establishment: Not Applicable. (See point 8)

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around a central graphic. The graphic consists of three stylized human profiles facing to the right, with flowing lines that suggest movement or connection.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 20, 2015

Natus Medical Incorporated DBA Excel-Tech Ltd. (Xltek) Sanjay Mehta Quality and Regulatory Affairs Manager 2560 Bristol Circle Oakville, Ontario, Canada, L6H5S1

Re: K143440

Trade/Device Name: Natus Ouantum Amplifier Regulation Number: 21 CFR 882.1400 Regulation Name: Electroencephalograph Regulatory Class: Class II Product Code: GWO. OLV. GYC Dated: March 18, 2015 Received: March 19, 2015

Dear Mr. Mehta:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Isting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Felipe Aquel -S

Carlos Peña, PhD, MS for Division Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143440

Device Name Natus Quantum Amplifier

Indications for Use (Describe)

The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus New Yorks™ software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals. The amplifier is designed to facilitate functional mapping using a Digital Switch Matrix. The Digital Switch Matrix portion of the headbox is a combination of hardware relays and software controls allowing the user (physician or technologist) to switch electrode pairs between the EEG recording amplifier and the external cortical stimulator for stimulus delivery.

The Natus Quantum Amplifier is intended to be used by trained medical professionals, and is designed for use in clinical environments such as hospital rooms, epilepsy monitoring units, intensive care units, and operating rooms. It can be used with patients of all ages, but is not designed for fetal use.

Type of Use (Select one or both, as applicable)
-------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is written in blue, block letters, with a registered trademark symbol next to it. Below "natus" is the word "neurology" in smaller, italicized, black letters.

510K Summary

Date: April 15, 2015

Submitted by: Natus Medical Incorporated DBA Excel-Tech Ltd. (XLTEK) 2568 Bristol Circle Oakville, Ontario Canada L6H 5S1

Contact Person: Sanjay Mehta Senior Manager QA/RA Natus Medical Incorporated Tel .: (905) 829-5300 ext 388 Fax .: (905) 829-5304 E-mail: sanjay.mehta@natus.com

Proprietary Name: Natus Quantum

Common Name: Electroencephalograph

Regulation Number : 21CFR 882.1400

Classification Name: Full-montage standard electroencephalograph

Product code: GWQ, OLV

Device Class: II

Predicate Device: EMU128 (K040360); Neurolink IP 256 (K100683)

Description:

1. Overview: Natus Quantum Amplifier

The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks™ software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals. The amplifier is designed to facilitate functional mapping using a Digital Switch Matrix. The Digital Switch Matrix portion of the headbox is a combination of hardware relays and software controls allowing the user (physician or technologist) to switch electrode pairs between the EEG recording amplifier and the external cortical stimulator for stimulus delivery.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image contains the logo for "natus. neurology". The word "natus" is in a larger, bold, blue font on the top line. The word "neurology" is in a smaller, italicized, black font on the second line.

The Natus Quantum Amplifier is intended to be used by trained medical professionals. and is designed for use in clinical environments such as hospital rooms, epilepsy monitoring units, intensive care units, and operating rooms. It can be used with patients of all ages, but is not designed for fetal use.

2. Operating Principle of the Quantum Amplifier

The Natus Quantum amplifier is comprised of a base unit and several breakout boxes. It is part of a system that is made up of a personal computer, a photic stimulator, an isolation transformer, video and audio equipment, networking equipment, and mechanical supports. The amplifier also contains an internal switch matrix to allow for a connection to an external cortical stimulator.

EEG and other physiological signals, from scalp electrodes, grid or needle electrodes, and other accessories such as pulse oximeters can be acquired by the Natus Quantum amplifier. These signals are digitized and transmitted to the personal computer running the Natus NeuroWorks software. The signals are displayed on the personal computer and can be recorded to the computer's local storage or to remote networked storage for later review.

Image /page/4/Picture/6 description: The image shows three pieces of Natus Neurology equipment. On the left is a large, black, rectangular device with rounded corners and a small silver knob on the top. To the right are two smaller, square devices with gray tops and black sides. The Natus Neurology logo is visible on all three devices.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in blue, block letters, with a small registered trademark symbol after the "s". Below that, the word "neurology" is in a smaller, italicized, black font. The logo is simple and clean, with a focus on the company name.

Quantum features include

  • . 128 referential channels (cascade-able up to 256).
  • . 16 referential inputs per 128 can be configured as 8 differential inputs
  • 16 DC channels,
  • Pulse Oximeter including SpO2, Pulse Rate and Pleth signals .
  • . Digital Trigger Input
  • . Ability to initiate an impedance test, change the threshold, and view the results in the patient room
  • . Digital Switch Matrix
  • A small and lightweight wearable breakout box .
  • . TCP/IP and USB connectivity
  • . Patient-event switch interface on both the breakout box and base units
  • . Photic stimulator interface for EEG applications
  • . Modular pouch for belt or backpack harness

System Setup Overview

The Natus Quantum amplifier connects to the Natus NeuroWorks (K090019) & Natus SleepWorks (K090277) software for the acquisition storage, analysis, and review of Electroencephalographic & Polysomnographic data in conjunction with synchronized digital video. The system overview is as follows.

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image contains the logo for "natus neurology". The word "natus" is in a larger, blue font, while "neurology" is in a smaller, black font. The logo is simple and clean, with a focus on the company name.

Image /page/6/Figure/2 description: This image shows a diagram of a system for recording and digitizing a patient's physiological data. The diagram includes an acquisition PC connected to a network, which is connected to Natus Quantum Breakout Boxes. The breakout boxes are connected to electrodes and sensors placed on the patient's scalp, as shown in the "Patient Electrode Placement International 10-20 System (example)" diagram.

Device-patient interaction Accessories List:

The table below lists all accessories to the subject device. Accessories (1) to (9) enter in contact with the patient. These sensors guarantees acquisition of the physiological signals and passively transfer them to the head box. Characteristics of the sensors vary and are described (cleared) under their respective 510K submissions (see table).

DescriptionBody contact locationDevice connection
1Reusable gold disk electrode(K982053)Scalp (according to 10-20 & 10-10 system)Referential and differential inputs(labeled numerically)
2Single Use Intracranial Grids, Stripsand Depth Electrodes. (K082474)Intracranial recordingsReferential and differential inputs(labeled numerically)
3Body position sensor (K923033)ThoraxDC input
4Xactrace (K043132)Respiratory beltsDifferential input

{7}------------------------------------------------

Image /page/7/Picture/1 description: The image shows the logo for Natus Neurology. The word "natus" is written in blue, block letters. Below that, the word "neurology" is written in a smaller, italicized font. The logo is simple and professional.

Thorax/Abdomen
6Thermistor (K922112)Nasal/OralDC input
7Pulse Oximeter Sensor (K092101)FingerChannel labeled "oximeter/photic"
8Snoring microphone (K941759)Nasion/cheek/chin or side of the neckDC input
9Airflow Pressure Sensor (K922112)NasalDC input
10Photic Stimulator (K991903)NoneChannel labeled "oximeter/photic"
11Cortical Stimulator (K072964)Intracranial contactsAmplifier Digital Switch Matrix input port.

Indications for Use

The Natus Quantum Amplifier is intended to be used as an electroencephalograph: to acquire, display, store and archive electrophysiological signals. The amplifier should be used in conjunction with Natus NeuroWorks™/SleepWorks™ software to acquire scalp and intracranial electroencephalographic (EEG) signals as well as polysomnographic (PSG) signals. The amplifier is designed to facilitate functional mapping using a Digital Switch Matrix. The Digital Switch Matrix portion of the headbox is a combination of hardware relays and software controls allowing the user (physician or technologist) to switch electrode pairs between the EEG recording amplifier and the external cortical stimulator for stimulus delivery.

The Natus Quantum Amplifier is intended to be used by trained medical professionals, and is designed for use in clinical environments such as hospital rooms, epilepsy monitoring units, intensive care units, and operating rooms. It can be used with patients of all ages, but is not designed for fetal use.

SpecificationPredicate DeviceEMU128S (K040360)Predicate DeviceNeuroLink IP (K100683)Subject DeviceNatus Quantum
ManufacturerExcel-Tech Ltd.(Xltek)Excel-Tech Ltd.(Xltek)Excel-Tech Ltd.(Xltek)
Technological CharacteristicsAnalog Specifications EEG Channels
EEG Channels12864-25664-256
ReferenceChannelsDedicated separatereference and groundDedicated separate referenceand groundDedicated separate referenceand ground
Input Impedance>47 MOhms> 100 MOhms>1000 MOhm
Input Noise< 2uV pk to pk @ fullbandwidth< 1uV rms @ 1....70Hzbandwidth< 1.5uV pk to pk @ .1....100Hzbandwidth(<0.53uV rms@1....100Hzbandwidth)

Comparison to Predicate Device

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in blue, block letters, and the word "neurology" is in black, italicized letters. The logo is simple and modern.

MaximumOperational DCinput voltageelectrode offset±300V±300V±300V
Input Bias Current< 20pA< 1 nA<1nA
Common modeRejection Ratio>110dB@60Hz>40dB@60Hz>110dB@60Hz
Digital Specifications
SamplingFrequency256, 512, 1024, 2048256, 512, 1024 Hz256, 512, 1024, 2048, 4096,8192, 16384 Hz
SamplingResolution - EEGchannels22 bits16 bits24 bits
SamplingQuantization -EEG channels310 nV179 nV305nV
Storage Resolution- EEG Channels16 bits16 bits16 bits

For a detailed discussion covering any differences between devices and why they do not raised any new safety or effectiveness issues, the reviewer is referred to VOL_007_Substantial Equivalence discussion.

Brief Summary of Performance Tests

Non-clinical:

Testing of the Natus Quantum was performed in compliance with Natus Corporation design control process. The validation was carried out as part of verification. Testing included:

TestResults
Signal Quality Verification TestPass
Functionality Verification TestPass

It was found that the Natus Quantum system meets the design specification and performs as specified.

The device is in compliance with the following industrial standards

{9}------------------------------------------------

Safety Standard of Compliance and normative references

StandardsTitle
CAN /CSA-C22.2 No. 60601-1: 08(R2013) +C2:2011Medical electrical equipment - Part 1: Generalrequirements for basic safety and essentialperformance
ANSI/AAMI ES60601-1:2005/(R)2012 +C1:2009/(R)2012 and A2:2010/(R)2012
IEC 60601-1:2005 + C1:2006 and C2:2007,Third Edition
CENELEC EN 60601-1:2006 + A1:2013
IEC 60601-1-6:2010, Edition 3.0Medical electrical equipment - Part 1-6:General requirements for basic safety andessential performance - Collateral standard:Usability
IEC 62366:2007, Edition 1.0Medical devices - Application of usabilityengineering to medical devices
IEC 60601-2-26:2012, Edition 3CENELEC EN 60601-2-26L2003, Edition 2Medical electrical equipment - Part 2-26:Particular requirements for the safety ofelectroencephalographs
EN ISO 80601-2-61:2011, Edition 1Medical electrical equipment - Part 2-61:Particular requirements for basic safety andessential performance of pulse oximeterequipment

EMC Standard of Compliance and normative references

StandardsTitle
IEC 60601-1-2:2007, Edition 3.0Medical electrical equipment - Part 1-2:General requirements for safety - collateralstandard: electromagnetic compatibility –requirements and tests
IEC 61000-4-2:2008, ed 2.0Electromagnetic Compatibility (EMC) Part 4-2:Testing and Measurement Techniques -Electrostatic Discharge Immunity Test
IEC 61000-4-3 ed 3.0 with A1:2007+A2:2010Electromagnetic Compatibility (EMC) Part 4-3:Testing and Measurement Techniques -Radiated, Radio-frequency, ElectromagneticField Immunity Test
IEC 61000-4-4:2012, ed 3.0Electromagnetic Compatibility (EMC) Part 4-4:Testing and Measurement Techniques -Electrical Fast Transient/Burst Immunity Test
IEC 61000-4-5:2014, ed 3.0Electromagnetic Compatibility (EMC) Part 4-5:Testing and Measurement Techniques - SurgeImmunity Test
IEC 61000-4-6 ed 2.0 with A1:2004 + A2:2006Electromagnetic Compatibility (EMC) Part 4-6:Testing and Measurement Techniques -Immunity to Conducted Disturbances, Inducedby Radio-frequency Fields
IEC 61000-4-8:2009, ed 2.0Electromagnetic Compatibility (EMC) Part 4-8:Testing and Measurement Techniques -Power Frequency Magnetic Field ImmunityTest
IEC 61000-4-11:2004, ed 2.0Electromagnetic Compatibility (EMC) Part 4-11: Testing and Measurement Techniques -Voltage Dips, Short Interruptions and VoltageVariations Immunity Tests
IEC 61000-3-2:2014, ed 4.0Electromagnetic Compatibility (EMC) Part 3-2:Limits - Limits for Harmonic Current Emissions
IEC 61000-3-3:2013, ed 3.0Electromagnetic Compatibility (EMC) Part 3-3:Limits - Limitation of Voltage Changes,Voltage Fluctuations and Flicker in Public Low-voltage Supply Systems
CISPR 11 ed 5.0 with A1:2010Industrial, Scientific and Medical (ISM) Radio-Frequency Equipment - ElectromagneticDisturbance Characteristics - Limits andMethods of Measurement

{10}------------------------------------------------

Image /page/10/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in blue, blocky, sans-serif font. Below it, in a smaller, italicized, sans-serif font, is the word "neurology" in black. The logo is simple and modern.

Quality System Compliance:

{11}------------------------------------------------

Image /page/11/Picture/0 description: The image shows the logo for Natus Neurology. The word "natus" is in a larger, blue, sans-serif font. Below it, the word "neurology" is in a smaller, italicized, blue, sans-serif font. The logo is simple and clean, with a focus on the company name.

  • . ISO 13485:2003 Medical devices - Quality management systems - Requirements for regulatory purposes.
  • SOR/98-282 Canadian Medical Device Regulations .
  • . 21 CFR Part 820 US Food and Drug Administration's Quality System Requilation
  • . 93/42/EEG European Medical Device Directives
  • . ISO 14971:2007 Medical Devices - Application of Risk Management to Medical Devices
  • EN 980:2008 Medical Devices Symbols for Use of labeling of medical device .
  • EN 1041:2008 Information supplied by the manufacturer of medical devices .
  • ISO 15223-1:2012,Medical Devices Symbols to be Used With Medical Device Labels, Labeling, . and Information to be Supplied - Part 1: General requirements
  • . IEC 62304:2006 Medical device software - Software life-cycle processes

Conclusions

The substantial equivalence of the Natus Quantum with EMU128S and Neurolink IP products was demonstrated by testing in compliance with the Design Control process. The intended use and technology of the Natus Quantum is similar to that of the predicate device XLTEK-EMU128S and Neurolink IP. Verification and Validation was performed to ensure no new questions of safety or effectiveness are raised.

§ 882.1400 Electroencephalograph.

(a)
Identification. An electroencephalograph is a device used to measure and record the electrical activity of the patient's brain obtained by placing two or more electrodes on the head.(b)
Classification. Class II (performance standards).