K Number
K143344

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-04-09

(139 days)

Product Code
Regulation Number
876.5980
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The device is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body via extension sets and feeding tubes in neonatal and small pediatric patients.

Device Description

The NeoMed NeoConnect™ Enteral Syringes are standard piston style syringes consisting of a syringe barrel with integral ENFit syringe tip, syringe plunger, syringe gasket, and supplied with a syringe tip cap. They are provided in varying sizes ranging from 0.5ml to 100ml nominal capacity. The integral syringe tip is a female ENFit connector which is compatible only with feeding tubes and extension sets that have ENFit male connectors to form a dedicated system that prevents wrong-route administration of fluids. They possess translucent barrels to provide visualization of fluid contents and volume.

AI/ML Overview

The provided document is a 510(k) premarket notification for a medical device (NeoMed NeoConnect™ Enteral Syringes with ENFit Connector). It does not contain information about a study proving device performance against acceptance criteria in the context of an AI/ML powered device, which typically involves metrics like sensitivity, specificity, AUC, or reader performance.

The document describes bench testing for the physical properties and safety of the syringe. These are acceptance criteria for a physical medical device, not an AI algorithm.

Therefore, many of the requested categories (e.g., sample size for test set, ground truth experts, MRMC study, standalone performance) are not applicable to the information provided in this regulatory submission.

However, I can extract the acceptance criteria and related "performance" from the document based on the provided bench testing section.


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria CategorySpecific Test/StandardReported Device Performance/Compliance
Risk ManagementEN ISO 14971:2012Risk Analysis including design, user and process FMEA conducted.
Human Factors/UsabilityHuman Factors and Usability ValidationDesign Validation performed. Human Factors Validation Study performed (for ENFit connector).
BiocompatibilityISO 10993-5: CytotoxicityTesting conducted.
ISO 10993-10: Irritation and sensitizationTesting conducted.
ISO 10993-11: Acute ToxicityTesting conducted.
Chemical TestingExtractables and LeachablesTesting conducted.
Finished Device VerificationCritical Dimension verificationVerification conducted.
Ink AdhesionTesting conducted.
Capacity ToleranceTesting conducted.
Graduated ScaleTesting conducted.
Piston Fit in BarrelTesting conducted.
Air and Liquid Leakage TestingTesting conducted.
Syringe Tip (ENFit connector)Enteral Connector Misconnection AssessmentAssessment conducted.
ENFit Connector Risk Management ReportReport (including misconnections FMEA) conducted.
Dimensional verificationVerification conducted.
Liquid Leakage TestingTesting conducted.
Stress CrackingTesting conducted.
Resistance to separation from axial loadTesting conducted.
Resistance to separation from unscrewingTesting conducted.
Resistance to overridingTesting conducted.
Disconnection by unscrewingTesting conducted.
SterilizationEthylene Oxide (EO), 10^-6^ SALSterilization method is EO with 10^-6^ SAL.

2. Sample size used for the test set and the data provenance:

  • Sample Size: Not specified in the document. The document refers to "Finished Device Verification Testing" and various specific tests but does not disclose the number of units tested.
  • Data Provenance: Not explicitly stated, but based on the context of a 510(k) submission, this would be prospective bench testing conducted by the manufacturer, NeoMed, Inc. The document implies the tests were performed as part of the device's design validation prior to market submission.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not Applicable. This document describes testing for a physical medical device (syringe), not an AI/ML algorithm that requires expert-established ground truth for diagnostic or prognostic performance. The "ground truth" for these tests would be the measured physical properties or compliance with engineering standards.

4. Adjudication method for the test set:

  • Not Applicable. As above, this is not an AI/ML diagnostic or prognostic test. Adjudication methods like 2+1 or 3+1 are used for expert consensus in image interpretation or similar tasks, which are not relevant here.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • Not Applicable. This is for a physical medical device; no AI component is mentioned, and therefore, no MRMC study involving AI assistance would have been performed or reported here.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not Applicable. There is no algorithm or software component described in this device submission.

7. The type of ground truth used:

  • Direct Measurement / Standard Compliance. The ground truth for this device's performance is established by direct physical measurements against defined engineering standards (e.g., critical dimensions, leakage rates, strength of connections, biocompatibility testing against ISO standards). This is typical for traditional hardware medical devices.

8. The sample size for the training set:

  • Not Applicable. This device does not have a "training set" as it is not an AI/ML product. The design and manufacturing processes are developed based on engineering principles and regulatory standards, not trained on data.

9. How the ground truth for the training set was established:

  • Not Applicable. As there is no training set, there is no ground truth for a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 29, 2016

NeoMed, Inc. Melinda Harrison Smith, RAC, ASQ-CBA Director, Quality and Regulatory Affairs 100 Londonderry Court, Suite 112 Woodstock, GA 30188

K143344 Re:

Trade/Device Name: NeoMed NeoConnect™ Enteral Syringes with ENFit Connector Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PNR Dated: March 11, 2015 Received: March 13, 2015

Dear Melinda Harrison Smith,

This letter corrects our substantially equivalent letter of April 7, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use bared in the May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of I he goneral ocharing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Page 2 - Melinda Harrison Smith, RAC, ASQ-CBA

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely,

For Division

Douglas Silverstein -S 2016.09. -04'00'

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143344

Device Name

NeoMed NeoConnect™ Enteral Syringes with ENFit Connector

Indications for Use (Describe)

The device is indicated for use as a dispenser, a measuring device. It is used to deliver fluids into the body via extension sets and feeding tubes in neonatal and small pediatric patients.

Type of Use (Select one or both, as applicable)
---------------------------------------------------

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

TRADITIONAL 510(K) SUMMARY (21 CFR § 807.92)

I. SUBMITTERNeoMed, Inc.100 Londonderry CourtSuite 112Woodstock, GA 30188Tel: 770-516-2225Fax: 770-516-2448
Contact:Melinda Harrison Smith, RAC, CBAmharrison@neomedinc.com
Date Prepared:18 November 2014
EstablishmentRegistration Number:3006520777

ll. DEVICE

Trade Name:NeoConnect™ Enteral Syringes
Common Name:Enteral Syringe
Classification Name:Gastrointestinal tube and accessories (21 CFR § 876.5980)
Regulatory Class:II
Product Code :PIF

lll. PREDICATE DEVICE

NeoMed Oral / Enteral Syringe (0.5ml to 100ml) (K122373) This predicate had not been subject to a design-related recall.

IV. DEVICE DESCRIPTION

The NeoMed NeoConnect™ Enteral Syringes are standard piston style syringes consisting of a syringe barrel with integral ENFit syringe tip, syringe plunger, syringe gasket, and supplied with a syringe tip cap. They are provided in varying sizes ranging from 0.5ml to 100ml nominal capacity. The integral syringe tip is a female ENFit connector which is compatible only with feeding tubes and extension sets that have ENFit male connectors to form a dedicated system that prevents wrong-route administration of fluids. They possess translucent barrels to provide visualization of fluid contents and volume.

V. INDICATIONS FOR USE

The device is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body via extension sets and feeding tubes in neonatal and small pediatric patients.

{4}------------------------------------------------

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE

FeaturePREDICATE DEVICESUBJECT DEVICE
Indications for Use /Intended UseStatementThe device is indicated for use asa dispenser, a measuring deviceand an oral fluid transfer device. Itis used to inject fluids into the bodyvia extension sets and feedingtubes in neonatal and smallpediatric patients.The device is indicated for use as adispenser, a measuring device anda fluid transfer device. It is used todeliver fluids into the body viaextension sets and feeding tubes inneonatal and small pediatricpatients.
Patient Population /Environment of UseNeonates and Small Pediatrics/Hospital, Disposable and for singlepatient use onlyNeonates and Small Pediatrics/Hospital, Disposable and for singlepatient use only
Description of DeviceAn oral / enteral syringe consistingof a syringe barrel with integral tip(tapered), plunger, gasket, barrellubricant and supplied with asyringe tip cap.An enteral syringe consisting of asyringe barrel with integral tip(ENFit), plunger, gasket, barrellubricant and supplied with asyringe tip cap.
Syringe TypeOral / Enteral SyringeEnteral Syringe
Principle of OperationPiston Style SyringePiston Style Syringe
Syringe Tip TypeTaperedENFit(Dimensional compliance toAAMI/CN3:2014 (PS) Part 3 TableB.2 Female Enteral Small-BoreConnector).
Syringe Sizes(nominal volumes)0.5ml to 100ml sizes0.5ml to 100ml sizes
Syringe Barrel withintegral syringe tip andvolume graduationsBarrel - Polypropylene, class VIGraduation Ink - OrangeBarrel - Polypropylene, class VIGraduation Ink - Orange or Purple
Integral Syringe TipFlexural Modulus9500 Kg/cm² (932 MPa)9500 Kg/cm² (932 MPa)
Syringe PlungerInert Polypropylene, class VIWhite ColorantInert Polypropylene, class VIWhite Colorant
Syringe GasketSilicone, Class VI, Black ColorantSilicone, class VI, Black Colorant
Syringe Tip CapInert Polypropylene, class VIOrange ColorantInert Polypropylene, class VIOrange or Purple Colorant
Barrel LubricantPolydimethylsiloxane class VIPolydimethylsiloxane class VI
Packaging MaterialTyvek or medical grade paper /film primary pouch, Double Flutepaper corrugate case boxTyvek or medical grade paper / filmprimary pouch, Double Flute papercorrugate case box
Sterilization Method,SALEthylene Oxide (EO), 10-6 SALEthylene Oxide (EO), 10-6 SAL
Human Factors andUsabilityDesign ValidationDesign ValidationDevice Master File #MAF2258
Risk AnalysisDevice Risk Assessment includingDesign and User FMEADevice Risk Assessment includingDesign and User FMEADevice Master File #MAF2258

{5}------------------------------------------------

VII. PERFORMACE DATA (BENCH)

The following performance testing was conducted on the NeoConnect™ Enteral Syringe:

  • . Finished Device
    O

  • Risk Analysis including design, user and process FMEA (Failure Modes and o Effects Analysis) in accordance with EN ISO 14971:2012

  • o Human Factors and Usability Validation

  • Biocompatibility o

    • · ISO 10993-5: Cytotoxicity
    • ISO 10993-10: Irritation and sensitization .
    • . ISO 10993-11: Acute Toxicity
  • Chemical Testing o

    • Extractables and Leachables .
    • Finished Device Verification Testing
      • Critical Dimension verification .
      • . Ink Adhesion
      • =
        • Capacity Tolerance ●
        • Graduated Scale ●
        • Piston Fit in Barrel .
        • Air and Liquid Leakage Testing ●
  • Syringe Tip (ENFit connector) ●

    • Enteral Connector Misconnection Assessment o
    • ENFit Connector Risk Management Report (including misconnections FMEA) O
    • Human Factors Validation Study o
    • Dimensional verification O
    • Liquid Leakage Testing O
    • O Stress Cracking
    • Resistance to separation from axial load O
    • Resistance to separation form unscrewing O
    • Resistance to overriding O
    • Disconnection by unscrewing o

CONCLUSIONS VIII.

The NeoMed NeoConnect™ Enteral Syringes are substantially equivalent to the NeoMed Oral / Enteral Syringe (K122373).

§ 876.5980 Gastrointestinal tube and accessories.

(a)
Identification. A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).(b)
Classification. (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.