K Number
K143334

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-01-09

(50 days)

Product Code
Regulation Number
888.3070
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

RANGE /DENALI/MESA and SMALL STATURE and ARI are cleared for the following indications: Non-cervical fixation as an adjunct to the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); turnor: pseudarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion.

Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatric patients. The Range Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Device Description

The Range/Mesa/Denali Spinal System is a top-loading, multiple component, posterior (thoraciclumbar) spinal fixation system which consists of pedicle screws, rods, hooks and rod connectors The subject 510(k) adds Mesa 2 implants and connectors to the system.

Function: The system functions as a spinal fixation device to provide support and stabilization of the posterior thoracic and lumbar spine.

AI/ML Overview

The provided text is related to a 510(k) premarket notification for a spinal fixation system, not a device that would typically have acceptance criteria and performance evaluated through clinical studies with ground truth. This document is a regulatory clearance from the FDA for a medical device (Range/Denali/Mesa Spinal System) based on its substantial equivalence to previously marketed predicate devices.

Therefore, the requested information regarding acceptance criteria, study details, sample sizes, expert qualifications, adjudication methods, MRMC studies, standalone performance, and ground truth for training and test sets cannot be extracted from this document. The document describes a "non-clinical performance evaluation" based on mechanical testing, not a study involving human data or AI algorithms.

Here's why and what information is available:

  • Device Type: The device is a "Spinal Fixation System" (pedicle screws, rods, hooks, connectors) which is an implantable medical device used in orthopedic surgery. These devices are typically cleared based on mechanical performance, material compatibility, and substantial equivalence to existing devices, rather than human clinical studies with "ground truth" as you'd find for diagnostic AI algorithms.
  • Regulatory Pathway: This is a 510(k) submission, which means the manufacturer is demonstrating that their new device is substantially equivalent to a legally marketed predicate device. This often relies on comparative analysis of design, materials, and mechanical testing, rather than extensive human clinical trials to establish efficacy or diagnostic performance.
  • "Performance Evaluation" mentioned: The document states: "Performance evaluations were conducted on constructs representing the worst case components (including static torsion, static compression and dynamic compression bending in accordance with ASTM F1717) and the proposed implants were found to be substantially the same as predicate devices." This refers to mechanical testing in a lab setting, not a clinical study involving patients or AI.

In summary, there is no information in the provided text to fulfill your request for acceptance criteria and study details related to device performance in a clinical setting, as this document pertains to the regulatory clearance of a physical implantable device based on substantial equivalence and mechanical testing.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of three human profiles facing right, arranged in a way that they resemble a bird-like shape. The profiles are black and are set against a white background. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 9, 2015

K2M, Incorporated Ms. Nancy Giezen Manager, Regulatory Affairs 751 Miller Drive Southeast Leesburg, Virginia 20175

Re: K143334

Trade/Device Name: Range/Denali/Mesa Spinal System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class III Product Code: NKB, OSH, MNH, MNI, KWP, KWQ Dated: November 19, 2014 Received: November 20, 2014

Dear Ms. Giezen:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Nancy Giezen

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143334

Device Name RANGE /DENALI/MESA Spinal System

Indications for Use (Describe)

RANGE /DENALI/MESA and SMALL STATURE and ARI are cleared for the following indications: Non-cervical fixation as an adjunct to the following indications: degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); turnor: pseudarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion.

Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatric patients. The Range Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY Range/Mesa/Denali Spinal System K2M, Inc.

Submitter

K2M, Inc. 751 Miller Drive SE Leesburg, VA 20175 Contact Person: Nancy Giezen Telephone: 571 919-2168 Date Prepared: 1/7/2015

Classification

Trade Name:Range/Denali/Mesa Spinal System
Common Name:Spinal Fixation System
Regulatory Class:Class II, Class III (NKB)

Classification Name(s):

Pedicle Screw Spinal System (21 CFR 888.3070, Product Codes: NKB, OSH, MNH, MNI) Spinal Intervertebral body fixation Orthosis (21 CFR 888.3060, Product Code: KWQ) Spinal Interlaminal fixation Orthosis (21 CFR 888.3050, Product Code: KWP)

Predicate Device(s)

Primary Predicate: K2M Range/Mesa/Denali Spinal System (K141873) Additional Predicates:

  • . K2M Everest Spinal System (K140529)

Device Description

The Range/Mesa/Denali Spinal System is a top-loading, multiple component, posterior (thoraciclumbar) spinal fixation system which consists of pedicle screws, rods, hooks and rod connectors The subject 510(k) adds Mesa 2 implants and connectors to the system.

Function: The system functions as a spinal fixation device to provide support and stabilization of the posterior thoracic and lumbar spine.

Indications For Use

RANGE /DENALI/MESA and SMALL STATURE and ARI are cleared for the following indications:

Non-cervical fixation as an adjunct to fusion for the following indications; degenerative disc disease (defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies); spondylolisthesis; trauma (i.e., fracture or dislocation); spinal stenosis; curvatures (i.e. scoliosis, kyphosis and/or lordosis); tumor; pseudarthrosis; and/or failed previous fusion.

Except for hooks, when used as an anterolateral thoracic/lumbar system the Range Spinal System may also be used for the same indications as an adjunct to fusion.

{4}------------------------------------------------

Except for the ARI staples, the Range Spinal System is indicated as an adjunct to fusion to treat adolescent idiopathic scoliosis when used for posterior noncervical fixation in pediatric patients. The Range Spinal System for pediatric use is intended to be used with autograft and/or allograft. Pediatric pedicle screw fixation is limited to a posterior approach.

Technological Comparison to Predicate(s)

The Range/Mesa/Denali Spinal System was compared to predicate systems and the design features, materials and sizes were found to be substantially the same as these systems.

Non-clinical Performance Evaluation

Performance evaluations were conducted on constructs representing the worst case components (including static torsion, static compression and dynamic compression bending in accordance with ASTM F1717) and the proposed implants were found to be substantially the same as predicate devices.

Conclusion

There are no significant differences between the Range/Mesa/Denali System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to these other devices in design, function, material and intended use.

N/A