K Number
K143302

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-02-12

(87 days)

Product Code
Regulation Number
866.3373
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Xpert® MTB/RIF Assay, performed on the GeneXpert® Instrument Systems, is a qualitative, nested realtime polymerase chain reaction (PCR) in vitro diagnostic test for the detection of Mycobacterium tuberculosis complex DNA in raw sputum or concentrated sputum sediment prepared from induced or expectorated sputum. In specimens where Mycobacterium tuberculosis complex) is detected, the Xpert MTB/RIF Assay also detects the rifampin-resistance associated mutations of the rpoB gene.

The Xpert MTB/RIF Assay is intended for use with specimens for whom there is clinical suspicion of tuberculosis (TB) and who have received no antituberculosis therapy, or less than three days of therapy. This test is intended as an aid in the diagnosis of pulmonary tuberculosis when used in conjunction with clinical and other laboratory findings.

An Xpert MTB/RIF Assay result of "MTB NOT DETECTED" from either one or two sputum specimens is highly predictive of the absence of M. tuberculosis complex bacilli on serial fluorescent acid-fast sputum smears from patients with suspected active pulmonary tuberculosis and can be used as an aid in the decision of whether continued airborne infection isolation (AII) is warranted in patients with suspected pulmonary tuberculosis. The determination of whether testing of either one or two sputum specimens for decisions regarding removal from AII should be based on specific clinical circumstances and institutional guidelines. Clinical decisions regarding the need for continued AII should always occur in conjunction with other clinical and laboratory evaluations and Xpert MTB/RIF Assay results should not be the sole basis for infection control practices.

The Xpert MTB/RIF Assay must always be used in conjunction with mycobacterial culture to address the risk of false negative results and to recover organisms when MTB-complex is present for further characterization and drug susceptibility testing. However, decisions regarding the removal of patients from AII need not wait for culture results. Sputum specimens for TB culture, AFB smear microscopy, and Xpert MTB/RIF Assay testing should follow CDC recommendations with regard to collection methods and time frame between specimen collection.

The Xpert MTB/RIF Assay does not provide confirmation of rifampin susceptibility since mechanisms of rifampin resistance other than those detected by this device may exist that may be associated with a lack of clinical response to treatment.

Specimens that have both MTB-complex DNA and rifampin-resistance associated mutations of the rpoB gene detected by the Xpert MTB/RIF Assay must have results confirmed by a reference laboratory. If the presence of rifampin-resistance associated mutations of the rpoB gene is confirmed, specimens should also be tested for the presence of genetic mutations associated with resistance to other drugs.

The Xpert MTB/RIF Assay should only be performed in laboratories that follow safety practices in accordance with the CDC/NIH Biosafety in Microbiological and Biomedical Laboratories publication and applicable state or local regulations.

Device Description

The Xpert MTB/RIF Assay is a qualitative, automated, in vitro diagnostic test for qualitative detection of Mycobacterium tuberculosis (MTB) complex DNA in raw sputum samples or in concentrated sputum sediments prepared from induced or expectorated sputa that are either acid-fast bacilli (AFB) smear positive or negative. The assay is performed on the Cepheid GeneXpert Instrument Systems. The Xpert MTB/RIF Assay on the GeneXpert Instrument System automates and integrates sample preparation, nucleic acid amplification, and detection of the target sequences in simple or complex samples using real-time PCR. The system consists of an instrument, personal computer, and preloaded software for running the tests and viewing the results. The system requires the use of single-use disposable cartridges that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained, cross-contamination between samples is minimized.

The Xpert MTB/RIF Assay includes reagents for the detection of MTB and Rifampin (RIF) resistance from raw sputum samples and in prepared sputum sediments. A Sample Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge. The SPC is present to control for adequate processing of the target bacteria and to monitor for the presence of inhibitors in the PCR reaction. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dve stability.

The GeneXpert Instrument Systems, comprised of the GeneXpert Dx Systems, the GeneXpert Infinity-48 System, the GeneXpert Infinity-48s, and the GeneXpert Infinity-80 System, have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

AI/ML Overview

The provided document describes the Xpert® MTB/RIF Assay, an in vitro diagnostic test for the detection of Mycobacterium tuberculosis complex DNA and rifampin-resistance associated mutations. A clinical performance study (Study 2) was conducted to evaluate the device's effectiveness.

Here's a breakdown of the acceptance criteria and study information:

1. A table of acceptance criteria and the reported device performance

The document does not explicitly state pre-defined acceptance criteria in terms of specific performance metrics (e.g., "Sensitivity must be >= X%"). Instead, it presents the device's performance in a multi-center study and implicitly suggests that these results fulfill the requirements for substantial equivalence.

Based on the provided tables, we can extrapolate the reported device performance metrics:

MetricAcceptance Criteria (Implied)Reported Device Performance (Xpert MTB/RIF Assay)
For Smear Positive Cases:(Not explicitly stated, but high sensitivity/specificity desirable)Sensitivity: 98.5% (129/131); 95% CI: 94.6%, 99.6%
Specificity: 94.4% (17/18); 95% CI: 74.2%, 99.0%
For Smear Negative Cases:(Not explicitly stated, but good sensitivity/specificity desirable)Sensitivity: 54.8% (46/84); 95% CI: 44.1%, 65.0%
Specificity: 98.8% (718/727); 95% CI: 97.7%, 99.4%
Overall (One Xpert MTB/RIF Assay vs. MTB Culture):(Not explicitly stated, but good sensitivity/specificity desirable)Sensitivity: 81.4% (95% CI: 75.7%, 86.0%)
Specificity: 98.7% (95% CI: 97.5%, 99.3%)
Negative Predictive Value (NPV) for Single Test (U.S. Subjects, for AII removal):(High NPV for ruling out AFB smear-positive TB for AII removal)NPV: 97.6% (95% CI: 95.9%, 98.6%)
Difference in NPV (Xpert vs. Two AFB Smears, U.S. Subjects):(Improvement over AFB smears indicated)Difference in NPV: 2.6% (95% CI: 1.2%, 4.2%), indicating better performance for Xpert for U.S. subjects compared to two AFB smears

2. Sample size used for the test set and the data provenance

  • Sample Size for Clinical Performance Study (Study 2): 960 subjects.
  • Data Provenance: The study was a prospective multi-center study conducted at multiple sites in the United States, as well as South Africa and Brazil.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

The document does not specify the number of experts or their qualifications for establishing the ground truth. It states that the ground truth for MTB detection was based on mycobacterial culture (stated in Table 5.5). For rifampin resistance, it is stated that "Specimens that have both MTB-complex DNA and rifampin-resistance associated mutations of the rpoB gene detected by the Xpert MTB/RIF Assay must have results confirmed by a reference laboratory."

4. Adjudication method for the test set

The document does not describe a formal adjudication method (like 2+1 or 3+1) for resolving discrepancies in the ground truth. The ground truth for MTB detection was based on mycobacterial culture, an objective laboratory method. For rifampin resistance, a reference laboratory confirmation is required for positive results.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

This is not a MRMC study involving human readers or AI assistance in the traditional sense of image interpretation. The device is an automated, in vitro diagnostic test. However, the study does compare the performance of the Xpert MTB/RIF Assay (the device) against AFB smear microscopy which involves human interpretation (laboratory technicians reading smears).

  • Comparison between device and AFB smears: The document notes: "One Xpert MTB/RIF Assay was associated with a sensitivity of 81.4% (95% CI: 75.7%, 86.0%) for identifying MTB culture-positive subjects compared to a sensitivity of 60.9% (95% CI: 54.3%, 67.2%) for two AFB smears."
  • Effect Size (implicitly an improvement over human-read smears): For U.S. subjects, the NPV for one Xpert MTB/RIF Assay result was 97.6% compared to 95.0% for two AFB smears results. The difference in NPVs was 2.6% (95% CI: 1.2%, 4.2%). This suggests that the device performs better than two AFB smears in ruling out active pulmonary TB for decisions on airborne infection isolation in the U.S. context.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Yes, the primary clinical performance study (Study 2) evaluated the standalone performance of the Xpert MTB/RIF Assay. The assay is an automated in vitro diagnostic test that processes the sample and provides a result without direct human interpretation of the PCR data. Human involvement is primarily in sample collection, preparation, loading the cartridge, and interpreting the output from the instrument system.

7. The type of ground truth used

The primary ground truth for MTB detection was mycobacterial culture. For cases where MTB complex DNA and rifampin-resistance associated mutations were detected, confirmation by a reference laboratory served as the ground truth for rifampin resistance.

8. The sample size for the training set

The document does not explicitly mention the size of a "training set" for the device's development. This is typical for a diagnostic assay where algorithms are often based on well-established molecular biology principles and analytical validation rather than machine learning on a large training set in the same way an imaging AI algorithm would. Previous FDA-cleared 510(k) K131706 references are made for analytical sensitivity, reactivity, specificity, interfering substances, carry-over contamination, RIF resistance study, reproducibility, and instrument system precision. These earlier studies would have involved extensive analytical testing to establish the device's performance characteristics.

9. How the ground truth for the training set was established

As there isn't a clearly defined "training set" in the context of machine learning, the establishment of ground truth would refer to the methods used during the development and analytical validation of the assay. This would involve:

  • Known positive and negative samples: Using characterized bacterial strains (e.g., M. tuberculosis complex, non-tuberculous mycobacteria) and samples with known resistance profiles.
  • Sequencing and other molecular methods: To confirm the presence of M. tuberculosis DNA and specific rpoB gene mutations associated with rifampin resistance.
  • Culture results: As the gold standard for viability and identification of M. tuberculosis.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of a human figure, represented by three overlapping profiles facing to the right. The profiles are rendered in a dark color, contrasting with the white background of the seal.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

CEPHEID PAMELA JOHNSON EXECUTIVE DIRECTOR, CLINICAL AFFAIRS 904 CARIBBEAN DRIVE SUNNYVALE CA 94089-1189

February 19. 2015

Re: K143302

Trade/Device Name: Xpert® MTB/RIF Regulation Number: 21 CFR 866.3373 Regulation Name: Nucleic acid-based in vitro diagnostic devices for the detection of Mycobacterium tuberculosis complex and the genetic mutations associated with Mycobacterium tuberculosis complex antibiotic resistance in respiratory specimens

Regulatory Class: Class II Product Code: PEU Dated: November 13, 2014 Received: November 19, 2014

Dear Ms. Johnson:

This letter corrects our substantially equivalent letter of February 12, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If vou desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours, Uwe Sonerf -S for

Sally A. Hojvat, M. Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K143302

Device Name

Xpert® MTB/RIF

Indications for Use (Describe)

The Xpert® MTB/RIF Assay, performed on the GeneXpert® Instrument Systems, is a qualitative, nested realtime polymerase chain reaction (PCR) in vitro diagnostic test for the detection of Mycobacterium tuberculosis complex DNA in raw sputum or concentrated sputum sediment prepared from induced or expectorated sputum. In specimens where Mycobacterium tuberculosis complex) is detected, the Xpert MTB/RIF Assay also detects the rifampin-resistance associated mutations of the rpoB gene.

The Xpert MTB/RIF Assay is intended for use with specimens for whom there is clinical suspicion of tuberculosis (TB) and who have received no antituberculosis therapy, or less than three days of therapy. This test is intended as an aid in the diagnosis of pulmonary tuberculosis when used in conjunction with clinical and other laboratory findings.

An Xpert MTB/RIF Assay result of "MTB NOT DETECTED" from either one or two sputum specimens is highly predictive of the absence of M. tuberculosis complex bacilli on serial fluorescent acid-fast sputum smears from patients with suspected active pulmonary tuberculosis and can be used as an aid in the decision of whether continued airborne infection isolation (AII) is warranted in patients with suspected pulmonary tuberculosis. The determination of whether testing of either one or two sputum specimens for decisions regarding removal from AII should be based on specific clinical circumstances and institutional guidelines. Clinical decisions regarding the need for continued AII should always occur in conjunction with other clinical and laboratory evaluations and Xpert MTB/RIF Assay results should not be the sole basis for infection control practices.

The Xpert MTB/RIF Assay must always be used in conjunction with mycobacterial culture to address the risk of false negative results and to recover organisms when MTB-complex is present for further characterization and drug susceptibility testing. However, decisions regarding the removal of patients from AII need not wait for culture results. Sputum specimens for TB culture, AFB smear microscopy, and Xpert MTB/RIF Assay testing should follow CDC recommendations with regard to collection methods and time frame between specimen collection.

The Xpert MTB/RIF Assay does not provide confirmation of rifampin susceptibility since mechanisms of rifampin resistance other than those detected by this device may exist that may be associated with a lack of clinical response to treatment.

Specimens that have both MTB-complex DNA and rifampin-resistance associated mutations of the rpoB gene detected by the Xpert MTB/RIF Assay must have results confirmed by a reference laboratory. If the presence of rifampin-resistance associated mutations of the rpoB gene is confirmed, specimens should also be tested for the presence of genetic mutations associated with resistance to other drugs.

The Xpert MTB/RIF Assay should only be performed in laboratories that follow safety practices in accordance with the CDC/NIH Biosafety in Microbiological and Biomedical Laboratories publication and applicable state or local regulations.

{3}------------------------------------------------

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 2 of 2

PSC Publishing Services (301) 443-6740 EF

{4}------------------------------------------------

5.0 510(k) Summary

As required by 21 CFR Section 807.92(c).

Submitted by:Cepheid®904 Caribbean DriveSunnyvale, CA 90489Phone number: (408) 400-8460Fax number: (847) 510-0539E-mail: kerry.flom@cepheid.com
Contact:Kerry J. Flom, Ph.D.
Date of Preparation:November 13, 2014
Device:
Trade name:Xpert® MTB/RIF
Common name:Xpert MTB/RIF Assay
Type of Test:Qualitative, nested real-time polymerase chain reaction (PCR)
Regulation number/866.3373/
Classification name/Product code:System, nucleic acid-based, mycobacterium tuberculosiscomplex, resistance marker, direct specimen/PEU
ClassificationAdvisory PanelMicrobiology (83)
Predicate DevicesName(s):Xpert MTB/RIF

Device Description:

The Xpert MTB/RIF Assay is a qualitative, automated, in vitro diagnostic test for qualitative detection of Mycobacterium tuberculosis (MTB) complex DNA in raw sputum samples or in concentrated sputum sediments prepared from induced or expectorated sputa that are either acid-fast bacilli (AFB) smear positive or negative. The assay is performed on the Cepheid GeneXpert Instrument Systems. The Xpert MTB/RIF Assay on the GeneXpert Instrument System automates and integrates sample preparation, nucleic acid amplification, and detection of the target sequences in simple or complex samples using real-time PCR. The system consists of an instrument, personal computer, and preloaded software for running the tests and viewing the results. The system requires the use of single-use disposable cartridges that hold the PCR reagents and host the PCR process. Because the cartridges are self-contained, cross-contamination between samples is minimized.

The Xpert MTB/RIF Assay includes reagents for the detection of MTB and Rifampin (RIF) resistance from raw sputum samples and in prepared sputum sediments. A Sample

{5}------------------------------------------------

Processing Control (SPC) and a Probe Check Control (PCC) are also included in the cartridge. The SPC is present to control for adequate processing of the target bacteria and to monitor for the presence of inhibitors in the PCR reaction. The Probe Check Control (PCC) verifies reagent rehydration, PCR tube filling in the cartridge, probe integrity, and dve stability.

The GeneXpert Instrument Systems, comprised of the GeneXpert Dx Systems, the GeneXpert Infinity-48 System, the GeneXpert Infinity-48s, and the GeneXpert Infinity-80 System, have 1 to 80 randomly accessible modules, depending upon the instrument, that are each capable of performing separate sample preparation and real-time PCR tests. Each module contains a syringe drive for dispensing fluids (i.e., the syringe drive activates the plunger that works in concert with the rotary valve in the cartridge to move fluids between chambers), an ultrasonic horn for lysing cells or spores, and a proprietary I-CORE® thermocycler for performing real-time PCR and detection.

Device Intended Use:

The Xpert® MTB/RIF Assay, performed on the GeneXpert® Instrument Systems, is a qualitative, nested real-time polymerase chain reaction (PCR) in vitro diagnostic test for the detection of Mycobacterium tuberculosis complex DNA in raw sputum or concentrated sputum sediment prepared from induced or expectorated sputum. In specimens where Mycobacterium tuberculosis complex (MTB-complex) is detected, the Xpert MTB/RIF Assay also detects the rifampin-resistance associated mutations of the rpoB gene.

The Xpert MTB/RIF Assay is intended for use with specimens from patients for whom there is clinical suspicion of tuberculosis (TB) and who have received no antituberculosis therapy, or less than three days of therapy. This test is intended as an aid in the diagnosis of pulmonary tuberculosis when used in conjunction with clinical and other laboratory findings.

An Xpert MTB/RIF Assay result of "MTB NOT DETECTED" from either one or two sputum specimens is highly predictive of the absence of M. tuberculosis complex bacilli on serial fluorescent acid-fast sputum smears from patients with suspected active pulmonary tuberculosis and can be used as an aid in the decision of whether continued airborne infection isolation (AII) is warranted in patients with suspected pulmonary tuberculosis. The determination of whether testing of either one or two sputum specimens is appropriate for decisions regarding removal from AII should be based on specific clinical circumstances and institutional guidelines. Clinical decisions regarding the need for continued AII should always occur in conjunction with other clinical and laboratory evaluations and Xpert MTB/RIF Assay results should not be the sole basis for infection control practices.

The Xpert MTB/RIF Assay must always be used in conjunction with mycobacterial culture to address the risk of false negative results and to recover organisms when MTB-complex is present for further characterization and drug susceptibility testing. However, decisions regarding the removal of patients from AII need not wait for culture

{6}------------------------------------------------

results. Sputum specimens for TB culture, AFB smear microscopy, and Xpert MTB/RIF Assay testing should follow CDC recommendations with regard to collection methods and time frame between specimen collection.

The Xpert MTB/RIF Assay does not provide confirmation of rifampin susceptibility since mechanisms of rifampin resistance other than those detected by this device may exist that may be associated with a lack of clinical response to treatment.

Specimens that have both MTB-complex DNA and rifampin-resistance associated mutations of the rpoB gene detected by the Xpert MTB/RIF Assay must have results confirmed by a reference laboratory. If the presence of rifampin-resistance associated mutations of the rpoB gene is confirmed, specimens should also be tested for the presence of genetic mutations associated with resistance to other drugs.

The Xpert MTB/RIF Assay should only be performed in laboratories that follow safety practices in accordance with the CDC/NIH Biosafety in Microbiological and Biomedical Laboratories publication and applicable state or local regulations.

Substantial Equivalence:

Predicate device name(s): Cepheid Xpert® MTB/RIF Assay

Predicate 510(k) number(s): K131706

Comparison with predicate:

The Xpert MTB/RIF Assay is substantially equivalent to the current Xpert MTB/RIF Assay [510(k) #K131706].

Similarities and differences between the Cepheid Xpert MTB/RIF Assay and the predicate device are shown in Table 5.1.

A prospective multi-center study (Study 2) was conducted at multiple sites in the United States, as well as South Africa and Brazil. Performance of the MTB/RIF Assay was assessed as an alternative to fluorescent stained AFB-smear microscopy as an aid in determining the need for continued airborne infection isolation in patients with suspected active pulmonary tuberculosis. The clinical data demonstrated that a single negative Xpert MTB/RIF result predicted the absence of AFB smear-positive pulmonary tuberculosis with an overall negative predictive value (NPV) of 99.7% (99.6% in the U.S. and 100% in non-U.S.). Two serial negative Xpert MTB/RIF Assay results predicted the absence of AFB smear-positive pulmonary tuberculosis with an overall NPV of 100%.

{7}------------------------------------------------

DevicePredicate
ItemCepheid Xpert MTB/RIFCurrent Cepheid Xpert MTB/RIF
510(k) NumberTo be assigned#K131706
RegulationSame866.3373
Product CodeSamePEU
Device ClassSameII
Technology/DetectionSameMultiplex real time RT/PCR
Indication for UseSamePatients for whom there is clinicalsuspicion of tuberculosis (TB) and whohave received no antituberculosistherapy, or less than 3 days of therapy
Assay TargetsSameMTB-complex DNA and Rif resistanceassociated mutations
Specimen TypesSameRaw sputum samples or concentratedsputum sediments
TechnologicalPrinciplesSameReal-time PCR
Nucleic AcidExtractionSameYes
ExtractionMethodsSameSample preparation integrated inGeneXpert Cartridge and GeneXpertInstrumentation System
Assay ResultsSameQualitative
Instrument SystemSameCepheid GeneXpert Instrument Systems
Assay ControlsSameSample Processing Control (SPC) andProbe Check Control (PCC).
Failures result in single sample repeat.
Rapid test resultsSameTotal 120 minutes for sample preparationand real-time PCR
Laboratory UsersSameCLIA Moderate or High Complexity
Differences
DevicePredicate
ItemCepheid Xpert MTB/RIFCurrent Cepheid Xpert MTB/RIF
IntendedUseThe Xpert® MTB/RIF Assay, performed on the GeneXpert® Instrument Systems, is a qualitative, nested real-time polymerase chain reaction (PCR) in vitro diagnostic test for the detection of Mycobacterium tuberculosis complex DNA in raw sputum or concentrated sputum sediment prepared from induced or expectorated sputum. In specimens where Mycobacterium tuberculosis complex (MTB-complex) is detected, the Xpert MTB/RIF Assay also detects the rifampin-resistance associated mutations of the rpoB gene.The Xpert MTB/RIF Assay is intended for use with specimens from patients for whom there is clinical suspicion of tuberculosis (TB) and who have received no antituberculosis therapy, or less than three days of therapy. This test is intended as an aid in the diagnosis of pulmonary tuberculosis when used in conjunction with clinical and other laboratory findings.The Xpert® MTB/RIF Assay, performed on the GeneXpert® Instrument Systems, is a qualitative, nested real-time polymerase chain reaction (PCR) in vitro diagnostic test for the detection of Mycobacterium tuberculosis complex DNA in raw sputum or concentrated sediments prepared from induced or expectorated sputum. In specimens where Mycobacterium tuberculosis complex (MTB-complex) is detected, the Xpert MTB/RIF Assay also detects the rifampin-resistance associated mutations of the rpoB gene.The Xpert MTB/RIF Assay is intended for use with specimens from patients for whom there is clinical suspicion of tuberculosis (TB) and who have received no antituberculosis therapy, or less than 3 days of therapy. This test is intended as an aid in the diagnosis of pulmonary tuberculosis when used in conjunction with clinical and other laboratory findings.The Xpert MTB/RIF Assay does not provide confirmation of rifampin susceptibility since mechanisms of
An Xpert MTB/RIF Assay result of "MTB NOT DETECTED" from either one or two sputum specimens is highly predictive of the absence of M. tuberculosis complex bacilli on serial fluorescent acid-fast sputum smears from patients with suspected active pulmonary tuberculosis and can be used as an aid in the decision of whether continued airborne infection isolation (AII) is warranted in patients with suspected pulmonary tuberculosis. The determination of whether testing of either one or two sputum specimens is appropriate for decisions regardingrifampin resistance other than those detected by this device may exist that may be associated with a lack of clinical response to treatment.Specimens that have both MTB-complex DNA and rifampin-resistance associated mutations of the rpoB gene detected by the Xpert MTB/RIF Assay must have results confirmed by a reference laboratory. If the presence of rifampin-resistance associated mutations of the rpoB gene is confirmed, specimens should also be tested for the presence of genetic mutations associated with resistance to other drugs.
Differences
DevicePredicate
ItemCepheid Xpert MTB/RIFCurrent Cepheid Xpert MTB/RIF
removal from AII should be based onspecific clinical circumstances andinstitutional guidelines. Clinicaldecisions regarding the need forcontinued AII should always occur inconjunction with other clinical andlaboratory evaluations and XpertMTB/RIF Assay results should not bethe sole basis for infection controlpractices.The Xpert MTB/RIF Assay mustalways be used in conjunction withmycobacterial culture to address therisk of false negative results and torecover organisms when MTB-complex is present for furthercharacterization and drugsusceptibility testing. However,decisions regarding the removal ofpatients from AII need not wait forculture results. Sputum specimens forTB culture, AFB smear microscopy,and Xpert MTB/RIF Assay testingshould follow CDC recommendationswith regard to collection methods andtime frame between specimencollection.The Xpert MTB/RIF Assay must be usedin conjunction with mycobacterialculture to address the risk of falsenegative results and to recover theorganisms for further characterizationand drug susceptibility testing.The Xpert MTB/RIF Assay should onlybe performed in laboratories that followsafety practices in accordance with theCDC/NIH Biosafety in Microbiologicaland Biomedical Laboratories publicationand applicable state or local regulations.
The Xpert MTB/RIF Assay does notprovide confirmation of rifampinsusceptibility since mechanisms ofrifampin resistance other than thosedetected by this device may exist thatmay be associated with a lack ofclinical response to treatment.Specimens that have both MTB-complex DNA and rifampin-resistanceassociated mutations of the rpoB gene
detected by the Xpert MTB/RIF Assaymust have results confirmed by a
Differences
DevicePredicate
ItemCepheid Xpert MTB/RIFCurrent Cepheid Xpert MTB/RIF
reference laboratory. If the presence ofrifampin-resistance associated mutationsof the rpoB gene is confirmed,specimens should also be tested for thepresence of genetic mutations associatedwith resistance to other drugs.The Xpert MTB/RIF Assay should onlybe performed in laboratories that followsafety practices in accordance with theCDC/NIH Biosafety in Microbiologicaland Biomedical Laboratories publicationand applicable state or local regulations.

Table 5.1. Comparison of Similarities and Differences of the proposed Xpert MTB/RIF Assay with the Predicate Device

{8}------------------------------------------------

{9}------------------------------------------------

{10}------------------------------------------------

Non-Clinical Studies:

Analytical Sensitivity

Please refer to the previously FDA-cleared 510(k) #K131706.

Analytical Reactivity (Inclusivity)

Please refer to the previously FDA-cleared 510(k) #K131706.

Analytical Specificity (Exclusivity)

Potential cross-reactivity of eight microorganisms was evaluated by in silico analysis. All eight microorganisms tested revealed no potential for cross-reactivity. See Table 5.2 below.

Table 5.2. Microorganisms Predicted to be Non-Cross Reactive by In Silico Analysis

Mycobacterium chimaeraMycobacterium immunogenum
Mycobacterium avium subsp.paratuberculosisMycobacterium massiliense
Mycobacterium avium subsp.silvaticumMycobacterium bolletii
Mycobacterium avium subsp.hominissuis
Mycobacterium franklinii

{11}------------------------------------------------

Please refer to the previously FDA-cleared 510(k) #K131706 for a description of additional analytical specificity testing performed.

Interfering Substances

Please refer to the previously FDA-cleared 510(k) #K131706.

Carry-Over Contamination Study

Please refer to the previously FDA-cleared 510(k) #K131706.

RIF Resistance Study

Please refer to the previously FDA-cleared 510(k) #K131706.

Linearity

Not applicable, the Xpert MTB/RIF Assay is a qualitative assay.

Clinical Performance Characteristics:

Reproducibility

Please refer to the previously FDA-cleared 510(k) #K131706.

Instrument System Precision

Please refer to the previously FDA-cleared 510(k) #K131706.

Clinical Performance Study

Xpert MTB/RIF Assay Performance in an HIV Population

To compare performance of the Xpert MTB/RIF Assay in HIV-infected and HIVuninfected subjects, data from Study 2 were analyzed by smear status of specimens and HIV status of the population. Tables 5.3 and 5.4 compare the sensitivities and specificities of one Xpert MTB/RIF Assay result in specimens obtained from HIVinfected and HIV-uninfected subjects stratified by AFB smear-positive and AFB smearnegative results, respectively. For both HIV-infected and HIV-uninfected subjects, the sensitivity of the Xpert MTB/RIF Assay for detection of MTB-complex was higher in AFB smear-positive specimens (100.0% and 97.8%, respectively) than in AFB smearnegative specimens (52.1% and 58.3%. respectively). These data are summarized in Table 5.4.

{12}------------------------------------------------

XpertMTB/RIFOverallHIV-infectedHIV-uninfectedDifference(95% CI)
Sensitivity98.5%(129/131)100%(39/39)97.8%(90/92)2.2%(-0.8%, 5.2%)
Specificity94.4%(17/18)100%(7/7)90.9%(10/11)9.1%(-7.9%, 26.1%)

Table 5.3. Comparison of Sensitivity and Specificity of One Xpert MTB/RIF Assay Result in HIV-Infected and HIV-Uninfected Subjects - AFB Smear Positive Only

Table 5.4. Comparison of Sensitivity and Specificity of One Xpert MTB/RIF Assay Result in HIV-Infected and HIV-Uninfected Subjects - AFB Smear Negative Only

XpertMTB/RIFOverallHIV-infectedHIV-uninfectedDifference(95% CI)
Sensitivity54.8%(46/84)52.1%(25/48)58.3%(21/36)-6.3%(-27.7%, 15.2%)
Specificity98.8%(718/721)98.2%(332/338)99.2%(386/389)-1.0%(-2.7%, 0.7%)

Xpert MTB/RIF Assay Performance as Basis for Removal of Patients from Respiratory Isolation

Tables 5.5 and 5.6 present the overall performance of one Xpert MTB/RIF Assay result compared to the results of MTB culture, stratified by AFB smear result (Table 5.5). Table 5.6 is a side-by-side comparison of the performance of one Xpert MTB/RIF Assay result versus the composite result of two AFB smears in U.S. and non-U.S. subjects (N=960). Overall sensitivity of one Xpert MTB/RIF Assay in AFB smear-positive and AFB smearnegative subjects (based on two AFB smears) was 98.5% (95% CI: 94.6%,99.6%) and 54.8% (95% CI: 44.1%, 65.0%) respectively, and overall specificity was 98.7% (95% CI: 97.5%, 99.3%). One Xpert MTB/RIF Assay result of MTB Not Detected was associated with a probability of MTB culture-positive/AFB smear-positive results of 0.4% for U.S. subjects and 0.0% for non-U.S. subjects.

{13}------------------------------------------------

Table 5.5. Performance of One Xpert MTB/RIF Assay Result Stratified by Two AFB Smears Relative to MTB Culture in U.S. and non-U.S. Subjects

CultureTotal
PositiveNegative
XpertMTB/RIFAssayAFBSmear+AFBSmear-OverallCulture+AFBSmear+AFBSmear-OverallCulture-
Positive129461751910a185
Negative2384017718735775
Total1318421518b727745960

Performance of Xpert MTB/RIF Assay for Smear Positive:

Sensitivity: 98.5% (129/131), 95% CI: 94.6%,99.6% Specificity: 94.4% (17/18), 95% CI: 74.2%, 99.0%

Performance of Xpert MTB/RIF Assay for Smear Negative:

Sensitivity: 54.8% (46/84), 95% CI: 44.1%, 65.0% Specificity: 98.8% (718/727), 95% CI: 97.7%, 99.4%

Prevalence of MTB Culture Positive: 22.4% (215/960) Prevalence of MTB Culture Positive in U.S. subjects: Prevalence of MTB Culture positive in non-U.S. subjects: 37.1% (127/342)

Percent of AFB smear positive subjects among subjects with MTB Culture Positive: 60.9% (131/215)

Overall Probability of MTB Culture Positive among subjects with an Xpert MTB/RIF Negative Result: 5.2% (40/775), 95% CI: 3.8%, 7.0%

Probability of MTB Culture Positive among subjects with an Xpert MTB/RIF Negative Result (U.S. subjects): 2.4% (13/539), 95% CI: 1.4%, 4.1%

Probability of MTB Culture Positive among subjects with an Xpert MTB/RIF Negative Result (non-U.S. subjects): 11.4% (27/236), 95% CI: 8.0%, 16.1%

Overall Probability of MTB Culture Positive and AFB smear positive among subjects with an Xpert MTB/RIF Negative Result: 0.3% (2/775), 95% CI: <0.1%, 0.9%

Probability of MTB Culture Positive and AFB smear positive among subjects with an Xpert MTB/RIF Negative Result (U.S. subjects): 0.4% (2/539), 95% CI: 0.1%, 1.3%

Probability of MTB Culture Positive and AFB smear positive among subjects with an Xpert MTB/RIF Negative Result (non-U.S.) subjects: 0.0% (0/236), 95% CI: 0.0%, 1.6%

40f the 10 MTB culture-negative specimens that were positive by Xpert MTB/RIF Assay, 5 grew non-tuberculosis mycobacteria (NTM). MTB-complex was isolated and identified using standard of care methods not associated with the study protocol in 4 of the 5 specimens.

0 of the 18 MTB culture-negative/AFB smear-positive specimens, 14 grew NTM.

One Xpert MTB/RIF Assay was associated with a sensitivity of 81.4% (95% CI: 75.7%, 86.0%) for identifying MTB culture-positive subjects compared to a sensitivity of 60.9% (95% CI: 54.3%, 67.2%) for two AFB smears.

{14}------------------------------------------------

Table 5.6. Comparison Of Performance of One Xpert MTB/RIF Assay Result vs Two AFB Smears Each Versus MTB Culture in U.S. and non-U.S. Subjects

One Xpert MTB/RIFAssay ResultsCultureTwo AFBSmearsCulture
XpertPositive17510185AFB SmearPositive13118149
Negative40735775Negative84727811
Total215745960Total215745960
Sensitivity:81.4% (95% CI: 75.7, 86.0)Sensitivity:60.9% (95% CI: 54.1, 67.5)
Specificity:98.7% (95% CI: 97.5, 99.3)Specificity:97.6% (95% CI: 96.2, 98.6)
U.S. prevalencePPV:NPV:14.2% (95% CI: 11.7, 17.2)94.9% (95% CI: 87.7, 98.0)97.6% (95% CI: 95.9, 98.6)U.S. prevalencePPV:NPV:14.2% (95% CI: 11.7, 17.2)77.2% (95% CI: 66.8, 85.1)95.0% (95% CI: 92.8, 96.5)
Non-U.S. prevalencePPVNPV37.1% (95% CI: 32.2, 42.4)94.3% (95% CI: 88.2, 97.4)88.6 (95% CI: 83.9, 92.0)Non-U.S. Prevalence:PPVNPV37.1% (95% CI: 32.2, 42.4)100% (95% CI: 94.8, 100)79.0% (95% CI: 73.8, 83.5)

In U.S. subjects, the NPV for one Xpert MTB/RIF Assay result was 97.6% (95% CI: 95.9%, 98.6%) while the NPV for two AFB smears results was 95.0% (95% CI: 92.8%, 96.5%) with a prevalence of TB in U.S. subjects of 14.2%. The difference in NPVs was 2.6% with 95% CI: 1.2%, 4.2%.

Proposed Labeling

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

Conclusions

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

§ 866.3373 Nucleic acid-based in vitro diagnostic devices for the detection of Mycobacterium tuberculosis complex (MTB-complex) and the genetic mutations associated with MTB-complex antibiotic resistance in respiratory specimens.

(a)
Identification. Nucleic acid-based in vitro diagnostic devices for the detection ofMycobacterium tuberculosis complex (MTB-complex) and the genetic mutations associated with MTB-complex antibiotic resistance in respiratory specimens are qualitative nucleic acid-based devices that detect the presence of MTB-complex-associated nucleic acid sequences in respiratory samples. These devices are intended to aid in the diagnosis of pulmonary tuberculosis and the selection of an initial treatment regimen when used in conjunction with clinical findings and other laboratory results. These devices do not provide confirmation of antibiotic susceptibility since other mechanisms of resistance may exist that may be associated with a lack of clinical response to treatment other than those detected by the device.(b)
Classification. Class II (special controls). The special controls for this device are:(1) The FDA document entitled “Class II Special Controls Guideline: Nucleic Acid-Based In Vitro Diagnostic Devices for the Detection of
Mycobacterium tuberculosis Complex and Genetic Mutations Associated with Antibiotic Resistance in Respiratory Specimens,” which addresses the mitigation of risks specific to the detection of MTB-complex. For availability of the document, see § 866.1(e).(2) The following items, which address the mitigation of risks specific to the detection of the genetic mutations associated with antibiotic resistance of MTB-complex:
(i) The device must include an external positive assay control as appropriate. Acceptable positive assay controls include MTB-complex isolates containing one or more antibiotic-resistance associated target sequences detected by the device.
(ii) The device must include internal controls as appropriate. An acceptable internal control may include human nucleic acid co-extracted with MTB-complex containing nucleic acid sequences associated with antibiotic resistance and primers amplifying human housekeeping genes (e.g., RNaseP, β-actin).
(iii) The device's intended use must include a description of the scope of antibiotic resistance targeted by the assay, i.e., the specific drugs and/or drug classes.
(iv) The specific performance characteristics section of the device's labeling must include information regarding the specificity of the assay oligonucleotides for detecting mutations associated with antibiotic resistance of MTB-complex, and any information indicating the potential for non-specific binding (e.g., BLAST search).
(v) In demonstrating device performance you must perform:
(A) Pre-analytical studies that evaluate:
(
1 )Frozen samples. If there is use of any frozen samples in the device performance studies, or if there is a device claim for the use of frozen samples for testing, the effect of freezing samples prior to testing and the effect of multiple freeze/thaw cycles on both antibiotic susceptible and antibiotic resistant strains of MTB-complex.(
2 )Nucleic acid extraction methods. Extraction methods must parallel those used in devices for the detection of MTB-complex nucleic acid and confirm that the detection of the genetic mutations associated with antibiotic resistance is not affected.(B) Analytical studies that analyze:
(
1 )Limit of Detection. Limit of Detection must be determined in the most challenging matrix (e.g., sputum) claimed for use with the device. The Limit of Detection must be determined using both antibiotic susceptible and antibiotic resistant strains of MTB-complex. The antibiotic resistant strains must be those with well characterized genetic mutations associated with antibiotic resistance.(
2 )Analytical Reactivity (Inclusivity). Testing must be conducted to evaluate the ability of the device to detect genetic mutations associated with antibiotic resistance in a diversity of MTB-complex strains. Isolates used in testing must be well characterized. Isolate strain characterization must be determined using standardized reference methods recognized by a reputable scientific body and appropriate to the strain lineage.(
3 )Within-Laboratory (Repeatability) Precision Testing. Within-laboratory precision studies, if appropriate, must include at least one antibiotic resistant and one antibiotic susceptible strain of MTB-complex.(
4) Between Laboratory Reproducibility Testing. The protocol for the reproducibility study may vary slightly depending on the assay format; however, the panel must include at least one antibiotic resistant and one antibiotic susceptible strain of MTB-complex.(C) Clinical Studies. Clinical performance of the device must be established by conducting prospective clinical studies that include subjects with culture confirmed active tuberculosis. Studies must attempt to enroll subjects at risk for antibiotic-resistant MTB-complex; however, it may be necessary to include supplemental antibiotic resistant retrospective and contrived samples. Clinical studies must compare device results to both phenotypic drug susceptibility testing and genotypic reference methods. The genotypic reference method must be a polymerase chain reaction based method that uses primers different from those in the experimental device and confirmed by bidirectional sequencing.