(60 days)
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Clinical Applications:
- Fetal (Sometimes called Obstetrics) .
- Abdominal ●
- Intra-operative ●
- Intra-operative (Neuro) (sometimes called Neuro Surgical) ●
- Pediatrics .
- Small Organ (Sometimes called Small Parts) .
- . Neonatal Cephalic (Cephalic is sometimes called trans-cranial)
- Adult Cephalic (Cephalic is sometimes called trans-cranial) ●
- Trans-rectal
- . Trans-vaqinal
- . Trans-urethral
- Musculo-skeletal (Conventional)
- Musculo-skeletal (Superficial)
- Cardiac Adult
- Peripheral Vessel (Sometimes called Peripheral Vascular) .
Ultrasound scanner and transducers for B, Tissue and Contrast Harmonic Imaging, M, PWD, CWD, Color Doppler, Vector Flow Imaging and combined mode imaging and Elastography.
Signal Analysis.
Guidance of biopsy needles, geometrical measurements and calculation of parameters. And optimal 3-D unit can reconstruct a series of 2-D images into a single 3-D volume and display this on the screen.
An optional Vector Flow Imaging (VFI) module: Color Flow Mapping (CFM) imaging mode with the ability to visualize both the axial and the transverse velocity.
This document is an FDA 510(k) premarket notification for the BK Medical Ultrasound Scanner System bk2300, dated January 16, 2015. It primarily details the device's indications for use, its substantial equivalence to predicate devices, and summaries of non-clinical tests. The document does not contain information about acceptance criteria or a study proving the device meets those criteria, specifically concerning clinical performance metrics. It focuses on safety and technological equivalence.
Therefore, many of the requested details cannot be extracted from the provided text.
Here is a summary of the available information:
1. A table of acceptance criteria and the reported device performance
- Not provided in the document. The document discusses regulatory compliance, safety standards, and intended uses, but does not specify performance acceptance criteria or report specific performance metrics from clinical studies.
2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)
- Not provided in the document. The document exclusively mentions "non-clinical tests" and does not refer to clinical studies or test sets.
3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)
- Not applicable. No clinical test set or ground truth establishment by experts is mentioned.
4. Adjudication method (e.g. 2+1, 3+1, none) for the test set
- Not applicable. No clinical test set or adjudication method is mentioned.
5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance
- Not applicable. The device is an ultrasound scanner system, not an AI assistance tool for human readers. No MRMC study is mentioned.
6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done
- Not applicable. This document describes a diagnostic ultrasound system, not a standalone algorithm.
7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)
- Not applicable. No clinical ground truth is mentioned. The document primarily relies on non-clinical testing for safety and technical characteristics.
8. The sample size for the training set
- Not applicable. No training set is mentioned as this document does not describe an AI/ML device relying on trained models.
9. How the ground truth for the training set was established
- Not applicable. No training set or ground truth for it is mentioned.
General Information from the Document (Relevant to Device and Regulatory Approval):
-
Device Name: Ultrasound Scanner System bk2300
-
Regulation Number: 21 CFR 892.1550
-
Regulation Name: Ultrasonic pulsed doppler imaging system
-
Regulatory Class: II
-
Key Features/Modes: B, Tissue and Contrast Harmonic Imaging, M, PWD, CWD, Color Doppler, Vector Flow Imaging, combined mode imaging, and Elastography.
-
New Modes (in this submission): Elastography, Vector Flow Imaging (VFI), Continuous Wave Doppler (CWD/CW).
-
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body.
-
Clinical Applications (Indications for Use): Fetal, Abdominal, Intra-operative (Neuro), Pediatrics, Small Organ, Neonatal Cephalic, Adult Cephalic, Trans-rectal, Trans-vaginal, Trans-urethral, Musculo-skeletal (Conventional & Superficial), Cardiac Adult, Peripheral Vessel.
-
Predicate Devices:
Non-Clinical Test Summaries (Relied upon for Substantial Equivalence):
- Acoustic Output: Evaluated according to AIUM/NEMA UD-2 and UD-3 standards, and FDA guidance "Information for Manufacturers Seeking Clearance of Diagnostic Ultrasound Systems and Transducers, FDA, CDRH, September 9, 2008". Output stays below pre-amendments upper limits (Ispta ≤ 720 mW/cm², MI ≤ 1.9, TI ≤ 6.0).
- Thermal, mechanical, and electrical safety: Tested by a recognized Certified Body and complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-37.
- Biocompatibility: Complies with EN ISO 10993-1.
- Software Lifecycle: Complies with IEC 62304.
- Specific Transducers mentioned with new indications:
- System bk2300 (general): Elastography, Vector Flow Imaging (VFI).
- Transducer 9018: New indication for Elastography (N24) for Trans-rectal and Trans-vaginal applications.
- Transducer 9051: New indication for Elastography (N24) and Vector Flow Imaging (N23) for Pediatric, Small Organ, Musculo-skeletal (Conventional), Musculo-skel. (Superficial), and Peripheral vessel applications.
- Transducer 9062: New indication for Elastography (N24) for Intra-operative, Intra-operative (Neuro), Pediatric, and Neonatal Cephalic applications.
- Transducer 9077: New indication for Continuous Wave Doppler (CWD) for Abdominal, Adult Cephalic, and Cardiac Adult applications. Also, CWD has been added as a "new mode" for these applications ("N" in the CWD column).
The document's purpose is to demonstrate substantial equivalence to existing legally marketed devices, primarily focusing on safety and technological characteristics rather than providing new clinical performance data against specific acceptance criteria.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 16, 2015
BK Medical ApS % Mrs. Randi Hauerberg Regulatory Affairs Lead Manager Mileparken 34 Herlev DK-2730 DENMARK
Re: K143298 Trade/Device Name: Ultrasound Scanner System bk2300 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: November 11, 2014 Received: November 26, 2014
Dear Mrs. Hauerberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
CEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Annroved: OMB No. 0910-0120
sweck fir rember iif known
Device Name
Ultrasound Scanner System bk2300
Indications for Use (Describe)
Ultrasound scanner and transducers for B, Tissue and Contrast Harmonic Imaging, M, PWD, CWD, Color Doppler, Vector Flow Imaging and combined mode imaging and Elastography.
Signal Analysis.
Guidance of biopsy needles, geometrical measurements and calculation of parameters. And optimal 3-D unit can reconstruct a series of 2-D imaqes into a single 3-D volume and display this on the screen.
An optional Vector Flow Imaging (VFI) module: Color Flow Mapping (CFM) imaging mode with the ability to visualize both the axial and the transverse velocity.
Clinical Applications:
- Fetal (Sometimes called Obstetrics) .
- Abdominal ●
- Intra-operative ●
- Intra-operative (Neuro) (sometimes called Neuro Surgical) ●
- Pediatrics .
- Small Organ (Sometimes called Small Parts) .
- . Neonatal Cephalic (Cephalic is sometimes called trans-cranial)
- Adult Cephalic (Cephalic is sometimes called trans-cranial) ●
- Trans-rectal
- . Trans-vaqinal
- . Trans-urethral
- Musculo-skeletal (Conventional)
- Musculo-skeletal (Superficial)
- Cardiac Adult
- Peripheral Vessel (Sometimes called Peripheral Vascular) .
| Type of Use (Select one or both, as applicable) |
|---|
| ------------------------------------------------- |
X Prescliption Use (Part 21 CFR 801 Subpart D)
Over-The Counter Use (21 CFR 801 Suppart C)
PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FOA USE ONLY
Concurrence of Certer for Devices and Radiclogical Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Recuction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response including the time to review instructions, search existing data sources, gather and maintain the dista needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burder, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Intornation Officer Paperwork Reduction Act (PRA) Staff PRAStaff@ida.hhs.gov
Ar agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number
{4}------------------------------------------------
Fill out one form for each ultrasound system and each transducer.
System: bk2300
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Modes | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks I & III) | B | M | PWD | CWD25) | TissueHarmonicImaging | ContrastHarmonicImaging26) | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined1) | Other |
| Ophthalmic | |||||||||||
| Fetal 2) | P | P | P | N | P | N | P | P | P | P | N24) |
| Abdominal 3) | P | P | P | N | P | N | P | P | P | P | N24) |
| Intra-operative (Specify 4) | P | P | P | N | P | N | P | P | P | P | N24) |
| Intra-operative (Neuro 5) | P | P | P | N | P | N | P | P | P | P | N24) |
| Laparoscopic 6) | |||||||||||
| Pediatric 7) | P | P | P | N | P | N | P | P | P | P | N24) |
| Small Organ (Specify 8) | P | P | P | N | P | N | P | P | P | P | N24) |
| Neonatal Cephalic 9) | P | P | P | N | P | N | P | P | P | P | N24) |
| Adult Cephalic 10) | P | P | P | N | P | N | P | P | P | P | N24) |
| Trans-rectal 11) | P | P | P | N | P | N | P | P | P | P | N24) |
| Trans-vaginal 12) | P | P | P | N | P | N | P | P | P | P | N24) |
| Trans-urethral 13) | P | P | P | N | P | N | P | P | P | P | N24) |
| Trans-esoph. (non-Card.) 14) | |||||||||||
| Musculo-skel. (Conventional 15) | P | P | P | N | P | N | P | P | P | P | N24) |
| Musculo-skel. (Superficial 16) | P | P | P | N | P | N | P | P | P | P | N24) |
| Intra-luminal 17) | |||||||||||
| Other (Specify 18) | |||||||||||
| Cardiac Adult 19) | P | P | P | N | P | N | P | P | P | P | |
| Cardiac Pediatric 20) | |||||||||||
| Trans-esoph. (Cardiac 21) | |||||||||||
| Other (Specify 22) | |||||||||||
| Peripheral vessel 23) | P | P | P | N | P | N | P | P | P | P | N23)N24) |
| Other (Specify 24) | P | P | P | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006 Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.
Additional Comments:
| 1. | B+M,B+D,B+C,B+D+C.B includes Tissue Harmonic ImagingD:PWD.C: Color Flow mapping Doppler incl.Amplitude(power)Doppler |
|---|---|
| Mode combinations: |
Vector Flow Imaging (VFI) 23)
- Elastography.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{5}------------------------------------------------
Fill out one form for each ultrasound system and each transducer.
System: bk2300
Transducer: 9018
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Modes | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks I & III) | B | M | PWD | CWD | TissueHarmonicImaging | ContrastHarmonicImaging | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined 1) | Other |
| Ophthalmic | |||||||||||
| Fetal 2) | |||||||||||
| Abdominal 3) | |||||||||||
| Intra-operative (Specify 4) | |||||||||||
| Intra-operative (Neuro 5) | |||||||||||
| Laparoscopic 6) | |||||||||||
| Pediatric 7) | |||||||||||
| Small Organ (Specify 8) | |||||||||||
| Neonatal Cephalic 9) | |||||||||||
| Adult Cephalic 10) | |||||||||||
| Trans-rectal 11) | P | P | P | P | P | P | P | P | N24) | ||
| Trans-vaginal 12) | P | P | P | P | P | P | P | P | N24) | ||
| Trans-urethral 13) | |||||||||||
| Trans-esoph. (non-Card.) 14) | |||||||||||
| Musculo-skel. (Conventional 15) | |||||||||||
| Musculo-skel. (Superficial 16) | |||||||||||
| Intra-luminal 17) | |||||||||||
| Other (Specify 18) | |||||||||||
| Cardiac Adult 19) | |||||||||||
| Cardiac Pediatric 20) | |||||||||||
| Trans-esoph. (Cardiac 21) | |||||||||||
| Other (Specify 22) | |||||||||||
| Peripheral vessel 23) | |||||||||||
| Other (Specify 24) | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.
Additional Comments:
| 1. | |
|---|---|
| Mode combinations: | |
| B+M, | |
| B+D, | |
| B+C, | |
| B+D+C. | |
| B includes Tissue Harmonic Imaging | |
| D:PWD. | |
| C: Color Flow mapping Doppler incl. Amplitude(power)Doppler |
- Elastography.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{6}------------------------------------------------
Fill out one form for each ultrasound system and each transducer.
System: bk2300
Transducer: 9051
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Modes | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks I & III) | B | M | PWD | CWD | TissueHarmonicImaging | ContrastHarmonicImaging | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined 1) | Other |
| Ophthalmic | |||||||||||
| Fetal 2) | |||||||||||
| Abdominal 3) | |||||||||||
| Intra-operative (Specify 4) | |||||||||||
| Intra-operative (Neuro 5) | |||||||||||
| Laparoscopic 6) | |||||||||||
| Pediatric 7) | P | P | P | P | P | P | P | P | N24) | ||
| Small Organ (Specify 8) | P | P | P | P | P | P | P | P | N24) | ||
| Neonatal Cephalic 9) | |||||||||||
| Adult Cephalic 10) | |||||||||||
| Trans-rectal 11) | |||||||||||
| Trans-vaginal 12) | |||||||||||
| Trans-urethral 13) | |||||||||||
| Trans-esoph. (non-Card.) 14) | |||||||||||
| Musculo-skel. (Conventional 15) | P | P | P | P | P | P | P | P | N24) | ||
| Musculo-skel. (Superficial 16) | P | P | P | P | P | P | P | P | N24) | ||
| Intra-luminal 17) | |||||||||||
| Other (Specify 18) | |||||||||||
| Cardiac Adult 19) | |||||||||||
| Cardiac Pediatric 20) | |||||||||||
| Trans-esoph. (Cardiac 21) | |||||||||||
| Other (Specify 22) | |||||||||||
| Peripheral vessel 23) | P | P | P | P | P | P | P | P | N23)N24) | ||
| Other (Specify 24) | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.
Additional Comments:
| 1. | Mode combinations: | B+M,B+D,B+C,B+D+C.B includes Tissue Harmonic ImagingD:PWD.C: Color Flow mapping Doppler incl.Amplitude(power)Doppler |
|---|---|---|
| ---- | -------------------- | -------------------------------------------------------------------------------------------------------------------------------------------------- |
Vector Flow Imaging (VFI) 23)
- Elastography.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR)
{7}------------------------------------------------
Fill out one form for each ultrasound system and each transducer.
System: bk2300
Transducer: 9062
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Modes | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks I & III) | B | M | PWD | CWD25) | TissueHarmonicImaging | ContrastHarmonicImaging26) | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined 1) | Other |
| Ophthalmic | |||||||||||
| Fetal 2) | |||||||||||
| Abdominal 3) | |||||||||||
| Intra-operative (Specify 4) | P | P | P | P | P | P | P | N24) | |||
| Intra-operative (Neuro 5) | P | P | P | P | P | P | P | N24) | |||
| Laparoscopic 6) | |||||||||||
| Pediatric 7) | P | P | P | P | P | P | P | N24) | |||
| Small Organ (Specify 8) | |||||||||||
| Neonatal Cephalic 9) | P | P | P | P | P | P | P | N24) | |||
| Adult Cephalic 10) | |||||||||||
| Trans-rectal 11) | |||||||||||
| Trans-vaginal 12) | |||||||||||
| Trans-urethral 13) | |||||||||||
| Trans-esoph. (non-Card.) 14) | |||||||||||
| Musculo-skel. (Conventional 15) | |||||||||||
| Musculo-skel. (Superficial 16) | |||||||||||
| Intra-luminal 17) | |||||||||||
| Other (Specify 18) | |||||||||||
| Cardiac Adult 19) | |||||||||||
| Cardiac Pediatric 20) | |||||||||||
| Trans-esoph. (Cardiac 21) | |||||||||||
| Other (Specify 22) | |||||||||||
| Peripheral vessel 23) | |||||||||||
| Other (Specify 24) | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.
Additional Comments:
| 1. | Mode combinations: | B+M,B+D,B+C,B+D+C.B includes Tissue Harmonic ImagingD:PWD.C: Color Flow mapping Doppler incl.Amplitude(power)Doppler |
|---|---|---|
| ---- | -------------------- | -------------------------------------------------------------------------------------------------------------------------------------------------- |
- Elastography.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{8}------------------------------------------------
Fill out one form for each ultrasound system and each transducer.
System: bk2300
Transducer: 9077
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Modes | ||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
| Specific(Tracks I & III) | B | M | PWD | CWD25) | TissueHarmonicImaging | ContrastHarmonicImaging26) | ColorDoppler | AmplitudeDoppler | ColorVelocityImaging | Combined 1) | Other |
| Ophthalmic | |||||||||||
| Fetal 2) | |||||||||||
| Abdominal 3) | P | P | P | N | P | P | P | P | |||
| Intra-operative (Specify 4) | |||||||||||
| Intra-operative (Neuro 5) | |||||||||||
| Laparoscopic 6) | |||||||||||
| Pediatric 7) | |||||||||||
| Small Organ (Specify 8) | |||||||||||
| Neonatal Cephalic 9) | |||||||||||
| Adult Cephalic 10) | P | P | P | N | P | P | P | P | |||
| Trans-rectal 11) | |||||||||||
| Trans-vaginal 12) | |||||||||||
| Trans-urethral 13) | |||||||||||
| Trans-esoph. (non-Card.) 14) | |||||||||||
| Musculo-skel. (Conventional 15) | |||||||||||
| Musculo-skel. (Superficial 16) | |||||||||||
| Intra-luminal 17) | |||||||||||
| Other (Specify 18) | |||||||||||
| Cardiac Adult 19) | P | P | P | N | P | P | P | P | |||
| Cardiac Pediatric 20) | |||||||||||
| Trans-esoph. (Cardiac 21) | |||||||||||
| Other (Specify 22) | |||||||||||
| Peripheral vessel 23) | |||||||||||
| Other (Specify 24) |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.
Additional Comments:
| 1. | Mode combinations: | B+M,B+D,B+C,B+D+C.B includes Tissue Harmonic ImagingD:PWD.C: Color Flow mapping Doppler incl.Amplitude(power)Doppler |
|---|---|---|
| ---- | -------------------- | -------------------------------------------------------------------------------------------------------------------------------------------------- |
- Continuous Wave Doppler (CWD/CW)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR)
{9}------------------------------------------------
510(k) Summary:
This summary is provided as part of this Premarket Notification in compliance with 21CRF, Section 807.92.
Submitters name: BK Medical Address: Mileparken 34, DK-2730 Herlev, Denmark Phone: +45 44528100 Fax: +45 44528199 Contact person: Randi Hauerberg, Regulatory Affairs Lead Manager Date prepared: November 6, 2014
Trade name: Ultrasound Scanner System bk 2300 Common name: Diagnostic Ultrasound System Classification names: Ultrasonic Pulsed Echo Imaging System (90 IYO, CFR 892.1560) Ultrasonic Pulsed Doppler Imaging System (90 IYN, CFR 892.1550) Diagnostic Ultrasonic Transducer (90 ITX, CFR 892.1570)
Identification of predicate, legally marketed device: B-K Medical Ultrasound Scanner Pro Focus 2202, K043524 B-K Medical Ultrasound Scanner Pro Focus 2202, K100919 BK Medical Ultrasound Scanner bk2300, K140428 Siemens Ultrasound Scanner System Acuson S1000, K130619
Device description:
Ultrasound scanner and transducers for B, Tissue and Contrast Harmonic Imaging, M, PWD, CWD, Color Doppler, Vector Flow Imaging and combined mode imaging and Elastography.
Signal Analysis.
Guidance of biopsy needles, geometrical measurements and calculation of parameters. And optimal 3-D unit can reconstruct a series of 2-D images into a single 3-D volume and display this on the screen.
An optional Vector Flow Imaging (VFI) module: Color Flow Mapping (CFM) imaging mode with the ability to visualize both the axial and the transverse velocity.
{10}------------------------------------------------
Transducers
All the transducers in this submission have previous been cleared.
Only the new modes
- Σ Elastography
- Σ Vector Flow Imaging (VFI).
- Continuous Wave Doppler (CWD/CW)
has been added to the functionality of relevant transducers as indicated in the Instructions For Use Section in VOL_004
There are no changes to the physical design of the transducers There are no changes to the patient contact materials. All patient contact materials are biocompatible.
All transducers used together with bk2300 are Track 3 transducers.
Acoustic output
The system controlling the Acoustic Output in the modified bk2300 is the same as the system in the predicate device Pro Focus 2202. The system will assure that the acoustic output always will stay below the pre-amendments upper limits i.e. Ispta [] 720 mW/cm² and MI [] 1.9 (Track 3, non ophthalmic).
The Thermal Index values are maximum 6.0, i.e. TI [] 6.0
Clinical measurement accuracy
Clinical measurements and calculations are described and accuracies are provided in the User Information.
Thermal, mechanical and electrical safety. The ultrasound scanner system bk2300 has been tested by a recognized Certified Body.
Acoustic Output Reporting
The Acoustic Output Reporting is made according to the standards required by "Information for Manufacturers Seeking Clearance of Diagnostic Ultrasound Systems and Transducers, FDA, CDRH, September 9, 2008"
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body.
{11}------------------------------------------------
| Summary of Technological Characteristics - Predicate Device Compared to Modified Device | |||||
|---|---|---|---|---|---|
| Devices andPredicates | Predicate deviceK140428,Ultrasoundscanner bk2300 | Predicate deviceK043524,Ultrasoundscanner ProFocus 2202 | Predicate deviceK100919,Ultrasoundscanner ProFocus 2202 | Predicate deviceK130619,Ultrasoundscanner systemAcuson S1000 | Modified device(thisapplication),Ultrasoundscanner bk2300 |
|---|---|---|---|---|---|
| Modes ofoperationRef.: [1]Appendix G | B | B | B | B | B |
| THIM,PWD,CFM 1)and combinations | THIM,PWD,CFM 1)and combinationsCHICWD, | THIM,PWD,CFM 1)and combinationsVFI | THIM,PWD,CFM 1)and combinationsElastography | THIM,PWD,CFM 1)and combinationsCHICWD,VFIElastography | |
| Intended Use: | Diagnosticultrasoundimaging or fluidflow analysis ofthe human bodyas follows: | Diagnosticultrasoundimaging or fluidflow analysis ofthe human bodyas follows: | Diagnosticultrasoundimaging or fluidflow analysis ofthe human bodyas follows: | Diagnosticultrasoundimaging or fluidflow analysis ofthe human bodyas follows: | Diagnosticultrasoundimaging or fluidflow analysis ofthe human bodyas follows: |
| Indications ForUse: | Fetal (inclObstetrics)AbdominalIntraoperativeNeurosurgery | Fetal (inclObstetrics)AbdominalIntraoperativeNeurosurgery | Fetal (inclObstetrics)AbdominalIntraoperativeNeurosurgery | Fetal (inclObstetrics)AbdominalIntraoperativeNeurosurgery | Fetal (inclObstetrics)AbdominalIntraoperativeNeurosurgery |
| Pediatrics | Pediatrics | Pediatrics | Pediatrics | Pediatrics | |
| Small Parts(organs) | Small Parts(organs) | Small Parts(organs) | Small Parts(organs) | Small Parts(organs) | |
| Adult Cephalic | Adult Cephalic | Adult Cephalic | Adult Cephalic | Adult Cephalic | |
| NeonatalCephalic | NeonatalCephalic | NeonatalCephalic | |||
| Transrectal | Transrectal | Transrectal | Transrectal | Transrectal | |
| Transvaginal | Transvaginal | Transvaginal | Transvaginal | Transvaginal | |
| Transurethral | Transurethral | Transurethral | Transurethral | Transurethral | |
| Muskulo-skeletal(conventional andsuperficial) | Muskulo-skeletal(conventional andsuperficial) | Muskulo-skeletal(conventional andsuperficial) | Muskulo-skeletal(conventional andsuperficial) | Muskulo-skeletal(conventional andsuperficial) | |
| Cardiac | Cardiac | Cardiac | Cardiac | Cardiac | |
| Peripheral Vessel | Peripheral Vessel | Peripheral Vessel | Peripheral Vessel | Peripheral Vessel | |
| Features | ECG (notmonitoring) | ECG (Notmonitoring) |
{12}------------------------------------------------
- CFM= Color Flow Mapping=Color Doppler and Amplitude (Power) Doppler.
{13}------------------------------------------------
A brief discussion of non-clinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence.
The device has been evaluated for acoustic output, thermal, electrical, electromagnetic and mechanical safety, biocompatibility and has been found to conform with applicable medical device safety standards. The system complies with the following voluntary standards:
- Σ AIUM/NEMA UD-2, Acoustic Output Measurement Standard For Diagnostic Ultrasound Equipment
- 2 AIUM/NEMA UD-3, Real-time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- A IEC 60601-1, Medical Electrical Equipment, Part 1: General requirements for safety
- IEC 60601-1-2, General requirements for safety, Collateral Standard, Electromagnetic Compatibility - Requirements and tests
- バ IEC 60601-2-37, Particular requirements for the safety of ultrasonic diagnostic medical and monitoring equipment
- ISO 14971, Application of Risk Management of Medical Devices
- EN ISO 10993-1. Biocompatibility
- IEC 62304, Medical Device Software Software lifecycle processes
- IEC 62359, Ultrasonics Field characterization Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields
Thermal, electrical, electromagnetic, mechanical and biocompatibility safety is unchanged.
Technological characteristics compared to the predicate device
The predicate devices have the same major technological characteristics as the subject device described above.
Minor differences consist with respect to the Ultrasound Scanner System: Addition of new modes.
N/A