K Number
K143281

Validate with FDA (Live)

Device Name
Ultimax-i
Date Cleared
2015-01-29

(76 days)

Product Code
Regulation Number
892.1650
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ultimax-i multipurpose digital X-ray system is designed for gastrointestinal studies, general radiography, and fluoroscopy.

The Ultimax-i system has medical applications ranging from but not limited to: contrast enhanced studies, support of endoscopic studies, nonvascular contrast-enhanced studies, nonvascular contrast-enhanced studies, support of vascular IVR, and general radiography.

Note: This system is not intended for cardiovascular contrast studies or interventional radiology procedures for the cardiac or cerebral blood vessels. This system is not intended for mammography studies in the US.

Device Description

This device is a fixed c-arm fluoroscopic device intended to provide radiographic and fluoroscopic images in a variety of studies. The unit consists of a c-arm with x-ray tube, beam limiter and digital flat panel detector (FPD); a patient table that can tilt in both directions, an 80kW x-ray generator, an image processor and both remote and table side control.

AI/ML Overview

The provided text does not contain a detailed study with acceptance criteria and device performance metrics in the format requested. The document is a 510(k) premarket notification letter and a summary of safety and effectiveness for the Toshiba Ultimax-i, an image-intensified fluoroscopic x-ray system.

It states that the device is "substantially equivalent" to predicate devices (Toshiba KALARE and Siemens Artis Zee Multi-purpose) and that "Comparison of image performance criteria demonstrate that the two devices perform in a similar manner. Clinical data was used to further demonstrate the effectiveness of the Ultimax-i system." However, it does not provide specific acceptance criteria, reported performance values, or details about the clinical study (like sample size, ground truth, experts used, or MRMC study results).

Here's a breakdown of what can be extracted and what is missing based on your request:

1. A table of acceptance criteria and the reported device performance

  • Missing. The document generally states that "Comparison of image performance criteria demonstrate that the two devices perform in a similar manner," but it does not specify what those criteria are or what the actual reported performance values are.

2. Sample sized used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Missing. The document mentions "clinical images" and "clinical data" were used to support efficacy and compare performance but does not specify the sample size or the provenance of this data.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

  • Missing. The document does not describe how ground truth was established or if any experts were involved in a formal capacity for a test set.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

  • Missing. There is no mention of an adjudication method.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • Missing. The document does not describe an MRMC comparative effectiveness study, nor does it mention AI assistance. The device is an X-ray system, not an AI-driven image analysis tool.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

  • Missing. This question is not directly applicable as the device is an imaging system, not a standalone algorithm. Performance is typically assessed on image quality and system functionality, not algorithmic output in isolation.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

  • Missing. The document mentions "clinical data" was used but does not specify the type of ground truth used (e.g., if it was a comparison to existing high-quality images, pathological findings, or patient outcomes).

8. The sample size for the training set

  • Missing. The document does not refer to a "training set" in the context of machine learning, as this is an imaging device submission.

9. How the ground truth for the training set was established

  • Missing. See response to #8.

Summary of available information:

  • Device Name: Ultimax-i
  • Purpose of Study/Testing: To demonstrate substantial equivalence to predicate devices (Toshiba KALARE, Siemens Artis Zee Multi-purpose) and confirm that changes to the table and FPD mounting assembly maintain performance.
  • Tests Conducted: Bench testing (for detector installation and specifications), and "additional testing... to compare the performance between the predicate device (KALARE) and the modified device that included testing directed at image quality, artifacts and motion/dynamic capabilities."
  • Support for Efficacy: "Clinical images in this submission to support the efficacy of the new device" and "Clinical data was used to further demonstrate the effectiveness of the Ultimax-i system."
  • Standards Followed: IEC60601-1 and its collateral standards (IEC60601-2-54, IEC60601-2-43, IEC 60601-2-28), 21 CFR Subchapter J, 21 CFR § 820, and ISO 13485.

In conclusion, while the document confirms testing was performed to demonstrate substantial equivalence and efficacy, it does not provide the detailed study information (specific acceptance criteria, sample sizes, expert involvement, and ground truth methodologies) that your request outlines. This level of detail is typically found in the full submission documents, not in the summary or FDA letter.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 29, 2015

Toshiba Medical Systems Corporation % Mr. Paul Biggins U.S. Agent/Director Regulatory Affairs Toshiba America Medical System, Inc. 2441 Michelle Drive TUSTIN CA 92780

Re: K143281

Trade/Device Name: Ultimax-i Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: OWB, JAA Dated: November 14, 2014 Received: November 17, 2014

Dear Mr. Biggins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Robert A Ochs

Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143281

Device Name Ultimax-i

Indications for Use (Describe)

The Ultimax-i multipurpose digital X-ray system is designed for gastrointestinal studies, general radiography, and fluoroscopy.

The Ultimax-i system has medical applications ranging from but not limited to: contrast enhanced studies, support of endoscopic studies, nonvascular contrast-enhanced studies, nonvascular contrast-enhanced studies, support of vascular IVR, and general radiography.

Note: This system is not intended for cardiovascular contrast studies or interventional radiology procedures for the cardiac or cerebral blood vessels. This system is not intended for mammography studies in the US.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) - SUMMARY OF SAFETY AND EFFECTIVENESS

1. CLASSIFICATION and DEVICE NAME:

Classification Name:Image Intensified Fluoroscopic X-ray System
Regulation Number:21 CFR 892.1650 (Class II)
Product CodeOWB-Interventional Fluoroscopic X-ray System (Primary)JAA - Image Intensified Fluoroscopic X-ray System(Secondary)
Trade Proprietary Name:Ultimax-i
Model Number:DREX-UI80

2. ESTABLISHMENT REGISTRATION: 9614698

3. CONTACT PERSON, U.S. AGENT and ADDRESS:

Contact Person/U.S. Agent:

Paul Biggins Director, Regulatory Affairs (714) 730-5000 Fax (714) 730-1310 pbiggins@tams.com

Establishment Name and Address:

Toshiba America Medical Systems, Inc. (TAMS) 2441 Michelle Drive Tustin, Ca. 92780

4. MANUFACTURING SITE

Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan Contact: Akinori Hatanaka

5. Date OF SUBMISSION:

November 13, 2014

6. PERFORMANCE STANDARD:

21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard

7. PREDICATE DEVICE:

Toshiba, KALARE (DREX-KL80) , K133553

{4}------------------------------------------------

Siemens Medical Systems, Artis Zee Multi-purpose K122664

8. REASON FOR SUBMISSION:

New device

9. SUBMISSION TYPE:

Traditional 510(k)

10. DEVICE DESCRIPTION:

This device is a fixed c-arm fluoroscopic device intended to provide radiographic and fluoroscopic images in a variety of studies. The unit consists of a c-arm with x-ray tube, beam limiter and digital flat panel detector (FPD); a patient table that can tilt in both directions, an 80kW x-ray generator, an image processor and both remote and table side control.

11. SUMMARY OF INTENDED USES:

The Ultimax-i multipurpose digital X-ray system is designed for gastrointestinal studies, vascular studies, general radiography, and fluoroscopy.

The Ultimax-i system has medical applications ranging from but not limited to: contrastenhanced studies, support of endoscopic studies, nonvascular contrast-enhanced studies, nonvascular IVR, vascular contrast-enhanced studies, support of vascular IVR, and general radiography.

Note: This system is not intended for cardiovascular contrast studies or interventional radiology procedures for the cardiac or cerebral blood vessels.

12. SUMMARY OF CHANGE(S)

The Ultimax has the same Image processor and employs the base software of the KALARE. Introduction of changes to the table and FPD mounting assembly have been implemented to provide an expansion to the type of studies performed on the Toshiba KARALE system. These changes are a new table and c-arm configuration for the mounting of the image receptor (digital flat panel detector) and the x-ray source.

This device is similar to the Siemens Artis Zee Multi-purpose as follows:

  • . Similarity in intended uses
  • Both designs are centered around an integrated C-arm and table configuration. .
  • Has similar movement of both table and FPD/tube assembly. ●
  • Both systems are floor mounted c-arm interventional x-ray systems ●

13. SUBSTANTIAL EQUIVALENCE:

This device is substantially equivalent to the KALARE, K133553, marketed by Toshiba America Medical Systems. Both systems have the same intended use as general fluoroscopy/radiographic systems. Ultimax-i includes modifications to the cleared device which adds functionality to the table and FPD movement, base software and

{5}------------------------------------------------

manufacturing process remain unchanged from the cleared device. Comparison of image performance criteria demonstrate that the two devices perform in a similar manner. Clinical data was used to further demonstrate the effectiveness of the Ultimax-i system.

This device is substantially equivalent to the Siemens Artis Zee Multi-purpose room in that the hardware configurations and thus intended uses are the similar.

14. SAFETY:

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 standards, its collateral standards and particular standards; IEC60601-2-54, IEC60601-2-43 and IEC 60601-2-28. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR §1020, that apply to this device, will be met and reported via product report.

15. TESTING

Testing of the modified system was conducted in accordance with the applicable standards published by the International Electromechanical Commission (IEC) for Medical Devices and XR Systems. Bench testing was done to confirm that the installation of the detector met the stated specifications of the component manufacturer. Additional testing was done to compare the performance between the predicate device (KALARE) and the modified device that included testing directed at image quality, artifacts and motion/dynamic capabilities. The sponsor has included clinical images in this submission to support the efficacy of the new device. Additionally, in support of this conclusion the detector component has been the subject of other 510k Premarket Notifications where the clinical aspects of the component have been addressed.

16. CONCLUSION

The Ultimax-i has the same intended use as the predicate devices. Safety and effectiveness have been verified via risk management and the application of design controls to the modifications.

§ 892.1650 Image-intensified fluoroscopic x-ray system.

(a)
Identification. An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.