K Number
K143238

Validate with FDA (Live)

Device Name
C3 Wave
Date Cleared
2015-04-24

(163 days)

Product Code
Regulation Number
880.5970
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The C3 Wave System is indicated for use as a supplemental aid in positioning for Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location information by using the patient's cardiac electrical activity. Confirmation of tip placement should be verified according to clinical judgement and established hospital protocol (e.g., Chest X-Ray, Fluoroscopy). Note: Limiting, but not contraindicated, situations for this technique are patients where cardiac rhythms may change presentation of the P wave: atrial fibrillation, atrial flutter, severe tachycardia, pacemaker driven rhythm, and cronic obstructive pulmonary disease (COPD). Such patients are easily identified prior to PICC insertion. Use of additional confirmation method is necessary to confirm catheter tip location.

Device Description

The C3 Wave system includes the iPad® monitor running the mobile application software, the C3 hub (a battery and power supply cord for the hub), a remote control, and an ECG clip cable (alligator clip). Procedural accessories include the ECG snap leads, ECG patient cable, ECG electrodes, remote cover, and prep pads which are provided as a convenience to the clinician.

AI/ML Overview

The provided text describes the C3 Wave device and its regulatory submission, but it does not contain the detailed study information required to fully answer your request.

Specifically, the document states:

  • "Design verification and validation activities were performed in accordance with Design Control requirements per 21 CFR 820.30 and demonstrate that the subject C3 Wave system meets predetermined performance specifications."
  • "The performance evaluation plan included testing per the following recognized standards to assess conformance to IEC 60601-1 (3rd Edition). IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety - IEC 60601-1-2 Medical Electrical Equipment - Part 2: General Requirements for Basic Safety and Essential Performance - Collateral Standard Electro Magnetic Compatibility - Requirements and Test"

However, it does not provide:

  • A table of acceptance criteria for its core function (PICC tip location) and the reported device performance against those criteria.
  • Details on the sample size for any clinical test set or its provenance.
  • The number or qualifications of experts used for ground truth.
  • Adjudication methods.
  • Whether MRMC, standalone, or training set studies were performed, or the details of these if conducted.
  • The type of ground truth used for assessing PICC tip location accuracy.

The document focuses on regulatory compliance, electrical safety standards (IEC 60601 series), and biocompatibility testing for materials that may contact the patient. These are important for device safety and general performance but do not detail the clinical performance validation for its primary "PICC tip location" function.

Therefore, I cannot fully complete the table or answer all your questions with the provided text.

Here's what can be extracted and what is missing:


Acceptance Criteria and Study Details for C3 Wave System

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (for PICC Tip Location Accuracy)Reported Device Performance
Not specified in textNot specified in text

Note: The document mentions performance testing relating to electrical safety (IEC 60601 series) and biocompatibility (ISO 10993-1), which the device passed. However, it does not provide acceptance criteria or performance data for the primary function of "real-time catheter tip location information by using the patient's cardiac electrical activity" in terms of accuracy or precision relative to a ground truth.

2. Sample size used for the test set and the data provenance:

  • Test Set Sample Size: Not specified for clinical efficacy.
  • Data Provenance: Not specified for clinical efficacy.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not specified.

4. Adjudication method for the test set:

  • Not specified.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, and the effect size of how much human readers improve with AI vs without AI assistance:

  • Not specified. The device is for "real-time catheter tip location information" and acts as a "supplemental aid," but no details of human-in-the-loop performance studies are provided.

6. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

  • Not specified for clinical performance. The functional descriptions imply standalone capability to detect P-wave changes, but explicit standalone clinical accuracy metrics are not provided.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

  • Not specified for clinical efficacy. The Indications for Use state, "Confirmation of tip placement should be verified according to clinical judgement and established hospital protocol (e.g., Chest X-Ray, Fluoroscopy)," implying these might be used as ground truth in a clinical study, but no study details are provided.

8. The sample size for the training set:

  • Not specified.

9. How the ground truth for the training set was established:

  • Not specified.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular emblem with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 24, 2015

Medcomp Ms. Courtney Nix Regulatory Associate 1499 Delp Drive Harleysville, PA 19438

Re: K143238

Trade/Device Name: C3 Wave Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, implanted, long-term intravascular catheter Regulatory Class: II Product Code: LJS Dated: March 6, 2015 Received: March 10, 2015

Dear Ms. Nix:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina
Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Form Approved: OMB No. 0910-0120
Expiration Date: January 31, 2017
See PRA Statement below.

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

Indications for Use

510(k) Number (if known)K143238
Device NameC3 Wave
Indications for Use (Describe)The C3 Wave System is indicated for use as a supplemental aid in positioning for Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location information by using the patient's cardiac electrical activity. Confirmation of tip placement should be verified according to clinical judgement and established hospital protocol (e.g., Chest X-Ray, Fluoroscopy). Note: Limiting, but not contraindicated, situations for this technique are patients where cardiac rhythms may change presentation of the P wave: atrial fibrillation, atrial flutter, severe tachycardia, pacemaker driven rhythm, and cronic obstructive pulmonary disease (COPD). Such patients are easily identified prior to PICC insertion. Use of additional confirmation method is necessary to confirm catheter tip location.
Type of Use (Select one or both, as applicable) Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)
-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)Page 1 of 1
-----------------------------------

PSC Publishing Services (301) 443-6740

EF

{3}------------------------------------------------

1499 Delp Drive

Harlevsville, PA 19438

Tel: 215-256-4201

Fax. 215-256-1787

www.medcompnet.com

Section 5510(k) SUMMARYTraditional 510K
K143238
A.Submitter Information:
Submitter:Medcomp®1499 Delp DriveHarleysville, PA 19438
Contact:Tel: (215) 256-4201, x 2285Fax: (215) 256-9191Courtney NixCnix@medcompnet.comRegulatory Associate
Date Prepared:11/11/2014
B.Trade Name:C3 Wave
Common Name:PICC placement accessory
Classification Name:Long Term Intravascular Catheter (80 LJS)
Regulation Name:Percutaneous, implanted, long-term intravascular catheter
C.F.R. Section:21 CFR 880.5970
Class:II
C.Predicate Devices:K140799 Medcomp, Celerity™ System

D. Device Description:

The C3 Wave system includes the iPad® monitor running the mobile application software, the C3 hub (a battery and power supply cord for the hub), a remote control, and an ECG clip cable (alligator clip). Procedural accessories include the ECG snap leads, ECG patient cable, ECG electrodes, remote cover, and prep pads which are provided as a convenience to the clinician.

E. Intended Use:

The C3 Wave System is intended to provide real time tip location information of a central venous catheter by utilization of ECG to observe P-wave changes as the tip approaches the right atrium of the heart via the superior vena cava.

แ Indications for Use:

The C3 Wave System is indicated for use as a supplemental aid in positioning for Peripherally Inserted Central Catheters (PICC) in adult patients. It provides real-time catheter tip location information by using the patient's cardiac electrical activity. Confirmation of tip placement should be verified according to clinical judgement and established hospital protocol (e.g., Chest X-Ray, Fluoroscopy). Note: Limiting, but not contraindicated, situations for this technique are patients where cardiac rhythms may change presentation of the P wave: atrial fibrillation, atrial flutter, severe tachycardia, pacemaker driven rhythm, and cronic obstructive pulmonary disease (COPD).

Such patients are easily identified prior to PICC insertion. Use of additional confirmation method is necessary to confirm catheter tip location.

  • G. Comparison to Predicate Devices:
    Medcomp C3 Wave system Summary Page 1 of 2

{4}------------------------------------------------

1499 Delp Drive

Harlevsville, PA 19438

H.

Tel: 215-256-4201

Fax. 215-256-1787

www.medcompnet.com

The C3 Wave system is substantially equivalent to the predicate device in terms of intended use, indication for use, anatomical location, basic design, performance, and labeling.

The difference between the C3 Wave system and the predicate device is in the use of Bluetooth® technology to digitally transfer the ECG waveform to the mobile application running on the iPad® monitor, and also the use of Bluetooth® technology in the remote control which allows the clinician to control the system without breaking the sterile field. The use of Bluetooth® technology eliminates the need to have cables running from the patient to the monitor and improves patient comfort.

Performance Testing:

Design verification and validation activities were performed in accordance with Design Control requirements per 21 CFR 820.30 and demonstrate that the subject C3 Wave system meets predetermined performance specifications.

The performance evaluation plan included testing per the following recognized standards to assess conformance to IEC 60601-1 (3rd Edition).

IEC 60601-1 Medical Electrical Equipment - Part 1: General Requirements for Safety

  • IEC 60601-1-2 Medical Electrical Equipment - Part 2: General Requirements for Basic Safety and Essential Performance - Collateral Standard Electro Magnetic Compatibility - Requirements and Test

​ដែ Biocompatibility:

C3 Wave patient contact is limited to the ECG cables. The C3 hub can have limited patient contact if the clinician places the hub on the patient's chest, however the hub can perform the exact same function if placed beside the patient therefore having no patient contact. As a result, biocompatibility testing was conducted on the C3 Wave hub and ECG cables (ECG Alligator Clip and Power Supply Cable) according to ISO 10993-1. The following test results demonstrate that the material used meets ISO 10993 requirements for an external communicating device with circulating blood path:

CYTOTOXICITY:

Cytotoxicity Test. Neutral Red Uptake 4 concentration- ISO- Passed-A: 14-03214-N1

SENSITIZATION:

B: Kligman Maximization - 2 extracts - Passed - Report -14-02840-G2

IRRITATION:

C: ISO Intracutaneous Injection - 2 extracts - Passed - 14-02840-G4

SYSTEMIC TOXICITY:

D: ISO Rabbit Pyrogen (material mediated) - Passed - Report- 14-02840-G5

CHEMICAL CHARACTERIZATION:

Medcomp C3 Wave system Summary Page 2 of 2

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the MedComp logo. The logo features a stylized, geometric shape in pink, resembling a house or an abstract letter. Below the shape, the word "medCOMP" is written in a stylized font, with "med" in pink and "COMP" in black. The logo appears to be clean and modern.

  • 1499 Delp Drive
    Harleysville, PA 19438

K.

Tel: 215-256-4201

Fax. 215-256-1787

www.medcompnet.com

  • E: Chemical Characterization test performed on the C3 Wave Hub and results were as expected.

J. Technological Characteristics:

Technological Characteristics of the C3 Wave system are equivalent to the predicate device with respect to the basic system design but differ in the technology employed to accomplish the same adjunct function. Even with the technology differences between the predicate and proposed devices, no new questions of effectiveness are raised.

Summary of Substantial Equivalence:

The proposed device is substantially equivalent to the predicate device based on:

  • The Intended Use .
  • Indications for Use .
  • e Basic system design
  • o Operating principle
  • o Results of performance testing

§ 880.5970 Percutaneous, implanted, long-term intravascular catheter.

(a)
Identification. A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.(b)
Classification. Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”