K Number
K143236

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-07-02

(232 days)

Product Code
Regulation Number
866.3305
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Theranos™ HSV-1 IgG Assay is a chemiluminescent immunoassay intended for the qualitative detection of IgG antibodies to herpes simplex virus type 1 (HSV-1) in human serum, in K2-EDTA anticoagulated human plasma from venous blood, and in human fingerstick K2-EDTA anticoagulated whole blood obtained with the Theranos Capillary Tubes and Nanotainer Tubes. The test is indicated for sexually active individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-1 infection. The predictive value of positive and negative results depends on the population's prevalence and the pretest likelihood of HSV-1.

The test is not FDA cleared for screening blood or plasma donors. The performance of this assay has not been established for use in a pediatric population, neonates and immunocompromised patients.

The Theranos HSV-1 IgG Assay is for use with the Theranos System which performs automated sample processing steps and result analysis.

Device Description

The Theranos HSV-1 IgG Assay is for use with the Theranos System. The Theranos System performs automated sample processing steps and analysis to produce the test results.

The Theranos HSV-1 IgG Assay is a three-step sandwich immunoassav with an HSV-1 glycoprotein G (gG) recombinant antigen coated surface, an anti-human IgG detection reagent conjugated to alkaline phosphatase (AP) and chemiluminescent substrate. During the first incubation step, the HSV-1 IgG antibodies present in the positive control and sample bind to the gG recombinant antigen on the coated surface. Following the first incubation step, unbound materials are removed with a wash cycle. Then the detection reagent-AP conjugate is added and during the second incubation step, the detection reagent-AP conjugate reacts with the HSV-1 IgG antibodies already bound to the capture surface. Following the second incubation, unbound materials are removed with a wash cycle. The chemiluminescent substrate is added to the capture-analyte-detection complex during the third incubation step to initiate the chemiluminescence reaction. Light generated by this reaction is detected and analyzed by the Theranos System using a calibration function to determine the cut-off index (COI) values for the sample and controls. The results for the Positive and Negative controls must be within specified limits for a run to be considered valid.

AI/ML Overview

The Theranos Herpes Simplex Virus-1 (HSV-1) IgG Assay is a chemiluminescent immunoassay for the qualitative detection of IgG antibodies to HSV-1 in human serum, K2-EDTA anticoagulated human plasma from venous blood, and human fingerstick K2-EDTA anticoagulated whole blood. It is intended to aid in the presumptive diagnosis of HSV-1 infection in sexually active individuals and expectant mothers.

Here's an analysis of the acceptance criteria and supporting study details:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state a consolidated table of acceptance criteria for all performance aspects. However, based on the studies described, we can infer some key criteria and the reported performance. The most critical performance metrics for a diagnostic assay like this are Sensitivity and Specificity in the intended use populations.

Performance MetricAcceptance Criteria (Inferred from Study Design & Conclusions)Reported Device Performance (Sexually Active Adults)Reported Device Performance (Pregnant Women)Reported Device Performance (Low Prevalence Population)
SensitivityNot explicitly stated but expected to be high for diagnostic aid. Implied acceptable range is within 95% Confidence Interval established by the study.95.1% (90.3-97.6% CI)97.9% (94.8-99.2% CI)97.0% (84.7-99.5% CI)
SpecificityNot explicitly stated but expected to be high for diagnostic aid. Implied acceptable range is within 95% Confidence Interval established by the study.97.4% (92.7-99.1% CI)95.2% (89.3-98.0% CI)100% (92.7-100% CI)
Precision (Venous Serum)Max %CV of 14.2% across different conditions (inferred from study conclusion).10.2% to 14.2%N/AN/A
Precision (Fingerstick Whole Blood)Max %CV of 12.2% across different conditions (inferred from study conclusion).N/A10.9% to 12.2%N/A
Reproducibility (Total Imprecision)%CV of 11.4% for COI >= 0.5, SD of 0.022 for COI < 0.5 (inferred from study conclusion).N/A%CV=11.4% (COI ≥ 0.5), SD=0.022 (COI < 0.5)N/A
Analyte Stability (Averaged Difference, COI > 0.5)< ±10%< 3.2% (largest average difference observed was -3.2% for Venous serum at room temp, 6 hrs)< 3.2% (largest average difference observed was -3.2% for Venous serum at room temp, 6 hrs)< 3.2% (largest average difference observed was -3.2% for Venous serum at room temp, 6 hrs)
Analyte Stability (Largest Observed Difference, COI > 0.5)< 15%< 13.9% (largest observed percent difference)< 13.9% (largest observed percent difference)< 13.9% (largest observed percent difference)
Analyte Stability (Averaged Difference, COI < 0.5)< 0.02 COI< 0.021 (largest average difference observed was 0.021 for Venous K2-EDTA plasma at 3 freeze/thaw cycles)< 0.021 (largest average difference observed was 0.021 for Venous K2-EDTA plasma at 3 freeze/thaw cycles)< 0.021 (largest average difference observed was 0.021 for Venous K2-EDTA plasma at 3 freeze/thaw cycles)
Analyte Stability (Largest Observed Difference, COI < 0.5)< 0.08 COI< 0.037 (largest observed difference)< 0.037 (largest observed difference)< 0.037 (largest observed difference)
Interfering Substances (Low Positive/High Negative)Mean recovery within +/- 20% of unspiked sample.All showed signal change < 15%.All showed signal change < 15%.All showed signal change < 15%.
Interfering Substances (Negative Pool)Deviation of < 0.02 COI.All showed mean deviation < 0.02 COI, except Intralipid (0.03 COI).All showed mean deviation < 0.02 COI, except Intralipid (0.03 COI).All showed mean deviation < 0.02 COI, except Intralipid (0.03 COI).
Cross-reactivityNo cross-reactivity with common infectious agents and conditions (i.e., Theranos HSV-1 Negative or Equivocal if reference is negative for HSV-1).Only one equivocal for Treponema pallidum and one positive for Rheumatoid Factor, but systematic cross-reactivity ruled out.Only one equivocal for Treponema pallidum and one positive for Rheumatoid Factor, but systematic cross-reactivity ruled out.Only one equivocal for Treponema pallidum and one positive for Rheumatoid Factor, but systematic cross-reactivity ruled out.
Cut-off Validation (NPA)Not explicitly stated but the reported 95% CI (86.3-98.9) indicates acceptance.96.0% (47/49)N/A (tested on serum samples to validate cut-off generally, not specific populations)N/A
Cut-off Validation (PPA)Not explicitly stated but the reported 95% CI (90.3-99.2) indicates acceptance.97.1% (69/71)N/A (tested on serum samples to validate cut-off generally, not specific populations)N/A
Method Comparison (Sensitivity, Fingerstick Plasma)Not explicitly stated but the reported 95% CI (86.8 - 99.6) indicates acceptance.97.4% (38/39)N/AN/A
Method Comparison (Specificity, Fingerstick Plasma)Not explicitly stated but the reported 95% CI (85.1 - 100) indicates acceptance.100% (22/22)N/AN/A
CDC Panel Agreement100% agreement.100% agreement100% agreement100% agreement

2. Sample Sizes and Data Provenance:

  • Clinical Performance Study (Sexually Active Adults): 260 retrospectively collected, archived venous serum samples. Samples obtained from multiple specimen sources covering 10 US states and Mexico.
  • Clinical Performance Study (Pregnant Women): 298 retrospectively collected, archived venous serum samples. Samples obtained from multiple specimen sources covering 10 US states and Mexico.
  • Low Prevalence Population Study: 82 samples. Samples obtained from multiple specimen sources covering 10 US states and Mexico.
  • CDC Panel Testing: 100 well-characterized serum samples. Origin: U.S. Centers for Disease Control and Prevention (CDC).
  • Precision (Venous Serum): 6 serum samples, with varying replicates (72-94 total replicates per panel member). One site (CLIA Laboratory model).
  • Precision (Fingerstick Whole Blood): 3 fingerstick plasma samples, with 170-171 total replicates per panel member. One site (CLIA Laboratory model).
  • Reproducibility (Fingerstick Whole Blood): 30 subjects (10 subjects at each of 3 collection sites). From each subject, 9 Capillary Tubes and Nanotainer Tubes and 2 serum separator tubes (SSTs) were collected.
  • Analyte Stability: Not explicitly stated but implied sufficient samples to represent different baseline COI values and matrices.
  • Interfering Substances: 3 serum samples (negative, high negative, low positive) contrived from a high positive sample and pooled negative serum. Each pool tested in duplicate.
  • Cross-reactivity: 21 different infectious agents/conditions, with at least 3 samples per agent (total of 139 samples). Samples acquired from commercial vendors.
  • Assay Cut-off Establishment/Validation: 192 serum samples to establish, then 120 independent serum samples to validate.

3. Number of Experts and Qualifications for Ground Truth:

The document primarily relies on reference methods for ground truth, rather than independent experts for adjudication of individual cases.

  • Clinical Performance (Sexually Active Adults & Pregnant Women): The FOCUS HerpeSelect Immunoblot was used as the reference method. When initial results were equivocal on the Focus HerpeSelect Immunoblot, a validated western blot reference test (University of Washington, Seattle) was used for resolution.
  • CDC Panel Testing: The CDC itself provided a panel of "well characterized serum samples" (implies expert characterization and consensus by the CDC).
  • Cross-reactivity: Banked serum samples "confirmed positive for IgG against the infectious agents of interest" and "negative for HSV-1 IgG on the reference method" were acquired from commercial vendors. The implicit experts are those who characterized these commercial samples and the reference method used.

No specific number or qualifications of experts are given beyond the implicit expertise associated with the CDC and the University of Washington laboratory performing the western blot.

4. Adjudication Method:

  • Clinical Performance Studies: For equivocal results on the primary reference method (Focus HerpeSelect Immunoblot), a secondary reference method (University of Washington western blot) was used. This acts as a 2+1 adjudication method, where the primary reference method's equivocal cases are resolved by a more definitive test. Specifically, for sexually active adults, 10 samples were initially equivocal, and a western blot resolved 9 (2 negative, 7 positive), with 1 unresolved. For pregnant women, 8 samples were initially equivocal, with 4 resolved by western blot (1 negative, 3 positive) and 4 unresolved due to insufficient volume.

5. Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study:

  • No MRMC study was done. This device is an automated, standalone diagnostic assay (algorithm only) that provides a qualitative result (positive, negative, equivocal). It does not involve human readers interpreting results in the loop, or any comparison of human readers with vs. without AI assistance.

6. Standalone Performance:

  • Yes, a standalone (algorithm only) performance study was done. The entire submission describes the performance of the Theranos HSV-1 IgG Assay and the Theranos System (which performs automated sample processing and result analysis) as a standalone device. The device itself generates the final qualitative result (positive, negative, equivocal) without human interpretation of the assay signal.

7. Type of Ground Truth Used:

  • The ground truth for the clinical performance studies (Sexually Active Adults, Pregnant Women, Low Prevalence Population) was established by a reference method, specifically the FOCUS HerpeSelect Immunoblot, with equivocal results resolved by a validated western blot reference test (from the University of Washington, Seattle).
  • For the CDC panel testing, the ground truth was based on the CDC's characterization of the serum samples.
  • For cross-reactivity, ground truth was based on third-party validated positive samples for other infectious agents and negative for HSV-1.

8. Sample Size for the Training Set:

The document does not explicitly state the sample size for a "training set." This type of regulatory submission (510(k)) focuses on validation rather than the development and training phases of an algorithm. Therefore, details about the internal development (training and internal validation) of the device's algorithm are typically not included or required in this format. The performance studies discussed are considered validation studies.

9. How the Ground Truth for the Training Set Was Established:

As no "training set" details are provided in this document, the method for establishing its ground truth is also not described. The document focuses on the validation of the device's performance against established reference methods.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features the department's emblem, which consists of a stylized representation of human profiles arranged in a cascading manner. The emblem is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 7, 2015

THERANOS, INC. BRAD ARINGTON ASSOCIATE DIRECTOR, REGULATORY 1701 PAGE MILL ROAD PALO ALTO, CA 94304

Re: K143236

Trade/Device Name: Theranos Herpes Simplex Virus-1 IgG Assav Regulation Number: 21 CFR 866.3305 Regulation Name: Herpes simplex virus serological assays Regulatory Class: II Product Code: MXJ Dated: June 29, 2015 Received: June 30, 2015

Dear Mr. Arington:

This letter corrects our substantially equivalent letter of July 2, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the

{1}------------------------------------------------

electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Sally A. Hojvat -S

Sally A. Hojvat, M.Sc., Ph.D. Director Division of Microbiology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143236

Device Name Theranos Herpes Simplex Virus-1 IgG Assay

Indications for Use (Describe)

The Theranos™ HSV-1 IgG Assay is a chemiluminescent immunoassay intended for the qualitative detection of IgG antibodies to herpes simplex virus type 1 (HSV-1) in human serum, in K2-EDTA anticoagulated human plasma from venous blood, and in human fingerstick K2-EDTA anticoagulated whole blood obtained with the Theranos Capillary Tubes and Nanotainer Tubes. The test is indicated for sexually active individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-1 infection. The predictive value of positive and negative results depends on the population's prevalence and the pretest likelihood of HSV-1.

The test is not FDA cleared for screening blood or plasma donors. The performance of this assay has not been established for use in a pediatric population, neonates and immunocompromised patients.

The Theranos HSV-1 IgG Assay is for use with the Theranos System which performs automated sample processing steps and result analysis.

Type of Use (Select one or both, as applicable)

☑ Prescription Use (Part 21 CFR 801 Subpart D)
☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY K143236

I. GENERAL INFORMATION

Submitter: Theranos, Inc. 1701 Page Mill Road Palo Alto, CA 94304 Phone: 650-838-9292 Fax: 650-838-9165

Contact Person: Brad Arington Associate Director, Regulatory Phone: 650-856-7304 650-838-9165 Fax: Email: barington@theranos.com

Date Prepared: June 29, 2015

II. DEVICE INFORMATION

Trade Name:Theranos™ Herpes Simplex Virus-1 (HSV-1) IgG Assay
Common Name:HSV-1 IgG assay
Regulation Number:21 CFR§866.3305
Regulation Name:Herpes simplex virus serological assays
Regulatory Class:Class II
Product Code:MXJ (Enzyme Linked Immunosorbent Assay, Herpes Simplex Virus, HSV-1)
Panel:Microbiology (83)

III. PREDICATE DEVICE

HerpeSelect® 1 and 2 Immunoblot IgG (K000238; Focus Diagnostics, Inc.)

{4}------------------------------------------------

IV. DEVICE DESCRIPTION

The Theranos HSV-1 IgG Assay is for use with the Theranos System. The Theranos System performs automated sample processing steps and analysis to produce the test results.

The Theranos HSV-1 IgG Assay is a three-step sandwich immunoassav with an HSV-1 glycoprotein G (gG) recombinant antigen coated surface, an anti-human IgG detection reagent conjugated to alkaline phosphatase (AP) and chemiluminescent substrate. During the first incubation step, the HSV-1 IgG antibodies present in the positive control and sample bind to the gG recombinant antigen on the coated surface. Following the first incubation step, unbound materials are removed with a wash cycle. Then the detection reagent-AP conjugate is added and during the second incubation step, the detection reagent-AP conjugate reacts with the HSV-1 IgG antibodies already bound to the capture surface. Following the second incubation, unbound materials are removed with a wash cycle. The chemiluminescent substrate is added to the capture-analyte-detection complex during the third incubation step to initiate the chemiluminescence reaction. Light generated by this reaction is detected and analyzed by the Theranos System using a calibration function to determine the cut-off index (COI) values for the sample and controls. The results for the Positive and Negative controls must be within specified limits for a run to be considered valid.

V. INDICATIONS FOR USE

The Theranos™ HSV-1 IgG Assay is a chemiluminescent immunoassay intended for the qualitative detection of IgG antibodies to herpes simplex virus type 1 (HSV-1) in human serum, in K2-EDTA anticoagulated human plasma from venous blood, and in human fingerstick K2-EDTA anticoagulated whole blood obtained with the Theranos Capillary Tubes and Nanotainer Tubes. The test is indicated for sexually active individuals and expectant mothers as an aid in the presumptive diagnosis of HSV-1 infection. The predictive value of positive and negative results depends on the population's prevalence and the pretest likelihood of HSV-1.

The test is not FDA cleared for screening blood or plasma donors. The performance of this assay has not been established for use in a pediatric population, neonates and immunocompromised patients.

The Theranos HSV-1 IgG Assay is for use with the Theranos System which performs automated sample processing steps and result analysis.

{5}------------------------------------------------

COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE VI.

Table 1: Similarities between the Theranos HSV-1 IgG Assay and the Predicate

CharacteristicTheranos™ HSV-1 IgG Assay(K143236)Focus HerpeSelect® 1 and 2Immunoblot IgG(K000238)
Intended useQualitative test to detect presence orabsence of IgG antibodies to HSV-1in human serum, in K2-EDTAanticoagulated human plasma fromvenous blood, and in humanfingerstick K2-EDTA anticoagulatedwhole blood obtained with theTheranos Capillary Tubes andNanotainer Tubes. Indicated fortesting sexually active individuals andexpectant mothers as an aid inpresumptive diagnosis of HSV-1infection. The predictive value ofpositive or negative results dependson the population's prevalence and thepretest likelihood of HSV-1 infection.Qualitative test to detect presence orabsence of IgG antibodies to HSV-1 andHSV-2 in human sera. Indicated fortesting sexually active adults or expectantmothers as an aid in presumptivediagnosis of HSV-1 and HSV-2 infection.The predictive value of positive ornegative results depends on thepopulation's prevalence and the pretestlikelihood of HSV-1 and HSV-2 infection.

Table 2: Differences between the Theranos HSV-1 IgG Assay and the Predicate

CharacteristicTheranos™ HSV-1 IgG Assay(K143236)Focus HerpeSelect® 1 and 2Immunoblot IgG(K000238)
SpecimenTypesVenous serum, K2-EDTA anticoagulated human plasma from venous blood, Human fingerstick K2-EDTA anticoagulated whole blood obtained with the Theranos Capillary Tubes and Nanotainer TubesVenous Serum
Type of AssayChemiluminescent immunoassayNitrocellulose Immunoblot
SampleHandlingAutomated sample handling/processingManual sample handling/processing
CaptureReagentHSV-1 recombinant antigen (gG1)HSV-1/HSV-2 antigen immobilized on nitrocellulose membrane

VII. PERFORMANCE

To characterize performance of the Theranos HSV-1 IgG immunoassay the following studies were conducted:

Precision - CLIA Laboratory Model, Venous Serum

{6}------------------------------------------------

A study for estimating the precision of the Theranos HSV-1 IgG Assay for venous serum samples in a CLIA Laboratory model was performed by testing a panel of 6 serum samples spanning the analytical range [negative (A), high negative (B), equivocal (C), low positive (D), moderate positive (E), and positive (F)]. The precision study was conducted at one site with thirty five (35) TSPU devices, three (3) lots of cartridges and sixteen (16) operators in total. The study duration was 13 days in total. Details of the study design for different samples are presented in Table 3 below.

PanelValid ReplicatesNo. ofNo. ofNo. ofNo. of Invalid
MemberTotalLot 1Lot 2Lot 3DevicesDaysOperatorsReplicates
A(Neg.)91263827357143
B(High Neg.)88243727287142
C(Equivocal)7827448*358163
D(Low Pos.)8025272811244
E(Mod. Pos.)6425132613261
F(Pos.)6925192515243

Table 3: Design of Precision Study: Numbers of Replicates, Devices, Days and Onerators

*Sufficient cartridges from reagent lot #3 were not available.

Results of the precision study are presented in Table 4.

PanelMemberMean(COI)Repeatability(same device,same lot)Between-deviceBetween-lotPrecision(samedevice,differentlot)Precision(differentdevice, samelot)Precision(differentdevice,differentlot)
A(Neg.)0.425SD0.0490.0070.0000.0490.0490.049
%CV11.5%1.6%0%11.5%11.6%11.6%
BHigh Neg.)0.648SD0.0860.0110.0290.0910.0870.092
%CV13.3%1.7%4.5%14.1%13.4%14.2%
CEquivocal)1.016SD0.0930.0620.0650.1130.1120.129
%CV9.1%6.1%6.4%11.1%11.0%12.7%
D(Low Pos.1.727SD0.2080.0980.0130.2080.2300.230
%CV12.0%5.7%0.8%12.0%13.3%13.3%
E(Mod. Pos.)3.809SD0.3050.2760.1080.3240.4110.425
%CV8.0%7.3%2.8%8.5%10.8%11.2%
F(Pos.)8.996SD0.8070.4370.0000.8070.9180.918
%CV9.0%4.9%0.0%9.0%10.2%10.2%

Table 4: Summary of Precision Study Results

{7}------------------------------------------------

Table 5 presents percent of invalid results and percents of negative, equivocal and positive among valid results for each sample.

PanelMemberMean(COI)Number ofReplicatesPercent ofInvalidPercent ofNegativeamongValidPercent ofEquivocalamongValidPercent ofPositiveamongValid
A(Neg.)0.425943.2%(3/94)100%(91/91)
B(High Neg.)0.648902.2%(2/90)100%(88/88)
C(Equivocal)1.016813.7%(3/81)17.9%(14/78)60.3%(47/78)21.8%(17/78)
D(Low Pos.)1.727844.8%(4/84)100%(80/80)
E(Mod. Pos.)3.809651.5%(1/65)100%(64/64)
F(Pos.)8.996724.2%(3/72)100%(69/69)

Table 5: Percent of Invalid Results and Percents of Negative, Equivocal and Positive among Valid Results

The results of the study demonstrate that the precision of the Theranos HSV-1 IgG Assay (including different TSPU devices, different lots of cartridges, and different operators) when performed in a CLIA Laboratory was in the range 10.2% to 14.2%.

Precision - CLIA Laboratory Model, Fingerstick Whole Blood

A study for estimating the precision of the Theranos HSV-1 IgG Assay for fingerstick whole blood samples in a CLIA Laboratory model was performed by testing a panel of 3 fingerstick plasma samples spanning the analytical range [high negative (P), equivocal (Q), moderate positive (R)]. The precision study was conducted at one site with thirty six (36) TSPU devices, three (3) lots of cartridges and nine (9) operators in total. The study duration was 4 days in total. Details of the study design for different samples are presented in Table 6 below.

Table 6: Design of Precision Study: Numbers of Replicates, Devices, Days and Operators

PanelMemberValid ReplicatesNo. ofDevicesNo. ofDaysNo. ofOperatorsNo. of InvalidReplicates
TotalLot 1Lot 2Lot 3
P(High Neg.)16856565630493*
Q(Equivocal)16856565629492*
R(Mod. Pos.)16856565627492*

*All invalid replicates were repeated

{8}------------------------------------------------

Table 7: Summary of Precision Study Results
PanelMemberMean(COI)Repeatability(same device,same lot)Between-deviceBetween-lotPrecision(samedevice,differentlot)Precision(differentdevice,same lot)Precision(differentdevice,differentlot)
P(High Neg.)0.888SD0.0830.0060.0500.0960.0830.097
%CV9.3%0.7%5.6%10.9%9.3%10.9%
Q(Equivocal)1.047SD0.0940.0250.0690.1170.0980.119
%CV9.0%2.4%6.6%11.1%9.3%11.4%
R(Mod. Pos.)3.241SD0.3420.1220.1570.3770.3630.396
%CV10.6%3.8%4.9%11.6%11.2%12.2%

Results of the precision study are presented in Table 7.

Table 8 presents percent of invalid results and percents of negative, equivocal and positive among valid results for each sample.

PanelMemberMean(COI)Number ofReplicatesPercent ofInvalidPercent ofNegativeamong ValidPercent ofEquivocalamong ValidPercent ofPositiveamongValid
P(High Neg.)0.8881711.8%(3/171)58.3%(98/168)40.5%(68/168)1.2%(2/168)
Q(Equivocal)1.0471701.2%(2/170)6.5% (11/168)63.1%(106/168)30.4%(51/168)
R(Mod. Pos.)1.0161701.2%(2/170)100%(168/168)

Table 8: Percents of Positive. Equivocal. Negative and Invalid Results

The results of the study demonstrate that precision of the Theranos HSV-1 IgG Assay (including different TSPU devices, different lots of cartridges, and different operators) when performed in a CLIA Laboratory was in the range from 10.9% to 12.2%.

Reproducibility

A study designed to process multiple fingerstick whole blood samples from individual subjects was performed to evaluate the reproducibility of the Theranos HSV-1 IgG Assay when used with Theranos Capillary Tubes and Nanotainer Tubes. The study was conducted at 3 collection sites with 10 subjects at each site. From each of 30 subjects, 9 Capillary Tubes and Nanotainer Tubes from 3 manufacturing lots (i.e. 3 Capillary Tubes and Nanotainer Tubes per lot) and 2 serum separator tubes (SSTs) were collected. Each subject had the following measurements:

  • Each of the 9 Capillary Tubes and Nanotainer Tubes was tested. These data were . used for evaluation of between-Capillary Tubes and Nanotainer Tubes imprecision, between-lot imprecision and total imprecision that includes between-Capillary Tubes and Nanotainer Tubes and between-lot imprecisions.
  • One Nanotainer Tube (from one of the 3rd lot of Capillary Tubes and Nanotainer Tubes for each subject) was tested in duplicate via recovering a sample from one

{9}------------------------------------------------

Capillary Tubes and Nanotainer Tubes and transferring a sample to another Capillary Tubes and Nanotainer Tubes. These data were used for evaluation of within-Capillary Tubes and Nanotainer Tubes imprecision.

  • . Each of the 2 SSTs was tested. These data were used for evaluation of between-SST imprecision.
    For samples with mean COI value at the baseline ≥0.5, percent differences were calculated and for samples with mean COI value at the baseline <0.5, differences were calculated. Table 9 summarizes the results of the reproducibility study broken down by collection site and by high or low COI subjects; the variability metrics are averaged across all subjects within the site.

{10}------------------------------------------------

Capillary Tubes and Nanotainer TubesSST
CollectionSiteSubjectsWithin-CapillaryTubes andNanotainerTubes(%CV orSD)Between-Capillary Tubesand NanotainerTubes(%CV orSD)Between-Lot(%CV orSD)Total(%CV orSD)Between-SST(%CV orSD)
16 subjectswith COIvalues1.4 - 13.5%CV=6.0%%CV = 9.0%%CV = 6.8%%CV=12.6%%CV=9.6%
4 subjectswith COIvalues0.03-0.28SD=0.008SD = 0.015SD = 0.016SD = 0.024SD=0.11
27 subjectswith COIvalues1.6 – 16.8%CV=8.2%%CV=9.2%%CV=3.2%%CV=10.8%%CV=12.5%
3 subjectswith COIvalues0.07-0.19SD=0.009SD=0.011SD=0.008SD=0.015SD=0.019
35 subjectswith COIvalues4.5 - 14.3%CV=8.2%%CV=8.1%%CV=6.0%%CV=11%%CV=12.4%
5 subjectswith COIvalues0.02-0.32SD=0.08SD=0.019SD=0.013SD=0.025SD=0.021
Combined18 subjectswith COIvalues1.4-16.8%CV=7.5%%CV=8.8%%CV=5.2%%CV=11.4%%CV=11.5%
12 subjectswith COIvalues0.02-0.32SD=0.008SD=0.015SD=0.013SD=0.022SD=0.017

Table 9: Summary of Results of the Reproducibility Study

{11}------------------------------------------------

  • . Within-Capillary Tubes and Nanotainer Tubes imprecision was %CV=7.5% for aggregated subjects with mean COI ≥ 0.5 and SD=0.008 for aggregated subjects with a mean COI < 0.5.
  • . Total imprecision including between-Capillary Tubes and Nanotainer Tubes and between-lot imprecisions was %CV= 11.4% for aggregated subjects with a mean COI ≥ 0.5 and SD=0.022 for aggregated subjects with a mean COI < 0.5.
  • Between-serum separator tubes imprecision was %CV=11.5% for aggregated subjects with a mean COI ≥ 0.5 and SD=0.017 for aggregated subjects with a mean COI < 0.5.

Analyte Stability

An analyte stability study was performed to characterize the stability of HSV-1 IgG in clinical matrices as measured by the Theranos HSV-1 IgG Assay under different sample storage conditions and time periods. A summary of analyte storage conditions and durations for different sample types and matrices claimed for the Theranos HSV-1 IgG Assay is presented in Table 10.

ConditionVenousSerumVenousK2-EDTAPlasmaFingerstickK2-EDTAPlasmaFingerstickK2-EDTAWhole Blood
Stored at 2-8°C48 hr48 hr48 hr48 hr
Stored at room temperature(20-25°C)6 hr6 hr6 hr6 hr
Stored at -20°C1 week1 week1 weekN/A
Freeze/thaw cycles333N/A
Table 10: Summary of Analyte Storage Conditions and Durations
Sample Type and MatrixSamples with a Baseline COI < 0.5Samples with a Baseline COI > 0.5
Difference averaged over all samplesThe largest observed difference among samplesPercent difference averaged over all samplesThe largest observed percent difference among samples
Stored at 2-8C, 48 hrsVenous serum0.0060.0061.0%13.6%
Venous K2-EDTA plasma-0.007-0.0072.3%13.3%
Fingerstick K2-EDTA plasma from whole blood0.0030.003-0.4%13.9%
Stored at -20, 1 weekVenous serum0.0080.0150.8%13.3%
Venous K2-EDTA plasma0.0030.005-1.0%12.7%
Fingerstick K2-EDTA plasma from whole blood0.0010.0081.8%-13.9%
Freeze thaw cycles, n=3Venous serum0.0070.021-0.1%13.6%
Venous K2-EDTA plasma0.0210.037-1.7%-13.4%
Fingerstick K2-EDTA plasma from whole blood0.0060.022-1.0%13.6%
Stored at room temp, 6 hrsVenous serum-0.001-0.011-3.2%-11.9%
Venous K2-EDTA plasma0.0020.0220.1%13.7%
Fingerstick K2-EDTA plasma from whole blood-0.004-0.0261.1%13.9%

Within 2 hours after collection, one aliquot of each sample type or matrix was tested with the Theranos HSV-1 IgG Assay in duplicate, to establish the value at baseline. The samples were stored in Nanotainer Tubes under the appropriate conditions. Comparison of an average of two replicates at the predetermined time points with the average of two replicates at baseline was performed. For samples with a mean COI value at the baseline ≥0.5, percent differences were calculated and for samples with a mean COI value at the baseline <0.5. differences were calculated.

Acceptance criteria were as follows: i) a difference averaged over all samples with baseline mean COI value >0.5 must be less than ±10% and a difference averaged over all samples with baseline COI mean <0.5 must be less than 0.02 and ii) for each sample, an observed difference must be less than 15% for the samples with baseline mean COI value >0.5 and must be less than 0.08 for the samples with baseline mean COI value <0.5 (the range of differences expected if there is no effect of storage on the HSV-1 IgG analyte). The results are summarized in Table 11.

{12}------------------------------------------------

Table 11: Summary of Mean Absolute Difference Measures for all Storage Conditions and Sample Types or Matrices ______________________________________________________________________________________________________________________________________________________________________________ ــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــــ

{13}------------------------------------------------

Interfering Substances

A study was designed and performed (in accordance with CLSI EP07-A2) to evaluate the performance of the Theranos HSV-1 IgG Assav in the presence of potentially interfering substances to assess the impact of these endogenous substances and commonly used drugs on the performance of the Theranos HSV-1 IgG Assay. Interferents were tested with three serum samples (negative (mean COI 0.024), high negative (mean COI 0.77) and low positive (mean COI 1.52)) that were contrived by using a high positive sample and diluting it with pooled negative serum. Samples were spiked with the interferent at levels shown in Table 12. Each serum pool was tested in duplicate.

For the low positive and the high negative pools, the acceptance criteria were a mean recovery within +/- 20% of the value of the unspiked sample (i.e., in the absence of the potential interferent or drug). All low positive and high negative samples showed a signal change of less than 15% for all interfering substances. All positive samples remained positive and all negative samples remained negative upon spiking of drug or other interferents. For the negative pool, the acceptance criterion was a deviation of less than 0.02 COI. All negative samples showed a mean deviation of <0.02 COI, except Intralipid. Intralipid spikes did not show any effect on recovery for near cut-off samples, high negative and low positive samples.

Negative PoolHigh Negative PoolLow Positive Pool
InterferentLevelMeanCOIΔCOIMeanCOI%RecoveryMeanCOI%Recovery
Hemoglobin1000 mg/dL0.0250.000.69901.71113
Bilirubin20 mg/dL0.0240.000.68881.61106
Intralipid2000 mg/dL0.0530.030.811051.60105
Acetylcysteine150 mg/L0.019-0.0040.68881.4092
Ampicillin-Na1000 mg/L0.0250.0010.76991.4495
Ascorbic acid300 mg/L0.0270.0030.75971.67110
Ca-Dobesilate200 mg/L0.0270.0040.70911.5199
Cyclosporine5 mg/L0.0310.0080.74971.53101
Cefoxitin2500 mg/L0.0270.0030.74971.52100
Heparin5000U0.020-0.0030.801031.52100

Table 12: Summary of Interfering Substances Studies: Endogenous Interferents and Drug Interferents

{14}------------------------------------------------

Negative PoolHigh Negative PoolLow Positive Pool
InterferentLevelMeanCOIΔCOIMeanCOI%RecoveryMeanCOI%Recovery
Levodopa20 mg/L0.0300.0060.68881.4294
Methyldopa+1.5h2020 mg/L0.0240.0000.74971.3790
Metronidazole200 mg/L0.0390.0160.74961.3891
Phenylbutazone400 mg/L0.021-0.0020.74961.4294
Doxycycline50 mg/L0.0240.0000.71921.3589
Acetylsalicylic acid1000 mg/L0.0260.0020.75971.3790
Rifampicin60 mg/L0.014-0.0090.69901.3589
Acetaminophen200 mg/L0.0340.0100.64831.68111
Control0.0240.0000.771001.52100

Cross-reactivity

A study was performed to evaluate the performance of the Theranos HSV-1 IgG Assay in the presence of IgG antibodies against twenty-one (21) infectious agents defined as potential cross-reactants in the FDA guidance on HSV serological assays. Banked serum samples confirmed positive for IgG against the infectious agents of interest were acquired from commercial vendors. At least three (3) samples, independently confirmed as positive for that agent and negative for HSV-1 IgG on the reference method, were tested on the Theranos HSV-1 IgG Assay in order to rule out cross-reactivity of the Theranos HSV-1 IgG Assay with IgG against a potential cross reactant. The results of this study are displayed in Table 13.

{15}------------------------------------------------

Organism/ConditionNo.ReferenceHSV-1 AssayTheranosHSV-1PositiveTheranosHSV-1NegativeTheranosHSV-1Equivocal
Epstein Barr Virus (IgG)6Negative060
Epstein Barr Virus (IgM)1Negative010
HPV4Negative040
Rubella (IgG)13Negative0130
HSV-2 (IgG)40Negative0400
HAMA samples4Negative040
Treponema pallidum8Negative071*
Rheumatoid Factor (RF)8Negative1**70
Anti-nuclear antibody (ANA)8Negative080
Sjogren's Syndrome3Negative030
CMV (IgG)5Negative050
CMV (IgM)2Negative020
Chlamydia trachomatis (IgG)10Negative0100
HCV (IgG)3Negative030
HBsAg3Negative030
VZV IgG5Negative050
Measles IgG5Negative050
HIV-1 (IgG)4Negative040
Toxoplasma IgG4Negative040
Candida albicans Ag3Negative030
Systemic Lupus3Negative030

Table 13: Summary of Cross-reactivity Study on Theranos HSV-1 IgG Assay

*Systematic cross-reactivity ruled out (7/8 samples in same category tested negative)

**Confirmed as positive upon retest by Theranos HSV-1 assay, systematic cross-reactivity ruled out (7/8 samples in same category
tested negative)

{16}------------------------------------------------

Assay Cut-off

A study was performed to establish the cut-off and the limits of the equivocal zone for the Theranos HSV-1 IgG Assay using 192 serum samples. Then 120 independent serum samples were analyzed to validate the established cut-off. The calibrators were assigned COI values based on the established assay cut-off. the cut-off for positive results a COI of 1.1 and cut-off for negative results a COI of 0.9. The results of the cut-off validation study are displayed in Table 14 below.

Agreement ClassificationPercent Agreement95% ConfidenceInterval
NPA96.0% (47/49)86.3-98.9
PPA97.1% (69/71)90.3-99.2

Fingerstick Plasma - CLIA Laboratory Model

To demonstrate the performance of the Theranos HSV-1 IgG Assay for fingerstick whole blood samples collected at 3 Theranos Patient Service Centers (TPSCs) and processed at the CLIA-certified laboratory.

At each site, fingerstick whole blood samples were collected into a pair of Theranos Capillary Tubes and Nanotainer Tubes, and venous samples were collected into serum tubes from each of 20, 16 and 25 adult subjects at three collection sites.

Samples were shipped refrigerated to the Theranos CLIA-certified laboratory in Palo Alto. CA. Upon receipt, fingerstick whole blood samples in the Nanotainer Tubes were centrifuged at 1200g for 5 minutes. Plasma was extracted and processed and analyzed on the Theranos System. All samples were processed or frozen as plasma within 48 hours of draw.

A summary of the performance information is shown in Table 15.

{17}------------------------------------------------

Table 15: Summary of Method Comparison for Samples Collected at 3 Theranos Patient Service Centers

Reference Result
POSNEG
Theranos ResultPOS380
NEG122
Point Estimate95% Confidence Interval

97.4% (38/39)

100% (22/22)

Matrix Comparison

Sensitivity

Specificity

The effect of anticoagulants and different sample types (fingerstick and venous) on the performance of the Theranos HSV-1 IgG Assay was determined by comparing matched venous serum, venous K2-EDTA plasma, and fingerstick K2-EDTA plasma samples from 70 donors. Forty-three matched sample sets were contrived to have analyte values close to the cut-off. The acceptance criterion was a recovery of positive plasma samples within ± 20% of the corresponding serum reference value (serum drawn into primary tubes without gel). For negative samples, the acceptance criteria was a difference of <= 0.02 COI from the corresponding serum value. All anticoagulant-treated plasma samples met this criterion. Weighted Deming regression was performed. The slope and an intercept of the regression line and their 95% confidence intervals along with correlation coefficients are shown in Table 16 and a graphical depiction is shown in Figure 1.

86.8 - 99.6

85.1 - 100

{18}------------------------------------------------

Table 16: Summary of Weighted Deming Regression Analysis Performed on Matrix
Equivalency Data for Venous Plasma and Fingerstick Plasma Samples
SampleType/MatrixCorrelationcoefficientSlope95% confidenceinterval onslopeIntercept95% confidenceinterval on intercept
Venous plasma0.9920.993[0.967, 1.019]0.000[-0.003, 0.003]
Fingerstick plasma0.9951.009[0.973, 1.044]-0.003[-0.006, -0.001]

Figure 1: Regression Analysis for Matrix Comparison Study

Image /page/18/Figure/5 description: The image contains two scatter plots comparing serum reference values with venous plasma and fingerstick plasma, respectively. Both plots show a weighted Deming regression with a high Pearson's r value, indicating a strong positive correlation. The regression equation for venous plasma is y = 0.993x + 0, with an intercept confidence interval of [0, 0] and a slope confidence interval of [0.97, 1.02], and the Pearson's r is 0.992. For fingerstick plasma, the regression equation is y = 1.009x + -0.003, with an intercept confidence interval of [-0.01, 0] and a slope confidence interval of [0.97, 1.04], and the Pearson's r is 0.995.

Expected Values

The Theranos HSV-1 IgG Assay was used to evaluate the prevalence of HSV-1 IgG antibodies in individuals for whom an HSV-1 IgG test was ordered by a physician including pregnant women. The study population for the Theranos HSV-1 IgG Assay consisted of a total of 558 subjects, with 260 sexually active adults and 298 individuals identified as pregnant women. The result for 1 out of the 558 subjects is not reported, as indicated in Table 17 (1 subject), giving a total of 557 subjects. The data for the intended use population (557 specimens) have been summarized according to age group in decades, gender, number of reactive results, number of equivocal results, and number of non-reactive results. The data for the intended use population have been summarized in Table 17 (259 specimens from sexually active adult subjects) and Table 18 (298 specimens from pregnant subjects).

{19}------------------------------------------------

AgeRangeGenderReactiveN/Total (%)EquivocalN/Total (%)Non-ReactiveN/Total (%)
16 to 19Male0/0 (0)0/0 (0)0/0 (0)
16 to 19Female1/4 (25)0/4 (0)3/4 (75)
20 to 29Male8/18 (44.4)1/18 (5.6)9/18 (50)
20 to 29Female29/73 (39.7)0/73 (0)44/73 (60.3)
30 to 39Male5/10 (50)0/10 (0)5/10 (50)
30 to 39Female33/62 (53.2)0/62 (0)29/62 (46.8)
40 to 49Male5/10 (50)0/10 (0)5/10 (50)
40 to 49Female16/27 (59.3)0/27 (0)11/27 (40.7)
50 to 59Male17/20 (85)0/20 (0)3/20 (15)
50 to 59Female9/11 (81.8)0/11 (0)2/11 (18.2)
60 to 69Male5/6 (83.3)0/6 (0)1/6 (16.7)
60 to 69Female5/10 (50)1/10 (10)4/10 (40)
70 to 79Male3/4 (75)0/4 (0)1/4 (25)
70 to 79Female1/3 (33.3)0/3 (0)2/3 (66.7)
80 to 89Male0/0 (0)0/0 (0)0/0 (0)
80 to 89Female1/1 (100)0/1 (0)0/1 (0)
Total*138/259 (53.3)2/259 (0.8)119/259 (45.9)

Table 17: Expected Results for Theranos HSV-1 IgG Assay in Sexually Active Adult Subjects

*1 sample not reported since age information was not available

AgeRangeGenderReactiveN/Total (%)EquivocalN/Total (%)Non-ReactiveN/Total (%)
18 to 19Female13/13 (100)0/13 (0)0/13 (0)
20 to 29Female114/175 (65.1)1/175 (0.6)60/175 (34.3)
30 to 39Female61/104 (58.7)0/104 (0)43/104 (41.3)
40 to 49Female5/6 (83.3)0/6 (0)1/6 (16.7)
Total193/298 (65)1/298 (0.3)104/298 (35)

The hypothetical positive and negative predictive values (PPV, NPV) for the two intended use populations are shown in Table 19. The calculations are based on the specificity and sensitivity values for the Theranos HSV-1 IgG Assay determined in the clinical study;

    1. Specificity of 97.4% and Sensitivity of 95.1% in sexually active adults
    1. Specificity of 95.2% and Sensitivity of 97.4% in pregnant women

{20}------------------------------------------------

Prevalence(%)Sexually Active AdultsPregnant Women
PPV (%)NPV (%)PPV (%)NPV (%)
5093.892.692.191.7
4593.293.291.392.4
4092.493.890.393.0
3591.494.289.193.5
3090.194.687.594.0
2588.394.985.394.3
2085.895.282.394.7
1582.095.577.795.0
1075.295.769.995.2
560.296.053.895.5

Table 19: Hypothetical Predictive Values

Clinical Performance in the Intended Use Populations (CLIA Laboratory Model)

A clinical study was conducted to characterize the performance of the Theranos HSV-1 IgG Assay in the Theranos CLIA-certified Laboratory in comparison to the FOCUS HerpeSelect Immunoblot (as the reference method for performance analysis).

Prospectively collected, archived venous serum samples collected from pregnant women and sexually active adults (18 years and older) who had a prescription for a HSV-1 IgG test. Samples were obtained from multiple specimen sources covering 10 US states and Mexico.

The equivocal results on the Focus HerpeSelect Immunoblot (that repeatedly tested equivocal) were resolved using a validated western blot reference test (University of Washington, Seattle) as per the instructions of the package insert for the reference method. Ten samples from the sexually active adult sub-population tested initially equivocal on the Focus HerpeSelect Immunoblot and were resolved by the University of Washington western blot as 2 negatives and 7 positives. One sample was not resolved. There were no samples in the sexually active adult sub-population that returned an invalid result.

In the pregnant women sub-population. 8 samples tested initially equivocal on the Focus HerpeSelect immunoblot. Of these, 4 samples could not be resolved by the University of Washington western blot due to insufficient volume availability. Of the remaining 4, 1 (one) were resolved as negative and 3 as positive. There were 3 samples that returned an invalid result on the Theranos HSV-1 IgG test. These samples were rerun and resulted in

{21}------------------------------------------------

valid results included in the analysis. The clinical performance results are summarized in Tables 20 and 21.

Table 20: Summary of Theranos HSV-1 IgG Assay Performance in Sexually Active Adult Population

Reference Method
PositiveEquivocalNegativeTotal
TheranosHSV-1IgGAssayPositive13702139
Equivocal1012
Negative51113119
Total1431116260
Point Estimate95% Confidence Interval
Sensitivity95.1% (137/144)90.3-97.6
Specificity97.4% (113/116)92.7-99.1

Table 21: Summary of Theranos HSV-1 IgG Assay Performance in Pregnant Women Population

PositiveEquivocalNegativeTotal
TheranosHSV-1IgGAssayPositive18814193
Equivocal0101
Negative22100104
Total1904104298
Point Estimate95% Confidence Interval
Sensitivity97.9% (188/192)94.8-99.2
Specificity95.2% (100/105)89.3-98.0

Reference Method

{22}------------------------------------------------

CDC Panel Testing

The objective of this study was to demonstrate agreement of the Theranos HSV-1 IgG Assay with the CDC panel. A panel of well characterized serum samples (n=100) was obtained from the U.S. Centers for Disease Control and Prevention (CDC). The CDC sample panel was tested with the HSV-1 IgG Assay and the results obtained by Theranos were sent to the CDC for confirmation. The panel consisted of 54 positives and 46 negatives. The Theranos HSV-1 IgG Assay demonstrated 100% agreement with the results provided by the CDC.

Low Prevalence Population

Serum samples were collected from a low prevalence population: Individuals who are not sexually active, and without a recent or current sexually transmitted disease (Hepatitis, Syphilis, HIV, HPV, Trichomonas, Chlamydia, Gonorrhoeae) as determined in an interview. Performance of the assay on this population is summarized in Table 22. The Samples were obtained from multiple specimen sources covering 10 US states and Mexico.

Table 22: Summary of Theranos HSV-1 IgG Assay Performance with Low Prevalence Population

Reference Method
PositiveEquivocalNegativeTotal
TheranosHSV-1 IgGAssayPositive320032
TheranosHSV-1 IgGAssayEquivocal0000
TheranosHSV-1 IgGAssayNegative014950
TheranosHSV-1 IgGAssayTotal3214982

Doforonao Mothod

Point Estimate95% Confidence Interval
Sensitivity97.0% (32/33)84.7-99.5
Specificity100% (49/49)92.7-100

CONCLUSIONS

The results of the analytical and clinical performance studies submitted in this premarket notification are complete and demonstrate that the Theranos HSV-1 IgG Assay meets the established specifications necessary for consistent performance during intended clinical use. The results support a decision that the Theranos Herpes Simplex Virus-1 (HSV-1) IgG Assay is substantially equivalent to the predicate.

§ 866.3305 Herpes simplex virus serological assays.

(a)
Identification. Herpes simplex virus serological assays are devices that consist of antigens and antisera used in various serological tests to identify antibodies to herpes simplex virus in serum. Additionally, some of the assays consist of herpes simplex virus antisera conjugated with a fluorescent dye (immunofluorescent assays) used to identify herpes simplex virus directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by herpes simplex viruses and provides epidemiological information on these diseases. Herpes simplex viral infections range from common and mild lesions of the skin and mucous membranes to a severe form of encephalitis (inflammation of the brain). Neonatal herpes virus infections range from a mild infection to a severe generalized disease with a fatal outcome.(b)
Classification. Class II (special controls). The device is classified as class II (special controls). The special control for the device is FDA's revised guidance document entitled “Class II Special Controls Guidance Document: Herpes Simplex Virus Types 1 and 2 Serological Assays.” For availability of the guidance revised document, see § 866.1(e).