K Number
K143216

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-03-17

(127 days)

Product Code
Regulation Number
870.2700
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Assurance™ Nasal / Alar SpO2 Sensor is indicated for single patient use for continuously noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients (weighing > 30kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision.

Device Description

The Alar / Nasal SpO2 Sensor is a disposable, single patient use Pulse Oximetry sensor designed to attach to the patient's nasal alar region - the fleshy region at the side of the nose. Skin contact and adhesive free sensor retention is via soft silicone rubber cushions encapsulating the optical components. The Alar / Nasal SpO2 Sensor with its associated patient cable, terminates in a DB-9 connector compatible with monitors employing Nellcor SpO2 technology.

The sensor utilizes red and IR LED light sources of 660 nm and 880 nm respectively along with a silicon photodiode detector to detect changes in oxygen saturation in the blood. Since oxygen saturated blood absorbs different amounts of light at each wavelength (red and infrared) as compared with unsaturated blood, the amount of light absorbed at each wavelength by the blood in each pulse can be used to calculate oxygen saturation.

AI/ML Overview

The provided document describes the Assurance™ Alar / Nasal SpO2 Sensor. Here's a breakdown of the acceptance criteria and study information:


1. A table of acceptance criteria and the reported device performance

Acceptance Criteria (Predicate Devices - 70-100%)Reported Device Performance (Proposed Device - 70-100%)
SpO2 Accuracy (A_RMS): ± 2%SpO2 Accuracy (A_RMS): ± 3% (ISO 80601-2-61 allows up to ± 3%)
BPM (Xhale Alar SpO2 sensor): 30-250 bpm ± 3 bpmBPM: 30-240 bpm ± 3 bpm

Note: The SpO2 A_RMS for the proposed device and one of the predicate devices is listed as ± 3%. However, elsewhere in the document, the predicate Xhale Alar SpO2 sensor is listed with an A_RMS of 70-100% ± 2%. The document specifically emphasizes that the proposed device's performance within ± 3% meets the ISO 80601-2-61 standard, which allows for this range.


2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

  • Sample Size for Test Set (Clinical Testing):
    • The "Clinical Testing" section mentions a "controlled desaturation study with healthy volunteers."
    • The table detailing SpO2 Accuracy for different oximetry platforms shows varying "pts" (points or patients) for each platform, ranging from 257 pts to 279 pts. It's unclear if "pts" refers to individual volunteers or data points. Given the context of a desaturation study with healthy volunteers, it likely refers to subjects, or instances of data collection from subjects. The overall study used healthy volunteers.
  • Data Provenance: Not explicitly stated (e.g., country of origin, retrospective or prospective). However, the nature of a "controlled desaturation study with healthy volunteers" implies a prospective design.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

This information is not provided in the document. The study involves a controlled desaturation study where arterial blood gas measurements (or a similar reference method) would typically establish the true SpO2. The ground truth for SpO2 measurements in such studies is usually derived from a co-oximeter analyzing arterial blood samples, not from expert interpretation of images or other data.


4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable to this type of device and study. Adjudication methods like 2+1 or 3+1 are typically used in studies involving expert review of medical images or data where there might be inter-reader variability in interpreting findings. For a pulse oximeter, the ground truth for SpO2 is a quantitative measurement, usually from a laboratory gold standard (e.g., co-oximetry of arterial blood), not subject to expert adjudication.


5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

There was no MRMC comparative effectiveness study and no mention of AI assistance. This device is a medical sensor, not an AI-powered diagnostic tool requiring human reader interpretation.


6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

This refers to the performance of the device itself (the sensor and its underlying algorithm) in measuring SpO2. The "Clinical Testing" section describes a controlled desaturation study where the device's SpO2 accuracy (A_RMS) was measured against a reference standard. This is essentially a standalone performance test for the algorithm within the device. The results are reported as A_RMS values for the device with various compatible monitors.


7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for the clinical SpO2 accuracy study would have been established by a reference co-oximeter performing arterial blood gas analysis, which is the gold standard for measuring arterial oxygen saturation. The document mentions "controlled desaturation study," which is the standard methodology for assessing SpO2 accuracy against a highly accurate reference.


8. The sample size for the training set

This information is not provided and is likely not applicable in the context of this 510(k) submission. Pulse oximeters operate on well-established spectrophotometric principles, and their algorithms are typically designed based on physics and physiological models, rather than trained on large datasets in the way that machine learning models are. If any calibration or tuning was done, the details are not disclosed as a "training set."


9. How the ground truth for the training set was established

As there's no mention of a traditional "training set" for a machine learning model, this information is not applicable. The device's operation is based on fundamental principles of light absorption by oxygenated and deoxygenated hemoglobin.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 17, 2015

Xhale Inc. c/o Mr. Paul Dryden Regulatory Consultant 3630 SW 47th Avenue, Suite 100 Gainesville, FL 32608

Re: K143216

Trade/Device Name: Assurance™ Alar / Nasal SpO2 Sensor Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: II Product Code: DQA Dated: February 12, 2015 Received: February 13, 2015

Dear Mr. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tejashri Purohit-Sheth, M.D.

Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR

Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K143216

Device Name

Assurance™ Nasal / Alar SpO2 Sensor

Indications for Use (Describe)

The Assurance™ Nasal / Alar SpO2 Sensor is indicated for single patient use for continuously noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2) and pulse rate from the nasal ala of adult and pediatric patients (weighing > 30kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision.

Type of Use (Select one or both, as applicable)

XX Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

FORM FDA 3881 (9/13)

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

{3}------------------------------------------------

510(k) Summary Page 1 of 6

Date Prepared:16-Mar-15
Xhale, Inc.
3630 SW 47th Ave, Suite 100
Gainesville, FL 32608
Phone: (352) 371-8488
Fax: (352) 375-3133
Official Contact:Jeffrey Hoebelheinrich
Director of Quality and Regulatory Affairs
Proprietary or Trade Name:Assurance™ Alar / Nasal SpO2 Sensor
Common/Usual Name:Oximeter (Accessory – sensor)
Classification Name:Oximeter
Product Classification – DQA
21 CFR 870.2700
Class II
Predicate Devices:K122996 – Xhale Assurance™ Aral SpO2 sensor
K060576- Nellcor OxiMax sensors

Device Description:

The Alar / Nasal SpO2 Sensor is a disposable, single patient use Pulse Oximetry sensor designed to attach to the patient's nasal alar region - the fleshy region at the side of the nose. Skin contact and adhesive free sensor retention is via soft silicone rubber cushions encapsulating the optical components. The Alar / Nasal SpO2 Sensor with its associated patient cable, terminates in a DB-9 connector compatible with monitors employing Nellcor SpO2 technology.

The sensor utilizes red and IR LED light sources of 660 nm and 880 nm respectively along with a silicon photodiode detector to detect changes in oxygen saturation in the blood. Since oxygen saturated blood absorbs different amounts of light at each wavelength (red and infrared) as compared with unsaturated blood, the amount of light absorbed at each wavelength by the blood in each pulse can be used to calculate oxygen saturation.

Indications for Use:

The Assurance™ Nasal / Alar SpO2 Sensor is indicated for single patient use for continuously noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO-) and pulse rate from the nasal ala of adult and pediatric patients (weighing > 30kg), who are well or poorly perfused. The sensor can be used in a variety of healthcare environments where compatible pulse oximetry monitors are indicated for use, under professional supervision.

Comparison to Predicates

The following table compares the predicates.

{4}------------------------------------------------

510(k) Summary

AttributeXhaleAlar SpO2 sensorK122996NellcorOxiMax SensorsK060576Proposed XhaleAlar / Nasal SpO2 Sensor
Indications for UseThe Assurance™ Alar SpO2 Sensor isindicated for single patient use forcontinuously noninvasive monitoring offunctional oxygen saturation of arterialhemoglobin (SpO2) and pulse rate fromthe nasal ala of adult and pediatricpatients (weighing > 30kg), who are wellor poorly perfused. The sensor can beused in a variety of healthcareenvironments where compatible pulseoximetry monitors are indicated for use,under professional supervision.Indicated for single patient use forcontinuously noninvasive monitoring offunctional oxygen saturation of arterialhemoglobin (SpO2) and pulse rate.Neonate, pediatric, and adult who arewell or poorly perfused, in hospitals,hospital-type facilities, intra-hospitaltransport, and home environments.The Assurance™ Nasal / Alar SpO2 Sensoris indicated for single patient use forcontinuously noninvasive monitoring offunctional oxygen saturation of arterialhemoglobin (SpO2) and pulse rate from thenasal ala of adult and pediatric patients(weighing > 30kg), who are wellor poorly perfused. The sensor can beused in a variety of healthcareenvironments where compatible pulseoximetry monitors are indicated for use,under professional supervision.
Environments of usevariety of healthcare environmentswhere compatible pulse oximetrymonitors are indicated for use, underprofessional supervisionhospitals, hospital-type facilities,intra-hospital transport, and homeenvironmentsvariety of healthcare environments wherecompatible pulse oximetry monitors areindicated for use, under professionalsupervision
Patient populationadult and pediatric patients (weighing >30kg)Neonate, pediatric, and adultadult and pediatric patients (weighing >30kg)
Single patient useYesYesYes
Technology, Features and Specifications
Principle of OperationSpectrophotometric measurement offunctional arterial oxygen saturation bytransmissive mode pulse oximetrySpectrophotometric measurement offunctional arterial oxygen saturation bytransmissive mode pulse oximetrySpectrophotometric measurement offunctional arterial oxygen saturation bytransmissive mode pulse oximetry
LEDSRed (~660 nm) and IR (~880 nm)Red (~660 nm) and IR (~880 nm)Red (~660 nm) and IR (~880 nm)
DetectorPhotodiodePhotodiodePhotodiode
Connector9 pin DB-9 style9 pin DB-9 style9 pin DB-9 style

{5}------------------------------------------------

510(k) Summary

AttributeXhaleAlar SpO2 sensorK122996NellcorOxiMax SensorsK060576Proposed XhaleAlar / Nasal SpO2 Sensor
Performance Testing - Clinical and Non-clinical
Patient Contact MaterialsClassificationMaterials in the gas pathwayExternal CommunicatingTissueProlonged durationMaterials in direct patient contactSurface ContactMucosalProlonged durationMaterials in the gas pathwayExternal CommunicatingTissueProlonged durationMaterials in direct patient contactSurface ContactMucosalProlonged durationMaterials in the gas pathwayExternal CommunicatingTissueProlonged durationMaterials in direct patient contactSurface ContactMucosalProlonged duration
SpO2 Accuracy ( $A_{RMS}$ )*70-100% ± 2%70-100% ± 2%70-100% ± 3%
BPM30-250 bpm ± 3 bpm20-250 bpm ± 3 bpm30-240 bpm ± 3 bpm
IEC 60601-1Yes-Yes
IEC 60601-1-2EMC-EMC
ISO 80601-2-61Mechanical strengthStorage and Operating Temperature andhumidityFluid ingress-Mechanical strengthStorage and Operating Temperatureand humidityFluid ingress
Pulse rate accuracyYes-Yes
Inter-device reliability andaccuracyYes-Yes

{6}------------------------------------------------

510(k) Summary Page 4 of 6 16-Mar-15

Substantial Equivalence Discussion

The table above compares the key features of the proposed Assurance™ Alar / Nasal SpO2 Sensor as compared to the predicate Xhale Assurance™ Alar SpO2 Sensor (K122996) and the predicate Nellcor OxiMax Sensor (K060576).

Discussion of Differences:

The differences between the proposed Alar / Nasal SpO- Sensor and the predicate Alar Sensor (K122996):

  • · The Alar / Nasal SpO2 Sensor has 1 (one) wire connecting the sensor to the DB-9 connector.
    • The predicate device had a separate AB-N Adaptor Cable that was necessary to o connect to a DB-9 port.
    • The proposed Alar / Nasal SpO2 Sensor removed the need for the additional o adaptor by direct connection of the sensor assembly to the cable with a DB-9 connector.
  • The proposed Alar / Nasal SpO2 Sensor has a PCB with embedded software that is ● directly connected to the DB-9 connector.
    • This allows the sensor to communicate with the compatible pulse oximetry o equipment and display the SpO2 and heart rate data on the screen.
  • Pulse Rate Accuracy - Testing was performed for the accuracy of pulse rate over the range from 30-240 bpm using a SpO2 simulator the results were identical pulse rate accuracy within + 3 bpm over the tested range. We tested 30-240 bpm when the predicate Alar (K122996) was tested at 30-250 bpm. This test method difference is not significant as long as it is disclosed in the labeling and is equivalent to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor.

Discussion - These differences can be viewed as substantially equivalent and do not raise any new safety or effectiveness concerns when compared to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor.

In summary one can conclude that substantial equivalence is met based upon the following:

Indications for Use – The indications for use are identical for the proposed device when compared to the predicate - K122996 - Xhale Assurance™ Alar SpO2 sensor. Discussion - Each device is indicated for use as a pulse oximeter sensor.

Technology and construction - The technology is identical to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor. The basic design, fabrication, constructions and materials are identical to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor. Discussion - The basic design, fabrication, constructions and materials are identical to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor.

Patient Population - The patient population is identical to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor.

Discussion - The patient population is identical - predicate and adults > 30kg.

{7}------------------------------------------------

510(k) Summary Page 5 of 6 16-Mar-15

Environment of Use - The environments of use are identical to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor. Discussion - Both devices are used in the identical settings.

Non-Clinical Testing Summary -

Pulse Rate Accuracy - Testing was performed for the accuracy of pulse rate over the range from 30-240 bpm using a SpO2 simulator the results were identical pulse rate accuracy within + 3 bpm over the tested range.

Discussion - The Pulse Rate Accuracy is equivalent to the predicate K122996 - Xhale Assurance™ Alar SpO2 sensor.

Inter-device reliability and accuracy - Testing of the sensor across a range of test conditions and the testing showed the pulse rate to be within + 3 bpm over the range.

Discussion - The inter-device reliability and accuracy is identical to the predicates K122996 -Xhale Assurance™ Alar SpO2 sensor and K060576 - Nellcor OxiMax sensor.

Skin Temperature - Per ISO 80601-2-61 the maximum temperature limit is 41°C. Testing showed that corrected for 35°C the proposed sensor did not exceed this limit. Results ranged from 34.3 to 39.1℃.

Discussion - The results are substantially equivalent to the predicates s K122996 – Xhale Assurance™ Alar SpO2 sensor and K060576 - Nellcor OxiMax sensor.

Materials - The materials were tested per ISO 10993-1 for the submission K122996 and we have not changed the materials.

Discussion - The materials are identical to the predicate K122996 - Xhale Assurance™M Alar SpO2.

Clinical Testing

We performed a controlled desaturation study with healthy volunteers per ISO 80601-2-61 and the results across the 8 oximetry platforms showed that the SpO2 was within ARMS specification of 3 under steady state / non-motion conditions for the range of 70-100%

Discussion - The clinical testing demonstrated that the performance is substantially equivalent to the predicates K122996 - Xhale Assurance™ Alar SpO2 sensor and K060576 - Nellcor OxiMax sensor. ISO 80601-2-61 allows for an ARMS range of up to + 3% across the range of 70-100%.

Xhale Assurance A1x sensor withmodel testedArms SpO270-100%
Nellcor N-600x1.8 (257 pts)
Philips Intellivue A041.5 (257 pts)
Nellcor N-5952.1 (258 pts)
Philips Intellivue A022.2 (258 pts)
Nellcor N-3951.7 (266 pts)
Philips Intellivue A011.7 (264 pts)
DataScope Passport 21.7 (279 pts)
GE Dash 30002.2 (279 pts)

{8}------------------------------------------------

510(k) Summary Page 6 of 6 16-Mar-15

Substantial Equivalence Conclusion -

The sponsor has demonstrated through performance testing, design and features, and non-clinical testing that the proposed device and predicate have been found to be substantially equivalent.

§ 870.2700 Oximeter.

(a)
Identification. An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.(b)
Classification. Class II (performance standards).