K Number
K143101

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-05-01

(184 days)

Product Code
Regulation Number
890.3850
Age Range
0 - 21
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The indications for use of the Helio A7 and Move Manual Wheelchairs is to provide mobility to persons limited to a sitting position.

Device Description

The Motion Composites Helio A7 and Move Manual Wheelchairs are manually operated, user propelled, manual, mechanical wheelchairs. Their intended function and use is to provide mobility to persons limited to a sitting position.

Both are traditional folding cross-brace wheelchairs. They are made of an aluminum frame which utilizes one cross brace system that, when opened, nestles inside of the frame onto 4 hooks to create a box like rigid assembly. Upon the outside of this framework, and to the rear, are assembled two aluminum axle plates. Wheels of varying size and type are connected to the stainless steel axle receivers via stainless steel axles.

On the front end of the frame are assembled two aluminum caster housings. Caster forks are mounted to these housings via steel axles. A variety of caster wheels and tires are then connected to the fork based on user preference.

AI/ML Overview

The provided text is a 510(k) premarket notification for the Motion Composites Helio A7 and Move Manual Wheelchairs. This document focuses on demonstrating substantial equivalence to a predicate device, the Sunrise Medical Quickie 2 Manual Wheelchair (K123975), rather than describing acceptance criteria for a new, unique device needing extensive clinical validation.

Therefore, the typical structure for acceptance criteria and a study proving device performance as commonly understood for AI/diagnostic devices (e.g., sensitivity, specificity, F1-score with defined thresholds and ground truth from experts) does not directly apply to this submission for a mechanical wheelchair.

Instead, the "acceptance criteria" here are compliance with recognized standards and demonstrating similarity to a legally marketed predicate device. The "study" proving performance is a series of engineering tests against these standards.

Here's an interpretation based on the provided document:


1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Performance Standard)Reported Device Performance (Compliance)
ISO 7176-1:1999 (Static Stability)Compliant
ISO 7176-3:2012 (Effectiveness of Brakes)Compliant
ISO 7176-5:2008 (Overall Dimensions, Mass, Maneuvering Space)Compliant
ISO 7176-7:1998 (Seating and Wheel Dimensions)Compliant
ISO-7176-8:1998 (Static Impact and Fatigue Strength)Compliant
ISO 7176-15 (Information Disclosure, Documentation, Labeling)Compliant
Cal 117 (Flammability for upholstery)Compliant
FDA 1995 Guidance (Mechanical & Powered Wheelchairs)Met or Addressed

2. Sample size used for the test set and the data provenance

  • Sample Size: The document does not specify the exact number of physical units tested. However, the testing involves physical prototypes of the Helio A7 and Move Manual Wheelchairs to determine compliance with the listed ISO standards.
  • Data Provenance: The tests were conducted on physical prototypes of the devices by the manufacturer (Motion Composites) or an accredited testing facility to ISO standards. This is prospective conformance testing of engineered products.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts

  • This is not applicable in the context of mechanical wheelchair testing against engineering standards. The "ground truth" is defined by the objective metrics and test procedures outlined in the ISO standards themselves, rather than expert consensus on a clinical diagnosis. The "experts" would be the engineers and technicians performing the tests and comparing results to the specified limits within the ISO standards.

4. Adjudication method for the test set

  • Not applicable in the sense of clinical adjudication. Compliance is determined by whether the physical measurements and performance during the tests fall within the acceptable ranges specified by each ISO standard.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

  • This is not applicable. This is a submission for a mechanical medical device (wheelchair), not an AI-enabled diagnostic or therapeutic device. No MRMC study was conducted.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

  • This is not applicable. The device is a manual wheelchair, operated by a human user. There is no algorithm or standalone performance in the sense of AI. The device function is "dependent solely upon the wheelchair user."

7. The type of ground truth used

  • The "ground truth" is adherence to international engineering standards (ISO 7176 series and Cal 117) for wheelchair performance and safety, along with FDA guidance documents for mechanical wheelchairs. These standards define objective, measurable criteria for stability, braking, dimensions, strength, flammability, and other safety/performance aspects.

8. The sample size for the training set

  • This is not applicable. There is no "training set" in the context of a mechanical device like a wheelchair. The design and manufacturing process would involve engineering principles and prototypes, but not machine learning training data.

9. How the ground truth for the training set was established

  • This is not applicable as there is no training set. The design of the wheelchair is based on established engineering principles for mechanical stability, durability, and user interface, aiming to meet the performance criteria outlined in the relevant ISO standards.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the logo is an abstract symbol that resembles a stylized caduceus or a series of interconnected human profiles.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 1, 2015

Motion Composites c/o Edward A. Kroll Spectre Solutions, Inc. 5905 Fawn Lane Cleveland, OH 44141

Re: K143101

Trade/Device Name: Helio A7 Manual Wheelchair and Move Manual Wheelchair Regulation Number: 21 CFR 890.3850 Regulation Name: Mechanical Wheelchair Regulatory Class: Class I Product Code: IOR Dated: March 23, 2015 Received: March 24, 2015

Dear Mr. Kroll,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Felipe Aquel -S

Carlos L. Peña, PhD, MS for Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K143101

Device Name

Helio A7 and Move Manual Wheelchairs

Indications for Use (Describe)

The indications for use of the Helio A7 and Move Manual Wheelchairs is to provide mobility to persons limited to a sitting position.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

|X | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for Motion Composites. The logo consists of a blue circle with a stylized blue wing inside of it, followed by the text "MOTION COMPOSITES" in blue. Below the logo is the text "510(k) Premarket Notification".

Motion Composites Helio A7 and Move Manual Wheelchairs 510(k) Summary (Modified 5/1/2015)

l. SUBMITTER

Motion Composites 519 J - Oswald Forest Suite 101 Saint-Roch-del'Achigan Quebec, Canada JOK 3H0 Phone: (450) 588-6555 Fax: (450) 588-0200 Email: info@motioncomposites.com Web Site: www.motioncomposites.com

II. DEVICE

Name of Device: Helio A7 and Move Manual Wheelchair Common or Usual Name: Manual Wheelchair Classification Name: Wheelchair, Mechanical (21 CFR 890.3850) Regulatory Class: I Product Code: IOR

III. PREDICATE DEVICE

Sunrise Medical Quickie 2 Manual Wheelchair K123975

IV. DEVICE DESCRIPTION

The Motion Composites Helio A7 and Move Manual Wheelchairs are manually operated, user propelled, manual, mechanical wheelchairs. Their intended function and use is to provide mobility to persons limited to a sitting position.

Both are traditional folding cross-brace wheelchairs. They are made of an aluminum frame which utilizes one cross brace system that, when opened, nestles inside of the frame onto 4 hooks to create a box like rigid assembly. Upon the outside of this framework, and to the rear, are assembled two aluminum axle plates. Wheels of varying size and type are connected to the stainless steel axle receivers via stainless steel axles.

On the front end of the frame are assembled two aluminum caster housings. Caster forks are mounted to these housings via steel axles. A variety of caster wheels and tires are then connected to the fork based on user preference.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Motion Composites, which includes a blue circle with a stylized feather inside, followed by the company name in blue text. Below the logo, the text "510(k) Premarket Notification" is present. The text is in a smaller font size and is also in blue.

Device Function

Device function is dependent solely upon the wheelchair user. It does not function on its' own in any manner. The wheelchair user controls motion, speed and direction by propelling themselves using the hand rims located on the rear wheels.

Scientific Concepts

There are no complex scientific concepts related to the Helio A7 and Move manual wheelchair. They are simple, basic, manually operated mobility devices.

Significant Physical and Performance Characteristics:

Design:

  • . The Helio A7 and Move utilize a monocoque aluminum frame that is entirely welded and heat threated for greatly improved rigidity over reqular interconnected side frames.
  • A completely symmetrical aluminum cross-brace connects the left and right mono side frames of the wheelchair together and serves as the mechanism that allows the frame to fold.
  • A set of adjustable axle plates are attached vertically to each side frame and provide the ● wheelchair with ability to move the rear wheel position forward and backward in order to give the user an adjustment as to the center of gravity of the wheelchair. This rear axle vertical position limits torsion and flexion for improved propulsion.

Materials:

Materials used are:

  • . Aluminum frame, support members, wheels and components
  • . Steel fasteners and components
  • Polyurethane tires
  • Fabric covered foam upholstery

Physical Properties:

The Helio A7 and Move consist primarily of an aluminum frame assembly, a back rest frame, seat and back rest upholstery, large rear wheels with hand rims for self-propelling the chairs and front swivel pivoting casters for turning.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for Motion Composites, which includes a blue circle with a stylized feather inside, followed by the company name in blue text. Below the logo, there is a horizontal line, and below that, the text "510(k) Premarket Notification" is written. The text is smaller than the logo and is also in blue.

INDICATIONS FOR USE V.

The indications for use of the A7 and Move Manual Wheelchairs are to provide mobility to persons limited to a sitting position. This identical to the predicate device.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The technology and principle of operation for the Helio A7 and Move and the Quickie 2 are identical. They all consist of an aluminum frame with a seat and large rear wheels with hand rims for propelling the device. Smaller, pivoting type casters are mounted on the front of the chairs for steering and turning.

Device function is dependent solely upon the wheelchair user. They do not function on their own in any manner. The wheelchair user controls motion, speed and direction by propelling themselves using the hand rims located on the rear wheels.

If both rear wheels are propelled at the same time, the chair will move forward in a straight direction. If only the left rear wheel is propelled the chair will turn to the right. If only the right rear wheel is propelled the chair will turn to the left.

Discussion of Similarities and Differences:

The Helio A7 and Move are substantially equivalent to the predicate Quickie 2 wheelchair (K123975) in technology, function, performance and materials. They have the same indications for use which is to provide mobility for persons restricted in a sitting position.

The frame width, depth and the back cane heights vary slightly between the devices, but since these are only used to better fit the device to the user's needs, these slight differences do not cause any concerns for the safety and effectiveness of the weight limit is the same for all devices and is clearly stated in the proposed device labeling.

With regard to accessories and add-ons, all devices offer the same types of armrests, backrests, hangers and footplates. These accessories are made of the exact same materials for all devices and thus do not raise any questions for the safety and effectiveness of the Helio A7 and Move wheelchairs. The same can be said of back types and extension tubes.

Axle plates for all devices are made out of the same material, however, the actual mountinq systems differ slightly. Where the Quickie 2 axle plate is mounted horizontally on the frame, the Helio A7 and Move are mounted vertically.

This mechanical change permits better handling and propulsion the inwards flexion of the wheel during wheel pushes. It does not raise any concerns for the safety and effectiveness of the device and actually augments the effectiveness of the Helio A7 and Move when compared to

{6}------------------------------------------------

Image /page/6/Picture/1 description: The image contains the logo for Motion Composites. The logo consists of a blue circle with a stylized blue feather inside of it. To the right of the logo is the text "MOTION COMPOSITES" in blue.

510(k) Premarket Notification

the predicate device. This mounting system is a patented technology exclusive to Motion Composites.

The same can be said for the caster housing that holds the front wheel into place. While the caster housing of the predicate device and those of the Helio A7 and Move are made all made of aluminum, the mounting system of the Helio A7 casters housing is patented and thus slightly different from the one found on the Quickie 2 wheelchair.

This mounting system embeds the caster housing directly into the frame, reducing the effects of wear and tear by minimizing the amount of parts used to fasten the caster to the frame, thus providing a better overall experience to the wheelchair user.

This change as well does not raise any concerns as far as the safety of the overall product against the predicate device, since all casters offer the same effectiveness as per the Helio A7 successfully obtaining its ISO certification. As for the Move caster housing, it uses the same exact design and function as the predicate device.

The seat and back upholstery offering of the Helio is equivalent to the one found on the predicate device. It is made of high-quality durable sailcloth, which consists of 1000 denier fill and 250 denier wrap.

The rear wheels and hand rims of the predicate device and the Helio A7 Move wheelchair are offered in a wide range of different options. They are materials and are widely used in the wheelchair industry. The specific type, material and size of the wheels are chosen depending on the user's needs. These differences do not raise any concerns for the safety and effectiveness of the Helio A7 and Move wheelchairs. We also note the wheel locks are the same on all devices.

Based on the above discussion, Motion Composites believes that the Helio A7 and Move wheelchairs are substantially equivalent to the Quickie 2 wheelchair (K123975). While there are some differences between the Helio A7 and Move and its predicate, these differences are minor and do not alter the intended function and use of the device, nor do they raise any new questions pertaining to safety or effectiveness.

VII. PERFORMANCE DATA

The Helio A7 and Move Manual Wheelchairs have been tested to the following standards;

  • ISO 7176-1:1999 Determination of Static Stability

  • ISO 7176-3:2012 Determination of Effectiveness of Brakes

  • ISO 7176-5:2008 Determination of Overall Dimensions, Mass and Maneuvering

  • Space

  • ISO 7176-7:1998 Determination of Seating and Wheel Dimensions

  • ISO-7176-8:1998 Requirements and Test Method for Static Impact and Fatique Strength

{7}------------------------------------------------

510(k) Premarket Notification

  • ISO 7176-15 Requirements for Information Disclosure, Documentation and Labeling

The upholstery used meets the requirements of Cal 117 for flammability. These are identical to the materials used in the Motion Composites Helio manual wheelchair. The Helio manual wheelchair was cleared by FDA December 19, 2012 under 510(k) Accession Number K120628.

The performance data for the Quickie 2 is not published. However, on page 6 of the Quickie 2 Owner's Manual, the following statement is made;

"Based on ANSI/RESNA testing, Sunrise Medical recommends the use of a caster wheel with a minimum diameter of 5″ if the wheelchair will be overcoming obstacles of up to ½″ on a reqular basis.

This statement indicates that the Sunrise Quickie 2 has been tested to at least some of the ANS//RESNA standards. These standards are comparable to the ISO 7176 standards which were used to test the Motion Composites A7 and Move wheelchairs. (Note that a copy of the Quickie 2 Owner's Manual is provided in Volume 10 of this submission).

Further, manual wheelchairs are subject to the 1995 FDA guidance document entitled: "Guidance Document for the Preparation of Premarket Notification [(510(k)] Applications Mechanical and Powered Wheelchairs, and Motorized Three-Wheeled Vehicles." This quidance includes various safety and performance requirements that the Quickie 2 would have to have met or addressed prior to being cleared by FDA. Meeting these requirements also serves as a basis for substantial equivalence of the A7 and Move to its' predicate.

VIII. CONCLUSIONS:

Performance testing and compliance with FDA quidance for manual wheelchairs supports the safety of the Helio A7 and Move manual wheelchairs. Since the predicate device was cleared based in part on these data, conformance with the same requirements demonstrate that the Helio A7 and Move will perform as intended in the specified use conditions.

Further, technology, principle of operation and indications for use between Helio A7 and Move and the Quickie 2, are identical. This demonstrates that they perform comparably to the predicate device that is currently marketed for the same intended use. Therefore, the Helio A7 and Move manual wheelchairs are substantially equivalent to the Quickie 2 device.

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image shows the logo for Motion Composites, a company that specializes in the design and manufacturing of high-performance wheelchairs. The logo features a stylized feather design in blue, along with the company name in a sans-serif font. Below the logo, the text "510(k) Premarket Notification" is displayed, indicating that the company has submitted a premarket notification to the FDA for one of its products.

Helio A7 and Move Manual Wheelchairs

Helio A7 and Move Predicate Device Comparison

Feature/SpecificationQuickie Wheelchair (K123975)Helio A7 WheelchairMove Wheelchair
Intended UseThe Quickie 2 Wheelchair is amanually operated device intended tobe used as a means of mobility forpersons restrict to a sitting position.The Helio A7 wheelchair is a manuallyoperated device intended to be usedas a means of mobility for personsrestrict to a sitting position.The Move wheelchair is a manuallyoperated device intended to be usedas a means of mobility for personsrestrict to a sitting position.
Primary MaterialsAluminum 6000 seriesAluminum 7000 seriesAluminum 6000 series
Folding MethodCollapsible Cross-BraceCollapsible Cross-braceCollapsible Cross-brace
Frame Width11" to 22"14" to 22"14, 16, 18, 20, 22"
Overall width20,5" to 28,5"21 1/2" to 29 ½"21 1/2" to 29 ½"
Seat Depth10" to 20"15" to 21"16",18",20
Back Heights8,5" to 19"8,5" to 19"16", 20"
Weight Limit :250lbs250 lbs.250 lbs.
Seat height :16.75" to 22.75".13" to 21 1/4"13" to 20 3/4"
Chair Weight (withoutfootrests)27 lbs.16.5 lbs. (without footrests)19 lbs. (without footrests)
WarrantyLifetime on frameFrame 5 years, Components 1 yearFrame 5 years, Components 1 year
ArmrestsFlip back Height adjustableT-Shaped Armrests - Height adjustableTubular Swing-away-Flip back armrest, removable, heightadjustable (by user, ne tool required)-T-Shaped armrest, removable, heightadjustable (by user, no tool required)-Tubular Swing-away-Flip back armrest, removable, heightadjustable (by user, ne tool required)-T-Shaped armrest, removable, heightadjustable (by user, no tool required)-Tubular Swing-away
Front end typesSwing-AwayNon swing-awaySwing-AwaySwing-Away
Back TypeStandardAngle AdjustableDepth Adjustable Standard-Straight-8 degree bend back-Adjustable angle- straight or 8 degreebend-8 degree bend back with integratedpush handle-Adjustable angle 8 degree bend withintegrated push handle
Footrest Hangers60,70,90 degrees, elevating legrestsFlip in, flip out; 60°, 70° , 90°,elevating legrestFlip out; 60°, 70° , elevating legrest
FootplatesComposites, Foam, Aluminum angleadjustable-Standard non adjustable-Angle Adjustable Oversized-Standard non adjustable-Angle Adjustable Oversized
Back UpholsteryLow, Medium, Tall, Adjustable,Standard, AdjustableStandard, Adjustable
Axle PlatesStandard, Curved, Amputee, OffsetStandardStandard, AmputeeStandard, Amputee
Wheel sizes20,22,25,26 "20 ,22,24, 25, 26 "20 ,22,24, 25, 26 "
Wheel typesSpoke, Composite Mag, Spinergy, Onearm drive-Plastic Mags-Spoke, Newton One-Spoke, Newton Gravity-Spinergy LX-Spinergy Spox-Plastic Mags-Spoke, Newton One-Spoke, Newton Gravity-Spinergy LX-Spinergy Spox
Tire typesPneumatic, Pneumatic w/ airlessinsertPolyurethane, Low Profile FullprofileIron Cap-Pneumatic regular and high pressure-Full and low profile polyurethane-Hard Urethane-Pneumatic regular and high pressure-Full and low profile polyurethane-Hard Urethane
HandrimsAluminum, Plastic Coated,Projections-Aluminum Anodized-Plastic Coated (regular or high friction)-Spinergy Flexrim-Natural Fit handrim-Surge Handrim-Aluminum Anodized-Plastic Coated (regular or high friction)-Spinergy Flexrim-Natural Fit handrim-Surge Handrim
Feature/SpecificationQuickie Wheelchair (K123975)Helio A7 WheelchairMove Wheelchair
Caster typesPolyurethane, semi-pneumatic, SoftComposites wheels, Pneumatic, SoftComposites wheels, Pneumatic, Soft
Roll, PneumaticRollRoll
Fork Sizes3″,4″,5″,6″,7″3″,4″,5″,7″3″,4″,5″,7″
Fork Stem SizesStd,+3/4",+1 ½"std, +1 , +2std, +1 , +2
Caster OptionsMulti-position fork caster, pin locksMulti-position fork caster, pin locksMulti-position fork caster, pin locks
Wheel LocksPush to lock, pull to lock, scissor lockPush to lock, pull to lock, scissor lockPush to lock, pull to lock, scissor lock
Anti-tip tubesYesYesYes
Target populationRestricted to a sitting positionRestricted to a sitting positionRestricted to a sitting position
StandardsUnknownISO 7176-(1, 3, 5, 7, 8 & 15)ISO 7176 - (1, 3, 5, 7, 8 & 15)

{9}------------------------------------------------

Image /page/9/Picture/0 description: The image shows the logo for Motion Composites, which includes a blue circle with a stylized feather inside. To the right of the logo is the company name, "MOTION COMPOSITES," also in blue. Below the logo and company name, the text "510(k) Premarket Notification" is displayed, indicating the document's purpose.

Helio A7 and Move Manual Wheelchairs

Helio A7 and Move Predicate Device Comparison

§ 890.3850 Mechanical wheelchair.

(a)
Identification. A mechanical wheelchair is a manually operated device with wheels that is intended for medical purposes to provide mobility to persons restricted to a sitting position.(b)
Classification. Class I (general controls).