K Number
K142907

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-03-25

(170 days)

Product Code
Regulation Number
874.3302
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Cochlear Baha 5 sound processor has the following indications for use:

  • Patients of any age for use with the Baha Softband or headband. Patients aged 5 and o older for use with the Baha auditory osseointegrated implant system.
  • Patients who have a conductive or mixed hearing loss and can still benefit from sound o amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 45 dB HL.
  • . Bilateral fitting is intended for patients who meet the criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conduction thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1, 2, and 3 kHz), or less than a 15 dB difference at individual frequencies.
  • . Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. Single-Sided Deafness; SSD). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.
  • Baha for SSD is also indicated for any patient who is indicated for an air-conduction ● contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
Device Description

The subject of this 510(k) is a modification to the Cochlear Baha 4 sound processor, which was cleared under 510(k) K132278 for unilateral or bilateral use with conductive and mixed hearing losses (as a result of congenital malformations such as atresia, or certain medical conditions such as chronic suppurative otitis media), and for cases of single-sided sensorineural deafness (SSD, caused by a congenital condition, surgery, trauma or disease). The Baha system has been marketed for more than 30 years throughout the world, and there are now more than 100,000 users of a Baha system globally.

Baha sound processors can be used with either the external Baha headband or Softband in persons of any age, or for children aged 5 or older with the Baha auditory osseointegrated implants. The external Baha headband or Softband system works via conventional transcutaneous bone conduction amplification. The Baha implant system works by combining the external sound processor with an abutment and a small titanium implant placed in the skull behind the ear in a simple surgical procedure. The system is based on the process of "osseointegration" through which living tissue integrates with titanium in the implant. Thus, the titanium implant becomes one with the surrounding bone, allowing high-quality amplified and processed sound from the Baha sound processor to be conducted via an abutment or a magnetic coupling to the skull bone and directly to a cochlea with residual functionality. For either form of transmission, the processed sound either bypasses a conductive block in those patients with conductive or mixed hearing loss, or transfers sound through the skull to the opposite-ear normal cochlea for patients with SSD.

The modified sound processor, the Baha 5 is an upgrade to the currently marketed Baha 4 sound processor. The Baha 5 sound processor has the same signal processing functionality as the Baha 4, but incorporates a new transducer into a smaller outer casing while providing comparable output. The modified device utilizes the same fundamental scientific principles, and has the same intended use and indications for use as the current legally marketed device it will replace. In addition, the Baha 5 is compatible with the same range of 2.4GHz devices currently marketed for use with the Baha 4, but also incorporates Bluetooth functionality directly into the device without the need for an external accessory. The Baha 5 also supports a mobile medical app, the Baha Smart App, which acts as an enhanced remote control for the device, and offers improved auditory streaming capabilities and mobile phone integration.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study information for the Cochlear™ Baha® 5 Sound Processor:

First, it's important to note that this document is a 510(k) summary for a premarket notification to the FDA. As such, it focuses on demonstrating substantial equivalence to a predicate device (the Baha 4 sound processor), rather than a comprehensive de novo clinical trial proving new acceptance criteria. Therefore, the "acceptance criteria" discussed are largely tied to demonstrating that the new device performs comparably to the predicate and meets general industry standards for medical devices.

1. Table of Acceptance Criteria and Reported Device Performance

Since this is a 510(k) summary for a modified device, the "acceptance criteria" are implied to be that the Baha 5 Sound Processor performs comparably to the Baha 4 Sound Processor and meets relevant safety, performance, and functionality standards for a bone conduction hearing aid. The document doesn't present a formal table of quantitative acceptance criteria with corresponding performance metrics like a typical clinical trial report would for a novel device. Instead, it describes general categories of testing and their successful outcomes as evidence of equivalence.

Acceptance Criteria Category (Implied)Reported Device Performance
Patient SafetyBiocompatibility testing successfully completed; Basic medical electrical safety testing successfully completed.
Product UsabilityDemonstrated through simulated use testing in a variety of use cases.
Product ReliabilityDemonstrated through exposure of the device to mechanical stresses, environmental extremes, and simulated long-term use followed by functional testing.
InteroperabilityFunctional testing with existing Baha fixation systems successful; Ability to perform a prescribed fitting through Baha Fitting Software successful.
Acoustic PerformanceCurves for maximum output force level at OFL90 and full on gain at OFL 60 generated and verified in comparison to target outputs and previous Baha processors. (Implied comparable performance to Baha 4)
Firmware ImplementationVerification of firmware implementation performed, including functional testing of firmware-enabled features.
Bluetooth FunctionalityFunctional testing of pairing, audio streaming, and remote control functionality of the Baha Smart App successful.
Wireless CompatibilityCompatibility with existing range of wireless accessories established; Testing of the range of the wireless link performed.
Electromagnetic Compatibility (EMC)Established that the device did not emit excessive amounts of electromagnetic energy; Operated as intended in the presence of interference sources; Coexistence testing performed with the device paired and streaming while exposed to in-band interference.

2. Sample Size Used for the Test Set and Data Provenance

The document does not specify a sample size for any "test set" in terms of human subjects or patient data. The testing described appears to be primarily non-clinical bench testing and simulated use testing of the device itself.

  • Data Provenance: Not applicable in the context of clinical patient data, as the focus is on device performance. The testing was conducted by Cochlear Americas/Cochlear Bone Anchored Solutions AB.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

This information is not provided as the document describes technical verification and validation testing, not a clinical study involving expert interpretation of patient data to establish ground truth. "Ground truth" in this context refers to the successful operation of the device against predefined engineering specifications and functional requirements.

4. Adjudication Method for the Test Set

This information is not provided and is generally not relevant for the type of technical verification and validation testing described. Adjudication methods like 2+1 or 3+1 are typically used in clinical studies where multiple human readers interpret medical images or data.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done

No, an MRMC comparative effectiveness study was not conducted or reported in this 510(k) summary. This document focuses on substantial equivalence based on technical and functional performance, not clinical effectiveness studies with human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The primary "device" here is a sound processor, which inherently involves a "human-in-the-loop" (the patient wearing it). While specific functional tests of the device (e.g., sound output, wireless connectivity) can be considered "standalone" tests of its technical performance, the device's overall purpose and function (hearing improvement) is not standalone from human interaction. The document describes tests of the device's components and integrated system, which are standalone in their verification against specifications, but not a standalone AI algorithm in the typical sense.

7. The Type of Ground Truth Used

The "ground truth" for the various tests described is based on:

  • Engineering specifications and design requirements: For safety, reliability, acoustic performance, firmware, Bluetooth, and wireless compatibility.
  • Industry standards: For biocompatibility and electromagnetic compatibility (EMC).
  • Performance of the predicate device (Baha 4): As a benchmark for "comparable output" and overall function.

8. The Sample Size for the Training Set

This information is not applicable/not provided. The Baha 5 Sound Processor is a hardware device with embedded firmware, not an AI/machine learning algorithm that requires a "training set" of data in the conventional sense. The "training" in the context of device development would refer to engineering design, iteration, and testing phases.

9. How the Ground Truth for the Training Set Was Established

This information is not applicable/not provided for the same reasons as point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's seal, which includes an abstract image of three human profiles facing right. The seal is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 25, 2015

Cochlear Americas Mr. Sean Bundy Director, Regulatory Strategy 13059 East Peakview Avenue Centennial. CO 80111

Re: K142907

Trade/Device Name: Baha 5 Sound Processor Regulation Number: 21 CFR 874.3300 Regulation Name: Hearing Aid, Bone Conduction Regulatory Class: Class II Product Code: LXB Dated: February 20, 2015 Received: February 23, 2015

Dear Mr. Bundy,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Image /page/1/Picture/8 description: The image shows the text "Deborah L. Falls -S" in a simple, sans-serif font. The text is arranged horizontally, with the name "Deborah L. Falls" appearing prominently, followed by a hyphen and the letter "S". The text is black against a white background, making it easily readable.

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications For Use

510(k) Number (if known): K142907 Device Name: Cochlear™ Baha® 5 Sound Processor Indications for Use Statement:

The Cochlear Baha 5 sound processor has the following indications for use:

  • Patients of any age for use with the Baha Softband or headband. Patients aged 5 and o older for use with the Baha auditory osseointegrated implant system.
  • Patients who have a conductive or mixed hearing loss and can still benefit from sound o amplification. The pure tone average bone-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) should be better than or equal to 45 dB HL.
  • . Bilateral fitting is intended for patients who meet the criterion in both ears, with bilaterally symmetric moderate to severe conductive or mixed hearing loss. Symmetrical bone-conduction thresholds are defined as less than a 10 dB average difference between ears (measured at 0.5, 1, 2, and 3 kHz), or less than a 15 dB difference at individual frequencies.
  • . Patients who suffer from unilateral sensorineural deafness in one ear with normal hearing in the other ear (i.e. Single-Sided Deafness; SSD). Normal hearing is defined as a pure tone average air-conduction hearing threshold (measured at 0.5, 1, 2, and 3 kHz) of better than or equal to 20 dB HL.
  • Baha for SSD is also indicated for any patient who is indicated for an air-conduction ● contralateral routing of signals (AC CROS) hearing aid, but who for some reason cannot or will not use an AC CROS.
Prescription UseX(21 CFR 801 Subpart C)AND/OROver-The-Counter Use(Part 21 CFR 801 Subpart D)
--------------------------------------------------------------------------------------------------------------

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

510(k) Summary

In accordance with the Safe Medical Devices Act (SMDA) of 1990 and Title of the Code of Federal Regulations Part 807 (21 CFR §807), and in particular §807.92, the following summary of safety and effectiveness information is provided.

Submitted by: Cochlear Americas 13059 East Peakview Ave. Centennial, CO 80111

Manufactured by: Cochlear Bone Anchored Solutions AB Konstruktionsvägen 14 SE-435 33 Mölnlycke Sweden

Contact Persons: Sean Bundy Director of Regulatory Strategy Phone: 303-524-7139 Facsimile: 303-524-6825 Email: SBundy@cochlear.com

Date Submission Prepared: December 2014

Device Name: Trade or Proprietary Name: Common or Usual Name: Classification Status: Product Codes: Panel:

Cochlear™ Baha® 5 Sound Processor Osseointegrated Auditory Prosthesis Sound Processor Class II, 21 CFR §874.3300 LXB Ear Nose and Throat Devices Panel

{4}------------------------------------------------

Device Description:

The subject of this 510(k) is a modification to the Cochlear Baha 4 sound processor, which was cleared under 510(k) K132278 for unilateral or bilateral use with conductive and mixed hearing losses (as a result of congenital malformations such as atresia, or certain medical conditions such as chronic suppurative otitis media), and for cases of single-sided sensorineural deafness (SSD, caused by a congenital condition, surgery, trauma or disease). The Baha system has been marketed for more than 30 years throughout the world, and there are now more than 100,000 users of a Baha system globally.

Baha sound processors can be used with either the external Baha headband or Softband in persons of any age, or for children aged 5 or older with the Baha auditory osseointegrated implants. The external Baha headband or Softband system works via conventional transcutaneous bone conduction amplification. The Baha implant system works by combining the external sound processor with an abutment and a small titanium implant placed in the skull behind the ear in a simple surgical procedure. The system is based on the process of "osseointegration" through which living tissue integrates with titanium in the implant. Thus, the titanium implant becomes one with the surrounding bone, allowing high-quality amplified and processed sound from the Baha sound processor to be conducted via an abutment or a magnetic coupling to the skull bone and directly to a cochlea with residual functionality. For either form of transmission, the processed sound either bypasses a conductive block in those patients with conductive or mixed hearing loss, or transfers sound through the skull to the opposite-ear normal cochlea for patients with SSD.

The modified sound processor, the Baha 5 is an upgrade to the currently marketed Baha 4 sound processor. The Baha 5 sound processor has the same signal processing functionality as the Baha 4, but incorporates a new transducer into a smaller outer casing while providing comparable output. The modified device utilizes the same fundamental scientific principles, and has the same intended use and indications for use as the current legally marketed device it will replace. In addition, the Baha 5 is compatible with the same range of 2.4GHz devices currently marketed for use with the Baha 4, but also incorporates Bluetooth functionality directly into the device without the need for an external accessory. The Baha 5 also supports a mobile medical app, the Baha Smart App, which acts as an enhanced remote control for the device, and offers improved auditory streaming capabilities and mobile phone integration.

Intended Use:

The new Baha 5 will be used as an external sound processor option (in the Cochlear Baha family of sound processors) to conduct sound energy directly to the cochlea via a Baha auditory implant, or via transcutaneous transmission with a Baha headband or Softband. This is the same intended use, and for the same patient population, as the current legally marketed, unmodified Baha 4 device (the predicate device that it will replace). The Baha Smart App is intended to supplement the existing range of accessories for the sound processor, and acts as a more fullyfeatured remote control than those currently available.

{5}------------------------------------------------

Technological Characteristics:

The modified sound processor will still be compatible with the currently marketed Softband/headband (cleared under K002913 and letters to file under this clearance), and the currently marketed auditory osseointegrated implant (BIA300 system, cleared under K100360 and BA400 cleared under K121317 as well as the Baha Attract cleared under K131240), and will also be backward compatible with the original auditory osseointegrated implant (cleared under K955713).

The primary modifications proposed are to the size and shape of the device, as well as the incorporation of a smaller vibrating transducer. The modified transducer provides comparable output to the currently marketed device, while allowing for a smaller overall design. In addition, the 2.4 GHz technology has been updated to incorporate Bluetooth in addition to the current wireless communications protocol. Compatibility with the existing range of wireless accessories is maintained.

The Baha Fitting Software (also cleared under K132278) has been updated to support the fitting of the modified device, but is largely unchanged in terms of capability.

Summary of Testing:

Testing of the Baha 5 included testing to confirm patient safety, product usability and reliability, interoperability, and electromagnetic compatibility. Patient safety testing included biocompatibility testing as well as basic medical electrical safety testing. Usability was demonstrated through simulated use testing in a variety of use cases. Reliability testing was conducted through exposure of the device to mechanical stresses, and environmental extremes, and simulated long-term use followed by functional testing. Interoperability testing was performed including functional testing with existing Baha fixation systems, as well as the ability to perform a prescribed fitting through the Baha Fitting Software. Curves for the maximum output force level at OFL90 and full on gain at OFL 60 were generated, and verified in comparison to target outputs and to previous Baha processors. Verification of the implementation of the firmware was also performed, including functional testing of the firmware-enabled features of the device. Testing of the Bluetooth capability of the device included functional testing of pairing, audio streaming and remote control functionality of the Baha Smart App. Compatibility with the existing range of wireless accessories was established, and testing of the range of the wireless link was performed. Electromagnetic compatibility testing established that the device did not emit excessive amounts of electromagnetic energy, and that the device operated as intended in the presence of interference sources. Coexistence testing was performed, with the device paired and in a streaming configuration while exposed to various sources of in-band interference. The results of the testing indicate that the Baha 5 is substantially equivalent to the predicate device.

Conclusions:

Despite the changes to the physical components of the Baha 5 sound processor has substantially equivalent function and technology, and the same intended use as the predicate Baha 4 sound processor cleared for marketing under K132278. The indication for use statement is also the same as that for the current legally marketed Baha 4 sound processor that it will replace.

§ 874.3302 Bone-conduction hearing aid.

(a)
Identification. A bone-conduction hearing aid is a wearable sound-amplifying device intended to compensate for impaired hearing and that conducts sound to the inner ear through the skull. The non-implantable components of a bone-conduction hearing aid, such as the external sound processor, are subject to the requirements in § 801.422 of this chapter.(b)
Classification. Class II.