K Number
K142830

Validate with FDA (Live)

Manufacturer
Date Cleared
2015-06-25

(268 days)

Product Code
Regulation Number
886.4390
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Ophthalmed Bending Laser Probes (Models A207000, and A307000) are indicated for performing laser endophotocoagulation in the posterior segment of the eye during vitreoretinal surgery at an operating wavelength range from 500 to 1100 nm.

Device Description

This laser probe is made out of 8ft fiberoptic, terminated on one end with Alcon compatible (SMA905) laser connector, and on another side with a handpiece for holding and manipulation during surgery. For protection against damages, a flexible plastic jacket covers the length of the fiber. The handpiece is terminated by a proximal stainless steel tubing that is 20, 23 or 25 gauge in size, and ending with a distal pre-curved PEEK memory tube that can change angle when activated by the sliding button on the handle side. When the sliding button on the handle is advanced, an internal straightening tube advances into the distal pre-curved PEEK memory tube, thus causing a reduction of the angle of the pre-curved PEEK memory tube, down to zero degrees at the maximum sliding position, allowing the selection of the desired angle during surgery. The maximum angle is 55 degrees for 20g and 23g, and 40 degrees for 25g. This device allows to transmit laser energy from the laser source to the surgical site, facilitated by the aiming beam that is provided by the laser source. This device does not provide illumination and should not be used to illuminate the surgical site.

AI/ML Overview

Here's an analysis of the provided text regarding the acceptance criteria and study for the Ophthalmed Bending Laser Probes.

Based on the provided 510(k) Summary for the Ophthalmed Bending Laser Probes (K142830), the device is a laser probe designed for endophotocoagulation in the posterior segment of the eye. The study supporting its substantial equivalence is primarily focused on bench testing rather than clinical or AI-centric performance. Therefore, many of the requested elements for AI/clinical studies (like sample size for test/training sets, data provenance, number of experts for ground truth, adjudication methods, MRMC studies, and standalone performance) are not applicable to this type of device and submission.

The acceptance criteria are established through comparison with predicate devices and demonstrated through various physical and performance tests.


Acceptance Criteria and Reported Device Performance

The acceptance criteria for this device are implicitly derived from its equivalence to predicate devices, particularly K021696 and K113857, in terms of various physical and performance characteristics. The device is expected to perform comparably in key metrics.

Acceptance Criteria CategorySpecific MetricPredicate Device (K021696/K050807)Predicate Device (K113857)Ophthalmed Bending Laser Probes (K142830)
BiocompatibilityCompliance with ISO 10993 for:SatisfiedSatisfiedSatisfied (Cytotoxicity, Irritation, Sensitization)
SterilitySterility Assurance Level (SAL)SAL 10-6N/ASAL 10-6
Operating WavelengthRange of laser wavelength supported500 to 1100 nm (initially 514-532 nm)500 to 900 nm500 to 1100 nm
Max Power InputAbility to handle laser powerUp to 1500 mWUp to 1500 mWUp to 1500 mW
Laser TransmissionPercentage of laser energy transmitted96 to 97%92 to 97%94 to 95%
Spot Size & DistributionMaintenance of uniform spot size and energy distribution after abuse testingConfirmed UniformConfirmed UniformConfirmed Uniform
Divergence AngleAverage laser beam divergence angle14 degrees20 degrees16 degrees
Linearity (Bending Angle)Achieved bending angles at various sliding positions for 20g/23gN/A (Fixed tip)22.5/42.5/60/75 deg (20g); 20/30/40/45-60 deg (25g)20/37.5/45/55 deg (20g/23g); 15/27.5/37.5/42.5 deg (25g)
Abuse Test ResistanceNo significant change in transmission, spot size, or linearity after 1500 shots at 1500mWDemonstratedDemonstratedDemonstrated

Study Information (as applicable to a non-AI/clinical device)

  1. Sample size used for the test set and the data provenance:

    • Test Set Sample Size: For mechanical and acoustic testing, three 20g, three 23g, and three 25g sterile probes (a total of 9 devices) of the Ophthalmed Bending Laser Probes (K142830) were used. The document also refers to testing "same laser probes (K021696, K113857 and K142830)" for spot size and energy distribution tests, implying predicate devices were tested alongside the proposed device for comparison.
    • Data Provenance: The studies were conducted internally or by external labs, as indicated by "internal and external" tests and references to GLP studies for biocompatibility. The data is thus prospective, proprietary to Ophthalmed LLC, and not tied to specific countries beyond the manufacturer's location (USA).
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

    • Not Applicable. This is a bench test and performance verification study for a physical device, not an AI or diagnostic tool needing expert ground truth. The "ground truth" is defined by established engineering parameters, standards (e.g., ISO 10993), and measurements from laboratory equipment.
  3. Adjudication method (e.g., 2+1, 3+1, none) for the test set:

    • None. Adjudication methods are relevant for subjective assessments, particularly in clinical or image-based diagnostic studies. These tests involve objective physical measurements and comparisons to defined performance targets or predicate device performance.
  4. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

    • Not Applicable. This device is a surgical instrument (laser probe), not an AI-assisted diagnostic or decision-making tool that involves human readers/interpreters.
  5. If a standalone (i.e., algorithm only without human-in-the-loop performance) was done:

    • Not Applicable. This is not an algorithm or AI device. Its performance is inherent to its physical and functional characteristics.
  6. The type of ground truth used (expert consensus, pathology, outcomes data, etc.):

    • The "ground truth" for this submission consists of objective physical measurements and standard compliance. For example, laser transmission rates are measured directly, bending angles are measured in degrees, and biocompatibility is confirmed through standardized laboratory tests (Cytotoxicity, Irritation, Sensitization) following ISO 10993. The performance of predicate devices also serves as a comparative "ground truth" to establish substantial equivalence.
  7. The sample size for the training set:

    • Not Applicable. This is not an AI/machine learning device; there is no training set.
  8. How the ground truth for the training set was established:

    • Not Applicable. As there is no training set for an AI model, this question is not relevant.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human profile in triplicate, suggesting a sense of community and care. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the emblem.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 15, 2015

Ophthalmed LLC Mr. Jay Mansour Director , QA/RA 1050 Northfield Ct., Suite 280 Roswell, GA 30076

Re: K142830

Trade/Device Name: Ophthalmed Directional Laser Probes Models A207000. A807000 and A307000 Regulation Number: 21 CFR 886.4390 Regulation Name: Ophthalmic Laser Regulatory Class: Class II Product Code: HQB, HQF Dated: September 22, 2014 Received: September 30, 2014

Dear Mr. Mansour:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"

(21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/1/Picture/10 description: The image shows the name "Deborah L. Falls -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or label. The background is plain, ensuring the text is easily readable. The letters are uniformly sized and spaced, contributing to the overall clarity of the name.

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142830

Device Name

Ophthalmed Bending Laser Probes (Models A207000, A807000, and A307000)

Indications for Use (Describe)

The Ophthalmed Bending Laser Probes (Models A207000, and A307000) are indicated for performing laser endophotocoagulation in the posterior segment of the eye during vitreoretinal surgery at an operating wavelength range from 500 to 1100 nm.

Type of Use (Select one or both, as applicable)
☐ Temporary hardship (illness, loss of job, etc.)
☐ Disaster-related hardship (declared disaster area)

X | Prescription Use (Part 21 CFR 801 Subpart D)

| | Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510k Summary

As required by 21 CFR 807.92

-SUBMITTER

OPHTHALMED LLC. 1050 NORTHFIELD COURT, SUITE 280. ROSWELL, GA 30076 USA. Tel 770 777 6613 Fax 678 623 3765

Contact person: Jay Mansour Date prepared: June 25, 2015

II. DEVICE

Name of device: Ophthalmed bending laser probes (Models A207000, and A307000) Common or usual name: laser probe Classification name: Ophthalmic photocoagulator Regulatory class: II Product code: HQB

III. PREDICATE DEVICE

510k numberTrade or Proprietary or Model NameManufacturerPrimary predicate device
K021696LF20, LF25OPHTHALMED LLCYes
K050807LF100OPHTHALMED LLCNo
K113857Synergetics Directional Endo OcularLaser ProbeSYNERGETICSNo
K121187Katalyst laser probesKATALYST SURGICALNo

IV. DEVICE DESCRIPTION

This laser probe is made out of 8ft fiberoptic, terminated on one end with Alcon compatible (SMA905) laser connector, and on another side with a handpiece for holding and manipulation during surgery. For protection against damages, a flexible plastic jacket covers the length of the fiber.

The handpiece is terminated by a proximal stainless steel tubing that is 20, 23 or 25 gauge in size, and ending with a distal pre-curved PEEK memory tube that can change angle when activated by the sliding button on the handle side.

When the sliding button on the handle is advanced, an internal straightening tube advances into the distal pre-curved PEEK memory tube, thus causing a reduction of the angle of the pre-curved PEEK memory tube, down to zero degrees at the maximum sliding position,

{4}------------------------------------------------

allowing the selection of the desired angle during surgery. The maximum angle is 55 degrees for 20g and 23g, and 40 degrees for 25g.

This device allows to transmit laser energy from the laser source to the surgical site, facilitated by the aiming beam that is provided by the laser source. This device does not provide illumination and should not be used to illuminate the surgical site.

The user should refer to the instructions for use of the laser source, for information about the increments for power settings, duty cycle, and ranges and increments for timing and interval.

This device does not include any lenses, and there is no focal distance to define. The laser spot size at 4mm away is 1mm.

v. INDICATIONS FOR USE

The Ophthalmed Bending Laser Probes (Models A207000, and A307000) are indicated for performing laser endophotocoagulation in the posterior segment of the eye during vitreoretinal surgery at an operating wavelength range from 500 to 1100 nm.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICE DEVICE:

Summary of technological characteristics of our device compared to predicate device, in design, material, chemical composition and energy source:

Design:

All predicate devices have the same main structure and functionality as our device: a fiber optic is terminated on one side via a laser connector that mates with a laser machine, and on the other side terminated by a tip that penetrates the eye so that the laser beam can reach into the surgical site. (K050807, in addition to the above, provides also illumination and aspiration functionalities).

The difference is that only our device and K113857 allow the tip to move, for the surgeon to select the desired angle. The other predicate devices have fixed tip. K113857 has a pre-curved Nitinol tube enclosed within the main (external) stainless steel shaft that projects out of the handle. As the Nitinol tube projects out of the main shaft, the fiber optic follows the curvature of the Nitinol tube. Our device, on the contrary, has a fixed and pre-curved PEEK tip, which changes angle, as a straightening (internal) stainless steel tube penetrates it internally. The fiber optic within the pre-curved tip follows the curvature of the pre-curved tip.

In terms of design, our device uses the same components as K021696 (and K050807, except for illumination and aspiration functions), but with two changes: (a) enhanced

{5}------------------------------------------------

fiber optic specifications, and (b) a movable tip that was tested to be substantially equivalent in its performance to K113857.

Material and chemical composition:

All predicate devices have the same component list as our device: glass fiber optic, plastic protective sheath that covers the fiber optic, metallic laser connector, plastic handle and stainless steel tip. In addition to the above, our device and K113857 have other materials, namely, PEEK tube for our device, and Nitinol tube for K113857.

Laser output properties for our device, K113857 and K021696 (and indirectly K050807, having the same fiber optic as K021696) were successfully tested. In other words, the intended functionality of our device was not affected by the change in materials.

Our device is sterile with sterility assurance level of SAL 10-6, as per documented evidence provided within the submission, by running an additional fractional cycle based on currently validated ethylene oxide sterilization cycle.

Energy used:

All predicate devices and our device do NOT emit energy, but only transmits laser energy from the connector side to the tip side.

The fiber optic of K021696 and K050807 was cleared originally for 514 to 532 nm, then subsequently for 500 to 1,100 nm.

K121187 was cleared for 500 to 900 nm.

Our device covers 500 to 1,100 nm, as for K021696 and K050807.

Our device, K113857 and K021696 (and indirectly K050807, having the same fiber optic as K021696) have been tested for up to 1,500 mW, with comparable and acceptable laser output characteristics.

Biocompatibility testing

The biocompatibility evaluation for Ophthalmed Bending Laser Probes (Models A207000, A807000, and A307000) was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO 10993, as recognized by FDA.

The battery of testing included the following tests:

  • Cytotoxicity (L929 Neutral red uptake, 4 concentrations, as per ISO 10993-5:2009, GLP study)
  • Irritation (intracutaneous injection, as per ISO 10993-10:2010, GLP study)
  • Sensitization (Kligman maximization, as per ISO 10993-10:2010, GLP study)

{6}------------------------------------------------

The device is considered tissue contacting for a duration of less than 24 hours.

Electrical safety and electromagnetic compatibility (EMC)

Not applicable

Software verification and validation testing

Not applicable

Mechanical and acoustic testing

Our device is substantially equivalent to K113857 and indirectly K050807, having the same fiber optic as K021696), as far as laser transmission, spot size, energy distribution and linearity, at baseline and after prolonged conditions.

Three 20g, three 23g and three 25g sterile probes were subjected to various tests, along with predicate devices, as described below, in accordance with 21 CFR 807.92(b)(1):

Laser transmission:

  • i-Ophthalmed probes as per K021696: 96 to 97%
  • ii-Synergetics probes as per K113857: 92 to 97%
  • iii-Probes under this submission K142830: 94 to 95%

The above transmission rates were substantially the same prior and after an abuse test based on 1,500 shots at 1,500mW, with pulse duration of 200 msec at 50 msec pulse intervals, with transmission rates measured at 1,000mW.

Spot size and energy distribution:

The same laser probes (K021696, K113857 and K142830) were also subjected to additional tests. Projecting the laser beam onto an appropriate surface, photos were taken at 1000mW, indicating no changes before and after the same abuse test of 1500mW, concerning the spot sizes and energy distribution.

The energy distribution was confirmed to be uniform for the three 510k submissions.

Also, the average divergence angle was determined to be 20 degrees for K113857, 14 degrees for K021696 and 16 degrees for K142830.

Linearity:

The laser probes of K113857 and K142830 were tested for linearity at 25%, 50%, 75% and 100% of knob sliding. Again, the same abuse test of 1500mW did not affect the results. It was determined that for sizes of 20g and 23g of K142830, the tip bends 20

{7}------------------------------------------------

deg, 37.5 deg, 45 deg and 55 deg, at 25%, 50%, 75% and 100% sliding, respectively, but 15 deg, 27.5 deg, 37.5 deg and 42.5 deg for 25g. In comparison, K113857 reflected 22.5 deg, 42.5 deg, 60 deg and 75 deg for 20g probe, but 20 deg, 30 deg, 40 deg, and 45 to 60 deg for 25g probe, respectively as well (at 25%, 50%, 75% and 100% of sliding).

Animal study

Not applicable.

Clinical studies

Not applicable.

VII. CONCLUSIONS

The various tests, internal and external, document that our device is substantially equivalent to the predicate device. Our laser output is similar, our mechanism is similar, our biocompatibility is tested, our sterility is documented, and our overall structure / components is similar.

§ 886.4390 Ophthalmic laser.

(a)
Identification. An ophthalmic laser is an AC-powered device intended to coagulate or cut tissue of the eye, orbit, or surrounding skin by a laser beam.(b)
Classification. Class II.