K Number
K142791

Validate with FDA (Live)

Date Cleared
2015-01-21

(117 days)

Product Code
Regulation Number
876.1075
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Spectra Soft Tissue Biopsy Needles are indicated for obtaining biopsy samples from soft tissue in percutaneous or open surgical procedures from various tissues through a combination of cutting and/or aspirating in such a manner that the biopsy sample is retained in the orifice of the needle.

Device Description

The Spectra Soft Tissue and Aspiration Biopsy Needles consist of a stainless steel needle and a translucent standard female Luer locking hub. A stylet rod assembly mates with the needle hub assembly. A depth stop is integral with each needle type to facilitate depth placement. The needle assembly is covered by a translucent removable needle guard. The needles are available in a range of wall thicknesses, gauges and lengths to match the end-user need. Needles are available with an echogenic treatment to help ensure strong reflection during ultrasound procedures. Spectra Soft Tissue Biopsy Needles will be marketed as sterile, non-pyrogenic, and single use devices.

AI/ML Overview

The provided document is a 510(k) Premarket Notification from the FDA for Spectra Soft Tissue Biopsy Needles. This type of submission focuses on demonstrating substantial equivalence to a predicate device, rather than proving novel effectiveness through clinical studies with acceptance criteria for device performance as would be expected for an AI/CADe device.

Therefore, the specific information requested, such as acceptance criteria table, sample size for test sets, number of experts for ground truth, adjudication methods, MRMC studies, standalone performance, training set details, and ground truth for training data, is not available in this document. The document describes performance testing for safety and basic functionality, not clinical performance for diagnostic accuracy.

Here's an analysis based on the information available in the document, framed against the requested items:

1. A table of acceptance criteria and the reported device performance

The document does not provide a table with specific acceptance criteria that relate to diagnostic performance metrics (e.g., sensitivity, specificity, AUC) or a clinical study proving those criteria were met for diagnostic accuracy. Instead, it details performance tests related to material safety and mechanical functionality.

Test TypeDescription / PerformanceAcceptance Criteria (Explicitly stated)
BiocompatibilityConducted in accordance with FDA Blue Book Memorandum #G95-1 and ISO 10993.Device is biocompatible (passed tests).
CytotoxicityPerformedPassed
SensitizationPerformedPassed
Irritation/Intracutaneous ReactivityPerformedPassed
Systemic ToxicityPerformedPassed
HemocompatibilityPerformedPassed
Standard ISO Luer HubsApplicable luer testing used ISO 594.Passed (met ISO 594 standards)
Needle Bond StrengthWith hubsPassed
Overall ConclusionThe Submission Devices have met all established acceptance criteria for performance testing. (General statement)Implied safety and functionality equivalence to predicate device.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

This information is not applicable and not provided. The document describes testing of the physical biopsy needles themselves (biocompatibility, mechanical performance), not an AI/CADe system evaluated on a dataset of patient images or clinical outcomes.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This information is not applicable and not provided. "Ground truth" in the diagnostic sense (e.g., presence or absence of disease in an image) is not relevant to the type of device and testing described here. The "ground truth" for the performance tests would be the established safety and mechanical standards.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

This information is not applicable and not provided. Adjudication methods are typically used in clinical studies for diagnostic devices to resolve discrepancies in expert interpretation, which is not the focus of this submission.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

No MRMC study was done. This device is a physical biopsy needle, not an AI/CADe system.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

A standalone performance study, as understood for AI/CADe systems, was not conducted because this is a physical medical device. The "standalone" performance here refers to the physical and chemical characteristics of the needle itself.

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

For the biocompatibility tests, the "ground truth" is defined by the regulatory standards and recognized biological responses to materials (e.g., no cytotoxicity, no sensitization). For mechanical tests, the "ground truth" is adherence to established engineering standards (e.g., ISO 594 for Luer hubs, bond strength specifications). No patient-level diagnostic "ground truth" (pathology, expert consensus) is involved in this documentation.

8. The sample size for the training set

This information is not applicable and not provided. There is no AI algorithm being trained.

9. How the ground truth for the training set was established

This information is not applicable and not provided because there is no AI algorithm being trained.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring three stylized human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 21, 2015

Spectra Medical Devices Mr. Agustin Turriza Operations Manager 260-F/H Fordham Road Wilmington, Massachusetts 01887

Re: K142791

Trade/Device Name: Spectra Soft Tissue Biopsy Needles Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: October 21, 2014 Received: October 23, 2014

Dear Mr. Turriza:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

for

Sincerely yours,

David Krause -S

Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Premarket Notification Submission: SPECTRA SOFT TISSUE BIOPSY NEEDLES

DEPARTMENT OF HEALTH AND HUMAN SERVICESForm Approved: OMB No. 0910-0120
Food and Drug AdministrationExpiration Date: January 31, 2017
Indications for UseSee PRA Statement below.
510(k) Number (if known)
K142791
Device Name
Spectra Medical Devices, Inc.
Indications for Use (Describe)
Device Name: Spectra Soft Tissue Biopsy Needles:
Chiba Soft Tissue Biopsy Needle / Fine Needle Aspiration Biopsy NeedleFranseen Soft Tissue Biopsy Needle
Westcott Soft Tissue Biopsy Needle
Turner Soft Tissue Biopsy Needle
Greene Soft Tissue Biopsy Needle
Indications For Use:
Spectra Soft Tissue Biopsy Needles are indicated for obtaining biopsy samples from soft tissue in percutancous or open
surgical procedures from various tissues through a combination of cutting in such a manner that the
biopsy sample is retained in the orifice of the needle.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
PRAStaff@fda.hhs.gov"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY OF SAFETY AND EFFECTIVENESS

(21 CFR 807.92) for Spectra Soft Tissue Biopsy Needles

SUBMITTER:

Spectra Medical Devices, Inc. 260-F/H Fordham Road Wilmington, Massachusetts, 01887

ESTABLISHMENT REGISTRATION NUMBER:

1224960

CONTACT:

Agustin Turriza Manager, Operations Telephone: (978) 657-0889 Fax: (978) 657-4339 Email: aturriza@SpectraMedical.com

DATE PREPARED:

Trade Name:

September 23, 2014

SUBMISSION DEVICE:

Spectra Soft Tissue Biopsy Needles

Regulation Name: Common/Usual Name: Classification Panel: Review Advisory Committee: Regulatory Class: Product Code: Regulation Number:

Gastroenterology-urology biopsy instrument Soft Tissue and Fine Needle Biopsy Needles Gastroenterology/Urology General & Plastic Surgery Class II KNW 21 CFR § 876.1075

The following predicate devices have not been the subject of a design related recall. No reference devices were used in this submission.

PREDICATE DEVICE:

Trade Name:Manan Soft Tissue Biopsy Needles K980122
Regulation Name:Gastroenterology-urology biopsy instrument
Common/Usual Name:Manual biopsy needles for soft tissues
Classification Panel:Gastroenterology/Urology
Review Advisory Committee:General & Plastic Surgery
Regulatory Class:Class II
Product Code:KNW
Regulation Number:21 CFR § 876.1075

{4}------------------------------------------------

Trade Name:Quick-Core™ Biopsy Needle K973565
Regulation Name:Gastroenterology-urology biopsy instrument
Common/Usual Name:Soft Tissue Biopsy, puncture and aspiration cannulas and needles
Review Advisory Committee:General & Plastic Surgery
Regulatory Class:Class II
Product Code:KNW
Regulation Number:21 CFR § 876.1075

DEVICE DESCRIPTION:

The Spectra Soft Tissue and Aspiration Biopsy Needles consist of a stainless steel needle and a translucent standard female Luer locking hub. A stylet rod assembly mates with the needle hub assembly. A depth stop is integral with each needle type to facilitate depth placement. The needle assembly is covered by a translucent removable needle guard. The needles are available in a range of wall thicknesses, gauges and lengths to match the end-user need. Needles are available with an echogenic treatment to help ensure strong reflection during ultrasound procedures.

Spectra Soft Tissue Biopsy Needles will be marketed as sterile, non-pyrogenic, and single use devices.

INTENDED USE:

The Spectra Soft Tissue Biopsy Needles are intended to be used by medical professionals to obtain biopsy samples from soft tissue in percutaneous or open surgical procedures from various tissues through a combination of cutting and/or aspirating in such a manner that the biopsy sample is retained in the orifice of the needle.

INDICATIONS FOR USE:

Chiba Soft Tissue Biopsy Needle / Fine Needle Aspiration Biopsy Needle Device Name: Franseen Soft Tissue Biopsy Needle Greene Soft Tissue Biopsy Needle Turner Soft Tissue Biopsy Needle Westcott Soft Tissue Biopsy Needle

The Spectra Soft Tissue Biopsy Needles are indicated for obtaining biopsy samples from soft tissue in percutaneous or open surgical procedures from various tissues through a combination of cutting and/or aspirating in such a manner that the biopsy sample is retained in the orifice of the needle.

The Indications for Use statement for the Spectra Soft Tissue Biopsy Needles is not identical to those of the predicate devices; however, the differences do not alter the intended use of the device nor do they affect the safety and efficacy of the device relative to the predicate. Both the submission and predicate devices have the same intended use.

{5}------------------------------------------------

TECHNOLOGICAL COMPARISON TO PREDICATE DEVICES:

The Spectra Soft Tissue Biopsy Needles utilize substantially equivalent technological elements compared to the predicate devices used in this submission:

  • the Spectra Soft Tissue Biopsy Needles are manual soft tissue biopsy needles
  • they use substantially equivalent materials
  • they are sterilized using ethylene oxide
  • same fundamental design: plastic injection molded hub, a stainless steel stylet secured in a ● stylet, a stainless steel cannula, echogenic options, industry-standard grinding and bevel designs
  • same method of operation
  • no change in patient population
  • no change in clinical context
  • no change in intended use ●

PERFORMANCE DATA

The following performance data were provided in support of the substantial equivalence determination.

Biocompatibility testing

The biocompatibility evaluation for the cap was conducted in accordance with the FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO 10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, " May 1, 1995. The submission device and predicate devices are identical. The battery of tests included the following:

  • Cytotoxicity
  • Sensitization
  • Irritation or Intracutaneous Reactivity
  • Systemic Toxicity ●
  • Hemocompatibility

Other performance tests

Physical tests were performed to ensure that the Spectra Soft Tissue Biopsy Needles had standard ISO Luer hubs with applicable luer testing using ISO 594. Needle bond strength tests with hubs were also conducted. The Spectra Soft Tissue Biopsy Needles passed all tests.

CONCLUSION

The Submission Devices, Spectra Soft Tissue Biopsy Needles, have met all established acceptance criteria for performance testing. The Submission Device is substantially equivalent to the predicate device in terms of intended use, design, materials, operation, function, and sterilization method. This demonstrates that the Submission Devices are safe and effective for the intended use, and based on FDA's 510(k) Decision-Making Flowchart is substantially equivalent to the indicated Predicate Devices:

  • Manan Soft Tissue Biopsy Needles K980122
  • Cook, Inc. Quick Core Soft Tissue Biopsy Needle K973565

§ 876.1075 Gastroenterology-urology biopsy instrument.

(a)
Identification. A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.(b)
Classification. (1) Class II (performance standards).(2) Class I for the biopsy forceps cover and the non-electric biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.