K Number
K142714

Validate with FDA (Live)

Date Cleared
2015-01-29

(129 days)

Product Code
Regulation Number
892.1550
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The SonoScape S8 Exp/S9 Pro system is a general-purpose ultrasonic intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Muscular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

The SonoScape S8 Exp/S9 Pro system is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

Device Description

This SonoScape S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body.

This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array with a frequency range of 2.0 MHz to 15.0 MHz. This system consists of a portable console with keyboard control panel, power supply module, color LCD monitor and optional probes.

This system is a portable, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination of these modes, Elastography, 3D/4D.

The subject of this submission is the addition of new indications, new probes and special function to original SonoScape S8 Exp and the addition of SonoScape S9 Pro.

AI/ML Overview

The provided document is a 510(k) premarket notification for the SonoScape S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System. It primarily focuses on demonstrating substantial equivalence to predicate devices rather than proving a device meets specific acceptance criteria through a study with the detailed information requested.

Therefore, many of the requested categories cannot be directly extracted from this document, as it outlines a regulatory submission process based on comparison, not a direct performance study to meet new acceptance criteria for a novel device.

However, I can extract the following information:

1. Table of Acceptance Criteria and Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" in the typical sense of a novel device's performance metrics. Instead, "acceptance" is implicitly tied to demonstrating substantial equivalence to existing predicate devices. Therefore, the "reported device performance" is primarily presented as its similarity and compliance with standards already met by predicate devices.

Acceptance Criteria (Implied)Reported Device Performance (as demonstrated by comparison to predicates)
General-purpose ultrasonic imaging capabilitiesCapable of B-Mode, M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or combinations thereof, Elastography, 3D/4D.
Intended UsesFetal, Abdominal, Pediatric, Small Organ, Cephalic, Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal, Cardiac, Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.
Technical Features (Design, Operation Controls, Display Modes, Measurement Items, Cine Loop, Power Supply, Operating/Storage Conditions, Screen Size)Same as original SonoScape S8 Exp (K132768).
Elastography Function PerformancePerformance met design specifications, including accuracy and repeatability of strain-ratio measurement, as verified by phantom tests.
Safety and Effectiveness (General)Complies with the same regulation and safety standards as predicate devices, consistent acoustic output levels, no new risks raised.
Compliance with specific standardsMet IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-5, ISO10993-10, UD2, and UD3.

2. Sample size used for the test set and the data provenance:

  • Sample Size for Test Set: Not applicable. The document discusses non-clinical tests (laboratory tests, phantom tests) but does not refer to a "test set" of clinical cases or patient data with a defined sample size for performance evaluation in the context of a new acceptance criterion.
  • Data Provenance: Not applicable in the context of a clinical test set. The non-clinical tests were conducted internally by SonoScape Company Limited.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • Not applicable. No clinical test set requiring expert ground truthing is described.

4. Adjudication method for the test set:

  • Not applicable. No clinical test set requiring adjudication is described.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No MRMC comparative effectiveness study was done or mentioned. This device is an ultrasound system, not an AI-assisted diagnostic tool in the sense of improving human reader performance.

6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done:

  • Not applicable. The document describes an ultrasound system, which is inherently a human-operated device. It's not an "algorithm only" device for standalone performance.

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc):

  • For the non-clinical tests mentioned, the "ground truth" would be established by physical measurements and expected outcomes from phantom models and engineering specifications (e.g., accuracy and repeatability of strain-ratio measurement for Elastography).

8. The sample size for the training set:

  • Not applicable. This document does not describe the development of an AI algorithm requiring a training set.

9. How the ground truth for the training set was established:

  • Not applicable. This document does not describe the development of an AI algorithm requiring a training set.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 29, 2015

Sonoscape Company Limited % Mrs. Toki Wu Regulatory Affairs Manager Yizhe Building, Yuquan Road, Nanshan Shenzhen 518051 P.R. CHINA

Re: K142714

Trade/Device Name: S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: September 18, 2014 Received: January 09, 2015

Dear Mrs. Wu:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Robert A Ochs

Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known) K142714

Device Name

S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System

Indications for Use (Describe)

The SonoScape S8 Exp/S9 Pro system is a general-purpose ultrasonic intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic (neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Muscular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

Type of Use (Select one or both, as applicable)
☑ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services

Food and Drug Administration

Office of Chief Information Officer

Paperwork Reduction Act (PRA) Staff

PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

{3}------------------------------------------------

510(k) Summary

1. Submitter [21 CFR807.92 (a) (1)]

Submitter:SonoScape Company Limited
Address:Yizhe Building, Yuquan Road, Nanshan, Shenzhen518051, P.R.China
Contact Person:Toki Wu
Tel:+86 755 26722890
Fax:+86 755 26722850
Email:wusq@sonoscape.net
Date PreparedSeptember 18, 2014

2. Device [21 CFR807.92 (a)(2)]

Trade Name:S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System
Common Name:Diagnostic Ultrasound System and Transducers

Classification Regulatory:

FR NumberProduct Code
Ultrasonic Pulsed Doppler Imaging System892.155090-IYN
Ultrasonic Pulsed Echo Imaging System892.156090-IYO
Diagnostic Ultrasound Transducer892.157090-ITX

Classification Panel: Radiology Device Class: ll

3. Predicate Device(s) [21 CFR 807.92(a) (3)]

The identified predicate devices within this submission are as follows:
SonoScape S8 Exp Portable Digital Color Doppler Ultrasound SystemK132768
SonoScape S30 Digital Color Doppler Ultrasound SystemK132527
Philips EPIQ Diagnostic Ultrasound SystemK132304
Philips CX50 3.0 Diagnostic Ultrasound SystemK123754

4. Device Description [21 CFR 807.92(a) (4)]

This SonoScape S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System is an integrated preprogrammed color ultrasound imaging system, capable of producing high detail resolution intended for clinical diagnostic imaging applications.

{4}------------------------------------------------

The basic principle is that system transmits ultrasonic energy into patient body and implements post processing of received echoes to generate onscreen display of anatomic structures and fluid flow within the body.

This system is a Track 3 device that employs a wide array of probes that include linear array, convex array and phased array with a frequency range of 2.0 MHz to 15.0 MHz. This system consists of a portable console with keyboard control panel, power supply module, color LCD monitor and optional probes.

This system is a portable, general purpose, software controlled, color diagnostic ultrasound system. Its basic function is to acquire ultrasound data and to display the image in B-Mode (including Tissue Harmonic Image), M-Mode, TDI, Color-Flow Doppler, Pulsed Wave Doppler, Continued Wave Doppler and Power Doppler, or the combination of these modes, Elastography, 3D/4D.

The subject of this submission is the addition of new indications, new probes and special function to original SonoScape S8 Exp and the addition of SonoScape S9 Pro.

New indicationsTrans-esoph.(Cardiac), Cerebral Vascular and Laparoscopic.
New probesAdd C613, MPTEE, MPTEE mini, CWD2.0, CWD5.0, PWD2.0 and LAP7 probes.
Special functionElastography.
New ModelAdd S9 Pro model which is identical to S8 Exp except for the difference of probes configuration.

5. Intended Use [21 CFR 807.92(a)(5)]

The SonoScape S8 Exp/S9 Pro system is a general-purpose ultrasonic imaging instrument intended for use by a qualified physician for evaluation of Fetal, Abdominal, Pediatric, Small Organ (breast, testes, thyroid), Cephalic(neonatal and adult), Trans-rectal, Trans-vaginal, Peripheral Vascular, Cerebral Vascular, Musculo-skeletal (Conventional and Superficial), Cardiac (neonatal and adult), Trans-esoph.(Cardiac), Laparoscopic, OB/Gyn and Urology.

6. Comparison with the Predicate Devices [21 CFR 807.92(a) (6)]

S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System is comparable with and substantially equivalent to the predicate devices:

SonoScape S8 Exp Portable Digital Color Doppler Ultrasound SystemK132768
SonoScape S30 Digital Color Doppler Ultrasound SystemK132527
Philips EPIQ Diagnostic Ultrasound SystemK132304

{5}------------------------------------------------

Philips CX50 3.0 Diagnostic Ultrasound System K123754

Intended Use Comparison:

Compared with SonoScape S30 (K132527), the Subject Device S8 Exp/S9 Pro has the same intended uses except for Laparoscopic indication which is comparable and equivalent to the Predicate Device Philips CX50 3.0 (K123754) and Cerebral Vascular indication which is comparable and equivalent to the Predicate Device Philips EPIQ (K132304). The detailed analysis can be found in the probes comparison.

Technical Characteristics Comparison:

The basic and main technical features of the Subject Device S8 Exp/S9 Pro are the same as the original SonoScape S8 Exp (K132768), including Design, Operation Controls, Display Modes, Measurement Items, Cine Loop, Power Supply, Operating and Storage Condition and Screen Size. For Operation Modes, Elastography is a special Operation Mode for the Subject Device S8 Exp/S9 Pro, but already employed by many marketed devices and considered Substantially Equivalent to the Predicate Device Philips EPIQ (K132304) and detailed comparison analysis can be found in Substantial Equivalence Comparison. The detailed technical features can be found in General Device Descriptions of the submission.

Probes Comparison:

Subject device S8 Exp/S9 Pro has the similar probes as the predicated device original SonoScape S8 Exp (K132768), SonoScape S30 (K132527), Philips EPIQ (K132304) and Philips CX50 3.0 (K123754).

Subject devicePredicate DeviceRemark
SonoScape S8 Exp/S9 ProSonoScape S8 Exp
C322 Micro-curved ArrayC322 Micro-curved ArraySame
C344 Curved ArrayC344 Curved Array
C353 Curved ArrayC353 Curved Array
C542 Curved ArrayC354 Curved Array
C362 Curved Array
C542 Curved Array
VC6-2 Curved ArrayVC6-2 Curved ArraySame
C613 Micro-curved ArrayC611 Micro-curved ArraySE
C311 Micro-curved ArrayAnalysis1a)
2P1 Phased Array2P1 Phased ArraySame
2P2 Phased Array2P2 Phased Array
3P1 Phased Array3P1 Phased Array

Table 1 Probes Comparison

{6}------------------------------------------------

5P1 Phased Array5P1 Phased ArraySame
5P2 Phased Array5P2 Phased Array
8P1 Phased Array8P1 Phased Array
L741 Linear ArrayL741 Linear Array
L742 Linear ArrayL742 Linear Array
L743 Linear ArrayL743 Linear Array
L752 Linear ArrayL752 Linear Array
10L1 Linear Array10L1 Linear Array
6V1 Micro-curved Array6V1 Micro-curved Array
6V3 Micro-curved Array6V3 Micro-curved ArraySame
EC9-5 Micro-curved ArrayEC9-5 Micro-curved Array
BCC9-5 Micro-curved ArrayBCC9-5 Micro-curved Array
BCL10-5 Biplane (Curved + Linear Array)BCL10-5 Biplane (Curved + Linear Array)
Subject devicePredicate DeviceRemark
SonoScape S8 Exp/S9 ProSonoScape S30
MPTEE Phased Array(Multi-plane)MPTEE Phased Array(Multi-plane)Same
MPTEE mini Phased Array(Multi-plane)MPTEE mini Phased Array(Multi-plane)
Subject devicePredicate DeviceRemark
SonoScape S8 Exp/S9 ProPhilips CX50 3.0
CWD2.0, 2.0 MHzD2cwc, 2.0 MHzSEAnalysis1b)
CWD5.0, 5.0 MHzD5cwc, 5.0 MHz
Continuous Wave DopplerContinuous Wave Doppler
LAP7, Linear Array, 10.0-5.0 MHzL10-4 lap, Linear Array, 10.0-4.0 MHz
Subject devicePredicate DeviceRemark
SonoScape S8 Exp/S9 ProPhilips EPIQ
PWD2.0, 2.0 MHzD2tcd, 2.0 MHzSE

Note: C344, 2P1, 2P2, 5P1, L741 and 6V1 only for S8 Exp; and 3P1, 5P2 and 8P1 only for S9 Pro.

SE Analysis 1:

  • a) Compared with the predicate device original SonoScape S8 Exp (K132768), there is a new probe (C613) for the subject device. And the frequency, performance and clinical application of C613 probe are the same as C611 for predicate device.
  • b) Compared with the predicate device Philips CX50 3.0 (K123754), the frequency and clinical application of CWD2.0 and CWD5.0 probe is the same as D2cwc and D5cwc cleared for use with predicate device; the clinical application of LAP7 probe is the same

{7}------------------------------------------------

as L10-4 lap cleared for use with predicate device, and though the frequency of them is different, they both meet the clinical use.

  • c) Compared with the predicate device Philips EPIQ (K132304), the frequency and clinical application of PWD2.0 probe is the same as D2tcd cleared for use with predicate device.
    Note: detailed description information about the new probes can be found in Substantial Equivalence Comparison.

Moreover, compared with predicated devices, the subject device (S8 Exp/S9 Pro) complies with the same regulation and safety standards and has the consistent acoustic output levels. Therefore they can be considered Substantially Equivalent in safety and effectiveness, and no new risk is raised, so the SE is not affected.

7. Non-Clinical Tests [21 CFR 807.92(b) (1)]

The S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System has been evaluated for electrical, mechanical, thermal and electromagnetic compatibility safety, biocompatibility and acoustic output.

Laboratory tests were conducted to verify that the S8 Exp/S9 Pro system met all design specifications and the S8 Exp/S9 Pro system conformed to applicable medical device standards. Phantom test was conducted to verify that the strain Elastography function was effective and Elastography performance met design specifications, including accuracy and repeatability of strainratio measurement and etc.

The S8 Exp/S9 Pro system has been designed and manufactured to meet the following standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993-5, ISO10993-10, UD2, and UD3.

8. Clinical Test [21 CFR 807.92(b) (2)]

No clinical testing was required.

9. Substantially Equivalent Conclusions [21 CFR 807.92(b) (3)]

In accordance with the 21 CFR Part 807 and based on the information provided in this premarket notification, SonoScape Company Limited concludes that S8 Exp/S9 Pro Portable Digital Color Doppler Ultrasound System is substantially equivalent to the predicate devices with regard to safety and effectiveness.

§ 892.1550 Ultrasonic pulsed doppler imaging system.

(a)
Identification. An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.(b)
Classification. Class II.