K Number
K142510

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-11-19

(72 days)

Product Code
Regulation Number
886.5916
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

UltraHealth™ (petrafocon A hem-larafilcon A) Hybrid Contact Lenses for keratoconus are indicated for the correction of hyperopic, myopic and astigmatic refractive error including presbyopia that manifest irregular corneas or irregular astigmatism, in aphakic and not aphakic, and otherwise non-diseased eyes. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with irregular astigmatism up to 6.00 D. For presbyopia, add powers between +1.00 and +4.00 D. The lenses may be disinfected using a chemical disinfecting system compatible with both silicone-hydrogel and rigid gas permeable lenses.

UltraHealth™ Flat Cornea SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lenses are for use in the correction of eyes with refractive errors resulting from corneal surgery or trauma including hyperopia and myopia, astigmatism and irregular astigmatism in aphakic, nondiseased eyes, with or without presbyopia. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with irregular astigmatism up to 6.00 D. For presbyopia, add powers between +1.00 and +4.00 D. The lenses may be disinfected using a chemical disinfecting system compatible with both silicone-hydrogel and rigid gas permeable lenses.

Device Description

The UltraHealth™ SiH Hybrid Contact lens for Keratoconus and the UltraHealth™ Flat Cornea SiH Hybrid Contact Lenses are contact lenses that combine a gas permeable optic surrounded by a soft hydrophilic skirt that straddles the limbus of the eye:

  • in the power range of -20.00 to +20.00 diopters for sphere,
  • with center thickness from 0.16 mm to 0.39 mm
  • with base curves of 5.5 mm to 9.00 mm
  • with diameter of 14.50mm

When placed on the human cornea, the UltraHealth™ Hybrid Contact lens for Keratoconus and the UltraHealth™ Flat Cornea act as a refracting medium to focus light rays onto the retina. The devices are available as lathe cut contact lenses in the following designs: spherical, toric, multifocal, and aspheric surfaces in blue visibility tinted material.

The optics of the UltraHealth™ Hybrid Contact lens for Keratoconus and the UltraHealthI™ Flat Cornea lenses are a gas permeable polymer material (petrafocon A). The soft skirt is comprised of silicone hydrogel (hem-larafilcon A) with a 32% water content. The junction between the rigid material and soft material is bound by a proprietary method.

AI/ML Overview

This document describes the premarket notification for UltraHealth™ SiH and UltraHealth™ Flat Cornea SiH Hybrid Contact Lenses. It details the device, its intended use, and its substantial equivalence to previously approved devices.

Here's a breakdown of the acceptance criteria and study information provided:

1. Acceptance Criteria and Reported Device Performance

The document doesn't explicitly state "acceptance criteria" in a tabulated format with numerical targets. Instead, it relies on demonstrating substantial equivalence to predicate devices that have already been proven safe and effective. The "reported device performance" is essentially that the new lenses are equivalent in design and material to previously cleared safe and effective lenses. The table below summarizes the characteristics of the new lenses compared to predicate and reference lenses, implying that similarity to the predicate is the primary criterion for acceptance.

CharacteristicUltraHealth™ SiH Hybrid Contact Lenses for Keratoconus (New Lens)Synergeyes KC (Predicate)Synergeyes SiH (Reference)
Material (RGP Center)petrafocon Apaflufocon Dpetrafocon A
Material (Soft Skirt)hem-larafilcon Ahem-iberfilcon Ahem-larafilcon A
UV BlockingYesNoYes
Base Curves5.5 mm to 9.0 mm5.5 to 7.0 mm7.1 to 9.0 mm
Base Curve Chord6.0 mm to 6.5 mm6.0 mm6.0 mm to 6.5 mm
DesignReverse geometry with anterior aspheric surfaceReverse geometry with anterior aspheric surfaceStandard geometry with anterior aspheric surface
RGP Center8.40mm8.40mm8.40 mm
Diameters14.5mm14.5mm14.5 mm
Power Range-20.00 to +20.00-20.00 to +20.00-20.00 to +20.00
Add Powers (multifocal)+1.00 D to +4.00 D+1.00 D to +4.00 D+1.00 D to +4.00 D
Refractive Index (RGP)1.4421.4421.442
Refractive Index (S/H Skirt)1.4351.4481.435
Specific Gravity (RGP)1.151.101.15
Hardness (Shore D) RGP767976
Modulus (kgf/cm²) RGP Center14515.1 ±32411137.9 ± 36714515.1 ±324
TintVisibility violet (D&C 2)Visibility green (D&C Green #6)Visibility violet (D&C 2)
Water Content (Soft Skirt)<1%<1%28%
Lens TypeGroup 1 Low WaterGroup 1 Low WaterGroup 1 Low Water
CharacteristicUltraHealth™ Flat Cornea SiH Hybrid Contact Lenses (New Lens)SynergEyes® PS (Predicate)Synergeyes® SiH (Reference)
Material (RGP Center)petrafocon Apaflufocon Dpetrafocon A
Material (Soft Skirt)hem-larafilcon Ahem-iberfilcon Ahem-larafilcon A
UV BlockingYesNoYes
Base Curves5.5 mm to 9.0 mm5.5 to 7.0 mm7.1 to 9.0 mm
Base Curve Chord6.0 mm to 6.5 mm6.0 mm6.0 mm to 6.5 mm
DesignReverse geometry with anterior aspheric surfaceReverse geometry with anterior aspheric surfaceStandard geometry with anterior aspheric surface
RGP Center8.40mm8.40mm8.40 mm
Diameters14.5mm14.5mm14.5 mm
Power Range-20.00 to +20.00-20.00 to +20.00-20.00 to +20.00
Add Powers (multifocal)+1.00 D to +4.00 DNone+1.00 D to +4.00 D
Refractive Index (RGP)1.4421.4421.442
Refractive Index (S/H Skirt)1.4351.4481.435
Hardness (Shore D) RGP767976
Modulus (kgf/cm²) RGP Center14515.1 ±32411137.9 ± 36714515.1 ±324
TintVisibility violet (D&C 2)Visibility green (D&C Green #6)Visibility violet (D&C 2)
Water Content (Soft Skirt)<1%<1%28%
Lens TypeGroup 1 Low WaterGroup 1 Low WaterGroup 1 Low Water

The key "performance" is that the new lenses are made of materials and have designs that match, or are substantially equivalent to, previously cleared devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document refers to "[c]linical studies submitted in K052675 and K060102" which "were shown to be safe and effective." These predicate devices' studies were conducted "per the FDA Contact Lens Daily Wear Guidance Document, May 1994."

  • Sample size: Not explicitly stated in this document. It refers to the clinical studies of predicate devices (K052675 and K060102) without detailing their sample sizes.
  • Data provenance: Not explicitly stated in this document. Details would be in the original submissions for K052675 and K060102. It's safe to assume they were prospective clinical studies, given the nature of a medical device approval process and the reference to FDA guidance for contact lens clinical trials.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

This is not applicable as this is a contact lens submission, not an AI or diagnostic imaging device submission. Safety and effectiveness are determined through clinical trials and material biocompatibility, not by expert interpretation of data for ground truth establishment in a diagnostic context.

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

Not applicable for this type of medical device submission. Clinical trial outcomes are typically assessed by investigators and monitored for adverse events and visual acuity changes, not through an adjudication method like those used for diagnostic device ground truth.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

Not applicable. This is not an AI-assisted diagnostic device.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

Not applicable. This is a physical medical device (contact lens).

7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

For the predicate devices, safety and effectiveness were established through outcomes data from clinical studies. This would typically include:

  • Visual acuity measurements with the lenses.
  • Assessment of ocular health (e.g., absence of adverse events like corneal staining, infections, edema) by optometrists/ophthalmologists.
  • Patient comfort and satisfaction.

8. The sample size for the training set

Not applicable. This is a contact lens submission, not an AI/machine learning device. The "training set" conceptually for a physical device would be the product development and testing phases, which do not involve a "training set" in the AI sense.

9. How the ground truth for the training set was established

Not applicable. See point 8.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring three stylized human profiles facing to the right, representing health and human services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 19, 2014

Synergeyes, Inc. Steven L. Ziemba, M.Sc. V.P., Quality Assurance & Regulatory Affairs 2232 Rutherford Road Carlsbad, CA 92008

Re: K142510

Trade/Device Name: UltraHealth™ SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lenses for Keratoconus UltraHealth™ Flat Cornea SiH (petrafocon A Hem-larafilcon A) Hybrid Contact Lenses Regulation Number: 21 CFR 886.5916 Regulation Name: Rigid Gas Permeable Contact Lenses Regulatory Class: Class II Product Code: HQD Dated: September 18, 20014 Received: September 22, 2014

Dear Mr. Ziemba:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act

{1}------------------------------------------------

Page 2 - Steven L. Ziemba, M.Sc.

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Kesia Y. Alexander -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known) K142510

Device Name

Device Names: UltraHealth SiH (petrafocon A hem-larafilcon A) Hybrid Contact lens for Keratoconus UltraHealth Flat Cornea SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lenses

Indications for Use (Describe)

UltraHealth™ (petrafocon A hem-larafilcon A) Hybrid Contact Lenses for keratoconus are indicated for the correction of hyperopic, myopic and astigmatic refractive error including presbyopia that manifest irregular corneas or irregular astigmatism, in aphakic and not aphakic, and otherwise non-diseased eyes. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with irregular astigmatism up to 6.00 D. For presbyopia, add powers between +1.00 and +4.00 D. The lenses may be disinfected using a chemical disinfecting system compatible with both silicone-hydrogel and rigid gas permeable lenses.

UltraHealth™ Flat Cornea SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lenses are for use in the correction of eyes with refractive errors resulting from corneal surgery or trauma including hyperopia and myopia, astigmatism and irregular astigmatism in aphakic, nondiseased eyes, with or without presbyopia. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with irregular astigmatism up to 6.00 D. For presbyopia, add powers between +1.00 and +4.00 D. The lenses may be disinfected using a chemical disinfecting system compatible with both silicone-hydrogel and rigid gas permeable lenses.

Type of Use (Select one or both, as applicable)

☒ Prescription Use (Part 21 CFR 801 Subpart D)☐ Over-The-Counter Use (21 CFR 801 Subpart C)
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

CONTINUE ON A SEPARATE PAGE IF NEEDEDThis section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY K142510

Date Prepared

04 November 2014

Submitter: Company Name: Address: Telephone: Fax:

Synergeyes, Inc. 5927 Priestly Avenue, Suite 210, Carlsbad, CA 92008 (760) 476-9410 (760) 431-0828

Official Correspondent:

Name:Steven L. Ziémba, M. Sc.
Title:V.P., Quality Assurance, Clinical & Regulatory Affairs, Synergeyes, Inc.
Address:2232 Rutherford Road, Carlsbad, CA 92008
E-mail:sziemba@synergeyes.com
Telephone:(760) 444-9687
Fax:(877) 819-9279

Establishment Registration Number: 3005087645

Device Name: Trade & USAN Names: TM UltraHealth SiH Hybrid Contact Lenses for Keratoconus TM UltraHealth Flat Cornea SiH Hybrid Contact Lenses

Contact Lens Common Name: Classification Panel: Ophthalmic Device Classification: Daily Wear Soft (hydrophilic)/RGP (hydrophobic) Combination Contact lens, Class II (21 CFR 886.5916) Device Product Code: 86 HQD

Indications for Use:

The UltraHealth™ Hybrid Contact lens for Keratoconus for daily wear are indicated for use in the correction of hyperopic, myopic and astigmatic refractive error including presbyopia, in aphakic and not aphakic, non-diseased eyes. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with astigmatism up to 6.00 D. For presbyopia, add powers between +1.00 and +4.00 D. The lenses may be disinfected using only a chemical disinfecting system compatible with both hydrogel and rigid gas permeable lenses.

The UltraHealth™ Flat Cornea Hybrid Contact Lenses for daily wear are indicated for use in the correction of eves with refractive errors resulting from corneal surgery or trauma including hyperopia and myopia, astigmatism, and irregular astigmatism, including presbyopia, in aphakic and not aphakic, non-diseased eyes. The lenses are indicated for daily wear for the correction of up to +20.00 and -20.00 D in eyes with astigmatism up to 6.00 D. For presbyopia, add powers between +1.00 and

{4}------------------------------------------------

+4.00 D. The lenses may be disinfected using only a chemical disinfecting system compatible with both hydrogel and rigid gas permeable lenses.

Device Description

The UltraHealth™ SiH Hybrid Contact lens for Keratoconus and the UltraHealth™ Flat Cornea SiH Hybrid Contact Lenses are contact lenses that combine a gas permeable optic surrounded by a soft hydrophilic skirt that straddles the limbus of the eye:

  • in the power range of -20.00 to +20.00 diopters for sphere, ●
  • with center thickness from 0.16 mm to 0.39 mm
  • with base curves of 5.5 mm to 9.00 mm
  • with diameter of 14.50mm ●

When placed on the human cornea, the UltraHealth™ Hybrid Contact lens for Keratoconus and the UltraHealth™ Flat Cornea act as a refracting medium to focus light rays onto the retina. The devices are available as lathe cut contact lenses in the following designs: spherical, toric, multifocal, and aspheric surfaces in blue visibility tinted material.

The optics of the UltraHealth™ Hybrid Contact lens for Keratoconus and the UltraHealthI™ Flat Cornea lenses are a gas permeable polymer material (petrafocon A). The soft skirt is comprised of silicone hydrogel (hem-larafilcon A) with a 32% water content. The junction between the rigid material and soft material is bound by a proprietary method.

Substantial Equivalence

For design: The hybrid (RGP Center and soft skirt) reverse geometry designs were cleared for the predicate SynergEyes™ (paflufocon D-hem iberfilcon A) Hybrid Contact Lenses in K056275 and K060102. The new lenses, the UltraHealth™ SiH (petrafocon A hem-larafilcon A) hybrid contact lenses for keratoconus, and the UltraHealth™ Flat Cornea SiH (petrafocon A hem-larafilcon A) hybrid contact lenses have the same substantially equivalent reverse geometry lens designs as the predicate lenses.

For material: The lens materials are comprised of a silicone Rigid Gas Permeable (RGP) center optic, and a soft silicone hydrogel skirt that surrounds the center optic zone which match the materials cleared under K083921. Biocompatibility, optical performance and safety of the materials for use in daily wear contact lenses were addressed in that submission.

For Optics:

The aspheric anterior optic surface matches that of the lenses cleared under K083921 and K051035.

{5}------------------------------------------------

Characteristics:
UltraHealth™ SiH Hybrid Contact Lenses for Keratoconus

NEW LENSPREDICATE LENSREFERENCE LENS
Lens CharacteristicsUltraHealth™ SiH HybridContact Lensesfor KeratoconusSynergeyes KC (paflufocon Dhem-iberfilcon A) Hybrid ContactLens for Keratoconus (K052675)Synergeyes SiH (petrafoconA hem-larafilcon A) HybridContact Lens (K083921)
ManufacturerSynergeyes, Inc.Synergeyes, Inc.Synergeyes, Inc.
MaterialRGP Center: petrafocon ASoft Skirt: hem-larafilcon ARGP Center: paflufocon DSoft Skirt: hem-iberfilcon ARGP Center: petrafocon ASoft Skirt: hem-larafilcon A
UV BlockingYesNoYes
Base Curves (varieswith vault)5.5 mm to 9.0 mm5.5 to 7.0 mm7.1 to 9.0 mm
Base Curve Chord6.0 mm to 6.5 mm6.0 mm6.0 mm to 6.5 mm
DesignReverse geometry withanterior aspheric surfaceReverse geometry with anterioraspheric surfaceStandard geometry withanterior aspheric surface
RGP Center8.40mm8.40mm8.40 mm
Diameters:14.5mm14.5mm14.5 mm
Power Range-20.00 to +20.00-20.00 to +20.00-20.00 to +20.00
Add Powers (formultifocal)+1.00 D to +4.00 D+1.00 D to +4.00 D+1.00 D to +4.00 D
Refractive Index(RGP)RGP Center:1.442S/H Skirt:1.435RGP Center:1.442S/H Skirt:1.448RGP Center:1.442S/H Skirt:1.435
Specific Gravity(RGP)1.151.101.15
Hardness (Shore D)RGP: 76RGP: 79RGP: 76
Modulus (kgf/cm²)RGP Center14515.1 ±32411137.9 ± 36714515.1 ±324
TintVisibility violet (D&C 2)Visibility green (D&C Green #6)Visibility violet (D&C 2)
Water Content(Soft Skirt)<1%<1%28%
Lens TypeGroup 1 Low WaterGroup 1 Low WaterGroup 1 Low Water

{6}------------------------------------------------

UltraHealth™ Flat Cornea

NEW LENSPREDICATE LENSREFERENCE LENS
Lens CharacteristicsUltraHealth™ Flat CorneaSiH Hybrid ContactLensesSynergEyes® PS (paflufocon Dhem-iberfilcon A) Hybrid ContactLenses for Post Surgical RefractiveError and Trauma (K060102)Synergeyes® SiH(petrafocon A hem-larafilconA) Hybrid Contact Lens(K083921)
ManufacturerSynergeyes®, Inc.Synergeyes®, Inc.Synergeyes®, Inc.
MaterialRGP Center: petrafocon ASoft Skirt: hem-larafilcon ARGP Center: paflufocon DSoft Skirt: hem-iberfilcon ARGP Center: petrafocon ASoft Skirt: hem-larafilcon A
UV BlockingYesNoYes
Base Curves (varies withvault)5.5 mm to 9.0 mm5.5 to 7.0 mm7.1 to 9.0 mm
Base Curve Chord6.0 mm to 6.5 mm6.0 mm6.0 mm to 6.5 mm
DesignReverse geometry withanterior aspheric surfaceReverse geometry with anterior asphericsurfaceStandard geometry withanterior aspheric surface
RGP Center8.40mm8.40mm8.40 mm
Diameters:14.5mm14.5mm14.5 mm
Power Range-20.00 to +20.00-20.00 to +20.00-20.00 to +20.00
Add Powers (formultifocal)+1.00 D to +4.00 DNone+1.00 D to +4.00 D
Refractive Index (RGP)RGP Center:1.442S/H Skirt:1.435RGP Center:1.442S/H Skirt:1.448RGP Center:1.442S/H Skirt:1.435
Hardness (Shore D)RGP: 76RGP: 79RGP: 76
Modulus (kgf/cm²) RGPCenter14515.1 ±32411137.9 ± 36714515.1 ±324
TintVisibility violet (D&C 2)Visibility green (D&C Green #6)Visibility violet (D&C 2)
Water Content (Soft Skirt)<1%<1%28%
Lens TypeGroup 1 Low WaterGroup 1 Low WaterGroup 1 Low Water

Non-Clinical Studies

The lens materials are comprised of the same silicone Gas Permeable optic, and a soft silicone hydrogel skirt that surrounds the optic as was described in K083921. Biocompatibility testing (including Cytotoxicity, Ocular Irritation, Systemic Toxicity, and 21 Day Ocular Safety in Rabbits) as well as chemical, mechanical and optical characteristics, clinical performance and safety of the silicone hydrogel materials for use in daily wear contact lenses were addressed in K083921 per the FDA Contact Lens Daily Wear Guidance Document, May 1994.

Clinical Studies

The safety and effectiveness of the design of the new lenses is the same as the design for lenses cleared under K056275 and K060102. Daily wear contact lenses for keratoconus with the Synergeyes reverse geometry design were shown to be safe and effective in clinical studies submitted in K052675 and daily wear contact lenses for post surgical refractive error and trauma were shown to be safe and effective in clinical studies submitted in K060102; both conducted per the FDA Contact Lens Daily Wear Guidance Document, May 1994.

{7}------------------------------------------------

The materials, processing, lens specifications, packaging and sterilization for the UltraHealth™ SiH (petrafocon A hem-larafilcon A) Hybrid Contact lens for Keratoconus and the UltraHealth™ Flat Cornea SiH (petrafocon A hem-larafilcon A) Hybrid Contact Lenses are identical to those described in K083921, which includes data to support that the materials and processing methods are biocompatible, safe and effective for use in daily wear contact lenses.

Packaging

The primary lens container for shipping is a sterile enclosed medical grade glass vial with a screw cap and rubber stopper and sterilized using an e-beam process. The lens is immersed in a sterile, phosphate-buffered saline.

Manufacturing & Sterilization

The same manufacturing and sterilization locations described for the predicate lenses are used for these new lenses.

Relationship to Special Controls (Guidance)

The Contact Lens Daily Wear Guidance Document, May 1994, is the relevant document to which this submission is based.

§ 886.5916 Rigid gas permeable contact lens.

(a)
Identification. A rigid gas permeable contact lens is a device intended to be worn directly against the cornea of the eye to correct vision conditions. The device is made of various materials, such as cellulose acetate butyrate, polyacrylate-silicone, or silicone elastomers, whose main polymer molecules generally do not absorb or attract water.(b)
Classification. (1) Class II if the device is intended for daily wear only.(2) Class III if the device is intended for extended wear.
(c)
Date PMA or notice of completion of a PDP is required. As of May 28, 1976, an approval under section 515 of the act is required before a device described in paragraph (b)(2) of this section may be commercially distributed. See § 886.3.