K Number
K142435

Validate with FDA (Live)

Date Cleared
2015-02-20

(175 days)

Product Code
Regulation Number
880.5860
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The 3S Safety Syringe with/without Needle is a sterile, single-use, disposable and non-reusable, manual retractable safety syringe intended for injection of fluids into the body, while reducing the risk of sharps injuries and the potential for syringe reuse.

Device Description

The 3S Safety Syringe with/without Needle is an integrated needle and piston syringe with an innovative anti-needle-stick mechanism. The design incorporates the ideal features desired in a safety syringe. No special techniques are required to use the safety mechanism. The mechanism allows clear visualization of the injection site at all times. The mechanism clearly shows the needle is contained within the syringe barrel. After standard techniques for injection, the plunger is withdrawn, snapped off and used syringe is discarded into a sharps container per regulatory provision.

AI/ML Overview

This document is primarily a 510(k) summary for a medical device (3S Safety Syringe). It focuses on establishing substantial equivalence to a predicate device rather than detailing extensive clinical study results as one might find in a PMA or a more comprehensive clinical trial report.

Therefore, many of your requested items, particularly those related to AI algorithm performance, sample sizes for test/training sets in an AI context, expert ground truth establishment for AI, MRMC studies, and standalone AI performance, are not applicable to this document as it describes a physical safety syringe, not an AI/software device.

However, I can extract the acceptance criteria and performance related to the physical device.


1. Table of Acceptance Criteria and Reported Device Performance

For the 3S Safety Syringe, the acceptance criteria are generally based on conformity to international standards and regulations. The "reported device performance" is essentially a statement of this conformity.

Acceptance Criteria (Standard/Requirement)Reported Device Performance (Conformity)
ISO 7864 (Sterile hypodermic needles)Conforms to ISO 7864
ISO 7886 (Sterile hypodermic syringes)Conforms to ISO 7886
ISO 10993 (Biological evaluation of medical devices)Conforms to ISO 10993
21 CFR Part 801 (Labeling requirements)Meets the requirements of 21 CFR Part 801
Sharps Injury Prevention FeaturesManual Retractable (similar to predicate)
Intended UseIntended for injection, reducing sharps injuries, and preventing reuse (similar to predicate)
Material CompatibilityMaterials (PP, Silicone, Stainless Steel) are consistent with predicate and suitable for medical use.

Note: This table reflects the information present in the provided document, which is a 510(k) summary. It asserts equivalence to established standards rather than providing raw performance data or a detailed study report.


Information Not Applicable or Not Provided in the Document:

The following information is not contained within the provided 510(k) summary, as it pertains to AI/software device studies or detailed clinical trial breakdowns not typically required for a Class II physical medical device 510(k) submission focused on substantial equivalence.

  1. Sample size used for the test set and the data provenance: Not applicable for this type of physical device submission. Performance is assessed against engineering standards, not a data test set in the AI sense.
  2. Number of experts used to establish the ground truth for the test set and the qualifications of those experts: Not applicable. Ground truth for a physical syringe's performance is established by engineering testing against standards (e.g., fluid leakage, plunger force, sterility, biocompatibility), not by expert consensus on data interpretation.
  3. Adjudication method for the test set: Not applicable.
  4. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance: Not applicable, as this is not an AI device.
  5. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done: Not applicable.
  6. The type of ground truth used: For a physical device like a syringe, "ground truth" refers to the objective physical and chemical properties and performance characteristics measured against established international standards (ISO, biological safety) and regulatory requirements (CFR). It's not an "expert consensus" or "pathology" in the diagnostic sense.
  7. The sample size for the training set: Not applicable, as there is no AI training set.
  8. How the ground truth for the training set was established: Not applicable.

{0}------------------------------------------------

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 20, 2015

Sincere Medical Device Company, Ltd. Mr. Garfield Wang RA Manager Room 218, 2F., No. 183, Zhouzi St. Taipei City, Taipei City 114 TAIWAN (R.O.C)

Re: K142435

Trade/Device Name: 3S Safety Syringe with/without needle Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: August 15, 2015 Received: January 23, 2015

Dear Mr. Wang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang
-s

for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo of Sincere Medical Device Co., Ltd. The logo consists of a green circular emblem with the letter 'S' inside, surrounded by stylized green leaves. To the right of the emblem is the company name, 'Sincere Medical Device Co., Ltd.,' written in a clear, sans-serif font.

510(k) Submission

Rev 0.00 15/08/14

Section_004 Indications for Use

K142435 510(k) Number (if known):

Device Name: 3S SAFETY SYRINGE WITH/WITHOUT NEEDLE

Indications for Use

[3S Safety Syringe]

The 3S Safety Syringe with/without Needle is a sterile, single-use, disposable and non-reusable, manual retractable safety syringe intended for injection of fluids into the body, while reducing the risk of sharps injuries and the potential for syringe reuse.

Prescription Use V AND/OR Over-The-Counter Use (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo of Sincere Medical Device Co., Ltd. The logo consists of a green circular emblem with the letter "S" inside, along with the company name written in black text to the right of the emblem. The text is clear and easy to read, and the overall design is simple and professional.

510(k) Submission

Rev 0.02

05/02/15

[3S Safety Syringe]

Section_005 510(K) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA and 21 CFR §807.92

Date Prepared: 15. 08.2014

1. Submitter Name and Address:

Owner Name:Sincere Medical Device Co., Ltd.
Address:Rm. 218, 2F., No. 183, Zhouzi St., Neihu Dist., Taipei City 114.Taiwan(R.O.C)
Contactor Name:Garfield Wang, Vivian Lu
TEL:+86-13564751751
E-mail:wangxuebo_11@hotmail.com
Web:http://sincere-med.com
Fax:+886-2-2659-8698

US Agent:

US Agent:Benjamin Lee.
Address:8692 9th Street, unit 42, Rancho Cucamonga, ca 91730
TEL:1909-985-0648

2. Submission Devices Information:

Trade/Proprietary Name: 3S Safety Syringe with/without Needle Common Name: Syringe, Antistick Piston Syringe Classification name: Piston Syringe.. Class: 2. Panel: 80. Product codes: MEG Submission Type: 510(k) Requlation Number: 880.5860

3. Predicate Devices Information:

Trade Name:Otter Safety Syringe
510(K) Number:K040545

4. Devices Description:

The 3S Safety Syringe with/without Needle is an integrated needle and piston syringe with an innovative anti-needle-stick mechanism. The design incorporates the ideal features desired in a safety syringe. No special techniques are required to use the safety mechanism. The mechanism allows clear visualization of the injection site at all times. The mechanism clearly shows the needle is contained within the syringe barrel. After standard techniques for injection, the plunger is withdrawn, snapped off and used syringe is discarded into a sharps container per regulatory provision.

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo of Sincere Medical Device Co., Ltd. The logo consists of a green circular emblem with the letter 'S' inside, accompanied by the company name in a simple, sans-serif font. The text is in a dark gray color, providing a clear contrast against the white background.

510(k) Submission

[3S Safety Syringe]

Rev 0.02 05/02/15

The 3S safety syringe with/without Needle will be available in numerous sizes and combinations between the smallest (0.5cc/ml + 28G x 1/2" to 20G x 1 1/2") and the largest (10cc/ml)

5. Intended Use:

The 3S Safety Syringe with/without Needle is a sterile, single-use, disposable and nonreusable, manual retractable safety syringe intended for injection of fluids into the body, while reducing the risk of sharps injuries and the potential for syringe reuse.

6. Technological Characteristics:

Through comparisons between the submitted devices with the predicate devices as follows tables. We believe the applicant devices are substantially equivalent with the predicate devices.

Element of ComparisonSubmission DevicePredicate DeviceK040545
Intended UseThe 3S Safety Syringe is anintegrated needle and pistonsyringe with an innovative anti-needle-stick mechanism. Thedesign incorporates the idealfeatures desired in a safetysyringe. No special techniquesare required to use the safetymechanism. The mechanismallows clear visualization of theinjection site at all times. Themechanism clearly shows thatthe needle has been crushedwithin the syringe barrel. Afterstandard techniques forinjection, the plunger iswithdrawn completely into thebarrel. Snap off plunger anddiscard used syringe in a sharpcontainer per regulatoryprovision.The Otter Safety Syringe is anintegrated needle and pistonsyringe with an innovative anti-needle-stick mechanism. Thedesign incorporates the idealfeatures desired in a safetysyringe. No special techniquesare required to use the safetymechanism. The mechanismallows clear visualization of theinjection site at all times. Themechanism clearly shows thatthe needle has been crushedwithin the syringe barrel. Afterstandard techniques forinjection, the plunger iswithdrawn completely into thebarrel. A forward compressionof the plunger again and theneedle is crushed within thepuncture-resistant barrel. Thisrenders the needle unusableand safe from accidentalneedle sticks.
Principle of OperationNormalNormal
Syringe CapacityVarious SizesVarious Sizes
Nozzle TypeLuer Slip & Luer LockLuer Slip & Luer Lock
Lubricant for BarrelSilicone OilSilicone Oil
Barrel TransparencyTransparent and ClearTransparent and Clear
Gradations LegibilityLegibleLegible

Safety Syringe Comparison Table

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows a logo with a green circular design. Inside the circle is a stylized letter 'S' in brown. Below the circle, the words 'SINCERE MEDICAL' are written in a curved fashion, following the shape of the circle. The logo appears to be for a medical organization or company.

Sincere Medical Device Co., Ltd.

510(k) Submission

[3S Safety Syringe]

Rev 0.02 05/02/15

Materials
BarrelPPPP
PlungerPPPP
PistonSILICONESILICONE
Needle HubPPPP
NeedleStainless SteelStainless Steel
Needle SheathPPPP
Needle Gauge and LengthVarious SizesVarious Sizes
Lubricant for NeedleSilicone OilSilicone Oil
Sharps Injury Prevention FeaturesManual RetractableManual Retractable
PerformancesConforms to ISO7864 ISO7886Conforms to ISO7864 ISO7886
BiocompatibilityConforms to ISO10993Conforms to ISO10993
LabelingMeet the requirements of 21 CFR Part 801Meet the requirements of 21 CFR Part 801

7. Conclusion:

The materials, performance, and operational features of both the submitted device and the predicate device are substantially equivalent.

END

§ 880.5860 Piston syringe.

(a)
Identification. A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.(b)
Classification. Class II (performance standards).