K Number
K142434

Validate with FDA (Live)

Date Cleared
2014-11-21

(84 days)

Product Code
Regulation Number
892.5050
Age Range
2 - 120
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.

The linear accelerator systems are high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT), modulated arc therapy (mARC) and precision stereotactic radiation therapy for lesions, tumors and conditions anywhere in the head and body where radiation therapy is indicated.

The syngo® RT Therapist Workspace is a component of the linear accelerator system and is based on the syngo® architecture. The syngo® RT Therapist workspace contains software applications to support patient selection/setup. patient positioning verification. treatment delivery/verification, and treatment recording.

The syngo® RT Oncologist Workspace is an optional accessory to the linear accelerator system and permits localization, contouring, segmentation, image review, and review and approval of treatment plan parameters. In addition, it includes tools and administrative functions to aid in the diagnosis, staging, and prescription of radiation therapy.

The syngo@ RT Therapist and the syngo@ RT Oncologist Workspaces v4.3.1 MR1, can be interfaced with third party OIS, V&R, TPS, and PACS devices conforming to the DICOM Standard.

Device Description

The project Sys_VC10C, Phase 2 software update for the ARTISTE™ Solution linear accelerator system with Control Console v13.0.301 with the syngo® RT Therapist Workspace v4.3.1 MR1 release is intended to update customers with the currently cleared and released syngo® RT Therapist Workspace software with versions v4.1, v4.2, v4.3 or v4.3.1 for ARTISTE, and is backwards compatible with the ONCOR & PRIMUS Linear Accelerator systems with the COHERENCE RT Therapist v2.3. The currently cleared syngo® RT Therapist workspace software v4.3.1 supports Siemens branded and third party OIS, V&R, TPS, and PACS systems.

Additionally, the Phase 2 software update includes software modifications and defect fixes for the syngo® RT Oncologist workspace. The Oncologist software is developed on the same syngo® software code platform as the RT Therapist and the workspace differentiation is license key controlled for clinically focused workspace options. The syngo® RT Oncologist workspace is considered a medical device in its own right had has multiple previous FDA 510(k) clearances.

Note: Several syngo® clinical modules are considered medical devices in their own right and may have separate clearances. For example, the syngo® VSim module, v2.7 received FDA clearance under K101119 and is licensed on several syngo® workspaces for different modalities. The syngo® RT Oncologist Workspace has hosted this module under multiple clearances with the latest being the current predicate, K121295.

These previous software versions can be directly migrated to the current release (syngo® RTT/ONC v4.3.1 MR1) and Control Console v13.0.301. The ARTISTE Solution with VC10C. Phase 1 and the syngo® RT Therapist v4.3.1 cleared under K132935 serves as a predicate in this submission.

The syngo® RT Oncologist workspace is an optional system for Siemens Linac customers. The syngo® RT Oncologist also hosts software based products such as the VSim v2.7 feature [VSIM] and other oncology based features. The syngo® RT Oncologist workspace v4.3 was previously cleared under K121295 which serves as a predicate in this submission.

The technological characteristics and the fundamental technology of the LINAC Control Console and the syngo® RT Therapist Workspace v4.3.1 MR1 and the syngo® RT Oncologist Workspace v4.3.1 MR1 (including the syngo® VSim module) remain unchanged from the currently cleared products. This is a software update only. The materials of use remain unchanged from the predicate devices.

AI/ML Overview

The provided text describes a 510(k) premarket notification for a software update to a medical device, specifically the Siemens ARTISTE Solution. The document contains information about acceptance criteria and how the device meets them through verification and validation activities.

Here's a breakdown of the requested information based on the provided text:

1. A table of acceptance criteria and the reported device performance

Acceptance Criteria CategoryReported Device Performance
Software Performance & FunctionalityAll testable requirements in the System Requirements Specifications (SRS), Sub-system Requirements Specifications (SSRS), and Component Requirements (CRS) for Sys VC10C project, Phase 2, and specific requirements for third-party OIS implementation have been successfully verified and traced. Software verification and regression testing have been performed successfully to meet their previously determined acceptance criteria as stated in the Test Concept.
System Level Validation (Clinical Workflow)Validation of the syngo® RT Therapist Workspace v4.3.1 MR1, the syngo® RT Oncologist Workspace v4.3.1 MR1 with hosting of optional third-party OIS, and regression testing with Control Consoles 13.0.301 has been performed at the System test level on production prototype devices. System level validation and regression testing has been performed successfully, demonstrating that the software meets the acceptance criteria as noted in the system test plans.
Risk Management & SafetyRisk management is ensured via a risk analysis, identifying potential hazards and mitigations. These hazards are controlled by software means, user instructions, verification of requirements, and validation of clinical workflow. The update does not introduce any new potential safety risks.
Compliance to Industry StandardsThe software update and Control Console have been tested (as needed) to meet the requirements for conformity (where applicable) to multiple industry standards. The R&D evaluation of relevant testing to consensus standards is documented.
Substantial Equivalence (to Predicate Devices)The verification testing to system requirements, validation of intended use, and regression testing to existing functional requirements, support the claim of substantial equivalence to previously cleared predicate devices (K121295 and K132935). The update has the same intended use, indications for use, and fundamental scientific technical characteristics as the predicate devices.

2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

The document mentions "production prototype devices" were used for system-level validation and regression testing. However, it does not specify the sample size for the test set (e.g., number of cases, number of devices tested).

Regarding data provenance, the document does not provide information on the country of origin of the data or whether the testing was retrospective or prospective. It focuses on in-house engineering and system-level validation.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

The document mentions that validation was performed by "appropriately trained and knowledgeable test personnel." However, it does not specify the number of experts used to establish ground truth or their specific qualifications (e.g., radiologists, oncologists with specific years of experience).

4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

The document does not describe any adjudication method for the test set. The validation appears to be based on successful completion of predefined test criteria by test personnel.

5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

The document does not mention a multi-reader, multi-case (MRMC) comparative effectiveness study nor any AI assistance to human readers. This submission is for a software update to a linear accelerator system and related workspaces, focusing on functionality, safety, and equivalence to predicate devices, not on AI-driven diagnostic or interpretative performance by human readers.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done

The device is a linear accelerator system and its associated software workspaces for radiation therapy planning and delivery. The testing described (unit, integration, system integration, and system-level validation) assesses the performance of the software and system components. While this is "algorithm only" in the sense that it's testing the software's inherent function, it's not a standalone diagnostic algorithm being evaluated for sensitivity/specificity in a clinical context. It's the performance of the tools that support human operators in delivering radiation therapy. The document does not present metrics typical of standalone diagnostic algorithm performance (e.g., AUC, sensitivity, specificity).

7. The type of ground truth used (expert consensus, pathology, outcomes data, etc)

The ground truth for the verification and validation activities appears to be based on pre-defined system requirements, specifications, and established acceptance criteria as outlined in the "Test Concept" and "system test plans." This does not involve clinical ground truth like pathology or outcomes data, as the device is not an AI diagnostic tool but a control and planning system for radiation therapy.

8. The sample size for the training set

The document does not mention a training set as this is a software update for a medical device control and planning system, not a machine learning or AI algorithm submission that typically requires training data.

9. How the ground truth for the training set was established

As no training set is mentioned or implied, the document does not describe how ground truth for a training set was established.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right. The profiles are stacked on top of each other, with the top profile being the most prominent.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 21, 2014

Siemens Medical Solutions USA, Inc. % Mr. Ken Duplantis Regulatory Affairs 757 "A" Arnold Drive MARTINEZ CA 94553

Re: K142434

Trade/Device Name: ARTISTE Solution, Sys VC10C Phase 2 update, with the syngo RT Therapist and RT Oncologist Workspaces, v4.3.1 MR1 update Regulation Number: 21 CFR 892.5050 Regulation Name: Medical Charged-Particle Radiation Therapy System Regulatory Class: II Product Code: IYE Dated: August 23, 2014 Received: August 29, 2014

Dear Mr. Duplantis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

K142434

Device Name

ARTISTE™ Solution. SYS VC10C. Phase 2 Update with the syngo® RT Therapist & RT Oncologist Workspaces. v4.3.1 MR1

Indications for Use (Describe)

The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.

The linear accelerator systems are high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT), modulated arc therapy (mARC) and precision stereotactic radiation therapy for lesions, tumors and conditions anywhere in the head and body where radiation therapy is indicated.

The syngo® RT Therapist Workspace is a component of the linear accelerator system and is based on the syngo® architecture. The syngo® RT Therapist workspace contains software applications to support patient selection/setup. patient positioning verification. treatment delivery/verification, and treatment recording.

The syngo® RT Oncologist Workspace is an optional accessory to the linear accelerator system and permits localization, contouring, segmentation, image review, and review and approval of treatment plan parameters. In addition, it includes tools and administrative functions to aid in the diagnosis, staging, and prescription of radiation therapy.

The syngo@ RT Therapist and the syngo@ RT Oncologist Workspaces v4.3.1 MR1, can be interfaced with third party OIS, V&R, TPS, and PACS devices conforming to the DICOM Standard.

Type of Use (Select one or both, as applicable) ☐ ☐
---------------------------------------------------------------------------------------------

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

Section 5

510(k) Summary

Date Prepared:

August 20, 2014

Submitter:Siemens Medical Solutions USA, Inc.Radiation Oncology757A Arnold DriveMartinez, CA 94553
Contact:Christine DunbarSenior Manager, Regulatory Affairs
Phone:Fax:(925) 293-5442(925) 602-8008

(925) Email: christine.dunbar@siemens.com

  • ARTISTE with Sys_VC10C, Phase 2 update, including the Proprietary Name: syngo® RT Therapist Workspace, v4.3.1 MR1 Update syngo® RT Oncologist Workspace, v4.3.1_MR1 Update
  • Medical Charged-Particle Radiation Therapy System Common Name:

Classification: 892.5050

Product Code: IYE

Substantial Equivalence Claimed To (21 CFR §807.92(a)(3))

Product510(k) Clearance /DateClaim of Equivalence for:
ARTISTE™ Solution with SYS-VC10C, Phase 1 for the syngo®RT Therapist (RTT v4.3.1, CC13.0.112) supporting 3rd partyOIS, V&R, TPS, and PACS.K132935 / Feb.21, 2014ARTISTE™ Solution with SYS-VC10C, Phase 2 update for thesyngo® RT Therapist (RTTv4.3.1_MR1, & CC 13.0.301)supporting 3rd party OIS, V&R, TPS,and PACS and new CC featureAutomatic Movement Protection.Additionally, serves as a predicatefor the hosting of third party OIS onthe syngo® based workspaces, thesyngo® RT Oncologist v4.3.1_MR1.
ARTISTE™ Solution with SYS-VC10A for the syngo® RTTherapist (RTT v4.3, CC13.0.65) supporting 3rd partyOIS, V&R, TPS, and PACS.&syngo® RT Oncologist v4.3 withVSim v2.7K121295 / June12, 2012ARTISTE™ Solution with SYS-VC10C, Phase 2 update for thesyngo® RT Therapist (RTTv4.3.1_MR1 & CC 13.0.301)supporting 3rd party OIS, V&R, TPS,and PACS.& syngo® RT Oncologistv4.3.1_MR1 and the VSim v2.7module.
Product510(k) Clearance /DateClaim of Equivalence for:
ARTISTE™ Solution with SYS-VC10C, Phase 1 for the syngo®RT Therapist (RTT v4.3.1, CC13.0.112) supporting 3rd partyOIS, V&R, TPS, and PACS.K132935 /Feb. 21, 2014ARTISTE™ Solution with SYS-VC10C, Phase 2 for the syngo® RTTherapist (RTT v4.3.1x, CC13.0.301) supporting 3rd party OIS,V&R, TPS, and PACS.
Sys_VC10A update backwardscompatible and an optionalupgrade for the ONCOR andPRIMUS systems.Sys_VC10C, version RTT 4.3.1x isbackwards compatible and is anoptional upgrade for the ONCORand PRIMUS systems.
ARTISTE™ Solution with SYS-VC10A for the syngo® RTTherapist (RTT v4.3, CC13.0.65) supporting 3rd partyOIS, V&R, TPS, and PACS.Andsyngo® RT Oncologist v4.3with VSIM v2.7 with SiemensOIS.K121295 /June 12, 2012ARTISTE™ Solution with SYS-VC10C, Phase 2 for the syngo® RTTherapist (RTT v4.3.1_MR1, CC13.0.301) supporting 3rd party OIS,V&R, TPS, and PACS.Andsyngo® RT Oncologist v4.3.1xwith VSIM v2.7 and third party OIS
Sys_VC10A update isbackwards compatible and is anoptional upgrade for the ONCORand PRIMUS systems.Sys_VC10C, RTT and ONC version4.3.1_MR1 is backwards compatibleand is an optional upgrade for theONCOR and PRIMUS systems.

{4}------------------------------------------------

The update to the ARTISTE Linear Accelerator's Control Console, the syngo® RT Therapist Workspace, v4.3.1 MR1 and the syngo® RT Oncologist Workspace, v4.3.1 MR1 as described in this premarket notification has the same intended use, indications for use, and fundamental scientific technical characteristics as the predicate devices listed above.

{5}------------------------------------------------

Description Summary (21 CFR §807.92(a)(4))

ARTISTE™ Solution Linac with Control Console v13.0.301, including the syngo® RT Therapist Workspace v4.3.1 MR1, and the syngo® RT Oncologist Workspace v4.3.1 MR1 Updates

The project Sys_VC10C, Phase 2 software update for the ARTISTE™ Solution linear

accelerator system with Control Console v13.0.301 with the syngo® RT Therapist Workspace v4.3.1 MR1 release is intended to update customers with the currently cleared and released syngo® RT Therapist Workspace software with versions v4.1, v4.2, v4.3 or v4.3.1 for ARTISTE, and is backwards compatible with the ONCOR & PRIMUS Linear Accelerator systems with the COHERENCE RT Therapist v2.3. The currently cleared syngo® RT Therapist workspace software v4.3.1 supports Siemens branded and third party OIS, V&R, TPS, and PACS systems.

Additionally, the Phase 2 software update includes software modifications and defect fixes for the syngo® RT Oncologist workspace. The Oncologist software is developed on the same syngo® software code platform as the RT Therapist and the workspace differentiation is license key controlled for clinically focused workspace options. The syngo® RT Oncologist workspace is considered a medical device in its own right had has multiple previous FDA 510(k) clearances.

Note: Several syngo® clinical modules are considered medical devices in their own right and may have separate clearances. For example, the syngo® VSim module, v2.7 received FDA clearance under K101119 and is licensed on several syngo® workspaces for different modalities. The syngo® RT Oncologist Workspace has hosted this module under multiple clearances with the latest being the current predicate, K121295.

These previous software versions can be directly migrated to the current release (syngo® RTT/ONC v4.3.1 MR1) and Control Console v13.0.301. The ARTISTE Solution with VC10C. Phase 1 and the syngo® RT Therapist v4.3.1 cleared under K132935 serves as a predicate in this submission.

The syngo® RT Oncologist workspace is an optional system for Siemens Linac customers. The syngo® RT Oncologist also hosts software based products such as the VSim v2.7 feature [VSIM] and other oncology based features. The syngo® RT Oncologist workspace v4.3 was previously cleared under K121295 which serves as a predicate in this submission.

The technological characteristics and the fundamental technology of the LINAC Control Console and the syngo® RT Therapist Workspace v4.3.1 MR1 and the syngo® RT Oncologist Workspace v4.3.1 MR1 (including the syngo® VSim module) remain unchanged from the currently cleared products. This is a software update only. The materials of use remain unchanged from the predicate devices.

Risk Management and General Safety and Effectiveness

The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.

{6}------------------------------------------------

Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by software means, user instructions, verification of requirements and validation of the clinical workflow to ensure that the product meets its intended uses. To minimize electrical, mechanical and radiation hazards, SIEMENS adheres to recognized and established industry practice and relevant international standards.

Refer to Section 21 for the Risk Management documentation.

Intended Use / Indications for Use (21 CFR §807.92(a)(5))

The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer.

The linear accelerator systems are high-dose and high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT), modulated arc therapy (mARC) and precision stereotactic radiation therapy for lesions, tumors and conditions anywhere in the head and body where radiation therapy is indicated.

The syngo® RT Therapist Workspace is a component of the linear accelerator system and is based on the syngo® architecture. The syngo® RT Therapist workspace contains software applications to support patient selection/setup, patient positioning verification, treatment delivery/verification, and treatment recording.

The syngo® RT Oncologist Workspace is an optional accessory to the linear accelerator system and permits localization, contouring, segmentation, image review and review and approval of treatment plan parameters. In addition, it includes tools and administrative functions to aid in the diagnosis, staging, and prescription of radiation therapy.

The syngo® RT Therapist and the syngo® RT Oncologist Workspaces v4.3.1 MR1, can be interfaced with third party OIS, V&R, TPS, and PACS devices conforming to the DICOM Standard.

Technological Characteristics and Substantial Equivalence (21 CFR §807.92(a)(6)):

The Substantial Equivalence comparison chart (Section 12) demonstrates the comparison of the technological characteristics of the syngo® RT Therapist Workspace, the syngo® RT Oncologist Workspace, and the LINAC Control Console v13 to their currently cleared predicate versions.

The syngo® RT Therapist Workspace v4.3.1 MR1 (project Sys VC10C, Phase 2) does not change the intended use or Indications for Use of the original ARTISTE™ Solution Linac with the syngo® RT Therapist Workspace v4.3.1 (project Sys VC10C, Phase 1) or the previously cleared ONCOR™ or PRIMUS™ Siemens branded Linear Accelerator Systems.

The syngo® RT Oncologist Workspace v4.3.1 MR1 (project Sys VC10C, Phase 2) does not change the Intended Use or Indications for Use of the currently cleared syngo® RT Oncologist Workspace v4.3. (Sys VC10A, K121295).

{7}------------------------------------------------

Summary of Verification and Validation Activity (21 CFR §807.92(b)):

Bench Testing (21 CFR §807.92(b)(1))

Bench testing in the form of Unit, Integration and System Integration testing was performed to evaluate the performance and functionality of the software update v4.3.1 MR1 for the RT Therapist and RT Oncologist software and regression testing the Control Console version 13.0.301.

All testable requirements in the System Requirements Specifications (SRS), Sub-system Requirements Specifications (SSRS) and Component Requirements (CRS) for the Sys VC10C project. Phase 2, and additionally the specific requirements for the implementation of the third party OIS have been successfully verified and traced in accordance with the Siemens product development (lifecycle) process (PDP) described in Section 11.

The software verification and regression testing has been performed successfully to meet their previously determined acceptance criteria as stated in the Test Concept.

Non-Clinical Test Results (21 CFR §807.92(b)(2))

Validation of the syngo® RT Therapist Workspace v4.3.1 MR1. the syngo® RT Oncologist Workspace v4.3.1 MR1 with the hosting of the optional third party OIS and regression testing with Control Consoles 13.0.301 has been performed at the System test level on production prototype devices by appropriately trained and knowledgeable test personnel, therefore, clinical testing is not required. System level validation and regression testing has been performed successfully, demonstrating that the software meets the acceptance criteria as noted in the system test plans.

Testing to Consensus Standards (21 CFR §807.92(b)(1))

The syngo® RT Therapist Workspace v4.3.1 MR1, the syngo® RT Oncologist Workspace v4.3.1 MR1 and the Control Console 13.0.301 update have been tested (as needed) to meet the requirements for conformity (where applicable) to multiple industry standards. The R&D evaluation of the relevant testing to consensus standards is documented. Refer to Section 9 for this documentation.

Substantial Equivalence to Predicates (21 CFR §807.92(b)(1))

The verification testing to the system requirements (SRS) for the syngo® RT Therapist v4.3.1 MR1 and the RT Oncologist v4.3.1 MR1 workspaces, validation of the intended use, and the regression testing to the existing syngo® RT Therapist / Oncologist software and Control Console v13.0.301 functional requirements, is intended to support the claim of substantial equivalence to the following predicates:

{8}------------------------------------------------

  • The ARTISTE™ Solution with SYS-VC10A for the syngo®RT Therapist (RTT v4.3, . CC 13.0.65) supporting 3d party OIS & V&R (MOSAIQ and Sequencer V&R ') and third party TPS and PACs systems. Additionally, the syngo®RT Oncologist workspace (ONC v4.3 with VSIM v2.7), (K121295). Primary predicate for the syngo® RT Oncologist workspace v4.3.1 MR1 updates including VSim v2.7 bug fixes, and support for the third party OIS connectivity.
  • The ARTISTE™ Solution with SYS-VC10C for the synqo®RT Therapist (RTT v4.3.1, ● CC 13.0.112) supporting 30 party OIS & V&R (ARIA and Varian Treat V&R²) and third party TPS and PACs systems. (K132935). Primary predicate for the syngo® RT Therapist workspace v4.3.1 MR1 with Control Console v13.0.301 / FC updates and support for the third party OIS connectivity.

1 MOSAIQ OIS and Sequencer V&R systems are manufactured by IMPAC Medical / Elekta AB. 2 ARIA OIS and Varian Treatment console are manufactured by Varian Medical Systems.

{9}------------------------------------------------

510(k) Summary (21 CFR §807.92(c))

In summary, based on the successful verification and validation testing to the software acceptance criteria, it is SIEMENS' opinion that the syngo® RT Therapist Workspace update to v4.3.1 MR1 including the update to the LINAC Control Console v13.0.301 to support the new Automatic Movement Protection feature and the syngo® RT Oncologist Workspace update to v4.3.1_MR1, including hosting of the optional 300 party OIS and bug fixes, does not introduce any new potential safety risks and are substantially equivalent to, and performs as well as, the predicate devices.

N/A