K Number
K142361

Validate with FDA (Live)

Date Cleared
2014-12-05

(102 days)

Product Code
Regulation Number
892.1000
Age Range
All
Reference & Predicate Devices
Predicate For
N/A
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Discovery MR750 3.0T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMI, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.

The images produced by the Discovery MR750 3.0Tsystem reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Device Description

The Discovery MR750 3.0T features a superconducting magnet operating at 3.0 Tesla. The data acquisition system accommodates up to 32 independent receive channels in various increments, and multiple independent coil elements per channel during a single acquisition series. The system uses a combination of time-varying magnetic fields (gradients) and RF transmissions to obtain information regarding the density and position of elements exhibiting magnetic resonance. The system can image in the sagittal, coronal, axial, oblique and double oblique planes, using various pulse sequences and reconstruction algorithms. The Discovery MR750 3.0T uses multi-drive RF transmit for imaging of the head and body regions.

AI/ML Overview

Here's a breakdown of the acceptance criteria and study information for the GE Discovery MR750 3.0T, based on the provided document:

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Stated Goal)Reported Device Performance
Maintains same imaging performance results as predicate device (K132376)"The clinical results demonstrated that the Discovery MR750 3.0T maintains the same imaging performance results as its predicate device (K132376)."
Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33The device "complies with the following voluntary standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-33"
Compliance with NEMA DICOM standards (NEMA PS3.1-3.20)The device is "in compliance with the applicable NEMA standards, including NEMA PS3.1-3.20 for DICOM conformance."
Successful completion of verification testing with defined pass/fail criteria"The testing was completed with passing results per the pass/fail criteria defined in the test cases."

2. Sample Size Used for the Test Set and the Data Provenance

The document states: "Internal scans were conducted as part of validation for workflow and image quality for the addition of the new feature."

  • Sample Size: Not explicitly stated. The term "internal scans" typically implies a smaller, controlled set of images used for engineering validation rather than a large clinical trial.
  • Data Provenance: "Internal scans" suggests the data was generated internally by GE Healthcare, likely on their own test subjects or phantoms. There is no mention of country of origin or whether it was retrospective or prospective in a clinical sense; it appears to be primarily technical validation.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and the Qualifications of Those Experts

This information is not provided in the document. The study described focuses on technical and imaging performance comparison to a predicate, rather than an assessment requiring expert ground truth in a diagnostic context.

4. Adjudication Method for the Test Set

This information is not provided in the document. Given the nature of the "internal scans" for workflow and image quality validation, a formal adjudication process as seen in clinical trials with multiple expert readers is unlikely to have been employed.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, If So, What Was the Effect Size of How Much Human Readers Improve with AI vs Without AI Assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. The study described is a technical and imaging performance comparison, not an AI-assisted diagnostic study. Therefore, there is no information on human reader improvement with or without AI assistance.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

The device itself is a Magnetic Resonance Diagnostic Device, not an AI algorithm. The tests performed were on the imaging system's performance, not a standalone algorithm.

7. The Type of Ground Truth Used (Expert Consensus, Pathology, Outcomes Data, etc.)

The document implies that the ground truth for performance was the performance of the predicate device (K132376). The goal was to demonstrate that the new device "maintains the same imaging performance results as its predicate device." This suggests a comparative "ground truth" rather than an independent gold standard like pathology. For technical performance, compliance with standards (IEC, NEMA) would also serve as a form of "ground truth."

8. The Sample Size for the Training Set

There is no mention of a training set. This type of device (a magnetic resonance scanner) does not typically involve a machine learning training set in the way an AI diagnostic algorithm would. The development involves engineering and software updates, and validation against established technical standards and predicate performance.

9. How the Ground Truth for the Training Set Was Established

As there is no mention of a training set, this information is not applicable/not provided.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a stylized emblem. The emblem consists of three abstract human profiles facing right, layered on top of each other. The profiles are rendered in black, creating a sense of depth and unity.

December 5, 2014

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Regulatory Affairs Leader, Magnetic Resonance

Re: K142361

GE Medical Systems, LLC

3200 N. Grandview Blvd. WAUKESHA WI 53188

% Ms. Jenny Wong

Trade/Device Name: Discovery MR750 3.0T Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: November 21, 2014 Received: November 24, 2014

Dear Ms. Wong:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D. O'Hara
for

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known)

K142361

Device Name Discovery MR750 3.0T

Indications for Use (Describe)

The Discovery MR750 3.0T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-to-noise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMI, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.

The images produced by the Discovery MR750 3.0Tsystem reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font inside. The letters are also blue, and they are surrounded by a swirling pattern. The logo is simple and recognizable, and it is often used to represent the company.

GE Healthcare 510(k) Premarket Notification Submission

510(k) Summary

In accordance with 21 CFR 807.92 the following summary of information is provided: Date: August 22, 2014

Submitter:GE Healthcare, (GE Medical Systems, LLC)3200 N Grandview Blvd.Waukesha, WI 53188
Primary Contact Person:Jenny WongRegulatory Affairs Leader, MRGE Healthcare, (GE Medical Systems, LLC)Phone: 262-548-2899Fax: 414-908-9585
Secondary Contact Person:Glen SabinRegulatory Affairs Director, MRGE Healthcare, (GE Medical Systems, LLC)Phone: 262-521-6848Fax: 414-908-9585
Device:Trade Name:Discovery MR750 3.0T
Common/Usual Name:Magnetic Resonance Diagnostic Device
Classification Names:892.1000
Product Code:LNH
Predicate Device(s):Discovery MR750 3.0T and Discovery MR750w 3.0T [K132376]
Device Description:The Discovery MR750 3.0T features a superconducting magnetoperating at 3.0 Tesla. The data acquisition system accommodates up to32 independent receive channels in various increments, and multipleindependent coil elements per channel during a single acquisitionseries. The system uses a combination of time-varying magnetic fields(gradients) and RF transmissions to obtain information regarding thedensity and position of elements exhibiting magnetic resonance. Thesystem can image in the sagittal, coronal, axial, oblique and doubleoblique planes, using various pulse sequences and reconstructionalgorithms. The Discovery MR750 3.0T uses multi-drive RF transmitfor imaging of the head and body regions.

The Discovery MR750 3.0T is designed to conform to NEMA DICOM

up to

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font inside. The circle has a decorative border with swirling lines. The logo is simple and recognizable, representing the company's brand.

GE Healthcare 510(k) Premarket Notification Submission

standards.

Intended Use: The Discovery MR750 3.0T system is a whole body magnetic resonance scanner designed to support high resolution, high signal-tonoise ratio, and short scan times. It is indicated for use as a diagnostic imaging device to produce axial, sagittal, coronal, and oblique images, spectroscopic images, parametric maps, and/or spectra, dynamic images of the structures and/or functions of the entire body, including, but not limited to, head, neck, TMJ, spine, breast, heart, abdomen, pelvis, joints, prostate, blood vessels, and musculoskeletal regions of the body. Depending on the region of interest being imaged, contrast agents may be used.

The images produced by the Discovery MR750 3.0T system reflect the spatial distribution or molecular environment of nuclei exhibiting magnetic resonance. These images and/or spectra when interpreted by a trained physician yield information that may assist in diagnosis.

  • Technology: The Discovery MR750 3.0T employs the same fundamental scientific technology as its predicate devices.
    The following is a summary of the differences between the proposed Discovery MR750 3.0T and its predicate Discovery MR750 3.0T and Discovery MR750w 3.0T (K132376).

  • New base software (version DV25) ●

  • . Software and hardware modifications to enable of the multidrive RF transmit technology for body and head

  • Hardware modifications to the host computer and reconstruction engine for obsolescence and compatibility with a new operating system

  • . Revised and updated labeling to support new feature

Determination of SubstantialEquivalence:Summary of Non-Clinical Tests:
The Discovery MR750 3.0T system with the addition of the new multi-drive RF transmit function complies with the following voluntary standards:IEC 60601-1 IEC 60601-1-2
  • IEC 60601-2-33

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for General Electric (GE). The logo is a blue circle with the letters "GE" in a stylized font in the center. The letters are white, and there are swirling designs around the letters. The logo is simple and recognizable, and it is associated with a well-known company.

GE Healthcare 510(k) Premarket Notification Submission

In addition, the Discovery MR750 3.0T system in compliance with the applicable NEMA standards, including NEMA PS3.1-3.20 for DICOM conformance.

The following quality assurance measures were applied to the development of the system:

  • . Risk Analysis
  • Requirements Reviews ●
  • Design Reviews ●
  • . Testing on unit level (Module verification)
  • . Integration testing (System verification)
  • Performance testing (Verification)
  • Safety testing (Verification)
  • . Simulated use testing (Validation)

The non-clinical tests have been summarized in the verification testing that was completed for the addition of the multi-drive RF transmit mode. The testing was completed with passing results per the pass/fail criteria defined in the test cases. This supports substantial equivalence to its predicates because it was also developed under quality assurance Design Controls. In addition, Discovery MR750 3.0T is in compliance to the same Standards.

Summary of Clinical Tests:

The subject of this premarket submission, Discovery MR750 3.0T, did not require external clinical studies to support substantial equivalence. Internal scans were conducted as part of validation for workflow and image quality for the addition of the new feature. The clinical results demonstrated that the Discovery MR750 3.0T maintains the same imaging performance results as its predicate device (K132376). Sample clinical images are included in this submission.

  • Conclusion: GE Healthcare considers the Discovery MR750 3.0T to be as safe, as effective, and performance is substantially equivalent to the predicate device(s).

§ 892.1000 Magnetic resonance diagnostic device.

(a)
Identification. A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).(b)
Classification. Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.