K Number
K142344

Validate with FDA (Live)

Date Cleared
2015-03-20

(210 days)

Product Code
Regulation Number
882.4560
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Disposable Passive Accessory is intended to be used for anatomy palpation in image guided surgeries with Medtronic StealthStation® System. The device is sterile and designed for single use.

Device Description

The IZI Disposable Passive Accessory contains one disposable point probe mounted with five passive reflective sphere markers, as well as ten additional passive reflective sphere markers. The Disposable Passive Accessory is intended to be used for anatomy palpation and registration in image guided surgeries with the Medtronic StealthStation® system. The device is sterile and designed for single use.

The Disposable Passive Accessory provides a single-use alternative to Medtronic reusable parts. The Disposable Passive Accessory is constructed by a dimension stable thermoplastic handle where five passive reflective sphere markers are mounted on, with a stainless steel tip that is used to register and locate patient anatomical structure in an image guided surgery using the Medtronic StealthStation® System. The permanently mounted passive reflective spheres are at the same three dimensional locations as the Medtronic predicate product. In an IGS procedure, the two cameras of the StealthStation® captures the reflective light from the five sphere markers for the System to calculate and determine the location of the stainless tip and the position of the probe.

AI/ML Overview

The provided document, K142344, describes the substantial equivalence of the IZI Disposable Passive Accessory to predicate devices for use in image-guided surgeries with the Medtronic StealthStation System. The acceptance criteria and the study proving the device meets these criteria are detailed below.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (What the device claims to do, or what needs to be met)Reported Device Performance (How the device performed in testing)
Compatibility with Medtronic StealthStation® SystemThe "IZI Disposable Accessory with the Medtronic StealthStation S7" study was conducted to ensure compatibility. It verified product performance in the StealthStation® patient registration process and measured navigation accuracy.
Accuracy and Precision (Dimensional Analysis)A "Design Verification - Dimensional Analysis" study was performed. All critical dimensions affecting passive reflective marker locations and probe tip location were measured against the predicate device. The CpK (Process Capability Index) of all dimensions were reported to be greater than 1.0, indicating the process is capable of producing items within specifications.
Comparable Performance to Predicate Device (Patient Registration & Navigation Accuracy)The "IZI Disposable Accessory with the Medtronic StealthStation S7" study concluded that the IZI Disposable Passive Accessory performs comparably to the Medtronic Planar Probe with manually mounted Spherz regarding patient registration and navigation error.
BiocompatibilityA cytotoxicity test was performed. The test confirmed that "the product does not cause any cell lysis or toxicity." The materials and components were known to be biocompatible or have a history of use in similar medical devices. The manufacturing process does not use cleaning agents or chemicals.
SterilityThe device is provided as sterile. (No specific test results from this document, but implied by classification and indication).
Single UseThe device is designed for single use. (This is a design characteristic, not a performance metric to be tested against, but an acceptance criterion for its intended use model).
Physical Dimensions and 3D Locations of Marker Post Relative to Probe TipIn the comparison table, it is stated to be "Same as the Medtronic Passive Planar Probe" and the comparison of technology is "Identical". The dimensional analysis study confirmed that "All critical dimensions that affect the passive reflective marker locations and the probe tip location were measured in the Study against the predicate device. The CpK of all dimensions are greater than 1.0." indicating that this specific criterion was met.
Probe Tip MaterialThe subject device uses Stainless steel (303 stainless steel), which is comparable to the predicate's Stainless steel (630 stainless steel). Both are stainless steel.
Passive Reflective MarkerThe subject device's markers are "Identical to Spherz, permanently mounted onto the probe post". This is deemed identical to the predicate's markers, which are provided separately for manual mounting.

2. Sample Size Used for the Test Set and Data Provenance

The document explicitly states that two non-clinical bench tests were conducted. However, it does not provide specific details on the sample sizes (e.g., number of devices tested, number of measurements taken) for these tests.

  • Design Verification - Dimensional Analysis: No specific sample size is mentioned, only that "All critical dimensions...were measured in the Study."
  • Design Verification - IZI Disposable Accessory with the Medtronic StealthStation S7: No specific sample size is mentioned for the number of accessories or the number of tests performed on the dummy head. The study used "a pre-scanned Medtronic IGS dummy head."

The data provenance is retrospective, as these are bench tests conducted by the manufacturer for regulatory submission, rather than prospective clinical trials. The country of origin for the data is implicitly the United States, given the submission to the FDA by a US-based company (IZI Medical Products LLC, Maryland).

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications of Those Experts

No explicit information is provided regarding the use of external experts to establish ground truth for the bench tests. The ground truth for the dimensional analysis would likely be engineering specifications and measurements. For the StealthStation compatibility test, the ground truth for navigation accuracy would be established by the known true positions within the pre-scanned dummy head, and the operational functionality would be against the expected behavior of the Medtronic StealthStation System. These are internal engineering and performance benchmarks rather than expert consensus on medical images or diagnoses.

4. Adjudication Method for the Test Set

No adjudication method is mentioned, as the studies are non-clinical bench tests. The performance metrics (CpK, navigation error) are quantifiable and objective, not requiring human interpretation or adjudication in the medical sense.

5. If a Multi Reader Multi Case (MRMC) Comparative Effectiveness Study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

A Multi-Reader Multi-Case (MRMC) comparative effectiveness study was not done. This device is a passive accessory (a probe) for an image-guided surgery system, not an AI-based diagnostic tool that assists human readers in interpreting medical images. Therefore, the concept of "human readers improving with AI assistance" does not apply to this device.

6. If a Standalone (i.e. algorithm only without human-in-the-loop performance) was done

This device is a hardware accessory (a probe), not an algorithm or AI system. Therefore, the concept of "standalone (algorithm only without human-in-the-loop performance)" does not apply. The device's function is inherently human-in-the-loop as it is used by a surgeon for palpation and registration.

7. The Type of Ground Truth Used (expert consensus, pathology, outcomes data, etc.)

The ground truth used for these non-clinical bench tests was based on:

  • Engineering Specifications and Measurements: For the "Design Verification - Dimensional Analysis," the ground truth was the engineering specifications of the predicate device and the physical measurements taken.
  • Known Physical Coordinates/System Functionality: For the "Design Verification - IZI Disposable Accessory with the Medtronic StealthStation S7," the ground truth for navigation accuracy was derived from the known positions within the pre-scanned Medtronic IGS dummy head and the expected operational parameters of the StealthStation® System itself.

8. The Sample Size for the Training Set

This device is not an AI/machine learning model, so there is no training set.

9. How the Ground Truth for the Training Set Was Established

As there is no training set for an AI/machine learning model, this question is not applicable.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, stacked on top of each other. The profiles are connected by a flowing, ribbon-like shape at the bottom. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the image.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 20, 2015

IZI Medical Products LLC % Mr. Qiang Cao Ouality Assurance and Regulatory Affairs Manager 5 Easter Court, Suite J Owings Mills, Maryland 21117

Re: K142344

Trade/Device Name: Disposable Passive Accessory for Medtronic StealthStation System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: HAW Dated: February 16, 2015 Received: February 18, 2015

Dear Mr. Cao,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Carlos L. Pena -S/Δ

Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.

510(k) Number (if known) K142344

Device Name

Disposable Passive Accessory for Medtronic StealthStation® System

Indications for Use (Describe)

The Disposable Passive Accessory is intended to be used for anatomy palpation in image guided surgeries with Medtronic StealthStation® System. The device is sterile and designed for single use.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

] Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{4}------------------------------------------------

510(k) Summary

I. SUBMITTER

IZI Medical Products LLC 5 Easter Court, Suite J Owings Mills, MD 21117

Phone: (410) 594-9403 Fax: (410) 594-0540 Contact Person: Qiang Cao Date Prepared: March 19, 2015

II. DEVICE

Name of Device: Disposable Passive Accessory for Medtronic StealthStation® System Common or Usual Name: Disposable Passive Accessory Classification Name: Neurological Stereotaxic Instrument (21 CFR 882.4560) Regulatory Class: II Product Code: HAW (Neurological Stereotaxic Instrument)

III. PREDICATE DEVICE

StealthStation® System Update (K050438) Spherz® (Passive Reflective Marker) (K022074)

IV. DEVICE DESCRIPTION

The IZI Disposable Passive Accessory contains one disposable point probe mounted with five passive reflective sphere markers, as well as ten additional passive reflective sphere markers. The Disposable Passive Accessory is intended to be used for anatomy palpation and registration in image guided surgeries with the Medtronic StealthStation® system. The device is sterile and designed for single use.

The Disposable Passive Accessory provides a single-use alternative to Medtronic reusable parts. The Disposable Passive Accessory is constructed by a dimension stable thermoplastic handle where five passive reflective sphere markers are mounted on, with a stainless steel tip that is used to register and locate patient anatomical structure in an image guided surgery using the Medtronic StealthStation® System. The permanently mounted passive reflective spheres are at the same three dimensional locations as the Medtronic predicate product. In an IGS procedure, the two cameras of the StealthStation® captures the reflective light from the five sphere markers

{5}------------------------------------------------

for the System to calculate and determine the location of the stainless tip and the position of the probe.

V. INDICATIONS FOR USE

The Disposable Passive Accessory is intended to be used for anatomy palpation and registration in image guided surgeries with Medtronic StealthStation system. The device is sterile and designed for single use.

The indication for use statement for the Disposable Passive Accessory is very similar to the labeled indication for use of the Medtronic Passive Planar Probe, however, the predicate device is an accessory part of the Medtronic StealthStation System. The Medtronic 510(k) K050438 indication for use emphasizes the System and applicable procedures, and the specific indication of use for each accessory part is not clearly stated in the 510(k). The minor difference of the subject device and the predicate device is that the subject device is single use and provided to the user as a sterile device. This does not affect the intended surgical use of the device. The IZI Spherz are a component part of the subject device. The subject device uses the Spherz within the scope of indication for use of the Spherz.

Patient Population

The device is intended for use in patients undergoing stereotactic surgery.

Environment of Use

The device is intended for use by a trained healthcare professional in a healthcare facility.

VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE

The IZI Disposable Passive Accessory contains a disposable passive probe where five passive reflective markers are mounted, as well as ten additional passive reflective markers. The five passive reflective markers mounted on the probe to be recognized and registered in the StealthStation system. The predicate devices are the Medtronic StealthStation System and the Medtronic Passive Planar Probe (Medtronic part number 960-556).

Technical CharacteristicsIZI Disposable PassiveAccessory (subjectdevice)MedtronicStealthStation® andPassive Planar Probe(K050438)Comparison ofTechnology
Physical dimensions and3D locations of markerpost relative to probe tipSame as the MedtronicPassive Planar ProbeMedtronic PassivePlanar ProbeIdentical

{6}------------------------------------------------

Probe tip materialStainless steel (303stainless steel)Stainless steel (630stainless steel)Both materials arestainless steel
Passive Reflective MarkerIdentical to Spherz,permanently mountedonto the probe postMedtronic predicatedoes not include thepassive reflectivemarkers. The samepassive reflectivemarkers are providedseparately andmounted manually byuser.Identical
Single Use or ReusableSingle useReusableSubject Device isfor single use
SterileSterileNot sterileSubject Device isprovided as sterile

The Disposable Passive Accessory is comparable to the Medtronic StealthStation and reusable accessory (K050438) regarding its intended use, functional and dimensional characteristics, and overall performance. Minor differences between the disposable devices and reusable device have no impact on safety or effectiveness.

Biocompatibility Profile - The device is an externally communicating, tissue contacting device with limited patient contact. The patient contact time is less than 24 hours. The product is made from materials and components were known to be biocompatible or have a history of being used in similar medical devices. The product is made in a controlled manufacture process that no cleaning agent or chemicals are used. To ensure the safety of the product, we have performed a cytotoxicity test. The test confirmed that the product does not cause any cell lysis or toxicity.

VII. PERFORMANCE DATA

Two non-clinical bench tests were conducted to demonstrate that the IZI Disposable Passive Accessory performs as intended and is comparable to the predicate devices.

  1. Design Verification - Dimensional Analysis of the IZI Disposable Passive Accessory Probe. The System calculates and determines the position and angle of the probe based on the three dimensional location of the five passive reflective sphere markers that are mounted on the probe. Dimension accuracy and precision is an important technical characteristic that ensures the subject device is compatible with the StealthStation® System and that the subject device is comparable to the predicate device. A dimensional analysis of the Disposable Passive Accessory and the predicate device was performed to ensure the disposable part meets engineering specifications and the subject device is as precise and accurate as the predicate device. All critical dimensions that affect the passive reflective marker locations and the probe tip location

{7}------------------------------------------------

were measured in the Study against the predicate device. The CpK of all dimensions are greater than 1.0.

  1. Design Verification - IZI Disposable Accessory with the Medtronic StealthStation S7 The Study was conducted to ensure the compatibility of the devices with the Medtronic StealthStation® system and confirming it is suitable for its intended use. The Study verified product performance in the StealthStation® patient registration process and measured navigation accuracy using a pre-scanned Medtronic IGS dummy head. Regarding patient registration and navigation error, the IZI Disposable Passive Accessory performs comparably to the Medtronic Planar Probe with manually mounted Spherz.

The two studies verify that the subject device is as safe and effective as the predicate device, and that the subject device is comparable to the predicate device.

VIII. CONCLUSION

The non-clinical data support the safety of the device and the hardware verification and validation demonstrate that the IZI Disposable Passive Accessory device should perform as intended in the specified use conditions. The non-clinical data demonstrate that the IZI Disposable Passive Accessory device performs comparably to the predicate device that is currently marketed for the same intended use.

§ 882.4560 Stereotaxic instrument.

(a)
Identification. A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.(b)
Classification. Class II (performance standards).