K Number
K142280

Validate with FDA (Live)

Date Cleared
2014-10-16

(62 days)

Product Code
Regulation Number
862.3650
Age Range
1 - 17
Reference & Predicate Devices
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

Healgen Oxazepam Test is an immunochromatographic assay for the qualitative determination of Oxazepam (a drug in the benzodiazepine class) in human urine at a Cut-Off concentration of 300 ng/m L. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test may yield preliminary positive results even when prescription drug Oxazepam is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for oxazepam in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Healgen Morphine Test is an immunochromatographic assay for the qualitative determination of morphine (a drug in the opiate class) in human urine at a Cut-Off concentration of 2000 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the presult is positive. For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Device Description

Healgen Oxazepam Test and Healgen Morphine Test are immunochromatographic assays for Oxazepam and Morphine. Each assay test is a lateral flow system for the qualitative detection of Oxazepam and Morphine (target analyte) in human urine. The products are in vitro diagnostic devices, which come in the form of: Strips, Cassettes, DipCards, or Cups. Each product contains a Test Device (in one of the four formats), and a package insert. Each test device is sealed with a desiccant in an aluminum pouch.

AI/ML Overview

This document describes the performance characteristics and acceptance criteria for the "Healgen Oxazepam Test" and "Healgen Morphine Test," which are immunochromatographic assays for the qualitative determination of Oxazepam and Morphine in human urine.

Here's a breakdown of the requested information:

1. Table of Acceptance Criteria & Reported Device Performance:

The document doesn't explicitly state "acceptance criteria" as pass/fail thresholds in a dedicated table format. Instead, it presents performance data that implicitly serves as the proof of meeting intended performance. The key performance metrics evaluated are precision and method comparison (clinical sample testing) near the cut-off concentrations, and lay-user study accuracy.

I've extracted the relevant performance values. For "acceptance criteria," I'll interpret this as the implied high accuracy expected from such diagnostic tests.

Performance MetricSpecific Test and ConcentrationImplied Acceptance Criteria (e.g., high agreement with GC/MS, low false rates)Reported Device Performance (Oxazepam)Reported Device Performance (Morphine)
PrecisionSamples at -100% to -25% of cut-off100% agreement with expected negative result50-/0+ (50 negative, 0 positive) across all concentrations (-100%, -75%, -50%, -25%) and all formats (Strip, Cassette, Dip Card, CUP) for all 3 lots tested.50-/0+ across all concentrations (-100%, -75%, -50%, -25%) and all formats (Strip, Cassette, Dip Card, CUP) for all 3 lots tested.
PrecisionSamples at +25% to +100% of cut-off100% agreement with expected positive result50+/0- (50 positive, 0 negative) across all concentrations (+25%, +50%, +75%, +100%) and all formats (Strip, Cassette, Dip Card, CUP) for all 3 lots tested.50+/0- across all concentrations (+25%, +50%, +75%, +100%) and all formats (Strip, Cassette, Dip Card, CUP) for all 3 lots tested.
PrecisionSamples at Cut-off concentrationHigh percentage of correct classifications (mix of positive/negative expected)Oxazepam Strip: 22-/28+ (22 negative, 28 positive); Cassette: 24-/26+; Dip Card: 18-/32+; CUP: 20-/30+ (across 3 lots each). This indicates expected variability near the cutoff.Morphine Strip: 18-/32+; Cassette: 22-/28+; Dip Card: 22-/28+; CUP: 20-/30+ (across 3 lots each). This indicates expected variability near the cutoff.
Method Comparison (Clinical Samples)Agreement with GC/MS for negative samples (Negative, Low Negative, Near Cutoff Negative)Very high agreement with GC/MSOxazepam (all formats, all 3 viewers): All samples in these categories (10 + 15 + 15 = 40 samples per viewer) were correctly identified as negative (e.g., "10", "15", "15" in negative rows).Morphine (all formats, all 3 viewers): All samples in these categories (10 + 16 + 14 = 40 samples per viewer) were correctly identified as negative (e.g., "10", "16", "14" in negative rows).
Method Comparison (Clinical Samples)Agreement with GC/MS for positive samples (Near Cutoff Positive, High Positive)Very high agreement with GC/MSOxazepam (all formats, all 3 viewers): Generally high agreement, with some discordant results near cutoff. For example, Strip format: 14/16 positive in "Near Cutoff Positive", 24/24 positive in "High Positive." See detailed discordant rows for exceptions.Morphine (all formats, all 3 viewers): Generally high agreement, with some discordant results near cutoff. For example, Strip format: 13/17 positive in "Near Cutoff Positive", 23/23 positive in "High Positive." See detailed discordant rows for exceptions.
Lay-user Study (Accuracy)Overall correct results for each concentration level tested (-100% Cutoff to +75% Cutoff)≥ 90% correct for all concentrationsGenerally 90-100% correct results across all concentration levels and all formats. For example, -25% Cutoff: 90-95% correct; +25% Cutoff: 90-95% correct. Others 100%.Generally 90-100% correct results across all concentration levels and all formats. For example, -25% Cutoff: 90% correct; +25% Cutoff: 90-95% correct. Others 100%.
Lay-user Study (Instructions)Ease of understanding instructionsEasily understood by lay usersAll lay users indicated that the device instructions can be easily followed. Flesch-Kincaid Grade Level: 7.All lay users indicated that the device instructions can be easily followed. Flesch-Kincaid Grade Level: 7.

2. Sample sizes used for the test set and the data provenance:

  • Precision Studies (Analytical Performance):

    • Test Set Size: For each drug (Oxazepam and Morphine), for each format (Strip, Cassette, Dip Card, CUP), and for each of the 3 lots, 50 samples were tested at each of 9 concentrations (-100%, -75%, -50%, -25%, Cut-off, +25%, +50%, +75%, +100% of cut-off).
      • Total precision samples per drug: (9 concentrations * 50 samples/conc * 4 formats * 3 lots) = 5400 samples
      • Total precision samples (Oxazepam + Morphine): 10,800 samples.
    • Data Provenance: The document states "samples were prepared by spiking drug in negative samples." "These samples were tested using three batches of each device." The provenance (country of origin), whether retrospective or prospective, is not explicitly stated for these analytical samples, but they appear to be experimentally prepared rather than naturally occurring clinical samples.
  • Method Comparison Studies (Clinical Samples):

    • Test Set Size: For each drug (Oxazepam and Morphine), and for each format (Strip, Cassette, Dip Card, CUP), 80 "unaltered clinical samples" (40 negative and 40 positive) were tested.
      • Total method comparison samples per drug: (80 samples * 4 formats) = 320 samples.
      • Total method comparison samples (Oxazepam + Morphine): 640 samples.
    • Data Provenance: "unaltered clinical samples." The geographic origin of these clinical samples is not specified. They appear to be retrospective as they were collected and then blind labeled for testing.
  • Lay-user Study:

    • Test Set Size: 140 lay persons for Oxazepam devices, and 140 lay persons for Morphine devices. Each participant was given 1 blind-labeled sample.
      • For Oxazepam and Morphine, each lay person tested one sample at one of 7 concentrations (-100%, -75%, -50%, -25%, +25%, +50%, +75%, of cut-off).
      • For each concentration level, 20 samples were tested. So, 20 * 7 = 140 samples were tested per drug.
    • Data Provenance: The study was "performed at three intended user sites." The geographic origin is not specified. The samples were "prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens." This indicates a prospective experimental design using prepared urine samples, not naturally occurring clinical samples.

3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts:

  • For Precision Studies & Lay-user Studies: The document states that drug concentrations in the prepared samples were "confirmed by GC/MS." GC/MS (Gas Chromatography/Mass Spectrometry) is a highly reliable analytical chemistry method, and its results are considered the gold standard for confirming drug concentrations. The "experts" here are the calibrated GC/MS instrument and the analytical chemists operating and interpreting it. Specific expert qualifications are not detailed, but it's implied they adhere to standard laboratory practices for GC/MS.

  • For Method Comparison Studies (Clinical Samples): The ground truth was established by "GC/MS results." Similar to above, GC/MS is the reference method, implying that the ground truth was derived from this analytical technique rather than human expert interpretation of the test device itself.

4. Adjudication method for the test set:

  • Precision Studies & Lay-user Studies: No explicit adjudication method is mentioned. The ground truth (GC/MS confirmation) is considered definitive for these prepared samples.
  • Method Comparison Studies: The comparison was directly against "GC/MS results," which serves as the independent truth. The results from each of the three "laboratory assistants" (viewers) were compared individually to the GC/MS result. There was no listed adjudication process among the viewers; their individual agreement/disagreement with GC/MS is reported.

5. If a multi-reader multi-case (MRMC) comparative effectiveness study was done, if so, what was the effect size of how much human readers improve with AI vs without AI assistance:

  • No, an MRMC comparative effectiveness study was not done. This document describes a traditional in-vitro diagnostic device (lateral flow immunoassay for drugs of abuse) and its validation. It is not an AI-assisted diagnostic device, nor does it involve human readers interpreting images or data with and without AI assistance. The "viewers" in the method comparison study are "laboratory assistants" performing the test and visually reading the results, not interpreting complex medical images.

6. If a standalone (i.e. algorithm only without human-in-the loop performance) was done:

  • Not applicable in the context of AI/algorithms. This device is a rapid, visual immunochromatographic assay. Its "performance" is inherently linked to human observation of the test lines. There is no underlying algorithm separate from the chemical reaction displayed visually. The "standalone" performance is essentially the listed "Precision" and "Method Comparison" results, where the device itself (the chemical immunoassay) is assessed for its ability to correctly manifest positive/negative results under various conditions, which are then read by a human.

7. The type of ground truth used:

  • GC/MS (Gas Chromatography/Mass Spectrometry) results were used as the gold standard ground truth for both the precision studies (prepared samples) and the method comparison studies (clinical samples). This is a highly accurate and quantitative analytical method for drug concentration determination.

8. The sample size for the training set:

  • Not applicable. This document describes a traditional immunochromatographic assay, not a machine learning or artificial intelligence model. Therefore, there is no "training set" in the computational sense. The device's "training" is inherent in its chemical design and manufacturing process, optimized through R&D and quality control, not through data-driven learning.

9. How the ground truth for the training set was established:

  • Not applicable. As established in point 8, there is no "training set" for an AI model. For the development and verification of such an immunoassay, the "ground truth" would be established through established chemical and biological assays, often using purified standards and reference methods like GC/MS (as seen in the performance validation), during the manufacturing and R&D phases to ensure the reagents and test strip components function as intended. However, this is part of the product development and quality control, not a "training set" for an algorithm.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of three human profiles facing to the right, arranged in a cascading manner. The profiles are depicted in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the profile images.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 16, 2014

HEALGEN SCIENTIFIC LLC C/O JOE SHIA LSI INTERNATIONAL INC. 504 EAST DIAMOND AVE., SUITE F GAITHERSBURG MD 20877

Re: K142280

Trade/Device Name: Healgen Oxazepam Test (Strip, Cassette, Cup, Dio Card): Healgen Morphine Test (Strip, Cassette, Cup, Dip Card) Regulation Number: 21 CFR 862.3170 Regulation Name: Benzodiazepine test system Regulatory Class: II Product Code: JXM, DJG Dated: August 12, 2014 Received: August 15, 2014

Dear Mr. Joe Shia:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulations (21 CFR Parts 801 and 809), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Courtney

Courtney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

{2}------------------------------------------------

Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known) K142280

Device Name

Healgen Oxazepam Test (Strip, Cassette, Cup, Dip Card) Healgen Morphine Test (Strip, Cassette, Cup, Dip Card)

Indications for Use (Describe)

Healgen Oxazepam Test is an immunochromatographic assay for the qualitative determination of Oxazepam (a drug in the benzodiazepine class) in human urine at a Cut-Off concentration of 300 ng/m L. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test may yield preliminary positive results even when prescription drug Oxazepam is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for oxazepam in urine. The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Healgen Morphine Test is an immunochromatographic assay for the qualitative determination of morphine (a drug in the opiate class) in human urine at a Cut-Off concentration of 2000 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the presult is positive. For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

510(k) SUMMARY

1. Date:October 6, 2014
2. Submitter:HEALGEN SCIENTIFIC LLC5213 Maple StBellaire, TX77401
3. Contact person:Jianqiu FangHEALGEN SCIENTIFIC LLC5213 Maple StBellaire, TX77401Telephone: 713-733-8088

Telephone: 713-733-8088 Fax: 713-733-8088 Email: bryan@healgen.com

    1. Device Name: Healgen Oxazepam Test (Strip, Cassette. Cup, Dip Card) Healgen Morphine Test (Strip, Cassette, Cup, Dip Card)

Classification:

ProductCodeCFR #Panel
JXM21 CFR, 862.3170 Benzodiazepine Test SystemToxicology
DJG21 CFR, 862.3650 Opiate Test SystemToxicology
    1. Predicate Devices:
      K052115 First Check Multi Drug Cup 12
    1. Intended Use
      Healgen Oxazepam Test is an immunochromatographic assay for the qualitative determination of Oxazepam (a drug in the benzodiazepine class) in human urine at a Cut-Off concentration of 300 ng/m L. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test may yield preliminary positive results even when prescription drug Oxazepam is ingested, at prescribed doses; it is not intended to distinguish between prescription use or abuse of this drug. There is no uniformly recognized cutoff concentration level for oxazepam in urine. The test provides only preliminary test results. A more specific alternative chemical must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and

{4}------------------------------------------------

professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

Healgen Morphine Test is an immunochromatographic assay for the qualitative determination of morphine (a drug in the opiate class) in human urine at a Cut-Off concentration of 2000 ng/mL. The test is available in a Strip format, a Cassette format, a Dip Card format and a Cup format.

The test provides only preliminary test results. A more specific alternative chemical method must be used in order to obtain a confirmed analytical result. GC/MS is the preferred confirmatory method. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary result is positive.

For in vitro diagnostic use only. It is intended for over-the-counter and for prescription use.

7. Device Description

Healgen Oxazepam Test and Healgen Morphine Test are immunochromatographic assays for Oxazepam and Morphine. Each assay test is a lateral flow system for the qualitative detection of Oxazepam and Morphine (target analyte) in human urine. The products are in vitro diagnostic devices, which come in the form of: Strips, Cassettes, DipCards, or Cups. Each product contains a Test Device (in one of the four formats), and a package insert. Each test device is sealed with a desiccant in an aluminum pouch.

8. Substantial Equivalence Information

A summary comparison of features of the Healgen Oxazepam Test and Healgen Morphine Test and the predicate device is provided in Table 1 & Table 2.

ItemDevicePredicate -K052115
Intended UseFor the qualitative determination of drugsof abuse in human urine.Same
Drug AnalyteOxazepamBenzodiazepineDrug Class
MethodologyCompetitive binding, lateral flowimmunochromatographic assays based onthe principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values300 ng/mLSame

Table 1: Features Comparison of Healgen Oxazepam Test and the Predicate Device

{5}------------------------------------------------

ItemDevicePredicate -K052115
IntendedPopulationFor over-the-counter and prescriptionuses.Forover-the-counteruse.
ConfigurationsStrip, Cassette, Cup, Dip CardCup
Table 2: Features Comparison of Healgen Morphine Test and the Predicate Device
--------------------------------------------------------------------------------------------
ItemDevicePredicate - K052115
Intended UseFor the qualitative determination ofdrugs of abuse in human urine.Same
Drug AnalyteMorphineOpiate Drug Class
MethodologyCompetitive binding, lateral flowimmunochromatographic assays basedon the principle of antigen antibodyimmunochemistry.Same
Specimen TypeHuman UrineSame
Cut-Off Values2000 ng/mLSame
IntendedFor over-the-counter and prescriptionFor over-the-counter
Populationuses.use.
ConfigurationsStrip, Cassette, Cup, Dip CardCup

9. Test Principle

Healgen Oxazepam Test and Healgen Morphine Test are rapid tests for the qualitative detection of Oxazepam and Morphine in urine samples. Each assay test is a lateral flow chromatographic immunoassay. During testing, a urine specimen migrates upward by capillary action. If target drugs are present in the urine specimen below its cut-off concentration, it will not saturate the binding sites of its specific antibody (monoclonal mouse antibody) coated on the particles. The antibody-coated particles will then be captured by immobilized drug-conjugate and a visible colored line will show up in the test line region. The colored line will not form in the test line region if the target drug level exceeds its cut-off concentration because it will saturate all the binding sites of the antibody coated on the particles. A band should form in the control region of the devices regardless of the presence of drug or metabolite in the sample.

10. Performance Characteristics

{6}------------------------------------------------

1. Analytical Performance

a.Precision

Precision studies were carried out for samples with concentrations of -100% cut-off, -50% cut-off, -25% cut-off, at the cut-off, +25% cut-off, +50% cut-off, +75% cut-off and +100% cut-off. These samples were prepared by spiking drug in negative samples. Each drug concentration was confirmed by GC/MS. All sample aliquots were blind labeled and randomized. For each concentration, tests were performed two runs per day for 25 days. The results obtained are summarized in the following tables:

Oxazepam

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot: 120100150-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot: 120100250-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot: 120100350-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

Strip Format

Cassette Format

Result -100%-75% --50%-25%+25% +50% +75% +100%
DrugCut-off Cut-off Cut-off Cut-offCut-offCut-off Cut-off Cut-off Cut-off
Lot: 120100450-10+50-/0+ 50-/0+ 50-/0+ 24-/26+ 50+/0- 50+/0- 50+/0- 50+/0-
Lot: 120100550-10+50-/0+ 50-/0+ 50-/0+ 24-/26+ 50+/0- 50+/0- 50+/0- 50+/0-
Lot: 120100650-/0+ 50-/0+ 50-/0+ 50-/0+ 24-/26+ 50+/0- 50+/0- 50+/0- 50+/0-

Dip Card Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
DrugLot: 120100750-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-
Lot: 120100850-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-
Lot: 120100950-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-

CUP Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot: 120101050-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot: 120101150-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot: 120101250-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-

{7}------------------------------------------------

Morphine

Strip Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
DrugLot: 111200150-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-
Lot: 111200250-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-
Lot: 111200350-/0+50-/0+50-/0+50-/0+18-/32+50+/0-50+/0-50+/0-50+/0-

Cassette Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
DrugLot: 111200450-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot:111200550-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot: 111200650-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

Dip Card Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot: 111200750-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot: 111200850-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-
Lot: 111200950-/0+50-/0+50-/0+50-/0+22-/28+50+/0-50+/0-50+/0-50+/0-

CUP Format

Result-100%Cut-off-75%Cut-off-50%Cut-off-25%Cut-offCut-off+25%Cut-off+50%Cut-off+75%Cut-off+100%Cut-off
Drug
Lot: 111201050-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot: 111201150-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-
Lot: 111201250-/0+50-/0+50-/0+50-/0+20-/30+50+/0-50+/0-50+/0-50+/0-

b. Linearity

Not applicable.

c.Stability

The devices are stable at 4-30℃ for 24 months based on the accelerated stability study at 45℃ and real time stability determination at both 4 ℃ and 30℃.

{8}------------------------------------------------

Control materials are not provided with the device. The labeling provides information on how to obtain control materials.

d. Cut-off

A total of 150 samples equally distributed at concentrations of -50% cut-off; -25% cut-off; cut-off; +25% cut-off; +50% cut-off were tested using three different lots of each device by three different operators. Results were all positive at and above +25% cut-off and all negative at and below -25% cut-off for both Oxazepam and Morphine. The following cut-off values for the test devices have been verified.

TestCalibratorCut-off(ng/mL)
Oxazepam TestOxazepam300
Morphine TestMorphine2000

e. Interference

Potential interfering substances found in human urine of physiological conditions were added to drug-free urine and target drugs urine with concentration at 25% above cut-off levels. These urine samples were tested using three batches of each device for all formats.

Compounds that showed no interference at a concentration of 100µg/mL are summarized in the following tables. There were no differences observed for different formats.

4-AcetamidophenolDoxylamineOxolinic acid
AcetophenetidinEcaonine dydrochloridePentobarbital
N-AcetylprocainamideEcgonine methylesterPerphenazine
Acetylsalicylic acid(-)-Ψ-EphedrinePhencyclidine
AminoprineFenoprofenPhenelzine
AmitriptylineFurosemidePhenobarbital
AmorbarbitalGentisic acidPhentermine
AmoxicillinHemoglobinL-Phenylephrine
AmpicillinHydrocortisonePhenylethylamine
1-Ascorbic AcidO-Hydroxyhippuric acidPhenylpropanotamine
D.L-Amphetaminep-Hydroxy-methamphetaminePrednisone
Aporphine3-HydroxytyramineD.L-Propanolol

Oxazepam

{9}------------------------------------------------

AspartameIbuprofenD-Propoxyphene
AtropineImipramineD-Pseudoephedrine
Benzillic acidIproniazidQuinine
Benzoic acid(±)IsoproterenolRanitidine
BenzoylecaonineIsoxsuprineSalicylic acid
BenzphetamineKetamineSecobarbital
BilirubinKetoprofenSerotonin(5-Hydroxytyramine)
(±) ChlorpheniramineLabetalolSertraline
CaffeineLoperamideSulfamethazine
CannabidiolMaprotilineSulindac
ChloralhydrateMeperidineTetrahydrocortisone,3 Acetate
ChloramphenicolMeprobamateTetrahydrocortisone,(β-Dglucuronide)
ChlorothiazideMethadoneTetrahydrozoline
(±)ChlorpheniramineMethoxyphenamineThiamine
Chlorpromazine(+) 3,4-Methylenedioxy-amphetamineThioridazine
Chlorquine(+)3,4-Methylenedioxy-methamphetamineD.L-Tyrosine
CholesterolNalidixic acidTolbutamide
ClomipramineNalorphineTriamterene
ClonidineNaloxoneTrifluoperazine
Cocaine hydrochlorideNaltrexoneTrimethoprim
CortisoneNaproxenTryptamine
(-)cotinineNiacinamideD.L-Tryptophan
CreatinineNifedipineTyramine
DextromethlorphanNorethindroneUric acid
DiclolrfenacD-NorpropoxypheneVerapamil
DiflunisalNoscapineZomepirac
DiaoxinD.L-Octopamine
DiphenhydramineOxalic acid

Morphine

4-AcetamidophenolEcgonine methylesterOxolinic acid
Acetophenetidin(-) -Y -EphedrineOxymetazoline
N-AcetylprocainamideErythromycinPapaverine
Acetylsalicylic acidβ-EstradiolPenicillin-G
AminopyrineEstrone-3-sulfatePentazocine
AmitryptylineEthyl-p-aminobenzoatePentobarbital
AmobarbitalFenoprofenPerphenazine
AmoxicillinFurosemidePhencyclidine
AmpicillinGentisic acidPhenelzine
Ascorbic acidHemoglobinPhenobarbital
D,L-AmphetamineHydralazinePhentermine
ApomorphineHydrochlorothiazideL-Phenylephrine
AspartameHydrocortisoneβ-Phenylethylamine
AtropineO-Hydroxyhippuric acidPhenylpropanolamine
Benzilic acidp-Hydroxy methamphetaminePrednisone
Benzoic acid3-HydroxytyramineD,L-Propanolol
BenzoylecgonineIbuprofenD-Propoxyphene
BenzphetamineImipramineD-Pseudoephedrine
Bilirubin (±)IproniazidQuinidine
BrompheniramineIsoproterenolQuinine
CaffeineIsoxsuprineRanitidine
CannabidiolKetamineSalicylic acid
ChloralhydrateKetoprofenSecobarbital
ChloramphenicolLabetalolSerotonin (5-Hydroxytyramine)
ChlordiazepoxideLoperamideSulfamethazine
ChlorothiazideMaprotilineSulindac
(±) ChlorpheniramineMeperidineTemazepam
ChlorpromazineMeprobamateTetracycline
ChlorquineMethadoneTetrahydrocortisone, 3Acetate
CholesterolMethoxyphenamineTetrahydrocortisone3 (ß-Dglucuronide)
Clomipramine(+) 3,4-Methylenedioxy-amphetamineTetrahydrozoline
Clonidine(+)3,4-Methylenedioxy-methamphetamineThiamine
Cocaine hydrochlorideNalidixic acidThioridazine
CortisoneNalorphineD, L-Tyrosine
(-) CotinineNaloxoneTolbutamide
CreatinineNaltrexoneTriamterene
DeoxycorticosteroneNaproxenTrifluoperazine
DextromethorphanNiacinamideTrimethoprim
DiazepamNifedipineTrimipramine
DiclofenacNorethindroneTryptamine
DiflunisalD-NorpropoxypheneD, L-Tryptophan
DigoxinNoscapineTyramine
DiphenhydramineD,L-OctopamineUric acid
DoxylamineOxalic acidVerapamil
Ecgonine hydrochlorideOxazepamZomepirac

{10}------------------------------------------------

{11}------------------------------------------------

f. Specificity

To test the specificity, drug metabolites and other components that are likely to interfere in urine samples were tested using three batches of each device for all formats. The obtained lowest detectable concentration was used to calculate the cross-reactivity. There were no differences observed for different formats.

Result%Cross-Reactivity
Oxazepam(Cut-off=300 ng/mL)Positive at 300 ng/mL100%
AlprazolamPositive at 200 ng/mL150%
BromazepamPositive at 1560 ng/mL19%
Chlordiazepoxide HCLPositive at 1560 ng/mL19%
ClobazamPositive at 100 ng/mL300%
ClonazepamPositive at 780 ng/mL38%
Clorazepate DipotassiumPositive at 200 ng/mL150%
DelorazepamPositive at 1560 ng/mL19%
DesalkylflurazepamPositive at 400 ng/mL75%
DiazepamPositive at 200 ng/mL150%
EstazolamPositive at 2500 ng/mL12%
FlunitrazepamPositive at 400 ng/mL75%
a-HydroxyalprazolamPositive at 1260 ng/mL24%
(±) LorazepamPositive at 1560 ng/mL19%
RS-Lorazepam glucuronidePositive at 160 ng/mL188%
MidazolamPositive at 12500 ng/mL2.4%
NitrazepamPositive at 100 ng/mL300%
NorchlordiazepoxidePositive at 200 ng/mL150%
NordiazepamPositive at 400 ng/mL75%
TemazepamPositive at 100 ng/mL300%
TriazolamPositive at 2500 ng/mL12%

{12}------------------------------------------------

Result% Cross-Reactivity
Morphine(Cut-off=2000 ng/mL)Positive at 2000 ng/mL100%
O6-AcetylmorphinePositive at 2500 ng/mL80%
CodeinePositive at 1000 ng/mL200%
EthylMorphinePositive at 250 ng/mL800%
HeroinPositive at 5000 ng/mL40%
HydromorphonePositive at 2500 ng/mL80%
HydrocodonePositive at 5000 ng/mL40%
OxycodonePositive at 75000 ng/mL3%
ThebainePositive at 13,000 ng/mL15%

g. Effect of Urine Specific Gravity and Urine pH

To investigate the effect of urine specific gravity and urine pH, urine samples with of 1.000 to 1.035 specific gravity or urine samples with pH 4 to 9 were spiked with target drugs at 25% below and 25% above cut-off levels. These samples were tested using three batches of each device for all formats. Results were all positive for samples at and above +25% cut-off and all negative for samples at and below -25% Cut-Off. There were no differences observed for different formats.

2. Comparison Studies

The method comparison studies for the Oxazepam Test, and the Morphine Test were performed in-house with three different laboratory assistants for each format of the device. Operators ran 80 (40 negative and 40 positive) unaltered clinical samples. The samples were blind labeled and compared to GC/MS results. The results are presented in the tables below:

Oxazepam
Strip formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cut-off)Near Cutoff Positive by GC/MS (Between the cut-off and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001324
Negative10151530
Viewer CPositive0001424
Negative10151520

Oxazepam

{13}------------------------------------------------

ViewerSample NumberGC/MS ResultStrip FormatViewer Results
Viewer A549306Negative
Viewer A523303Negative
Viewer B549306Negative
Viewer B523303Negative
Viewer B579312Negative
Viewer C549306Negative
Viewer C523303Negative

Discordant Results of Oxazepam Strip

CassetteformatNegativeLowNegativeby GC/MS(less than-50%)NearCutoffNegativeby GC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Positive0001324
Viewer ANegative10ાં રાં ર30
Viewer BPositive0001324
Negative10ા રા ર30
Viewer CPositive0001324
Negative10ા રા ર30

Discordant Results of Oxazepam Cassette

ViewerSample NumberGC/MS ResultCassette FormatViewer Results
Viewer A549306Negative
Viewer A523303Negative
Viewer A579312Negative
Viewer B549306Negative
Viewer B523303Negative
Viewer B579312Negative
Viewer C549306Negative
Viewer C523303Negative
Viewer C579312Negative

{14}------------------------------------------------

Cup formatNegativeLow Negative by GC/MS (less than -50%)Near Cutoff Negative by GC/MS (Between -50% and cut-off)Near Cutoff Positive by GC/MS (Between the cut-off and +50%)High Positive by GC/MS (greater than +50%)
Viewer APositive0001424
Negative10151520
Viewer BPositive0001424
Negative10151520
Viewer CPositive0001424
Negative10151520

Discordant Results of Oxazepam Cup

ViewerSample NumberGC/MS ResultCup FormatViewer Results
Viewer A549306Negative
Viewer A523303Negative
Viewer B523303Negative
Viewer B579312Negative
Viewer C549306Negative
Viewer C523303Negative
Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001224
Negative10151540
Viewer BPositive0001224
Negative10151540
Viewer CPositive0001424
Negative10151520

Discordant Results of OxazepamDip Card

Discordant Results of OxazepamDip Card
ViewerSample NumberDip Card FormatGC/MS ResultViewer Results
Viewer Aરે રેણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, આંગણવાડી તેમ જ દૂધની ડેરી જેવી સવલતો પ્રાપ્ય થયેલી છે. આ ગામનાં પ્રાથમિક શાળા, આંગણવાડી તેમ315Negative
Viewer A549306Negative

{15}------------------------------------------------

ViewerSample NumberGC/MS ResultDip Card FormatViewer Results
Viewer A523303Negative
Viewer A579312Negative
Viewer B505315Negative
Viewer B549306Negative
Viewer B523303Negative
Viewer B579312Negative
Viewer C549306Negative
Viewer C523303Negative

Morphine

StripformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001323
Negative10161440
Viewer BPositive0001323
Negative10161440
Viewer CPositive0001523
Negative10161420

Discordant Results of Morphine Strip

ViewerSample NumberGC/MS ResultStrip Format Viewer Results
Viewer A4682036Negative
Viewer A4662019Negative
Viewer A4532066Negative
Viewer A4492005Negative
Viewer B4682036Negative
Viewer B4662019Negative
Viewer B4532066Negative
Viewer B4492005Negative
Viewer C4662019Negative

{16}------------------------------------------------

ViewerSample NumberGC/MS ResultStrip Format Viewer Results
Viewer C4492005Negative
CassetteformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001423
Negative10161430
Viewer BPositive0001323
Negative10161440
Viewer CPositive0001323
Negative10161440

Discordant Results of Morphine Cassette

ViewerSample NumberGC/MS ResultCassette Format Viewer Results
Viewer A4682036Negative
Viewer A4662019Negative
Viewer A4492005Negative
Viewer B4052053Negative
Viewer B4682036Negative
Viewer B4662019Negative
Viewer B4492005Negative
Viewer C4052053Negative
Viewer C4682036Negative
Viewer C4662019Negative
Viewer C4492005Negative

{17}------------------------------------------------

Dip CardformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001323
Viewer ANegative10161440
Viewer BPositive0001223
Viewer BNegative10161450
Viewer CPositive0001223
Viewer CNegative10161450

Discordant Results of Morphine Dip Card

ViewerSample NumberGC/MS ResultDip Card FormatViewer Results
Viewer A4052053Negative
Viewer A4682036Negative
Viewer A4662019Negative
Viewer A4492005Negative
Viewer B4052053Negative
Viewer B4682036Negative
Viewer B4662019Negative
Viewer B4492005Negative
Viewer B4292067Negative
Viewer C4052053Negative
Viewer C4682036Negative
Viewer C4662019Negative
Viewer C4492005Negative
Viewer C4292067Negative

{18}------------------------------------------------

CupformatNegativeLowNegativeby GC/MS(less than-50%)Near CutoffNegative byGC/MS(Between-50% andcut-off)Near CutoffPositive byGC/MS(Betweenthe cut-offand +50%)HighPositive byGC/MS(greaterthan+50%)
Viewer APositive0001523
Negative10161420
Viewer BPositive0001423
Negative10161430
Viewer CPositive0001323
Negative10161440

Discordant Results of Morphine Cup

ViewerSample NumberGC/MS ResultCup Format Viewer Results
Viewer A4662019Negative
Viewer A4492005Negative
Viewer B4682036Negative
Viewer B4662019Negative
Viewer B4492005Negative
Viewer C4052053Negative
Viewer C4682036Negative
Viewer C4662019Negative
Viewer C4492005Negative

Lay-user study

A lay user study was performed at three intended user sites with 140 lay persons testing the Oxazepam devices and another set of 140 persons testing the morphine devices. A total of 44 females and 96 males tested the Oxazepam samples, and 41 females and 99 males tested the Morphine samples. They had diverse educational and professional backgrounds and ranged in age from 21 to > 50 years. Urine samples were prepared at the following concentrations; negative, +/-75%, +/-50%, +/-25% of the cutoff by spiking drugs into drug free-pooled urine specimens. The concentrations of the samples were confirmed by GC/MS. Each sample was aliquoted into individual containers and blind-labeled. Each participant was provided with the package insert, 1 blind labeled sample and a device. The results are summarized below.

{19}------------------------------------------------

% of CutoffNumber of samplesOxazepam Concentration by GC/MS (ng/mL)Lay person results No. of PositiveLay person results No. of NegativeThe percentage of correct results (%)
-100% Cutoff200020100%
-75% Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022511995%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%

Comparison between GC/MS and Lay Person Results (Oxazepam Strip)

Comparison between GC/MS and Lay Person Results (Oxazepam Cassette)

NumberOxazepam ConcentrationLay person resultsThe
% of Cutoffofsamplesby GC/MS(ng/mL)No. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022521890%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%

Comparison between GC/MS and Lay Person Results (Oxazepam DipCard)

NumberofsamplesOxazepam Concentrationby GC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022511995%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%

{20}------------------------------------------------

NumberofsamplesOxazepam Concentrationby GC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff2075020100%
-50% Cutoff20150020100%
-25% Cutoff2022521890%
+25% Cutoff2037519195%
+50% Cutoff20450200100%
+75% Cutoff20525200100%

Comparison between GC/MS and Lay Person Results (Oxazepam Cup)

Comparison between GC/MS and Lay Person Results (Morphine Strip)

% of CutoffNumber of samplesMOP Concentration by GC/MS (ng/mL)Lay person resultsThe percentage of correct results (%)
-100% Cutoff200020100%
-75% Cutoff20500020100%
-50% Cutoff201000020100%
-25% Cutoff20150021890%
+25% Cutoff20250019195%
+50% Cutoff203000200100%
+75% Cutoff203500200100%

Comparison between GC/MS and Lay Person Results (Morphine Cassette)

NumberofsamplesMOP Concentration byGC/MS(ng/mL)Lay person resultsThe
% of CutoffNo. ofPositiveNo. ofNegativepercentage ofcorrect results(%)
-100%Cutoff200020100%
-75%Cutoff20500020100%
-50% Cutoff201000020100%
-25% Cutoff20150011995%
+25% Cutoff20250019195%
+50% Cutoff203000200100%
+75% Cutoff203500200100%

{21}------------------------------------------------

% of CutoffNumberofsamplesMOP Concentration byGC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
No. ofPositiveNo. ofNegative
-100% Cutoff200020100%
-75% Cutoff20500020100%
-50% Cutoff201000020100%
-25% Cutoff20150021890%
+25% Cutoff20250018290%
+50% Cutoff203000200100%
+75% Cutoff203500200100%

Comparison between GC/MS and Lay Person Results (Morphine DipCard)

Comparison between GC/MS and Lay Person Results (Morphine Cup)

% of CutoffNumberofsamplesMOP Concentration byGC/MS(ng/mL)Lay person resultsThepercentage ofcorrect results(%)
-100% Cutoff200020100%
-75% Cutoff20500020100%
-50% Cutoff201000020100%
-25% Cutoff20150021890%
+25% Cutoff20250019195%
+50% Cutoff203000200100%
+75% Cutoff203500200100%

Lay-users were also given surveys on the ease of understanding the package insert instructions. All lay users indicated that the device instructions can be easily followed. A Flesch-Kincaid reading analysis was performed on each package insert and the scores revealed a reading Grade Level of 7.

    1. Clinical Studies
      Not applicable.

11.Conclusion

Based on the test principle and acceptable performance characteristics including precision, cut-off, interference, specificity and method comparison of the devices, it's concluded that the Healgen Oxazepam Test, and Healgen Morphine Test are substantially equivalent to the predicate.

§ 862.3650 Opiate test system.

(a)
Identification. An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.(b)
Classification. Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (e.g., programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).