K Number
K142249

Validate with FDA (Live)

Date Cleared
2015-01-27

(166 days)

Product Code
Regulation Number
874.4760
Age Range
All
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticPediatricDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Schoelly Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including thinology, and endoscopic plastic and reconstructive surgery.

Device Description

The Schoelly Sinuscope is a rigid reusable endoscope for visualization during Otolaryngology-Head and Neck surgery in conjunction with a commercially available and approved light guide, light source, video camera, monitor, and printer. Light source, video camera, monitor, and printer are not included in the scope of delivery and are further not within the scope of this application.

Schoelly Sinuscopes are manufactured in multiple configurations that differ in insertion tube outer diameter and working length and with respect to optical parameters (direction of view, field of view). Several models of the Schoelly Sinuscope have already been cleared for marketing by FDA (K133682) for the visualization during arthroscopic procedures.

Like other currently marketed sinuscopes, Schoelly Sinuscopes have outer surfaces mainly made from metal (Phynox cobalt-nickel-chromium alloy, 304 stainless steel) and further comprise fiber optics for light transmission and rigid rod-lenses for image transmission.

The Schoelly Sinuscope is delivered in a non-sterile condition and is already CE marked.

AI/ML Overview

The Schoelly Sinuscope is a rigid reusable endoscope intended for visualization during Otolaryngology-Head and Neck surgery. The device's performance was evaluated through non-clinical testing, including temperature, optical parameter, biocompatibility, and performance testing, demonstrating its substantial equivalence to the predicate device, SHARPSITE Ac (K965233).

1. Table of Acceptance Criteria and Reported Device Performance

Test CategoryAcceptance CriteriaReported Device Performance
TemperatureMeet requirements specified in IEC 60601-2-18 and IEC 60601-1 for surface temperatures at various locations over time using different light sources.The device was measured for surface temperatures at various locations over time using different light sources and found to meet requirements specified in IEC 60601-2-18 and IEC 60601-1.
Optical ParametersMeet minimum requirements defined in internal specifications and as specified in the ISO 8600 series of standards (field of view, direction of view, diopters, image eccentricity, size of view, and vignetting).The device was tested for all relevant optical parameters and found to meet the minimum requirements defined in internal specifications and as specified in the ISO 8600 series of standards (field of view, direction of view, diopters, image eccentricity, size of view, and vignetting).
BiocompatibilityDemonstrate biocompatibility of device components in contact with the patient, according to ISO 10993 (including cytotoxicity, sensitization, irritation, and acute systemic toxicity).A series of biocompatibility testing according to ISO 10993, including cytotoxicity, sensitization, irritation, and acute systemic toxicity, demonstrated that the device components that are in contact with the patient are biocompatible.
PerformanceDemonstrate comparable bonding strength for the Sinuscope distal glass to the distal glass bonding strength of the currently marketed predicate device. The specific quantitative acceptance criteria for bonding strength are not explicitly stated, but the criterion is implicitly comparability to the predicate.Performance testing consisted of measurement of the bonding strength for the Sinuscope distal glass, with comparison to the distal glass bonding strength of the currently marketed predicate device. The results demonstrated that the Schoelly Sinuscope has met pre-determined acceptance criteria (implying comparability to the predicate).
ReprocessingValidated manual cleaning, automated cleaning, steam sterilization, and STERRAD® 100S, STERRAD® 100NX, and STERRAD® NX sterilization, in accordance with AAMI TIR12:2010, AAMI TIR30:2011, ANSI/AAMI ST15883-1: 2009, ISO 14937:2009, ANSI/AAMI ST81:2004, ISO 17664:2004, and ANSI/AAMI/ISO 17665-1:2006.The Schoelly Sinuscope is the subject of completed reprocessing validations including manual cleaning, automated cleaning, steam sterilization, as well as STERRAD® 100S, STERRAD® 100NX and STERRAD® NX sterilization, following the aforementioned standards. Devices used for testing underwent multiple reprocessing cycles and artificial surface marring to simulate end-of-lifetime use.

2. Sample Size Used for the Test Set and Data Provenance

The document does not explicitly state a specific "sample size" for a test set in the traditional sense of a clinical study with patient data. The testing described is primarily non-clinical, involving laboratory-based evaluations of the device's physical and material properties.

  • Test Set: Not applicable in the context of patient data. The "test sets" refer to the Schoelly Sinuscope devices themselves that were subjected to the various physical and chemical tests (e.g., specific units for temperature testing, optical testing, biocompatibility testing, reprocessing validation). The exact number of devices used for each specific test is not detailed.
  • Data Provenance: The studies are non-clinical, laboratory-based tests of the device. There is no mention of geographical provenance (e.g., country of origin) or retrospective/prospective classification as it relates to human subject data. The testing was conducted by or on behalf of Schoelly Fiberoptic GmbH, located in Denzlingen, Germany.

3. Number of Experts Used to Establish the Ground Truth for the Test Set and Qualifications

Not applicable. The ground truth for the non-clinical tests (e.g., temperature, optical parameters, biocompatibility, bonding strength, reprocessing effectiveness) is established by adherence to recognized international standards (e.g., IEC, ISO, AAMI) and internal specifications, rather than expert consensus on medical images or clinical outcomes. The "ground truth" is defined by the objective physical or chemical properties of the device and its demonstrated ability to meet the quantitative and qualitative requirements of these standards.

4. Adjudication Method for the Test Set

Not applicable. As the testing is non-clinical and objective (e.g., measurements against a standard), there is no need for adjudication by multiple human observers to establish a ground truth.

5. Multi Reader Multi Case (MRMC) Comparative Effectiveness Study

No, a Multi Reader Multi Case (MRMC) comparative effectiveness study was not conducted. This type of study is typically performed to evaluate the diagnostic accuracy or effectiveness of an AI system or diagnostic device when interpreted by multiple human readers, often comparing performance with and without AI assistance. The Schoelly Sinuscope is a physical medical device (an endoscope) for direct visualization, not an AI-powered diagnostic tool, and its evaluation did not involve human reader interpretation of data for diagnostic purposes in an MRMC setting.

6. Standalone (Algorithm Only Without Human-in-the-Loop Performance) Study

Not applicable. The Schoelly Sinuscope is a physical medical device, not an algorithm or AI system. Therefore, a standalone performance study without human-in-the-loop is not relevant.

7. Type of Ground Truth Used

The ground truth used for performance evaluation is derived from:

  • International Standards and Regulations: e.g., IEC 60601-2-18, IEC 60601-1 (for temperature), ISO 8600 series (for optical parameters), ISO 10993 (for biocompatibility), AAMI TIRs, ANSI/AAMI STs, and ISO 14937, ISO 17664, ANSI/AAMI/ISO 17665-1 (for reprocessing).
  • Internal Specifications: These specifications set the minimum requirements for the device's optical parameters.
  • Predicate Device Performance: For performance testing (e.g., bonding strength), the "ground truth" or acceptance criterion was defined by comparison to the established performance of the legally marketed predicate device, SHARPSITE Ac (K965233).

8. Sample Size for the Training Set

Not applicable. Since the Schoelly Sinuscope is a physical medical device and not an AI/machine learning system, there is no "training set" in the context of algorithm development.

9. How the Ground Truth for the Training Set Was Established

Not applicable. There is no training set as this is not an AI/machine learning device.

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized representation of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in black and are connected by flowing lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

January 27, 2015

Schoelly Fiberoptic Gmbh % Ms. Pamela Papineau, RAC President Delphi Medical Device Consulting, Inc. 5 Whitcomb Ave Ayer, Massachusetts 01432

K142249 Trade/Device Name: Schoelly Sinuscope Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible or Rigid) And Accessories Regulatory Class: Class II Product Code: EOB Dated: December 22, 2014 Received: December 24, 2014

Dear Ms. Papineau:

Re:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

Indications for Use

510(k) Number (if known)

Device Name Schoelly Sinuscope

Indications for Use (Describe)

The Schoelly Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including thinology, and endoscopic plastic and reconstructive surgery.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

K142249 - 510(k) Summary

General Information

Preparation date:11/24/2014
Owner's Name:Schoelly Fiberoptic GmbH (Registration: 8043903)
Address:Robert-Bosch-Str. 1 - 379211 DenzlingenGermany
Telephone Number:+49-7666-980-0
Fax Number:+49-7666-908-380
Contact Person:Dr. Sandra Baumann
Subject Device Name:Schoelly Sinuscope
Trade Name:Schoelly Sinuscope
Common/Usual Name:Sinuscope
Classification Name:EOB - Nasopharyngoscope (flexible or rigid)21 CFR 874.4760; Class II
Predicate Device Name:SHARPSITE Ac
Trade Name:SHARPSITE Ac
Common/Usual Name:Rigid Rod Lense Endoscope
Classification Name:EOB - Nasopharyngoscope (flexible or rigid)21 CFR 874.4760; Class II
Premarket Notification:K965233, Medtronic Xomed Inc., SE date April 4, 1997

Device Description

The Schoelly Sinuscope is a rigid reusable endoscope for visualization during Otolaryngology-Head and Neck surgery in conjunction with a commercially available and approved light guide, light source, video camera, monitor, and printer. Light source, video camera, monitor, and printer are not included in the scope of delivery and are further not within the scope of this application.

Schoelly Sinuscopes are manufactured in multiple configurations that differ in insertion tube outer diameter and working length and with respect to optical parameters (direction of view, field of view). Several models of the Schoelly Sinuscope have already been cleared for marketing by FDA (K133682) for the visualization during arthroscopic procedures.

Like other currently marketed sinuscopes, Schoelly Sinuscopes have outer surfaces mainly made from metal (Phynox cobalt-nickel-chromium alloy, 304 stainless steel) and further comprise fiber optics for light transmission and rigid rod-lenses for image transmission.

The Schoelly Sinuscope is delivered in a non-sterile condition and is already CE marked.

{4}------------------------------------------------

Technological Characteristics

Light that is created by an external light source is transmitted from the Sinuscope light guide connector through the Sinuscope itself to the tip via a fiber optic system. Images are transferred the other way back through a rigid rod lens system. Technical parameters of the Schoelly Sinuscopes that characterize the optical view are the Direction of View (0°-70°) and the Field of View (80°-100°). The image can be displayed by a camera/monitor system which can be connected to the Sinuscope eyepiece. The Schoelly Sinuscopes are available in two different configurations for the insertion tube: 4mm diameter, working length 175mm and 2.7mm diameter, working length 110mm. All configurations of the Schoelly Sinuscope do not have any working channel.

A detailed comparison of the technological characteristics with the predicate device is shown in the table below.

ProposedDevice(current submission)PredicateDevice (K965233)
Attribute
LighttransmissionFiber opticsFiber optics
Light sourceExternal, connected via lightguide to light guide connectorExternal, connected via lightguide to light guide connector
ImagetransmissionRigid rod lensesRigid rod lenses
Direction ofview$0°-70°$$0°-70°$
Field of view$80°-100°$$80°-100°$
Image displayCamera/monitor connected viathe endoscope eyepieceCamera/monitor connected viathe endoscope eyepiece
Workingchannelnonenone
Single Use /ReusableReusableReusable
ReprocessingCleaning, sterilization (steam,STERRAD®)Cleaning, sterilization (steam,STERRAD®, EtO)
Materials
- insertion tube(outer surface)stainless steel, Co-Cr-Ni alloystainless steel
- insertion part(distal surface)stainless steel, glassstainless steel, glass
- fiber opticsglass fibersglass fibers
- lens bondingAuSn brazing alloy, epoxyepoxy

Table 5.1 Technological Characteristics of the Proposed Device and the Predicate Device

{5}------------------------------------------------

AttributeProposedDevice(current submission)PredicateDevice (K965233)
TemperaturetestingIEC 60601-2-18 compliantIEC 60601-2-18 compliant
PackagingCase with foam insertCase with foam insert

Indications for Use

The Schoelly Sinuscope is intended for use in otolaryngology and Head and Neck procedures, including rhinology, and endoscopic plastic and reconstructive surgery.

The indications for use of the proposed devices are identical to the indications for use of the Medtronic predicate device.

Non-clinical Performance Testing

The Schoelly Sinuscope was subjected to temperature, optical parameter, biocompatibility and performance testing with identical acceptance criteria for the Schoelly and the predicate devices. The results of this testing demonstrated that the Schoelly Sinuscope has met pre-determined acceptance criteria and is substantially equivalent to the predicate device. The risks associated with the use of the new device were found acceptable when evaluated in accordance with ISO 14971. Risks and benefits of predicate device are the same as compared to the Schoelly Sinuscope.

Temperature testing: The device was measured for surface temperatures at various locations over time using different light sources and found to meet requirements specified in IEC 60601-2-18 and IEC 60601-1.

Optical parameter testing: The device was tested for all relevant optical parameters and found to meet the minimum requirements defined in internal specifications and as specified in the ISO 8600 series of standards (field of view, direction of view, diopters, image eccentricity, size of view, and vignetting).

Biocompatibility testing: A series of biocompatibility testing according to ISO 10993, including cytotoxicity, sensitization, irritation, and acute systemic toxicity, demonstrated that the device components that are in contact with the patient are biocompatible.

Performance testing: Performance testing consisted of measurement of the bonding strength for the Sinuscope distal glass, with comparison to the distal glass bonding strength of the currently marketed predicate device.

Reprocessing

The Schoelly Sinuscope is the subject of completed reprocessing validations including manual cleaning, automated cleaning, steam sterilization, as well as STERRAD® 100S, STERRAD® 100NX and STERRAD® NX sterilization.

Cleaning studies have been performed in accordance with AAMI TIR12:2010 (Designing. Testing and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A

{6}------------------------------------------------

Guide for Device Manufacturers), AAMI TIR30:2011 (A compendium of processes, materials, test methods, and acceptance criteria for cleaning reusable devices), and ANSI/AAMI ST15883-1: 2009 (Washer-disinfectors - Part 1 - General requirements, terms, and definitions and tests). Devices that have been used for testing had been undergo multiple reprocessing cycles and surface marring artificially created by scratching with metal tools to simulate end of lifetime use and to address the FDA Draft Guidance Processing Medical Devices in Healthcare Settings (dated: May 2, 2011).

Sterilization studies have been performed in accordance with ISO 14937:2009 (Sterilization of health care products - General requirements for characterization of sterilizing agent and the development, validation and routine control of a sterilization process for medical devices), ANSI/AAMI ST81:2004 (Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical devices), ISO 17664:2004 (Sterilization of medical devices - Information to be provided by the manufacturer for the processing of resterilizable medical devices), and ANSI/AAMI/ISO 17665-1:2006 (Sterilization of health care products - Moist heat - Requirements for the development, validation and routine control of sterilization process for medical devices).

Conclusion

The Schoelly Sinuscope meets all the pre-determined acceptance criteria of the testing performed to demonstrate substantial equivalence to the predicate device.

§ 874.4760 Nasopharyngoscope (flexible or rigid) and accessories.

(a)
Identification. A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.(b)
Classification. Class II.