K Number
K142230

Validate with FDA (Live)

Manufacturer
Date Cleared
2014-09-10

(28 days)

Product Code
Regulation Number
888.3040
Reference & Predicate Devices
Predicate For
AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
Intended Use

The Apollo Suture Anchors and Titan Screws are indicated for use in fixation of ligament, tendon, bone, or soft tissue to bone in knee, shoulder, foot/ankle, elbow, and hand/wrist procedures. The screws are intended for use in the following procedures:
Indications - Apollo Medial Suture Anchor and Apollo XT Suture Anchors: Shoulder (Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction), Foot/Ankle (Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair), Knee (Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis), Elbow (Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction), Hip (Capsular Repair, Acetabular Labral Repair).
Indications - Apollo Labral Suture Anchor: Shoulder (Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction), Wrist (Scapholunate Ligament Reconstruction), Elbow (Biceps Tendon Reattachment, Ulnar or Radial Collateral Ligament Reconstruction), Hip (Capsular Repair, Acetabular Labral Repair), Knee (Extracapsular Repair, Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Joint Capsule Closure, Iliotibial Band Tenodesis Reconstruction, Patellar Ligament/Tendon Repair, Vastus Medialis Obliquus Muscle Advancement).
Indications - Apollo Lateral Anchor: Shoulder (Rotator Cuff Repair, Bankart Repair, SLAP Lesion Repair, Biceps Tenodesis, Acromio-Clavicular Separation Repair, Deltoid Repair, Capsular Shift or Capsulolabral Reconstruction), Wrist/Hand (Scapholunate Ligament Reconstruction, Ulnar/Radial Collateral Ligament Reconstruction), Foot/Ankle (Lateral Stabilization, Medial Stabilization, Achilles Tendon Repair/Reconstruction, Hallux Valgus Reconstruction, Mid and Forefoot Reconstruction), Elbow (Biceps Tendon Reconstruction, Ulnar or Radial Collateral Ligament Reconstruction, Lateral Epicondylitis Repair (PEEK Anchor Only)), Knee (Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Posterior Oblique Ligament Repair, Joint Capsule Closure, Iliotibial Band Tenodesis, Patellar Ligament/Tendon Repair).
Indications –Interference Screws: The Titan Interference Screws are indicated for the reattachment of ligament, tendon, soft tissue, or bone to bone during cruciate ligament reconstruction surgeries of the knee. All screws with a diameter of 9 mm or less and a length of 23 mm or less are also intended for the use in the following procedures: Knee (ACL repairs, PCL repairs, Extra-capsular repairs (Medial collateral ligament, Lateral collateral ligament, Posterior oblique ligament), Patellar realignment and tendon repairs (Vastus medialis oliquous advancement), Iliotibial band tenodesis), Shoulder (Capsular stabilization (Bankart repair, Anterior shoulder instability, SLAP lesion repairs, Capsular shift of capsulolabral reconstructions), Acromioclavicular separation repairs, Deltoid repairs, Rotator cuff tear repairs, Biceps tenodesis), Foot and Ankle (Hallux valgus repairs, Medial or lateral instability repairs/reconstructions, Achilles tendon repairs/reconstructions, Midfoot reconstructions, Metatarsal ligament/tendon repairs/reconstructions, Bunionectomy, Flexor Hullucis Longus, Tendon transfers), Elbow, Wrist, and Hand (Biceps tendon reattachment, Ulnar or radial collateral ligament reconstructions, Lateral epicondylitis repair, Scapholunate ligament reconstruction, Tendon transfers).
Indications –Titan Mini-Interference Screws: The Titan Mini-Interference Screws are intended to be used for fixation of tissue, including ligament or tendon to bone, or a bone/tendon to bone. The Mini-Interference Screws are intended to provide soft tissue reattachment (i.e. fixation of ligament and tendon graft tissue). See below for specific indications. Shoulder (Capsular stabilization (Bankart repair, Anterior shoulder instability, SLAP lesion repairs, Capsular shift of capsulolabral reconstructions), Acromioclavicular separation repairs, Deltoid repairs, Rotator cuff tear repairs, Biceps tenodesis), Foot and Ankle (Hallux valgus reconstruction, Medial stabilization, Lateral stabilization, Achilles Tendon Repair, Midfoot reconstructions, Metatarsal ligament repair, Bunionectomy, Flexor Hullucis Longus for Achilles Tendon reconstruction, Tendon transfers in the foot and ankle), Knee (Medial Collateral Ligament Repair, Lateral Collateral Ligament Repair, Patellar Tendon Repair, Posterior Oblique Ligament Repair, Illiotibial Band Tenodesis, Posterior Cruciate Ligament Repair), Elbow (Biceps tendon reattachment, Ulnar or radial collateral ligament reconstruction), Wrist and Hand (Scapholunate Ligament Reconstruction, Ulnar Collateral Ligament Reconstruction, Radial Collateral Ligament Reconstruction, Carpometalcarpal joint arthroplasty (basal thumb joint arthroplasty), Carpal Ligament Reconstructions and repairs, Tendon transfer in the hand/wrist, Lateral Epicondylitis repair).

Device Description

Apollo Family: The Apollo Medial Suture Anchor, XT Suture Anchor, Lateral Anchor, and Labral Suture Anchor are delivery systems for anchors for use in fixation of ligament, tendon, bone, or soft tissue to bone in knee, shoulder, foot/ankle, elbow, and hand/wrist procedures. These anchors consist of cannulated anchors with integrated suture attachment or separate suture punch eyelet. The Anchors are provided loaded on individual inserters with and without integrated sutures, sterile, for single use only.
Titan Family: The Titan Interference and Mini-Interference Screws are interference screws for use in fixation of ligament, tendon, bone, or soft tissue to bone in knee, shoulder, foot/ankle, elbow, and hand/wrist procedures. The screws are provided sterile, for single use only.
Screw and anchor implants are made from either a titanium alloy (6AI4V) per ASTM F136, or PEEK (Zeniva ZA-500) per ASTM F2026 from Solvay Advanced Polymers.

AI/ML Overview

Here's a breakdown of the requested information based on the provided text.

This document is a 510(k) premarket notification for a medical device called the "Apollo Suture Anchor System and Titan Screws." It's primarily concerned with demonstrating substantial equivalence to a legally marketed predicate device, rather than proving a device meets specific acceptance criteria through a standalone study of its performance in a clinical setting against a gold standard.

Therefore, the typical acceptance criteria and study data associated with AI/software devices (e.g., sensitivity, specificity, clinical outcome measures) are not directly applicable or present in this type of regulatory submission. This submission focuses on demonstrating the device is as safe and effective as a previously cleared device.

However, I can extract the relevant information regarding performance testing that was done to support the safety and effectiveness for this type of device.

1. Table of Acceptance Criteria and Reported Device Performance

Acceptance Criteria (Measured Performance)Reported Device Performance (Reference Standard/Predicate)
Axial Pull-Out (comparison to predicate)Confirmed that modifications did not introduce new risks when compared to the predicate device.
Insertion Torque (comparison to predicate)Confirmed that modifications did not introduce new risks when compared to the predicate device.

Explanation: The "acceptance criteria" here are not numerical thresholds for clinical performance or diagnostic accuracy. Instead, the acceptance criterion for this 510(k) submission is that the modified device performs comparably to the predicate device and that the modifications (adding a 3rd suture portal) do not introduce new risks. The reported device performance is that this comparison was confirmed through mechanical testing.

2. Sample Size Used for the Test Set and Data Provenance

  • Sample Size: Not explicitly stated as a number of devices or cases. The testing conducted was "Axial Pull-Out and Insertion Torque per ASTM F543-7 testing." ASTM F543-7 is a standard test method, but the specific number of samples tested by Valeris Medical is not provided in this summary.
  • Data Provenance: Not applicable in the context of clinical data provenance (country of origin, retrospective/prospective). This refers to mechanical testing data.

3. Number of Experts Used to Establish Ground Truth for the Test Set and Qualifications of Those Experts

Not applicable. This device is a surgical implant (suture anchor and screws). The "ground truth" for its performance is assessed via engineering and mechanical testing standards (ASTM F543-7) rather than expert consensus on medical images or clinical outcomes.

4. Adjudication Method for the Test Set

Not applicable. Adjudication methods are typically relevant for studies involving human interpretation or clinical outcomes. The testing here is mechanical.

5. If a Multi-Reader Multi-Case (MRMC) Comparative Effectiveness Study Was Done, and the Effect Size of How Much Human Readers Improve with AI vs. Without AI Assistance

Not applicable. This document does not describe an AI/software device or a clinical study involving human readers.

6. If a Standalone (i.e., algorithm only without human-in-the-loop performance) Was Done

Not applicable. This is not an algorithm or AI device.

7. The Type of Ground Truth Used

The "ground truth" for the performance evaluation in this context is established by validated mechanical testing standards (ASTM F543-7). This standard provides methods for determining the axial pull-out and insertion torque of metallic bone screws. The goal was to show that the modified device's performance aligned with these established standards and did not significantly differ from the predicate device.

8. The Sample Size for the Training Set

Not applicable. This is not an AI/machine learning device that requires a training set.

9. How the Ground Truth for the Training Set Was Established

Not applicable. This is not an AI/machine learning device that requires a training set.


Summary of the Study Discussed in the Document:

The study described is a performance testing study.

  • Purpose: To confirm that a modification (addition of a 3rd suture portal) to the Apollo Suture Anchor System and Titan Screws did not introduce new risks compared to the predicate device.
  • Methodology: Axial Pull-Out and Insertion Torque testing was conducted according to ASTM F543-7.
  • Conclusion: The testing confirmed that the modification did not introduce any new risks, supporting the substantial equivalence of the subject device to the predicate device (K133036).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the seal of the U.S. Department of Health and Human Services. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined snakes and a staff. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 10, 2014

Valeris Medical % Ms. Cheryl Wagoner Principal Consultant Wagoner Consulting LLC P O Box 15729 Wilmington, North Carolina 28408

Re: K142230

Trade/Device Name: Apollo Suture Anchor System and Titan Screws Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: MBI Dated: August 8, 2014 Received: August 13, 2014

Dear Ms. Wagoner,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related

{1}------------------------------------------------

adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

Indications for Use

510(k) Number (if known): K142230

Device Name:

Apollo Suture Anchor System and Titan Screws

Indications for Use:

The Apollo Suture Anchors and Titan Screws are indicated for use in fixation of ligament, tendon, bone, or soft tissue to bone in knee, shoulder, foot/ankle, elbow, and hand/wrist procedures. The screws are intended for use in the following procedures:

Indications - Apollo Medial Suture Anchor and Apollo XT Suture Anchors
The Apollo Medial Suture Anchor and Apollo XT Suture Anchors are intended for:
Shoulder
Rotator Cuff Repair
Bankart Repair
SLAP Lesion Repair
Biceps Tenodesis
Acromio-Clavicular Separation Repair
Deltoid Repair
Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle
Lateral Stabilization
Medial Stabilization
Achilles Tendon Repair
Knee
Medial Collateral Ligament Repair
Lateral Collateral Ligament Repair
Posterior Oblique Ligament Repair
Illiotibial Band Tenodesis
Elbow
Biceps Tendon Reattachment
Ulnar or Radial Collateral Ligament Reconstruction
Hip
Capsular Repair
Acetabular Labral Repair
Indications - Apollo Labral Suture Anchor
Shoulder
Rotator Cuff Repair
Bankart Repair
SLAP Lesion Repair
Biceps Tenodesis
Acromio-Clavicular Separation Repair
Deltoid Repair
Capsular Shift or Capsulolabral Reconstruction
Indications - Apollo Lateral Anchor
The Apollo Lateral Anchor is indicated for:
Shoulder
Rotator Cuff Repair
Bankart Repair
SLAP Lesion Repair
Biceps Tenodesis
Acromio-Clavicular Separation Repair
Deltoid Repair
Capsular Shift or Capsulolabral Reconstruction
Wrist/Hand
Scapholunate Ligament Reconstruction
Ulnar/Radial Collateral Ligament Reconstruction
Foot/Ankle
Lateral Stabilization
Medial Stabilization
Achilles Tendon Repair/Reconstruction
Hallux Valgus Reconstruction
Mid and Forefoot Reconstruction
Elbow
Biceps Tendon Reconstruction
Ulnar or Radial Collateral Ligament Reconstruction
Lateral Epicondylitis Repair (PEEK Anchor Only)
Knee
Medial Collateral Ligament Repair
Lateral Collateral Ligament Repair
Posterior Oblique Ligament Repair
Joint Capsule Closure
Iliotibial Band Tenodesis
Patellar Ligament/Tendon Repair

{3}------------------------------------------------

Wrist
[ Scapholunate Ligament Reconstruction
Elbow
Biceps Tendon Reattachment
□ Ulnar or Radial Collateral Ligament Reconstruction
Hip
□ Capsular Repair
■ Acetabular Labra1 Repair
Knee
□ Extracapsular Repair
□ Medial Collateral Ligament Repair
■ Lateral Collateral Ligament Repair
□ Posterior Oblique Ligament Repair
□ Joint Capsule Closure
□ Iliotibial Band Tenodesis Reconstruction
[ Patellar Ligament/Tendon Repair
[ Vastus Medials Obliquus Muscle Advancement
Indications -Interference ScrewsIndications -Titan Mini-Interference Screws
The Titan Interference Screws are indicated for thereattachment of liqament, tendon, soft tissue, or bone toThe Titan Mini-Interference Screws are intended to beused for fixation of tissue, including ligament or tendon to
bone during cruciate ligament reconstruction surgeries ofbone, or a bone/tendon to bone. See below for specific
the knee. All screws with a diameter of 9 mm or less and aindications.
length of 23 mm or less are also intended for the use in theThe Mini-Interference Screws are intended to provide soft
following procedures:tissue reattachment (i.e. fixation of ligament and tendon
Kneegraft tissue).See below for specific indications.
ACL repairsShoulder
□ PCL repairs□ Extra-capsular repairs□ Capsular stabilization
o Medial collateral ligamento Bankart repair
o Lateral collateral ligamento Anterior shoulder instability
o Posterior oblique ligamento SLAP lesion repairs
□ Patellar realignment and tendon repairso Capsular shift of capsulolabral reconstructions
o Vastus medialis oliquous advancement_ Acromioclavicular separation repairs
■ Iliotibial band tenodes is□ Deltoid repairs
■ Rotator cuff tear repairs
Shoulder□ Capsular stabilization■ Biceps tenodesis
o Bankart repairFoot and Ankle
o Anterior shoulder instability■ Hallux valgus reconstruction
o SLAP lesion repairs□ Medial stabilization
o Capsular shift of capsulolabral reconstructions□ Lateral stabilization
□ Acromioclavicularseparation repairs_ Achilles Tendon Repair
□ De ltoid re pairs■ Midfoot reconstructions
■ Rotator cuff tear repairs_ Metatarsal ligament repair
□ Biceps tenodesis■ Bunionectomy
■ Flexor Hullucis Longus for Achilles Tendon
Foot and Anklereconstruction
□ Hallux valgus repairs■ Medial or lateral instability repairs/reconstructions□ Tendon transfers in the foot and ankle
□ Achilles tendon repairs/reconstructions
Midfoot reconstructionsKnee
□ Metatars al ligament/tendon repairs/recons tructions□ Medial Collateral Ligament Repair
■ BunionectomyLateral Collateral Ligament Repair
■ Flexor Hullucis Longus■ Patellar Tendon Repair
□ Tendon transfers□ Posterior Oblique Ligament Repair
Elbow, Wrist, and HandIlliotibial Band Tenodesis
□ Biceps tendon reattachmentPosterior Cruciate Ligament Repair
□ Ulnar or radial collateral ligament reconstructionsElbow
■ Lateralepicondylitis repair_ Biceps tendon reattachmentUlnar or radial collateral ligament reconstruction

{4}------------------------------------------------

□ Scapholunate ligament reconstruction□ Tendon transfersWrist and Hand□ Scapholunate Ligament Reconstruction□ Ulnar Collateral Ligament Reconstruction□ Radial Collateral Ligament Reconstruction□ Carpometalcarpal joint arthroplasty (basal thumb joint arthroplasty)□ Carpal Ligament Reconstructions and repairs□ Tendon transfer in the hand/wrist□ Lateral Epicondylitis repair
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Prescription Use ________________________________________________(21 CFR 801 Subpart (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

{5}------------------------------------------------

510(k) Summary

(as required by 21 CFR 807.92)

SubmitterValeris Medical
Address200 Cobb Pkwy NBuilding 200, Suite 210Marietta, GA 30062
Telephone888-404-3980 Ext 101
Fax678-669-2188
Contact PersonDaniel LanoisGeneral Manager
AddressValeris Medical200 Cobb Pkwy NBuilding 200, Suite 210Marietta, GA 30062
Telephone888-404-3980 Ext 101
Fax678-669-2188
emaildaniel@valerismedical.com

Date Prepared August 8, 2014

Trade NameApollo Suture Anchor System and Titan Screws
Common NameScrew, Fixation, Bone
Panel CodeOrthopaedics/87
Classification NameSmooth or threaded metallic bone fixation fastener
ClassClass II
Regulation Number21 CFR 888.3040
Product CodeMBI
Name of Predicate Device510(k) #Manufacturer
Apollo Suture Anchor System and Titan ScrewsK133036Amendia (transferred product to Valeris Medical)
DescriptionApollo FamilyThe Apollo Medial Suture Anchor, XT Suture Anchor, Lateral Anchor, andLabral Suture Anchor are delivery systems for anchors for use in fixationof ligament, tendon, bone, or soft tissue to bone in knee, shoulder,foot/ankle, elbow, and hand/wrist procedures. These anchors consist ofcannulated anchors with integrated suture attachment or separate suturepunch eyelet. The Anchors are provided loaded on individual inserterswith and without integrated sutures, sterile, for single use only.
Titan FamilyThe Titan Interference and Mini-Interference Screws are interferencescrews for use in fixation of ligament, tendon, bone, or soft tissue to bonein knee, shoulder, foot/ankle, elbow, and hand/wrist procedures. Thescrews are provided sterile, for single use only.
Screw and anchor implants are made from either a titanium alloy (6AI4V)per ASTM F136, or PEEK (Zeniva ZA-500) per ASTM F2026 fromSolvay Advanced Polymers.

{6}------------------------------------------------

Indications andIntended UseThe Apollo Suture Anchors and Titan Screws are indicated for usein fixation of ligament, tendon, bone, or soft tissue to bone in knee,shoulder, foot/ankle, elbow, and hand/wrist procedures. Thescrews are intended for use in the following procedures:
--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Indications - Apollo Medial Suture Anchor and Apollo XT Suture Anchors
The Apollo Medial Suture Anchor and Apollo XT Suture Anchors are intended for:
Shoulder
☐ Rotator Cuff Repair
☐ Bankart Repair
☐ SLAP Lesion Repair
☐ Biceps Tenodesis
☐ Acromio-Clavicular Separation Repair
☐ Deltoid Repair
☐ Capsular Shift or Capsulolabral Reconstruction
Foot/Ankle
☐ Lateral Stabilization
☐ Medial Stabilization
☐ Achilles Tendon Repair
Knee
☐ Medial Collateral Ligament Repair
☐ Lateral Collateral Ligament Repair
☐ Posterior Oblique Ligament Repair
☐ Illiotibial Band Tenodesis
Elbow
☐ Biceps Tendon Reattachment
☐ Ulnar or Radial Collateral Ligament Reconstruction
Hip
☐ Capsular Repair
☐ Acetabular Labral Repair
Indications - Apollo Lateral Anchor
The Apollo Lateral Anchor is indicated for:
Shoulder
☐ Rotator Cuff Repair
☐ Bankart Repair
☐ SLAP Lesion Repair
☐ Biceps Tenodesis
☐ Acromio-Clavicular Separation Repair
☐ Deltoid Repair
☐ Capsular Shift or Capsulolabral Reconstruction
Wrist/Hand
☐ Scapholunate Ligament Reconstruction
☐ Ulnar/Radial Collateral Ligament Reconstruction
Foot/Ankle
☐ Lateral Stabilization
☐ Medial Stabilization
☐ Achilles Tendon Repair/Reconstruction
☐ Hallux Valgus Reconstruction
☐ Mid and Forefoot Reconstruction
Elbow
☐ Biceps Tendon Reconstruction
☐ Ulnar or Radial Collateral Ligament Reconstruction
☐ Lateral Epicondylitis Repair (PEEK Anchor Only)
Knee
☐ Medial Collateral Ligament Repair
☐ Lateral Collateral Ligament Repair
☐ Posterior Oblique Ligament Repair
☐ Joint Capsule Closure
☐ Iliotibial Band Tenodesis
☐ Patellar Ligament/Tendon Repair
Indications - Apollo Labral Suture Anchor
--------------------------------------------------
Shoulder
☐ Rotator Cuff Repair
☐ Bankart Repair
☐ SLAP Lesion Repair
☐ Biceps Tenodesis
☐ Acromio-Clavicular Separation Repair
☐ Deltoid Repair
☐ Capsular Shift or Capsulolabral Reconstruction
Wrist
☐ Scapholunate Ligament Reconstruction
Elbow
☐ Biceps Tendon Reattachment
☐ Ulnar or Radial Collateral Ligament Reconstruction
Hip
☐ Capsular Repair
☐ Acetabular Labral Repair

{7}------------------------------------------------

Knee
☐ Extracapsular Repair
☐ Medial Collateral Ligament Repair
☐ Lateral Collateral Ligament Repair
☐ Posterior Oblique Ligament Repair
☐ Joint Capsule Closure
☐ Iliotibial Band Tenodesis Reconstruction
☐ Patellar Ligament/Tendon Repair
☐ Vastus Medialis Obliquus Muscle Advancement
Indications –Interference ScrewsIndications –Titan Mini-Interference Screws
The Titan Interference Screws are indicated for thereattachment of ligament, tendon, soft tissue, or bone toThe Titan Mini-Interference Screws are intended to beused for fixation of tissue, including ligament or tendon to
bone during cruciate ligament reconstruction surgeries ofbone, or a bone/tendon to bone. See below for specific
the knee. All screws with a diameter of 9 mm or less and aindications.
length of 23 mm or less are also intended for the use in theThe Mini-Interference Screws are intended to provide soft
following procedures:tissue reattachment (i.e. fixation of ligament and tendon
Kneegraft tissue).See below for specific indications.
☐ ACL repairsShoulder
☐ PCL repairs☐ Capsular stabilization
☐ Extra-capsular repairso Bankart repair
o Medial collateral ligamento Anterior shoulder instability
o Lateral collateral ligamento SLAP lesion repairs
o Posterior oblique ligamento Capsular shift of capsulolabral reconstructions
☐ Patellar realignment and tendon repairs☐ Acromioclavicular separation repairs
o Vastus medialis oliquous advancement☐ Deltoid repairs
☐ Iliotibial band tenodesis☐ Rotator cuff tear repairs
Shoulder☐ Biceps tenodesis
☐ Capsular stabilizationFoot and Ankle
o Bankart repair☐ Hallux valgus reconstruction
o Anterior shoulder instability☐ Medial stabilization
o SLAP lesion repairs☐ Lateral stabilization
o Capsular shift of capsulolabral reconstructions☐ Achilles Tendon Repair
☐ Acromioclavicular separation repairs☐ Midfoot reconstructions
☐ Deltoid repairs☐ Metatarsal ligament repair
☐ Rotator cuff tear repairs☐ Bunionectomy
☐ Biceps tenodesis☐ Flexor Hullucis Longus for Achilles Tendonreconstruction
Foot and Ankle☐ Tendon transfers in the foot and ankle
☐ Hallux valgus repairs
☐ Medial or lateral instability repairs/reconstructionsKnee
☐ Achilles tendon repairs/reconstructions☐ Midfoot reconstructions☐ Medial Collateral Ligament Repair
☐ Metatarsal ligament/tendon repairs/reconstructions☐ Lateral Collateral Ligament Repair
☐ Bunionectomy☐ Patellar Tendon Repair
☐ Flexor Hullucis Longus☐ Posterior Oblique Ligament Repair
☐ Tendon transfers☐ Illiotibial Band Tenodesis
Elbow, Wrist, and Hand☐ Posterior Cruciate Ligament Repair
☐ Biceps tendon reattachmentElbow
☐ Ulnar or radial collateral ligament reconstructions☐ Biceps tendon reattachment
☐ Lateral epicondylitis repair☐ Ulnar or radial collateral ligament reconstruction
☐ Scapholunate ligament reconstruction
☐ Tendon transfersWrist and Hand
☐ Scapholunate Ligament Reconstruction
☐ Ulnar Collateral Ligament Reconstruction
☐ Radial Collateral Ligament Reconstruction
☐ Carpometalcarpal joint arthroplasty (basal thumb jointarthroplasty)
☐ Carpal Ligament Reconstructions and repairs
☐ Tendon transfer in the hand/wrist
☐ Lateral Epicondylitis repair

{8}------------------------------------------------

TechnologicalCharacteristicsand SubstantialEquivalenceDocumentation was provided to demonstrate that the Subject device, Apollo Suture Anchor System and Titan Screws is substantially equivalent to the Predicate Apollo Suture Anchor System and Titan Screws (K133036), The Subject device is substantially equivalent to the predicate device in intended use, indications for use, materials, technological characteristics, and labeling.
---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Performance DataAxial Pull-Out and Insertion Torque per ASTM F543-7 testing wereconducted to confirm that the modification to add a 3rd suture portal didnot introduce any new risk.
ConclusionBased on the indications for use, technological characteristics, materials,and comparison to predicate devices, the Subject Apollo Suture AnchorSystem and Titan Screws has been shown to be substantially equivalentto legally marketed predicate devices, and is safe and effective for itsintended use.

§ 888.3040 Smooth or threaded metallic bone fixation fastener.

(a)
Identification. A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.(b)
Classification. Class II.